A Polarization-Insensitive Recirculating Delayed Self-Heterodyne Method for Sub-Kilohertz Laser Linewidth Measurement

Photonics 2021, 8(5), 137; https://doi.org/10.3390/photonics8050137
by Jing Gao 1,2,3, Dongdong Jiao 1,3, Xue Deng 1,3, Jie Liu 1,3, Linbo Zhang 1,2,3, Qi Zang 1,2,3, Xiang Zhang 1,2,3, Tao Liu 1,3,* and Shougang Zhang 1,3
Reviewer 1: Anonymous
Reviewer 2: Anonymous
Reviewer 3: Anonymous
Photonics 2021, 8(5), 137; https://doi.org/10.3390/photonics8050137
Submission received: 15 March 2021 / Revised: 15 April 2021 / Accepted: 16 April 2021 / Published: 23 April 2021
(This article belongs to the Section Lasers, Light Sources and Sensors)

Round 1

Reviewer 1 Report

Line 57: optical isolator and not optical isolate

Author Response

Thank you very much for your reviews and comments to our manuscript (manuscript ID: photonics-1163770, Type: Communication, Title: A polarization-insensitive recirculating delayed self-heterodyne method for sub-kilohertz laser linewidth measurement). We have carefully revised the manuscript according to the comments.

Author Response File: Author Response.docx

Reviewer 2 Report

The manuscript by J. Gao, et al. describes a polarization-insensitive recirculating delayed self-heterodyne method for precision measurement of 
sub-kHz linewidths of a fiber laser. The method was demonstrated clearly by the detailed calculation and experimental results. Before the consideration of publications, I have a few comments that need to be addressed.

  1. In the introduction, the author mentioned the improved delayed self-heterodyne method (DSHM) for narrow-linewidth measurements, such as the recirculating-DSHM used in the paper. This conception and method have been proposed and investigated very early by H. Tsuchida. Tsuchida H. Simple technique for improving the resolution of the delayed self-heterodyne method[J]. Optics letters, 1990, 15(11): 640-642. 
  2. The experimental setup shown in Fig. 1 is not intuitive enough. Could the author add the arrows to depict the light propagation direction?
  3. In Fig. 2 and Fig. 3, it is better to add some gab between (a) and (b), (b)and (c) to avoid any misunderstanding by the readers.
  4. When the authors present the linewidth results, is it for 3-dB or 20-dB linewidth? It should be clarified.
  5. The authors compared the Lorentzian linewidths and Gaussian linewidths for high-order beat notes (S10-S13). Could the author also add the results of low-order beat notes S1-S9? I presume that the low-order (maybe S1 and S2) can not give the precision measurement results due to the limited delay length. However, more data in Fig. 6 will help to support their theoretical analysis.
  6. In line 57, "isolate" should be "isolator"
  7. In line 58, "mw" should be "mW"

Author Response

Thank you very much for your reviews and comments to our manuscript (manuscript ID: photonics-1163770, Type: Communication, Title: A polarization-insensitive recirculating delayed self-heterodyne method for sub-kilohertz laser linewidth measurement). We have carefully revised the manuscript according to the comments.

Author Response File: Author Response.docx

Reviewer 3 Report

The paper presents a polarization-insensitive recirculating delayed self-heterodyne method (PI-RDSHM) used to measure the linewidth of sub-kilohertz lasers. Conventional DSHM systems have the disadvantage of requiring polarization controller to control the state of polarization otherwise the spectrum SNR could degrade in time. This work propose a combination of three Faraday rotator mirrors to replace the polarization controller and have passive polarization insensitive technique.

The experimental results show that the 10th and 11th order beat-notes are independent of the polarization and linewidth measurement by fitting with a Voigt function is presented.

At Line 128  probably there is a typo regarding the word “with”

Author Response

Thank you very much for your reviews and comments to our manuscript (manuscript ID: photonics-1163770, Type: Communication, Title: A polarization-insensitive recirculating delayed self-heterodyne method for sub-kilohertz laser linewidth measurement). We have carefully revised the manuscript according to the comments.

Author Response File: Author Response.docx