Miniature Ultrasound Transducer Incorporating Sm-PMN-PT 1-3 Composite

1
Department of Applied Physics, The Hong Kong Polytechnic University, Hong Kong, China
2
Department of Electrical and Electronic Engineering, The Hong Kong Polytechnic University, Hong Kong, China
3
College of Physics, University-Industry Joint Center for Ocean Observation and Broadband Communication, National Demonstration Center for Experimental Applied Physics Education, Qingdao University, Qingdao 266071, China
4
Centre for Medical and Industrial Ultrasonics, James Watt School of Engineering, University of Glasgow, Glasgow G12 8QQ, UK
*
Author to whom correspondence should be addressed.
J. Compos. Sci. 2024, 8(3), 80; https://doi.org/10.3390/jcs8030080
Submission received: 15 January 2024 / Revised: 4 February 2024 / Accepted: 19 February 2024 / Published: 22 February 2024
(This article belongs to the Special Issue Feature Papers in Journal of Composites Science in 2024)

Abstract

:
Piezoelectric 1-3 composite materials have become extensively utilized in diagnostic ultrasound transducers owing to their high electromechanical coupling coefficient, low acoustic impedance, and low dielectric loss. In this study, Sm-doped PMN-PT ceramic/epoxy 1-3 composite with a ceramic volume fraction of 60% is fabricated by the dice-and-fill method, resulting in a high piezoelectric constant (650 pC/N) and clamped dielectric constant (2350). Utilizing the exceptionally high clamped dielectric constant, a low-frequency (12.4 MHz) ultrasound transducer is developed with a miniature aperture size (0.84 mm × 0.84 mm), exhibiting a −6 dB bandwidth of 70% and an insertion loss of −20.5 dB. The imaging capability of the miniature composite transducer is validated through both phantom and ex vivo imaging. The satisfactory results indicate that Sm-doped ceramic/epoxy composites possess significant potential for miniature devices in biomedical imaging applications.

1. Introduction

Ultrasound imaging holds a crucial role in the biomedical field, owing to its non-invasive and radiation-free characteristics [1,2]. The ultrasound transducer stands out as a critical factor in determining imaging quality, as it facilitates the conversion of electrical energy into mechanical energy and vice versa [3]. Piezoelectric ceramic materials, notably lead zirconate titanate (PZT), exhibit remarkable piezoelectric properties and stability [4,5], making them a popular choice as the active layer in ultrasound transducers. However, one limitation of ceramic materials is their relatively high acoustic impedance (>30 MRayl), in contrast to the low acoustic impedance of biological tissues (~1.5 Mrayl). This acoustic impedance mismatch can result in ultrasound transducer performance degradation. To address this issue, carefully designed matching layers with specific thicknesses become essential. These layers aim to enhance the acoustic transmission rate between the transducer’s front surface and biological tissues, ensuring optimal ultrasound imaging performance [6].
To overcome the aforementioned challenge, piezoelectric composite materials have been employed, aiming for lower acoustic impedance, higher electromechanical coupling coefficients, and reduced dielectric losses [7]. These composites involve the integration of both piezoelectric and polymer materials, forming either 1-3 or 2-2 connectivity configurations [8]. The 1-3 composite, known for its superior effective thickness coupling performance, has found widespread use in ultrasound transducer development [9,10]. Initially, PZT/epoxy-based 1-3 composites were extensively investigated and are now still widely used for biomedical ultrasound imaging applications [11,12]. In addition to conventional PZT ceramics, extensive research has been conducted on lead-free piezoelectric ceramics for the application of ultrasound transducers, particularly in the form of 1-3 composites. Barium-modified bismuth sodium titanate, 0.94(Bi0.5Na0.5)TiO3-0.06BaTiO3 (BNT) fibers were derived using the sol-gel method to develop a 15-MHz 1-3 composite ultrasound transducer [13]. Chen et al. explored the use of manganese-doped 0.7BiFeO3-0.3BaTiO3 ceramics for ultrasound transducer development, demonstrating its potential for practical applications [14]. Potassium sodium niobate (KNN)-based ceramics, known for their high piezoelectric performance among lead-free ceramics, were also investigated. With Fe2O3-modified KNN ceramics, a broadband ultrasound transducer based on the corresponding composite was successfully developed [15]. While these composites demonstrated an improved electromechanical coupling coefficient compared to bulk ceramic materials, the piezoelectric constant was compromised due to the passive nature of epoxy, which did not contribute to vibration during the piezoelectric effect. Consequently, the sensitivity of 1-3 composite-based ultrasound transducers, particularly unfocused ones, was deemed unsatisfactory [16].
To enhance the performance of 1-3 composites, there arose a demand for piezoelectric ceramics with high piezoelectric constants. In recent years, rare-earth-doped (1 − x)Pb(Mg1/3Nb2/3)O3xPbTiO3 (PMN-PT) has emerged as a promising candidate, demonstrating substantial improvements in piezoelectric constants. For instance, ceramics doped with europium (Eu) or samarium (Sm) can achieve piezoelectric constants exceeding 1000 pC/N [17,18]. These materials also exhibit high dielectric constants, a desirable feature for miniature biomedical ultrasound transducers.
In this study, 2.5%Sm-0.72PMN-0.28PT with high piezoelectric coefficient and dielectric permittivity was synthesized. A high-frequency 1-3 composite was then fabricated using the dice-and-fill method for the development of a miniature ultrasound transducer. With the high dielectric permittivity, the aperture size of the transducer was significantly reduced but retained a satisfying bandwidth and sensitivity. Spatial resolution was measured to verify the transducer performance. The ex vivo imaging was also conducted to show potential for biomedical imaging applications.

2. Materials and Methods

2.1. Preparation and Characterizations

With Pb3O4, Sm2O3, Mg2(OH)2CO3, Nb2O5, and TiO2 as raw materials, 2.5%Sm-0.72PMN-0.28PT was synthesized along with our previous optimized method [19]. Two kinds of pre-calcination powders, one with a rhombohedral phase and the other with a tetragonal phase were initially prepared at 880 °C for 4 h. Subsequently, they were mixed and sintered at 1250 °C for 1.5 h, following the designated proportions to synthesize the target composition of Pb0.9625Sm0.025(Mg1/3Nb2/3)0.72Ti0.28O3. To test the piezoelectric and dielectric properties of the ceramics, silver paste was applied to both surfaces and sintered at 650 °C for 30 min to form electrodes. After being poled at 3 kV/mm for 10 min and undergoing an aging treatment for 24 h, the electrical properties were measured.
The crystal structures and morphologies of the samples were characterized using X-ray powder diffraction (XRD, Rigaku, Tokyo, Japan) and scanning electron microscopy (SEM, JSM-6510, JEOL, Tokyo, Japan), respectively. After co-fired silver electrodes on both surfaces of ceramic disk, the ferroelectric hysteresis loop was measured at room temperature. The piezoelectric coefficient (d33) was measured using a quasi-static d33 tester (ZJ-6A, Chinese Academy of Science, Beijing, China). The dielectric properties and electromechanical coupling coefficient (kt) were measured by an impedance analyzer (HP4294A, Agilent, Santa Clara, CA, USA).
To determine the acoustic impedance of ceramics, the measurements of longitudinal sound velocity (c) and density (ρ) were conducted. The density of samples was determined using the Archimedes drainage method. For the longitudinal sound velocity, two commercial 5-MHz ultrasound transducers were positioned at opposite sides with a specific distance and immersed in distilled water. One transducer was excited by a pulser/receiver to transmit signals, while the other received the signals. The time-of-flight results of ultrasound signals were recorded both with and without inserting the ceramic sample between the two transducers. The longitudinal sound velocity (csample) was then calculated as:
τ w a t e r τ s a m p l e = t 1 c w a t e r 1 c s a m p l e ,
where cwater and csample are sound velocities in water and sample, respectively, τ water and τ sample are the times for receiving signals without and with the sample, respectively, and t is the thickness of the sample.

2.2. Design and Fabrication of 1-3 Composites

To achieve high-frequency 1-3 composites, the diced kerfs should be sufficiently thin. In this study, a ZH05-AA (DISCO Co., Ltd., Omori-Kita, Tokyo, Japan) blade was employed to achieve a kerf width of 21 μm. After determining the kerf width, the ceramic pillar width can be calculated through the below equation of ceramic volume fraction ( ϕ ):
ϕ = ( c e r a m i c   p i l l a r   w i d t h ) 2 ( c e r a m i c   p i l l a r   w i d t h + k e r f   w i d t h ) 2 .
The volume fraction was determined as ~60% to retain good piezoelectric performance such that the ceramic pillar width was set as 72 μm. To mitigate the lateral vibration modes, the final thickness of 1-3 composites was lapped down to 96 μm, ensuring the attainment of a specific width-to-height aspect ratio.
A conventional dice-and-fill method, as depicted in Figure 1, was commonly employed to fabricate the 1-3 composites [20,21]. The unpoled bulk piezoelectric ceramic (1 mm thick) was initially diced with a pitch of 93 μm in one direction, with a dicing depth of ~300 μm. Subsequently, low-viscosity epoxy (EPO-TEK 301, Epoxy Tech., Billerica, MA, USA) was applied to fill the diced kerfs. After curing for 24 h at room temperature, excess epoxy was removed through lapping and polishing. The same dicing procedure was then repeated in the direction perpendicular to the initial dicing, followed by the same filling method. Finally, the entire stack was lapped from the other side of the ceramic to remove the bulk ceramic part. Poling was conducted after dicing and lapping, as these processes may lead to the depolarization of ceramic with a relatively low Curie temperature [22].
The characteristics of composite material were measured, including d33, kt, and clamped dielectric constant (εs). Based on electrical impedance/phase spectra, kt can be calculated as:
k t = π 2 f r f a tan π 2 f a f r f a ,
where fr and fa are resonant frequency and anti-resonant frequency, respectively. In transducer design, achieving an electrical impedance of near 50 Ω for effective electrical matching, and this impedance is closely related to εs. The measurement of εs involves analyzing series capacitance and dissipation curves, as expressed:
ε s = C s t A ,
where Cs is the series capacitance at a frequency that is twice the resonant frequency, and t and A are the thickness and aperture size, respectively, of the composite material. To achieve effective electrical matching, the aperture size was calculated as:
A = t 2 π ε s ε 0 f r Z ,
where ε 0 is the permittivity at vacuum (8.854 × 10−12 F/m), and Z was set as 50 Ω.

2.3. Fabrication and Evaluation of Transducer

PiezoCAD software (3.06) was employed for transducer performance simulation. Based on the Krimboltz, Leedom, and Mattaei (KLM) model, PiezoCAD could provide an optimized design for ultrasound transducers. In this study, E-solder 3022 (Von Roll Isola Inc., Schenetady, NY, USA) silver epoxy was employed as the backing layer due to its electrical conductivity and high acoustic attenuation. A single matching layer has been commonly employed to enhance the performance of ultrasound transducer, whose ideal acoustic impedance could be expressed as:
Z m a t c h i n g   l a y e r = Z 0 1 / 3 Z 1 2 / 3 ,
where Z0 and Z1 are acoustic impedances of piezoelectric composite and biological tissues (1.5 MRayl), respectively [23]. Vapor-deposited parylene was selected as the single matching layer not only for its proper acoustic impedance but also for its waterproof properties.
The transducer fabrication process follows the conventional ultrasound transducer technology: An unpolarized 2.5%Sm-0.72PMN-0.28PT was firstly diced and filled to form a 1-3 composite structure, which was then lapped to 96 μm. After polishing the surface, a Cr/Au electrode was sputtered onto it. E-solder was subsequently cast on the surface, using a centrifuge at 2000 rpm for 15 min to ensure connectivity throughout the whole aperture. After curing overnight, the backing layer was lapped down to ~0.5 mm. The other side of composite was also polished and electroded. The acoustic stacks were diced to the desired aperture size with a square shape. A co-axial cable was used for electrode connection on both sides of composite with E-solder. The stack connected to the cable was then placed into a stainless-steel tube, with epoxy for isolation and stabilization. The parylene layer was deposited with the desired thickness. Finally, an electric field was applied through the co-axial cable to pole the transducer in the air. A schematic cross-section diagram of the transducer is illustrated in Figure 2.
The pulse–echo response was measured to determine the center frequency (fc) and bandwidth (BW) of the developed transducer. The transducer was immersed into distilled water and excited by a pulse/receiver (Panametrics 5900PR, Olympus, Tokyo, Japan) with 1 μJ energy, while the echo signal was acquired by a digital oscilloscope with 50 Ω damping (HP Infinium DSO-S204A, Keysight Technologies, Santa Rosa, CA, USA). The fast Fourier transform (FFT) was employed to transfer the time-domain echo signals to a frequency-domain spectrum. fc and BW can be determined as:
f c = f 1 + f 2 2 ,
B W = f 2 f 1 f c × 100 % ,
where f1 and f2 are lower and upper −6 dB frequencies, respectively. The effective electromechanical coupling coefficient (keff) was calculated as:
k e f f = 1 f r 2 f a 2 .
Two-way insertion loss (IL) was also measured by exciting a transducer through a function generator (AFG3251, Tektronix, Beaverton, AL, USA), where a burst of a 10-cycle sinusoidal wave with a frequency of fc was generated with 50 Ω coupling. The echo signal, reflected by a quartz plate, was recorded by a digital oscilloscope with 1 MΩ coupling. IL was then calculated as [24]:
I L = 20 l o g ( V 1 V 0 ) + 1.9   d B + 2.2 × 10 4 × f c 2 × 2 d ,
where V0 and V1 are peak-to-peak voltages of the emitted signal and received echo signal, respectively. A total of 1.9 dB represents the loss of ultrasound signal at the quartz surface, while 2.2 × 10−4 (dB/cm·MHz2) is the attenuation coefficient of ultrasound signal in water.
Imaging experiments were then conducted to evaluate the performance of 1-3 composite transducers, wherein lateral and axial resolutions were measured. Theoretically, the lateral resolution of ultrasound mainly depends on the dimensions of the transducer, while fc and BW determine the axial resolution as follows:
R l a t e r a l = λ F d ,
R a x i a l   = 0.5 c f c × B W ,
where λ and c are the acoustic wavelength and acoustic velocity in water, respectively, d is the diameter of the acoustic stack, and F is the focal length. For an unfocused transducer, its focal length is the near-field distance (N) as:
N = d 2 f c 4 c .
Four 10 μm-diameter tungsten wires were arranged at different heights in a 3D-printed holder. The wire phantom was immersed in distilled water for imaging acquired by the transducer. With a step motor (a step size of 5 μm) triggered by the pulse/receiver, the transducer was horizontally moved for scanning. A dynamic range of 26 dB was employed during imaging. Analyzing a cross-sectional ultrasound image of tungsten wire, the lateral and axial resolutions were assessed along specific directions using the full width at half maximum (FWHM).
The ex vivo sample was prepared with biological tissue and metal [25]. Raw chicken breast purchased from the local market was used as the tissue. A 1.6 mm-diameter metal needle was inserted into the chicken breast for evaluating the imaging capability of transducer.

3. Results and Discussion

3.1. Ceramic Properties

The XRD pattern and Rietveld refinement of 2.5%Sm-0.72PMN-0.28PT ceramics are shown in Figure 3. The XRD result of 2.5%Sm-0.72PMN-0.28PT ceramics shows a pure perovskite structure (PDF card No. 39-1487) without the involvement of a second phase. Both (110) and (002) peaks were single peaks with good symmetry, indicating the coexistence of rhombohedral and tetragonal phases [26]. This was attributed to the preparation process in which a mixture of two different crystalline phases was sintered to precisely control the phase proportion, thus forming a crystalline structure that resembles morphotropic phase boundaries (MPB). It is generally recognized that ceramics perform excellent electrical properties at the MPB [27]. The Rietveld refinement was performed to further investigate the crystalline structure of the ceramics. The results show that the 2.5%Sm-0.72PMN-0.28PT ceramics featured a mixture of the 40% tetragonal phase and the 60% rhombohedral phase, which was completely in agreement with the above analysis.
The SEM micrograph of 2.5%Sm-0.72PMN-0.28PT ceramic is shown in Figure 4. The inset shows the distribution of grain size that obeys the Gaussian-shaped distribution and gives an average grain size of 4.95 μm. All cross-sectional morphology showed homogeneous and dense ceramic grains. Moreover, there were no visible pores and impurities in the morphology, indicating the well-developed ceramic grains.
The temperature-dependent dielectric spectra of 2.5%Sm-0.72PMN-0.28PT ceramics at different frequencies are shown in Figure 5. It is obvious that the phenomena of dispersive phase transition (broadened dielectric and loss peaks) and frequency dispersion (dielectric and loss peaks move towards higher temperatures with increasing frequency) are the characteristics of relaxor ferroelectrics [28]. The Curie temperature Tc of 2.5%Sm-0.72PMN-0.28PT ceramics was 81 °C, representing the temperature at which the phase transition occurs from a ferroelectric phase to a paraelectric phase at 1 kHz. The corresponding dielectric peak value reached 30,192 at 1 kHz.
The temperature dependence of the dielectric constant of ceramics can be characterized by the Curie–Weiss law [29]:
ε = C T T 0
or quadratic law [30]:
1 ε = 1 ε m a x + ( T T c ) γ C
where C is the Curie–Weiss constant, T0 is the Curie–Weiss temperature, γ is the relaxation index, and C′ is a constant. The fitting results of the Curie–Weiss law and quadratic law for the temperature-dependent dielectric spectra at 1 kHz are shown in Figure 5b. According to the Curie–Weiss plot, the ferroelectric transition temperature (TC) and Curie–Weiss temperature (T0) were determined to be 102 °C and 81 °C, respectively. The deviations of TC and T0 manifest the typical characteristics of relaxor ferroelectric materials: broadened dielectric peaks rather than sharp ones. Regarding the quadratic law, when γ approaches 1, it indicates a normal ferroelectric, whereas an approach towards 2 indicates a relaxing ferroelectric. The fitting result yielded a relaxation index of 1.83, indicative of obvious relaxation characteristics.
The ferroelectric properties of 2.5%Sm-0.72PMN-0.28PT ceramics are shown in Figure 6. It can be seen that the P-E loop exhibits great symmetry with the remanent polarization strength of 23.21 μC/cm2, showing good ferroelectric properties. Moreover, the coercive field was 0.27 kV/mm, indicating that the polarization process could be completed at a relatively low electric field [31]. The overall piezoelectric and dielectric properties of 2.5%Sm-0.72PMN-0.28PT ceramic at room temperature are listed in Table 1.

3.2. Composite Properties

Figure 7 shows a photograph of a fabricated 1-3 composite taken with high-resolution optical microscopy. The kerf width was measured as 21 μm, and the ceramic pillar was measured as 72 μm. The ceramic volume fraction was calculated as 59.9%, which correlates well with the designed parameters.
Piezoelectric and dielectric properties of 1-3 composites are summarized in Table 2. Compared to other reported ceramic/epoxy 1-3 composites, d33 and εs of composite in this study were significantly high due to the excellent properties of Sm-PMN-PT ceramics, while maintaining the coupling coefficient. Thus, the 1-3 composites demonstrate great potential for the development of miniature ultrasound transducers with high sensitivity.

3.3. Simulation and Measurement of Transducer

Based on the high εs value, a miniature ultrasound transducer was designed with an aperture size of 0.84 mm × 0.84 mm. The design parameters of materials used for transducer development are listed in Table 3, while the simulated pulse–echo response and impedance spectra are shown in Figure 8. According to the simulation results, electrical matching could be achieved with a miniature aperture size of 1-3 composites. The simulated fc and −6 dB BW were 14.3 MHz and 79%, respectively.
The fabricated transducer prototype is shown in Figure 9. Figure 10a,b shows the measured pulse–echo response and impedance/phase spectra of the transducer, respectively. It was found that fc was 12.4 MHz and BW was ~70%. Furthermore, the electrical impedance at fc was close to the optimized value (~50 Ω). In addition, keff was calculated as 0.66.
Table 4 compares the performance of transducers fabricated using various 1-3 composites. Among the transducers with similar fc, the aperture size presented in this work was significantly small but with the optimized electrical impedance, which is attributed to the high εs of Sm-PMN-PT ceramics. Compared with the reported work as listed in Table 4, keff and BW of transducer incorporating Sm-PMN-PT based 1-3 composite was relatively high.
The sensitivity of the transducer was also measured, showing an IL value of −20.5 dB. As listed in Table 5, the composite transducer developed in this work exhibited high sensitivity when compared to other reported transducers with similar configurations, which is attributed to the outstanding piezoelectric performance of Sm-PMN-PT ceramics.
It was observed that the sensitivity of the transducer was affected by the miniature aperture, limiting the emitted acoustic pressure. The miniature ultrasound transducer exhibited lower emitted acoustic pressure compared to a transducer with a similar fc but a larger aperture size. Consequently, the received acoustic pressure of the echo signal was relatively weak, resulting in a high IL value. Additionally, miniature transducers, in general, tend to show a larger measured IL (magnitude) compared to simulation results based on the KLM model [37]. Despite these challenges, the sensitivity of the Sm-PMN-PT-based 1-3 composites transducer remained satisfactory, demonstrating comparable performance to miniature transducers fabricated using piezoelectric ceramics [38].
A scanned image of four 10 μm-diameter tungsten wires is shown in Figure 11a. The first tungsten wire was analyzed with the aforementioned method. Both the axial and lateral resolutions were 70 μm and 350 μm, respectively, which were analyzed from the first tungsten wire image with the corresponding FWHMs shown in Figure 11b,c. Both measured axial and lateral resolutions correlated well with theoretical values (85 μm; 317 μm). The chicken breast inserted with a metal needle, as shown in Figure 12a, was employed as the biological sample of ex vivo imaging. The image (Figure 12b) clearly shows the tissue structure of the chicken breast and the location of the metal needle. The needle was clearly displayed with significantly different signal intensity compared with surrounding tissues, demonstrating the in vitro imaging capability of a miniature transducer. Moreover, the transducer was configured as the prototype of a side-looking catheter. The aforementioned results indicate promising potential for its application in endoscopy for the gastrointestinal tract [39].

4. Conclusions

In this study, 2.5Sm-0.72PMN-0.28PT was synthesized for the development of 1-3 ceramic/epoxy composite material. With a ceramic volume fraction of 60%, the composite was successfully fabricated through the dice-and-fill method, showing a high piezoelectric constant and clamped dielectric constant. A miniature 12.4-MHz ultrasound transducer with an aperture size of 0.84 mm × 0.84 mm was fabricated using the developed composite. In comparison to ultrasound transducers fabricated using alternative piezoelectric 1-3 composites, the aperture size could be reduced to only 1/7 while maintaining an optimized electrical matching design. The achieved −6 dB bandwidth and insertion loss were found to be satisfactory when compared to other transducers with similar structures. Tungsten wire imaging was conducted to evaluate the axial and lateral resolutions, and ex vivo imaging was conducted to prove the capability of the miniature transducer for biomedical imaging applications.
This study has some limitations. Firstly, the imaging quality was suboptimal, possibly attributed to the low frequency of the transducer. Additionally, the experimental waveforms of pulse–echo response and bandwidth exhibited slight discrepancies compared to the simulation results. One potential explanation could be the width-to-height aspect ratio approaching the threshold value, where transverse vibration modes may impact the transducer performance.
The future work involves advancing transducer technology by incorporating higher frequencies (>30 MHz) through the utilization of Sm-PMN-PT composite materials, aiming to enhance imaging quality. However, achieving this goal requires the integration of picosecond laser cutting technology to produce narrower kerfs [40].
Concurrently, as ultrasound transducers with higher frequencies inherently demand smaller optimal apertures, there will arise a potential compromise in sensitivity. To address this challenge, exploring alternative methods such as alternating current polarization and conducting in-depth investigations into the composition and synthesis of piezoelectric ceramics could offer viable solutions to improve the overall transducer performance. Furthermore, the optimization of the matching layer is imperative, involving a thorough study of material characteristics and thickness [34]. This exploration has the potential to enable transducer operation away from transverse vibration mode, thereby maximizing ultrasound transmission efficiency.

Author Contributions

Conceptualization, K.-H.L. and J.-M.Z.; Methodology, J.-M.Z., W.G., R.-Q.L. and G.-C.B.; Data curation, J.-M.Z., G.-C.B., R.-Q.L. and F.Y.; Writing—original draft preparation, J.-M.Z. and G.-C.B.; Writing—review and editing, K.-H.L.; Supervision, Funding acquisition, K.-H.L. All authors have read and agreed to the published version of the manuscript.

Funding

The authors gratefully acknowledge support from the Hong Kong Research Impact Fund (R5029-19) and the Hong Kong General Research Fund (15220920).

Data Availability Statement

Data are contained within the article.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Chen, Y.; Lam, K.H.; Zhou, D.; Cheng, W.; Dai, J.; Luo, H.; Chan, H. High-frequency PIN-PMN-PT single crystal ultrasonic transducer for imaging applications. Appl. Phys. A 2012, 108, 987–991. [Google Scholar] [CrossRef]
  2. Wells, P.N.; Liang, H.-D. Medical ultrasound: Imaging of soft tissue strain and elasticity. J. R. Soc. Interface 2011, 8, 1521–1549. [Google Scholar] [CrossRef]
  3. Chan, V.; Perlas, A. Basics of ultrasound imaging. In Atlas of Ultrasound-Guided Procedures in Interventional Pain Management; Springer: New York, NY, USA, 2011; pp. 13–19. [Google Scholar]
  4. Haertling, G.H. Ferroelectric ceramics: History and technology. J. Am. Ceram. Soc. 1999, 82, 797–818. [Google Scholar] [CrossRef]
  5. Lu, X.M.; Proulx, T. Single Crystals vs. PZT Ceramics for Medical Ultrasound Applications. In Proceedings of the IEEE Ultrasonics Symposium, Rotterdam, The Netherlands, 18–21 September 2005; pp. 227–230. [Google Scholar]
  6. Edwards, G.; Chan, H.; Batten, A.; Lam, K.H.; Luo, H.; Scott, D. PMN-PT single-crystal transducer for non-destructive evaluation. Sens. Actuator A Phys. 2006, 132, 434–440. [Google Scholar] [CrossRef]
  7. Gururaja, T.; Schulze, W.A.; Cross, L.E.; Newnham, R.E.; Auld, B.A.; Wang, Y.J. Piezoelectric composite materials for ultrasonic transducer applications. Part I: Resonant modes of vibration of PZT rod-polymer composites. IEEE Trans. Sonics Ultrason 1985, 32, 481–498. [Google Scholar] [CrossRef]
  8. Newnham, R.E.; Skinner, D.P.; Cross, L.E. Connectivity and piezoelectric-pyroelectric composites. Mater. Res. Bull. 1978, 13, 525–536. [Google Scholar] [CrossRef]
  9. Smith, W.A.; Auld, B.A. Modeling 1-3 composite piezoelectrics: Thickness-mode oscillations. IEEE Trans. Ultrason Ferroelectr. Freq. Control 1991, 38, 40–47. [Google Scholar] [CrossRef]
  10. Ritter, T.; Geng, X.; Shung, K.K.; Lopath, P.D.; Park, S.-E.; Shrout, T.R. Single crystal PZN/PT-polymer composites for ultrasound transducer applications. IEEE Trans. Ultrason Ferroelectr. Freq. Control 2000, 47, 792–800. [Google Scholar] [CrossRef]
  11. Klicker, K.; Biggers, J.; Newnham, R. Composites of PZT and epoxy for hydrostatic transducer applications. J. Am. Ceram. Soc. 1981, 64, 5–9. [Google Scholar] [CrossRef]
  12. Chan, H.L.-W.; Li, K.; Choy, C.-L. Piezoelectric ceramic fibre/epoxy 1-3 composites for high-frequency ultrasonic transducer applications. Mater Scs. Eng. B 2003, 99, 29–35. [Google Scholar] [CrossRef]
  13. Wang, D.Y.; Li, K.; Chan, H.L.W. High frequency 1-3 composite transducer fabricated using sol–gel derived lead-free BNBT fibers. Sens. Actuator A Phys. 2004, 114, 1–6. [Google Scholar] [CrossRef]
  14. Chen, Y.; Mei, K.; Wong, C.-M.; Lin, D.; Chan, H.L.W.; Dai, J. Ultrasonic Transducer Fabricated Using Lead-Free BFO-BTO+Mn Piezoelectric 1-3 Composite. Actuators 2015, 4, 127–134. [Google Scholar] [CrossRef]
  15. Zheng, T.; Zhang, Y.; Ke, Q.; Wu, H.; Heng, L.W.; Xiao, D.; Zhu, J.; Pennycook, S.J.; Yao, K.; Wu, J. High-performance potassium sodium niobate piezoceramics for ultrasonic transducer. Nano Energy 2020, 70, 104559. [Google Scholar] [CrossRef]
  16. Ke, Q.; Liew, W.H.; Tao, H.; Wu, J.; Yao, K. KNNS-BNZH lead-free 1–3 piezoelectric composite for ultrasonic and photoacoustic imaging. IEEE Trans. Ultrason Ferroelectr. Freq. Control 2019, 66, 1395–1401. [Google Scholar] [CrossRef]
  17. Yan, P.; Qin, Y.; Xu, Z.; Han, F.; Wang, Y.; Wen, Z.; Zhang, Y.; Zhang, S. Highly transparent Eu-doped 0.72 PMN-0.28 PT ceramics with excellent piezoelectricity. Acs Appl. Mater Inter. 2021, 13, 54210–54216. [Google Scholar] [CrossRef]
  18. Li, C.; Xu, B.; Lin, D.; Zhang, S.; Bellaiche, L.; Shrout, T.R.; Li, F. Atomic-scale origin of ultrahigh piezoelectricity in samarium-doped PMN-PT ceramics. Phys. Rev. B 2020, 101, 140102. [Google Scholar] [CrossRef]
  19. Bao, G.-C.; Shi, D.-L.; Zhang, J.-M.; Yang, F.; Yang, G.; Li, K.; Fang, B.-J.; Lam, K.-H. Samarium-doped lead magnesium niobate-lead titanate ceramics fabricated by sintering the mixture of two different crystalline phases. Materials 2023, 16, 6781. [Google Scholar] [CrossRef]
  20. Savakus, H.; Klicker, K.; Newnham, R. PZT-epoxy piezoelectric transducers: A simplified fabrication procedure. Mater. Res. Bull. 1981, 16, 677–680. [Google Scholar] [CrossRef]
  21. Jiang, L.; Chen, R.; Xing, J.; Lu, G.; Li, R.; Jiang, Y.; Kirk Shung, K.; Zhu, J.; Zhou, Q. Fabrication of a (K,Na)NbO3-based lead-free 1–3 piezocomposite for high-sensitivity ultrasonic transducers application. J. Appl. Phys. 2019, 125, 214501. [Google Scholar] [CrossRef]
  22. Wong, C.; Chan, S.; Liu, R.; Zhang, J.; Wu, W.; Liang, Z.; Yau, H.; Wang, D.; Li, S.; Lam, K. 20-MHz phased array ultrasound transducer for in vivo ultrasound imaging of small animals. Ultrasonics 2022, 126, 106821. [Google Scholar] [CrossRef]
  23. Cannata, J.M.; Ritter, T.A.; Chen, W.-H.; Silverman, R.H.; Shung, K.K. Design of efficient, broadband single-element (20–80 MHz) ultrasonic transducers for medical imaging applications. IEEE Trans. Ultrason Ferroelectr. Freq. Control 2003, 50, 1548–1557. [Google Scholar] [CrossRef]
  24. Fan, D.; Niu, H.; Liu, K.; Sun, X.; Wang, H.; Shi, K.; Mo, W.; Jian, Z.; Wen, L.; Shen, M. Nb and Mn co-modified Na0.5Bi4.5Ti4O15 bismuth-layered ceramics for high-frequency transducer applications. Micromachines 2022, 13, 1246. [Google Scholar] [CrossRef]
  25. Chiu, C.T.; Kang, B.J.; Eliahoo, P.; Abraham, T.; Shung, K.K. Fabrication and characterization of a 20-MHz microlinear phased-array transducer for intervention guidance. IEEE Trans. Ultrason Ferroelectr. Freq. Control 2017, 64, 1261–1268. [Google Scholar] [CrossRef]
  26. Guo, Q.; Hou, L.; Li, F.; Xia, F.; Wang, P.; Hao, H.; Sun, H.; Liu, H.; Zhang, S. Investigation of dielectric and piezoelectric properties in aliovalent Eu3+-modified Pb(Mg1/3Nb2/3)O3-PbTiO3 ceramics. J. Am. Ceram. Soc. 2019, 102, 7428–7435. [Google Scholar] [CrossRef]
  27. Li, F.; Lin, D.; Chen, Z.; Cheng, Z.; Wang, J.; Li, C.; Xu, Z.; Huang, Q.; Liao, X.; Chen, L.-Q. Ultrahigh piezoelectricity in ferroelectric ceramics by design. Nat. Mater. 2018, 17, 349–354. [Google Scholar] [CrossRef] [PubMed]
  28. Yang, Y.; Sun, E.; Xu, Z.; Zheng, H.; Yang, B.; Zhang, R.; Cao, W. Elimination of pyrochlore phase in high-concentration Sm3+-doped PMN-PT piezoelectric ceramics by excessive MgO. J. Am. Ceram. Soc. 2022, 105, 4180–4190. [Google Scholar] [CrossRef]
  29. M’Peko, J.-C.; Peixoto, A.G.; Jiménez, E.; Gaggero-Sager, L.M. Electrical properties of Nb-doped PZT 65/35 ceramics: Influence of Nb and excess PbO. J. Electroceramics 2005, 15, 167–176. [Google Scholar] [CrossRef]
  30. Uchino, K.; Nomura, S. Critical exponents of the dielectric constants in diffused-phase-transition crystals. Ferroelectrics 1982, 44, 55–61. [Google Scholar] [CrossRef]
  31. Fang, Z.; Tian, X.; Zheng, F.; Jiang, X.; Ye, W.; Qin, Y.; Wang, X.; Zhang, Y. Enhanced piezoelectric properties of Sm3+-modified PMN-PT ceramics and their application in energy harvesting. Ceram. Int. 2022, 48, 7550–7556. [Google Scholar] [CrossRef]
  32. Zhang, Q. Ultrasonic Transducers for Acoustic Vortex Tweezing Application. Ph.D. Thesis, Hong Kong Polytechnic University, Hong Kong, China, 2020. [Google Scholar]
  33. Shen, Z.Y.; Li, J.F.; Chen, R.; Zhou, Q.; Shung, K.K. Microscale 1–3-type(Na,K)NbO3-based Pb-free piezocomposites for high-frequency ultrasonic transducer applications. J. Am. Ceram. Soc. 2011, 94, 1346–1349. [Google Scholar] [CrossRef] [PubMed]
  34. Li, Z.; Shao, W.; Zhu, X.; Lv, J.; Han, Z.; Cui, Y. Parylene coating for 13 MHz 1-3 composite transducer performance enhancement. Appl. Acoust. 2021, 174, 107696. [Google Scholar] [CrossRef]
  35. Jiang, L.; Chen, H.; Zeng, Y.; Tan, Z.; Wu, J.; Xing, J.; Zhu, J. Potassium sodium niobate-based lead-free high-frequency ultrasonic transducers for multifunctional acoustic tweezers. Acs Appl. Mater. Inter. 2022, 14, 30979–30990. [Google Scholar] [CrossRef] [PubMed]
  36. Jadidian, B.; Hagh, N.M.; Winder, A.A.; Safari, A. 25 MHz ultrasonic transducers with lead-free piezoceramic, 1–3 PZT fiber-epoxy composite, and PVDF polymer active elements. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2009, 56, 368–378. [Google Scholar] [CrossRef] [PubMed]
  37. Sun, P.; Zhou, Q.; Zhu, B.; Wu, D.; Hu, C.; Cannata, J.M.; Tian, J.; Han, P.; Wang, G.; Shung, K.K. Design and fabrication of PIN-PMN-PT single-crystal high-frequency ultrasound transducers. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2009, 56, 2760–2763. [Google Scholar] [PubMed]
  38. Zhang, Q.; Pang, X.; Zhang, Z.; Su, M.; Hong, J.; Zheng, H.; Qiu, W.; Lam, K.H. Miniature transducer using PNN-PZT-based ceramic for intravascular ultrasound. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2019, 66, 1102–1109. [Google Scholar] [CrossRef] [PubMed]
  39. Liu, J.-B.; Miller, L.S.; Bagley, D.H.; Goldberg, B.B. Endoluminal sonography of the genitourinary and gastrointestinal tracts. J. Ultrasound Med. 2002, 21, 323–337. [Google Scholar] [CrossRef]
  40. Lei, Z.; Xu, G.; Liu, J.; Liu, Y.; Ji, X. Micromachining of high-quality PMN-PT/epoxy 1-3 composite for high-frequency (>30 MHz) ultrasonic transducer applications. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2023, 70, 1563–1573. [Google Scholar] [CrossRef]
Figure 1. Dice-and-fill method for the fabrication of 1-3 composite [16].
Figure 1. Dice-and-fill method for the fabrication of 1-3 composite [16].
Jcs 08 00080 g001
Figure 2. Cross-sectional illustration of transducer.
Figure 2. Cross-sectional illustration of transducer.
Jcs 08 00080 g002
Figure 3. (a) XRD pattern and (b) Rietveld refinement of 2.5%Sm-0.72PMN-0.28PT ceramics.
Figure 3. (a) XRD pattern and (b) Rietveld refinement of 2.5%Sm-0.72PMN-0.28PT ceramics.
Jcs 08 00080 g003
Figure 4. SEM micrograph and (inset) grain size statistics of 2.5%Sm-0.72PMN-0.28PT ceramics.
Figure 4. SEM micrograph and (inset) grain size statistics of 2.5%Sm-0.72PMN-0.28PT ceramics.
Jcs 08 00080 g004
Figure 5. (a) Temperature-dependent dielectric spectra and (b) Curie–Weiss plot for the temperature-dependent dielectric spectra at 1 kHz (the inset shows the quadratic law fitting result for the temperature-dependent dielectric spectra at 1 kHz) of 2.5%Sm-0.72PMN-0.28PT ceramics.
Figure 5. (a) Temperature-dependent dielectric spectra and (b) Curie–Weiss plot for the temperature-dependent dielectric spectra at 1 kHz (the inset shows the quadratic law fitting result for the temperature-dependent dielectric spectra at 1 kHz) of 2.5%Sm-0.72PMN-0.28PT ceramics.
Jcs 08 00080 g005
Figure 6. P-E hysteresis loop of 2.5%Sm-0.72PMN-0.28PT ceramics.
Figure 6. P-E hysteresis loop of 2.5%Sm-0.72PMN-0.28PT ceramics.
Jcs 08 00080 g006
Figure 7. Optical image of Sm-PMN-PT 1-3 composite.
Figure 7. Optical image of Sm-PMN-PT 1-3 composite.
Jcs 08 00080 g007
Figure 8. Simulated (a) pulse–echo response and (b) electrical spectra of transducer incorporating Sm-PMN-PT 1-3 composite.
Figure 8. Simulated (a) pulse–echo response and (b) electrical spectra of transducer incorporating Sm-PMN-PT 1-3 composite.
Jcs 08 00080 g008
Figure 9. Photograph of fabricated miniature ultrasound transducer using Sm-PMN-PT 1-3 composite.
Figure 9. Photograph of fabricated miniature ultrasound transducer using Sm-PMN-PT 1-3 composite.
Jcs 08 00080 g009
Figure 10. Measured (a) pulse–echo response and (b) electrical impedance spectra of transducer incorporating Sm-PMN-PT 1-3 composite.
Figure 10. Measured (a) pulse–echo response and (b) electrical impedance spectra of transducer incorporating Sm-PMN-PT 1-3 composite.
Jcs 08 00080 g010
Figure 11. (a) Ultrasound image of tungsten wires with a dynamic range of 26 dB, and the measured (b) axial and (c) lateral resolutions of miniature transducer.
Figure 11. (a) Ultrasound image of tungsten wires with a dynamic range of 26 dB, and the measured (b) axial and (c) lateral resolutions of miniature transducer.
Jcs 08 00080 g011
Figure 12. (a) Photograph of a biological sample with a needle inserted, with a red dashed line denoting the scan line. (b) Ex vivo ultrasound B-scan image of cross-section of the needle (highlighted with a red oval) inserted into the sample.
Figure 12. (a) Photograph of a biological sample with a needle inserted, with a red dashed line denoting the scan line. (b) Ex vivo ultrasound B-scan image of cross-section of the needle (highlighted with a red oval) inserted into the sample.
Jcs 08 00080 g012
Table 1. Properties of 2.5%Sm-0.72PMN-0.28PT ceramics at room temperature.
Table 1. Properties of 2.5%Sm-0.72PMN-0.28PT ceramics at room temperature.
d33 (pC/N)ktε33/ε0ρ (kg/m3)c (m/s)Z (MRayl)
10250.5112,0067.9410032.4
Table 2. Properties comparison of reported 1-3 composite materials and this work.
Table 2. Properties comparison of reported 1-3 composite materials and this work.
Materiald33ε33/ε0ktϕ (%)
PZT-5H 1-3 composite [32]3905570.6860
Mn-doped BFO-BTO 1-3 composite [13]45/0.4660
NKLKT 1-3 composite [33]140//25
Fe-KNN-based 1-3 composite [21]3405130.5962
KNNS-BNZH 1-3 composite [15]350/0.7061
Sm-PMN-PT 1-3 composite (This work)65023500.6760
Table 3. Properties of 1-3 composite and passive materials.
Table 3. Properties of 1-3 composite and passive materials.
Materialt (μm)A (mm)Z (MRayl)Usage
Sm-PMN-PT Composite960.84 × 0.8419.4Piezoelectric layer
E-solder 30225000.84 × 0.845.6Backing layer
Parylene30N/A2.7Matching layer
Table 4. Performance comparison of ultrasound transducers fabricated using various 1-3 composites.
Table 4. Performance comparison of ultrasound transducers fabricated using various 1-3 composites.
Transducer Materialfc
(MHz)
BW
(%)
A (mm2)keffElectrical Impedance (Ω)
PZT-5H 1-3 composite [32]13.5/3.5 × 3.50.56~50
PZT-5H 1-3 composite [34]13.0593.0 × 1.5//
Fe-KNN 1-3 composite [21]16.083Ø 2.70.55~50
KNN 1-3 composite [35]23.075Ø 2.50.50~50
Sm-PMN-PT 1-3 composite
(This work)
12.4700.84 × 0.840.66~50
Table 5. Comparisons of matching layer material and sensitivity of ultrasound transducers fabricated using various 1-3 composites.
Table 5. Comparisons of matching layer material and sensitivity of ultrasound transducers fabricated using various 1-3 composites.
Transducer MaterialMatching LayerIL (dB)
PZT 1-3 composite [36]Silver epoxy−34.0
BNT 1-3 composite [13]N/A−34.8
BTO 1-3 composite [13]Al2O3/epoxy−33.5
KNN 1-3 composite [33]1st: Silver epoxy & 2nd: Polymer−25.1
KNNS-BNZH 1-3 composite [15]Epoxy−30.0
Sm-PMN-PT 1-3 composite (This work)Parylene−20.5
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Zhang, J.-M.; Bao, G.-C.; Gao, W.; Lin, R.-Q.; Yang, F.; Lam, K.-H. Miniature Ultrasound Transducer Incorporating Sm-PMN-PT 1-3 Composite. J. Compos. Sci. 2024, 8, 80. https://doi.org/10.3390/jcs8030080

AMA Style

Zhang J-M, Bao G-C, Gao W, Lin R-Q, Yang F, Lam K-H. Miniature Ultrasound Transducer Incorporating Sm-PMN-PT 1-3 Composite. Journal of Composites Science. 2024; 8(3):80. https://doi.org/10.3390/jcs8030080

Chicago/Turabian Style

Zhang, Jia-Ming, Guo-Cui Bao, Wen Gao, Ri-Qiang Lin, Fan Yang, and Kwok-Ho Lam. 2024. "Miniature Ultrasound Transducer Incorporating Sm-PMN-PT 1-3 Composite" Journal of Composites Science 8, no. 3: 80. https://doi.org/10.3390/jcs8030080

Article Metrics

Back to TopTop