Next Article in Journal
Microstructural Characterization of Laser Weld of Hot-Stamped Al-Si Coated 22MnB5 and Modification of Weld Properties by Hybrid Welding
Next Article in Special Issue
High Mechanic Enhancement of Chopped Carbon Fiber Reinforced-Low-Density Unsaturated Polyester Resin Composite at Low Preparation Temperature with Facile Polymerization
Previous Article in Journal
Polymer-Based Carriers in Dental Local Healing—Review and Future Challenges
Previous Article in Special Issue
A Micro-Computed Tomography Comparison of the Porosity in Additively Fabricated CuCr1 Alloy Parts Using Virgin and Surface-Modified Powders
 
 
Article
Peer-Review Record

Improved Biological Responses of Titanium Coating Using Laser-Aided Direct Metal Fabrication on SUS316L Stainless Steel

Materials 2021, 14(14), 3947; https://doi.org/10.3390/ma14143947
by Tae-In Kim 1,†, Se-Won Lee 2,3,†, Woo-Lam Jo 2,4, Yong-Sik Kim 2,4, Seung-Chan Kim 2,5, Soon-Yong Kwon 2,5 and Young-Wook Lim 2,4,*
Reviewer 1: Anonymous
Reviewer 2: Anonymous
Materials 2021, 14(14), 3947; https://doi.org/10.3390/ma14143947
Submission received: 3 June 2021 / Revised: 29 June 2021 / Accepted: 13 July 2021 / Published: 14 July 2021
(This article belongs to the Special Issue Porous Materials and Advanced Manufacturing Technologies)

Round 1

Reviewer 1 Report

The introduction is very limited. In particular, the publications focused on laser machining of Ti and austenitic stainless steels should be given. Now, it is unclear whether similar attempts have been carried out (there has been plenty of them) and whether this scientific problem is justified.

Materials and Methods lack information on the chemical composition of both steels, Ti powder (also on average grain size), and applied laser equipment (name, manufacturer, details of laser deposition. There is neither name nor manufacturer for EDS. The same is true for the push-out tests – give a standard or describe the method.   And also for the light microscope. Most importantly, there is no mention of an obligatory allowance for tests on animals (rabbits), please give the number of such a document. If not, such tests are illegal and should not be taken into account.

Results demonstrate also some errors. Figure 4: change Mpa to MPa. Line 211: micrometers, the unit, is improper.

Author Response

Please see the attachment.

Author Response File: Author Response.pdf

Reviewer 2 Report

Title:

Does Titanium Coating Using Laser-aided Direct Metal Fabrication Improve Biological Responses of SUS316L Stainless Steel?

 

This manuscript mentioned that the investigation of the biological responses on the Ti-coated SUS316 and SUS316. However, there are some unclear explanations, especially metal preparation and properties. The manuscript needs to be major changed for publication.

 

  1. The title is a question form. I wonder that it is inappropriate for an article in the field of Materials because this journal is not a medical journal. This article is not a review one.
  2. 88, “The average surface roughness (Ra value) was 6.1 µm±0.23 lm”. What the unit of lm? The quantitative values of the surface roughness need to be mentioned to compare the size of the osteoblast cell. The authors need a more detailed discussion about this point.
  3. Ti coating by laser sintering is difficult because Ti is easily oxidized. In the 3D printing process, what was the atmosphere? What’s the pressure? EDS mapping is shown in Fig.1(B). What’s the concentration of oxygen in the coated Ti? Is it increased after the laser sintering by comparing it with the raw Ti powder? The basic data for the preparation of the samples are unclear.
  4. The SUS-DMF is effective in push-out test shown in Fig.4. However, the surface of Ti must be oxidized without specific treatment or in the atmosphere. My question is whether the improvement by SUS-DMS was caused by Ti-coating or Ti oxide coating. Furthermore, is the kind of coated materials important or NOT-SUS surface important for the improvement? The authors should mention the mechanism for the improvement.

Author Response

Please see the attachment.

Author Response File: Author Response.pdf

Round 2

Reviewer 2 Report

Title:

Does Titanium Coating Using Laser-aided Direct Metal Fabrication Improve Biological Responses of SUS316L Stainless Steel?

 

This manuscript mentioned that the investigation of the biological responses on the Ti-coated SUS316 and SUS316. All the reviewer’s concerns became clear in the revised manuscript. The revised manuscript is sufficient interest for publication.

 

Back to TopTop