Review

Applications of Smartphone-Based Aptasensor for Diverse Targets Detection

1
Academy of Medical Engineering and Translational Medicine, Tianjin University, Tianjin 300072, China
2
Tianjin Key Laboratory of Brain Science and Neuroengineering, Tianjin 300072, China
3
Department of Biomedical Engineering, College of Precision Instruments and Optoelectronics Engineering, Tianjin University, Tianjin 300072, China
*
Authors to whom correspondence should be addressed.
These authors contributed equally to this work.
Biosensors 2022, 12(7), 477; https://doi.org/10.3390/bios12070477
Submission received: 16 June 2022 / Revised: 27 June 2022 / Accepted: 27 June 2022 / Published: 30 June 2022
(This article belongs to the Special Issue Smartphone-Based Sensors for Biomedical Applications)

Abstract

:
Aptamers are a particular class of functional recognition ligands with high specificity and affinity to their targets. As the candidate recognition layer of biosensors, aptamers can be used to sense biomolecules. Aptasensors, aptamer-based biosensors, have been demonstrated to be specific, sensitive, and cost-effective. Furthermore, smartphone-based devices have shown their advantages in binding to aptasensors for point-of-care testing (POCT), which offers an immediate or spontaneous responding time for biological testing. This review describes smartphone-based aptasensors to detect various targets such as metal ions, nucleic acids, proteins, and cells. Additionally, the focus is also on aptasensors-related technologies and configurations.

1. Introduction

Biosensors can recognize the target analyte or a specific group of target molecules in complex samples [1]. A biosensor comprises three parts: molecular recognition elements (such as nucleic acids, receptors, antibodies, enzymes, organelles, tissues, microorganisms, and animal or plant cells, among others) [2], a transducer, and a signal processing unit. With the advancement of science and technology, recently developed biosensors are more portable, rapid, and user-friendly [3]. Hence, biosensors have been widely applied in the food industry [4,5], environmental monitoring [6,7], clinical monitoring [8,9], biomedicine [10,11], and other fields.
Given the rapid advances in personalized medicine and portable health care, point-of-care testing (POCT) based on biosensors has excellent potential. Advances in nanomaterials, molecular biotechnology, and optoelectronic integration have promoted the development of biosensing toward low-cost, easy-to-use, and on-site POCT [12]. Earlier POCT usually required high-cost and complex peripheral devices for analysis and evaluation, limiting its global and public health applications [13]. Therefore, POCT with low cost and easy operation urgently needs to be developed. Smartphones can solve these problems effectively solved by the integration of sensing, processing, and connectivity. Smartphones appeared in the consumer market in the late 1990s and quickly replaced the mobile phone market [14]. Typically, combining basic features such as voice and messaging with powerful computing technology, smartphones can support third-party applications, sensing, Internet access, and wireless connections with other devices [15]. Furthermore, smartphones can also equip with various sensors for various applications in multiple fields and providing substantial resources [16]. Additionally, smartphones with powerful computing and analysis functions can take high-resolution images, which have the characteristics of image storage, portability, and easy accessibility [17]. Hence, many scientific research institutions integrate smartphones into detection systems, such as colorimetric [18], electrochemical [19,20], optical [21,22], and other aspects of signal processing and image acquisition.
Aptamers are short oligonucleotide sequences obtained by in vitro screening, which can bind to multiple targets with high affinity and specificity, and have unique physical and chemical characteristics [23]. Aptamers with stable performance can specifically bind small substances such as ions [24,25], molecules [26], proteins [27,28], and cells [29]. Aptamers can be selected by the systematic evolution of ligands by exponential enrichment (SELEX), which is an iterative process that involves repetitive cycles of binding, partitioning, and amplification steps. The new SELEX strategies were developed to simplify procedures and improve aptamers species [30], such as microfluidic SELESX, in silico SELEX, bead-based SELEX, in vivo SELEX, and so on. As the most effective bioreceptor sensing component, the aptasensors become a powerful sensing instrument [31]. Aptamer has been an attractive tool for various applications compared with antibodies and enzymes due to its cost-effectiveness, strong specificity, suitable pretreatment, simple operation, and short detection time [32]. Recently, aptasensors have been applied to cell imaging [33,34], drug delivery [35,36], disease treatment [37,38], microbial detection [39,40], and other fields. Hence, a smartphone-based aptasensor will be an ideal tool for POCT and analysis to obtain a fast and precise response in various fields. This review summarizes the advancements in smartphone-based aptasensor, emphasizing various detection methods of aptasensors for the different target analytes.

2. Smartphone-Based Aptasensor

Smartphone-based aptasensors have been demonstrated to detect five types of targets: metal ions, small molecules, nucleic acids, proteins or glycoproteins, and bacteria. Furthermore, multiple-type targets can also be detected by a multi-channel system simultaneously.

2.1. Metal Ions Detection

Heavy metals are non-biodegradable and ubiquitously distributed in the biosphere with a highly toxic property [41]. Heavy metal ions discharged into the environment from household and industrial wastewater are likely to accumulate in living organisms through the alimentary canal, thereby harming human health [42,43]. Therefore, developing a fast and sensitive heavy metal ions detection method is very urgent and essential.
Many studies have reported the detection of heavy metal ions such as Cd2+ and Hg2+ based on their specific aptamers [44,45,46,47]. For Cd2+ detection, Wu et al. [44] presented a SELEX strategy for selecting Cd2+ aptamers and further adopted the selected aptamer as a recognition molecule to obtain the colorimetric detection of Cd2+. To improve the convenience and rapidity of detection, Gan et al. [48] introduced a smartphone-based colorimetric system to detect Cd2+. As shown in Figure 1a, in the presence of Cd2+, Cd-aptamers tend to combine with Cd2+ rather than gold nanoparticles(AuNPs), causing the dispersion of AuNPs. After adding NaCl solution, the AuNPs could aggregate to a different degree due to the number difference of bare AuNPs, making the solution changeable in color. While, in the absence of Cd2+ in the sample, aptamers immobilized on the AuNPs protected AuNPs from aggregation under NaCl conditions. Hence, the color of the last reaction solution changed from red to purple and finally to gray as the Cd2+ concentration increased. As shown in Figure 1b, a self-developed smartphone-based colorimetric system was used to capture the colorimetric image of a 96-well microplate, extract the pixel of the image, and then analyze the image data by the image processing algorithm. At last, the concentrations of Cd2+ in the sample were calculated by comparison with the colorimetric response standard curve of the Cd2+-spiked tap water sample. The linear range of Cd2+ was 2–20 μg/L, while the limit of detection (LOD) was 1.12 ug/L, showing a high selectivity and sensitivity of the smartphone-based colorimetric system.
Employing the same colorimetry principle based on the color change of AuNPs after adding NaCl solution, Xiao et al. [49] and Sajed et al. [50] proposed distinct smartphone-based aptasensors to detect Hg2+ or Cd2+. As depicted in Figure 1c, Xiao et al. [49] designed a microwell reader consisting of a resistor, a 520 nm light-emitting diode (LED), and a microwell to detect Hg2+. The microwell reader containing the sample was positioned on the top of the ambient light sensor. Following pressing the switch to form exciting light, the transmitted light intensity data were recorded and displayed on the light meter app, and the concentrations of Hg2+ were determined with the linear range of 1–32 ng/mL, and the LOD of 0.28 ng/mL in both Pearl River water and tap water samples.
Subsequently, Sajed [50] et al. used colorimetry at multiple points of the visible spectrum (470, 540, 640 nm). The group fabricated a gadget using three-dimensional printing technology applicable to any type of smartphone and integrated it with optical components, as revealed in Figure 1d. The gadget consisted of a thin film transistor liquid crystal display (TFTLED) board chamber, a focusing chamber of the camera, and two cuvette chambers containing the reference solution and the sample solution. A full-color TFTLCD was adopted as the light source, providing a uniform irradiance condition with any predefined pattern. Hg2+ concentration was estimated using multiple linear regression (MLR), a statistical technique for analyzing the connection between a dependent variable and multiple independent variables. In the MLR model, red, green, and blue (RGB) values extracted from images of samples by smartphone were applied as the input, while the concentration of Hg2+ in the sample was applied as the regression target. Light source enhancement, RGB analysis, and a novel image processing protocol make it possible to achieve an excellent level of sensitivity (1 nM).
Apart from NaCl solution, polydiene dimethyl ammonium chloride (PDDA) also leads to the aggregation of AuNPs, which can make the color of the response solution shift reported by Xu et al. [51]. The aptasensor combined with smartphone colorimetry to detect Cd2+ based on an aptamer competitive binding assay of PDDA and Cd2+. The directional light on the sealed box is reflected on the measured object by aluminum foil, as shown in Figure 1e. The images of samples were captured by smartphone, and a ColorAssist app read the RGB values. The linear correlation between Cd2+ concentration and red value (R-value) was plotted in Figure 1e, with a linear range of 1–400 ng/mL. The aptasensor of this work showed the potential for practical application as its results were in accordance with those from the atomic absorption spectrometer.

2.2. Small Molecules Detection

Many studies have reported small-molecule assays based on specific aptamers, such as cyanotoxin [52,53,54], kanamycin [55,56,57], and ATP [58]. In recent years, with an emphasis on POCT detection, many researchers have combined aptamer and smartphones to identify small molecules. Table 1 summarizes small-molecule detection using smartphone-based aptasensors. For example, Li et al. [59] developed an easy-to-use and miniaturized detector, using a fluorescent aptasensor array composed of specific aptamers and dyes for the multiplexed detection of four common cyanotoxins as environmentally hazardous substances. As shown in Figure 2a, a blue 462 nm laser illuminates single-stranded DNA (ssDNA) dyes without cyanotoxins, and the green fluorescence decreases upon adding analyte solutions, resulting from aptamer conformational changes. The authors also developed a smartphone-based fluorescence readout platform that integrated a microfluidic chip where the aptamer binding assays occurred with a 3D-printed attachment (Figure 2b). The emitted green fluorescence from the microfluidic chip was filtrated by a filter in the attachment and detected by the smartphone camera. In the raw images, the fluorescent intensity of the reaction was reversely proportional to the concentration of anatoxin-a (ATX). At the same time, the other four areas did not show any distinct changes in fluorescence signals, which certified the high specificity of aptamers (Figure 2c). As a result, the smartphone-based sensor platform realized accurate detection and measurement of four common cyanotoxin species, with the LOD lower than 3 nM, in parallel.
In addition to the change in intensity of single-color fluorescence, the analytes can be detected by the signal conversion of two-color fluorescence. Saurabh et al. [60] demonstrated ratiometric Förster resonance energy transfer (FRET) between fluorescent dye pairs on aptamers for rapid-response and sensitive kanamycin detection. As depicted in Figure 2d, the upper half of dye-labeled kanamycin DNA aptamer (KBA) is the kanamycin binding aptamer, while the bottom is double-stranded DNA used for surface modification of glass slide. When kanamycin was bound to KBAFRET, KBA displayed a lower level of FRET efficiency due to the unique placement of a donor-acceptor dye pair. The ratiometric FRET-based estimation could be implemented on a custom-made smartphone-based fluorescence setup with a LOD of 28 nM (Figure 2e).
Another energy transfer model superior to FRET is nanometal surface energy transfer (NSET), which has a higher energy transfer efficiency and a more extended energy transfer distance than dipole-dipole FRET [61,62]. Employing the NSET strategy, Nie et al. [63] developed an electrochemiluminescence (ECL) biosensor for point-of-care detection of ATP. As shown in Figure 2f, Fe3O4 NP@ZIF-8 was used as an effective catalyst, and its poor electron transfer capabilities could not influence the other reagents. MoS2 QDs produced NSET with AuNPs by the specific binding of cDNA and ATP aptamer, which decreased the ECL signal of MoS2 QDs. When ATP bound to the aptamer, the ECL-NSET system was deconstructed, resulting in the recovery of the ECL signal. Furthermore, a self-developed software for smartphones could capture changes in the ECL signal, which achieved point-of-care detection of ATP with a LOD of 0.015 nmol L−1 (Figure 2g).
Table 1. A summary of small-molecule detection using smartphone-based aptasensors.
Table 1. A summary of small-molecule detection using smartphone-based aptasensors.
Target AnalyteDetection ProbeDetection MethodLODReference
Cyanotoxinsaptamer−dyeFluorescence detection<3 nM[59]
Kanamycinaptamer-dyesFluorescence detection28 nM[60]
ATPaptamer-MoS2 QDElectrochemiluminescence0.015 nmol L−1[63]
ATPaptamer-Fe(CN)63−ColorimetryNR[64]
Mycotoxinaptamer−dyesFluorescence detection0.1 ng/mL[65]
ChloramphenicolAuNPs-aptamerColorimetry5.88 nM[66]
17-β-estradioldye-split aptamer fluorescent beadsFluorescence detection1 pg/mL (in spiked wastewater samples)[67]
Streptomycinaptamer−dyeFluorescence analysis94 nM[68]
StreptomycinsAuNPs-aptamerColorimetry12.3 nM[69]
IbuprofenAuNPs-aptamerColorimetry1.24 pg/mL (S-Ibu)
3.91 pg/mL (R-Ibu)
[70]
SulfadimethoxineAuNPs-aptamerColorimetry0.023 ppm[71]
CocaineAuNPs-aptamer-UCNPsLuminescence detection10 nM (in aqueous solution) 50 nM (in human saliva)[72]
Figure 2. (a) Detection theory of the fluorescent aptasensor. Reprinted with permission from Ref. [59], copyright 2019, American Chemical Society. (b) Schematic illustration of the smartphone−based fluorescent aptasensor. Reprinted with permission from Ref. [59], copyright 2019, American Chemical Society. (c) Raw fluorescence images in response to different concentrations of ATX. Reprinted with permission from Ref. [59], copyright 2019, American Chemical Society. (d) Schematic of FRET-based kanamycin aptasensor. Reprinted with permission from Ref. [60], copyright 2019, Royal Society of Chemistry. (e) Calibration curve for Kanamycin detection. Reprinted with permission from Ref. [60], copyright 2019, Royal Society of Chemistry. (f) Schematic of the ECL biosensor. Reprinted with permission from Ref. [63], copyright 2020, Elsevier. (g) Correlation between the ATP concentration and ECL intensity. Reprinted with permission from Ref. [63], copyright 2020, Elsevier.
Figure 2. (a) Detection theory of the fluorescent aptasensor. Reprinted with permission from Ref. [59], copyright 2019, American Chemical Society. (b) Schematic illustration of the smartphone−based fluorescent aptasensor. Reprinted with permission from Ref. [59], copyright 2019, American Chemical Society. (c) Raw fluorescence images in response to different concentrations of ATX. Reprinted with permission from Ref. [59], copyright 2019, American Chemical Society. (d) Schematic of FRET-based kanamycin aptasensor. Reprinted with permission from Ref. [60], copyright 2019, Royal Society of Chemistry. (e) Calibration curve for Kanamycin detection. Reprinted with permission from Ref. [60], copyright 2019, Royal Society of Chemistry. (f) Schematic of the ECL biosensor. Reprinted with permission from Ref. [63], copyright 2020, Elsevier. (g) Correlation between the ATP concentration and ECL intensity. Reprinted with permission from Ref. [63], copyright 2020, Elsevier.
Biosensors 12 00477 g002

2.3. Nucleic Acids Detection

The principle of detecting nucleic acids using aptamers is complementary base pair, and then specific aptamers can be designed by target sequences [20,73]. MicroRNAs (miRNAs), a kind of nucleic acid, can play as biomarkers of some tumors and cancers [74]. Hence, realizing portable and efficient on-site miRNAs detection is critical for clinical prevention and therapy of many cancer patients [75,76]. Aptasensors for miRNA detection offer higher sensitivity, lower costs, and require less sample than other approaches, opening new possibilities for detecting circulating miRNA in POCT [77,78]. Furthermore, smartphones with portability, high-resolution camera, a global positioning system (GPS), and Internet connection capability can provide healthcare diagnostics in resource-limited circumstances.
Consequently, the combination of smartphone and molecular diagnostics is particularly critical. Although the achievements in this research direction are relatively few, some progress has been made. For example, in the study presented by Shin Low et al. [20], aptamer appeared as a complementary strand of miR-21 binding with each other by complementary base pairs. Then, circulating miR-21 was detected in saliva using a smartphone-based biosensing system. The system consisted of a disposable electrochemical aptasensor, a self-designed circuit board, a smartphone with Bluetooth function, and a specially developed Android application (Figure 3a). Furthermore, the biosensing system demonstrated equivalent performance to a commercial electrochemical workstation, suitable selectivity, and an acceptable recovery rate of 96.2% to 107.2% in spiked artificial saliva.
Tian et al. [79] presented a smartphone-based detection platform for the detection of multiple miRNAs simultaneously. The quantitative detection platform included number-coded hydrogel microparticles, a microscope, and a smartphone with fluorescence image recognition software (Figure 3b). After the reaction shown in Figure 3c, hydrogel microparticles prepared by flow lithography technology displayed different fluorescence intensities. Images were achieved by a smartphone camera and then analyzed by an image processing procedure installed on the smartphone. In this work, the LOD of the detection platform reached the femtomole level in spiked healthy human serum samples, which demonstrated the POCT of tumor biomarkers.
In addition to playing as biomarkers, nucleic acids can constitute viruses. Type A influenza viruses are the most notorious and virulent human pathogens with high morbidity and mortality, which can cause an epidemic or even pandemic [80,81,82]. Lai et al. [83] developed a microfluidic system to screen a specific H1N1 virus-aptamer automatically and efficiently based on the SELEX technique. Furthermore, the selected aptamer completed specific and sensitive detection of the influenza A (H1N1) virus in biological samples by a magnetic-bead assay. Subsequently, Lee et al. [73], based on a specific aptamer, employed a built-in camera of a smartphone and a low-cost homemade setup (Figure 3d) to identify the H1N1 virus successfully with a LOD of 70 ng mL−1 in spiked phosphate-buffered saline (PBS) solution. Compared to sophisticated and bulky instruments, this portable device allows users to measure targets anytime and anywhere.

2.4. Proteins or Glycoproteins Detection

Carcinoembryonic antigen (CEA), a 180 kDa glycoprotein and a tumor biomarker, plays a significant role in predicting, prognosis, and tumor screening of many malignancies [84,85]. The CEA concentration of healthy individuals is less than 5 μg L−1, while patients containing cancer cells are more than 20 μg L−1 in serum [86]. Therefore, a rapid, reliable, and sensitive CEA detection method is of particular need to the public. Shu et al. [87] established an electrochemical aptasensor to detect CEA. Based on the signal amplification of AuNPs, the biosensor demonstrated excellent sensitivity and selectivity toward CEA, which indicated a high potential for practical application. Several years later, combining with a smartphone, Qiu et al. [88] developed a paper-based analytical device (PAD) for the fluorescence detection of CEA. As depicted in Figure 4a, the assay was conducted in a centrifuge tube where glucose molecules were gated into the mesoporous silica nanocontainers (MSNs) with the help of the CEA aptamer. Upon adding the analyte, CEA reacted with aptamers contributing to opening the pore and releasing glucose. The released glucose was catalyzed by glucose oxidase (GOD) immobilized on paper to produce hydrogen peroxide and gluconic acid, which quenched the fluorescence of CdTe/CdSe QDs and then measured by a smartphone and a commercial fluorospectrometer. Under optimal conditions, the system enabled sensitive detection of target CEA with a LOD of 6.7 pg mL−1 in spiked PBS solution, lower than most commercial enzyme-linked immunosorbent assay (ELISA) kits for CEA.
Enzymes, most of which are proteins, can be detected using aptasensor such as Plasmodium falciparum glutamate dehydrogenase (PfGDH) [89,90] and creatine kinase isoenzymes (CK-MB) [91]. To further increase the convenience of detection, Sanjay et al. [92] constructed a multi-channel, smartphone-based optic fiber platform for quantitative detection of PfGDH. As depicted in Figure 4b, the smartphone flashlight and camera were coupled to two ends of optic fiber by a four-channel smartphone accessory. U-bent optic fiber probes were prepared by gold-plating and PfGDH aptamer modification on the surface of plastic optic fibers. After the addition of simulated malaria samples, the images of the light spots were captured by a smartphone camera and analyzed by ImageJ. software. Consequently, the smartphone-based aptasensor exhibited a LOD of 264 pM in spiked buffer and 352 pM in spiked serum. For the CK-MB, Zhang et al. [93] proposed a dual sensitization smartphone colorimetric strategy to detect CK-MB gathering rolling circle amplification (RCA) coils and Au tetrahedra. The platform demonstrated reasonable specificity with a low LOD of 0.8 pM and performed well in spiked human serum samples.
The above reports were all for the detection of a single target, Mahmoud et al. [94] proposed a novel system for the simultaneous determination of two proteins, interleukin-6 (IL-6) and thrombin, using corresponding antibodies and aptamers, respectively. The system was composed of a 3D-printed smartphone imager (Figure 4c) and lateral flow strips modified with QD labels thrombin aptamers [95] and interleukin-6 antibodies. Through RGB channels, colored pictures of lateral flow strips were split to achieve optical multiplexing, which reduced turnaround time. With the LOD of 100 pM for IL-6 and 3 nM for thrombin in spiked buffer solutions, the smartphone-based system was suitable for duplex detection, particularly in low-resource conditions.
Figure 4. (a) Visual fluorescence detection mechanism of CEA biomarker based on paper-based analytical device (PAD). Reprinted with permission from Ref. [88], copyright 2017, American Chemical Society. (b) Schematic illustration of the smartphone-based optic fiber platform. Reprinted with permission from Ref [92], copyright 2020, RSC Publishing. (c) Smartphone dark box comprising (1) a bottom part fitted with a UV-LED light source, (2) a lid fitted with a replaceable smartphone adapter, and (3) a sample plate for seven lateral flow strips. Reprinted with permission from Ref [94], copyright 2021, Elsevier.
Figure 4. (a) Visual fluorescence detection mechanism of CEA biomarker based on paper-based analytical device (PAD). Reprinted with permission from Ref. [88], copyright 2017, American Chemical Society. (b) Schematic illustration of the smartphone-based optic fiber platform. Reprinted with permission from Ref [92], copyright 2020, RSC Publishing. (c) Smartphone dark box comprising (1) a bottom part fitted with a UV-LED light source, (2) a lid fitted with a replaceable smartphone adapter, and (3) a sample plate for seven lateral flow strips. Reprinted with permission from Ref [94], copyright 2021, Elsevier.
Biosensors 12 00477 g004

2.5. Bacteria Detection

Pathogenic bacteria cause various infectious diseases and kill millions of people each year [96]. Hence, to prevent and monitor diseases, it is crucial to detect the causative pathogens accurately and timely [97,98]. Specific recognition of Mycobacterium tuberculosis (M.tb) [99], Staphylococcus aureus (S. aureus) [100], and Salmonella Typhimurium (S. Typhimurium) [101] based on aptamers have been broadly reported. Recently, the detections of three above pathogenic bacteria based on smartphones and aptamers have been published for POCT detection. Li et al. [102] developed a reliable and cost-effective system based on enzyme-linked aptamers assay containing direct and indirect dot-blot assay (Figure 5A) for the detection of M.tb. Furthermore, an Android application performed colorimetric analysis on the photos captured by smartphone, and diagnostic reports were generated and transmitted through the Internet. This assay provided a lower limit of quantitation and a higher accuracy (Figure 5B) compared to traditional acid fast-staining assay, and greatly reduced the detection time compared to traditional bacterial culture.
For S. aureus, Shrivastava et al. [103] established a rapid, culture-free, smartphone-based fluorescence imaging platform for the quantitative detection of S. aureus based on the specific aptamer conjugated with fluorescent magnetic nanoparticles (FMNPs). As shown in Figure 5C, the S. aureus aptamer (Sap)-functionalized FMNPs were incubated with S. aureus, and then the fluorescence-tagged bacteria were spiked in complex, minimally processed samples. Next, the spiked samples were loaded into the bacterial detection cassette, followed by imaging using a smartphone fluorescence microscope comprised of a smartphone and a 3D-printed magnet-mounted holder. The top-right inset image represents the fluorescent intensity distribution of area-of-interest segmented from the captured image. By counting individual bacteria cells, the detection platform captured S. aureus cells efficiently from a peanut milk sample, with a minimum LOD of 10 cfu/mL within 10 min.
For S. Typhimurium, Man et al. [104] proposed a microfluidic colorimetric biosensor for on-site and rapid detection of Salmonella typhimurium based on aptamers and gold nanoparticles (AuNPs). Solutions mixing, reaction, and Salmonella detection occurred in the microfluidic chip (Figure 5D), where inlets (a, b, and d) were connected with a syringe pump to drive fluids, and a microvalve was used to control the flow direction of the fluids. Moreover, the colorimetric signals of the aggregated AuNPs were captured and processed by a smartphone imaging APP. The microfluidic aptasensor exhibited a low LOD of 6.0 × 101 cfu/mL in spiked AuNPs solutions and may further improve the sensitivity and enable high-throughput detection with immunomagnetic bead technique and an optimized microfluidic device.

2.6. Multiple-Type Targets Detection

Multiple-type target simultaneous analysis detects two or more targets such as metal ions, small molecules, nucleic acids, proteins, glycoproteins, and bacteria using the same system. Gu et al. [105] developed a dual-signal sensing system of naked-eye colorimetry and magneto-electrochemical detection for the detection of ATP [106] and Hg2+ [107] based on aptamers. The electrochemical sensing unit included a homemade magneto-electrochemical detection module, a custom-built hand-held workstation, and a smartphone. The electrode chip-equipped electrochemical detection module is inserted into the embedded connector and controlled by a smartphone linked to the workstation via Bluetooth (Figure 6A). In the electrochemical unit, the LOD of ATP and Hg2+ were 5.203 pmol L−1 and 1.597 pmol L−1 (S/N = 3), respectively. More significantly, the dual-signal platform is well extensible, enabling on-demand rapid quantification of other targets by substituting specific aptamers.
Additionally, multiple-type target detection is often implemented based on lateral flow assays (LFA). For instance, Jin et al. [108] designed a lateral flow strip coupled with a smartphone-based portable device for the simultaneous and specific detection of multiple targets with three different aptamers. As depicted in Figure 6B, aptamers immobilized with three-colored upconversion nanoparticles (UCNPs) were used as capture probes and complementary DNA sequences as test and control probes in lateral flow aptamer assay. In the absence of targets (Salmonella, ochratoxin A, mercury ion (SE [109], OTA [110], and Hg2+ [111])), aptamers attached to UCNP probes are hybridized with the corresponding complementary DNA. In the presence of targets, the aptamers tend to bind the corresponding targets, thereby liberating UCNPs and decreasing fluorescence signals. Hence, the concentration of each target is inversely related to the corresponding color intensity of the colored band. Then, the detection results were read by the smartphone equipped with a portable device. Overall, the detection platform was further operated successfully in tap water samples within 30 min, offering a potential approach for convenient, simultaneous, sensitive detection in various fields.
Gong et al. [112] developed a UCNP-LFA platform that involved a similar homemade reader (Figure 6C(a,b)) of the above literature [108], realizing the quantitative of SE [109], OTA [113], and Hg2+ [111] by using custom-designed software. From the detection results in Figure 6C(c,d), the UCNP-LFA platform has a precision and accuracy comparable to the gold-standard method, and the correlation coefficients between them are over 0.992. Subsequently, the universality of this platform was successfully demonstrated by the detection of other targets involved in public health. The platform with quantitative and highly sensitive detection capability meets the requirement of clinical testing, which makes it highly promising for environmental pollution monitoring and food safety testing.
Figure 6. (A) Schematic of the electrochemical sensing unit. Reprinted with permission from Ref. [105], copyright 2020, Elsevier. (B) Schematic of lateral flow aptamer assay for simultaneous three kinds of targets detection. Reprinted with permission from Ref. [108], copyright 2018, Elsevier. (C) Photograph (a) and schematic (b) of UCNP-LFA reader. (c,d) Detection results of SE, Hg2+, and OTA using the UCNP-LFA platform and the gold-standard method. Reprinted with permission from Ref. [112], copyright 2019, Elsevier. (D) Detection result readout process based on a smartphone-based portable gadget and custom-designed app. Reprinted with permission from Ref. [114], copyright 2021, Elsevier.
Figure 6. (A) Schematic of the electrochemical sensing unit. Reprinted with permission from Ref. [105], copyright 2020, Elsevier. (B) Schematic of lateral flow aptamer assay for simultaneous three kinds of targets detection. Reprinted with permission from Ref. [108], copyright 2018, Elsevier. (C) Photograph (a) and schematic (b) of UCNP-LFA reader. (c,d) Detection results of SE, Hg2+, and OTA using the UCNP-LFA platform and the gold-standard method. Reprinted with permission from Ref. [112], copyright 2019, Elsevier. (D) Detection result readout process based on a smartphone-based portable gadget and custom-designed app. Reprinted with permission from Ref. [114], copyright 2021, Elsevier.
Biosensors 12 00477 g006
Apart from DNA complementary sequences to the aptamers, antibodies are also acted as test and control probes in LFAs. Zhong et al. [114] constructed a multiple targets detection platform for rapid detection of Pb2+ [115], chloramphenicol [116], and β-lactoglobulin [117], based on aptamers and pregnancy test strips. In this LFA, aptamers were used as capture probes, while antibodies as the control probes. The pregnancy test strip shows two red lines when the target analytes are present and one red line when they are not. After the emergence of color, a portable gadget and a smartphone app were used to quantify the analytes (Figure 6D). Due to the maturity and ease of purchasing the pregnancy test strip, this detection platform based on LFA has a high potential for commercial application even in resource-limited settings.

3. Conclusions and Future Perspectives

Smartphone-based aptasensor is developing rapidly, which facilitates accurate, portable, rapid, and cost-effective detection. In this review, we discussed the detection methods of several substances using aptasensors on the platform of smartphones, including metal ions, small molecules, nucleic acids, proteins or glycoproteins, bacteria, and multiple-type targets. At present, most studies use aptamers as molecular probes of sensors to recognize and capture substances with their strong binding specificity. Smartphones, as data carriers, have been applied for colorimetric or electrochemical detection, in which data were sent to a smartphone via camera or Bluetooth for intelligent processing. Based on this method combined with smartphones, POCT can be used for medical, food, environment, and other fields, which makes the analysis and evaluation of detection results facility and convenient.
Aptasensors possess many advantages compared to other biosensors, such as enzyme biosensors and immunosensors. First, Aptamers can theoretically be selected in vitro for any particular target; thus, aptasensors have a more comprehensive range of applications in the detection field [118]. Second, high chemical stability and resistance to ambient temperature characterize the aptasensors, making them prosper in POCT. Third, the design of aptasensors is flexible due to the properties of aptamers with easy labeling and conformational changes upon target binding [119].
Smartphone-based aptasensor detection has broad prospects and great potential, but most aptasensor detection methods are still in the laboratory stage. Meanwhile, ELISA, electrochemiluminescence immunoassay (ECLIA), and other methods are primarily used in the clinical detection of antibodies. Aptamers have high specificity and trapping ability; however, we have not thoroughly investigated the properties of most aptamers [120]. In addition, the current research on aptamer is only to determine whether it can be used for substance detection, and the study of the modification and conformation change of aptamer is insufficient. The modification and assembly of aptasensors are also complicated, which takes time and effort to optimize the operating conditions. The cost of developing smartphone-based detection systems is very limited in remote areas with limited resources. This also requires professionals to develop, operate and expand the program, which gives users relative more difficulties in the process of participation [121].
Some designs need to be improved to develop smartphone-based aptasensors. For optical detection of smartphone-based aptasensor, further efforts are needed to reduce the volume and weight of the optical accessories by simplifying optical components and paths. Similarly, detectors of smartphone-based electrochemical aptasensor also need to be miniaturized. However, the fabrication of the integrated circuit chips is costly and complex. The built-in function modules of smartphones may replace them to eliminate hardware requirements [121].
In general, a smartphone-based aptasensor for diverse target detection has great advantages over the traditional method. Future efforts should focus on more advanced nanomaterials and molecular biotechnology to develop a more widely used and easily modified aptamer suitable for clinical application. In addition, it should also research photoelectric sensing systems and wireless communication equipment for smartphones to optimize the sensors. Furthermore, intelligent sensing technology will be combined with artificial intelligence, big data, the Internet of Things, cloud computing, and blockchain to become a more advantageous and promising detection technology. At last, costs should be reduced so that detection methods can be widely used in all regions to benefit all people. It is believed that smartphone-based aptasensors will play a more significant role in POCT in various fields after overcoming all challenges.

Author Contributions

Conceptualization, Y.L., B.H., C.S.T. and D.M.; methodology, Y.L. and B.H.; software, Y.L. and B.H.; investigation, Y.L. and B.H.; writing—original draft preparation, Y.L. and B.H.; writing—review and editing, Y.L. and B.H.; visualization, Y.L. and B.H.; supervision, C.S.T. and D.M.; project administration, C.S.T. and D.M.; funding acquisition, C.S.T. and D.M.; Y.L. and B.H. contributed equally to this work. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the National Natural Science Foundation of China (21904094 and 21950410516) and the Natural Science Foundation of Tianjin (20JCQNJC01100).

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Wang, X.D.; Wolfbeis, O.S. Fiber-Optic Chemical Sensors and Biosensors (2015–2019). Anal. Chem. 2020, 92, 397–430. [Google Scholar] [CrossRef] [PubMed]
  2. Mohankumar, P.; Ajayan, J.; Mohanraj, T.; Yasodharan, R. Recent developments in biosensors for healthcare and biomedical applications: A review. Measurement 2021, 167, 108293. [Google Scholar] [CrossRef]
  3. Huang, X.; Xu, D.; Chen, J.; Liu, J.; Li, Y.; Song, J.; Ma, X.; Guo, J. Smartphone-based analytical biosensors. Analyst 2018, 143, 5339–5351. [Google Scholar] [CrossRef] [PubMed]
  4. Meshram, B.D.; Agrawal, A.K.; Adil, S.; Ranvir, S.; Sande, K.K. Biosensor and its Application in Food and Dairy Industry: A Review. Int. J. Curr. Microbiol. Appl. Sci. 2018, 7, 3305–3324. [Google Scholar] [CrossRef]
  5. Turasan, H.; Kokini, J. Novel Nondestructive Biosensors for the Food Industry. Annu. Rev. Food Sci. Technol. 2021, 12, 539–566. [Google Scholar] [CrossRef]
  6. Hashem, A.; Hossain, M.A.M.; Marlinda, A.R.; Mamun, M.A.; Simarani, K.; Johan, M.R. Nanomaterials based electrochemical nucleic acid biosensors for environmental monitoring: A review. Appl. Surf. Sci. Adv. 2021, 4, 100064. [Google Scholar] [CrossRef]
  7. Bertani, P.; Lu, W. Cyanobacterial toxin biosensors for environmental monitoring and protection. Med. Nov. Technol. Devices 2021, 10, 100059. [Google Scholar] [CrossRef]
  8. Mavrikou, S.; Tsekouras, V.; Hatziagapiou, K.; Paradeisi, F.; Bakakos, P.; Michos, A.; Koutsoukou, A.; Konstantellou, E.; Lambrou, G.I.; Koniari, E.; et al. Clinical Application of the Novel Cell-Based Biosensor for the Ultra-Rapid Detection of the SARS-CoV-2 S1 Spike Protein Antigen: A Practical Approach. Biosensors 2021, 11, 224. [Google Scholar] [CrossRef]
  9. Song, Z.; Ma, Y.; Chen, M.; Ambrosi, A.; Ding, C.; Luo, X. Electrochemical Biosensor with Enhanced Antifouling Capability for COVID-19 Nucleic Acid Detection in Complex Biological Media. Anal. Chem. 2021, 93, 5963–5971. [Google Scholar] [CrossRef]
  10. Mauriz, E.; Lechuga, L.M. Plasmonic Biosensors for Single-Molecule Biomedical Analysis. Biosensors 2021, 11, 123. [Google Scholar] [CrossRef]
  11. Gheorghiu, M. A short review on cell-based biosensing: Challenges and breakthroughs in biomedical analysis. J. Biomed. Res. 2020, 35, 255–263. [Google Scholar] [CrossRef] [PubMed]
  12. Wang, Y.; Zhou, J.; Li, J. Construction of Plasmonic Nano-Biosensor-Based Devices for Point-of-Care Testing. Small Methods 2017, 1, 1700197. [Google Scholar] [CrossRef]
  13. Xu, X.; Akay, A.; Wei, H.; Wang, S.; Pingguan-Murphy, B.; Erlandsson, B.E.; Li, X.; Lee, W.; Hu, J.; Wang, L.; et al. Advances in Smartphone-Based Point-of-Care Diagnostics. Proc. IEEE 2015, 103, 236–247. [Google Scholar] [CrossRef]
  14. Cecere, G.; Corrocher, N.; Battaglia, R.D. Innovation and competition in the smartphone industry: Is there a dominant design? Telecommun. Policy 2015, 39, 162–175. [Google Scholar] [CrossRef]
  15. Bender, J.L.; Yue, R.Y.K.; To, M.J.; Deacken, L.; Jadad, A.R. A lot of action, but not in the right direction: Systematic review and content analysis of smartphone applications for the prevention, detection, and management of cancer. J. Med. Internet Res. 2013, 15, e287. [Google Scholar] [CrossRef]
  16. Krichen, M. Anomalies Detection Through Smartphone Sensors: A Review. IEEE Sens. J. 2021, 21, 7207–7217. [Google Scholar] [CrossRef]
  17. Mahato, K.; Chandra, P. Paper-based miniaturized immunosensor for naked eye ALP detection based on digital image colorimetry integrated with smartphone. Biosens. Bioelectron. 2019, 128, 9–16. [Google Scholar] [CrossRef]
  18. Aydindogan, E.; Ceylan, A.E.; Timur, S. Paper-based colorimetric spot test utilizing smartphone sensing for detection of biomarkers. Talanta 2020, 208, 120446. [Google Scholar] [CrossRef]
  19. Bai, Y.; Guo, Q.; Xiao, J.; Zheng, M.; Zhang, D.; Yang, J. An inkjet-printed smartphone-supported electrochemical biosensor system for reagentless point-of-care analyte detection. Sens. Actuators B Chem. 2021, 346, 130447. [Google Scholar] [CrossRef]
  20. Shin Low, S.; Pan, Y.; Ji, D.; Li, Y.; Lu, Y.; He, Y.; Chen, Q.; Liu, Q. Smartphone-based portable electrochemical biosensing system for detection of circulating microRNA-21 in saliva as a proof-of-concept. Sens. Actuators B Chem. 2020, 308. [Google Scholar] [CrossRef]
  21. Hou, L.; Qin, Y.; Li, J.; Qin, S.; Huang, Y.; Lin, T.; Guo, L.; Ye, F.; Zhao, S. A ratiometric multicolor fluorescence biosensor for visual detection of alkaline phosphatase activity via a smartphone. Biosens. Bioelectron. 2019, 143, 111605. [Google Scholar] [CrossRef] [PubMed]
  22. Shan, Y.; Wang, B.; Huang, H.; Jian, D.; Wu, X.; Xue, L.; Wang, S.; Liu, F. On-site quantitative Hg2+ measurements based on selective and sensitive fluorescence biosensor and miniaturized smartphone fluorescence microscope. Biosens. Bioelectron. 2019, 132, 238–247. [Google Scholar] [CrossRef] [PubMed]
  23. Bashir, A.; Yang, Q.; Wang, J.; Hoyer, S.; Chou, W.; McLean, C.; Davis, G.; Gong, Q.; Armstrong, Z.; Jang, J.; et al. Machine learning guided aptamer refinement and discovery. Nat. Commun. 2021, 12, 2366. [Google Scholar] [CrossRef]
  24. Guo, W.; Zhang, C.; Ma, T.; Liu, X.; Chen, Z.; Li, S.; Deng, Y. Advances in aptamer screening and aptasensors’ detection of heavy metal ions. J. Nanobiotechnology 2021, 19, 166. [Google Scholar] [CrossRef] [PubMed]
  25. Rahnama, S.; Shariati, S.; Divsar, F. Selective aptamer conjugation to silver-coated magnetite nanoparticles for magnetic solid-phase extraction of trace amounts of Pb2+ ions. RSC Adv. 2021, 11, 4971–4982. [Google Scholar] [CrossRef] [PubMed]
  26. Huang, H.; Li, J.; Pan, S.; Wang, H.; Liang, A.; Jiang, Z. A novel small molecular liquid crystal catalytic amplification-nanogold SPR aptamer absorption assay for trace oxytetracycline. Talanta 2021, 233, 122528. [Google Scholar] [CrossRef]
  27. Li, H.; Xing, S.; Xu, J.; He, Y.; Lai, Y.; Wang, Y.; Zhang, G.; Guo, S.; Deng, M.; Zeng, M.; et al. Aptamer-based CRISPR/Cas12a assay for the ultrasensitive detection of extracellular vesicle proteins. Talanta 2021, 221, 121670. [Google Scholar] [CrossRef]
  28. Li, S.; Liu, X.; Liu, S.; Guo, M.; Liu, C.; Pei, M. Fluorescence sensing strategy based on aptamer recognition and mismatched catalytic hairpin assembly for highly sensitive detection of alpha-fetoprotein. Anal. Chim. Acta 2021, 1141, 21–27. [Google Scholar] [CrossRef]
  29. Zhang, X.H.; Wang, W.; Chen, X. Selection and identification of an ssDNA aptamer to NB4 cell. J. Clin. Lab. Anal. 2021, 35, e23718. [Google Scholar] [CrossRef]
  30. Lyu, C.; Khan, I.M.; Wang, Z. Capture-SELEX for aptamer selection: A short review. Talanta 2021, 229, 122274. [Google Scholar] [CrossRef]
  31. Nur, Y.; Gaffar, S.; Hartati, Y.W.; Subroto, T. Applications of electrochemical biosensor of aptamers-based (APTASENSOR) for the detection of leukemia biomarker. Sens. Bio-Sens. Res. 2021, 32, 100416. [Google Scholar] [CrossRef]
  32. Khoshbin, Z.; Davoodian, N.; Taghdisi, S.M.; Abnous, K. Metal organic frameworks as advanced functional materials for aptasensor design. Spectrochim. Acta Part A: Mol. Biomol. Spectrosc. 2022, 276, 121251. [Google Scholar] [CrossRef] [PubMed]
  33. Mo, X.; Linjie, G.; Shu, X.; Shuting, C.; Ziwei, Z.; Kaiyi, L.; Jiang, L.; Shihua, L.; Yu, Z.; Lianhui, W. Cell imaging with multi-color DNA framework probes. Chem. Commun. 2021, 86, 11318–11321. [Google Scholar]
  34. Gu, Y.; Huang, L.J.; Zhao, W.; Zhang, T.T.; Cui, M.R.; Yang, X.J.; Zhao, X.L.; Chen, H.Y.; Xu, J.J. Living-Cell MicroRNA Imaging with Self-Assembling Fragments of Fluorescent Protein-Mimic RNA Aptamer. ACS Sens 2021, 6, 2339–2347. [Google Scholar] [CrossRef] [PubMed]
  35. Sathiyaseelan, A.; Saravanakumar, K.; Mariadoss, A.V.A.; Wang, M.-H. pH-controlled nucleolin targeted release of dual drug from chitosan-gold based aptamer functionalized nano drug delivery system for improved glioblastoma treatment. Carbohydr. Polym. 2021, 262, 117907. [Google Scholar] [CrossRef] [PubMed]
  36. Song, X.; Yan, T.; Tian, F.; Li, F.; Ren, L.; Li, Q.; Zhang, S. Aptamer Functionalized Upconversion Nanotheranostic Agent With Nuclear Targeting as the Highly Localized Drug-Delivery System of Doxorubicin. Front. Bioeng. Biotechnol. 2021, 9, 639487. [Google Scholar] [CrossRef] [PubMed]
  37. Scott, T.P.; Nel, L.H. Rabies Prophylactic and Treatment Options: An In Vitro Study of siRNA- and Aptamer-Based Therapeutics. Viruses 2021, 13, 881. [Google Scholar] [CrossRef]
  38. Wan, Q.; Liu, X.; Zu, Y. Oligonucleotide aptamers for pathogen detection and infectious disease control. Theranostics 2021, 11, 9133–9161. [Google Scholar] [CrossRef]
  39. Vishwakarma, A.; Lal, R.; Ramya, M. Aptamer-based approaches for the detection of waterborne pathogens. Int. Microbiol. 2021, 24, 125–140. [Google Scholar] [CrossRef]
  40. Guo, K.; Song, Z.; Wang, G.; Tang, C. Detecting Redox Potentials Using Porous Boron Nitride/ATP-DNA Aptamer/Methylene Blue Biosensor to Monitor Microbial Activities. Micromachines 2022, 13, 83. [Google Scholar] [CrossRef]
  41. Gumpu, M.B.; Sethuraman, S.; Krishnan, U.M.; Rayappan, J.B.B. A review on detection of heavy metal ions in water—An electrochemical approach. Sens. Actuators B Chem. 2015, 213, 515–533. [Google Scholar] [CrossRef]
  42. Rajaganapathy, V.; Xavier, F.; Sreekumar, D.; Mandal, P.K. Heavy metal contamination in soil, water and fodder and their presence in livestock and products: A review. J. Environ. Sci. Technol. 2011, 4, 234–249. [Google Scholar] [CrossRef] [Green Version]
  43. Chunhabundit, R. Cadmium Exposure and Potential Health Risk from Foods in Contaminated Area, Thailand. Toxicol. Res. 2016, 32, 65–72. [Google Scholar] [CrossRef]
  44. Wu, Y.; Zhan, S.; Wang, L.; Zhou, P. Selection of a DNA aptamer for cadmium detection based on cationic polymer mediated aggregation of gold nanoparticles. Analyst 2014, 139, 1550–1561. [Google Scholar] [CrossRef] [PubMed]
  45. Li, J.; Sun, M.; Wei, X.; Zhang, L.; Zhang, Y. An electrochemical aptamer biosensor based on "gate-controlled" effect using beta-cyclodextrin for ultra-sensitive detection of trace mercury. Biosens. Bioelectron. 2015, 74, 423–426. [Google Scholar] [CrossRef] [PubMed]
  46. Li, L.; Li, B.; Qi, Y.; Jin, Y. Label-free aptamer-based colorimetric detection of mercury ions in aqueous media using unmodified gold nanoparticles as colorimetric probe. Anal. Bioanal. Chem. 2009, 393, 2051–2057. [Google Scholar] [CrossRef]
  47. Zhu, Y.F.; Wang, Y.S.; Zhou, B.; Yu, J.H.; Peng, L.L.; Huang, Y.Q.; Li, X.J.; Chen, S.H.; Tang, X.; Wang, X.F. A multifunctional fluorescent aptamer probe for highly sensitive and selective detection of cadmium(II). Anal. Bioanal Chem. 2017, 409, 4951–4958. [Google Scholar] [CrossRef]
  48. Gan, Y.; Liang, T.; Hu, Q.; Zhong, L.; Wang, X.; Wan, H.; Wang, P. In-situ detection of cadmium with aptamer functionalized gold nanoparticles based on smartphone-based colorimetric system. Talanta 2020, 208, 120231. [Google Scholar] [CrossRef]
  49. Xiao, W.; Xiao, M.; Fu, Q.; Yu, S.; Shen, H.; Bian, H.; Tang, Y. A Portable Smart-Phone Readout Device for the Detection of Mercury Contamination Based on an Aptamer-Assay Nanosensor. Senser 2016, 16, 1871. [Google Scholar] [CrossRef]
  50. Sajed, S.; Arefi, F.; Kolahdouz, M.; Sadeghi, M.A. Improving sensitivity of mercury detection using learning based smartphone colorimetry. Sens. Actuators B Chem. 2019, 298, 126942. [Google Scholar] [CrossRef]
  51. Xu, L.; Liang, J.; Wang, Y.; Ren, S.; Wu, J.; Zhou, H.; Gao, Z. Highly Selective, Aptamer-Based, Ultrasensitive Nanogold Colorimetric Smartphone Readout for Detection of Cd(II). Molecules 2019, 24, 2745. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  52. Li, X.; Cheng, R.; Shi, H.; Tang, B.; Xiao, H.; Zhao, G. A simple highly sensitive and selective aptamer-based colorimetric sensor for environmental toxins microcystin-LR in water samples. J. Hazard Mater 2016, 304, 474–480. [Google Scholar] [CrossRef] [PubMed]
  53. Lin, Z.; Huang, H.; Xu, Y.; Gao, X.; Qiu, B.; Chen, X.; Chen, G. Determination of microcystin-LR in water by a label-free aptamer based electrochemical impedance biosensor. Talanta 2013, 103, 371–374. [Google Scholar] [CrossRef] [PubMed]
  54. Wang, F.; Liu, S.; Lin, M.; Chen, X.; Lin, S.; Du, X.; Li, H.; Ye, H.; Qiu, B.; Lin, Z.; et al. Colorimetric detection of microcystin-LR based on disassembly of orient-aggregated gold nanoparticle dimers. Biosens. Bioelectron. 2015, 68, 475–480. [Google Scholar] [CrossRef]
  55. Bai, X.; Hou, H.; Zhang, B.; Tang, J. Label-free detection of kanamycin using aptamer-based cantilever array sensor. Biosens. Bioelectron. 2014, 56, 112–116. [Google Scholar] [CrossRef]
  56. Song, K.M.; Cho, M.; Jo, H.; Min, K.; Jeon, S.H.; Kim, T.; Han, M.S.; Ku, J.K.; Ban, C. Gold nanoparticle-based colorimetric detection of kanamycin using a DNA aptamer. Anal. Biochem. 2011, 415, 175–181. [Google Scholar] [CrossRef]
  57. Zhu, Y.; Chandra, P.; Song, K.M.; Ban, C.; Shim, Y.B. Label-free detection of kanamycin based on the aptamer-functionalized conducting polymer/gold nanocomposite. Biosens. Bioelectron. 2012, 36, 29–34. [Google Scholar] [CrossRef]
  58. Wang, J.; Wang, L.; Liu, X.; Liang, Z.; Song, S.; Li, W.; Li, G.; Fan, C. A Gold Nanoparticle-Based Aptamer Target Binding Readout for ATP Assay. Adv. Mater. 2007, 19, 3943–3946. [Google Scholar] [CrossRef]
  59. Li, Z.; Zhang, S.; Yu, T.; Dai, Z.; Wei, Q. Aptamer-Based Fluorescent Sensor Array for Multiplexed Detection of Cyanotoxins on a Smartphone. Anal. Chem. 2019, 91, 10448–10457. [Google Scholar] [CrossRef]
  60. Umrao, S.; Jain, V.; Chakraborty, B.; Roy, R. Smartphone-based kanamycin sensing with ratiometric FRET. RSC Adv. 2019, 9, 6143–6151. [Google Scholar] [CrossRef] [Green Version]
  61. Chen, C.; Midelet, C.; Bhuckory, S.; Hildebrandt, N.; Werts, M.H.V. Nanosurface Energy Transfer from Long-Lifetime Terbium Donors to Gold Nanoparticles. J. Phys. Chem. C 2018, 122, 17566–17574. [Google Scholar] [CrossRef] [Green Version]
  62. Halder, S.; Aggrawal, R.; Jana, S.; Saha, S.K. Binding interactions of cationic gemini surfactants with gold nanoparticles-conjugated bovine serum albumin: A FRET/NSET, spectroscopic, and docking study. J. Photochem. Photobiol. B 2021, 225, 112351. [Google Scholar] [CrossRef] [PubMed]
  63. Nie, Y.; Liu, Y.; Zhang, Q.; Zhang, F.; Ma, Q.; Su, X. Fe3O4 NP@ZIF-8/MoS2 QD-based electrochemiluminescence with nanosurface energy transfer strategy for point-of-care determination of ATP. Anal. Chim. Acta 2020, 1127, 190–197. [Google Scholar] [CrossRef] [PubMed]
  64. Zhang, X.; Lazenby, R.A.; Wu, Y.; White, R.J. Electrochromic, Closed-Bipolar Electrodes Employing Aptamer-Based Recognition for Direct Colorimetric Sensing Visualization. Anal. Chem. 2019, 91, 11467–11473. [Google Scholar] [CrossRef]
  65. Ji, W.; Zhang, Z.; Tian, Y.; Yang, Z.; Cao, Z.; Zhang, L.; Qi, Y.; Chang, J.; Zhang, S.; Wang, H. Shape Coding Microhydrogel for a Real-Time Mycotoxin Detection System Based on Smartphones. ACS Appl. Mater. Interfaces 2019, 11, 8584–8590. [Google Scholar] [CrossRef]
  66. Wu, Y.Y.; Liu, B.W.; Huang, P.; Wu, F.Y. A novel colorimetric aptasensor for detection of chloramphenicol based on lanthanum ion-assisted gold nanoparticle aggregation and smartphone imaging. Anal. Bioanal. Chem. 2019, 411, 7511–7518. [Google Scholar] [CrossRef]
  67. Lee, W.I.; Shrivastava, S.; Duy, L.T.; Yeong Kim, B.; Son, Y.M.; Lee, N.E. A smartphone imaging-based label-free and dual-wavelength fluorescent biosensor with high sensitivity and accuracy. Biosens. Bioelectron. 2017, 94, 643–650. [Google Scholar] [CrossRef]
  68. Lin, B.; Yu, Y.; Cao, Y.; Guo, M.; Zhu, D.; Dai, J.; Zheng, M. Point-of-care testing for streptomycin based on aptamer recognizing and digital image colorimetry by smartphone. Biosens. Bioelectron. 2018, 100, 482–489. [Google Scholar] [CrossRef]
  69. Liu, Z.; Zhang, Y.; Xu, S.; Zhang, H.; Tan, Y.; Ma, C.; Song, R.; Jiang, L.; Yi, C. A 3D printed smartphone optosensing platform for point-of-need food safety inspection. Anal. Chim Acta 2017, 966, 81–89. [Google Scholar] [CrossRef]
  70. Ping, J.; He, Z.; Liu, J.; Xie, X. Smartphone-based colorimetric chiral recognition of ibuprofen using aptamers-capped gold nanoparticles. Electrophoresis 2018, 39, 486–495. [Google Scholar] [CrossRef]
  71. Zhang, X.; Wang, L.; Li, X.; Li, X. AuNP aggregation-induced quantitative colorimetric aptasensing of sulfadimethoxine with a smartphone. Chin. Chem. Lett. 2021, 33, 3078–3082. [Google Scholar] [CrossRef]
  72. He, M.; Li, Z.; Ge, Y.; Liu, Z. Portable Upconversion Nanoparticles-Based Paper Device for Field Testing of Drug Abuse. Anal. Chem. 2016, 88, 1530–1534. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  73. Lee, N.; Wang, C.; Park, J. User-friendly point-of-care detection of influenza A (H1N1) virus using light guide in three-dimensional photonic crystal. RSC Adv. 2018, 8, 22991–22997. [Google Scholar] [CrossRef] [Green Version]
  74. Heneghan, H.M.; Miller, N.; Kerin, M.J. MiRNAs as biomarkers and therapeutic targets in cancer. Curr. Opin. Pharm. 2010, 10, 543–550. [Google Scholar] [CrossRef] [PubMed]
  75. Zhou, M.; Teng, X.; Li, Y.; Deng, R.; Li, J. Cascade Transcription Amplification of RNA Aptamer for Ultrasensitive MicroRNA Detection. Anal. Chem. 2019, 91, 5295–5302. [Google Scholar] [CrossRef] [PubMed]
  76. Tang, X.; Deng, R.; Sun, Y.; Ren, X.; Zhou, M.; Li, J. Amplified Tandem Spinach-Based Aptamer Transcription Enables Low Background miRNA Detection. Anal. Chem. 2018, 90, 10001–10008. [Google Scholar] [CrossRef]
  77. Ye, J.; Xu, M.; Tian, X.; Cai, S.; Zeng, S. Research advances in the detection of miRNA. J. Pharm. Anal. 2019, 9, 217–226. [Google Scholar] [CrossRef]
  78. Ciui, B.; Jambrec, D.; Sandulescu, R.; Cristea, C. Bioelectrochemistry for miRNA detection. Curr. Opin. Electrochem. 2017, 5, 183–192. [Google Scholar] [CrossRef]
  79. Tian, Y.; Zhang, L.; Wang, H.; Ji, W.; Zhang, Z.; Zhang, Y.; Yang, Z.; Cao, Z.; Zhang, S.; Chang, J. Intelligent Detection Platform for Simultaneous Detection of Multiple MiRNAs Based on Smartphone. ACS Sens. 2019, 4, 1873–1880. [Google Scholar] [CrossRef]
  80. Tseng, Y.T.; Wang, C.H.; Chang, C.P.; Lee, G.B. Integrated microfluidic system for rapid detection of influenza H1N1 virus using a sandwich-based aptamer assay. Biosens. Bioelectron. 2016, 82, 105–111. [Google Scholar] [CrossRef]
  81. Bai, C.; Lu, Z.; Jiang, H.; Yang, Z.; Liu, X.; Ding, H.; Li, H.; Dong, J.; Huang, A.; Fang, T.; et al. Aptamer selection and application in multivalent binding-based electrical impedance detection of inactivated H1N1 virus. Biosens. Bioelectron. 2018, 110, 162–167. [Google Scholar] [CrossRef] [PubMed]
  82. Bhardwaj, J.; Chaudhary, N.; Kim, H.; Jang, J. Subtyping of influenza A H1N1 virus using a label-free electrochemical biosensor based on the DNA aptamer targeting the stem region of HA protein. Anal. Chim. Acta 2019, 1064, 94–103. [Google Scholar] [CrossRef] [PubMed]
  83. Lai, H.C.; Wang, C.H.; Liou, T.M.; Lee, G.B. Influenza A virus-specific aptamers screened by using an integrated microfluidic system. Lab. Chip. 2014, 14, 2002–2013. [Google Scholar] [CrossRef] [PubMed]
  84. Hasanzadeh, M.; Shadjou, N.; Lin, Y.; De La Guardia, M. Nanomaterials for use in immunosensing of carcinoembryonic antigen (CEA): Recent advances. TrAC Trends Anal. Chem. 2017, 86, 185–205. [Google Scholar] [CrossRef]
  85. Miao, H.; Wang, L.; Zhuo, Y.; Zhou, Z.; Yang, X. Label-free fluorimetric detection of CEA using carbon dots derived from tomato juice. Biosens. Bioelectron. 2016, 86, 83–89. [Google Scholar] [CrossRef]
  86. Guo, C.; Su, F.; Song, Y.; Hu, B.; Wang, M.; He, L.; Peng, D.; Zhang, Z. Aptamer-Templated Silver Nanoclusters Embedded in Zirconium Metal-Organic Framework for Bifunctional Electrochemical and SPR Aptasensors toward Carcinoembryonic Antigen. ACS Appl. Mater. Interfaces 2017, 9, 41188–41199. [Google Scholar] [CrossRef]
  87. Shu, H.; Wen, W.; Xiong, H.; Zhang, X.; Wang, S. Novel electrochemical aptamer biosensor based on gold nanoparticles signal amplification for the detection of carcinoembryonic antigen. Electrochem. Commun. 2013, 37, 15–19. [Google Scholar] [CrossRef]
  88. Qiu, Z.; Shu, J.; Tang, D. Bioresponsive Release System for Visual Fluorescence Detection of Carcinoembryonic Antigen from Mesoporous Silica Nanocontainers Mediated Optical Color on Quantum Dot-Enzyme-Impregnated Paper. Anal. Chem. 2017, 89, 5152–5160. [Google Scholar] [CrossRef]
  89. Singh, N.K.; Chakma, B.; Jain, P.; Goswami, P. Protein-Induced Fluorescence Enhancement Based Detection of Plasmodium falciparum Glutamate Dehydrogenase Using Carbon Dot Coupled Specific Aptamer. ACS Comb. Sci. 2018, 20, 350–357. [Google Scholar] [CrossRef]
  90. Singh, N.K.; Thungon, P.D.; Estrela, P.; Goswami, P. Development of an aptamer-based field effect transistor biosensor for quantitative detection of Plasmodium falciparum glutamate dehydrogenase in serum samples. Biosens. Bioelectron. 2019, 123, 30–35. [Google Scholar] [CrossRef]
  91. Zhang, J.; Lv, X.; Feng, W.; Li, X.; Li, K.; Deng, Y. Aptamer-based fluorometric lateral flow assay for creatine kinase MB. Microchim. Acta 2018, 185, 364. [Google Scholar] [CrossRef] [PubMed]
  92. Sanjay, M.; Singh, N.K.; Ngashangva, L.; Goswami, P. A smartphone-based fiber-optic aptasensor for label-free detection of Plasmodium falciparum glutamate dehydrogenase. Anal. Methods 2020, 12, 1333–1341. [Google Scholar] [CrossRef]
  93. Zhang, J.; Wang, Y.; Zhao, X.; Chen, M.; Peng, Y.; Bai, J.; Li, S.; Han, D.; Ren, S.; Qin, K.; et al. Dual Sensitization Smartphone Colorimetric Strategy Based on RCA Coils Gathering Au Tetrahedra and Its Application in the Detection of CK-MB. Anal. Chem. 2021, 93, 16922–16931. [Google Scholar] [CrossRef] [PubMed]
  94. Mahmoud, M.; Ruppert, C.; Rentschler, S.; Laufer, S.; Deigner, H.-P. Combining aptamers and antibodies: Lateral flow quantification for thrombin and interleukin-6 with smartphone readout. Sens. Actuators B Chem. 2021, 333, 129246. [Google Scholar] [CrossRef]
  95. Nagatoishi, S.; Tanaka, Y.; Tsumoto, K. Circular dichroism spectra demonstrate formation of the thrombin-binding DNA aptamer G-quadruplex under stabilizing-cation-deficient conditions. Biochem. Biophys. Res. Commun. 2007, 352, 812–817. [Google Scholar] [CrossRef]
  96. Lozano, R.; Naghavi, M.; Foreman, K.; Lim, S.; Shibuya, K.; Aboyans, V.; Abraham, J.; Adair, T.; Aggarwal, R.; Ahn, S.Y.; et al. Global and regional mortality from 235 causes of death for 20 age groups in 1990 and 2010: A systematic analysis for the Global Burden of Disease Study 2010. Lancet 2012, 380, 2095–2128. [Google Scholar] [CrossRef]
  97. Dong, J.; Olano, J.P.; McBride, J.W.; Walker, D.H. Emerging Pathogens: Challenges and Successes of Molecular Diagnostics. J. Mol. Diagn. 2008, 10, 185–197. [Google Scholar] [CrossRef] [Green Version]
  98. Caliendo, A.M.; Gilbert, D.N.; Ginocchio, C.C.; Hanson, K.E.; May, L.; Quinn, T.C.; Tenover, F.C.; Alland, D.; Blaschke, A.J.; Bonomo, R.A.; et al. Better Tests, Better Care: Improved Diagnostics for Infectious Diseases. Clin. Infect. Dis. 2013, 57, S139–S170. [Google Scholar] [CrossRef] [Green Version]
  99. Pan, Q.; Wang, Q.; Sun, X.; Xia, X.; Wu, S.; Luo, F.; Zhang, X.-L. Aptamer Against Mannose-capped Lipoarabinomannan Inhibits Virulent Mycobacterium tuberculosis Infection in Mice and Rhesus Monkeys. Mol. Ther. 2014, 22, 940–951. [Google Scholar] [CrossRef] [Green Version]
  100. Abbaspour, A.; Norouz-Sarvestani, F.; Noori, A.; Soltani, N. Aptamer-conjugated silver nanoparticles for electrochemical dual-aptamer-based sandwich detection of staphylococcus aureus. Biosens. Bioelectron. 2015, 68, 149–155. [Google Scholar] [CrossRef]
  101. Duan, N.; Wu, S.; Zhu, C.; Ma, X.; Wang, Z.; Yu, Y.; Jiang, Y. Dual-color upconversion fluorescence and aptamer-functionalized magnetic nanoparticles-based bioassay for the simultaneous detection of Salmonella Typhimurium and Staphylococcus aureus. Anal. Chim. Acta 2012, 723, 1–6. [Google Scholar] [CrossRef] [PubMed]
  102. Li, L.; Liu, Z.; Zhang, H.; Yue, W.; Li, C.-W.; Yi, C. A point-of-need enzyme linked aptamer assay for Mycobacterium tuberculosis detection using a smartphone. Sens. Actuators B Chem. 2018, 254, 337–346. [Google Scholar] [CrossRef]
  103. Shrivastava, S.; Lee, W.I.; Lee, N.E. Culture-free, highly sensitive, quantitative detection of bacteria from minimally processed samples using fluorescence imaging by smartphone. Biosens. Bioelectron. 2018, 109, 90–97. [Google Scholar] [CrossRef] [PubMed]
  104. Man, Y.; Ban, M.; Li, A.; Jin, X.; Du, Y.; Pan, L. A microfluidic colorimetric biosensor for in-field detection of Salmonella in fresh-cut vegetables using thiolated polystyrene microspheres, hose-based microvalve and smartphone imaging APP. Food Chem. 2021, 354, 129578. [Google Scholar] [CrossRef]
  105. Gu, Y.; Li, J.; Qian, K.; Zhang, Z.; Wang, S.; Wang, J. Integrated dual-signal aptasensor based on magnet-driven operations and miniaturized analytical device for on-site analysis. Sens. Actuators B Chem. 2020, 310. [Google Scholar] [CrossRef]
  106. Zhang, H.; Liu, Y.; Zhang, K.; Ji, J.; Liu, J.; Liu, B. Single Molecule Fluorescent Colocalization of Split Aptamers for Ultrasensitive Detection of Biomolecules. Anal. Chem. 2018, 90, 9315–9321. [Google Scholar] [CrossRef]
  107. Guo, H.; Li, J.; Li, Y.; Wu, D.; Ma, H.; Wei, Q.; Du, B. Exciton energy transfer-based fluorescent sensor for the detection of Hg2+ through aptamer-programmed self-assembly of QDs. Anal. Chim. Acta 2019, 1048, 161–167. [Google Scholar] [CrossRef]
  108. Jin, B.; Yang, Y.; He, R.; Park, Y.I.; Lee, A.; Bai, D.; Li, F.; Lu, T.J.; Xu, F.; Lin, M. Lateral flow aptamer assay integrated smartphone-based portable device for simultaneous detection of multiple targets using upconversion nanoparticles. Sens. Actuators B Chem. 2018, 276, 48–56. [Google Scholar] [CrossRef]
  109. Ma, X.; Jiang, Y.; Jia, F.; Yu, Y.; Chen, J.; Wang, Z. An aptamer-based electrochemical biosensor for the detection of Salmonella. J. Microbiol. Methods 2014, 98, 94–98. [Google Scholar] [CrossRef]
  110. Cruz-Aguado, J.A.; Penner, G. Determination of Ochratoxin A with a DNA Aptamer. J. Agric. Food Chem. 2008, 56, 10456–10461. [Google Scholar] [CrossRef]
  111. Li, Q.; Michaelis, M.; Wei, G.; Colombi Ciacchi, L. A novel aptasensor based on single-molecule force spectroscopy for highly sensitive detection of mercury ions. Analyst 2015, 140, 5243–5250. [Google Scholar] [CrossRef] [PubMed]
  112. Gong, Y.; Zheng, Y.; Jin, B.; You, M.; Wang, J.; Li, X.; Lin, M.; Xu, F.; Li, F. A portable and universal upconversion nanoparticle-based lateral flow assay platform for point-of-care testing. Talanta 2019, 201, 126–133. [Google Scholar] [CrossRef] [PubMed]
  113. Ma, W.; Yin, H.; Xu, L.; Xu, Z.; Kuang, H.; Wang, L.; Xu, C. Femtogram ultrasensitive aptasensor for the detection of Ochratoxin A. Biosens. Bioelectron. 2013, 42, 545–549. [Google Scholar] [CrossRef] [PubMed]
  114. Zhong, Z.T.; Wang, H.B.; Zhang, T.; Li, C.Q.; Liu, B.; Zhao, Y.D. Quantitative analysis of various targets based on aptamer and functionalized Fe3O4@graphene oxide in dairy products using pregnancy test strip and smartphone. Food Chem. 2021, 352, 129330. [Google Scholar] [CrossRef] [PubMed]
  115. Wang, H.-B.; Ma, L.-H.; Fang, B.-Y.; Tan, F.; Cao, Y.-C.; Zhao, Y.-D.; Hu, X.-B. Visual detection of Pb2+ using strip biosensor based on PS2M aptamer and sensitivity enhancement probe. Sens. Actuators B Chem. 2018, 261, 307–315. [Google Scholar] [CrossRef]
  116. Wu, S.; Zhang, H.; Shi, Z.; Duan, N.; Fang, C.; Dai, S.; Wang, Z. Aptamer-based fluorescence biosensor for chloramphenicol determination using upconversion nanoparticles. Food Control. 2015, 50, 597–604. [Google Scholar] [CrossRef]
  117. Eissa, S.; Zourob, M. In vitro selection of DNA aptamers targeting β-lactoglobulin and their integration in graphene-based biosensor for the detection of milk allergen. Biosens. Bioelectron. 2017, 91, 169–174. [Google Scholar] [CrossRef]
  118. Song, S.; Wang, L.; Li, J.; Fan, C.; Zhao, J. Aptamer-based biosensors. TrAC Trends Anal. Chem. 2008, 27, 108–117. [Google Scholar] [CrossRef]
  119. O’Sullivan, C.K. Aptasensors—The future of biosensing? Anal. Bioanal. Chem. 2002, 372, 44–48. [Google Scholar] [CrossRef]
  120. Wu, L.; Wang, Y.; Xu, X.; Liu, Y.; Lin, B.; Zhang, M.; Zhang, J.; Wan, S.; Yang, C.; Tan, W. Aptamer-Based Detection of Circulating Targets for Precision Medicine. Chem. Rev. 2021, 121, 12035–12105. [Google Scholar] [CrossRef]
  121. Zhang, D.; Liu, Q. Biosensors and bioelectronics on smartphone for portable biochemical detection. Biosens. Bioelectron. 2016, 75, 273–284. [Google Scholar] [CrossRef] [PubMed]
Figure 1. (a) Colorimetric detection principle based on AuNPs. Reprinted with permission from Ref. [48], copyright 2020, Elsevier. (b) Schematic illustration of the smartphone-based colorimetric system. Reprinted with permission from Ref. [48], copyright 2020, Elsevier. (c) Optical microwell reader picture and schematic of a microwell reader mounted on an Android smartphone. Reprinted with permission from Ref. [49], copyright 2016, MDPI, Basel, Switzerland, under the Creative Commons Attribution License. (d) Schematic of the designed smartphone-based colorimeter. Reprinted with permission from Ref. [50], copyright 2019, Elsevier. (e) Schematic illustration and the standard curve with different Cd2+ concentrations using the proposed smartphone-based colorimetric method. Reprinted with permission from Ref. [51], copyright 2019, MDPI, Basel, Switzerland, under the Creative Commons Attribution License.
Figure 1. (a) Colorimetric detection principle based on AuNPs. Reprinted with permission from Ref. [48], copyright 2020, Elsevier. (b) Schematic illustration of the smartphone-based colorimetric system. Reprinted with permission from Ref. [48], copyright 2020, Elsevier. (c) Optical microwell reader picture and schematic of a microwell reader mounted on an Android smartphone. Reprinted with permission from Ref. [49], copyright 2016, MDPI, Basel, Switzerland, under the Creative Commons Attribution License. (d) Schematic of the designed smartphone-based colorimeter. Reprinted with permission from Ref. [50], copyright 2019, Elsevier. (e) Schematic illustration and the standard curve with different Cd2+ concentrations using the proposed smartphone-based colorimetric method. Reprinted with permission from Ref. [51], copyright 2019, MDPI, Basel, Switzerland, under the Creative Commons Attribution License.
Biosensors 12 00477 g001
Figure 3. (a) Real image of the smartphone-based electrochemical biosensing system for detection of miR-21. Reprinted with permission from Ref. [20], copyright 2020, Elsevier. (b,c) Schematic illustration of the miR-21, let-7a simultaneous detection based on aptamer and smartphone. Reprinted with permission from Ref. [79], copyright 2019, American Chemical Society. (d) Internal illustration and appearance of self-made equipment with a smartphone. Reprinted with permission from Ref. [73], copyright 2018, Royal Society of Chemistry.
Figure 3. (a) Real image of the smartphone-based electrochemical biosensing system for detection of miR-21. Reprinted with permission from Ref. [20], copyright 2020, Elsevier. (b,c) Schematic illustration of the miR-21, let-7a simultaneous detection based on aptamer and smartphone. Reprinted with permission from Ref. [79], copyright 2019, American Chemical Society. (d) Internal illustration and appearance of self-made equipment with a smartphone. Reprinted with permission from Ref. [73], copyright 2018, Royal Society of Chemistry.
Biosensors 12 00477 g003
Figure 5. (A) Schematic illustrations of direct (a) and indirect (b) dot-blot assay. Reprinted with permission from Ref. [102], copyright 2018, Elsevier. (B) The color spots with varying M.tb concentration and application interfaces of the calibration curves in the direct (a,c) and indirect (b,d) dot-blot system, respectively. Reprinted with permission from Ref. [102], copyright 2018, Elsevier. (C) Schematic illustration of Staphylococcus aureus capture and quantitative detection in minimally processed samples. Reprinted with permission from Ref. [103], copyright 2018, Elsevier. (D) The schematic representation of the proposed microfluidic colorimetric biosensor based on a smartphone imaging APP for detecting Salmonella typhimurium. Reprinted with permission from Ref. [104], copyright 2021, Elsevier.
Figure 5. (A) Schematic illustrations of direct (a) and indirect (b) dot-blot assay. Reprinted with permission from Ref. [102], copyright 2018, Elsevier. (B) The color spots with varying M.tb concentration and application interfaces of the calibration curves in the direct (a,c) and indirect (b,d) dot-blot system, respectively. Reprinted with permission from Ref. [102], copyright 2018, Elsevier. (C) Schematic illustration of Staphylococcus aureus capture and quantitative detection in minimally processed samples. Reprinted with permission from Ref. [103], copyright 2018, Elsevier. (D) The schematic representation of the proposed microfluidic colorimetric biosensor based on a smartphone imaging APP for detecting Salmonella typhimurium. Reprinted with permission from Ref. [104], copyright 2021, Elsevier.
Biosensors 12 00477 g005
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Lan, Y.; He, B.; Tan, C.S.; Ming, D. Applications of Smartphone-Based Aptasensor for Diverse Targets Detection. Biosensors 2022, 12, 477. https://doi.org/10.3390/bios12070477

AMA Style

Lan Y, He B, Tan CS, Ming D. Applications of Smartphone-Based Aptasensor for Diverse Targets Detection. Biosensors. 2022; 12(7):477. https://doi.org/10.3390/bios12070477

Chicago/Turabian Style

Lan, Ying, Baixun He, Cherie S. Tan, and Dong Ming. 2022. "Applications of Smartphone-Based Aptasensor for Diverse Targets Detection" Biosensors 12, no. 7: 477. https://doi.org/10.3390/bios12070477

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop