Next Article in Journal
Progress in Plasmonic Sensors as Monitoring Tools for Aquaculture Quality Control
Previous Article in Journal
Sensitive and Rapid Detection of Aspartic Acid with Co3O4-ZnO Nanorods Using Differential Pulse Voltammetry
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

The Integration of Reference Electrode for ISFET Ion Sensors Using Fluorothiophenol-Treated rGO

1
Department of Medical IT Convergence Engineering, Kumoh National Institute of Technology, Gumi 39177, Republic of Korea
2
Center for Nano Bio Development, National NanoFab Center (NNFC), 291 Daehak-ro, Yuseong-gu, Daejeon 34141, Republic of Korea
3
Department of Research and Development, MCK Tech Co., Ltd., Daejeon 34013, Republic of Korea
4
Department of Internal Medicine, Keimyung University School of Medicine, Daegu 41931, Republic of Korea
*
Author to whom correspondence should be addressed.
Biosensors 2023, 13(1), 89; https://doi.org/10.3390/bios13010089
Submission received: 13 December 2022 / Revised: 3 January 2023 / Accepted: 3 January 2023 / Published: 5 January 2023
(This article belongs to the Section Biosensor and Bioelectronic Devices)

Abstract

:
Ion-sensitive field-effect transistors (ISFETs) detect specific ions in solutions that enable straightforward, fast, and inexpensive sensors compared to other benchtop equipment. However, a conventional reference electrode (RE) such as Ag/AgCl is limited on the miniaturization of the sensor. We introduce reduced graphene oxide (rGO), which serves as a new RE, when fluorinated (F-rGO) using fluorothiophenol through the π–π interaction. The circular RE is integrated between a fabricated microscale two-channel ISFET, which is capable of detecting two kinds of ions on an indium tin oxide (ITO) thin-film substrate, using the photolithography process. F-rGO bound to this circular region to function as an RE in the ISFETs sensor, which operated stably in solution and showed a relatively high transconductance (gm) value (1.27 mS), low drift characteristic (3.2 mV), and low hysteresis voltage (±0.05 mV). It detected proton (H+) ions in a buffer solution with high sensitivity (67.1 mV/pH). We successfully detected Na+ (62.1 mV/dec) and K+ (57.6 mV/dec) ions in human patient urine using a two-channel ISFET with the F-rGO RE. The F-rGO RE will be a suitable component in the fabrication of low-cost, mass-produced, and disposable ISFETs sensors.

1. Introduction

Graphene oxide (GO) is generally produced using the modified Hummer’s method. This chemical process is used to oxidize the graphite and is exploited to produce GO, which contains oxygen functional groups [1]. Removing the surface functional groups of GO or the reduction of GO produces reduced graphene oxide (rGO). rGO has been extensively used in diverse applications as a substitute for pristine graphene because of its relatively high production efficiency, electronic conductivity, large surface area, and thermal stability [2,3]. rGO has a close chemical similarity to pristine graphene, although it has a number of sp3 sites [4].
Ion-sensitive field-effect transistors (ISFETs) can be used to detect many kind of ions depending on the ion-selective membrane (ISM) on the gate channel surface [5,6,7,8]. The PVC membrane that makes up the ISM is composed of molecules that react with specific ions. In an ISFET, the ISM is located on the gate surface and reacts with the specific ions that have entered through the membrane and acts as a barrier to prevent other ions from reaching the sensor surface. Only the specific ions functionalized with the membrane can pass through and transfer charge to the gate of the ISFET [6]. These ISFET sensors require a reference electrode (RE) to accurately determine the potential in an electrolyte solution [7,8]. An Ag/AgCl RE has been widely used because it has high reproducibility and stable working characteristics, even in harsh environments [9,10,11,12]. However, Ag/AgCl Res have limitations in terms of miniaturization because the structure requires an Ag wire and a filling solution, which hinders the possibility of integration with ISFETs and leads to high production costs. There is a need for a new RE that can be miniaturized and integrated into the ISFET manufacturing process. We propose a new RE and fabricate an RE-integrated ISFET ion sensor that can stably detect proton (H+), sodium (Na+), and potassium (K+) ions in human urine. Using a radio frequency (RF) sputtering method, we deposited an indium tin oxide (ITO) thin film on silicon substrate and fabricated the uniform ISFETs using a photolithography process. The graphene RE was manufactured using fluorinated rGO (F-rGO RE) treated through the π–π interaction between fluorothiophenol (FTP) and rGO. This F-rGO RE can be manufactured using a semiconductor manufacturing process and produces a miniaturized RE that can be integrated with ISFETs.

2. Materials and Methods

2.1. Fabrication of Graphene RE and ITO Deposition

The average thickness of the rGO (Graphene Supermarket, New York, NY, USA), which consists of three to eight graphene monolayers, was 3 nm. The rGO was sonicated in an FTP solution for 5 min, heated at 180 °C for 20 min, and cooled at 25 °C. Subsequently, it was centrifuged at 3000 rpm at 25 °C to obtain fluorinated rGO (F-rGO). The resulting F-rGO was dispersed in FTP at 1 mg/mL and then dropped onto the RE surface. The excess unadsorbed F-rGO on the substrate was rinsed off with flowing distilled water. The formed F-rGO layer had an area of 1.77 mm2 in contact with the electrolyte.
ITO thin films were deposited on the silicon wafer (Si/SiO2) using RF sputtering equipment (Korea Vacuum Tech. KVS-2000L, Korea). The pressure in the chamber was 5 × 10−6 Torr, and Ar and O2 gas were injected at a ratio of 20:1. Subsequently, the temperature of the substrate was maintained at 100 °C, and the pressure in the chamber was 2 × 10−2 Torr. The internal gas was ionized with a power of 200 W, followed by 10 min of pre-sputtering, and thereafter sputtered to deposit the ITO thin film on the Si/SiO2 substrate for 210 s. After deposition was completed, the Si/SiO2 substrate was held in the chamber for 30 min at 100 °C to ensure the uniformity of the ITO thin film, and then cooled slowly in the chamber in an Ar gas environment.

2.2. Ion-Selective Membrane and Buffer Solution

An ion-selective membrane (ISM) for Na+ ion detection in solution was made from 2 mg sodium ionophore III and 33 mg PVC, with 65 mg bis(2-ethylhexyl) sebacate as a plasticizer, and 100 mg of this mixture was dissolved in 1 mL tetrahydrofuran (THF). In the ISM for K+ ion detection, 2 mg valinomycin and 0.5 mg potassium tetrakis(4-chlorophenyl) borate were added to the lipophilic salt. Then, 32.8 mg PVC and 64.7 mg bis(2-ethylhexyl) sebacate were dissolved in 1 mL THF. The dissolved mixture was dispersed through sonication, and 2 µL was used by drop-casting onto the gate channel surface. Carmody buffer solution (0.2 M boric acid, 0.05 M citric acid and 0.1 M trisodium phosphate) was used as a pH buffer solution adjusted from pH 2 to 12. Ultrapure water (18.2 MΩ⋅cm−1) was used. Na+ and K+ ion solutions were prepared by dissolving NaCl and KCl in a 50 mM Tris-HCl buffer solution. The chemicals were used without any further purification, and pH measurements were performed using a digital pH meter.

2.3. Method of Analysis

The crystallinity of the ITO thin films was evaluated using an X-ray diffractometer (XRD, Rigaku SmartLab, Tokyo, Japan). The Cu-α emission line (λ = 1.5418 Å) was scanned in the range of 20° < 2θ < 70°. The thickness of the ITO thin film was measured using a field-emission scanning electron microscope (FE-SEM, JEOL 7610F, JEOL Ltd, Tokyo, Japan). A Fourier-transform infrared microscope (FT-IR, Bruker Hyperion 2000, Bruker, Billerica, Germany) was used to analyze the surface functional groups of the rGO and F-rGO. The properties of rGO and F-rGO were analyzed by Raman spectroscopy (Renishaw 1000, Renishaw PLC, Gloucestershire, UK) using an argon-ion laser at a wavelength of 514 nm.
The electrical properties of the ISFET sensor in the electrolyte solution were measured using a SourceMeter (Keithley 2400, Keithley instruments, Inc., Cleveland, OH, USA) and characterized by the drain–source current (IDS), drain–source voltage (VDS), and gate–source voltage (VGS) in the electrolyte solution. After the electrolyte solution was exchanged, the ISFET sensor was stabilized for 2 min before the steady-state electrical measurements. All electrical measurements were carried out at 25 °C and biased within the potential window of the ITO and F-rGO electrodes to prevent redox reaction.

3. Results and Discussion

3.1. Evaluation of ITO and F-rGO

The benzene group of FTP is bound to rGO via a π–π bond and functionalizes rGO with fluorine and thiol. The thiol group at the end of the FTP allows the binding of rGO to the Au electrode through thiol bonding (Figure 1a). The F-rGO did not contain solvent residue and was continuous over the entire electrode surface. The rGO exhibited characteristic peaks at 1089, 1261, 1394, 1645, and 3433 cm−1 in the FT-IR spectra, as shown in Figure 1b. These peaks were caused by C–O–C bending, C–O stretching, C–OH stretching, the C=C stretching vibration of the unoxidized graphitic domain, and the OH stretching group, respectively. Compared to rGO, F-rGO showed additional characteristic peaks at 1006 and 1217 cm−1 that were caused by the C–S stretching vibration and C–F stretching vibration in the benzene ring, respectively [13]. The C=C stretching vibration in the benzene ring was shown at 1483 and 1585 cm−1.
The Raman spectra of the rGO and F-rGO are shown in Figure 1c. The D and G bands of the rGO were centered at 1352 and 1596 cm−1, respectively, and the intensity ratio (ID/IG) was 1.42. After fluorination using FTP, the D and G bands were centered at 1349 and 1580 cm−1, respectively, and the intensity ratio (ID/IG) of F-rGO was 0.52. The intensity ratio (ID/IG) of F-rGO decreased because the disordered rGO that existed on the surface was removed and the oxygen functional groups at the edge states were removed by the FTP treatment. The 2D band increased at 2728 cm−1 and the shape of the D band changed after FTP modification. The 2D band is active for crystalline graphitic materials and it is sensitive to the π band in the graphitic electronic structure [14]. The layer of rGO (three to eight graphene monolayers) changed to a light layer (two to five graphene monolayers) after fluorination, and FTP modification was caused by π–π bonding between the benzene ring of FTP and the rGO surface.
Figure S1 is an FE-SEM image of the ITO thin film on the Si/SiO2 substrate showing that the film thickness was 35 nm and its crystallinity had a columnar structure. The pattern showed a peak (2θ = 31°) that appeared normal to the (222) plane, indicating ITO thin film, and it demonstrated the typical ITO diffraction pattern, such as (211), (400), (440), and (622) peaks. The peak intensity ratio of the (222) orientation and the (400) orientation peak (I222/I400) was 1.455 due to insufficient O2 flow in the thin-film deposition process [15]. In an environment where the ratio of O2/Ar does not exceed 1/13 during deposition, depending on the structure of the chamber, the sheet resistance and lattice constant tended to be lowered [16]. However, we set the O2/Ar ratio to 1/20 to increase the resistance of the ITO thin film. As shown previously, high sensitivity to ions has been observed in ITO-ISFET sensors with high gate channel resistance [17].

3.2. Characteristics of the ITO-ISFET and F-rGO RE

The photolithography process was used to fabricate the drain, source, and gate channel. The RE pattern on the Si/SiO2 substrate was fabricated using the same photolithography process, as shown in Figure 2. The gate channel was formed using a photoresist (DNR L300), and ITO was deposited on the Si/SiO2 substrate through RF sputtering. After the ITO gate channel was formed on the Si/SiO2 substrate, the remaining photoresist was lifted off. The open area of the gate channel in direct contact with the electrolyte solution was 2000 µm in width and 80 µm in length. Patterns of the source and drain electrodes were formed using the DNR L300. Au/Ti was evaporated using a thermal evaporator in a vacuum chamber to make ohmic contact on the source and drain electrodes. An RE with a radius of 0.75 mm was formed by Au/Ti at a location 3 mm away from the gate channel of the ITO-ISFET, and the area of RE in contact with the electrolyte was 1.77 mm2. A circular RE electrode was positioned between two ITO-ISFETs at one sensor tip. If the ISM applied to each gate channel is different, two types of ions can be detected simultaneously. It was reported that the single-gate and the double-gate transistors reveal the sub-threshold swings quite close to the theoretical limit at room temperature [18,19]. This means that for a thin channel, there is no necessity to resort to a double-gate structure, which is very inconvenient for fabrication using the semiconductor process.
The transfer characteristics of the ITO-ISFET with Ag/AgCl and F-rGO REs were evaluated in Carmody buffer solution (CBS) with a fixed pH value (pH 6). Figure 3a shows that the ITO-ISFET with Ag/AgCl RE demonstrated traditional characteristics [20]. The IDS increased depending on the VGS and VDS, as shown in Figure 3b. The transconductance (gm) of the ITO-ISFET with Ag/AgCl RE was 1.79 mS (VDS = 0.2 V), as shown in Figure 3c.
A schematic diagram of the ITO-ISFET with the F-rGO RE is shown in Figure 3d. As for the sensor structure, a three-dimensional structure (using the Ag/AgCl RE) changed to a two-dimensional structure when the F-rGO RE was used. The IDS-VDS and IDS-VGS of the ITO-ISFET with the F-rGO RE were the same as those of the general ITO-ISFET with Ag/AgCl, as shown in Figure 3e,f. The gm value of the ITO-ISFET with F-rGO RE was 1.27 mS (VDS = 0.2 V), which was comparable to the gm value when the Ag/AgCl RE was used, as shown in Figure 3f.
When VGS was applied using the F-rGO RE, the ions in the electrolyte moved and formed electrical double layers (<5 nm) on both sides between the gate channel surface of the ITO-ISFET and F-rGO RE. The current flowing through the electrolyte from the ITO-ISFET to the F-rGO RE was negligible (51.2 nA). The electrical double layers had no charge transfer and acted as thin capacitors on both sides. The IDS of the ITO-ISFET is expressed by
I DS = ( μ n C G ) ( W L ) [ 2 ( V GS V th ) V DS V DS 2 ]
where μn is the electron mobility of ITO, CG is the gate capacitance, L is the channel length, and W is the channel width of the ITO-ISFET. The electron mobility of the ITO is 30 cm2∙V−1∙s−1 [21,22]. The CG of the ITO-ISFET (VDS = 0.4 V) with Ag/AgCl RE was 1.16 × 10−5 F∙cm−2. The CG is the electric double-layer capacitor of the ITO gate channel (CITO-EDL) on the ITO-ISFET with Ag/AgCl RE because the gate channel of the ITO-ISFET was in direct contact with the electrolyte solution. This capacitor acts as a nanogap capacitor at the electrode–electrolyte interface and has high capacitance values (1–100 µF∙cm−2) to create a strong electric field at the interface [23]. A high gate capacitance has been pursued to develop high-output currents and to develop biochemical sensors with high sensitivity [24,25]. The CG of the ITO-ISFET with the F-rGO RE was the series connection of CITO-EDL and the capacitor of F-rGO RE. The capacitor of the F-rGO RE is the electric double-layer capacitor (CF-rGO-EDL) because the F-rGO electrode was in direct contact with the electrolyte solution. The CG of the ITO-ISFET (VDS = 0.4 V) with the F-rGO RE was 1.07 × 10−5 and the CF-rGO-EDL was 1.42 × 10−4 F∙cm−2, respectively. The CF-rGO-EDL was high because the F-rGO RE has a large area of overlapping fluorinated graphene nano-electrodes. The value of CG follows the capacitor with small capacitance because the two capacitors connect in series. F-rGO RE has the potential to be used as an RE for ISFETs made of various materials, because the CF-rGO-EDL is larger than the CEDL of other gate channels in the electrolyte.

3.3. Ion Detection Using ITO-ISFET and F-rGO RE

We fabricated a pH sensor that can detect proton concentrations using a two-dimensional ITO-ISFET sensor based on the F-rGO RE. Figure 4a shows the IDS-VDS curve of the ITO-ISFET with the F-rGO RE when the pH value of the buffer solution changes from 2 to 12. The VGS was fixed at 0.25 V, and the VDS was swept from 0.0 V to 0.5 V in each pH buffer solution. The protonation of the gate channel surface of the ITO-ISFET decreases the channel resistance, because ITO is an n-type semiconductor. As a result, IDS increases depending on the proton concentration of the electrolyte. Figure 4b shows the IDS-VGS curve of the ITO-ISFET with the F-rGO RE in pH buffer solutions at a fixed VDS (0.5 V). As the pH increased, the VGS of the ITO-ISFET with the F-rGO RE shifted in direction to the right. The ΔVGS was 67.1 mV/pH, which was over the ideal Nernst equation. Similar results were obtained from the ITO-ISFET with the Ag/AgCl RE (ΔVGS was 66.5 mV/pH) (Figure S3 in Supplementary Materials). We evaluated the long-term stability of the ITO-ISFET with F-rGO and Ag/AgCl REs in a buffer solution (pH 6) in real time, which was similar to the drift characteristic of the ISFET [18]. The VGS was continuously measured to keep the IDS at 500 µA with the VDS fixed at 0.5 V on the IDS–VGS characteristic, and was continuously maintained at the voltage of −229.0 + 3.2 mV for 360 min, as shown in Figure 4c. The shift voltage of the VGS (0.53 mV/h) was comparable to that of the Ag/AgCl RE (2.01 mV/h) used for 6 h. The hysteresis characteristics of the ITO-ISFET using the F-rGO RE were evaluated. The VGS was continuously measured at fixed values of IDS (500 µA) and VDS (0.5 V) on the IDS–VGS characteristics with a periodic injection of CBS at different pH values every 5 min for 25 min, as shown in Figure 4d. The VDS was maintained constant so as to bias the device to work reliably. In the n-channel region, the VGS increased at the high pH to maintain the IDS at a fixed value of VDS on the ITO-ISFET because the surface charge on the gate channel was negative owing to deprotonation at the high pH. In contrast, the surface charge was positive owing to protonation at the low pH, and the VGS decreased to maintain IDS at a fixed value of VDS. The hysteresis voltage of the ITO-ISFET using the F-rGO RE was ±0.05 mV at pH 7. The long-term stability and hysteresis voltage of the ITO-ISFET using the F-rGO RE have smaller values than those of the ITO-ISFET using an Ag/AgCl RE [19,26]. The F-rGO RE operates stably in the electrolyte solution and is useful for detecting chemical and biological molecules with high stability.
We fabricated ion sensors by pasting ISM to the gate channel surface of the ITO-ISFET (ITO-ISFET-ISM) and evaluated the ion sensitivity using Ag/AgCl and F-rGO REs. The stability of the F-rGO was evaluated according to the K+ and Na+ ion solutions in order to confirm the change of potential according to the ionic strength. The F-rGO had stable potential in the ion solution for each concentration (Figure S6). The sensitivity of the ITO-ISFET-ISM using Ag/AgCl and F-rGO REs in Na+ and K+ ions dissolved in Tris-HCl buffer solution is shown in Figure 5a. The sensor with sodium ISM exhibited a sensitivity of 62.1 mV/decade and linearity in the Na+ ion concentration range of 10−5 to 1 M, although the interfering ion (100 mM KCl) was dissolved in the buffer solution. The ΔVGS of the ITO-ISFET with a potassium ISM using the F-rGO RE depended on the K+ ion concentration in the Tris-HCl buffer solution. The sensitivity of the K+ ion was 57.6 mV/decade, although the interfering ion (100 mM NaCl) was dissolved in the buffer solution. Among various ions contained in the electrolyte, only target ions were detected by the ITO-ISFET-ISM in the Tris-HCl buffer solution. Target ions were captured on the ITO gate channel surface by the ISM. As the concentration of Na+ and K+ ions increased, the positive charge on the ITO gate channel surface increased; hence, the VGS of the ITO-ISFET-ISM shifted in the negative direction because ITO-ISFET-ISM is an n-type gate channel. We conducted real-time detection of K+ ions in Tris-HCl buffer. The VGS was continuously measured with a fixed VDS (0.5 V) and IDS (500 μA) while adjusting the K+ ion concentration by periodically injecting a high-concentration KCl solution. The ITO-ISFET-ISM using the F-rGO RE exhibited an immediate and linear response in real time to changes in the K+ ion concentration from 10−5 to 1 M, as shown in Figure 5b. The ITO-ISFET-ISM using the F-rGO RE has sufficient sensitivity and stability even compared to other ion detection sensors using Ag/AgCl RE, as shown in Table 1. In another study, where the voltage was set as sensitivity using ISFET, Garcia et al. developed a sodium and potassium sensor using a miniaturized Ag/AgCl reference electrode [6]. Fakih et al. and Li et al. confirmed high performance by fabricating the ISFET for potassium detection using graphene [7,8]. Bao et al. developed a 3D-printed hybrid ISFET for sensing ammonium potassium calcium ions [9]. Hu et al. confirmed high sensitivity to Na ions and MB ions through ISM doping of the developed ISFET [10]. In addition, various types of ion sensors have been developed using Ag/AgCl reference electrodes.
In the human body, it is important to monitor concentrations of cations (Na+, K+, and H+ ions), which are found in the blood and extracellular fluid [29] and play an important role in maintaining electrolyte balance, fluid balance, and pH balance in the body [30]. The urine cation test may be necessary after abnormal results have been obtained from a blood cation test, which can identify acute kidney failure [31]. The 24-hour urine cation test can help distinguish between the two common causes of prerenal acute kidney injury (dehydration), and acute tubular necrosis [32]. However, testing using the 24-hour urine analysis is cumbersome and inconvenient. Hence, a robust, easy, and accurate test for monitoring cations is important. We evaluated the ITO-ISFET-ISM using the F-rGO RE on urine samples of patients provided by Keimyung University Dongsan Hospital (IRB No. 2021-08-117). The Na+, K+, and Cl- ion concentrations in the urine samples are summarized in Table 2. The concentration of each ion in the urine sample was characterized using ion-selective electrode (ISE)-based analytical equipment (ADVIA 2400, Siemens Healthcare, We removed the background color.USA) at Keimyung University Dongsan Hospital.
The sensitivity was simultaneously evaluated by variations in the concentrations of Na+ and K+ ions in the same urine sample. Figure S4a denotes the IDS-VGS of the ITO-ISEFT with sodium ISM using the F-rGO RE in the urine. The VDS was fixed at 0.5 V when VGS was swept from −1.0 to 0.5 V and each concentration was measured five times. The Na+ ion concentration in the urine increased from 76 mM to 81, 86, 96, 106, 116, 136, 156, and 176 mM.
The high concentration of Na+ ions in the urine sample caused a decrease in the VGS of the ITO-ISEFT with the sodium ISM and shifted the value to the left. Conversely, the IDS-VGS of the ITO-ISEFT with the potassium ISM showed irregular shifts with the Na+ ion concentration in the urine, as shown in Figure S4b. The sodium sensor in the urine revealed a sensitivity of 69.4 mV/dec, as shown in Figure 6a. Figure S5a shows the IDS-VGS of the ITO-ISEFT with the potassium ISM, using the F-rGO RE in the urine at a fixed VDS (0.5 V). The K+ ion concentration in the urine increased from 56.8 mM to 61.8, 66.8, 76.8, 86.8, 96.8, 116.8, 136.8, and 156.8 mM; the VGS shifted to the left only for the potassium ISM sensor with the K+ ion increment. On the other hand, the IDS-VGS of the ITO-ISEFT with the sodium ISM showed irregular shifts with the K+ ion concentration, as shown in Figure S5b. The sensitivity of the K+ ion sensor was 59.6 mV/dec, as shown in Figure 6b. ISM captured the target ions based on ion size. In the case of the K+ ion sensor, since the Na+ ion is smaller than the K+ ion, it was captured by the potassium ISM on the gate channel surface, resulting in a large interference effect. As in the previous Tris buffer solution as shown in Figure 5b, the detection sensitivity for the K+ ions was smaller than that of the Na+ ion sensor.

4. Conclusions

The rGO was fluorinated using FTP, and the F-rGO was shown to be usable as an RE by sustaining a stable potential despite the change in the solution. The F-rGO RE was integrated into the ITO-ISFET using a semiconductor device fabrication process. These sensors showed the voltage shifts following the Nernstian equation due to changes in the concentrations of cations (K+ Na+, and H+) in the electrolyte solution and showed high linearity. The ISFET with the F-rGO RE has potential as a platform to accurately and quickly diagnose other ions present in human urine.
The imbalance of ions must be carefully managed and accurately diagnosed as some deficiencies (or excesses) may lead to imbalances of other ions. Hospitals demand smart diagnostic equipment to test the pH and anion gap in blood or urine. The ISFET-based ion sensor integrated with the F-rGO RE is one of the best ways to satisfy these requirements. In the present study, although we limited the use of the F-rGO RE to ITO-ISFET sensors, it can also be integrated into silicon, carbon, and other semiconductor-based ISFETs. Therefore, if F-rGO is used as the RE for the ISFETs of various materials, the three-dimensional ISFET type of ion sensor could be converted into a two-dimensional structure and its integration can be broadened, making it possible to develop a low-cost, disposable ion diagnostic sensor.

Supplementary Materials

The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/bios13010089/s1, Figure S1: FE-SEM image of the ITO thin film on Si/SiO2 substrate; Figure S2: (a) Schematic diagram of the experiment. pH sensitivity of (b) the pristine graphene ion-sensitive field-effect transistor (PG-ISFET) and (c) fluorinated graphene ion-sensitive field-effect transistor (FG-ISFET) with Ag/AgCl RE in the pH buffer solution; Figure S3: IDS-VGS and IDS-VDS transfer characteristics of the ITO-ISFET with Ag/AgCl RE in the pH buffer solution; Figure S4: Detection of Na+ and K+ ions using ITO-ISFET-ISM with F-rGO RE sensors in patient urine samples with varying Na+ ion concentrations. The sensitivities of Na+ ion sensor (a) and K+ ion sensor (b); Figure S5: Detection of Na+ and K+ ions using ITO-ISFET-ISM with F-rGO RE sensors in patient urine samples with varying K+ ion concentrations. The sensitivities of K+ ion sensor (a) and Na+ ion sensor (b); Figure S6. Stability comparison of Ag/AgCl reference electrode and F-rGO reference electrode in solution A and solution B.

Author Contributions

Conceptualization, H.S.C., D.H.K. and K.S.S.; methodology, H.S.C. and H.G.O.; software, H.G.O.; validation, B.K.J. and K.S.S.; formal analysis, D.H.K. and K.S.S.; investigation, J.H.K., D.A.J. and B.K.J.; resources, D.A.J.; writing—original draft preparation, H.S.C.; writing—review and editing, D.H.K. and K.S.S.; supervision, K.S.S.; project administration, J.H.K.; funding acquisition, K.S.S. All authors have read and agreed to the published version of the manuscript.

Funding

This work was supported by the National Research Foundation of Korea (NRF) grant funded by the Korean government (MSIT) (No. 2020R1F1A1067581) and by the Ministry of Small and Medium-Sized Enterprises (SMEs) and Startups (MSS), Korea, under the “Regional Specialized Industry Development Plus Program (R&D, S3092048)” supervised by the Korea Institute for Advancement of Technology (KIAT).

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Morozov, S.V.; Novoselov, K.S.; Katsnelson, M.I.; Schedin, F.; Elias, D.C.; Jaszczak, J.A.; Geim, A.K. Giant Intrinsic Carrier Mobilities in Graphene and Its Bilayer. Phys. Rev. Lett. 2008, 100, 16602. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  2. Gómez-Navarro, C.; Weitz, R.T.; Bittner, A.M.; Scolari, M.; Mews, A.; Burghard, M.; Kern, K. Electronic Transport Properties of Individual Chemically Reduced Graphene Oxide Sheets. Nano Lett. 2007, 7, 3499–3503. [Google Scholar] [CrossRef] [PubMed]
  3. Feng, H.; Cheng, R.; Zhao, X.; Duan, X.; Li, J. A low-temperature method to produce highly reduced graphene oxide. Nat. Commun. 2013, 4, 1539. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  4. Zhang, X.; Zhang, Y.; Liao, Q.; Song, Y.; Ma, S. Reduced Graphene Oxide-Functionalized High Electron Mobility Transistors for Novel Recognition Pattern Label-Free DNA Sensors. Small 2013, 9, 4045–4050. [Google Scholar] [CrossRef]
  5. Ahn, M.-S.; Ahmad, R.; Bhat, K.S.; Yoo, J.-Y.; Mahmoudi, T.; Hahn, Y.-B. Fabrication of a solution-gated transistor based on valinomycin modified iron oxide nanoparticles decorated zinc oxide nanorods for potassium detection. J. Colloid Interface Sci. 2018, 518, 277–283. [Google Scholar] [CrossRef]
  6. Garcia-Cordero, E.; Bellando, F.; Zhang, J.; Wildhaber, F.; Longo, J.; Guérin, H.; Ionescu, A.M. Three-Dimensional Integrated Ultra-Low-Volume Passive Microfluidics with Ion-Sensitive Field-Effect Transistors for Multiparameter Wearable Sweat Analyzers. ACS Nano 2018, 12, 12646–12656. [Google Scholar] [CrossRef]
  7. Fakih, I.; Centeno, A.; Zurutuza, A.; Ghaddab, B.; Siaj, M.; Szkopek, T. High resolution potassium sensing with large-area graphene field-effect transistors. Sensors Actuators B Chem. 2019, 291, 89–95. [Google Scholar] [CrossRef]
  8. Li, H.; Walsh, K.B.; Bayram, F.; Koley, G. Direct measurement of K+ ion efflux from neuronal cells using a graphene-based ion sensitive field effect transistor. RSC Adv. 2020, 10, 37728–37734. [Google Scholar] [CrossRef]
  9. Bao, C.; Kaur, M.; Kim, W.S. Toward a highly selective artificial saliva sensor using printed hybrid field effect transistors. Sensors Actuators B: Chem. 2019, 285, 186–192. [Google Scholar] [CrossRef]
  10. Hu, Q.; Chen, S.; Wang, Z.; Zhang, Z. Improving selectivity of ion-sensitive membrane by polyethylene glycol doping. Sensors Actuators B Chem. 2020, 328, 128955. [Google Scholar] [CrossRef]
  11. Yoon, J.H.; Park, H.J.; Park, S.H.; Lee, K.G.; Choi, B.G. Electrochemical characterization of reduced graphene oxide as an ion-to-electron transducer and application of screen-printed all-solid-state potassium ion sensors. Carbon Lett. 2019, 30, 73–80. [Google Scholar] [CrossRef]
  12. Pirovano, P.; Dorrian, M.; Shinde, A.; Donohoe, A.; Brady, A.J.; Moyna, N.M.; Wallace, G.; Diamond, D.; McCaul, M. A wearable sensor for the detection of sodium and potassium in human sweat during exercise. Talanta 2020, 219, 121145. [Google Scholar] [CrossRef]
  13. Cao, N.; Lyu, Q.; Li, J.; Wang, Y.; Yang, B.; Szunerits, S.; Boukherroub, R. Facile synthesis of fluorinated polydopa-mine/chitosan/reduced graphene oxide composite aerogel for efficient oil/water separation. Chem. Eng. J. 2017, 326, 17–28. [Google Scholar] [CrossRef]
  14. Zhan, D.; Ni, Z.; Chen, W.; Sun, L.; Luo, Z.; Lai, L.; Yu, T.; Wee, A.T.S.; Shen, Z. Electronic structure of graphite oxide and thermally reduced graphite oxide. Carbon 2011, 49, 1362–1366. [Google Scholar] [CrossRef]
  15. Park, Y.-C.; Kim, Y.-S.; Seo, H.-K.; Ansari, S.; Shin, H.-S. ITO thin films deposited at different oxygen flow rates on Si(100) using the PEMOCVD method. Surf. Coatings Technol. 2002, 161, 62–69. [Google Scholar] [CrossRef]
  16. Kim, J.; Bae, J.; Kim, H.; Lee, N.-E.; Yeom, G.; Oh, K. Effects of oxygen radical on the properties of indium tin oxide thin films deposited at room temperature by oxygen ion beam assisted evaporation. Thin Solid Films 2000, 377, 103–108. [Google Scholar] [CrossRef]
  17. Gianti, M.S.; Oh, H.G.; Cho, H.S.; Jo, D.A.; Indriatmoko, M.N.; Lim, J.M.; Jang, B.K.; Song, K.S. Low-cost fabrication of indium tin oxide (ITO) FETs for sodium detection in electrolytes and human urine. Electron. Lett. 2021, 57, 682–684. [Google Scholar] [CrossRef]
  18. Oh, H.; Jeon, D.; Gianti, M.; Cho, H.; Jo, D.; Indriatmoko, M.; Jang, B.; Lim, J.; Cho, S.; Song, K. Two–Dimensional Disposable Graphene Sensor to Detect Na+ Ions. Nanomaterials 2021, 11, 787. [Google Scholar] [CrossRef]
  19. Lue, C.-E.; Wang, I.-S.; Huang, C.-H.; Shiao, Y.-T.; Wang, H.-C.; Yang, C.-M.; Hsu, S.-H.; Chang, C.-Y.; Wang, W.; Lai, C.-S. pH sensing reliability of flexible ITO/PET electrodes on EGFETs prepared by a roll-to-roll process. Microelectron. Reliab. 2011, 52, 1651–1654. [Google Scholar] [CrossRef]
  20. Bergveld, P. Thirty years of ISFETOLOGY: What happened in the past 30 years and what may happen in the next 30 years. Sens. Actuators B Chem. 2003, 88, 1–20. [Google Scholar] [CrossRef]
  21. Hamberg, I.; Granqvist, C.G. Evaporated sn-doped in2o3 films: Basic optical properties and applications to energy-efficient windows. J. Appl. Phys. 1986, 60, R123–R160. [Google Scholar] [CrossRef]
  22. Shao, Y.; Xiao, X.; Wang, L.; Liu, Y.; Zhang, S. Anodized ito thin-film transistors. Adv. Funct. Mater. 2014, 24, 4170–4175. [Google Scholar] [CrossRef]
  23. Ono, S.; Miwa, K.; Seki, S.; Takeya, J. A comparative study of organic single-crystal transistors gated with various ionic-liquid electrolytes. Appl. Phys. Lett. 2009, 94, 63301. [Google Scholar] [CrossRef]
  24. Liu, N.; Chen, R.; Wan, Q. Recent Advances in Electric-Double-Layer Transistors for Bio-Chemical Sensing Applications. Sensors 2019, 19, 3425. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  25. Han, S.; Yamamoto, S.; Polyravas, A.G.; Malliaras, G.G. Microfabricated Ion-Selective Transistors with Fast and Super-Nernstian Response. Adv. Mater. 2020, 32, 2004790. [Google Scholar] [CrossRef]
  26. Chou, J.C.; Tsai, H.M.; Shiao, C.N.; Lin, J.S. Study and simulation of the drift behaviour of hydrogenated amorphous silicon gate ph-isfet. Sens. Actuators B Chem. 2000, 62, 97–101. [Google Scholar] [CrossRef]
  27. Lim, H.-R.; Kim, Y.-S.; Kwon, S.; Mahmood, M.; Kwon, Y.-T.; Lee, Y.; Lee, S.M.; Yeo, W.-H. Wireless, Flexible, Ion-Selective Electrode System for Selective and Repeatable Detection of Sodium. Sensors 2020, 20, 3297. [Google Scholar] [CrossRef]
  28. Lim, H.-R.; Lee, S.; Mahmood, M.; Kwon, S.; Kim, Y.-S.; Lee, Y.; Yeo, W.-H. Development of Flexible Ion-Selective Electrodes for Saliva Sodium Detection. Sensors 2021, 21, 1642. [Google Scholar] [CrossRef]
  29. Kraut, J.A.; Madias, N.E. Metabolic acidosis: Pathophysiology, diagnosis and management. Nat. Rev. Nephrol. 2010, 6, 274–285. [Google Scholar] [CrossRef]
  30. Fried, L.F.; Palevsky, P.M. Hyponatremia and hypernatremia. Med. Clin. North Am. 1997, 81, 585–609. [Google Scholar] [CrossRef]
  31. Zou, J.; Zhang, K.; Cai, W.; Chen, T.; Nathan, A.; Zhang, Q. Optical-reconfigurable carbon nanotube and indium-tin-oxide complementary thin-film transistor logic gates. Nanoscale 2018, 10, 13122–13129. [Google Scholar] [CrossRef]
  32. Kovesdy, C.P.; Lott, E.H.; Lu, J.L.; Malakauskas, S.M.; Ma, J.Z.; Molnar, M.Z.; Kalantar-Zadeh, K. Hyponatremia, Hypernatremia, and Mortality in Patients With Chronic Kidney Disease With and Without Congestive Heart Failure. Circulation 2012, 125, 677–684. [Google Scholar] [CrossRef]
Figure 1. (a) Schematic diagram of fluorinated F−rGO and FE-SEM image of the F−rGO on the RE surface. (b) FT−IR and (c) Raman spectra of the rGO and F−rGO.
Figure 1. (a) Schematic diagram of fluorinated F−rGO and FE-SEM image of the F−rGO on the RE surface. (b) FT−IR and (c) Raman spectra of the rGO and F−rGO.
Biosensors 13 00089 g001
Figure 2. Integrated two-channel ITO-ISFET with F-rGO RE on the sensor tip.
Figure 2. Integrated two-channel ITO-ISFET with F-rGO RE on the sensor tip.
Biosensors 13 00089 g002
Figure 3. (a) Schematic diagram of the Ag/AgCl RE. (b) IDS−VDS and (c) IDS−VGS transfer characteristics of ITO−ISFET with the Ag/AgCl RE. (d) Schematic diagram of F−rGO RE. (e) IDS−VDS and (f) IDS−VGS transfer characteristics of the ITO-ISFET with the F−rGO RE.
Figure 3. (a) Schematic diagram of the Ag/AgCl RE. (b) IDS−VDS and (c) IDS−VGS transfer characteristics of ITO−ISFET with the Ag/AgCl RE. (d) Schematic diagram of F−rGO RE. (e) IDS−VDS and (f) IDS−VGS transfer characteristics of the ITO-ISFET with the F−rGO RE.
Biosensors 13 00089 g003
Figure 4. (a) IDS−VDS, (b) IDS−VGS, (c) long-term stability, and (d) hysteresis characteristics of ITO−ISFET with F−rGO RE in a pH buffer solution.
Figure 4. (a) IDS−VDS, (b) IDS−VGS, (c) long-term stability, and (d) hysteresis characteristics of ITO−ISFET with F−rGO RE in a pH buffer solution.
Biosensors 13 00089 g004
Figure 5. (a) Na+ and K+ ion sensitivities of the ITO−ISFET−ISM with the Ag/AgCl and F−rGO REs. (b) Real−time detection of K+ ions using the ITO−ISFET−ISM with the F−rGO RE.
Figure 5. (a) Na+ and K+ ion sensitivities of the ITO−ISFET−ISM with the Ag/AgCl and F−rGO REs. (b) Real−time detection of K+ ions using the ITO−ISFET−ISM with the F−rGO RE.
Biosensors 13 00089 g005
Figure 6. Detection of (a) Na+, (b) K+ ions, and the sensitivities of ITO−ISFET−ISM with F−rGO RE sensors in patient urine samples.
Figure 6. Detection of (a) Na+, (b) K+ ions, and the sensitivities of ITO−ISFET−ISM with F−rGO RE sensors in patient urine samples.
Biosensors 13 00089 g006
Table 1. Other ion sensors using Ag/AgCl RE.
Table 1. Other ion sensors using Ag/AgCl RE.
Reference
Electrode
Detection IonSensitivityDetection LimitSensor TypeRef.
F-rGONa+
K+
62.1 mV/dec
57.6 mV/dec
10 μMISFETThis work
Ag/AgClK+4.65 uA/uM0.04 μMISFET[5]
Ag/AgClNa+
K+
62 mV/dec
55 mV/dec
5 mMISFET[6]
Ag/AgClK+37 mV/dec10 nMISFET[7]
Ag/AgClK+67 mV/dec10 μMISFET[8]
Ag/AgClNH4+
K+
Ca2+
98 mV/dec
104 mV/dec
42 mV/dec
10 μMISFET[9]
Ag/AgClNa+60 mV/dec60 μMISFET[10]
Ag/AgClK+53.34 mV/dec0.06 mMrGO[11]
Ag/AgClNa+K+56.4 mV/dec
54.3 mV/dec
100 μM100 μMISE[12]
Ag/AgClNa+56.1 mV/dec4 μMISE[27]
Ag/AgClNa+58.9 mV/dec42.7 μMISE[28]
Table 2. Information regarding patient urine sample.
Table 2. Information regarding patient urine sample.
Urine
Na+ ion (mM)K+ ion (mM)Cl ion (mM)
7656.849
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Kim, D.H.; Cho, H.S.; Kim, J.H.; Jo, D.A.; Oh, H.G.; Jang, B.K.; Song, K.S. The Integration of Reference Electrode for ISFET Ion Sensors Using Fluorothiophenol-Treated rGO. Biosensors 2023, 13, 89. https://doi.org/10.3390/bios13010089

AMA Style

Kim DH, Cho HS, Kim JH, Jo DA, Oh HG, Jang BK, Song KS. The Integration of Reference Electrode for ISFET Ion Sensors Using Fluorothiophenol-Treated rGO. Biosensors. 2023; 13(1):89. https://doi.org/10.3390/bios13010089

Chicago/Turabian Style

Kim, Dae Hoon, Hae Shin Cho, Jin Heung Kim, Da Ae Jo, Hong Gi Oh, Byoung Kuk Jang, and Kwang Soup Song. 2023. "The Integration of Reference Electrode for ISFET Ion Sensors Using Fluorothiophenol-Treated rGO" Biosensors 13, no. 1: 89. https://doi.org/10.3390/bios13010089

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop