Article

Study of Rotation and Bending Effects on a Flexible Hybrid Implanted Power Transfer and Wireless Antenna System

1
Electrical Engineering Department, Applied Science Private University, Amman 11931, Jordan
2
School of Electrical Engineering and Computer Science, University of North Dakota, Grand Forks, ND 58202, USA
3
Phoenix Analysis and Design Technologies, Tempe, AZ 85284, USA
*
Author to whom correspondence should be addressed.
Sensors 2020, 20(5), 1368; https://doi.org/10.3390/s20051368
Submission received: 13 January 2020 / Revised: 19 February 2020 / Accepted: 26 February 2020 / Published: 2 March 2020
(This article belongs to the Special Issue Wearable Antennas)

Abstract

:
We present rotational misalignment and bending effects on a hybrid system to transfer power and data wirelessly for an implantable device. The proposed system consists of a high-frequency coil (13.56 MHz) to transfer power and an ultra-high frequency antenna (905 MHz) for data communication. The system performance and the transmitted power were studied under two misalignment conditions: (1) receiver rotation around itself with reference to the transmitter, and (2) bending of the implanted receiver under three different radii. Implanted receiver was printed on a flexible Kapton substrate and placed inside a layered body tissue model at a 30 mm depth. It is shown that the inductive link is stable under rotational misalignment and three bending conditions, whereas the communication data link is suitable to be used if the rotation angle is less than 75° or larger than 150°. The results show that the resonance frequency varies by 1.6%, 11.05%, and 6.62% for the bending radii of 120 mm, 80 mm, and 40 mm, respectively. Moreover, transmission efficiency varies by 4.3% for the bending radius of 120 mm. Decreasing the bending radius has more effects on antenna transmission efficiency that may cause severe losses in the communication link.

1. Introduction

Recently, industrial and academic researchers have shown significant interest in the field of designing unified systems consisting of wireless power transfer (WPT) and a communication data links, especially for biomedical applications.
In the last few years, research on implanted and flexible electronics has been focused on electrodes, substrate materials, and relevant electronic components. The authors of Reference [1] presented a slot-patch flexible antenna working at 434 MHz, used as an implant for animals and ingestible for humans. In Reference [2], the authors present a near-field capacitive coupling to transfer 100 mW wirelessly to a flexible power receiver. However, using a capacitive coupled powering scheme needs to be investigated for long term subcutaneous implant applications to limit its harmful effect. In Reference [3], the authors designed a fully implantable stimulator with WPT and data transmission. Their design consisted of Litz wire and it was not flexible. The authors in Reference [4] fabricated a flexible printed dual-band dipole antenna, operating at 900 MHz/2.44 GHz, a low-cost fabrication process. This antenna was considered a receiving antenna and was connected to a rectifier. The authors in References [5,6] analyzed the efficiency and investigated the effects of the shape on magnetic flux density in wireless power transfer systems. The authors in Reference [7] designed a printed dipole antenna on a flexible substrate for the ultra-high frequency (UHF) band. The total dimension was 120 mm × 50 mm × 0.05 mm.
The critical challenge in implantable devices is miniaturization. To minimize the footprint of the implanted wireless system, we introduced a hybrid system that combines coils used for wireless power transfer at 13.56 MHz and antennas for data communication at 905 MHz. The hybrid system was presented in Reference [8]. The implanted system was named “IM”, and the external antenna/coil combination, which was placed on the skin layer outside the body, was named “EX”. Ideally, the maximum coupling occurred when EX and IM were perfectly aligned. However, various misalignments may occur. From Reference [9], we studied the effect of lateral and angular misalignments on the hybrid system design. We analyzed the misalignment effects on the link budget in Reference [10].
The rotational misalignment and bending are two conditions that may happen because of changes in the IM location due to body movement or inaccuracy in the placement of the EX. Moreover, bending of the IM can easily occur due to the person’s movement or anatomical conditions, such as skin mobility and variations in the thickness of subcutaneous fatty tissue [11]. Additionally, conformal and flexible shapes are of interest for implanted devices in order to minimize the discomfort and scar effects, which may require bending of the IM. This work focuses on the study of the rotational misalignment and bending effect on the IM performance.
The paper is organized as follows. Section 2 provides the antenna and coil design. The rotational misalignment study based on numerical simulation is presented in Section 3. Bending effects were studied numerically in Section 4. In Section 5, fabrication and measurements results are provided. Conclusions are summarized in Section 6. Information about phantom fabrication and measurement is given in Appendix A.

2. Design

The proposed system consisted of a hybrid coil/antenna named “EX”, which was placed directly at the outer surface of the body, and another coil/antenna system named “IM”, which was embedded inside a layer of muscle at 30 mm distance away from the EX. The EX and IM microstrip antennas had a G-shape profile (Figure 1).
The coils had a spiral shape. Details of the design and optimization of the dimensions of the EX and IM are presented in Reference [12]. IM was considered to be printed on flexible Kapton material. Kapton was chosen due to its specification, and, in addition to its flexibility and biocompatibility, it has a low loss factor over a wide frequency range, high tensile strength, and dielectric strength [13]. This Kapton substrate (εr = 3.9, σ = 0.00524 siemens/m) is considered to be covered by a layer of silicone material. Silicon is biocompatible and is needed for an implanted device. A drawing of the hybrid system design is shown in Figure 1. The location of the shorting pin was chosen such that it tuned the EX and IM antennas to operate at 905 MHz. The locations of the shorting pins were optimized for achieving the maximum power efficiency for the selected antenna dimension parameters for both EX and IM parts, and the results are summarized in Table 1. The origin of the coordinate system was assumed to be at the antenna’s center.
The body was modeled as a layered structure of skin, fat, and muscle. The electromagnetic properties of each layer at each frequency were obtained from the Institute of Applied Physics (IFAC) database and are summarized in Table 2. The thickness of the layers is shown in Figure 1.

3. Rotational Misalignment

The WPT technique, based on the inductive links, is one of the promising solutions for powering biomedical implanted devices. Ensuring a stable power transfer and data communication in implanted device, under conditions, such as misalignment, is challenging. Therefore, the performance of our proposed hybrid system is investigated and studied in two cases:
First, since the G-shape was not symmetrical in the xy plane, we needed to investigate the performance of the hybrid system by rotating the IM combination around the z-axis while keeping the EX plane unchanged, as shown in Figure 2. The rotating angle (Φ) was changed from 0° to 360°. The distance between the EX and IM combinations was kept at 30 mm. This study was based on simulation. For this part of the study, the Kapton layer was considered to have 0.8 mm thickness.
We studied the changes in the coupling coefficient k, reflection coefficients S11 (Tx), S22 (IM), transmission coefficient S21, and transmission efficiency η. ANSYS HFSS (high frequency structure simulator) software was used to simulate the G-shape antenna, whereas ANSYS Maxwell was used for simulating the coils around 13.56 MHz, then the simulated design was imported to ANSYS Twin Builder to calculate S21 and η = |S21|2 × 100%. All simulations were done using the 2019R2 version of the above-mentioned simulation tools.
The results of the simulation of the bending scenario are shown in Figure 3. These graphs show the relationship between S21 and η versus Φ for antenna and coil pairs when the IM was rotated around the z-axis. The tissue model shown in Figure 1 was used. It is noted that S21 and η for the coil set were not affected by angular misalignment due to their symmetrical shape. However, S21 values for the antenna pair were more stable (S21 varied around 20% from the perfectly aligned case) for misalignment angles less than 75° or more than 150°. If the angles were not within these ranges of S21 drops, the worst case occurred at 105°.

4. Bending

The impact of bending the IM coil/antenna on the performance was investigated under three bending conditions (Rad = 40 mm, 80 mm, and 120 mm), as shown in Figure 4. Figure 5 shows the simulation results of the effects of bending on the EX and IM performance.
Figure 5 shows the simulation results of S22 and S21 for the IM antenna based on the tissue model shown in Figure 1 under bending conditions. Table 3 shows these parameters at the resonance frequency for various bending radii. It can be noted that the IM resonance frequency (foIM) varied as much as –100 MHz at Rad = 80 mm to +15 MHz at Rad = 120 mm, compared to its value at foIM for the flat case. Please note that the larger Rad value, the closer the coil was to the flat case.
While S21 did not vary more than 1.57 dB, bending the IM had significant effects on the resonance frequency of the antenna and the level of S22. This value was increased by 15.51 dB, 15.89 dB, and 12.74 dB at Rad values of 40 mm, 80 mm, and 120 mm, respectively. Moreover, the –9 dB bandwidths of antennas are given in Table 3. It can be seen the maximum bandwidth happened at Rad = 120 mm.
Table 4 shows coupling (k) values under all bending condition cases at 13.56 MHz. It can be noted that bending had a very small and negligible effect on the performance of the inductive link.

5. Fabrication and Measurements

The EX and IM were fabricated using a Voltera V-One automated printed circuit board (PCB) printer system [14]. A flexible conductive ink was used that was suitable for printing on a flexible substrate, such as Kapton (DuPont, Wilmington, DE, USA) (polyimide), polycarbonate, and polyethylene terephthalate (PET). This specific ink had a curing temperature of 140 °C for 10 min or 120 °C for 30 min of curing. The printed prototypes of the EX and IM are shown in Figure 6. To create it, the shorting pin a hole was created by drilling and filled by the same conductor ink. It is important to note that we were not able to obtain Kapton at 0.8 mm. The thickness of the Kapton substrate used for measurement was 0.15 mm. To compare the results, a new set of simulations was conducted based on the corresponding design. While the thickness had no effect on the coil operation, it may have caused changes in the frequency of the antenna.
To mimic the effect of tissue on the implanted system, we used a muscle phantom. A phantom is a tissue mimicking material. In general, phantoms should represent the dielectric permittivity and conductivity (losses) of the tissue of interest at the working frequency. The muscle phantom was created based on ingredients and steps described in Reference [15] to represent the muscle tissue model; more details are mentioned in Index I.
Figure 7 and Figure 8 show the simulated and measured values of S22 and S21 for the IM antenna, respectively. There is a small difference between the measured and simulated resonance frequencies. S22 varied from 1.35% at Rad = 80 mm to 6.84% at Rad = 40 mm. This could be because the IM was not completely covered by the phantom in the back of the hybrid design.

6. Conclusions

In this paper, we studied the rotation misalignment and bending effects on the performance of a hybrid power transfer and communication system design. Two main conclusions are as follows. First, for the inductive link, the system was not affected by angular misalignment because of the symmetry of the coils. Additionally, the bending did not have much of an effect on the coil system. Second, for the data communication link through antennas, the system performance was acceptable if the rotation angle was less than 75° or larger than 150°. Moreover, the results show that the resonance frequency varied from the flat shape by 1.6%, 11.05%, and 6.62% at Rad = 120 mm, 80 mm, and 40 mm, respectively, whereas the maximum variation of S21 was around 4.3% at Rad = 120 mm. However, increasing the bending (decreasing Rad) may have caused severe losses in the communication link and is not recommended.

Author Contributions

Conceptualization, R.S. and S.N.; methodology, R.S. and S.N.; simulation, R.S.; validation, M.H.; investigation, R.S., M.H., and S.N.; resources, S.N.; writing—original draft preparation, R.S. and M.H.; writing—review and editing, S.N. and M.H.; visualization, R.S., M.H., and S.N.; supervision, S.N.; project administration, S.N.; funding acquisition, S.N. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the North Dakota Department of Commerce Venture Grant, grant number 16-02-J1-112.

Acknowledgments

Authors would like to acknowledge the University of North Dakota College of Engineering and Mines, where this research work took place.

Conflicts of Interest

The authors declare no conflict of interest.

Appendix A. Phantom Ingredients and Fabrication (Muscle Type)

a. Muscle Phantom Ingredients and Permittivity Measurements:
Table A1 summarizes the material type and amount needed for muscle phantom fabrication. The phantom material was measured at room temperature using a Keysight 85070E dielectric probe. The results are shown in Figure A1.
Table A1. Ingredients of muscle tissue-mimicking phantom.
Table A1. Ingredients of muscle tissue-mimicking phantom.
TypeWaterGelatinNaClOilDetergent Ultra IvoryFood Coloring
Ingredients (g)23034.11.235401.3
Figure A1. Measured relative permittivity and loss tangent of fabricated muscle phantom.
Figure A1. Measured relative permittivity and loss tangent of fabricated muscle phantom.
Sensors 20 01368 g0a1
b. Fabrication process:
The fabrication of the muscle phantoms requires the following steps:
  • Mix the gelatin and 100 g de-ionized water in a beaker.
  • Cover the beaker with cling film and heat the mixture up to 80 °C in a hot water jacket or in a double boiler.
  • Leave the mixture to cool until 35 °C.
  • Add the remaining room temperature deionized water and NaCl whilst stirring the mixture slowly.
  • Add the room temperature dish-washing detergent to the mixture and keep stirring slowly.
  • When the mixture reaches 28 °C, add room temperature oil and keep stirring slowly until the mixture becomes homogeneous. At this point, the color of the mixture changes to white. We can also add food coloring at this point.
  • Pour the homogeneous mixture into containers and leave it to solidify overnight.
Stir the mixture slowly to avoid air bubbles when making the tissue-mimicking phantoms. Moreover, the mixture should be handled with care.

References

  1. Nikolayev, D.; Zhadobov, M.; Sauleau, R. Immune-To-Detuning wireless in-body platform for versatile biotelemetry applications. IEEE Trans. Biomed. Circuits Syst. 2019, 13, 403–412. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  2. Jegadeesan, R.; Agarwal, K.; Guo, Y.X.; Yen, S.C.; Thakor, N.V. Wireless power delivery to flexible subcutaneous implants using capacitive coupling. IEEE Trans. Microw. Theory Tech. 2017, 65, 280–292. [Google Scholar] [CrossRef]
  3. Xu, Q.; Hu, D.; Duan, B.; He, J. A fully implantable stimulator with wireless power and data transmission for experimental investigation of epidural spinal cord stimulation. IEEE Trans. Neural Syst. Rehabil. Eng. 2015, 23, 683–692. [Google Scholar] [CrossRef] [PubMed]
  4. Berges, R.; Fadel, L.; Oyhenart, L.; Vigneras, V. A Flexible Printed Dual-Band Antenna Dedicated to RF Energy Harvesting Application. Available online: https://hal.archives-ouvertes.fr/hal-01484377/document (accessed on 31 May 2016).
  5. Haerinia, M.; Mosallanejad, A.; Afjei, E.S. Electromagnetic analysis of different geometry of transmitting coils for wireless power transmission applications. Prog. Electromagn. Res. M 2016, 50, 161–168. [Google Scholar] [CrossRef] [Green Version]
  6. Haerinia, M.; Afjei, E.S. Resonant inductive coupling as a potential means for wireless power transfer to printed spiral coil. J. Electr. Eng. 2016, 16, 68–74. [Google Scholar]
  7. El Maleky, O.; Abdelouahab, F.B.; Essaaidi, M.; Ennasar, M.A. Design of simple printed dipole antenna on flexible substrate for UHF band. Procedia Manuf. 2018, 22, 428–435. [Google Scholar] [CrossRef]
  8. Shadid, R.; Noghanian, S. Hybrid power transfer and wireless antenna system design for biomedical implanted devices. In Proceedings of the 2018 International Applied Computational Electromagnetics Society Symposium (ACES), Denver, CO, USA, 25–29 March 2018; pp. 1–2. [Google Scholar]
  9. Shadid, R.; Noghanian, S. Study of misalignment effects on hybrid power transfer and wireless antenna system for implanted devices. In Proceedings of the 2018 IEEE International Symposium on Antennas and Propagation & USNC/URSI National Radio Science Meeting, Boston, MA, USA, 8–13 July 2018; pp. 2553–2554. [Google Scholar]
  10. Haerinia, M.; Noghanian, S. Analysis of misalignment effects on link budget of an implantable antenna. In Proceedings of the 2019 URSI International Symposium on Electromagnetic Theory (EMTS), San Diego, CA, USA, 27–31 May 2019; pp. 1–4. [Google Scholar]
  11. Fotopoulou, K.; Flynn, B.W. Wireless power transfer in loosely coupled links: Coil misalignment model. IEEE Trans. Magn. 2011, 47, 416–430. [Google Scholar] [CrossRef]
  12. Shadid, R.; Haerinia, M.; Sayan, R.; Noghanian, S. Hybrid inductive power transfer and wireless antenna system for biomedical implanted devices. Prog. Electromagn. Res. C 2018, 88, 77–88. [Google Scholar]
  13. Monne, M.A.; Lan, X.; Chen, M.Y. Material selection and fabrication processes for flexible conformal antennas. Int. J. Antennas Propag. 2018. [Google Scholar] [CrossRef] [Green Version]
  14. Voltera. Available online: http://www.voltera.io (accessed on 11 November 2018).
  15. Yilmaz, T.; Foster, R.; Hao, Y. Broadband tissue mimicking phantoms and a patch resonator for evaluating noninvasive monitoring of blood glucose levels. IEEE Trans. Antennas Propag. 2014, 62, 3064–3075. [Google Scholar] [CrossRef]
Figure 1. Coil and antenna configuration for the external antenna/coil combination (EX) and implanted system (IM) and their location, with respect to the layered body model.
Figure 1. Coil and antenna configuration for the external antenna/coil combination (EX) and implanted system (IM) and their location, with respect to the layered body model.
Sensors 20 01368 g001
Figure 2. Rotational misalignment setting.
Figure 2. Rotational misalignment setting.
Sensors 20 01368 g002
Figure 3. S21 and η under rotational misalignment for the IM around z axis (the IM was at 30 mm depth) for (a) the antenna pair and (b) the coil pair.
Figure 3. S21 and η under rotational misalignment for the IM around z axis (the IM was at 30 mm depth) for (a) the antenna pair and (b) the coil pair.
Sensors 20 01368 g003
Figure 4. The IM hybrid design under three bending conditions. (a) three dimensional views of bent RX, (b) Rad = 40 mm, (c) Rad = 80 mm, and (d) Rad = 120 mm.
Figure 4. The IM hybrid design under three bending conditions. (a) three dimensional views of bent RX, (b) Rad = 40 mm, (c) Rad = 80 mm, and (d) Rad = 120 mm.
Sensors 20 01368 g004
Figure 5. Simulated S21 and S22 for the IM under three bending cases (Kapton thickness of 0.8 mm).
Figure 5. Simulated S21 and S22 for the IM under three bending cases (Kapton thickness of 0.8 mm).
Sensors 20 01368 g005
Figure 6. The fabricated prototype of the EX (left) and IM (right).
Figure 6. The fabricated prototype of the EX (left) and IM (right).
Sensors 20 01368 g006
Figure 7. Simulated and measured IM reflection coefficients (S22) for flat and three bending cases (Kapton thickness of 0.15 mm).
Figure 7. Simulated and measured IM reflection coefficients (S22) for flat and three bending cases (Kapton thickness of 0.15 mm).
Sensors 20 01368 g007
Figure 8. Simulated and measured transmission coefficient (S21) for flat and three bending cases (Kapton thickness of 0.15 mm).
Figure 8. Simulated and measured transmission coefficient (S21) for flat and three bending cases (Kapton thickness of 0.15 mm).
Sensors 20 01368 g008
Table 1. Shorting pin location for 905 MHz resonance.
Table 1. Shorting pin location for 905 MHz resonance.
Antenna Locationx (mm)y (mm)
EX6.75−13
IM0.30−9
Table 2. Layered body tissue assumed electromagnetic properties at different frequencies.
Table 2. Layered body tissue assumed electromagnetic properties at different frequencies.
Frequency13.56 MHz905 MHz
TissueMuscleFatSkinMuscleFatSkin
εr138.411.83285.25555.4641.4
Loss tangent6.013.401.110.340.190.42
Table 3. Antenna characteristics under bending conditions.
Table 3. Antenna characteristics under bending conditions.
Rad (mm)f0IM(MHz)S22(dB)S21(dB)Bandwidth (Hz)
40845−10.31−37.12202
80805−9.93−37.80148
120920−13.08−38.25318
flat905−25.82−36.68372
Table 4. Coils coupling coefficient (k) under bending conditions.
Table 4. Coils coupling coefficient (k) under bending conditions.
Rad (mm)k
400.054
800.052
1200.054
flat0.057

Share and Cite

MDPI and ACS Style

Shadid, R.; Haerinia, M.; Noghanian, S. Study of Rotation and Bending Effects on a Flexible Hybrid Implanted Power Transfer and Wireless Antenna System. Sensors 2020, 20, 1368. https://doi.org/10.3390/s20051368

AMA Style