Next Article in Journal
Tapered Optical Fiber for Hydrogen Sensing Application Based on Molybdenum Trioxide (MoO3)
Previous Article in Journal
ZnO and TiO2 Assisted Photocatalytic Degradation of Butachlor in Aqueous Solution under Visible Light
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Proceeding Paper

A System-on-Chip Assay for Bilirubin Levels Measurement in Whole Blood †

by
Jean Pierre Ndabakuranye
1,2,*,
Steven Prawer
1 and
Arman Ahnood
1,2
1
School of Physics, University of Melbourne, Parkville, VIC 3010, Australia
2
School of Engineering, RMIT University, Bundoora, VIC 3083, Australia
*
Author to whom correspondence should be addressed.
Presented at the 8th International Electronic Conference on Sensors and Applications, 1–15 November 2021; Available online: https://ecsa-8.sciforum.net.
Eng. Proc. 2021, 10(1), 74; https://doi.org/10.3390/ecsa-8-11295
Published: 1 November 2021

Abstract

:
Bilirubin (BR) is clinically confirmed as a biomarker for liver health and is used to assess the prognosis of cirrhosis. Optical and chemical methods have been utilized for blood BR biosensing. While optical methods offer real-time monitoring and are handy and immune to infection, measurements may not be practical due to the instrument complexity and space requirements. This study investigated the dual-wavelength (DWL) technique for BR estimation using a system-on-chip (SoC). The SoC includes an optical module with blue (455 nm) and green (530 nm) LEDs which were used for DWL measurement. Porcine blood was used as a surrogate of human blood and BR levels were kept within the pathophysiological ranges projected from healthy individuals (<1.2 mg/dL) to cirrhotic patients (up to 50 mg/dL). Our findings show a high BR sensitivity in blood. This lays the groundwork for point-of-care testing for BR levels, primarily for hyperbilirubinemia infants and cirrhotic adults out in homes or in-community settings.

1. Introduction

Bilirubin is a by-product of heme catabolism [1], where less than 1.2 mg/dL is always present in the blood of healthy individuals [2]. It exists as conjugated and unconjugated bilirubin (UCB) in the bloodstream. Unconjugated bilirubin (UCB) is toxic and water-insoluble; hence, it should be excreted. The excretion starts with UCB binding with albumin and is transported to the liver. At this point, UCB is enzymatically converted into conjugated bilirubin (CB). The latter is less toxic and water-soluble, allowing for renal or intestinal excretion [3]. However, in the case of pathophysiological events, bilirubin levels are elevated, and the liver fails to live up to its conjugation ability, leading to irreversible neurological damage or death [4]. High bilirubin levels have been correlated with hepatic and hemolytic disorders [5,6].
Several studies on BR levels estimation have been reported [7,8], but these techniques suffer from the instrument complexity and cost. This mini-paper investigated the dual-wavelength (DWL) technique for BR estimation using a system-on-chip (SoC).

2. Materials and Methods

Materials: Bilirubin (PN: 14370), Dimethyl Sulfoxide (DMSO) and sodium citrate (NaCHO) commercial standards were supplied from Sigma-Aldrich, Australia.
Blood sample preparation: Porcine blood was procured from the abattoir immediately after sacrifice and was preferred as a prominent replacement for human blood due to its biochemical resemblance [9] and low cost. Solutions were prepared by mixing BR at varied concentrations and blood with anticoagulant (4% w/v)-to-blood ratio of 1:9 (v/v) [10]. BR-blood samples were stored at ~5 °C in the dark to prevent photodegradation. For extended preparation steps, refer to the study performed by Ndabakuranye et al. [11].
Optical measurement: A SoC platform was used for BR measurement. The setup consisted of a MAX86916 optical module, a MAX32630 host and an optical stage designed to operate in transflection mode. Transflection mode implies that a rear optical reflector (E39-R42, Omron) was used on top of the sample. Therefore, the spectral response contained both light intensity backscattered from the sample and light transmitted through the sample then reflected by the reflector to be retransmitted back through the sample, thereby doubling the path length (Figure 1). Major features of the system-on-chip are summarized in Table 1.
Feature selection: Several features were optimally selected using theoretical investigations and simulations to ensure the accuracy, reliability, and safety of bilirubin measurements. These features are summarized in Table 2.

3. Results and Discussions

The analytical (470 nm) and reference (525 nm) wavelengths were obtained by analyzing the distinct optical signatures of blood and BR. It does not significantly absorb green light but strongly absorbs blue light (ε525 = 214, ε470 = 46188 [cm−1 M−1]) [12]. However, although the LEDs on the SoC emit the maximum power at the wavelengths of 455 nm and 530 nm, the LEDs’ emission bands are sufficiently wide to also cover 470 nm and 525 nm, since the LEDs are not ideal monochromatic sources.
The DWL measurement was performed using a revised 2-dimensional Beer’s law as shown in Equation (1) which suggests that BR concentration can be correlated with its absorbance ratio at 470 and 525 nm (t: path length [cm], ε: extinction coefficient, C: concentration).
( A 470 nm A 525 nm ) = [ ε B R ,   470 ε H b ,   470 ε B R ,   525 ε H b ,   525 ] ( C B R C H b ) × t
To investigate BR’s sensitivity in blood, R-parameters were deduced by collecting the MAX86916 ADC counts data and calculated using Equation (2). Figure 1 shows the SoC platform used to measure the ADC data used to compute R-parameters ( R ˜ ). The plot of R-parameters as a function of BR concentrations is shown in Figure 2, and results showed a strong linear relationship with the coefficient of determination greater than 0.991.
R ˜ = log [ A D C   c o u n t B l u e ] log [ A D C   c o u n t G r e e n ]
Although this technique provides a robust and simple way of measuring BR, it may be susceptible to errors due to LEDs’ spectral and spatial distribution inadequacies, residual bilirubin, other hemoglobin forms (COHb or MetHb), SaO2 and Hb levels variability.

4. Conclusions

The feasibility of BR monitoring by the DWL method was investigated, and BR’s sensitivity in blood was explored at pathophysiological ranges (1.2–50 mg/dL) using an Soc. The SoC includes a miniature MAX86916 optical module with integrated signal conditioning and processing capabilities. Results showed a strong correlation between R-parameters and BR concentration. Our findings lay the groundwork for point-of-care testing for BR levels, primarily for hyperbilirubinemia infants and cirrhotic adults out of clinical settings.

Author Contributions

Methodology, J.P.N.; Formal Analysis, J.P.N.; Investigation, J.P.N.; Data Curation, J.P.N.; Original Draft Preparation and Writing, J.P.N.; Conceptualisation, J.P.N., A.A. and S.P.; Writing—review and editing, J.P.N., A.A. and S.P.; Supervision, S.P. and A.A.; Project Administration, S.P. and A.A.; Funding Acquisition, S.P. and A.A. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the Australian Research Council, grant number LP160101052.

Institutional Review Board Statement

Experiments were carried out on samples from the abattoir, and thus, neither ethics application nor written consent was required.

Acknowledgments

Jean Pierre gratefully acknowledges the Melbourne Research Scholarship (the University of Melbourne), The School of Physics (The University of Melbourne) and the STEM college (RMIT University).

Conflicts of Interest

Authors declare that they have no conflicting financial or non-financial interests.

References

  1. Ahmad, H.B.; Ahmad, S.; Shad, M.A.; Hussain, M. Kinetic measurements for photodecomposition of bilirubin. Asian J. Chem. 2013, 25, 7945–7948. [Google Scholar] [CrossRef]
  2. Dufour, D.R.; Lott, J.A.; Nolte, F.S.; Gretch, D.R.; Koff, R.S.; Seeff, L.B. Diagnosis and monitoring of hepatic injury. I. Performance characteristics of laboratory tests. Clin. Chem. 2000, 46, 2027–2049. [Google Scholar] [CrossRef] [PubMed]
  3. Berk, P.D. Bilirubin metabolism and the hereditary hyperbilirubinemias. Semin. Liver Dis. 1994, 14, 321–322. [Google Scholar] [CrossRef] [PubMed]
  4. AAP. Management of hyperbilirubinemia in the newborn infant 35 or more weeks of gestation. Pediatrics 2004, 114, 297–316. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  5. Lopez-Velazquez, J.A.; Chavez-Tapia, N.C.; Ponciano-Rodriguez, G.; Sanchez-Valle, V.; Caldwell, S.H.; Uribe, M.; Mendez-Sanchez, N. Bilirubin alone as a biomarker for short-term mortality in acute-on-chronic liver failure: An important prognostic indicator. Ann. Hepatol. 2014, 13, 98–104. [Google Scholar] [CrossRef]
  6. D’Amico, G.; Garcia Tsao, G.; Pagliaro, L. Natural history and prognostic indicators of survival in cirrhosis: A systematic review of 118 studies. J. Hepatol. 2006, 44, 217–231. [Google Scholar] [CrossRef] [PubMed]
  7. Hulzebos, C.V.; Vitek, L.; Zabetta, C.D.C.; Dvořák, A.; Schenk, P.; van der Hagen, E.A.; Cobbaert, C.; Tiribelli, C. Diagnostic methods for neonatal hyperbilirubinemia: Benefits, limitations, requirements, and novel developments. Pediatr. Res. 2021, 90, 277–283. [Google Scholar] [CrossRef] [PubMed]
  8. Rawal, R.; Kharangarh, P.R.; Dawra, S.; Tomar, M.; Gupta, V.; Pundir, C. A comprehensive review of bilirubin determination methods with special emphasis on biosensors. Process Biochem. 2020, 89, 165–174. [Google Scholar] [CrossRef]
  9. Pond, W.G.; Mersmann, H.J. Biology of the Domestic Pig; Comstock Pub. Associates, Cornell University Press: Ithaca, NY, USA, 2001. [Google Scholar]
  10. Sorapukdee, S.; Narunatsopanon, S. Comparative study on compositions and functional properties of porcine, chicken and duck blood. Korean J. Food Sci. Anim. Resour. 2017, 37, 228–241. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  11. Ndabakuranye, J.P.; Rajapaksa, A.E.; Burchall, G.; Li, S.; Prawer, S.; Ahnood, A. A novel optical assay system for bilirubin concentration measurement in whole blood. IEEE Trans. Biomed. Eng. 2021, in press. [CrossRef]
  12. Yao, J.; Wang, L.V. Photoacoustic microscopy. Laser Photonics Rev. 2013, 7, 758–778. [Google Scholar] [CrossRef]
Figure 1. Illustration of the DWL measurement in transflection mode.
Figure 1. Illustration of the DWL measurement in transflection mode.
Engproc 10 00074 g001
Figure 2. The graph of R-parameters as a function of bilirubin concentrations as obtained from DWL measurements on whole blood using the SoC platform.
Figure 2. The graph of R-parameters as a function of bilirubin concentrations as obtained from DWL measurements on whole blood using the SoC platform.
Engproc 10 00074 g002
Table 1. Major features of MAX86916 optical module.
Table 1. Major features of MAX86916 optical module.
FeatureValue
ADC resolution19-bit
Size3.5 × 7.0 × 1.5 mm
Spectral sensitivity400–1100 nm
Data communication protocolI2C
Radiant sensitive area1.51 mm2
Table 2. Optimal features for DWL measurement.
Table 2. Optimal features for DWL measurement.
Method ParameterOptimized Parameter
Operating wavelengthsBlue and green
Path length200 µm
SolventDMSO
Bilirubin concentration range1.2–50 mg/dL
AnticoagulantSodium citrate
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Ndabakuranye, J.P.; Prawer, S.; Ahnood, A. A System-on-Chip Assay for Bilirubin Levels Measurement in Whole Blood. Eng. Proc. 2021, 10, 74. https://doi.org/10.3390/ecsa-8-11295

AMA Style

Ndabakuranye JP, Prawer S, Ahnood A. A System-on-Chip Assay for Bilirubin Levels Measurement in Whole Blood. Engineering Proceedings. 2021; 10(1):74. https://doi.org/10.3390/ecsa-8-11295

Chicago/Turabian Style

Ndabakuranye, Jean Pierre, Steven Prawer, and Arman Ahnood. 2021. "A System-on-Chip Assay for Bilirubin Levels Measurement in Whole Blood" Engineering Proceedings 10, no. 1: 74. https://doi.org/10.3390/ecsa-8-11295

Article Metrics

Back to TopTop