Article

Fabrication of CeO2/GCE for Electrochemical Sensing of Hydroquinone

1
Department of Chemistry, Medi-Caps University, AB Road, Pigdamber, Rau, Indore 453331, M.P., India
2
Department of Chemistry, M.M.D.C, Moradabad, M.J.P. Rohilkhand University, Bareilly 244001, U.P., India
3
Department of Chemistry, College of Science, King Saud University, Riyadh 11451, Saudi Arabia
4
School of Materials Science and Engineering, Yeungnam University, Gyeongsan 38541, Korea
*
Authors to whom correspondence should be addressed.
These authors contributed equally to this work.
Biosensors 2022, 12(10), 846; https://doi.org/10.3390/bios12100846
Submission received: 14 September 2022 / Revised: 29 September 2022 / Accepted: 1 October 2022 / Published: 8 October 2022

Abstract

:
Hydroquinone is a widely used derivative of phenol which has a negative influence on human beings and the environment. The determination of the accurate amount of hydroquinone is of great importance. Recently, the fabrication of an electrochemical sensing device has received enormous attention. In this study, we reported on the facile synthesis of cerium dioxide (CeO2) nanoparticles (NPs). The CeO2 NPs were synthesized using cerium nitrate hexahydrate as a precursor. For determining the physicochemical properties of synthesized CeO2 NPs, various advanced techniques, viz., powder X-ray diffraction (PXRD), scanning electron microscopy (SEM), energy-dispersive X-ray spectroscopy (EDX), and X-ray photoelectron spectroscopy (XPS), were studied. Further, these synthesized CeO2 NPs were used for the modification of a glassy carbon electrode (CeO2/GCE), which was utilized for the sensing of hydroquinone (HQ). A decent detection limit of 0.9 µM with a sensitivity of 0.41 µA/µM cm2 was exhibited by the modified electrode (CeO2/GCE). The CeO2/GCE also exhibited good stability, selectivity, and repeatability towards the determination of HQ.

1. Introduction

Hydroquinone (benzene 1,4-diol, HQ) is a positional isomer of a phenolic compound and is known for its high toxicity to the ecological environment and humans, even if it is used at very low concentrations [1,2]. If recent reports from the United States Environmental Protection Agency (UPA) and the European Union (EU) are taken into account, HQ is harmful to the environment due to its lower rate of degradability in ecological environments and its biohazard nature [3,4]. HQ finds a variety of applications in bleaching creams, pesticides, medicines, cosmetics, secondary coloring materials, photography chemicals, and flavoring compounds [5,6,7,8]. Owing to the large industrial use of HQ, it is being released into the environment and aquatic systems; therefore, it becomes essential to detect the traces of HQ [7].
For determining and quantifying the concentration of HQ in a system, a number of analytical ways are being applied, viz., gas chromatography-mass spectrometry (GCMS), high-performance liquid chromatography (HPLC), capillary electrochromatography, chemiluminescence, the fluorimetric method, flow injection analysis, spectrophotometry, electrochemical methods, etc. [9,10,11,12,13,14,15]. However, gas chromatography and high-performance liquid chromatography involve costly, sophisticated instruments and skilled staff alongside the usage of extra-pure and large volumes of organic solvents; hence, they are inappropriate for routine field analysis. On the other hand, in the determination of HQ via the spectrophotometric technique, the presence of correlated compounds interferes in the result. Additionally, the reproducibility of the results remains a prime concern in the fluorescence-based methods.
On the contrary, merits such as cost effectiveness, easy handling, rapid response, stability, high sensitivity, high selectivity, a wider detection range, and a lower limit of detection (LOD) makes the electrochemical method a promising analytical approach for the detection of pollutants [1,16,17,18]. Unfortunately, if bare electrodes such as glassy carbon/platinum/gold/graphite electrodes are used for the quantification and determination of HQ, they show a poor response and, sometimes, even a redox peak is not observed. Therefore, it becomes significant to fabricate new electrode materials which are cost effective, easy to handle, stable, and may exhibit excellent electrocatalytic activity and selectivity for the HQ.
In recent years, researchers have made significant efforts to determine different pollutants electrochemically using chemically modified electrodes (CMEs), as CMEs are consistent, economic, give a rapid response, and consume low energy, particularly during the in situ determination of pollutants [19,20]. For the assembly of CMEs, it is vital to select a suitable modifier with enhanced electrocatalytic activity or excellent conductivity. The modification of GC, gold and platinum electrodes by nanoparticles/nanocomposites of transition metal oxides, their composites, and conducting polymers have emerged as a leading area of research during the last decade [20,21,22,23,24,25].
In 2020, Mobin and the research team at IIT Indore developed a novel HQ sensor using the N-rGO/SrZrO3 composite as an electrode material [1]. In 2019, Ahmad et al. [21] also reported the fabrication of an HQ and catechol sensor using SnO2 as the electrode material. In another work, Meskher et al. fabricated NiO/rGO/f-MWCNTs on a platinum (Pt) electrode for the determination of HQ and catechol simultaneously [26]. This indicated that electrode materials have a significant role in the development of electrochemical sensors [1].
Cerium dioxide (CeO2) is a rare-earth metal oxide which possesses excellent biocompatibility, oxygen vacancy formation, and natural abundance [27]. CeO2 has been widely used in a variety of applications such as catalysis, solar cells, electrochemical sensors, energy storage, batteries, light-emitting diodes, photodetectors, and photocatalysis [28,29,30,31,32,33,34]. Temerk et al. modified a GC paste electrode using doped CeO2 NPs and applied it for electrochemical sensing of uric acid in a biological fluid [35]. The interest in the ceria nanoparticles (CeO2 NPs) is due to the presence of Ce3+ ions and generated oxygen-vacancies. The Ce3+ ions on the surface act as sites for catalytic reactions and generated oxygen vacancies help in the transformation and migration of active species present in that system [36].
Herein, we report a “benign approach for the synthesis of CeO2 NPs for electrochemical sensing of HQ”. The fabricated HQ sensor exhibits excellent sensing performance in terms of detection limit, sensitivity, and repeatability. Moreover, the fabricated CeO2/GCE demonstrates excellent selectivity towards the sensing of HQ in the presence of various electroactive species.

2. Materials and Methods

2.1. Chemicals

The chemicals, reagents, and solvents used for the experiment, viz., cerium nitrate hexahydrate (Ce(NO3)3.6H2O), glucose, ascorbic acid, dopamine, uric acid, potassium hydroxide (KOH), phosphate-buffered saline (PBS) solutions, and hydroquinone, were purchased from Alfa Aesar, Loba, SRL, Fisher Scientific, Sigma, Merck, and TCI. All the chemicals were used as received without any further purification or treatment. The stock solution of a 0.1 M concentration of HQ was prepared and used for the electrochemical sensing measurement.

2.2. Synthesis of CeO2

The CeO2 NPs were prepared using the precipitation method, followed by calcination. In brief, we dissolved 3.26 g of Ce(NO3)3.6H2O in 25 mL of distilled water via stirring at room temperature (RT). After that, a 1 M solution of KOH was prepared using distilled water (D.W.). We slowly added the prepared 1M KOH solution to the aqueous solution of Ce(NO3)3.6H2O. The reaction mixture was stirred for 2 h at RT. Further, the precipitate was collected through centrifugation and dried at 60 °C overnight. Finally, it was calcinated at 450 °C for 3 h, which yielded CeO2 NPs (Scheme 1).

2.3. Fabrication of Electrode

The bare GCE was cleaned with an alumina slurry (0.5 µm) and velvet pad. Further, 3.5 mg of CeO2 NPs was dispersed in 3 mL of DW (0.1 wt% Nafion) using ultrasonication for 1 h. The cleaned GCE (3 mm) was modified with 7.5 µL of CeO2 ink and dried at room temperature for 3 h (Scheme 2).
This CeO2-modified GCE is denoted as CeO2/GCE and used as the working electrode in this experiment. Here, the reference electrode is silver/silver chloride (Ag/AgCl), whereas the counter electrode is a platinum electrode (the Ag/AgCl electrode was filled with a 3 M KCl solution). We performed all the electrochemical measurements on three electrode computer-controlled potentiostat/galvanostat systems (CH Instruments).

3. Results

3.1. Physiochemical Properties of CeO2

We used the powder X-ray diffraction (PXRD, Rigaku) method to confirm the structural phase of CeO2. The PXRD pattern of the obtained CeO2 NPs between the 2θ range of 10–80° is presented in Figure 1. The diffraction peaks are observed at ~28.66°, 33.03°, 47.56°, 56.39°, 59.05°, 69.34°, 76.61°, and 79.27°, which relate to (111), (200), (220), (311), (222), (400), (331), and (420) planes. The observed reflections exhibit the formation of the face-centered cubic phase of CeO2 where the lattice parameters are a = b = c = 0.5412 nm. The obtained PXRD pattern of the CeO2 NPs was well-matched with the previously reported JCPDS file No: 81–0792. The absence of any other additional peaks in the PXRD reveals the phase purity of the CeO2 sample. Thus, it can be said that CeO2 NPs have been prepared with high-phase purity by the precipitation method.
The average crystallite size of the synthesized CeO2 has been calculated employing the Debye–Scherrer equation given below [37].
D   = 0.9   λ β   cos   θ
where D = Average crystallite size (Å), λ = X-ray wavelength (0.154 nm), θ = diffraction angle, β = full width at half maximum (FWHM) of the observed peak.
The average crystallite size of the prepared CeO2 NPs was found to be 23.83 nm. Our results related to the crystallographic properties of the prepared CeO2 are in accordance with the previously reported literature [37]. The surface morphological/structural properties play a significant role in the development of electrochemical and optoelectronic applications. Therefore, it is required to examine the morphological features of the prepared CeO2 NPs. Thus, the surface morphology of the prepared CeO2 sample has been investigated using scanning electron microscopy (SEM, Zeiss). The recorded microscopic SEM images of the CeO2 sample are displayed in Figure 2a,b. The SEM results indicated the presence of an agglomeration of the CeO2 NPs. Furthermore, we have also examined the phase purity of the prepared CeO2 NPs using energy-dispersive X-ray spectroscopy (EDX). The EDX spectrum of the prepared CeO2 NPs was collected on Horiba EDX Instruments connected with the SEM instrument. The collected EDX spectrum of the Horiba EDX Instruments has been presented in Figure 2c. The EDX results indicated the presence of Ce and O elements in the prepared CeO2 NPs (Figure 2c). The atomic and weight percentage of the Ce and O elements in the CeO2 NPs are presented in Figure 2d. The EDX results confirmed the formation of CeO2 NPs with good purity.
Furthermore, surface atomic compositions and valence states of the prepared CeO2 NPs have been studied via XPS analysis. According to the survey spectrum, CeO2 NPs contain only cerium, oxygen, and carbon elements (Figure 3a). The Ce 3D spectrum shows the peaks for both Ce4+ and Ce3+ species, revealing the reduction of Ce4+ to Ce3+. Under the conditions of high temperatures and low oxygen partial pressures, Ce4+ is partially reduced to Ce3+ [38]. The XPS spectrum of the O1s region generally reveals knowledge about absorbed oxygen, oxygen vacancies, lattice oxygen, and the surface hydroxyl group (OH) [36,38]. On the other side, the O1s spectrum shows asymmetric peaks for the CeO2 sample. Lower binding energy peaks (~528.6–528.8 eV) could be ascribed to O2− ions associated with the lattice oxygen in the cubic structure, whereas a higher binding energy peak (~530.9 eV) could be ascribed to the O2− ions present in the oxygen-deficient region [39]. The above studies confirmed the successful formation of CeO2 NPs with good phase purity [39].

3.2. Electrochemical Sensing Properties of CeO2/GCE

Initially, electrocatalytic activities of the GCE and CeO2/GCE were studied in a 5 mM [Fe(CN)6]3−/4− redox solution. The cyclic voltammetry (CV) was used to obtained the CV graph of the GCE and CeO2/GCE in a 5 mM [Fe(CN)6]3−/4− redox solution (scan rate = 50 mV/s). The obtained CV graphs of the GCE and CeO2/GCE in the 5 mM [Fe(CN)6]3−/4− redox solution are presented in Figure S1a. The CV results exhibited the presence of a higher electrocatalytic current response for CeO2/GCE compared to the bare GCE (Figure S1a). Furthermore, electrochemical impedance spectroscopy (EIS) was also used to examine the electrocatalytic behavior of GCE and CeO2/GCE in the 5 mM [Fe(CN)6]3−/4− redox solution. Figure S1b demonstrates the Nyquist plot of the GCE and CeO2/GCE in the 5 mM [Fe(CN)6]3−/4− redox solution. The observations revealed that the bare GCE showed a large semicircle, which demonstrates the presence of a high charge-transfer resistance (Rct) of 833.4 Ω, whereas CeO2/GCE showed a small semicircle with an Rct of 512.3 Ω. Thus, it can be said that CeO2/GCE has better electrocatalytic activity, which may be due to the presence of CeO2 NPs.
The electrochemical sensing behavior of the CeO2/GCE and bare GCE were examined using the CV technique. The CV graphs of the CeO2/GCE and bare GCE were collected in the presence of 15 µM of HQ in 0.1 M PBS (pH of the PBS was 7.0) at the applied scan rate of 50 mV/s. Figure 4 demonstrates the collected CV graphs of the CeO2/GCE and bare GCE in the presence of 15 µM of HQ in 0.1 M PBS (pH = 7.0) at the applied scan rate of 50 mV/s in the potential range from −0.2 to 0.6 V. The bare GCE exhibits the oxidation peak at 0.27 V with an electrocatalytic current response of 4.65 µA (Figure 4). On the other hand, CeO2/GCE exhibits the oxidation peak at 0.27 V with an enhanced electrocatalytic current response of 7.01 µA (Figure 4). This indicated the successful surface modification of GCE with CeO2 NPs. This improved current response for CeO2/GCE was attributed to the presence of good electrochemical properties of CeO2 NPs. The CV of the CeO2/GCE was also obtained in the absence of HQ, which does not showed any peaks related to the HQ (Figure 4). Thus, it can be understood that CeO2/GCE can be further used as a potential electrochemical sensor for the detection of HQ. It is well-known that the concentration of the HQ can influence the electrochemical performance of the fabricated electrochemical sensor. Thus, we have further investigated the effect of various concentrations of the HQ on the sensing ability of the modified CeO2/GCE using the CV technique. The CV graphs of the CeO2/GCE were collected at various concentrations of HQ from 15 to 225 µM in 0.1 M PBS (pH = 7.0) at a fixed applied scan rate of 50 mV/s. Figure 5a showed the collected CV graphs of the CeO2/GCE at various concentrations of HQ from 15 to 225 µM in 0.1 M PBS (pH = 7.0) at a fixed applied scan rate of 50 mV/s. The observations indicated that the current response for CeO2/GCE increases with the increasing concentration of the HQ from 15 to 225 µM at a fixed scan rate of 50 mV/s (Figure 5a). The good sensing response of CeO2/GCE may be attributed to Ce3+ ions [40].
The calibration plot between the current responses against the concentrations of the HQ has been plotted. The calibration curve of the current response versus the concentration of the HQ has been presented in Figure 5b. The calibration plot suggests that the current response was increased linearly, where R2 = 0.99 (Figure 5b).
The influence of various applied scan rates on the electrochemical performance/ability of the CeO2/GCE was also checked using the CV technique. The CV graphs of the CeO2/GCE were obtained at a fixed concentration of 15 µM of HQ in 0.1 M PBS (pH = 7.0) at various applying scan rates (50 to 500 mV/s). The collected CV graphs of the CeO2/GCE in 15 µM of HQ in 0.1 M PBS at a pH of 7.0 are presented in Figure 6a. The CV results indicated that the current response for CeO2/GCE increases when applying scan rate changes from 50 mV/s to 500 mV/s (Figure 6a). The calibration curve between current responses and the square root of the applied scan rates was plotted, which is displayed in Figure 6b. The calibration plot between the current response and square root of the scan rate suggested that the current response increases linearly with an increasing applied scan rate, where R2 = 0.99 (Figure 6b).
According to the recently published literature, it has been observed that differential pulse voltammetry (DPV) is a highly sensitive and efficient electrochemical sensing technique compared to the CV or linear sweep voltammetry. Thus, we have also applied the DPV technique to detect the HQ using CeO2/GCE as the electrochemical sensor. The DPV graph of the CeO2/GCE and bare GCE were obtained in the presence of 15 µM of HQ in 0.1 M PBS (pH = 7.0) at the applied scan rate of 50 mV/s. The collected DPV graphs of the CeO2/GCE and bare GCE have been displayed in Figure 7.
The bare GCE demonstrates the current response of 5.57 µA towards the detection of 15 µM of HQ in 0.1 M PBS (pH = 7.0, scan rate = 50 mV/s). On the other side, CeO2/GCE exhibits significant improvement in the current response for the detection of 15 µM of HQ in 0.1 M PBS (pH = 7.0, scan rate = 50 mV/s). The highest DPV current response of 18.67 µA was obtained for CeO2/GCE, which is higher than that of the bare GCE (Figure 7). This clearly revealed that CeO2/GCE has better electrochemical sensing properties for the detection of HQ compared to the bare GCE using the DPV technique. The obtained DPV current response was found to be much better compared to the CV technique. Therefore, we have used the DPV technique for further electrochemical sensing measurements using CeO2/GCE as the HQ sensor.
The influence of the different concentrations of the HQ (15 µM, 35 µM, 55 µM, 75 µM, 100 µM, 125 µM, 150 µM, 175 µM, 200 µM, and 225 µM) on the electrochemical ability of the CeO2/GCE was also investigated. The DPV graphs of the CeO2/GCE were collected for various concentrations of HQ (15 µM, 35 µM, 55 µM, 75 µM, 100 µM, 125 µM, 150 µM, 175 µM, 200 µM, and 225 µM) in 0.1 M PBS (pH = 7.0) at a fixed scan rate of 50 mV/s. Figure 8a showed the DPV graphs of the CeO2/GCE in various concentrations of HQ (15 µM, 35 µM, 55 µM, 75 µM, 100 µM, 125 µM, 150 µM, 175 µM, 200 µM, and 225 µM) in 0.1 M PBS (pH = 7.0) at a fixed scan rate of 50 mV/s. The DPV results showed that the current response increases when increasing the concentration of HQ from 15 to 225 µM (Figure 8a). The calibration curve between the current response and concentration of HQ is presented in Figure 8b. The calibration plot between the current response and concentration of HQ suggests that the current response increases linearly, where R2 = 0.99.
The plausible sensing mechanism for the sensing of HQ by GCE/CeO2 has been drawn based on the previous literature [41]. The CV investigations suggest the involvement of the reversible oxidation–reduction reaction during the electrochemical detection of HQ.
Initially, oxygen–hydrogen bonds of both the phenolic -OH break, and subsequently, HQ is converted to quinoid by the release of two electrons and two protons (Scheme 2). Hence, it can be concluded that in the first step, the HQ is converted to quinone by liberating two protons that are later reverted to HQ and accept the liberated protons. The schematic representation is depicted in Scheme 2.
For selecting an ideal and efficient sensor, selectivity remains one of the most important parameters. Various interfering species create an interfering atmosphere, causing improper and inaccurate results during the detection of the desired analyte. Thus, in order to check the selectivity of CeO2/GCE for HQ sensing, we have applied the DPV method. First, the DPV curve of CeO2/GCE was recorded in the presence of 15 µM of HQ (Figure 9a); after that, the DPV curve of CeO2/GCE was recorded in the presence of 15 µM of HQ + various interfering molecules (glucose, dopamine, H2O2, urea, uric acid, resorcinol, chlorophenol, nitrophenol, acetone, methanol, ascorbic acid, and hydrazine). The concentration of the interfering species was five times higher than HQ. The DPVs were recorded at a scan rate of 50 mV/s. Insignificant variations in the current response were observed, as suggested by the obtained DPV curves, indicating the higher selectivity of CeO2/GCE against HQ. For the practical application of any sensor, its repeatability and stability are also to be taken into consideration. The repeatability of the designed CeO2/GCE electrode for sensing of HQ was also investigated by recording five consecutive DPV curves. DPV curves were recorded in the presence of 15 µM of HQ at a scan rate of 50 mV/s. The obtained DPVs exhibited good repeatability of CeO2/GCE for HQ sensing. In further studies, 100 DPV cycles of CeO2/GCE were obtained in 75 µM of HQ at a scan rate of 50 mV/s. The DPVs have been displayed in Figure 9c and observations revealed that CeO2/GCE has excellent stability up to 100 cycles.
The electrochemical sensing performance of any sensor can be evaluated by calculating the limit of detection (LoD) and sensitivity.
The LoD and sensitivity of the CeO2/GCE for HQ sensing were determined by using the following equations given below:
LoD = 3 × (σb/S)
where σb = standard deviation or error of blank, and S = slope of the calibration curve.
Sensitivity = S/A
A = area of the working electrode.
The LoD of 0.9 µM and sensitivity of 0.41 µA µM−1cm−2 were obtained for CeO2/GCE, which is presented in Table 1.
In the past few years, a large number of HQ sensors have been reported which demonstrated good sensing performance. Through this connection, carbon spherical shells based on an HQ sensor were reported by Martoni et al. [42]. In another report, Maciel et al. [43] prepared poly (butylene adipate-co-terephthalate)/graphite as an HQ sensor. This fabricated sensor (poly (butylene adipate-co-terephthalate)/graphite) exhibited a decent LoD of 1.04 µM and sensitivity of 0.07 µA µM−1cm−2. In 2017, Zhang et al. [44] used a polarized GCE as the HQ sensor. Authors used the CV approach for the determination of HQ and obtained an LoD of 3.57 µM with a sensitivity of µA µM−1cm−2 [44]. Feng et al. [45] designed and synthesized carboxylic functional multi-walled carbon nanotubes using a layer-by-layer covalent-self-assembly approach. Further, the HQ sensor was developed using carboxylic functional multi-walled carbon nanotubes as the electrode modifier [45]. Authors modified the GCE with carboxylic functional multi-walled carbon nanotubes and investigated its sensing ability towards HQ detection using CV and DPV techniques [45]. The fabricated carboxylic functional multi-walled carbon nanotubes/GCE exhibited an LoD of 2.3 µM [45]. In another work, Hu et Al. [46] fabricated the rGO/MWCNTs composite and modified the GCE with rGO/MWCNTs as the electrode modifier. This modified GCE (rGO/MWCNTs/GCE) showed an LoD of 2.6 µM with excellent selectivity using CV and DPV techniques [46]. Umasankar et al. [47] prepared MWCNT-poly-malachite green on a GCE using the potentiodynamic method. The fabricated MWCNT-poly-malachite green/GCE was employed as the HQ sensor, which demonstrated an LoD of 1.6 µM. In another work, Erogul et al. [48] obtained an iron oxide (Fe3O4)-functionalized graphene oxide-gold nanoparticle (AuNPs/Fe3O4/APTESGO/GCE) and applied it as the HQ sensor. This fabricated electrode (AuNPs/Fe3O4/APTESGO/GCE) exhibits an LoD of 1.1 µM. In another work, a CNT/GCE-based HQ sensor exhibited an LoD of 2.9 µM [49], whereas Li et al. [50] fabricated a Zn/Al-layered double-hydroxide (LDHf) film on the GCE (LDHf/GCE). This fabricated LDHf/GCE showed an LoD of 9 µM [50]. Peng et al. [51] developed a GCE-based HQ sensor. This developed sensor showed the LoD of 8 µM [51]. Our obtained results are comparable with previously published sensors in terms of the LoD, as shown in Table 1 [42,43,44,45,46,47,48,49,50,51].

4. Conclusions

In conclusion, we can summarize that CeO2 NPs have been synthesized at room temperature via the precipitation method. The physical and chemical features of the synthesized CeO2 NPs were examined by using advanced characterization tools. Further, a hydroquinone sensor was developed using a glassy carbon electrode as the working substrate. The as obtained CeO2 NPs were drop-casted on to the active carbon surface of the glassy carbon electrode. The fabricated CeO2 NP-based glassy carbon electrode (CeO2/GCE) exhibits a reasonable limit of detection of 0.9 µM. The fabricated CeO2/GCE also showed a decent sensitivity of 0.41 µA µM−1cm−2. These observations indicated that CeO2/GCE has good electrocatalytic properties towards the sensing of hydroquinone. In further investigations, it was also found that CeO2/GCE possesses good repeatability, stability, and selectivity for the determination of hydroquinone using differential pulse voltammetry. We believe that the fabricated CeO2/GCE may be employed as a potential sensor for the determination of hydroquinone.

Supplementary Materials

The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/bios12100846/s1, Figure S1: CVs (a) and Nyquist plot (b) of GCE and CeO2/GCE in 5 mM [Fe(CN)6]3−/4− and 0.1 M KCl in the applied frequency ranges of 0.1 Hz–100 kHz.

Author Contributions

Conceptualization, K.A. and A.C.; Methodology, R.A.K., M.Q.K. and A.C.; Formal analysis, K.A., R.A.K. and A.C.; Writing—original draft. A.C., M.Q.K. and K.A.; Writing—review & editing, H.K. and A.A.; Supervision, H.K. and A.A. Funding acquisition, H.K. and A.A. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Acknowledgments

A.A. acknowledges the researchers supporting project number (RSP-2021/78), King Saud University, Riyadh, Saudi Arabia. This work was supported by the Korea Innovation Foundation (INNOPOLIS) grant funded by the Korean government (MSIT) (2020-DD-UP-0278) and the National Research Foundation of Korea (NRF) grant funded by the Korean government (MSIT) (No. 2019R1A5A8080290).

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Ahmad, K.; Kumar, P.; Mobin, S.N. A highly sensitive and selective hydroquinone sensor based on a newly designed N-rGO/SrZrO3 composite. Nanoscale Adv. 2020, 2, 502–511. [Google Scholar] [CrossRef] [Green Version]
  2. He, J.; Qiu, R.; Li, W.; Xing, S.; Song, Z.; Li, Q.; Zhang, S. A voltammetric sensor based on eosin Y film modified glassy carbon electrode for simultaneous determination of hydroquinone and catechol. Anal. Methods 2014, 6, 6494–6503. [Google Scholar] [CrossRef]
  3. EPA, United States Environmental Protection Agency (USEPA). Health effect notebook for hazardous air pollutants. Available online: http://www.epa.gov/ttn/atw/hlthef/hydroqui.html (accessed on 29 September 2022).
  4. Verma, R.; Vinoda, K.S.; Papireddy, M.; Gowda, A.N.S. Toxic Pollutants from Plastic Waste—A Review. Procedia Environ. Sci. 2016, 35, 701–708. [Google Scholar]
  5. Mohanadas, D.; Tukimin, N.; Sulaiman, Y. Simultaneous electrochemical detection of hydroquinone and catechol using poly(3,4-ethylenedioxythiophene)/reduced graphene oxide/manganese dioxide. Synth. Met. 2019, 252, 76–81. [Google Scholar] [CrossRef]
  6. Huang, R.; Liao, D.; Chen, S.; Yu, J.; Jiang, X. A strategy for effective electrochemical detection of hydroquinone and catechol: Decoration of alkalization-intercalated Ti3C2 with MOF-derived N-doped porous carbon. Sens. Actuator B Chem. 2020, 320, 128386. [Google Scholar] [CrossRef]
  7. Huang, R.; Chen, S.; Yu, J.; Jiang, X. Self-assembled Ti3C2/MWCNTs nanocomposites modified glassy carbon electrode for electrochemical simultaneous detection of hydroquinone and catechol. Ecotoxicol. Environ. Saf. 2019, 184, 109619. [Google Scholar] [CrossRef]
  8. Chetankumar, K.; Swamy, B.E.K.; Sharma, S.C.; Hariprasad, S.A. An efficient electrochemical sensing of hazardous catechol and hydroquinone at direct green 6 decorated carbon paste electrode. Sci. Rep. 2021, 11, 15064. [Google Scholar] [CrossRef]
  9. Karthika, A.; Raja, V.R.; Karuppasamy, P.; Suganthi, A.; Rajarajan, M. A novel electrochemical sensor for determination of hydroquinone in water using FeWO4/SnO2 nanocomposite immobilized modified glassy carbon electrode. Arab. J. Chem. 2020, 13, 4065–4081. [Google Scholar] [CrossRef]
  10. Ahmed, J.; Rahman, M.M.; Siddiquey, I.A.; Asiri, A.M.; Hasnat, M.A. Efficient hydroquinone sensor based on zinc, strontium and nickel based ternary metal oxide (TMO) composites by differential pulse voltammetry. Sens. Actuats B 2018, 256, 383–392. [Google Scholar] [CrossRef]
  11. Liu, Y.; Liao, H.; Zhou, Y.; Du, Y.; Wei, C.; Zhao, J.; Sun, S.; Loo, J.S.C.; Xu, Z.J. Fe2O3 Nanoparticle/SWCNT Composite Electrode for Sensitive Electrocatalytic Oxidation of Hydroquinone. Electrochim. Acta 2015, 180, 1059–1067. [Google Scholar] [CrossRef]
  12. Soltani, H.; Pardakhty, A.; Ahmadzadeh, S. Determination of hydroquinone in food and pharmaceutical samples using a voltammetric based sensor employing NiO nanoparticle and ionic liquids. J. Mol. Liq. 2016, 219, 63–67. [Google Scholar] [CrossRef]
  13. Guo, H.L.; Peng, S.; Xu, J.H.; Zhao, Y.Q.; Kang, X. Highly stable pyridinic nitrogen doped graphene modified electrode in simultaneous determination of hydroquinone and catechol. Sens. Actuators B 2014, 193, 623–629. [Google Scholar] [CrossRef]
  14. Du, H.; Ye, J.; Zhang, J.; Huang, X.; Yu, C. A voltammetric sensor based on grapheme-modified electrode for simultaneous determination of catechol and hydroquinone. J. Electroanal. Chem. 2011, 650, 209–213. [Google Scholar] [CrossRef]
  15. El-Azazy, M.; Ahsan, I.; Bensalah, N. Electrochemical Analysis of Sulfisoxazole Using Glassy Carbon Electrode (GCE) and MWCNTs/Rare Earth Oxide (CeO2 and Yb2O3) Modified-GCE Sensors. Molecules 2022, 27, 2033. [Google Scholar] [CrossRef] [PubMed]
  16. Safavi, A.; Maleki, N.; Moradlou, O. A selective and sensitive method for simultaneous determination of traces of paracetamol and p-aminophenol in pharmaceuticals using carbon ionic liquid electrode. Electroanalysis 2008, 20, 2158. [Google Scholar] [CrossRef]
  17. Huang, W.; Hu, W.; Song, J. Adsorptive stripping voltammetric determination of 4-aminophenol at a single-wall carbon nanotubes film coated electrode. Talanta 2003, 61, 411. [Google Scholar] [CrossRef]
  18. Rahman, M.M.; Khan, S.B.; Jamal, A.; Faisal, M.; Asiri, A.M. Highly sensitive methanol chemical sensor based on undoped silver oxide nanoparticles prepared by a solution method. Microchim. Acta 2012, 178, 99. [Google Scholar] [CrossRef]
  19. Kumunda, C.; Adekunle, A.S.; Mamba, B.B.; Hlongwa, N.W.; Nkambule, T.T.I. Electrochemical Detection of Environmental Pollutants Based on Graphene Derivatives: A Review. Front. Mater. 2021, 7, 616787. [Google Scholar] [CrossRef]
  20. Chang, F.; Wang, H.; He, S.; Gu, Y.; Zhu, W.; Li, T.; Ma, R. Simultaneous determination of hydroquinone and catechol by a reduced graphene oxide-polydopamine-carboxylated multi-walled carbon nanotube nanocomposite. RSC Adv. 2021, 11, 31950–31958. [Google Scholar] [CrossRef] [PubMed]
  21. Ahmad, K.; Kumar, P.; Mobin, S.M. Hydrothermally Grown SnO2 Flowers as Efficient Electrode Modifier for Simultaneous Detection of Catechol and Hydroquinone. J. Electrochem. Soc. 2019, 166, B1577. [Google Scholar] [CrossRef]
  22. Ahmad, K.; Mobin, S.M. Shape controlled synthesis of high surface area MgO microstructures for highly efficient congo red dye removal and peroxide sensor. J. Environ. Chem. Eng. 2019, 7, 103347. [Google Scholar] [CrossRef]
  23. Ahmad, K.; Kumar, P.; Mobin, S.M. Hydrothermally grown novel pyramids of the CaTiO3 perovskite as an efficient electrode modifier for sensing applications. Mater. Adv. 2020, 1, 2003–2009. [Google Scholar] [CrossRef]
  24. Ahmad, K.; Mobin, S.M. High surface area 3D-MgO flowers as the modifier for the working electrode for efficient detection of 4-chlorophenol. Nanoscale Adv. 2019, 1, 719–727. [Google Scholar] [CrossRef] [Green Version]
  25. Ahmad, K.; Shinde, M.A.; Kim, H. Molybdenum disulfide/reduced graphene oxide: Progress in synthesis and electro-catalytic properties for electrochemical sensing and dye sensitized solar cells. Microchem. J. 2021, 169, 106583. [Google Scholar] [CrossRef]
  26. Meskhar, H.; Achi, F.; Zouaoui, A.; Ha, S.; Peacock, M.; Belkhalfa, H. Simultaneous and Selective Electrochemical Determination of Catechol and Hydroquinone on A Nickel Oxide (NiO) Reduced Graphene Oxide (rGO) Doped Multiwalled Carbon Nanotube (fMWCNT) Modified Platinum Electrode. Anal. Lett. 2022, 55, 1466–1481. [Google Scholar] [CrossRef]
  27. Wu, J.; Wu, Y.; Lu, L.; Zhang, D.; Wang, X. Single-atom Au catalyst loaded on CeO2: A novel single-atom nanozyme electrochemical H2O2 sensor. Talanta Open 2021, 4, 100075. [Google Scholar] [CrossRef]
  28. Rajendran, S.; Manoj, D.; Suresh, R.; Vasseghian, Y.; Ghfar, A.A.; Sharma, G.; Soto-Moscoso, M. Electrochemical detection of hydrogen peroxide using micro and nanoporous CeO2 catalysts. Environ. Res. 2022, 214, 113961. [Google Scholar] [CrossRef] [PubMed]
  29. Gayathri, R.; Raja, G.; Rajeswaran, P. A simple and one step low cost microwave induced low cost grapheme modified CeO2 photo electrodes for high-efficiency dye-sensitized solar cells. Inorg. Chem. Commun. 2020, 120, 108132. [Google Scholar] [CrossRef]
  30. Das, H.T.; Balaji, T.E.; Dutta, S.; Das, N.; Das, P.; Mondal, A.; Imran, M. Recent trend of CeO2-based nanocomposites electrode in supercapacitor: A review on energy storage applications. J. Energy Storage 2022, 50, 104643. [Google Scholar] [CrossRef]
  31. Habib, I.Y.; Burhan, J.; Jaladi, F.; Lim, C.M.; Usman, A.; Kumara, N.T.R.N.; Tsang, S.C.E.; Mahadi, A.J. Effect of Cr doping in CeO2 nanostructures on photocatalysis and H2O2 assisted methylene blue dye degradation. Catal. Today 2021, 375, 506–513. [Google Scholar] [CrossRef]
  32. Zinzuvadiya, S.; Pandya, N.C.; Joshi, U.S. Optoelectronic response of (111) oriented CeO2 films for UV photodetector. Thin Solid Film. 2019, 669, 525–530. [Google Scholar] [CrossRef]
  33. Basavaraj, R.B.; Navami, D.; Deepthi, N.H.; Venkataravanappa, M.; Lokesh, R.; Sudheer Kumar, K.H.; Sreelakshmi, T.K. Novel orange-red emitting Pr3+ doped CeO2 nanopowders for white light emitting diode applications. Inorg. Chem. Commun. 2020, 120, 108164. [Google Scholar] [CrossRef]
  34. Cheng, P.; Guo, P.; Sun, K.; Zhao, Y.; Liu, D.; He, D. CeO2 decorated graphene as separator modification material for capture and boost conversion of polysulfide in lithium-sulfur batteries. J. Membr. Sci. 2021, 619, 118780. [Google Scholar] [CrossRef]
  35. Temerk, Y.; Ibrahim, H. A new sensor based on In doped CeO2 nanoparticles modified glassy carbon paste electrode for sensitive determination of uric acid in biological fluids. Sens. Actuators B Chem. 2016, 224, 868–877. [Google Scholar] [CrossRef]
  36. Soni, S.; Vats, V.S.; Kumar, S.; Dalela, B.; Mishra, M.; Meena, R.S.; Gupta, G.; Alvi, P.A.; Dalela, S. Structural, optical and magnetic properties of Fe-doped CeO2 samples probed using X-ray photoelectron spectroscopy. J. Mater. Sci. Mater. Electron. 2018, 29, 10141–10153. [Google Scholar] [CrossRef]
  37. Tamizhdurai, P.; Sakthinathan, S.; Chen, S.-M.; Shanthi, K.; Sivasanker, S.; Sangeetha, P. Environmentally friendly synthesis of CeO2 nanoparticles for the catalytic oxidation of benzyl alcohol to benzaldehyde and selective detection of nitrite. Sci. Rep. 2017, 7, 46372. [Google Scholar] [CrossRef] [Green Version]
  38. Korjus, O.; Aruvali, J.; Kivi, I.; Kodu, M.; Lust, E.; Nurk, G. Simultaneous Operando Characterization of Crystallographic and Electrochemical Properties of Ni-Ce0.9Gd0.1O2-δ Solid Oxide Fuel Cell Anode. J. Electrochemi. Soc. 2018, 165, F1043–F1050. [Google Scholar] [CrossRef]
  39. Maslakov, K.I.; Teterin, Y.A.; Popel, A.J.; Teterin, A.Y.; Ivanov, K.E.; Kalmykov, S.M.; Petrov, V.G.; Petrov, P.K.; Farnan, I. XPS study of ion irradiated and unirradiated CeO2 bulk and thin film samples. Appl. Surf. Sci. 2018, 448, 154–162. [Google Scholar] [CrossRef]
  40. Rajendran, S.; Khan, M.M.; Gracia, F.; Qin, J.; Gupta, V.K.; Arumainathan, S. Ce3+-ion-induced visible-light photocatalytic degradation and electrochemical activity of ZnO/CeO2 nanocomposite. Sci. Rep. 2016, 6, 31641. [Google Scholar] [CrossRef] [Green Version]
  41. Domínguez-Aragón, A.; Dominguez, R.B.; Zaragoza-Contreras, E.A. Simultaneous Detection of Dihydroxybenzene Isomers Using Electrochemically Reduced Graphene Oxide-Carboxylated Carbon Nanotubes/Gold Nanoparticles Nanocomposite. Biosensors 2021, 11, 321. [Google Scholar] [CrossRef]
  42. Martoni, L.V.L.; Gomes, N.O.; Prado, T.M.; Calegaro, M.L.; Oliveira, O.N., Jr.; Machado, S.A.S.; Raymundo-Pereira, P.A. Carbon spherical shells in a flexible photoelectrochemical sensor to determine hydroquinone in tap water. J. Environ. Chem. Eng. 2022, 10, 107556. [Google Scholar] [CrossRef]
  43. Maciel, C.C.; de Lima, L.F.; Ferreira, A.L.; de Araujo, W.R.; Ferreira, M. Development of a flexible and disposable electrochemical sensor based on poly (butylene adipate-co-terephthalate) and graphite for hydroquinone sensing. Sens. Actuators Rep. 2022, 4, 100091. [Google Scholar] [CrossRef]
  44. Zhang, H.; Li, S.; Zhang, F.; Wang, M.; Lin, X.; Li, H. Simultaneous detection of hydroquinone and catechol on electrochemical-activated glassy carbon electrode by simple anodic and cathodic polarization. J. Solid State Electrochem. 2017, 21, 735–745. [Google Scholar] [CrossRef]
  45. Feng, S.; Zhang, Y.; Zhong, Y.; Li, Y.; Li, S. Simultaneous determination of hydroquinone and catechol using covalent layer-by-layer self-assembly of carboxylated-MWNTs. J. Electroanal. Chem. 2014, 733, 1–5. [Google Scholar] [CrossRef]
  46. Hu, F.; Chen, S.; Wang, C.; Yuan, R.; Yuan, D.; Wang, C. Study on the application of reduced graphene oxide and multiwall carbon nanotubes hybrid materials for simultaneous determination of catechol, hydroquinone, p-cresol and nitrite. Anal. Chim. Acta 2012, 724, 40–46. [Google Scholar] [CrossRef]
  47. Umasankar, Y.; Periasamy, A.P.; Chen, S.-M. Electrocatalysis and simultaneous determination of catechol and quinol by poly(malachite green) coated multiwalled carbon nanotube film. Anal. Biochem. 2011, 411, 71–79. [Google Scholar] [CrossRef]
  48. Erogul, S.; Bas, S.Z.; Ozmen, M.; Yildiz, S. A new electrochemical sensor based on Fe3O4 functionalized graphene oxide-gold nanoparticle composite film for simultaneous determination of catechol and hydroquinone. Electrochim. Acta 2015, 186, 302–313. [Google Scholar] [CrossRef]
  49. Ziyatdinova, G.; Gainetdinova, A.; Morozov, M.; Budnikov, H.; Grazhulene, S.; Redkin, A. Voltammetric detection of synthetic water-soluble phenolic antioxidants using carbon nanotube based electrodes. J. Solid State Electrochem. 2012, 16, 127–134. [Google Scholar] [CrossRef]
  50. Li, M.; Ni, F.; Wang, Y.; Xu, S.; Zhang, D.; Chen, S.; Wang, L. Sensitive and Facile Determination of Catechol and Hydroquinone Simultaneously Under Coexistence of Resorcinol with a Zn/Al Layered Double Hydroxide Film Modified Glassy Carbon Electrode. Electroanalysis 2009, 21, 1521–1526. [Google Scholar] [CrossRef]
  51. Peng, J.; Gao, Z.N. Influence of micelles on the electrochemical behaviors of catechol and hydroquinone and their simultaneous determination. Anal. Bioanal. Chem. 2006, 384, 1525–1532. [Google Scholar] [CrossRef]
Scheme 1. Schematic representation for the synthesis of CeO2 NPs.
Scheme 1. Schematic representation for the synthesis of CeO2 NPs.
Biosensors 12 00846 sch001
Scheme 2. Schematic picture shows the fabrication and working of CeO2/GCE.
Scheme 2. Schematic picture shows the fabrication and working of CeO2/GCE.
Biosensors 12 00846 sch002
Figure 1. PXRD pattern of the synthesized CeO2.
Figure 1. PXRD pattern of the synthesized CeO2.
Biosensors 12 00846 g001
Figure 2. SEM images (a,b), EDX spectrum (c), and elemental composition (d) of the synthesized CeO2 NPs.
Figure 2. SEM images (a,b), EDX spectrum (c), and elemental composition (d) of the synthesized CeO2 NPs.
Biosensors 12 00846 g002
Figure 3. XPS of CeO2. Survey scan (a), Ce3d (b), and O1s (c) of the synthesized CeO2 NPs.
Figure 3. XPS of CeO2. Survey scan (a), Ce3d (b), and O1s (c) of the synthesized CeO2 NPs.
Biosensors 12 00846 g003
Figure 4. CV graph of GCE (a) and CeO2/GCE in presence (b) and absence (c) of 15 µM HQ in 0.1 M PBS of pH 7.0 (applied scan rate = 50 mV/s).
Figure 4. CV graph of GCE (a) and CeO2/GCE in presence (b) and absence (c) of 15 µM HQ in 0.1 M PBS of pH 7.0 (applied scan rate = 50 mV/s).
Biosensors 12 00846 g004
Figure 5. CVs (a) of CeO2/GCE in presence of HQ (15 µM, 35 µM, 55 µM, 75 µM, 100 µM, 125 µM, 150 µM, 175 µM, 200 µM, and 225 µM) in 0.1 M PBS (pH = 7.0; applied scan rate = 50 mV/s). Calibration plot (b) of peak current versus concentration.
Figure 5. CVs (a) of CeO2/GCE in presence of HQ (15 µM, 35 µM, 55 µM, 75 µM, 100 µM, 125 µM, 150 µM, 175 µM, 200 µM, and 225 µM) in 0.1 M PBS (pH = 7.0; applied scan rate = 50 mV/s). Calibration plot (b) of peak current versus concentration.
Biosensors 12 00846 g005
Figure 6. CVs (a) of CeO2/GCE in presence of 15 µM HQ in 0.1 M PBS (pH = 7.0) at different applied scan rates (50, 100, 150, 200, 250, 300, 350, 400, 450, and 500 mV/s). Calibration curve of peak current versus scan rate (b).
Figure 6. CVs (a) of CeO2/GCE in presence of 15 µM HQ in 0.1 M PBS (pH = 7.0) at different applied scan rates (50, 100, 150, 200, 250, 300, 350, 400, 450, and 500 mV/s). Calibration curve of peak current versus scan rate (b).
Biosensors 12 00846 g006
Figure 7. DPV graph of GCE (a) and CeO2/GCE in presence (b) and absence (c) of 15 µM HQ in 0.1 M PBS of pH 7.0 (applied scan rate = 50 mV/s).
Figure 7. DPV graph of GCE (a) and CeO2/GCE in presence (b) and absence (c) of 15 µM HQ in 0.1 M PBS of pH 7.0 (applied scan rate = 50 mV/s).
Biosensors 12 00846 g007
Figure 8. DPVs (a) of CeO2/GCE in presence of HQ (15, 35, 55, 75, 100, 125, 150, 175, 200, and 225 µM) in 0.1 M PBS (pH = 7.0) (applied scan rate = 50 mV/s). Calibration curve of peak current versus concentration (b).
Figure 8. DPVs (a) of CeO2/GCE in presence of HQ (15, 35, 55, 75, 100, 125, 150, 175, 200, and 225 µM) in 0.1 M PBS (pH = 7.0) (applied scan rate = 50 mV/s). Calibration curve of peak current versus concentration (b).
Biosensors 12 00846 g008
Figure 9. DPVs of CeO2/GCE. (a) Selectivity test: DPVs in presence of 15 µM HQ with (black) and without (red) interfering molecules (glucose, dopamine, H2O2, urea, uric acid, chlorophenol, nitrophenol, acetone, methanol, ascorbic acid, and hydrazine) at scan rate of 50 mV/s (concentration of interfering molecules was 5 times higher than that of HQ). (b) Repeatability test: 5 consecutive cycles of DPVs in presence of 15 µM HQ (scan rate = 50 mV/s). (c) Stability test: 1st, 50th, and 100th DPV cycle in 75 µM HQ (applied scan rate = 50 mV/s).
Figure 9. DPVs of CeO2/GCE. (a) Selectivity test: DPVs in presence of 15 µM HQ with (black) and without (red) interfering molecules (glucose, dopamine, H2O2, urea, uric acid, chlorophenol, nitrophenol, acetone, methanol, ascorbic acid, and hydrazine) at scan rate of 50 mV/s (concentration of interfering molecules was 5 times higher than that of HQ). (b) Repeatability test: 5 consecutive cycles of DPVs in presence of 15 µM HQ (scan rate = 50 mV/s). (c) Stability test: 1st, 50th, and 100th DPV cycle in 75 µM HQ (applied scan rate = 50 mV/s).
Biosensors 12 00846 g009
Table 1. Comparison of LoD of CeO2/GCE with previous reports [42,43,44,45,46,47,48,49,50,51].
Table 1. Comparison of LoD of CeO2/GCE with previous reports [42,43,44,45,46,47,48,49,50,51].
MaterialLoD (µM)Sensitivity (µA µM−1 cm−2)References
CeO2/GCE0.90.41Present study
Carbon spherical shells2.70.0742
poly (butylene adipate-co-terephthalate)/graphite1.04-43
Polarized GCE3.570.13644
Carboxylic functional multi-walled carbon nanotubes2.3-45
rGO/MWNTs/GCE2.6-46
MWCNT–poly-malachite green/GCE1.6-47
AuNPs/Fe3O4/APTESGO/GCE1.1-48
CNTs/GCE2.9-49
LDHf/GCE9-50
GCE8-51
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Chaudhary, A.; Khan, M.Q.; Khan, R.A.; Alsalme, A.; Ahmad, K.; Kim, H. Fabrication of CeO2/GCE for Electrochemical Sensing of Hydroquinone. Biosensors 2022, 12, 846. https://doi.org/10.3390/bios12100846

AMA Style

Chaudhary A, Khan MQ, Khan RA, Alsalme A, Ahmad K, Kim H. Fabrication of CeO2/GCE for Electrochemical Sensing of Hydroquinone. Biosensors. 2022; 12(10):846. https://doi.org/10.3390/bios12100846

Chicago/Turabian Style

Chaudhary, Archana, Mohd Quasim Khan, Rais Ahmad Khan, Ali Alsalme, Khursheed Ahmad, and Haekyoung Kim. 2022. "Fabrication of CeO2/GCE for Electrochemical Sensing of Hydroquinone" Biosensors 12, no. 10: 846. https://doi.org/10.3390/bios12100846

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop