Article

Electrochemical Ultrasensitive Sensing of Uric Acid on Non-Enzymatic Porous Cobalt Oxide Nanosheets-Based Sensor

1
Sensors Lab, Centre for Nanoscience and Nanotechnology, Jamia Millia Islamia, New Delhi 110025, India
2
Quantum and Nanophotonics Research Laboratory, Centre for Nanoscience and Nanotechnology, Jamia Millia Islamia, New Delhi 110025, India
3
Department of Life Science, Shakuntala Memorial Educational Institute, Bahraich 271870, India
4
Department of Pharmacology & Therapeutics, Rosewell Park Cancer Institute, Elm Street & Carlton Street, Buffalo, NY 14263, USA
5
Department of Bioresources, University of Kashmir, Hazratbal, Srinagar 190006, India
6
Singapore-MIT Alliance for Research and Technology (SMART), Critical Analytics for Manufacturing Personalized-Medicine (CAMP), Create Way 138602, Singapore
7
Department of Applied Chemistry, School of Advanced Materials and Nanotechnology, Xidian University, Xi’an 710126, China
*
Author to whom correspondence should be addressed.
Biosensors 2022, 12(12), 1140; https://doi.org/10.3390/bios12121140
Submission received: 29 October 2022 / Revised: 21 November 2022 / Accepted: 29 November 2022 / Published: 7 December 2022
(This article belongs to the Special Issue Construction of Biosensors Using Nano- and Microtechnology)

Abstract

:
Transition metal oxide (TMO)-based nanomaterials are effectively utilized to fabricate clinically useful ultra-sensitive sensors. Different nanostructured nanomaterials of TMO have attracted a lot of interest from researchers for diverse applications. Herein, we utilized a hydrothermal method to develop porous nanosheets of cobalt oxide. This synthesis method is simple and low temperature-based. The morphology of the porous nanosheets like cobalt oxide was investigated in detail using FESEM and TEM. The morphological investigation confirmed the successful formation of the porous nanosheet-like nanostructure. The crystal characteristic of porous cobalt oxide nanosheets was evaluated by XRD analysis, which confirmed the crystallinity of as-synthesized cobalt oxide nanosheets. The uric acid sensor fabrication involves the fixing of porous cobalt oxide nanosheets onto the GCE (glassy carbon electrode). The non-enzymatic electrochemical sensing was measured using CV and DPV analysis. The application of DPV technique during electrochemical testing for uric acid resulted in ultra-high sensitivity (3566.5 µAmM−1cm−2), which is ~7.58 times better than CV-based sensitivity (470.4 µAmM−1cm−2). Additionally, uric acid sensors were tested for their selectivity and storage ability. The applicability of the uric acid sensors was tested in the serum sample through standard addition and recovery of known uric acid concentration. This ultrasensitive nature of porous cobalt oxide nanosheets could be utilized to realize the sensing of other biomolecules.

1. Introduction

Uric acid (UA) plays an important role in various biological processes and physiological functions in humans and higher species [1]. Bodily fluids (i.e., serum, urine, and saliva) contain UA. Mainly, the liver and intestines are the primary sites of UA production. However, the majority of UA is removed by urate transporters in the kidneys and intestines. Unusually low, excessive, or variable UA concentration is an indicator of various diseases (i.e., pneumonia, gout, leukemia, type 2 diabetes, chronic renal diseases, toxemia during pregnancy, multiple sclerosis, metabolic disorders, and hypertension) [2,3,4,5]. Therefore, monitoring UA concentration becomes crucial for early disease diagnosis since UA acts as a diagnostic marker for UA-concentration associate diseases.
Several standard analytical methods (i.e., flow-injection analysis, chromatography, HPLC, chemiluminescence, mass spectrometry, colorimetry, capillary electrophoresis-amperometry, and enzyme test kits) are utilized to determine UA concentration [6,7,8,9,10]. However, complex sample preparation, high cost, slow preparation, and skilled technician requirements are the major drawbacks of the above-mentioned analytical methods [11]. On the other hand, electrochemical-based sensors have the potential for rapid detection and offer high-sensing performance during biomolecules detection [8,9,10,12,13,14]. Recently, considerable attention has been given to electrochemical-based UA sensors due to their high selectivity, accuracy, less interference, and low-cost [15,16,17,18,19,20].
There are two types of electrodes (i.e., metal-based and carbon-based electrodes) that are mostly used in electrochemical sensor fabrication. The carbon-based electrodes (i.e., glassy carbon, screen-printed, and carbon-paste electrodes) are the preferred working electrodes due to the minimum over potential requirement compared to the metal-based electrodes [19]. Due to their low costs and ease of preparation, the carbon-based electrodes have been extensively used in electrochemical sensing applications. These conventional bare electrodes are not preferred due to their weak electrochemical oxidation kinetics [20]. Hence, nanomaterial/electrocatalyst (as redox dynamic site) modification is needed to improve the electrochemical oxidation kinetics of the bare electrodes [21]. The choice of nanomaterial/electrocatalyst modifiers (for example, metals/metal oxides or carbon-based nanomaterials) depends on the type of the electrode and analytical requirements.
The improved electro-catalytic activity and charge transfer capabilities of metal/metal oxide have attracted considerable interest for their applications in electrode modifications [15,16,17,18,19,20,21,22,23,24,25,26,27]. Among different metal oxides, cobalt oxide nanomaterials have been utilized to fabricate enzymatic/non-enzymatic sensors where the use of cobalt oxide nanostructures enhances the desired electrochemical properties [23,24,25,26,27,28,29]. Different methods (i.e., thermal decomposition, sol-gel, surfactant-mediated synthesis, spray-pyrolysis, and polymer-matrix assisted) are utilized for the synthesis of various kinds of cobalt oxide nanostructures. Recently, cobalt oxide has attained interest for different applications (i.e., heterogeneous catalysis, lithium-ion batteries, gas sensing, electrochemical sensors, and solar cells). Cobalt oxide nanostructures are the preferred material to modify the electrode, which offers abundant active sites for reaction and easy adsorption/electroactive species diffusion [24]. Hence, cobalt oxide nanostructures modified sensor electrodes are suitable for the detection of different analytes. For example, Nagal et al. described the use of nano berry-like cobalt oxide nanostructures for the electrochemical-based enzyme-less uric acid sensor [24]. Kogularasu et al. investigated the impact of cobalt oxide polyhedrons to develop an enzyme-free biosensor detect H2O2 [25]. Zhang and Liu utilized cobalt oxide nanosheets for enzyme-free detection of glucose [26]. Kang et al. utilized cobalt oxide nanowires for the fabrication of an enzyme-less glucose sensor [27]. Mondal et al. fabricated glucose sensors using different nanostructures (like spherical nanoparticles, porous nanorods, and nanoflowers) of cobalt oxide [28]. Chang et al. evaluated the electrochemical sensing performance of lactic acid using cobalt oxide nanostructures [29]. Therefore, there is a demand for designing cobalt oxide nanostructures having excellent surface area, which can be utilized to fabricate high-performance sensors.
In this work, we demonstrated the synthesis of porous cobalt oxide nanosheets using a simple and low-temperature-based hydrothermal method. The porous cobalt oxide nanosheets were comprehensively characterized using XRD, FE-SEM, and TEM. The electrochemical properties of porous cobalt oxide nanosheet-based non-enzymatic UA sensors were studied by cyclic voltammetry (CV), EIS (electrochemical-impedance-spectroscopy), and DPV (differential-pulse voltammetry). The porous cobalt oxide nanosheets-based non-enzymatic UA sensor exhibited ultra-high sensitivity (3566.5 µAmM−1cm−2) when analyzed with the DPV technique. Additionally, selectivity, stability, and applicability in serum samples were evaluated. This porous cobalt oxide nanosheet-based non-enzymatic UA sensor offers better sensitivity when compared to CV-measured sensitivity (470.4 µAmM−1cm−2).

2. Materials and Methods

2.1. Chemicals

All analytical chemicals were obtained from Sigma Aldrich and used. Cobalt nitrate hexahydrate (≥99.99%; Co(NO3)2.6H2O), sodium hydroxide pellets (≥97%; NaOH), potassium chloride (KCl), uric acid (≥99%), ethylene glycol (99.8%), sodium chloride (NaCl), potassium hexacyanoferrate [K3Fe(CN)6]3−/4− (≥99%), glucose, fructose, lactic acid, L-cysteine, urea, phosphate buffer saline (PBS; pH = 7.4) solution, 2-(2-Butoxyethoxy)ethyl acetate (≥99.2%), dopamine, and ascorbic acid were purchased. In our experiments, ultrapure water was utilized.

2.2. Porous Cobalt Oxide Nanosheets Synthesis

A simple low-cost hydrothermal process is used to synthesize porous cobalt oxide nanosheets (Scheme 1). For synthesis, first, a precursor solution of Co(NO3)2.6H2O (0.58 g) was prepared in 10 mL DI water, and then NaOH (0.2 g) solution was dropwise added while vigorous stirring for 30 min to obtain a homogenous mixture. The prepared solution was transferred into an autoclave vessel (Teflon-lined stainless steel) and put into a hot air oven at 150 °C for 6h. On completion of the reaction, the autoclave was cooled down and the powder sample was washed multiple times with DI and ethanol. Finally, the black precipitate was dried (at 60 °C) and annealed (at 500 °C) for 3h before characterizing it in detail.

2.3. Sensor Fabrication

The porous cobalt oxide nanosheets-based non-enzymatic UA sensor was fabricated using a conductive binder along with cobalt oxide nanosheets (Scheme 1). In brief, a slurry of 0.01 g porous cobalt oxide nanosheets and 50 µL conductive binder (2-(2-Butoxyethoxy)ethyl acetate) was prepared using mortar and pestle. The slurry was sonicated for 10 min before fixing onto the working electrode surface. Different amounts of slurry were fixed on the working electrode to optimize the most suitable amount of slurry, which gives the best sensing performance (for example, 4 µL was the optimized amount). The porous cobalt oxide nanosheet-based sensor was dried at 60 °C for 6h and kept at room temperature.

2.4. Materials Characterization and Electrochemical Analysis of Sensor

The FESEM (field-emission-scanning electron microscope; Zeiss, Sigma) was utilized to analyze porous sheet-like cobalt oxide nanostructures morphology. The structural analysis was examined using XRD (Rigaku), where Cu-K α X-ray radiation ( λ = 1.5418 Å), current (30 mA), and voltage (40 kV) were used. A more detailed study of porous sheet-like cobalt oxide nanostructures was characterized with TEM (TECNAI G20; accelerating voltage = 200 kV). The ASAP 2010 analyzer and Barrett-Joyner-Halenda (BJH) method were utilized for nitrogen adsorption-desorption analysis (at 77 K) and pore size distribution determination, respectively.
For the electrochemical measurements of the fabricated porous cobalt oxide nanosheets-based non-enzymatic UA sensor, a compact and portable potentiostat/impedance analyzer “PalmSens4” was used. The CV and EIS were used to evaluate the best-performing sensor using the optimum amount of porous cobalt oxide nanosheets in a redox probe solution ([Fe(CN)6]3−/4−). During EIS measurement, frequency range and applied potential were set to 0.01 Hz to 100 kHz and 0.25 V, respectively. The sensing performance of the fabricated porous cobalt oxide nanosheet-based non-enzymatic UA sensor was evaluated using CV and DPV techniques.

3. Results

3.1. Characterizations of Porous Cobalt Oxide Nanostructures

The surface morphology and crystallinity of as-synthesized cobalt oxide nanostructure were characterized using FESEM, TEM, and XRD. Figure 1a–c displays the surface morphology of the as-synthesized cobalt oxide nanostructure. The low-magnification images of cobalt oxide nanostructures reveal that the synthesized nanomaterial is obtained in bulk amounts with irregular shapes and sizes (Figure 1a,b). The high-magnification image shows that cobalt oxide nanostructures bear porous sheet-like morphology (Figure 1c). Only the surface morphology is smooth, with uniform pores present on the surface. Figure 2d shows the XRD pattern of as-synthesized cobalt oxide nanomaterial. The XRD pattern confirms the crystalline nature of nanomaterial, and the obtained pattern is well indexed (JCPDS card no. 42-1467) [30]. The diffraction peaks of cobalt oxide are observed at 2Ɵ values of 22.1°, 31.1°, 36.6°, 38.1°, 44.6°, 55.4°, 59.6°, and 66.7°, corresponding to the miller indices of [111], [220], [311], [222], [400], [422], [511], and [440], respectively. The highest intensity diffraction peak is seen along [311] lattice plane. To confirm the porosity, TEM analysis of the porous cobalt oxide sheet-like nanostructure was done (Figure 2). The TEM images show the porous sheet-like nanostructure of cobalt oxide. Additionally, the SAED pattern (Figure 2c) of the porous cobalt oxide sheet-like nanostructure suggests the crystal nature. These observations are supported by FESEM images. The surface area and pore size of the porous cobalt oxide sheet-like nanostructure was analyzed using BET (Brunauer-Emmett-Teller) analysis, shown in Figure 2d. The isotherm pattern indicates the porous nature of nanomaterial. The obtained BET surface area of porous cobalt oxide sheet-like nanostructure is around ~166 m2/g. A narrow pore size distribution from 3 to 7 nm was obtained with an average pore size of ~5 nm (Inset of Figure 2d). The high surface area and small pore size of the porous cobalt oxide sheet-like nanostructure will offer a better catalytic for electrochemical reactions.

3.2. Electrochemical Studies using CV and EIS Techniques

The electrochemical properties of bare GCE and cobalt oxide/GCE electrodes were analyzed using EIS and CV techniques. The electron transfer reaction of bare GCE and cobalt oxide/GCE electrodes for the redox probe solution of [Fe(CN)6]3−/4− with KCl (0.1 M) determine the kinetic parameters (i.e., electron transfer rate and charge transfer resistance) (Figure 3). The EIS was carried out at 0.25 V in the frequency range of 0.01 Hz to 100 kHz. Nyquist plots in Figure 3a show two frequency regions, one at a higher and another at a lower frequency region. Cobalt oxide/GCE electrode showed a small semicircle (at a higher frequency region) and a straight line (at a lower frequency region), which suggests a perfect diffusion-controlled process during the electron transfer reaction. The bare GCE electrode showed a higher charge transfer resistance (Rct) compared to the modified GCE electrode. The bode plots of bare GCE and cobalt oxide/GCE electrodes are illustrated in Figure 3b,c, respectively. It can be seen from these bode plot curves that the modifying GCE with cobalt oxide decreased the interfacial impedance. Additionally, the bode angle was decreased after GCE surface modification with cobalt oxide. The shift of lower frequency peak for cobalt oxide modified GCE suggests the more prominent electron transfer process compared to the bare GCE electrode [31].
The CV data obtained for the bare GCE and cobalt oxide/GCE electrodes agree with the EIS data. The CV response curves were recorded from −0.2V to + 0.8V (vs. Ag/AgCl) at a fixed scan rate (50 mV/s) (Figure 4a). Clear redox peaks (i.e., oxidation and reduction peaks) can be seen in the obtained CV response curves, where the cobalt oxide/GCE electrode showed improved oxidation peak current value compared to bare GCE. Additionally, the effect of scan rate on the cobalt oxide/GCE electrode’s electron transfer characteristics was investigated by measuring CV response curves at different scan rates (i.e., 10–250 mV/s) (Figure 4b). It can be seen from the CV curves that the oxidation and reduction peak current values increase with the scan rate increase. The CV curve shape and the values of peak potential separation indicate a diffusion-controlled process in the redox probe solution. To verify the diffusion-controlled process over the cobalt oxide/GCE electrode, a plot of current peak vs. square root of scan rate is plotted in Figure 4c. A perfect linear relationship was observed between the square root of the scan rate and the value of the peak current. This further confirms the diffusion-controlled process over the surface of the modified electrode [32,33,34].

3.3. Sensing Performance Characterization using CV Technique

The response of the cobalt oxide/GCE sensor was characterized towards uric acid before a detailed analysis of sensing performance. The CV analysis of the cobalt oxide/GCE sensor was performed in PBS without and with uric acid (10 µM) at 50 mV/s (Figure 5a). When CV analysis was done in PBS, there was no noticeable peak in the CV curve. However, in 10 µM uric acid, a noticeable peak of uric acid oxidation was present at 0.6 voltage. The possible and most accepted detection mechanism for uric acid oxidation involves the transfer of two-electron/two-proton, which enhances the response during sensing measurement (Scheme 2) [35,36,37]. Additionally, the porous nature of nanosheets provided abundant catalytic sites due to the large surface area. Moreover, cobalt oxide is a p-type semiconductor, which could provide excess hole concentration and help to capture the electrons during uric acid oxidation. Also, we observed the irreversible oxidation peak in the CV curve that indicated swift electron transfer between the GCE and porous cobalt oxide during the electrochemical detection of uric acid.
To evaluate the sensing performance (i.e., sensitivity, detection range, and detection limit), the CV response curves of the cobalt oxide/GCE sensor were measured with increasing concentrations of uric acid (0–2500 µM) as shown in Figure 5b. In this figure, the CV curves showed an increase in current with increasing uric acid concentration. A graph of peak current (µA) vs. uric acid concentration (µM) was drawn (Figure 5c). From this graph, two regions (linear and non-linear) can be seen. In general, the non-linear region signifies the saturation of the current response of the cobalt oxide/GCE sensor on those uric acid concentrations (i.e., high uric acid concentration). Further, the linear region of the sensor response is taken and a calibration plot (peak current (µA) vs. uric acid concentration (µM) is plotted, shown in the inset of Figure 4c. The sensor responded linearly up to 1000 µM of uric concentration (regression coefficient (R2) = 0.9978). From the slope, we calculated the sensitivity of 470.4 µAmM−1cm−2 [38]. Additionally, based on the S/N ratio = 3, the detection limit was calculated to be 10 µM. The obtained sensitivity, linear range, and detection limit were comparatively better than most of the previously reported literature (Table 1). The good sensing performance is due to high surface-to-volume ratio and the presence of large active sites on as-synthesised porous nanosheets like nanomaterial.

3.4. Sensing Performance Characterization using DPV Technique

The DPV technique is more sensitive than CV due to the minimization of capacitive current. For this reason, we utilized the DPV technique to evaluate the sensing performance of the cobalt oxide/GCE sensor. Initially, the electrochemical behavior of the cobalt oxide/GCE sensor was measured. Figure 6a shows DPV curves obtained in PBS buffer (pH 7.4) without and with 10 µM uric acid. The appearance of the uric acid oxidation peak at 0.45 potential (vs. Ag/AgCl), compared to the DPV curve recorded in PBS buffer solution, indicated the sensitivity nature of the cobalt oxide/GCE sensor towards uric acid. Additionally, when measuring CV response (as shown in Figure 5a), it was seen that uric acid oxidation is an irreversible process.
Then, DPV was performed with increasing uric acid concentration (up to 2500 µM) in PBS buffer. The obtained DPV curves are shown in Figure 6b, where an increase in current can be seen with increased uric acid concentration. A plot of peak current (µA) vs. uric acid concentration (µM) is shown in Figure 6c along with the calibrated plot of the linear range of the sensor in the inset. The cobalt oxide/GCE sensor showed a linear range of up to 800 µM uric acid concentration with R2 of 0.9929. However, the current level was decreased at a higher uric acid concentration due to the saturation of electrocatalysis of uric acid on the electrode surface. The sensitivity of the cobalt oxide/GCE sensor was calculated by using the standard equation of slope of the calibrated plot/working electrode surface area. The sensor showed the highest sensitivity of 3566.5 µAmM−1cm−2. The limit of detection was 12 µM. The achieved sensing performance results are shown in Table 1. As shown in Table 1, the cobalt oxide/GCE sensor showed ultra-high sensitivity compared to previously reported literature [11,24,30,39,40,41,42,43,44,45,46]. Furthermore, the DPV technique showed ~7.58 times high sensitivity (3566.5 µAmM−1cm−2) compared to CV’s measured sensitivity (470.4 µAmM−1cm−2). These results confirmed the fact that DPV is a more sensitive technique as compared to CV. However, the main reason for getting high sensing performance is attributed to large active sites and the surface-to-volume ratio of as-synthesised porous nanosheets like nanomaterial.

3.5. Interference and Stability Tests of Cobalt Oxide/GCE Sensor

We investigated the selectivity of the cobalt oxide/GCE sensor for uric acid detection in the presence of possible interferences. Eight possible interfering species, including lactic acid, L-cysteine, glucose, urea, fructose, sodium chloride, and potassium chloride, were taken for the selectivity study. Figure 7a illustrates CV curves for 25 µM uric acid only and 25 µM uric acid and 100 µM of each interfering species (i.e., lactic acid, L-cysteine, glucose, urea, fructose, sodium chloride, and potassium chloride). There is a slight increase (positive interference) in CV response with a high concentration of interfering species. Based on this result, the cobalt oxide/GCE sensor is selective for uric acid determination. Additionally, we evaluated the sensor stability after storing sensor at room temperature and measuring the response after 30 and 45 days for 25 µM uric acid (Figure 7b). As shown in Figure 7b, the sensor showed good stability and maintained 97.4% of its current peak after 45 days of storage. Additionally, the low RSD of 2.6% indicated good stability. Finally, we tested selectivity tests in the presence of ascorbic acid, dopamine, and urea (Figure 7c). In the presence of these species, a slight increase in current response was noticed. However, no other peaks were noticed.

3.6. Analysis of Real Serum Sample

To determine whether the cobalt oxide/GCE sensor was suitable for uric acid detection in human serum samples (obtained from Sigma−Aldrich; H4522). We used a standard addition- (known uric acid concentration) based method to estimate the recovery results of the added uric acid concentration in a serum sample. The recovery (%) was calculated using formula [Recovery (%) = Calculated uric acid concentration × 100/Added uric acid concentration]. The obtained data are shown in Table 2. Recovery results showed that the cobalt oxide/GCE sensor was suitable for uric acid determination in the real sample.

4. Conclusions

In this study, a low temperature-based hydrothermal method was utilized to synthesize porous nanosheets-like cobalt oxide nanostructures. The crystallinity and morphology of as-synthesized cobalt oxide nanostructures were tested using direct techniques (i.e., FESET, TEM, and XRD). The obtained results showed the successful formation of the porous nanosheet-like nanostructure that bears good crystallinity. The possibility of using such porous nanosheets-like cobalt oxide nanostructures in the sensor was tested by using electrochemical methods, such as CV and EIS. Based on the obtained data, sensing performance evaluation using CV and DPV techniques indicated high sensitivity. The DPV is a more sensitive technique as compared to CV. In this context, DVP data showed ultra-high sensitivity (3566.5 µAmM−1cm−2), which was ~7.58 times better than CV-based sensitivity (470.4 µAmM−1cm−2). Additionally, the cobalt oxide/GCE sensor exhibited good selectivity during uric acid measuring in the interfering species. The stability and applicability of the cobalt oxide/GCE sensor were tested, which showed good stability and applicability in a serum sample. Nevertheless, this work contributes to obtaining ultra-high sensitivity using porous cobalt oxide nanosheet-like nanostructures and provides the further possibility to improve sensing performance with surface modification of nanosheets using other metal/metal oxides.

Author Contributions

Conceptualization, S.M., V.N. and R.A.; methodology, V.N. and I.M.; software, S.M.; validation, S.M., V.N. and M.K.; formal analysis, V.N. and M.K.; resources, R.A.; writing—original draft preparation, S.M. and V.N.; writing—review and editing, R.A., S.A., K.S.B. and A.K.; supervision, R.A.; funding acquisition, R.A. All authors have read and agreed to the published version of the manuscript.

Funding

R.A. gratefully acknowledges the Ramalingaswami Fellowship Award (BT/RLF/Re-entry/28/2018) and research funding from the Department of Biotechnology (DBT), Government of India.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. El Ridi, R.; Tallima, H. Physiological Functions and Pathogenic Potential of Uric Acid: A Review. J. Adv. Res. 2017, 8, 487–493. [Google Scholar] [CrossRef] [PubMed]
  2. Guan, Y.; Chu, Q.; Ye, J. Determination of Uric Acid in Human Saliva by Capillary Electrophoresis with Electrochemical Detection: Potential Application in Fast Diagnosis of Gout. Anal. Bioanal. Chem. 2004, 380, 913–917. [Google Scholar] [CrossRef] [PubMed]
  3. Leiba, A.; Vinker, S.; Dinour, D.; Holtzman, E.J.; Shani, M. Uric Acid Levels Within the Normal Range Predict Increased Risk of Hypertension: A Cohort Study. J. Am. Soc. Hypertens. 2015, 9, 600–609. [Google Scholar] [CrossRef] [PubMed]
  4. King, C.; Lanaspa, M.A.; Jensen, T.; Tolan, D.R.; Sanchez-Lozada, L.G.; Johnson, R.J. Uric Acid in Chronic Kidney Disease. Contrib. Nephrol. 2018, 192, 88–102. [Google Scholar]
  5. Virdis, A.; Masi, S.; Casiglia, E.; Tikhonoff, V.; Cicero, A.F.; Ungar, A.; Rivasi, G.; Salvetti, M.; Barbagallo, C.M.; Bombelli, M.; et al. Identification of the Uric Acid Thresholds Predicting an Increased Total and Cardiovascular Mortality Over 20 Years. Hypertension 2020, 75, 302–308. [Google Scholar] [CrossRef]
  6. Matos, R.C.; Augelli, M.A.; Lago, C.L.; Angnes, L. Flow Injection Analysis-Amperometric Determination of Ascorbic and Uric Acids in Urine Using Arrays of Gold Microelectrodes Modified by Electrodeposition of Palladium. Anal. Chim. Acta 2000, 404, 151–157. [Google Scholar] [CrossRef]
  7. Dai, X.; Fang, X.; Zhang, C.; Xu, R.; Xu, B. Determination of Serum Uric Acid using High-Performance Liquid Chromatography (HPLC)/Isotope Dilution Mass Spectrometry (ID-MS) as a Candidate Reference Method. J. Chromatogr B. Analyt. Technol. Biomed. Life Sci. 2007, 857, 287–295. [Google Scholar] [CrossRef]
  8. Ahmad, R.; Tripathy, N.; Jang, N.K.; Khang, G.; Hahn, Y.-B. Fabrication of Highly Sensitive Uric Acid Biosensor based on Directly Grown ZnO Nanosheets on Electrode Surface. Sens. Actuators B Chem. 2015, 206, 146–151. [Google Scholar] [CrossRef]
  9. Ahmad, R.; Tripathy, N.; Ahn, M.-S.; Hahn, Y.-B. Solution Process Synthesis of High Aspect Ratio ZnO Nanorods on Electrode Surface for Sensitive Electrochemical Detection of Uric Acid. Sci. Rep. 2017, 7, 46475. [Google Scholar] [CrossRef] [Green Version]
  10. Nagal, V.; Kumar, V.; Khan, M.; Alomar, S.; Tripathy, N.; Singh, K.; Khosla, A.; Ahmad, N.; Hafiz, A.K.; Ahmad, R. A Highly Sensitive Uric Acid Biosensor Based on Vertically Arranged ZnO Nanorods on ZnO Nanoparticles Seeded Electrode. New J. Chem. 2021, 45, 18863–18870. [Google Scholar] [CrossRef]
  11. Murugan, N.; Chan-Park, M.B.; Sundramoorth, A.K. Electrochemical Detection of Uric Acid on Exfoliated Nanosheets of Graphitic-Like Carbon Nitride (g-C3N4) Based Sensor. J. Electrochem. Soc. 2019, 166, B3163–B3170. [Google Scholar] [CrossRef]
  12. Ahmad, R.; Mahmoudi, T.; Ahn, M.-S.; Hahn, Y.-B. Recent Advances in Nanowires-Based Field-Effect Transistors for Biological Sensor Applications. Biosens. Bioelectron. 2018, 100, 312–325. [Google Scholar] [CrossRef]
  13. Bhat, K.S.; Ahmad, R.; Mahmoudi, T.; Hahn, Y.-B. High Performance Chemical Sensor with Field-Effect Transistors Array for Selective Detection of Multiple Ions. Chem. Eng. J. 2021, 417, 128064. [Google Scholar] [CrossRef]
  14. Yassin, M.A.; Shrestha, B.K.; Ahmad, R.; Shrestha, S.; Parka, C.H.; Kim, C.S. Exfoliated Nanosheets of Co3O4 Webbed with Polyaniline Nanofibers: A Novel Composite Electrode Material for Enzymeless Glucose Sensing Application. J. Industrial Eng. Chem. 2019, 73, 106–117. [Google Scholar] [CrossRef]
  15. Jayaprakash, G.K.; Swamy, B.E.K.; Sánchez, J.P.M.; Li, X.; Sharma, S.C.; Lee, S.-L. Electrochemical and Quantum Chemical Studies of Cetylpyridinium Bromide Modified Carbon Electrode Interface for Sensor Applications. J. Molecular Liquids 2020, 315, 113719. [Google Scholar] [CrossRef]
  16. Piedras, J.; Dominguez, R.B.; Gutiérrez, J.M. Determination of Uric Acid in Artificial Saliva with Compact AMP3291 Reader and Au Nanoparticles Modified Electrode. Chemosensors 2021, 9, 73. [Google Scholar] [CrossRef]
  17. Kudo, H.; Takagi, Y. Electrochemical Biosensor for Simplified Determination of Salivary Uric Acid. Sens. Mater. 2018, 30, 1187–1195. [Google Scholar]
  18. Wang, C.; Du, J.; Wang, H.; Zou, C.; Jiang, F.; Yang, P.; Du, Y. A Facile Electrochemical Sensor based on Reduced Graphene Oxide and Au Nanoplates Modified Glassy Carbon Electrode for Simultaneous Detection of Ascorbic Acid, Dopamine and Uric Acid. Sensors Actuators B Chem. 2014, 204, 302–309. [Google Scholar] [CrossRef]
  19. Hadia, Z.; Ghanbari, K. A Novel Electrochemical Sensor for Determination of Uric Acid in the Presence of Ascorbic Acid and Dopamine Based on a Carbon Paste Electrode Modified with an Electrochemically Reduced Para-Nitrobenzoic Acid/Graphene Oxide Nanocomposite. New J. Chem. 2022, 46, 12941–12951. [Google Scholar] [CrossRef]
  20. Metto, M.; Eramias, S.; Gelagay, B.; Washe, A.P. Voltammetric Determination of Uric Acid in Clinical Serum Samples Using DMF Modified Screen Printed Carbon Electrodes. Int. J. Electrochem. 2019, 2019, 6318515. [Google Scholar] [CrossRef] [Green Version]
  21. Boumya, W.; Taoufik, N.; Achak, M.; Barka, N. Chemically Modified Carbon-Based Electrodes for the Determination of Paracetamol in Drugs and Biological Samples. J. Pharmaceutical Analysis 2021, 11, 138–154. [Google Scholar] [CrossRef] [PubMed]
  22. Tasić, Ž.Z.; Mihajlović, M.B.P.; Radovanović, M.B.; Simonović, A.T.; Medić, D.V.; Antonijević, M.M. Electrochemical Determination of L-tryptophan in Food Samples on Graphite Electrode Prepared from Waste Batteries. Sci. Rep. 2022, 12, 5469. [Google Scholar] [CrossRef] [PubMed]
  23. Ahmad, R.; Wolfbeis, O.S.; Hahn, Y.-B.; Alshareef, H.N.; Torsi, L.; Salama, K.N. Deposition of Nanomaterials: A Crucial Step in Biosensor Fabrication. Materials Today Commun. 2018, 17, 289–321. [Google Scholar] [CrossRef] [Green Version]
  24. Nagal, V.; Tuba, T.; Kumar, V.; Alam, S.; Ahmad, A.; Alshammari, M.B.; Hafiz, A.K.; Ahmad, R. A Non-Enzymatic Electrochemical Sensor Composed of Nano-Berry Shaped Cobalt Oxide Nanostructures on a Glassy Carbon Electrode for Uric Acid Detection. New J. Chem. 2022, 46, 12333. [Google Scholar] [CrossRef]
  25. Kogularasu, S.; Govindasamy, M.; Chen, S.-M.; Akilarasan, M.; Mani, V. 3D graphene Oxide-Cobalt Oxide Polyhedrons for Highly Sensitive Non-Enzymatic Electrochemical Determination of Hydrogen Peroxide. Sens. Actuators B Chem. 2017, 253, 773–783. [Google Scholar] [CrossRef]
  26. Zhang, H.; Liu, S. A Combined Self-Assembly and Calcination Method for Preparation of Nanoparticles-Assembled Cobalt Oxide Nanosheets using Graphene Oxide as Template and their Application for Non-Enzymatic Glucose Biosensing. J. Colloid Interface Sci. 2017, 485, 159–166. [Google Scholar] [CrossRef]
  27. Kang, L.; He, D.; Bie, L.; Jiang, P. Nanoporous Cobalt Oxide Nanowires for Non-Enzymatic Electrochemical Glucose Detection. Sens. Actuators B Chem. 2015, 220, 888–894. [Google Scholar] [CrossRef]
  28. Mondal, S.; Madhuri, R.; Sharma, P.K. Probing the Shape Specific Electrochemical Properties of Cobalt Oxide Nanostructures for their Application as Selective and Sensitive Nonenzymatic Glucose Sensors. J. Mater. Chem. C 2017, 5, 6497–6505. [Google Scholar] [CrossRef]
  29. Chang, A.S.; Memon, N.N.; Amin, S.; Chang, F.; Aftab, U.; Abro, M.I.; Chandio, A.; Shah, A.A.; Ibupoto, M.H.; Ansari, M.A.; et al. Facile Non-Enzymatic Lactic Acid Sensor based on Cobalt Oxide Nanostructures. Electroanalysis 2019, 31, 1296–1303. [Google Scholar] [CrossRef] [Green Version]
  30. Chang, A.S.; Tahira, A.; Chang, F.; Memon, N.N.; Nafady, A.; Kasry, A.; Ibupoto, Z.H. Silky Co3O4 nanostructures for the selective and sensitive enzyme free sensing of uric acid, RSC Adv. 2021, 11, 5156–5162. 11.
  31. Sayyad, P.W.; Ansari, T.R.; Ingle, N.N.; Al-Gahouari, T.; Bodkhe, G.A.; Mahadik, M.M.; Shirsat, S.M.; Shirsat, M.D. L-Cysteine Peptide-Functionalized PEDOT-PSS/rGO Nanocomposite for Selective Electrochemical Detection of Lead Pb(II) Ions. Applied Physics A 2021, 127, 381. [Google Scholar] [CrossRef]
  32. Ognjanović, M.; Nikolić, K.; Bošković, M.; Pastor, F.; Popov, N.; Marciuš, M.; Krehula, S.; Antić, B.; Stanković, D.M. Electrochemical Determination of Morphine in Urine Samples by Tailoring FeWO4/CPE Sensor. Biosensors 2022, 12, 932. [Google Scholar] [CrossRef] [PubMed]
  33. Verma, S.; Singh, A.; Shukla, A.; Kaswan, J.; Arora, K.; Ramirez-Vick, J.; Singh, P.; Singh, S.P. Anti-IL8/AuNPs-rGO/ITO as an Immunosensing Platform for Noninvasive Electrochemical Detection of Oral Cancer. ACS Appl. Mater. Interfaces 2017, 9, 27462–27474. [Google Scholar] [CrossRef] [PubMed]
  34. Khan, M.; Nagal, V.; Nakate, U.T.; Khan, M.R.; Khosla, A.; Ahmad, R. Engineered CuO Nanofibers with Boosted Non-Enzymatic Glucose Sensing Performance. J. Electrochem. Soc. 2021, 168, 067507. [Google Scholar] [CrossRef]
  35. Numan, A.; Shahid, M.M.; Omar, F.S.; Ramesh, K.; Ramesh, S. Facile Fabrication of Cobalt Oxide Nanograin-Decorated Reduced Graphene Oxide Composite as Ultrasensitive Platform for Dopamine Detection. Sens. Actuators B Chem. 2017, 238, 1043–1051. [Google Scholar] [CrossRef]
  36. Ma, L.; Zhang, Q.; Wu, C.; Zhang, Y.; Zeng, L. PtNi Bimetallic Nanoparticles Loaded MoS2 Nanosheets: Preparation and Electrochemical Sensing Application for the Detection of Dopamine and Uric Acid. Anal. Chim. Acta 2019, 1055, 17–25. [Google Scholar] [CrossRef]
  37. Turkkan, G.; Bas, S.Z.; Atacan, K.; Ozmen, M. An Electrochemical Sensor based on a Co3O4-ERGO Nanocomposite Modified Screen-Printed Electrode for Detection of Uric Acid in Artificial Saliva. Anal. Methods 2022, 14, 67–75. [Google Scholar] [CrossRef]
  38. Ahmad, R.; Ahn, M.-S.; Hahn, Y.-B. A Highly Sensitive Non enzymatic Sensor Based on Fe2O3 Nanoparticle Coated ZnO Nanorods for Electrochemical Detection of Nitrite. Adv. Mater. Interfaces 2017, 4, 1700691. [Google Scholar] [CrossRef]
  39. Ponnaiah, S.K.; Periakaruppan, P.; Vellaichamy, B. New Electrochemical Sensor Based on a Silver-Doped Iron Oxide Nanocomposite Coupled with Polyaniline and Its Sensing Application for Picomolar Level Detection of Uric Acid in Human Blood and Urine Samples. J. Phys. Chem. B 2018, 122, 3037–3046. [Google Scholar] [CrossRef]
  40. Zhao, Y.; Yan, X.; Kang, Z.; Lin, P.; Fang, X. Highly Sensitive Uric Acid Biosensor based on Individual Zinc Oxide Micro/Nanowires. Microchim. Acta 2013, 180, 759–766. [Google Scholar] [CrossRef]
  41. Yan, Q.; Zhi, N.; Yang, L.; Xu, G.; Feng, Q.; Zhang, Q.; Sun, S. A Highly Sensitive Uric Acid Electrochemical Biosensor based on a Nano-Cube Cuprous Oxide/Ferrocene/Uricase Modifed Glassy Carbon Electrode. Sci. Rep. 2020, 10, 10607. [Google Scholar] [CrossRef]
  42. Arvand, M.; Hassannezhad, M. Magnetic Core-Shell Fe3O4@SiO2/MWCNT Nanocomposite Modified Carbon Paste Electrode for Amplified Electrochemical Sensing of Uric Acid. Mater. Sci. Eng. C 2014, 36, 160–167. [Google Scholar] [CrossRef] [PubMed]
  43. Han, S.H.; Ha, Y.-J.; Kang, E.H.; Shin, K.; Lee, Y.J.; Lee, G.-J. Electrochemical Detection of Uric Acid in Undiluted Human Saliva using Uricase Paper Integrated Electrodes. Sci. Rep. 2022, 12, 12033. [Google Scholar] [CrossRef] [PubMed]
  44. Mazzaraa, F.; Patellaa, B.; Aiello, G.; O’Riordan, A.; Torino, C.; Vilasi, A.; Inguant, R. Electrochemical Detection of Uric Acid and Ascorbic Acid using r-GO/NPs based Sensors. Electrochim. Acta 2021, 388, 138652. [Google Scholar] [CrossRef]
  45. Santos, P.L.D.; Katic, V.; Toledo, K.C.; Bonacin, J.A. Photochemical One-Pot Synthesis of Reduced Graphene Oxide/Prussian Blue Nanocomposite for Simultaneous Electrochemical Detection of Ascorbic Acid, Dopamine, and Uric Acid. Sens. Actuators B Chem. 2018, 255, 2437. [Google Scholar] [CrossRef]
  46. Thanh, T.S.; Qui, P.T.; Tu, N.T.T.; Toan, T.T.T.; Hoa, T.T.B.; Son, L.V.T.; Nguyen, D.M.; Tuyen, T.N.; Khieu, D.Q. Electrochemical Determination of Uric Acid in Urine by Using Zeolite Imidazolate Framework-11 Modified Electrode. J. Nanomater. 2021, 2021, 9914062. [Google Scholar] [CrossRef]
Scheme 1. Schematic showing the synthesis process of porous cobalt oxide nanosheets, sensor fabrication process, and measurement techniques used.
Scheme 1. Schematic showing the synthesis process of porous cobalt oxide nanosheets, sensor fabrication process, and measurement techniques used.
Biosensors 12 01140 sch001
Figure 1. FESEM images at low- (a,b) and high- (c) magnifications and XRD analysis (d) of porous cobalt oxide nanostructure.
Figure 1. FESEM images at low- (a,b) and high- (c) magnifications and XRD analysis (d) of porous cobalt oxide nanostructure.
Biosensors 12 01140 g001
Figure 2. TEM images at low- (a), high- (b) magnifications, SAED pattern (c), and BET Nitrogen adsorption-desorption isotherms of porous cobalt oxide sheet-like nanostructure. Inset (d) shows the pore size distribution plot of cobalt oxide sheet-like nanostructure.
Figure 2. TEM images at low- (a), high- (b) magnifications, SAED pattern (c), and BET Nitrogen adsorption-desorption isotherms of porous cobalt oxide sheet-like nanostructure. Inset (d) shows the pore size distribution plot of cobalt oxide sheet-like nanostructure.
Biosensors 12 01140 g002
Figure 3. EIS spectra of bare GCE and cobalt oxide/GCE electrodes recorded in the redox probe solution of [Fe(CN)6]3−/4− with KCl (0.1 M) “as supporting electrolyte”. (a) Nyquist plots of bare GCE (black line) and cobalt oxide/GCE (red line) electrodes and their respective Bode plots (b,c). Inset a shows the Randle circuit.
Figure 3. EIS spectra of bare GCE and cobalt oxide/GCE electrodes recorded in the redox probe solution of [Fe(CN)6]3−/4− with KCl (0.1 M) “as supporting electrolyte”. (a) Nyquist plots of bare GCE (black line) and cobalt oxide/GCE (red line) electrodes and their respective Bode plots (b,c). Inset a shows the Randle circuit.
Biosensors 12 01140 g003
Figure 4. (a) CV response curves of bare GCE and cobalt oxide/GCE electrodes recorded in the redox probe solution of [Fe(CN)6]3−/4− with KCl (0.1 M) “as supporting electrolyte”. The CV curves were measured at the fixed scan rate (50 mV/s). (b) CV analysis of cobalt oxide/GCE electrode at varying scan rates and respective calibration plot (c). In figure (c), the arrow shows the scan rate increasing direction from 10 mV/s to 250 mV/s.
Figure 4. (a) CV response curves of bare GCE and cobalt oxide/GCE electrodes recorded in the redox probe solution of [Fe(CN)6]3−/4− with KCl (0.1 M) “as supporting electrolyte”. The CV curves were measured at the fixed scan rate (50 mV/s). (b) CV analysis of cobalt oxide/GCE electrode at varying scan rates and respective calibration plot (c). In figure (c), the arrow shows the scan rate increasing direction from 10 mV/s to 250 mV/s.
Biosensors 12 01140 g004
Figure 5. (a) CV response curves recorded for cobalt oxide/GCE electrode in PBS without and with uric acid (10 µM) at 50 mV/s, (b) CV responses of cobalt oxide/GCE electrode with increasing uric acid concentrations (0−2500 µM), and (c) plot of current response vs. uric acid concentration showing linear and non−linear regions. Inset b shows the magnified CV response curves for low-concentration uric acid. Inset c shows the calibration plot of the linear region (i.e., response vs. uric acid concentration).
Figure 5. (a) CV response curves recorded for cobalt oxide/GCE electrode in PBS without and with uric acid (10 µM) at 50 mV/s, (b) CV responses of cobalt oxide/GCE electrode with increasing uric acid concentrations (0−2500 µM), and (c) plot of current response vs. uric acid concentration showing linear and non−linear regions. Inset b shows the magnified CV response curves for low-concentration uric acid. Inset c shows the calibration plot of the linear region (i.e., response vs. uric acid concentration).
Biosensors 12 01140 g005
Scheme 2. Scheme showing uric acid oxidation over the porous cobalt oxide nanosheets/GCE sensor.
Scheme 2. Scheme showing uric acid oxidation over the porous cobalt oxide nanosheets/GCE sensor.
Biosensors 12 01140 sch002
Figure 6. (a) DPV response curves recorded for cobalt oxide/GCE sensor in PBS without and with uric acid (10 µM), (b) DPV responses of cobalt oxide/GCE electrode with increasing uric acid concentrations (0–2500 µM), and (c) plot of current response vs. uric acid concentration showing linear and non-linear regions. Inset b shows the magnified DPV response curves for low-concentration uric acid. Inset c shows the calibration plot of the linear region (i.e., response vs. uric acid concentration).
Figure 6. (a) DPV response curves recorded for cobalt oxide/GCE sensor in PBS without and with uric acid (10 µM), (b) DPV responses of cobalt oxide/GCE electrode with increasing uric acid concentrations (0–2500 µM), and (c) plot of current response vs. uric acid concentration showing linear and non-linear regions. Inset b shows the magnified DPV response curves for low-concentration uric acid. Inset c shows the calibration plot of the linear region (i.e., response vs. uric acid concentration).
Biosensors 12 01140 g006
Figure 7. (a) CV response curves recorded for cobalt oxide/GCE sensor in PBS with 25 µM uric acid only (black curve) and 25 µM uric acid and 100 µM of each interfering species (i.e., lactic acid, L−cysteine, glucose, urea, fructose, sodium chloride, and potassium chloride) (red curve). (b) CV response curves of cobalt oxide/GCE sensor for 25 µM uric acid showing stability of sensor after 30 and 45 days. (c) DPV response curves for 25 µM uric acid only (black curve) and 25 µM uric acid and 100 µM of each interfering species (i.e., ascorbic acid, dopamine, and urea).
Figure 7. (a) CV response curves recorded for cobalt oxide/GCE sensor in PBS with 25 µM uric acid only (black curve) and 25 µM uric acid and 100 µM of each interfering species (i.e., lactic acid, L−cysteine, glucose, urea, fructose, sodium chloride, and potassium chloride) (red curve). (b) CV response curves of cobalt oxide/GCE sensor for 25 µM uric acid showing stability of sensor after 30 and 45 days. (c) DPV response curves for 25 µM uric acid only (black curve) and 25 µM uric acid and 100 µM of each interfering species (i.e., ascorbic acid, dopamine, and urea).
Biosensors 12 01140 g007
Table 1. Sensing performance comparison with previous reports based on electrochemical studies.
Table 1. Sensing performance comparison with previous reports based on electrochemical studies.
Modified ElectrodeMethod of DetectionSensitivity
(µAcm−2mM−1)
Linear range (µM)Detection limit (µM)Ref.
g-C3N4 NSs/GCEDPV-100–10004.5[11]
Co3O4 nanoberries/GCECV2065–3000~2.4[24]
Co3O4 nanostructures/GCECV-500–530100[30]
Ag-Fe2O3@PANIDPV128.290.001–0.9000.000102[39]
Nafion/Uricase/ZnO/AuAmperometry89.74100–59025.6[40]
Cu2O/ferrocene/uricase/GCEDPV1.90.1–10000.0596[41]
Fe3O4@SiO2/MWCNTSWV0.3030.60–1000.13[42]
Uricase/PPD/PrB-SPCECV4.950–100018.7[43]
ITO-rGO-AuNPsLSV0.3110–5003.6[44]
PrGO/PB 100/GCECV-40–4158.0[45]
ZIF-11/GCEDPV17420–5400.48[46]
Co3O4 porous NSs/GCECV470.40–100010This work
Co3O4 porous NSs/GCEDPV3566.50–80012This work
Abbreviations: Co3O4, cobalt oxide; GCE, glassy carbon electrode; CV, cyclic voltammetry; Ag, silver; Fe2O3, iron oxide; PANI, polyaniline; DPV, differential pulse voltammetry; ZnO, zinc oxide; Au, gold; CVD, chemical vapour deposition; Cu2O, cuprous oxide; SiO2, silicon dioxide; MWCNT, multiwalled carbon nanotubes; SWV, square-wave voltammetry; PPD, p-phenylenediamine; PrB, Prussian blue; SPCE, screen-printed-carbon-electrode; ITO, indium-doped tin oxide; rGO, reduced graphene oxide; NPs, nanoparticles; LSV, linear-sweep voltammetry, ZIF-11, zeolite imidazole framework; g-C3N4, graphitic-like carbon nitride; NSs, nanosheets.
Table 2. Uric acid detection in human serum with cobalt oxide/GCE sensor.
Table 2. Uric acid detection in human serum with cobalt oxide/GCE sensor.
SampleAdded Uric Acid (µM)Found (µM)Recovery (%)RSD (%) (n = 3)
Human serum0282--
10291.6962.38
50329.895.62.92
100379973.45
50077198.24.25
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Masrat, S.; Nagal, V.; Khan, M.; Moid, I.; Alam, S.; Bhat, K.S.; Khosla, A.; Ahmad, R. Electrochemical Ultrasensitive Sensing of Uric Acid on Non-Enzymatic Porous Cobalt Oxide Nanosheets-Based Sensor. Biosensors 2022, 12, 1140. https://doi.org/10.3390/bios12121140

AMA Style

Masrat S, Nagal V, Khan M, Moid I, Alam S, Bhat KS, Khosla A, Ahmad R. Electrochemical Ultrasensitive Sensing of Uric Acid on Non-Enzymatic Porous Cobalt Oxide Nanosheets-Based Sensor. Biosensors. 2022; 12(12):1140. https://doi.org/10.3390/bios12121140

Chicago/Turabian Style

Masrat, Sakeena, Vandana Nagal, Marya Khan, Iqra Moid, Shamshad Alam, Kiesar Sideeq Bhat, Ajit Khosla, and Rafiq Ahmad. 2022. "Electrochemical Ultrasensitive Sensing of Uric Acid on Non-Enzymatic Porous Cobalt Oxide Nanosheets-Based Sensor" Biosensors 12, no. 12: 1140. https://doi.org/10.3390/bios12121140

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop