Review

The State of the Art on Graphene-Based Sensors for Human Health Monitoring through Breath Biomarkers

by
Pedro Catalão Moura
,
Paulo António Ribeiro
,
Maria Raposo
and
Valentina Vassilenko
*
Laboratory for Instrumentation, Biomedical Engineering and Radiation Physics (LIBPhys-NOVA), Department of Physics, NOVA School of Science and Technology, NOVA University of Lisbon, Campus FCT-NOVA, 2829-516 Caparica, Portugal
*
Author to whom correspondence should be addressed.
Sensors 2023, 23(22), 9271; https://doi.org/10.3390/s23229271
Submission received: 27 October 2023 / Revised: 13 November 2023 / Accepted: 17 November 2023 / Published: 19 November 2023
(This article belongs to the Special Issue Graphene-Based Sensors: Design, Development and Application)

Abstract

:
The field of organic-borne biomarkers has been gaining relevance due to its suitability for diagnosing pathologies and health conditions in a rapid, accurate, non-invasive, painless and low-cost way. Due to the lack of analytical techniques with features capable of analysing such a complex matrix as the human breath, the academic community has focused on developing electronic noses based on arrays of gas sensors. These sensors are assembled considering the excitability, sensitivity and sensing capacities of a specific nanocomposite, graphene. In this way, graphene-based sensors can be employed for a vast range of applications that vary from environmental to medical applications. This review work aims to gather the most relevant published papers under the scope of “Graphene sensors” and “Biomarkers” in order to assess the state of the art in the field of graphene sensors for the purposes of biomarker identification. During the bibliographic search, a total of six pathologies were identified as the focus of the work. They were lung cancer, gastric cancer, chronic kidney diseases, respiratory diseases that involve inflammatory processes of the airways, like asthma and chronic obstructive pulmonary disease, sleep apnoea and diabetes. The achieved results, current development of the sensing sensors, and main limitations or challenges of the field of graphene sensors are discussed throughout the paper, as well as the features of the experiments addressed.

1. Introduction

1.1. Biomarkers

The medical field is constantly seeking newer technologies and procedures that can enable successful, accurate and rapid diagnostics. The current diagnostic methodologies often rely on interventions that are invasive, painful, time-consuming and, at times, even unsafe to the patient, not to mention the commonly elevated costs and the lack of repeatability of the results [1,2]. To remedy these limitations, newer options have been studied regarding their suitability to diagnose pathologies and health conditions. Among those options, the detection, identification and quantification of biomarkers is gaining relevance [2,3].
A biomarker can be defined as any biological-borne molecule or organic feature, like the temperature, whose characteristics or detection can be indicative of the presence or development of a vast range of health conditions and pathologies [4]. In fact, the identification of biomarkers is often used to assess eventual health risks, to screen health state, to determine prognostics, to evaluate the response to treatments and even to evaluate the progress of a disease [5,6].
The most often studied biomarkers are compounds that are produced by the human organism as an outcome of the development or evolution of a specific pathology or health condition [7,8]. These molecules, once produced, interact with the biological tissues and are commonly released to the exterior of the organism through several body fluids. Breath, urine, faeces, perspiration, blood and even lacrimal fluid are examples of fluids that often carry molecules whose characteristics can make them act as biomarkers for diseases [9,10,11].
Among all the potential molecules studied as potential biomarkers, some of them can be linked to a vast range of conditions while others are studied regarding their exclusive relationship with a single disease. Acetone, for example, has been deeply addressed in the medical community since it can be used as a biomarker for at least 11 pathologies, namely sleep apnoea [12], malaria [13], lung cancer [14], gastric cancer [15], diabetes [16], cystic fibrosis [17], COVID-19 [18], colorectal cancer [19], chronic liver disease [20], chronic kidney disease (CKD) [21], and asthma [22]. Coincidently, the detection of isoprene in human emissions can equally be linked to 11 health conditions, which are chronic obstructive pulmonary disease (COPD) [23], chronic kidney disease [24], chronic liver disease [25], breast cancer [26], cystic fibrosis [27], COVID-19 [28], gastric cancer [29], diabetes [30], malaria [31], lung cancer [32], and sleep apnoea [12]. On the other hand, compounds like 2-acetylpyridine (chronic obstructive pulmonary disease [33]), ethanal (COVID-19 [28]), ethyl acrylate (cystic fibrosis [34]), 2-, 3- and 4-ethyltoluene (squamous cell cancer [35]), 5-ethyl-2-methylheptane (tuberculosis [36]), methyl butyrate (prostate cancer [37]) and many others are relatable with a single disease, as per the moment this paper was published.
Some of the mentioned biomarkers are well known and are even certified for medical use; nonetheless, many other organic-borne molecules still require deeper studies to fully assess their real dependence on a pathology or health condition and judge their suitability for diagnostic purposes. To do so, many scientific studies have been developed around several collection, separation and analysis techniques [38,39,40]. Among these, one can find works developed around ion mobility spectrometry [41], mass spectrometry [42], gas chromatography [43], and, as in for the scope of this paper, sensor array-based electronic noses built around several types of sensors [44], including graphene and graphene derivate-based sensors [45].

1.2. Graphene-Based Sensors

Graphene-based sensors have been largely explored regarding their vast range of applications, their adaptability to distinct scenarios and their proven results in the identification and quantification of specific molecules. Nonetheless, this exploration is only possible due to the characteristics of graphene whose excitability and sensitivity allow its application in the development of molecular electronic devices for sensing purposes.
Initially discovered by Geim et al. (2007) after peeling off the thin (atomic size) layers from pieces of graphite, graphene and graphene-derivative materials have played an important role in sensing-based scientific fields [46]. In fact, graphene is often described, at an atomic level, as the lightest, thinnest and strongest material, which presents hydrophobic behaviour, is aggregable in aqueous solutions, insoluble in organic solvents, and can be stable at temperatures as high as 200 °C [47]. Regarding its structure, graphene presents symmetric arrangements of bonds between carbon, which form a honeycomb pattern with a surface area of around 2600 m2/g [48].
Besides the mentioned features of graphene, it is worth saying that its major advantage is related to the fact that graphene characteristics can be easily altered if exposed to several scenarios. In fact, graphene can be chemically altered through π–π interactions or via electron transfer processes, for example, when exposed to scenarios rich in organic compounds whose functional groups are effortlessly attachable to graphene. In addition, graphene is an electrical conductor whose optical and electrical properties can be modified in several ways that include chemical, electrochemical and thermal options [47,49]. This reactivity and excitability of the graphene and graphene derivates materials make them ideal to be employed in several types of sensors, namely, optical fibre sensors, physical sensors, chemical sensors, electrochemical sensors, and wearable sensors [48,49].
Evidently, each one of these types of sensors can be assembled through different techniques. One of the preparation methodologies of graphene-based sensors often used to analyse gaseous or volatile samples is the layer-by-layer technique. Here, alternated electrically charged thin films of polyelectrolytes are applied to a solid base containing interdigitated electrodes [50]. Examples of those polyelectrolytes are polyethyleneimine and, evidently, graphene oxide. The application consists of successive immersions of the base in the solutions. Once produced, the sensors can, then, be exposed to the volatile samples targeted during the experiment and, specifically, exhaled air samples. As addressed, there are several types of sensors and several ways of utilizing them. One of the options is reported by Moura et al. (2023) [50]. In their work, the authors registered the electrical response of the sensors through the impedance variance of the sensors when exposed to the target volatile organic compounds. Further details regarding the operating principle of the several types of graphene-based sensors can be found elsewhere [49]. Figure 1 schematizes a generic system based on graphene sensors to analyse exhaled breath through the impedance variation. Here, a graphene oxide-based sensor is exposed to the analytes emitted in breath and the impedance-frequency relationship is registered by a dedicated software. The graph illustrates the registered curves, in different colours, of two random compounds.
As addressed, each atom of graphene and graphene derivates is extremely sensitive and can easily be excited and react once exposed to a vast and varied range of scenarios that extend from environmental evaluation to health monitoring and assessment. These applications include examples as mixed as the study of industrially relevant alcohols [50], monitoring of volatile organic compounds emitted during extreme wildfires [51], identification of organic contaminants in aqueous matrices [52,53], detection of indoor air pollutants [54,55], and many other applications [56,57,58]. Among these topics, medical applications have been gaining relevance and importance in the field of graphene-based sensors, namely, the detection, identification and quantification of organic-borne biomarkers whose detection can represent an open window to the interior of the human organism and represent a valuable source of information on the health condition [45,59].

2. Graphene Sensors for Biomarker Detection

Graphene-based sensors have gained relevance in the field of the detection, identification and quantification of compounds for medical purposes and, specifically, for diagnostics of pathologies via organic-borne biomarkers. In fact, the number of scientific publications on the topic has presents an evident positive rate of increase, as shown in Figure 2. Here, the number of publications indexed in one of the main scientific databases of publications, Web of Science, under the scope “Graphene Sensors” and the scope “Graphene Sensors and Biomarkers”, proves the increasing interest in this field and the growing dedication of the scientific community to the development and application of graphene-based sensors in the assessment of organic-borne biomarkers.
Having in mind all the facts and information previously addressed, one can deduce the pertinence of assessing the state of the art of the field. In this way, this review aims to share the current level of development in the field of graphene-based sensors for detecting, identifying, and quantifying volatile organic compounds present in the exhaled breath that, due to their organic origin, may act as biomarkers of specific pathologies and health conditions. To do so, the most relevant papers published under the scope mentioned above were considered and consulted for this review. The research was limited to the last 10 years of publication, i.e., 2012–2022. More than 3000 papers were initially gathered. Review papers were disregarded and the most relevant of the cohort were considered and reviewed in this article.
During the bibliographic search, one could realize that graphene-based sensors have been developed and employed to identify biomarkers in the breath of patients suffering from six specific pathologies. These six health conditions are represented in Figure 3. They are lung cancer, gastric cancer, chronic kidney diseases, respiratory diseases that involve inflammatory processes of the airways, like asthma and COPD), sleep apnoea and diabetes.

2.1. Asthma and COPD

Asthma, chronic obstructive pulmonary disease and other health conditions that involve inflammatory processes of the respiratory tract are among the most common and rather mortal pathologies worldwide. In fact, the World Health Organization (WHO) has reported a staggering number of 500,000 deaths directly caused by asthma, and 3 million deaths provoked by COPD worldwide every year [60,61]. These kinds of pathologies impose a very poor quality of life on the patients, whose state depends directly on the rapidness and effectiveness of the treatments [62,63]. In order to prescribe the proper treatment procedures, the physicians have to accurately diagnose the pathology in question; however, the current methodologies to do so are often time-consuming, expensive, invasive and very discomforting to the patients [64,65].
The detection, identification and quantification of volatile organic compounds in the breath of patients has gained relevance as a rapid, accurate, non-invasive, painless and low-cost procedure to identify inflammatory pathologies of the respiratory tract. Compounds such as acetone [22], decane [66], propanol [67], hexane [67], dodecane [66], ethylbenzene [68] and many others have been linked to the diagnosis of asthma when detected in exhaled air. A similar logic can be applied to the diagnostic of COPD via the identification of acetaldehyde [33], benzaldehyde [23], benzene [23], butanal [33], isoprene [23], isopropanol [33], limonene [69], nonanal [70], and many other analytes in breath.
Besides the potentialities of the biomarkers in breath, one of the current challenges to be surmounted is the lack of standardized procedures for the detection of those analytes. The field of graphene oxide-based sensors has positioned itself as one of the potential candidates to help overcome this issue. In fact, several works have been developed in the area [45,71].
Aiming to study one specific marker for inflammatory processes of the respiratory tract, nitrite, Gholizadeh et al. (2017) developed graphene oxide-based sensors that were later tested in two scenarios. Initially, the authors exposed the sensors to standard nitrile solutions prepared with different concentrations; then, the sensors were tested with samples of exhaled air condensate. The variation in the impedance of the sensors during the tests was assessed through impedance spectroscopy. In both scenarios, the authors were able to characterize the optimal operation conditions for the sensors, as well as successfully identify nitrite in both standard and exhaled air samples [72]. In a follow-up study, the authors were able to validate the results, showing the appropriateness of graphene sensors for medical applications and, in specific, for the identification of biomarkers for the diagnostic of inflammatory conditions of the respiratory tract [73].
The field of graphene oxide-based sensors has been making its contribution to the development of systems for the analysis of breath samples. The work of Kumar et al. (2020) is an example of that. The authors have developed an array of sensors based on graphene oxide composite to study three specific volatile compounds, ammonia, ethanol and the potential asthma biomarker, acetone. Standard samples of acetone were previously prepared with concentration levels of 1000 and 2000 ppmv. The developed sensors were capable of detecting, identifying and quantifying acetone successfully, proving their suitability for the eventual analysis of breath samples and asthma diagnosis [74].
Another well-known asthma biomarker is propanol [67]. This volatile organic compound has been studied through sensors of graphene oxide in several scientific studies. One of those works was developed by Samadi et al. (2021). The authors developed sensors capable of detecting propanol at room temperature by scattering thin layers of a specific nanocomposite, ZnO@SiO2/rGO. Then, the sensor was tested by being exposed to standard samples of propanol previously prepared with concentration levels ranging from 150 ppmv to 450 ppmv. Propanol was successfully identified and quantified, proving the suitability of this kind of system for the assessment of biomarkers in breath [75].
Several works have reported a direct connection between the concentration levels of hexane in exhaled air, and the existence of inflammatory processes in the respiratory tract and, specifically, of asthma. Graphene-based sensors have been tested to assess their potential for hexane detection. Hussein et al. (2019) developed gas sensors by scattering thin layers of reduced graphene oxide (rGO) that were later used to detect three specific analytes, chloroform, ethanol and the potential asthma biomarker already mentioned, hexane. The tested volatile samples were previously prepared with known concentration levels of the target compounds. Hexane was successfully detected and quantified, proving that the system developed by the authors can eventually be tested and applied to real samples of exhaled air [76].
In order to detect the mentioned compounds in breath, some work has been performed regarding the graphene oxide sensing capacity. Murashima et al. (2016), for example, addressed the capacities of graphene-based sensors for the qualification and quantification of one of the aforementioned biomarkers, i.e., acetaldehyde. To do so, the authors developed the sensors by scattering thin films of graphene oxide that were later exposed to gaseous samples of acetaldehyde. The samples were prepared in 10 to 50% ratios between the target analyte and room air. All the samples were successfully detected, proving that the sensors developed by the authors can be utilized for clinical purposes [77].
As mentioned, another common biomarker of COPD is benzaldehyde. Coincidentally, Khan et al. (2015) studied this specific analyte with graphene oxide-based sensors. In fact, the authors developed an array of sensors that were later exposed to the standards of the target analytes. The authors were able to achieve sensitivity levels of ~1.2277 µAcm−2µM−2, proving the outstanding capacity of the developed gas sensors for the sensing of acetaldehyde [78].
Isoprene, a COPD biomarker often studied in breath, was successfully detected by an array of graphene oxide-based sensors developed by Chen et al. (2019). Authors were able to sense isoprene with limits of detection as low as 237 ppbv in the volatile emissions of several types of food and, specifically, fruits. Besides that, considering the outstanding detection limits, their system proved their suitability and eventual applicability for exhaled air analysis and biomarkers identification [79].
A final example of a biomarker usually detected in breath and often linked to COPD is nonanal, as mentioned. With that in mind, Liu et al. (2019) developed an electronic nose based on chemoresistive sensors of graphene that was later used to analyse some specific analytes, including nonanal. The other studied volatile organic compounds were ethanol, 2-ethoxyethanol and ethylbenzene. The authors aim to develop more versatile and sensitive sensors; nonetheless, the achieved results leave no doubts regarding the pertinency of graphene-based sensors in the field of health monitoring and disease diagnosis [80].
Thousands of deaths occur every year directly due to respiratory pathologies and, specifically, due to asthma and chronic obstructive pulmonary disease. A considerable portion of these deaths could be avoided if a faster and more accurate diagnosis was possible. In this way, the scientific community has focused its efforts on developing electronic noses based on arrays of graphene sensors to detect organic-borne biomarkers that can act as biomarkers for the mentioned pathologies. As seen in the reviewed works, the community is working to develop more accurate, standardized systems to analyse human breath and the results already achieved are, undoubtedly, promising and worthy of attention.
Table 1 includes all the details on the procedures adopted in each one of the studies addressed in this section, including the target biomarkers and respective references on the Human Metabolome Database (HMDB), the considered populations, the developed sensors, and the respective bibliographic source.

2.2. Chronic Kidney Diseases

A fast and accurate diagnostic is especially mandatory in the case of chronic kidney diseases. This group of pathologies is responsible for leading thousands of people to hospitals every year due to direct complications of the pathology but also due to all the secondary consequences and comorbidities [81]. When identified in later stages of development CKD often results in mandatory haemodialysis treatment, acute renal failures and even life-threatening cardiovascular episodes [82,83].
Considering the aforementioned facts, it is crucial that newer, faster and more accurate procedures to diagnose CKD are developed. Electronic noses, whose working principle is based on the sensing capacities of graphene-based sensors, have been developed to tackle this challenge by qualifying and quantifying the analytes emitted in the breath of the patients that can act as biomarkers of CKD [84]. It is worth stating that several scientific studies have been published regarding this issue and whose focuses are given to known CKD biomarkers, namely, ammonia [85], acetone [86], ethanol [87], and isoprene [24], among others.
Aiming to assess the suitability of graphene-based sensors for the detection of volatile organic compounds often detected in breath, Lee et al. (2021) developed and validated reduced graphene oxide-based sensors and machine learning algorithms. The system was then tested with three specific volatile organic compounds, isopropanol, acetone and ammonia. As mentioned, ammonia is a common biomarker of chronic kidney diseases in the exhaled air so, with that in mind, authors exposed the sensors to several standard samples of the analyte, at specific concentration levels, and were able to identify it with accuracy levels above 90%. They were equally capable of differentiating health volunteers from chronic kidney disease patients, gathered in a synthetic cohort of volunteers, with accuracy levels above 20% [88].
Having the goal of detecting biomarkers of kidney pathologies in exhaled air, Majidi et al. (2021) developed and tested nitrogenated holey graphene sheet-based sensors. The target analytes of the authors were isoprene, pentanal, hexanal and heptanal, analytes commonly associated with kidney diseases. In this way, the authors studied the adsorption of these molecules on the surface of the developed sensors. They were able to successfully detect the analytes, proving the suitability of their system for further applications in the medical field and, in specific, for breath analysis [89].
Two specific volatile compounds commonly related to the diagnosis of chronic kidney diseases are acetone and ethanol. With that in mind, Tung et al. (2020) developed graphene-based chemoresistive sensors that were later used to detect specific volatile organic compounds, including acetone and ethanol. Other analytes equally relevant to the field of carcinogenic biomarkers were also tested; they are methanol, chloroform, acetonitrile and tetrahydrofuran. The developed array of sensors could detect the target compounds in limits of detection as low as 2.82 ppbv, outstanding results that undoubtedly show the suitability of graphene-based systems for lung cancer diagnosis via biomarkers in breath [90].
Acetone was also the target compound of the study developed by Choi et al. (2014). As stated by the authors, human exhaled air has tremendous advantages and a high potential to act as a valuable source of information about the organism and about specific pathologies and health conditions. In this way, and being aware of the relationship between the presence of acetone in breath and an eventual diagnosis of conditions like diabetes or gastric cancer, among others, the authors developed an array of sensors using WO3 hemitubes functionalized by graphene-based electronic sensitizers. To test the sensors’ performance, solutions of acetone were prepared with 1 ppmv concentration. The sensors were then exposed to the solution and their response was assessed. A full detection of the target VOC was achieved with limits of detection as low as 100 ppbv and with response times ranging between 11.5 and 13.5 s. The outstanding sensitivity and overall behaviour of the sensors leave no doubts about their suitability for real medical scenarios [91].
As mentioned above, CKD often result in mandatory haemodialysis treatments. Several procedures have been developed aiming to identify biomarkers in breath to assess the evolution and impact of the treatments in CKD patients [92,93].
Ammonia is a well-known volatile organic compound that has been studied regarding its direct connection to haemodialysis treatments. This compound occurs naturally in the human organism since it is produced by protein metabolism and is often excreted through urine. Nonetheless, it can traverse biological tissue and is often emitted through breath after being transported to the lungs via the circulatory system [94].
Aiming to identify ammonia in the exhaled air of patients undertaking haemodialysis treatment, Shahmoradi et al. (2021) fabricated graphene-based sensors, namely, sulfonate graphene-, graphene oxide- and reduced graphene oxide-based sensors. Then, the authors exposed the produced sensors to previously prepared gaseous samples of ammonia whose concentrations ranged from 0.5 ppbv to 12 ppmv, similar concentration levels to the ones commonly found in the exhaled breath of haemodialysis patients. The system proved to be suitable for an accurate and sensitive detection of the target analyte in a non-invasive and painless way. The results prove the auspicious future of graphene-based sensors in the field of detection of biomarkers for the evaluation of the organism’s reaction to haemodialysis treatments and to diagnostic renal disease overall [95].
A final example of the work being developed in the field of graphene-based sensors for the detection, identification and quantification of biomarkers in breath is the study of Yabaş et al. (2023). To create the sensors, the authors synthesized 4-pyridynyl-oxadiazole tetrasubstituted zinc and cobalt phthalocyanine compounds that were, then, mixed with reduced graphene oxide to obtain hybrids, namely, rGO, rGO/ZnPc and rGO/CoPc. The performance of the array of sensors was tested against a total of five volatile compounds. Among these were ethanol and acetone, biomarkers for gastric cancer, and ammonia, a biomarker of haemodialysis treatment. Concentration levels ranging between 30 and 210 ppmv were considered during the study, and response times of 190, 230 and 250 s were achieved, respectively, to rGO/CoPc, rGO/ZnPc and rGO-based sensors [96].
Chronic kidney diseases are serious conditions that affect the quality of life of the patients extremely, not only due to their own consequences but also to secondary comorbidities. All these consequences can be grandly diminished if the diseases are diagnosed in an initial stage of development. The detection of biomarkers in breath with graphene-based sensors has gained relevance in the past years as a rapid, accurate and non-invasive way of diagnosing CKD. The achieved results show that, once the procedures are standardized, all the target analytes are defined and the detection limits are improved, the field of graphene-based sensors and biomarkers will have a central role in modern medicine.
Table 2 includes all the details on the procedures adopted in each one of the studies addressed in this section, including the target biomarkers and respective references on HMDB, the considered populations, the developed sensors, and the respective bibliographic source.

2.3. Diabetes

Diabetes is among the most common pathologies worldwide, with notable incidence in developed countries. The WHO estimates that the number of patients suffering from diabetes will reach the humongous number of 350 million cases by the year 2030. Coincidentally, a considerable portion of the cases are diagnosed in a later stage of development, preventing a proper treatment and leading to relevant and, in some cases, life-threatening comorbidities [97,98].
The current procedures for monitoring blood glucose, besides being effective, are often invasive, complicated, and even expensive, so medical academia is constantly seeking new and non-invasive procedures that allow us to control diabetes accurately and rapidly [99]. In order to tackle this limitation, the scientific and medical communities have given full attention to the detection and identification of volatile organic compounds emitted in the breath that can act as biomarkers for the diagnosis of diabetes [100]. The main diabetes biomarkers are acetone [101], methanol [102], ethanol [102], isoprene [30], isopropanol [101] and others.
In order to help the medical field detect and identify the compounds emitted in breath, the scientific community investigating graphene-based sensors has placed their efforts in developing novel sensors, devices and even methodologies that allow a full characterization of the emitted analytes [103].
A direct example of the utility of graphene-based sensors can be found in the work of Kalidoss et al. (2019). In one of the first papers on the matter, authors developed gas sensors based on a ternary (graphene oxide, tin dioxide and titanium dioxide) nanocomposite for the detection of acetone in the breath of diabetic patients. Then, the authors exposed the developed sensors to several concentration levels of the analyte and fully characterized the behaviours of the sensors, namely, their response and recovery times, their ideal operating temperature, and even their sensitivity. The results achieved by the authors prove the promising future of this type of sensor [104].
During a second work, the authors developed and tested graphene-based chemoresistive sensors with the single purpose of detecting acetone in exhaled air samples. To do so, a prototypic device was developed around the array of sensors and used to analyse breath samples of 17 diabetic patients and 13 healthy volunteers. The authors claim that their system allows the differentiation among both groups of volunteers with an accuracy of over 60%, proving the suitability of graphene-base sensors for this type of application. Additionally, the authors intend to keep developing their prototype to the level of becoming a single lab-on-chip pocket device [105].
In a third work developed by the same research group, the authors used the data provided by the electronic nose developed around the array of graphene-based sensors, to develop an algorithm capable of increasing the already-good levels of accuracy of the sensors. During the study, exhaled air samples from a cohort of 60 volunteers (30 diabetic patients and 30 healthy subjects) were analysed with the device. The new approach to the data enabled the authors to increase the accuracy by up to 70% [106].
Acetone was also the biomarker of diabetes targeted by Thakur et al. (2022). An array of six sensors comprising hybridized graphene oxide field-effect transistors was developed by the authors specifically for the detection of acetone. To do so, the array was exposed to a dummy breath, i.e., a synthetically prepared breath whose purpose was to mimic the human exhaled air. The samples were prepared with synthetic air and specific portions of acetone whose concentration levels ranged between 400 ppbv and 80 ppmv. Additionally, six other volatile compounds (formaldehyde, toluene, benzene, propanol, ethanol and methanol) were added to the mixture to act as interfering compounds during the analyses. Accuracy levels of 100% were achieved by the authors during the identification of acetone with the developed sensors [107].
In a direct application of graphene oxide nanosheets, Choi et al. (2014) focused their work on the sensing capacities of graphene oxide and developed gas sensors to detect the presence of acetone in the exhaled air of diabetic patients. Samples of acetone, previously prepared with concentration levels ranging between 1 and 5 ppmv, were used as target materials. The sensors were exposed to the samples and their behaviour was analysed. The authors state that the developed sensors enabled the detection of acetone with high levels of selectivity and limits of detection as low as 100 ppbv. This value proves the suitability of the described procedure to perform analyses of exhaled air, since the concentration of acetone in breath is often superior to 100 ppbv [108].
In a very recent study, Sen et al. (2023) developed sensors based on reduced graphene oxide nanocomposites to detect acetone in the breath of a diabetic patient. The electrical response of the sensors exposed to the target analyte was studied through impedance spectroscopy. The capacity to detect concentration levels lower than 1 ppmv and the excellent sensing response (response and recovery times of 10 and 30 s, respectively), prove the suitability of the assembled graphene-based sensors for the field of breath assessment [109].
Quantum resistive gas sensors to detect three specific analytes, acetone, methanol and ethanol, were prepared by Yempally et al. (2020) via the scattering of thin layers of graphene. To test the behaviour of the developed sensors when exposed to the target compounds, a mixture of the three compounds was previously prepared. The authors were capable of fully detecting and differentiating all the analytes with detection limits as low as 1 ppmv, exhibiting the potential of their system for real-scenario applications [110].
A final example of the suitability of graphene-based sensors for the detection of acetone in the breath of diabetic patients was developed by Rakkesh et al. (2022). The authors used a revolutionary microwave-assisted chemical reduction technique to extract layers of graphene from natural materials, specifically, coconut shells. Then, the synthesized nanostructure was evaluated regarding its capacity to sense acetone. Pure samples prepared with different concentration levels of the target compound were tested during the study for distinct conditions of analysis, namely, temperature and humidity conditions. The authors stated that the sensors enabled a successful detection of acetone and exhibited a response time/recovery time ratio of 1.11/41.25 s, proving the promising future of this field [111].
Interestingly, diabetes is among the health diseases most studied regarding the identification of biomarkers in exhaled air with graphene-based sensors. As for the other reviewed conditions, some challenges must be overcome, namely, the lack of standardized methodologies and the detection limit of the sensors that often fail to achieve concentration levels present in breath (low ppmv–ppbv). Once solved, this field will play a relevant role in the diagnosis and monitoring of diabetes in the near future.
Table 3 includes all the details on the procedures adopted in each one of the studies addressed in this section, including the target biomarkers and respective references on HMDB, the considered populations, the developed sensors, and the respective bibliographic source.

2.4. Gastric Cancer

The diagnostic of gastric cancer often involves a gastric endoscopy with a complementary biopsy and subsequent identification through histopathological analysis. As is known, this type of procedure is extremely invasive and leads the patient to scenarios of extreme discomfort. Additionally, the results often require some time to be available causing a delay in the treatment procedures. In this way, the development of non-invasive, rapid and accurate techniques for gastric cancer diagnosis is required to allow proper treatment of the disease as rapidly as possible [112,113].
Aiming to accelerate the diagnosis of the pathology, the identification of gastric cancer biomarkers in breath has gained pertinency [114]. One can find several works whose scopes were dedicated to the identification of biomarkers; they are acetic acid [115], acetone [116], 2-butanone [117], isoprene [115], propanal [115], phenyl acetate [116], furfural [118], toluene [119], and many others.
To overcome the challenges of detection and analysis of volatile organic compounds, the field of electronic noses based on graphene oxide sensors has played an important role. A considerable number of works have been published regarding the capacity of graphene for the sensing of breath biomarkers [120].
A total of 14 volatile organic compounds present in exhaled air samples were studied by Chen et al. (2016), during their search for gastric cancer biomarkers. To do so, the authors developed gas sensors based on the scattering of graphene oxide thin films that were later exposed to a cohort of 200 simulated breath samples of healthy volunteers (56 subjects), early-stage gastric cancer (55 subjects) and advanced-stage gastric cancer (89 subjects). The authors were able to successfully differentiate the samples with levels of sensitivity above 80%. As mentioned, a total of 14 biomarkers were studied during the study, namely, 2-methylhexane, 2-methylpentane, 3-methylpentane, dodecane, tetradecane, menthol, phenyl acetate, hexanol, pivalic acid, 3-methylhexane, 2,3-dimethylpentane, hexane, isoprene and acetone [116].
Acetic acid was one of the analytes targeted by Moura et al. (2023) in their recent study on volatile compounds. The authors developed graphene oxide sensors based on thin films scattering for the purpose of sensing four specific analytes, methanol, isopropanol, ethanol and acetic acid. Then, the authors exposed the sensors to samples of acetic acid ranging from 24 to 120 ppmv and successfully detected and quantified all the samples. The entire procedure and, specifically, the resolution of 0.04 ppmv achieved by the sensors prove their eventual applicability to analyse exhaled air samples [50].
Another example of the suitability of graphene oxide-based sensors for breath assessment is the work of Jia et al. (2022). Aiming to identify and quantify acetone in breath, the authors synthesized Fe3O4/rGO composites through a simple hydrothermal reaction that was later used to develop the sensors. Then, the array was exposed to samples of acetone previously prepared at several concentration levels (maximum of 800 ppmv). The main goal of the authors was to identify acetone as a diabetes biomarker in breath, and that was completely achieved; nonetheless, considering that acetone can equally act as gastric cancer biomarkers, the results are auspicious and also promising to this field of research [121].
As mentioned, acetone, isoprene and toluene are well-studied analytes often detected in breath, whose levels of concentration can be indicative of carcinogenic pathologies and, specifically, gastric cancer. Being aware of this fact, Tombel et al. (2021) decided to assess the suitability of graphene-based sensors to study the mentioned analytes. In this way, authors integrated reduced graphene oxide on Ti/Pt interdigitated electrode deposited on a SiO2/Si substrate that was later exposed to acetone, isoprene and toluene. The standard solutions used as samples were previously prepared with concentration levels in the ppmv range (up to 6 ppmv). The sensors showed outstanding capacities for detecting all the compounds; in fact, the identification of acetone, toluene and isoprene was achieved with R-squared values of 0.9312, 0.0093 and 0.9388 [119].
The diagnostic of gastric cancer, as mentioned in the initial portion of this chapter, often requires invasive and painful procedures. The field of graphene-based sensors has made a contribution to overcoming these issues. The detection of biomarkers in the breath of gastric cancer patients, in this way, has grown as a potential solution. Once overcome limitations like the lack of standardized procedures, the lack of portability of the developed systems and the full identification of the target analytes, graphene-based sensors can be, without doubt, a useful tool for the medicine of the future.
Table 4 includes all the details on the procedures adopted in each one of the studies addressed in this section, including the target biomarkers and respective references on HMDB, the considered populations, the developed sensors, and the respective bibliographic source.

2.5. Lung Cancer

Due to its direct connection to the respiratory system, lung carcinoma has been deeply studied regarding the possibility of a faster, more accurate, non-invasive and painless diagnostic that would allow a proper and effective treatment in expedited time. This is, in fact, one of the main necessities of current medicine, accurately and rapidly diagnosing lung cancer [122,123]. The high levels of incidence allied to the mortality of the pathology make it one of the most concerning conditions worldwide. In fact, every year, more than 2 million new cases are diagnosed on the planet [124].
Several approaches, procedures and methodologies have been used to detect lung cancer. The detection of biomarkers in the exhaled air is one of them [125,126]. One can list several analytes often linked to pulmonary carcinogenic conditions; they are heptane [127], hexanal [128], pentane [32], 2-butanone [129], furan [130], decane [14], acetone [32], isoprene [131], ethanol [132], or even formaldehyde [133], among many others. To supplant the issues of analysing and identifying the volatile analytes emitted in breath, the field of graphene oxide-based sensors has given its contribution by developing innovative electronic noses based on the sensing capacities of graphene [134]. In fact, the literature provides some examples of that contribution.
One of the main examples of the utilization of systems based on graphene-oxide sensors for the analysis of breath in the search for lung cancer biomarkers is the work of Emam et al. (2020). A total of nine volatile compounds (butyraldehyde, tetrahydrofuran, acetonitrile, heptane, hexanal, benzene, 2-butanone and furan) were analysed with electrochemical gas sensors developed via the deposition of thin layers of graphene and Prussian blue on a chromium-modified silicon substrate. To do so, the authors developed an entire system around the array of sensors that registered the resistance variation on the sensors once exposed to the target analytes. These variations were, then, saved for consultation in a mobile application specifically developed for this project. The tests were performed using pure standards of the nine biomarkers, a procedure that allowed the authors to certify the correct operability and suitability of their system for clinical purposes [135].
Graphene-based sensors were also the basis of the analytical technique employed by Shanmugasundaram et al. (2022) to study two specific compounds, decane and heptane. These compounds are well-known biomarkers often detected in the breath of lung cancer patients [6]. To detect them, the authors developed a methodology based on SnO2 nanospheres and a reduced-graphene-oxide-incorporated SnO2 nanocomposite. Then, the sensors were exposed to standard samples of the target gases, allowing the authors to fully assess the behaviour of the developed system during the exposure. The sensors proved to be capable of detecting concentration levels as low as 1 ppmv and the authors intend to test them in real samples of breath [136]. Coincidently, decane was also studied by Zhang et al. (2016). To do so, the authors developed surface acoustic wave gas sensors coated with thin films of graphene oxide and exposed them to standard samples of the target analyte. The authors were able to detect this well-known biomarker with sensitivity levels as low as 0.2 ppmv, proving the suitability of the technique for breath analyses [137].
An electronic nose was developed by Chen et al. (2020) to analyse breath samples and detect potential biomarkers for lung cancer in human exhaled air. The electronic nose consisted of an array of chemoresistive graphene oxide-based sensors whose resistance variation was assessed when exposed to real samples of breath. A cohort of 108 subjects was considered for the study and, with the developed E-nose the authors could differentiate both groups with sensitivity and specificity levels of 95.8% and 96.0%, respectively. Acetone, isoprene and ammonia were the analytes addressed as potential lung cancer biomarkers during the study [138].
Acetone and ethanol were equally addressed as potential biomarkers for lung cancer diagnosis through breath analysis by Sánchez-Vicente et al. (2020). Here, the authors used graphene-doped tin oxide nanofibres and nanoribbons as gas sensors to analyse previously developed synthetic samples of breath. The synthetic samples were prepared with known concentrations of acetone and ethanol, ranging between 0.5 and 4 ppmv, to simulate both healthy individuals and pathological volunteers. The developed system could successfully differentiate both groups with outstanding capacity, proving that the authors can now test it using real samples of exhaled air [139].
A final biomarker often linked to the diagnosis of lung cancer is formaldehyde, as mentioned. In order to assess its real suitability for helping to diagnose lung cancer in a rapid, painless, noninvasive and accurate way, Shanmugasundaram et al. (2022) developed an electronic nose whose operating principle was based on reduced graphene oxide superstructures. The authors simulated exhaled air samples of healthy people and lung cancer patients by preparing solutions of formaldehyde with specific concentration levels, namely, 49 ppbv to healthy samples and 83 ppbv to pathological samples. Once exposed to the cohort, the developed sensors were capable of fully differentiating between both groups with pinpoint accuracy. This outstanding evidence proves the suitability of graphene oxide-based systems to diagnose lung cancer through biomarkers in breath [140].
As addressed, several works have dedicated their scope to the development of graphene-based sensors for the detection of biomarkers in the breath of lung cancer patients. Nonetheless, several challenges have to be overcome, namely, the improvement in the detection systems, the standardization of the collection and analysis procedures, and the improvement in the detection limits and sensitivity of the sensors. All in all, the results already achieved prove that the future of lung cancer assessment through graphene sensors is auspicious and should not be disregarded.
Table 5 includes all the details on the procedures adopted in each one of the studies addressed in this section, including the target biomarkers and respective references on HMDB, the considered populations, the developed sensors, and the respective bibliographic source.

2.6. Sleep Apnoea

It is estimated that a staggering number of 936 million adults suffer from sleep apnoea worldwide [141]. This pathology is characterized by profound alterations in the breathing rhythm, i.e., variations between breathing and non-breathing periods during sleep due to the collapse of the airways. The patients are, then, exposed to a vast range of commodities and consequences from this successive oxygenation interruption [142,143].
As with other respiratory pathological conditions, sleep apnoea has been studied regarding eventual biomarkers emitted in the breath that can lead to a rapid, accurate and non-invasive diagnosis. The main analytes often linked to sleep apnoea are acetone [12], decanal [12], heptane [12], hexane [12], nonane [12], octane [12], β-pinene [144], toluene [12], and p-xylene [12], among some others. Due to the current challenges in the detection of these analytes in samples of breath, several procedures have been developed around an array of sensors based on the sensing capacities of a specific nanocomposite, graphene [145].
As mentioned above, toluene is one of the biomarkers often found in the breath of sleep apnoea patients. Coincidently, toluene was one of the volatile organic compounds studied by Casanova-Chafer et al. (2021). To do so, the authors based their work on the sensing capacities of nanohybrid composites comprising graphene loaded with perovskite nanocrystals. Then, the array of sensors was exposed to the standard solutions of toluene prepared in concentration levels of 2, 4, 6 and 8 ppmv. The behaviour of the sensors was assessed for all four situations, and they left no doubts about their suitability for the detection of gaseous samples of toluene and, consequently, their applicability for exhaled-air study [146].
Once again, acetone has shown itself as a common compound commonly present in the exhaled air and often linked to health conditions. Sleep apnoea is no exception cit. With that in mind, Sen et al. (2021) developed an array of sensors based on ZnO-SnO2 nanocomposites decorated with reduced graphene oxide. Then, the authors tested the sensors regarding their sensing capacities by exposing them to solutions of acetone prepared at concentration levels ranging between 1 and 10 ppmv. The outstanding accuracy of 91% and detection limits of 0.675 ppmv prove not only the perfect capacity for volatile acetone detection but also the suitability for the identification of biomarkers in breath [147].
Aiming to develop a wearable sensor for breath monitoring, Caccami et al. (2017) turned to the sensing capacities of graphene and developed graphene oxide-based sensors capable of adhering to the skin and continuously monitoring the exhaled air of the volunteer. The authors focused their work not on the variation of the sensors’ resistance when exposed to specific analytes, but on the respiratory (inspiration/expiration) cycles to assess conditions like apnoea and tachypnoea [148]. In a follow-up study, Caccami et al. (2018) were able to test their graphene-based sensors in 10 volunteers who were asked to reproduce a predefined set of normal and abnormal respiratory rhythms. Again, authors were able to prove the suitability of sensors based on graphene to fully characterize the exhaled breath and, consequently, assess diseases and health conditions like apnoea [149].
Isopropanol, also known as isopropyl alcohol, is another biomarker of sleep apnoea whose detection has been studied with graphene-based sensors. An example of that is the work of Ray et al. (2022). Aware that the breathomics is the future of non-invasive, painless, accurate and rapid medicine, the authors developed graphene-based sensors that were then used to analyse real breath samples. The breath samples were collected and spiked with concentrations of 0.5 ppmv of well-known volatile organic compounds: acetone, acetaldehyde, butanediol, cyclohexanone, decane, ethanol, methanol, octane, styrene, propyl benzene and, as previously addressed, the sleep apnoea biomarker, isopropanol. Outstanding levels of accuracy (ranging between 92.8 and 96%). Considering the achieved results, there is no doubt about the total suitability of graphene-based sensors for the identification of biomarkers in breath [150].
Without question, the most common pathology in the world from six reviewed in this paper, sleep apnoea is responsible for the lack of life quality of almost 1 billion people. Aiming to improve the life quality of all these people, academia has focused on developing newer electronic noses based on arrays of graphene sensors, which allow the diagnosis and monitoring of sleep apnoea in an accurate, rapid, non-invasive, painless and low-cost way. As seen throughout the chapter, these systems still present some challenges to overcome, namely, the lack of standards for the collection and analysis of breath samples, the lack of detection limit capable of equalizing the concentration levels often found in exhaled air, and evidently, the certification of all these technologies for real medical applications. Nonetheless, the future looks promising, and modern medicine will rely, once again, on the achievements of academia.
Table 6 includes all the details on the procedures adopted in each one of the studies addressed in this section, including the target biomarkers and respective references on HMDB, the considered populations, the developed sensors, and the respective bibliographic source.

3. Conclusions

Medicine of the future will deeply rely on modern methods of diagnosis that allow the identification of pathologies and health conditions in faster, more accurate, painless and non-invasive ways, to the detriment of the current procedures that often involve risks, pain, long periods of waiting and failure to offer repeatable results. The detection, identification and quantification of biomarkers emitted in the exhaled air of patients suffering from the target pathologies is a field of growing interest and one whose past results prove its pertinency.
In order to fully use the potentialities of breath biomarkers, accurate, standardized and sensitive detection systems are required. Nonetheless, not all the analytical techniques and electronic noses available have the necessary features to analyse such a complex matrix as it is the human breath. With this in mind, the scientific community has devoted considerable efforts to developing new systems based on the sensing capacities of a specific nanocomposite, graphene, for the qualification and quantification of organic-borne biomarkers.
The present review aimed to gather the most relevant published papers under the scopes “Graphene-based sensors” and “Biomarkers”, in order to assess the state of the art of the field, identify the developments achieved during the past years, and spot the main pathologies under exploration regarding the suitability of graphene sensors to detect biomarkers. A total of 3000 publications were initially considered and once reduced to the most relevant ones, one could identify a total of six main areas of work. They are lung cancer, gastric cancer, diabetes, sleep apnoea, chronic kidney diseases and respiratory conditions like asthma and chronic obstructive pulmonary disease.
Considering all the reviewed works, one can state that impressive results have been achieved in the field of graphene sensors for breath analysis. A considerably elevated number of systems have been successfully developed and tested both in laboratory scenarios and with real samples of exhaled air. Some of the sensors accurately detected the target analytes with detection limits comparable to the concentration levels often found in breath (low ppmv–ppbv). Some could provide repeatable results from the data produced during the measurements. All these facts prove the auspicious and promising future of the field.
Besides the advantages and positive results, some issues must be overcome in order to reinforce the relevancy of this field. The lack of standardized procedures and, consequently, the different results achieved by independent groups is one of those limitations. This deficiency in the standardization of the methodologies and systems employed for the fabrication of sensors and consequent detection of biomarkers leads to the impossibility of comparing and combining the data reported. Then, the complexity of the hardware utilized in the production of the systems is equally challenging since it often prevents on-site clinical studies and large medical trials. The reduced number of trials and, specifically, experiments with graphene-based sensors and real humans are another issue that must be assessed and overcome.
Still under the topic of limitations, this review paper proves that only some of the current systems exhibit detection limits that achieve the desired levels. Compounds in breath are often present in breath at low ppmv and even ppbv levels, and it is evident that not all the systems could reach such levels in order to be tested against biological samples in real scenarios.
Finally, one can conclude that another major challenge of this field is directly related to the biomarkers themselves. As noticeable throughout the review, the analytes are often linked to more than one pathology. An evident example of that is acetone, which has been studied regarding its suitability to act as a biomarker of all the eight pathologies here addressed, namely, asthma and COPD, CKD, diabetes, gastric and lung cancer, and sleep apnoea. This lack of specificity of the detected biomarkers prevents their contemporary utilization for medical purposes since further studies are required to fully assess how their presence and concentration levels can be related to a specific pathology.
One can foresee that when all these issues are fixed, the graphene-based sensors and the detection of organic-borne biomarkers will, undoubtedly, play a relevant role in the medicine of the future and will help improve medical care worldwide.

Author Contributions

Conceptualization, P.A.R., M.R. and V.V.; methodology, P.C.M.; software, P.C.M.; validation, P.A.R., M.R. and V.V.; formal analysis, P.C.M.; investigation, P.C.M.; resources, P.A.R., M.R. and V.V.; data curation, P.C.M.; writing—original draft preparation, P.C.M.; writing—review and editing, M.R. and V.V.; visualization, P.A.R. and M.R.; supervision, P.A.R., M.R. and V.V.; project administration, P.A.R., M.R. and V.V.; funding acquisition, P.A.R., M.R. and V.V. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by FUNDAÇÃO PARA A CIÊNCIA E TECNOLOGIA, grant number PD/BDE/150627/2020, UIDB/04559/2020(LIBPhys) and UIDP/04559/2020(LIBPhys).

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Etzioni, R.; Urban, N.; Ramsey, S.; McIntosh, M.; Schwartz, S.; Reid, B.; Radich, J.; Anderson, G.; Hartwell, L. The case for early detection. Nat. Rev. Cancer 2003, 3, 243–252. [Google Scholar] [CrossRef] [PubMed]
  2. Zhou, J.; Huang, Z.A.; Kumar, U.; Chen, D.D.Y. Review of recent developments in determining volatile organic compounds in exhaled breath as biomarkers for lung cancer diagnosis. Anal. Chim. Acta 2017, 996, 1–9. [Google Scholar] [CrossRef] [PubMed]
  3. Haworth, J.J.; Pitcher, C.K.; Ferrandino, G.; Hobson, A.R.; Pappan, K.L.; Lawson, J.L.D. Breathing new life into clinical testing and diagnostics: Perspectives on volatile biomarkers from breath. Crit. Rev. Clin. Lab. Sci. 2022, 59, 353–372. [Google Scholar] [CrossRef] [PubMed]
  4. Califf, R.M. Biomarker definitions and their application. Exp. Biol. Med. 2018, 243, 213–221. [Google Scholar] [CrossRef]
  5. Sawyers, C.L. The cancer biomarker problem. Nature 2008, 452, 548–552. [Google Scholar] [CrossRef] [PubMed]
  6. Moura, P.C.; Raposo, M.; Vassilenko, V. Breath Volatile Organic Compounds (VOCs) as Biomarkers for the Diagnosis of Pathological Conditions: A Review. Biomed. J. 2023, 46, 100623. [Google Scholar] [CrossRef]
  7. Amann, A.; Smith, D. Volatile Biomarkers: Non-Invasive Diagnosis in Physiology and Medicine, 1st ed.; Elsevier: Waltham, MA, USA, 2013. [Google Scholar]
  8. Daneshkhah, A.; Siegel, A.P.; Agarwal, M. Volatile organic compounds: Potential biomarkers for improved diagnosis and monitoring of diabetic wounds. In Wound Healing, Tissue Repair, and Regeneration Diabetes, 1st ed.; Bagchi, D., Das, A., Sashwati, R., Eds.; Academic Press: Cambridge, MA, USA, 2020; pp. 491–512. [Google Scholar]
  9. Amann, A.; Costello, B.L.; Miekisch, W.; Schubert, J.; Buszewski, B.; Pleil, J.; Ractiffe, N.; Risby, T. The human volatilome: Volatile organic compounds (VOCs) in exhaled breath, skin emanations, urine, feces and saliva. J. Breath Res. 2014, 8, 034001. [Google Scholar] [CrossRef]
  10. Cauchi, M.; Weber, C.M.; Bolt, B.J.; Spratt, P.B.; Bessant, C.; Turner, D.C.; Willis, C.M.; Britton, L.E.; Turner, C.; Morgan, G. Evaluation of gas chromatography mass spectrometry and pattern recognition for the identification of bladder cancer from urine headspace. Anal. Methods 2016, 8, 4037. [Google Scholar] [CrossRef]
  11. Bond, A.; Greenwood, R.; Lewis, S.; Corfe, B.; Sarkar, S.; O’Toole, P.; Rooney, P.; Burkitt, M.; Hold, G.; Probert, C. Volatile organic compounds emitted from faeces as a biomarker for colorectal cancer. Aliment. Pharmacol. Ther. 2019, 49, 1005–1012. [Google Scholar] [CrossRef]
  12. Aoki, T.; Nagaoka, T.; Kobayashi, N.; Kurahashi, M.; Tsuji, C.; Takiguchi, H.; Tomomatsu, K.; Oguma, T.; Kobayashi, N.; Magatani, K.; et al. Prospective Analyses of Volatile Organic Compounds in Obstructive Sleep Apnea Patients. Toxicol. Sci. 2017, 156, 362–374. [Google Scholar] [CrossRef]
  13. Berna, A.Z.; McCarthy, J.S.; Wang, R.X.; Saliba, K.J.; Bravo, F.G.; Cassells, J.; Padovan, B.; Trowell, S. Analysis of Breath Specimens for Biomarkers of Plasmodium falciparum Infection. J. Infect. Dis. 2015, 212, 1120–1128. [Google Scholar] [CrossRef] [PubMed]
  14. Chatterjee, S.; Castro, M.; Feller, J.F. An e-nose made of carbon nanotube based quantum resistive sensors for the detection of eighteen polar/nonpolar VOC biomarkers of lung cancer. J. Mater. Chem. B 2013, 1, 4563. [Google Scholar] [CrossRef] [PubMed]
  15. Hong, Y.; Che, X.; Su, H.; Mai, Z.; Huang, Z.; Huang, W.; Chen, W.; Liu, S.; Gao, W.; Zhou, Z.; et al. Exhaled breath analysis using on-line preconcentration mass spectrometry for gastric cancer diagnosis. J. Mass Spectrom. 2021, 56, e4588. [Google Scholar] [CrossRef] [PubMed]
  16. Halbritter, S.; Fedrigo, M.; Höllriegl, V.; Szymczak, W.; Maier, J.M.; Ziegler, A.G.; Hummel, M. Human Breath Gas Analysis in the Screening of Gestational Diabetes Mellitus. Diabetes Technol. Ther. 2012, 14, 917–925. [Google Scholar] [CrossRef]
  17. Neerincx, A.H.; Geurts, B.P.; Loon, J.; Tiemes, V.; Jansen, J.J.; Harren, F.J.M.; Kluijtmans, L.A.J.; Merkus, P.J.F.M.; Cristescu, S.M.; Buydens, L.M.C.; et al. Detection of Staphylococcus aureus in cystic fibrosis patients using breath VOC profiles. J. Breath Res. 2016, 10, 046014. [Google Scholar] [CrossRef]
  18. Chen, H.; Qi, X.; Zhang, L.; Li, X.; Ma, J.; Zhang, C.; Feng, H.; Yao, M. COVID-19 screening using breath-borne volatile organic compounds. J. Breath Res. 2021, 15, 047104. [Google Scholar] [CrossRef]
  19. Amal, H.; Leja, M.; Funka, K.; Lasina, I.; Skapars, R.; Sivins, A.; Ancans, G.; Kikuste, I.; Vanags, A.; Tolmanis, I.; et al. Breath testing as potential colorectal cancer screening tool. Int. J. Cancer 2016, 138, 229–236. [Google Scholar] [CrossRef]
  20. Hanouneh, I.A.; Zein, N.N.; Cikach, F.; Dababneh, L.; Grove, D.; Alkhouri, N.; Lopez, R.; Dweik, R.A. The Breathprints in Patients with Liver Disease Identify Novel Breath Biomarkers in Alcoholic Hepatitis. Clin. Gastroenterol. Hepatol. 2014, 12, 516–523. [Google Scholar] [CrossRef]
  21. Grabowska-Polanowska, B.; Skowron, M.; Miarka, P.; Pietrzycka, A.; Sliwka, I. The application of chromatographic breath analysis in the search of volatile biomarkers of chronic kidney disease and coexisting type 2 diabetes mellitus. J. Chromatogr. B 2017, 1060, 103–110. [Google Scholar] [CrossRef]
  22. Smolinska, A.; Klaassen, E.M.M.; Dallinga, J.W.; Kant, K.D.G.; Jobsis, Q.; Moonen, E.J.C.; Schayck, O.C.P.; Dompeling, E.; Schooten, F.J. Profiling of Volatile Organic Compounds in Exhaled Breath as a Strategy to Find Early Predictive Signatures of Asthma in Children. PLoS ONE 2014, 9, e95668. [Google Scholar] [CrossRef]
  23. Phillips, C.; Parthaláin, N.M.; Syed, Y.; Deganello, D.; Claypole, T.; Lewis, K. Short-Term Intra-Subject Variation in Exhaled Volatile Organic Compounds (VOCs) in COPD Patients and Healthy Controls and Its Effect on Disease Classification. Metabolites 2014, 4, 300–318. [Google Scholar] [CrossRef] [PubMed]
  24. Marom, O.; Nakhoul, F.; Tisch, U.; Shiban, A.; Abassi, Z.; Haick, H. Gold nanoparticle sensors for detecting chronic kidney disease and disease progression. Nanomedicine 2012, 7, 639–650. [Google Scholar] [CrossRef] [PubMed]
  25. Dadamio, J.; Velde, S.; Laleman, W.; Hee, P.; Coucke, W.; Nevens, F.; Quirynen, M. Breath biomarkers of liver cirrhosis. J. Chromatogr. B 2012, 905, 17–22. [Google Scholar] [CrossRef]
  26. Phillips, M.; Cataneo, R.N.; Saunders, C.; Hope, P.; Schmitt, P.; Wai, J. Volatile biomarkers in the breath of women with breast cancer. J. Breath Res. 2010, 4, 026003. [Google Scholar] [CrossRef] [PubMed]
  27. Dryahina, K.; Sovová, K.; Nemec, A.; Spanel, P. Differentiation of pulmonary bacterial pathogens in cystic fibrosis by volatile metabolites emitted by their in vitro cultures: Pseudomonas aeruginosa, Staphylococcus aureus, Stenotrophomonas maltophilia and the Burkholderia cepacia complex. J. Breath Res. 2016, 10, 037102. [Google Scholar] [CrossRef]
  28. Ruszkiewicz, D.M.; Sanders, D.; O’Brien, R.; Hempel, F.; Reed, M.J.; Riepe, A.C.; Bailie, K.; Brodrick, E.; Darnley, K.; Ellerkmann, R.; et al. Diagnosis of COVID-19 by analysis of breath with gas chromatography-ion mobility spectrometry—A feasibility study. EClinicalMedicine 2020, 29, 100609. [Google Scholar] [CrossRef]
  29. Xu, Z.; Broza, Y.Y.; Ionsecu, R.; Tisch, U.; Ding, L.; Liu, H.; Song, Q.; Pan, Y.; Xiong, F.; Gu, K.; et al. A nanomaterial-based breath test for distinguishing gastric cancer from benign gastric conditions. Br. J. Cancer 2013, 108, 941–950. [Google Scholar] [CrossRef]
  30. Neupane, S.; Peverall, R.; Richmond, G.; Blaikie, T.P.J.; Taylor, D.; Hancock, G.; Evans, M.L. Exhaled Breath Isoprene Rises During Hypoglycemia in Type 1 Diabetes. Diabetes Care 2016, 39, e97–e98. [Google Scholar] [CrossRef]
  31. Schaber, C.L.; Katta, N.; Bollinger, L.B.; Mwale, M.; Mlotha-Mitole, R.; Trehan, I.; Raman, B.; John, A.R.O. Breathprinting Reveals Malaria-Associated Biomarkers and Mosquito Attractants. J. Infect. Diss 2018, 217, 1553–1560. [Google Scholar] [CrossRef]
  32. Ma, H.; Li, X.; Chen, J.; Wang, H.; Cheng, T.; Chen, K.; Xu, S. Analysis of human breath samples of lung cancer patients and healthy controls with solid-phase microextraction (SPME) and flow-modulated comprehensive two-dimensional gas chromatography (GC-GC). Anal. Methods 2014, 6, 6841–6849. [Google Scholar] [CrossRef]
  33. Rodríguez-Aguilar, M.; Ramírez-García, S.; Ilizaliturri-Hernández, C.; Gómez-Gómez, A.; Van-Brussel, E.; Díaz-Barriga, F.; Medellín-Garibay, S.; Flores-Ramírez, R. Ultrafast gas chromatography coupled to electronic nose to identify biomarkers in exhaled breath from chronic obstructive pulmonary disease patients: A pilot study. Biomed. Chromatogr. 2019, 33, e4684. [Google Scholar] [CrossRef]
  34. Mastrigt, E.; Reyes-Reyes, A.; Brand, K.; Bhattacharya, N.; Urbach, H.P.; Stubbs, A.P.; Jongste, J.C.; Pijnenburg, M.W. Exhaled breath profiling using broadband quantum cascade laser-based spectroscopy in healthy children and children with asthma and cystic fibrosis. J. Breath Res. 2016, 10, 026003. [Google Scholar] [CrossRef] [PubMed]
  35. Zhou, W.; Huang, C.; Zou, X.; Lu, Y.; Shen, C.; Ding, X.; Wang, H.; Jiang, H.; Chu, Y. Exhaled breath online measurement for cervical cancer patients and healthy subjects by proton transfer reaction mass spectrometry. Anal. Bioanal. Chem. 2017, 409, 5603–5612. [Google Scholar] [CrossRef]
  36. Phillips, M.; Basa-Dalay, V.; Bothamley, G.; Cataneo, R.N.; Lam, P.K.; Natividad, M.P.R.; Schmitt, P.; Wai, J. Breath biomarkers of active pulmonary tuberculosis. Tuberculosis 2010, 90, 145–151. [Google Scholar] [CrossRef] [PubMed]
  37. Maiti, K.S.; Fill, E.; Strittmatter, F.; Volz, Y.; Sroka, R.; Apolonski, A. Towards reliable diagnostics of prostate cancer via breath. Sci. Rep. 2021, 11, 18381. [Google Scholar] [CrossRef]
  38. Moura, P.C.; Vassilenko, V.; Ribeiro, P.A. Ion Mobility Spectrometry Towards Environmental Volatile Organic Compounds Identification and Quantification: A Comparative Overview over Infrared Spectroscopy. Emiss. Control Sci. Technol. 2023, 9, 25–46. [Google Scholar] [CrossRef]
  39. Boeker, P. On ‘Electronic Nose’ methodology. Sens. Actuators B Chem. 2014, 204, 2–17. [Google Scholar] [CrossRef]
  40. Dragonieri, S.; Pennazza, G.; Carratu, P.; Resta, O. Electronic Nose Technology in Respiratory Diseases. Lung 2017, 195, 157–165. [Google Scholar] [CrossRef] [PubMed]
  41. Moura, P.C.; Vassilenko, V. Contemporary ion mobility spectrometry applications and future trends towards environmental, health and food research: A review. Int. J. Mass Spectrom. 2023, 486, 117012. [Google Scholar] [CrossRef]
  42. Ludwig, K.R.; Hummon, A.B. Mass spectrometry for the discovery of biomarkers of sepsis. Mol. Biosyst. 2017, 13, 648. [Google Scholar] [CrossRef]
  43. Kouremenos, K.A.; Johansson, M.; Marriott, P.J. Advances in Gas Chromatographic Methods for the Identification of Biomarkers in Cancer. J. Cancer 2012, 3, 404–420. [Google Scholar] [CrossRef] [PubMed]
  44. Binson, V.A.; Subramoniam, M.; Mathew, L. Discrimination of COPD and lung cancer from controls through breath analysis using a self-developed e-nose. J. Breath Res. 2021, 15, 046003. [Google Scholar] [CrossRef]
  45. Al-Hamry, A.; Panzardi, E.; Mugnaini, M.; Kanoun, O. Human Breathing Monitoring by Graphene Oxide Based Sensors. In Advanced Sensors for Biomedical Applications, 1st ed.; Kanoun, O., Derbel, N., Eds.; Springer: Cham, Switzerland, 2021; pp. 97–105. [Google Scholar]
  46. Geim, A.K.; Novoselov, K.S. The rise of graphene. Nat. Mater. 2007, 6, 183–191. [Google Scholar] [CrossRef] [PubMed]
  47. Sekhon, S.S.; Kaur, P.; Kim, Y.H.; Sekhon, S.S. 2D graphene oxide–aptamer conjugate materials for cancer diagnosis. NPJ 2D Mater. Appl. 2021, 5, 21. [Google Scholar] [CrossRef]
  48. Shahdeo, D.; Roberts, A.; Abbineni, N.; Gandhi, S. Graphene based sensors. Compr. Anal. Chem. 2020, 91, 175–199. [Google Scholar]
  49. Zagalo, P.M.; Sério, S.; Ribeiro, P.A.; Raposo, M. Graphene-based biosensors. In Recent Advances in Graphene-Based Technologies, 1st ed.; Chandran, A., Unnikrishan, N.V., Jayaraj, M.K., George, J., Eds.; IOP Publishing: Bristol, UK, 2023; pp. 1–20. [Google Scholar]
  50. Moura, P.C.; Pivetta, T.P.; Vassilenko, V.; Ribeiro, P.A.; Raposo, M. Graphene Oxide Thin Films for Detection and Quantification of Industrially Relevant Alcohols and Acetic Acid. Sensors 2023, 23, 462. [Google Scholar] [CrossRef]
  51. Magro, C.; Gonçalves, O.C.; Morais, M.; Ribeiro, P.A.; Sério, S.; Vieira, P.; Raposo, M. Volatile Organic Compound Monitoring during Extreme Wildfires: Assessing the Potential of Sensors Based on LbL and Sputtering Films. Sensors 2022, 22, 6677. [Google Scholar] [CrossRef]
  52. Magro, C.; Mateus, E.P.; Raposo, M.; Ribeiro, A.B. Overview of electronic tongue sensing in environmental aqueous matrices: Potential for monitoring emerging organic contaminants. Environ. Rev. 2019, 27, 202–214. [Google Scholar] [CrossRef]
  53. Magro, C.; Zagalo, P.; Pereira-da-Silva, J.; Mateus, E.P.; Ribeiro, A.B.; Ribeiro, P.; Raposo, M. Polyelectrolyte Based Sensors as Key to Achieve Quantitative Electronic Tongues: Detection of Triclosan on Aqueous Environmental Matrices. Nanomaterials 2020, 10, 640. [Google Scholar] [CrossRef]
  54. Pienutsa, N.; Roongruangsree, P.; Seedokbuab, V.; Yannawibut, K.; Phatoomvijitwong, C.; Srinives, S. SnO2-graphene composite gas sensor for a room temperature detection of ethanol. Nanotechnology 2020, 32, 115502. [Google Scholar] [CrossRef]
  55. Phasuksom, K.; Prissanaroon-Ouajai, W.; Sirivat, A. A highly responsive methanol sensor based on graphene oxide/polyindole composites. RSC Adv. 2020, 10, 15206–15220. [Google Scholar] [CrossRef] [PubMed]
  56. Minta, D.; González, Z.; Wiench, P.; Gryglewicz, S.; Gryglewicz, G. N-Doped Reduced Graphene Oxide/Gold Nanoparticles Composite as an Improved Sensing Platform for Simultaneous Detection of Dopamine, Ascorbic Acid, and Uric Acid. Sensors 2020, 20, 4427. [Google Scholar] [CrossRef] [PubMed]
  57. Vitoria, I.; Gallego, E.E.; Melendi-Espina, S.; Hernaez, M.; Zamarreño, C.R.; Matías, I.R. Gas Sensor Based on Lossy Mode Resonances by Means of Thin Graphene Oxide Films Fabricated onto Planar Coverslips. Sensors 2023, 23, 1459. [Google Scholar] [CrossRef] [PubMed]
  58. Hernaez, M.; Zamarreño, C.R.; Melendi-Espina, S.; Bird, L.R.; Mayes, A.G.; Arregui, F.J. Optical Fibre Sensors Using Graphene-Based Materials: A Review. Sensors 2017, 17, 155. [Google Scholar] [CrossRef]
  59. Li, D.; Shao, Y.; Zhang, Q.; Qu, M.; Ping, J.; Fu, Y.; Xie, J. A flexible virtual sensor array based on laserinduced graphene and MXene for detecting volatile organic compounds in human breath. Analyst 2021, 146, 5704. [Google Scholar] [CrossRef]
  60. Chronic Obstructive PULMONARY Disease (COPD). Available online: https://www.who.int/news-room/fact-sheets/detail/chronic-obstructive-pulmonary-disease-(copd) (accessed on 21 October 2023).
  61. Asthma. Available online: https://www.who.int/news-room/fact-sheets/detail/asthma (accessed on 21 October 2023).
  62. Hargreave, F.E.; Nair, P. The definition and diagnosis of asthma. Clin. Exp. Allergy 2009, 38, 1652–1658. [Google Scholar] [CrossRef]
  63. Mannino, D.M. COPD: Epidemiology, Prevalence, Morbidity and Mortality, and Disease Heterogeneity. Chest 2002, 121, 121S–126S. [Google Scholar] [CrossRef]
  64. Beach, J.; Russell, K.; Blitz, S.; Hooton, N.; Spooner, C.; Lemiere, C.; Tarlo, S.; Rowe, B. A Systematic Review of the Diagnosis of Occupational Asthma. Chest 2007, 131, 569–578. [Google Scholar] [CrossRef]
  65. Rennard, S.I.; Drummond, M.B. Early chronic obstructive pulmonary disease: Definition, assessment, and prevention. Lancet 2015, 385, 1778–1788. [Google Scholar] [CrossRef]
  66. Caldeira, M.; Perestrelo, R.; Barros, A.S.; Bilelo, M.J.; Morête, A.; Câmara, J.S.; Rocha, S.M. Allergic asthma exhaled breath metabolome: A challenge for comprehensive two-dimensional gas chromatography. J. Chromatogr. A 2012, 1254, 87–97. [Google Scholar] [CrossRef]
  67. Schleich, F.N.; Zanella, D.; Stefanuto, P.H.; Bessonov, K.; Smolinska, A.; Dallinga, J.W.; Henket, M.; Paulus, V.; Guissard, F.; Graff, S.; et al. Exhaled Volatile Organic Compounds Are Able to Discriminate between Neutrophilic and Eosinophilic Asthma. Am. J. Respir. Crit. Care Med. 2019, 15, 444–453. [Google Scholar] [CrossRef]
  68. Gahleitner, F.; Guallar-Hoyas, C.; Beardsmore, C.S.; Pandya, H.C.; Thomas, C.P. Metabolomics pilot study to identify volatile organic compound markers of childhood asthma in exhaled breath. Bioanalysis 2013, 5, 2239–2247. [Google Scholar] [CrossRef]
  69. Cazzola, M.; Segreti, A.; Capuano, R.; Bergamini, A.; Martinelli, E.; Calzetta, L.; Rogliani, P.; Ciaprini, C.; Ora, J.; Paolesse, R.; et al. Analysis of exhaled breath fingerprints and volatile organic compounds in COPD. COPD Res. Pract. 2015, 1, 7. [Google Scholar] [CrossRef]
  70. Jareño-Esteban, J.J.; Muñoz-Lucas, M.A.; Gómez-Martín, O.; Utrilla-Trigo, S.; Gutiérrez-Ortega, C.; Aguilar-Ros, A.; Collado-Yurrita, L.; Callol-Sánchez, L.M. Study of 5 Volatile Organic Compounds in Exhaled Breath in Chronic Obstructive Pulmonary Disease. Arch. Bronconeumol. 2017, 53, 251–256. [Google Scholar] [CrossRef]
  71. Bollella, P.; Fusco, G.; Tortolini, C.; Sanzò, G.; Favero, G.; Gorton, L.; Antiochia, R. Beyond graphene: Electrochemical sensors and biosensors for biomarkers detection. Biosens. Bioelectron. 2017, 89, 152–166. [Google Scholar] [CrossRef]
  72. Ghoizadeh, A.; Voiry, D.; Weisel, C.; Gow, A.; Laumbach, R.; Kipen, H.; Chhwalla, M. Toward point-of-care management of chronic respiratory conditions: Electrochemical sensing of nitrite content in exhaled breath condensate using reduced graphene oxide. Microsys. Nanoeng. 2017, 3, 17022. [Google Scholar] [CrossRef] [PubMed]
  73. Gholizadeh, A.; Black, K.; Kipen, H.; Laumbach, R.; Gow, A.; Weisel, C.; Javanmard, M. Detection of respiratory inflammation biomarkers in non-processed exhaled breath condensate samples using reduced graphene oxide. RSC Adv. 2022, 12, 35627. [Google Scholar] [CrossRef] [PubMed]
  74. Kumar, R.; Ghosh, R. Selective determination of ammonia, ethanol and acetone by reduced graphene oxide based gas sensors at room temperature. Sens. Bio-Sens. Res. 2022, 28, 100336. [Google Scholar] [CrossRef]
  75. Samadi, S.; Nouroozshad, M.; Zakaria, S.A. ZnO@SiO2/rGO core/shell nanocomposite: A superior sensitive, selective and reproducible performance for 1-propanol gas sensor at room temperature. Mater. Chem. Phys. 2021, 271, 124884. [Google Scholar] [CrossRef]
  76. Hussein, H.T.; Nayef, U.M.; Hussien, A.M.A. Synthesis of graphene on porous silicon for vapor organic sensor by using photoluminescence. Optik 2019, 180, 61–70. [Google Scholar] [CrossRef]
  77. Murashima, Y.; Karim, M.R.; Furue, R.; Matsui, T.; Takehira, H.; Wakata, K.; Toda, K.; Ohtani, R.; Nakamura, M.; Hayami, S. Reduced graphene oxide–transition metal hybrids as p-type semiconductors for acetaldehyde sensing. Inorg. Chem. Front. 2016, 3, 842–848. [Google Scholar] [CrossRef]
  78. Khan, A.A.P.; Khan, A.; Rahman, M.M.; Asiri, A.M. Conventional surfactant-doped poly (o-anisidine)/GO nanocomposites for benzaldehyde chemical sensor development. J. Solgel Sci. Technol. 2016, 77, 361–370. [Google Scholar] [CrossRef]
  79. Chen, Q.; Liu, D.; Lin, L.; Wu, J. Bridging interdigitated electrodes by electrochemical-assisted deposition of graphene oxide for constructing flexible gas sensor. Sens. Actuators B Chem. 2019, 286, 591–599. [Google Scholar] [CrossRef]
  80. Liu, B.; Huang, Y.; Kam, K.W.L.; Cheung, W.F.; Zhao, N.; Zheng, B. Functionalized graphene-based chemiresistive electronic nose for discrimination of disease-related volatile organic compounds. Biosens. Bioelectron. X 2019, 1, 100016. [Google Scholar] [CrossRef]
  81. Turner, J.M.; Bauer, C.; Abramowitz, M.K.; Melamed, M.L.; Hostetter, T.H. Treatment of chronic kidney disease. Kidney Int. 2012, 81, 351–362. [Google Scholar] [CrossRef]
  82. Hsu, C.Y.; Ordoñez, J.D.; Chertow, G.M.; Fan, D.; McCulloch, C.E.; Go, A.S. Thre risk of acute renal failure in patients with chronic kidney disease. Kidney Int. 2008, 74, 101–107. [Google Scholar] [CrossRef] [PubMed]
  83. Wright, J.; Hutchison, A. Cardiovascular disease in patients with chronic kidney disease. Vasc. Health Risk Manag. 2009, 5, 713–722. [Google Scholar]
  84. Taniselass, S.; Arshad, M.K.; Gopinath, S.C.B. Graphene-based electrochemical biosensors for monitoring noncommunicable disease biomarkers. Biosens. Bioelectron. 2019, 130, 276–292. [Google Scholar] [CrossRef]
  85. Bayrakli, I.; Turkmen, A.; Akman, H.; Sezer, M.T.; Kutluhan, S. Applications of external cavity diode laser-based technique to noninvasive clinical diagnosis using expired breath ammonia analysis: Chronic kidney disease, epilepsy. J. Biomed. Opt. 2016, 21, 087004. [Google Scholar] [CrossRef]
  86. Grabowska-Polanowska, B.; Faber, J.; Skowron, M.; Miarka, P.; Pietrzycka, A.; Sliwka, I.; Amann, A. Detection of potential chronic kidney disease markers in breath using gas chromatography with mass-spectral detection coupled with thermal desorption method. J. Chromatogr. A 2013, 1301, 179–189. [Google Scholar] [CrossRef]
  87. Obermeier, J.; Trefz, P.; Happ, J.; Schubbert, J.K.; Staude, H.; Fischer, D.C.; Miekisch, W. Exhaled volatile substances mirror clinical conditions in pediatric chronic kidney disease. PLoS ONE 2017, 12, e0178745. [Google Scholar] [CrossRef] [PubMed]
  88. Lee, B.M.; Eetemadi, A.; Tagkopoulos, I. Reduced Graphene Oxide-Metalloporphyrin Sensors for Human Breath Screening. Appl. Sci. 2021, 11, 11290. [Google Scholar] [CrossRef]
  89. Majidi, R.; Nadafan, M. Application of nitrogenated holey graphene for detection of volatile organic biomarkers in exhaled breath of humans with chronic kidney disease: A density functional theory study. J. Comput. Electron. 2021, 20, 1930–1937. [Google Scholar] [CrossRef]
  90. Tung, T.T.; Tran, M.T.; Feller, J.F.; Castro, M.; Ngo, T.V.; Hassan, K.; Nine, J.; Losic, D. Graphene and metal organic frameworks (MOFs) hybridization for tunable chemoresistive sensors for detection of volatile organic compounds (VOCs) biomarkers. Carbon 2020, 159, 333–344. [Google Scholar] [CrossRef]
  91. Choi, S.J.; Fuchs, F.; Demadrille, R.; Grévin, B.; Jang, B.H.; Lee, S.J.; Lee, J.H.; Tuller, H.; Kim, I.D. Fast Responding Exhaled-Breath Sensors Using WO3 Hemitubes Functionalized by Graphene-Based Electronic Sensitizers for Diagnosis of Diseases. Appl. Mater. Interfaces 2014, 6, 9061–9070. [Google Scholar] [CrossRef] [PubMed]
  92. Brannelly, N.T.; Hamilton-Shield, J.P.; Killard, A.J. The Measurement of Ammonia in Human Breath and its Potential in Clinical Diagnostics. Crit. Rev. Anal. Chem. 2016, 46, 490–501. [Google Scholar] [CrossRef]
  93. Ogimoto, Y.; Selyanchyn, R.; Takahara, N.; Wakamatsu, S.; Lee, S. Detection of ammonia in human breath using quartz crystal microbalance sensors with functionalized mesoporous SiO2 nanoparticle films. Sens. Actuators B Chem. 2015, 215, 428–436. [Google Scholar] [CrossRef]
  94. Narasimhan, L.R.; Goodman, W.; Patel, C.K. Correlation of breath ammonia with blood urea nitrogen and creatinine during hemodialysis. Proc. Natl. Acad. Sci. USA 2001, 98, 4617–4621. [Google Scholar] [CrossRef]
  95. Shahmoradi, A.; Hosseini, A.; Akbarinejad, A.; Alizadeh, N. Noninvasive Detection of Ammonia in the Breath of Hemodialysis Patients Using a Highly Sensitive Ammonia Sensor Based on a Polypyrrole/Sulfonated Graphene Nanocomposite. Anal. Chem. 2021, 93, 6706–6714. [Google Scholar] [CrossRef]
  96. Yabaş, E.; Biçer, E.; Altındal, A. Novel reduced graphene oxide/zinc phthalocyanine and reduced graphene oxide/cobalt phthalocyanine hybrids as high sensitivity room temperature volatile organic compound gas sensors. J. Mol. Struct. 2023, 1271, 134076. [Google Scholar] [CrossRef]
  97. Deshpande, A.; Harris-Hayes, M.; Schootman, M. Epidemiology of Diabetes and Diabetes-Related Complications. Phys. Ther. 2008, 88, 1254–1264. [Google Scholar] [CrossRef]
  98. Iglay, K.; Hannachi, H.; Howie, P.J.; Xu, J.; Li, X.; Engel, S.S.; Moore, L.M.; Rajpathak, S. Prevalence and co-prevalence of comorbidities among patients with type 2 diabetes mellitus. Curr. Med. Res. Opin. 2016, 32, 1243–1252. [Google Scholar] [CrossRef] [PubMed]
  99. Inzucchi, S.E. Diagnosis of Diabetes. N. Engl. J. Med. 2012, 367, 542–550. [Google Scholar] [CrossRef] [PubMed]
  100. Dixit, K.; Fardindoost, S.; Ravishankara, A.; Tasnim, N.; Hoorfar, M. Exhaled Breath Analysis for Diabetes Diagnosis and Monitoring: Relevance, Challenges and Possibilities. Biosensors 2021, 11, 476. [Google Scholar] [CrossRef] [PubMed]
  101. Li, W.; Liu, Y.; Liu, Y.; Cheng, S.; Duan, Y. Exhaled isopropanol: New potential biomarker in diabetic breathomics and its metabolic correlations with acetone. RSC Adv. 2017, 7, 17480. [Google Scholar] [CrossRef]
  102. Righettoni, M.; Schmid, A.; Amann, A.; Pratsinis, S.E. Correlations between blood glucose and breath components from portable gas sensors and PTR-TOF-MS. J. Breath Res. 2013, 7, 037110. [Google Scholar] [CrossRef]
  103. Yan, H.; Zhou, Y.G. Electrical sensing of volatile organic compounds in exhaled breath for disease diagnosis. Curr. Opin. Electrochem. 2022, 33, 100922. [Google Scholar] [CrossRef]
  104. Kalidoss, R.; Umapathy, S.; Sivalingam, Y. An investigation of GO-SnO2-TiO2 ternary nanocomposite for the detection of acetone in diabetes mellitus patient’s breath. Appl. Surf. Sci. 2018, 449, 677–684. [Google Scholar] [CrossRef]
  105. Kalidoss, R.; Umapathy, S. A comparison of online and offline measurement of exhaled breath for diabetes pre-screening by graphene-based sensors; from powder processing to clinical monitoring prototype. J. Breath Res. 2019, 13, 036008. [Google Scholar] [CrossRef] [PubMed]
  106. Kalidoss, R.; Umapathy, S.; Kothalam, R.; Sakthivelu, U. Adsorption kinetics feature extraction from breathprint obtained by graphene based sensors for diabetes diagnosis. J. Breath Res. 2021, 15, 016005. [Google Scholar] [CrossRef] [PubMed]
  107. Thakur, U.N.; Bhardwaj, R.; Hazra, A. A Multivariative Computational Approach with Hybrid Graphene Oxide Sensor Array for Partial Fulfillment of Breath Acetone Sensing. IEEE Sens. J. 2022, 22, 20207–20216. [Google Scholar] [CrossRef]
  108. Choi, S.J.; Jang, B.; Lee, S.; Min, B.K.; Rothschild, A.; Kim, I. Selective Detection of Acetone and Hydrogen Sulfide for the Diagnosis of Diabetes and Halitosis Using SnO2 Nanofibers Functionalized with Reduced Graphene Oxide Nanosheets. Appl. Mater. Interfaces 2014, 6, 2588–2597. [Google Scholar] [CrossRef] [PubMed]
  109. Sen, S.; Maity, S.; Kundu, S. Reduced graphene oxide (rGO) decorated NiO-SnO2 nanocomposite based sensor towards room temperature diabetic biomarker detection. J. Alloys Compd. 2023, 966, 171553. [Google Scholar] [CrossRef]
  110. Yempally, S.; Hegazy, S.M.; Aly, A.; Kannan, K.; Sadasivuni, K.K. Non-Invasive Diabetic Sensor Based on Cellulose Acetate/Graphene Nanocomposite. Macromol. Symp. 2020, 392, 2000024. [Google Scholar] [CrossRef]
  111. Rakkesh, R.A.; Durgalakshmi, D.; Balakumar, S. Scalable approach to fabricate paper-based biomass reduced graphene sensor for the detection of exhaled diabetic breath. Nanotechnology 2022, 33, 495703. [Google Scholar] [CrossRef] [PubMed]
  112. Catalano, V.; Labianca, R.; Beretta, G.D.; Gatta, G.; Braud, F.; Cutsem, E. Gastric Cancer. Crit. Rev. Oncol. Hematol. 2009, 71, 127–164. [Google Scholar] [CrossRef]
  113. Blair, V.; Martin, I.; Shaw, D.; Winship, I.; Kerr, D.; Arnold, J.; Harawira, P.; McLeod, M.; Parry, S.; Charlton, A.; et al. Hereditary Diffuse Gastric Cancer: Diagnosis and Management. Clin. Gastroenterol. Hepatol. 2006, 4, 262–275. [Google Scholar] [CrossRef]
  114. Xiang, L.; Wu, S.; Hua, Q.; Bao, C.; Liu, H. Volatile Organic Compounds in Human Exhaled Breath to Diagnose Gastrointestinal Cancer: A Metal-Analysis. Front. Oncol. 2021, 11, 606915. [Google Scholar] [CrossRef]
  115. Jung, Y.J.; Seo, H.S.; Kim, J.H.; Song, K.Y.; Park, C.H.; Lee, H.H. Advanced Diagnostic Technology of Volatile Organic Compounds Real Tiem Analysis from Exhaled Breath of Gastric Cancer Patients Using Proton-Transfer-Reaction Time-of-Flight Mass Spectrometry. Front. Oncol. 2021, 11, 560591. [Google Scholar] [CrossRef]
  116. Chen, Y.; Zhang, Y.; Pan, F.; Liu, J.; Wang, K.; Zhang, C.; Cheng, S.; Lu, L.; Zhang, W.; Zhang, Z.; et al. Breath Analysis Based on Surface-Enhanced Raman Scattering Sensors Distinguishes Early and Advanced Gastric Cancer Patients from Healthy Persons. ACS Nano 2016, 10, 8169–8179. [Google Scholar] [CrossRef]
  117. Zhang, Y.; Gao, G.; Liu, H.; Fu, H.; Fan, J.; Wang, K.; Chen, Y.; Li, B.; Zhang, C.; Zhi, X.; et al. Identification of Volatile Biomarkers of Gastric Cancer Cells and Ultrasensitive Electrochemical Detection based on Sensing Interface of Au-Ag Alloy coated MWCNTs. Theranostics 2014, 4, 154–162. [Google Scholar] [CrossRef]
  118. Amal, H.; Leja, M.; Funka, K.; Skarpars, R.; Sivins, A.; Ancans, G.; Liepniece-Karele, I.; Kikuste, I.; Lasina, I.; Haick, H. Detection of precancerous gastric lesions and gastric cancer through exhaled breath. Gut 2016, 65, 400–407. [Google Scholar] [CrossRef]
  119. Tombel, N.S.M.; Badaruddin, S.A.M.; Yakin, F.S.; Zaki, H.F.M.; Syono, M.I. Detection of low PPM of volatile organic compounds using nanomaterial functionalized reduced graphene oxide sensor. AIP Conf. Proc. 2021, 2368, 020004. [Google Scholar]
  120. Velumani, M.; Prasanth, A.; Narasimman, S.; Chandrasekhar, A.; Sampson, A.; Meher, S.R.; Rajalingam, S.; Rufus, E.; Alex, Z.C. Nanomaterial-Based Sensors for Exhaled Breath Analysis: A Review. Coatings 2022, 12, 1989. [Google Scholar] [CrossRef]
  121. Jia, X.; Yu, S.; Cheng, C.; Yang, J.; Li, Y.; Wang, S.; Song, H. Ag nanoparticles modified Fe3O4/reduced graphene oxide and their acetone sensing properties. Mater. Chem. Phys. 2022, 276, 125378. [Google Scholar] [CrossRef]
  122. Pastorino, U. Lung cancer screening. Br. J. Cancer 2010, 102, 1681–1686. [Google Scholar] [CrossRef] [PubMed]
  123. Schabath, M.B.; Cote, M.L. Cancer Progress and Priorities: Lung Cancer. Cancer Epidemiol. Biomark. Prev. 2019, 28, 1563–1579. [Google Scholar] [CrossRef]
  124. Lung Cancer Statistics. Available online: https://www.wcrf.org/cancer-trends/lung-cancer-statistics/ (accessed on 21 October 2023).
  125. Thriumani, R.; Zakaria, A.; Hashim, Y.; Jeffree, A.I.; Helmy, K.M.; Kamarudin, L.M.; Omar, M.I.; Shakaff, A.Y.; Adom, A.H.; Persaud, K.C. A study on volatile organic compounds emitted by in-vitro lung cancer cultured cells using gas sensor array and SPME-GCMS. BMC Cancer 2018, 18, 362. [Google Scholar] [CrossRef]
  126. Kischkel, S.; Miekisch, W.; Sawacki, A.; Straker, E.M.; Trefz, P.; Amann, A.; Schubert, J.K. Breath biomarkers for lung cancer detection and assessment of smoking related effects—Confounding variables, influence of normalization and statistical algorithms. Clin. Chim. Acta 2010, 411, 1637–1644. [Google Scholar] [CrossRef]
  127. Poli, D.; Goldoni, M.; Corradi, M.; Acampa, O.; Carbognani, P.; Internullo, E.; Casalini, A.; Mutti, A. Determination of aldehydes in exhaled breath of patients with lung cancer by means of on-fiber-derivatisation SPME-GC/MS. J. Chromatogr. B 2010, 878, 2643–2651. [Google Scholar] [CrossRef]
  128. Fuchs, P.; Loeseken, C.; Schubert, J.K.; Mieksich, W. Breath gas aldehydes as biomarkers of lung cancer. Int. J. Cancer 2010, 126, 2663–2670. [Google Scholar] [CrossRef] [PubMed]
  129. Fu, X.A.; Li, M.; Knipp, R.J.; Nantz, M.H.; Bousamra, M. Noninvasive detection of lung cancer using exhaled breath. Cancer Med. 2014, 3, 174–181. [Google Scholar] [CrossRef] [PubMed]
  130. Buszewski, B.; Ligor, T.; Jezierski, T.; Wenda-Piesik, A.; Walczak, M.; Rudnicka, J. Identification of volatile lung cancer markers by gas chromatography-mass spectrometry: Comparison with discrimination by canines. Anal. Bioanal. Chem. 2012, 404, 141–146. [Google Scholar] [CrossRef]
  131. Saalberg, Y.; Bruhns, H.; Wolff, M. Photoacoustic Spectroscopy for the Determination of Lung Cancer Biomarkers—A Preliminary Investigation. Sensors 2017, 17, 210. [Google Scholar] [CrossRef]
  132. Buszewski, B.; Ulanowska, A.; Kowalkowski, T.; Cieslinski, K. Investigation of lung cancer biomarkers by hyphenated separation techniques and chemometrics. Clin. Chem. Lab. Med. 2011, 50, 573–581. [Google Scholar] [CrossRef]
  133. Kushch, I.; Schwarz, K.; Schwentner, L.; Baumann, B.; Dzien, A.; Schmid, A.; Unterkofler, K.; Gastl, G.; Španěl, P.; Smith, D.; et al. Compounds enhanced in a mass spectrometric profile of smokers’ exhaled breath versus non-smokers as determined in a pilot study using PTR-MS. J. Breath Res. 2008, 2, 026002. [Google Scholar] [CrossRef]
  134. Zhou, X.; Xue, Z.; Chen, X.; Huang, C.; Bai, W.; Lu, Z.; Wang, T. Nanomaterial-based gas sensors used for breath diagnosis. J. Mater. Chem. B 2020, 8, 3231. [Google Scholar] [CrossRef] [PubMed]
  135. Emam, S.; Nasrollahpour, M.; Allen, J.P.; He, Y.; Hussein, H.; Shah, H.S.; Tavangarian, F.; Sun, N.X. A handheld electronic device with the potential to detect lung cancer biomarkers from exhaled breath. Biomed. Microdevices 2022, 24, 41. [Google Scholar] [CrossRef]
  136. Shanmugasundaram, A.; Lee, D.W. SnO2/rGO nanocomposite for the detection of biomarkers of lung cancer. Micro Nano Syst. Lett. 2022, 10, 13. [Google Scholar] [CrossRef]
  137. Zhang, X.; Zhang, Z.; He, Y.; Liu, Y.; Li, S.; Fang, J.; Zhang, X.; Peng, G. Sniffing lung cancer related biomarkers using an oxidized graphene SAW sensor. Front. Phys. 2016, 11, 116801. [Google Scholar] [CrossRef]
  138. Chen, Q.; Chen, Z.; Liu, D.; He, Z.; Wu, J. Constructing an E Nose Using Metal-Ion-Induced Assembly of Graphene Oxide for Diagnosis of Lung Cancer via Exhaled Breath. App. Mater. Interfaces 2020, 12, 17713–17724. [Google Scholar] [CrossRef] [PubMed]
  139. Sánchez-Vicente, C.; Santos, J.P.; Lozano, J.; Sayago, I.; Sanjurjo, J.L.; Azabal, A.; Ruiz-Valdepeñas, S. Graphene-Doped Tin Oxide Nanofibers and Nanoribbons as Gas Sensors to Detect Biomarkers of Different Diseases through the Breath. Sensors 2020, 20, 7223. [Google Scholar] [CrossRef] [PubMed]
  140. Shanmugasundaram, A.; Manorama, S.V.; Kim, D.S.; Jeong, Y.J.; Lee, D.W. Toward Point-of-Care chronic disease Management: Biomarker detection in exhaled breath using an E-Nose sensor based on rGO/SnO2 superstructures. Chem. Eng. J. 2022, 448, 137736. [Google Scholar] [CrossRef]
  141. Sleep Apnea Statistics. Available online: https://www.ncoa.org/adviser/sleep/sleep-apnea-statistics/ (accessed on 21 October 2023).
  142. Lévy, P.; Kohler, M.; McNicholas, W.T.; Barbé, F.; McEvoy, R.D.; Somers, V.K.; Lavie, L.; Pépin, J.L. Obstructive sleep apnoea syndrome. Nat. Rev. Dis. Primers 2015, 1, 15015. [Google Scholar] [CrossRef] [PubMed]
  143. Veasey, S.C.; Rosen, I.M. Obstructive Sleep Apnea in Adults. N. Engl. J. Med. 2019, 380, 1442–1449. [Google Scholar] [CrossRef] [PubMed]
  144. Dragonieri, S.; Porcelli, F.; Longobardi, F.; Carratù, P.; Aliani, M.; Ventura, V.A.; Tutino, M.; Quaranta, V.N.; Resta, O.; Gennaro, G. An electric nose in the discrimination of obese patients with and without obstructive sleep apnoea. J. Breath Res. 2015, 9, 026005. [Google Scholar] [CrossRef]
  145. Casanova-Chafer, J.; Garcia-Aboal, R.; Atienzar, P.; Bittencourt, C.; Lobet, E. Perovskite@Graphene Nanohybrids for Breath Analysis: A Proof-of-Concept. Chemosensors 2021, 9, 215. [Google Scholar] [CrossRef]
  146. Casanova-Chafer, J.; Garcia-Aboal, R.; Atienzar, P.; Llobet, E. The role of anions and cations in the gas sensing mechanisms of graphene decorated with lead halide perovskite nanocrystals. Chem. Comm. 2020, 56, 8956–8959. [Google Scholar] [CrossRef]
  147. Sen, S.; Kundu, S. Reduced graphene oxide (rGO) decorated ZnO-SnO2: A ternary nanocomposite towards improved low concentration VOC sensing performance. J. Alloys Compd. 2021, 881, 160406. [Google Scholar] [CrossRef]
  148. Caccami, M.C.; Miozzi, C.; Mulla, M.; Natale, C.; Marrocco, G. An epidermal graphene oxide-based RFID sensor for the wireless analysis of human breath. In Proceedings of the IEEE International Conference on RFID Technology & Application (RFID-TA), Warsaw, Poland, 20–22 September 2017. [Google Scholar]
  149. Caccami, M.C.; Mulla, M.; Occhiuzzi, C.; Natale, C.; Marrocco, G. Design and Experimentation of a Batteryless On-Skin RFID Graphene-Oxide Sensor for the Monitoring and Discrimination of Breath Anomalies. IEEE Sens. J. 2018, 18, 8893–8900. [Google Scholar] [CrossRef]
  150. Ray, B.; Parmar, S.; Vijayan, V.; Vishwakarma, S.; Datar, S. Detection of trace volatile organic compounds in spiked breath samples: A leap towards breathomics. Nanotechnology 2022, 33, 205505. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Schematic of a generic system based on graphene sensors for exhaled air analysis.
Figure 1. Schematic of a generic system based on graphene sensors for exhaled air analysis.
Sensors 23 09271 g001
Figure 2. Number of scientific publications indexed in Web of Science for the past 10 years under the scope “Graphene Sensors” (circle markers) and the scope “Graphene Sensors and Biomarkers” (square markers).
Figure 2. Number of scientific publications indexed in Web of Science for the past 10 years under the scope “Graphene Sensors” (circle markers) and the scope “Graphene Sensors and Biomarkers” (square markers).
Sensors 23 09271 g002
Figure 3. Six main pathologies are addressed under the scope of this review paper.
Figure 3. Six main pathologies are addressed under the scope of this review paper.
Sensors 23 09271 g003
Table 1. Details on the procedures adopted in each one of the studies addressed in the section “Asthma and COPD”, including the target biomarkers, the considered population, the developed sensors and the respective bibliographic source.
Table 1. Details on the procedures adopted in each one of the studies addressed in the section “Asthma and COPD”, including the target biomarkers, the considered population, the developed sensors and the respective bibliographic source.
Target BiomarkerPopulationSensorsReferences
NameHMDBTargetNotesSensitivityDetection LimitsResponse/Recovery TimeAccuracy
NitriteHMDB0002786Human subjectsExhaled air condensate100%[72]
NitriteHMDB0002786Human subjectsExhaled air condensate100%[73]
AcetoneHMDB0001659Standard solutionsTwo known concentrations1000–2000 ppmv100/–s[74]
PropanolHMDB0000820Standard solutionsKnown concentrations (150–450 ppmv)300 ppmv4 ppmv156.8595%[75]
HexaneHMDB0029600Standard solutionsSolutions of three compounds[76]
AcetaldehydeHMDB0000990Standard solutionsGaseous samples1.012–1.04330–70/45–85 s[77]
BenzaldehydeHMDB0006115Standard solutionsGaseous samples1.22770.03 nM10/–s[78]
IsopreneHMDB0253673Standard solutionsGaseous samples (5–160 ppmv)237 ppbv[79]
NonanalHMDB0059835Standard solutionsBinary mixture25 ppmv61–200/97–416 s[80]
Table 2. Details on the procedures adopted in each one of the studies addressed in the section “Chronic Kidney Diseases”, including the target biomarkers, the considered population, the developed sensors and the respective bibliographic source.
Table 2. Details on the procedures adopted in each one of the studies addressed in the section “Chronic Kidney Diseases”, including the target biomarkers, the considered population, the developed sensors and the respective bibliographic source.
Target BiomarkerPopulationSensorsReferences
NameHMDBTargetNotesSensitivityDetection LimitsResponse/Recovery TimeAccuracy
AmmoniaHMDB0000051Standard solutions + Synthetic Breath24 Levels of concentration–/–91.7%[88]
IsopreneHMDB0253673Standard solutionsKnown concentrations–/–[89]
PentanalHMDB0031206
HexanalHMDB0005994
HeptanalHMDB0031475
AcetoneHMDB0001659Standard solutionsGaseous samples2.82 ppbv–/–[90]
EthanolHMDB0000108
AcetoneHMDB0001659Standard solutionsSolutions of 1 ppmv1.7100 ppbv11.5–13.5/–s[91]
AmmoniaHMDB0000051Standard solutionsKnown concentrations0.2 ppbv–12 ppmv48/234 s[95]
AcetoneHMDB0001659Standard solutionsKnown concentrations (30–210 ppmv)82 ppbv190–250/–s[96]
EthanolHMDB0000108
AmmoniaHMDB0000051
Table 3. Details on the procedures adopted in each one of the studies addressed in the section “Diabetes”, including the target biomarkers, the considered population, the developed sensors and the respective bibliographic source.
Table 3. Details on the procedures adopted in each one of the studies addressed in the section “Diabetes”, including the target biomarkers, the considered population, the developed sensors and the respective bibliographic source.
Target BiomarkerPopulationSensorsReferences
NameHMDBTargetNotesSensitivityDetection LimitsResponse/Recovery TimeAccuracy
AcetoneHMDB0001659Standard solutions6.280.25–30 ppmv–/–[104]
AcetoneHMDB0001659Human subjects30 Volunteers: 17 diabetic patients and 13 healthy individuals5.6610/12 s60%[105]
AcetoneHMDB0001659Human subjects60 Volunteers: 30 diabetic patients and 30 healthy individuals0–3 ppmv–/–70%[106]
AcetoneHMDB0001659Synthetic breathFour distinct solutions0.5–3.5400 ppbv–80 ppmv–/–100%[107]
AcetoneHMDB0001659Standard solutions10.0>100 ppbv–/–[108]
AcetoneHMDB0001659Human subjects7.8<1 ppmv10/30 s[109]
AcetoneHMDB0001659Standard solutionsMixture of known concentrations1–100 ppmv–/–[110]
MethanolHMDB0001875
EthanolHMDB0000108
AcetoneHMDB0001659Standard solutions1–5 ppmv1.11/41.25 s[111]
Table 4. Details on the procedures adopted in each one of the studies addressed in the section “Gastric Cancer”, including the target biomarkers, the considered population, the developed sensors and the respective bibliographic source.
Table 4. Details on the procedures adopted in each one of the studies addressed in the section “Gastric Cancer”, including the target biomarkers, the considered population, the developed sensors and the respective bibliographic source.
Target BiomarkerPopulationSensorsReferences
NameHMDBTargetNotesSensitivityDetection LimitsResponse/Recovery TimeAccuracy
Acetic acidHMDB0000042Standard solutionsThree known concentrations30.30.04 ppmv[50]
AcetoneHMDB0001659Standard solutionsConcentrations up to 6 ppmv1 ppmv15/75[119]
IsopreneHMDB0253673
TolueneHMDB0034168
2-MethylhexaneHMDB0245230Synthetic breathMixture of 14 compounds at known concentrations83%–/–92%[116]
2-MethylpentaneHMDB0061884
3-MethylpentaneHMDB0061885
DodecaneHMDB0031444
TetradecaneHMDB0059907
MentholHMDB0003352
Phenyl acetateHMDB0040733
HexanolHMDB0012971
Pivalic acidHMDB0041992
3-MethylhexaneHMDB0245932
2,3-DimethylpentaneHMDB0245455
HexaneHMDB0029600
IsopreneHMDB0253673
AcetoneHMDB0001659
AcetoneHMDB0001659Standard solutionsConcentrations up to 800 ppmv35 ppmv[121]
Table 5. Details on the procedures adopted in each one of the studies addressed in the section “Lung Cancer”, including the target biomarkers, the considered population, the developed sensors and the respective bibliographic source.
Table 5. Details on the procedures adopted in each one of the studies addressed in the section “Lung Cancer”, including the target biomarkers, the considered population, the developed sensors and the respective bibliographic source.
Target BiomarkerPopulationSensorsReferences
NameHMDBTargetNotesSensitivityDetection LimitsResponse/Recovery TimeAccuracy
ButyraldehydeHMDB0003543Standard solutionsKnown concentrations1–20 ppbv–/–[135]
TetrahydrofuranHMDB0303508
AcetonitrileHMDB0061869
HeptaneHMDB0031447
HexanalHMDB0005994
BenzeneHMDB0001505
PentaneHMDB0029603
2-ButanoneHMDB0000474
FuranHMDB0013785
DecaneHMDB0031450Standard solutionsKnown concentrations1 ppmv15/90[136]
HeptaneHMDB003144719/48
DecaneHMDB0031450Standard solutionsKnown concentrations0.2 ppmv28/37 s[137]
AcetoneHMDB0001659Human subjects108 Volunteers: 48 healthy individuals and 60 lung cancer patients0.05–10 ppmv60/180 s96%[138]
IsopreneHMDB0253673
AmmoniaHMDB0000051
Ethanol Synthetic breathKnown concentrations0.5 ppmv50/60 s[139]
AcetoneHMDB0001659
FormaldehydeHMDB0001426Synthetic breathLung cancer samples with 83 ppbv and healthy samples with 49 ppbv100 ppbv10 ppbv[140]
Table 6. Details on the procedures adopted in each one of the studies addressed in the section “Sleep Apnoea”, including the target biomarkers, the considered population, the developed sensors and the respective bibliographic source.
Table 6. Details on the procedures adopted in each one of the studies addressed in the section “Sleep Apnoea”, including the target biomarkers, the considered population, the developed sensors and the respective bibliographic source.
Target BiomarkerPopulationSensorsReferences
NameHMDBTargetNotesSensitivityDetection LimitsResponse/Recovery TimeAccuracy
TolueneHMDB0034168Standard solutionsConcentrations of 2, 4, 6 and 8 ppmv0.5–/–[146]
AcetoneHMDB0001659Standard solutionsConcentrations of 1–10 ppmv0.675 ppmv10/10091%[147]
IsopropanolHMDB0000863Human subjectsBreath spiked with 0.5 ppmv of several compounds0.5 ppmv-92.8–96%[150]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Moura, P.C.; Ribeiro, P.A.; Raposo, M.; Vassilenko, V. The State of the Art on Graphene-Based Sensors for Human Health Monitoring through Breath Biomarkers. Sensors 2023, 23, 9271. https://doi.org/10.3390/s23229271

AMA Style

Moura PC, Ribeiro PA, Raposo M, Vassilenko V. The State of the Art on Graphene-Based Sensors for Human Health Monitoring through Breath Biomarkers. Sensors. 2023; 23(22):9271. https://doi.org/10.3390/s23229271

Chicago/Turabian Style

Moura, Pedro Catalão, Paulo António Ribeiro, Maria Raposo, and Valentina Vassilenko. 2023. "The State of the Art on Graphene-Based Sensors for Human Health Monitoring through Breath Biomarkers" Sensors 23, no. 22: 9271. https://doi.org/10.3390/s23229271

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop