Towards a Point-of-Care (POC) Diagnostic Platform for the Multiplex Electrochemiluminescent (ECL) Sensing of Mild Traumatic Brain Injury (mTBI) Biomarkers

Diagnostic Systems Research Group, Institute of Life Technologies, School of Engineering, University of Applied Sciences and Arts Western Switzerland (HES-SO Valais-Wallis), 1950 Sion, Switzerland
*
Author to whom correspondence should be addressed.
Biosensors 2022, 12(3), 172; https://doi.org/10.3390/bios12030172
Submission received: 14 February 2022 / Revised: 2 March 2022 / Accepted: 7 March 2022 / Published: 11 March 2022
(This article belongs to the Section Biosensors and Healthcare)

Abstract

:
Globally, 70 million people are annually affected by TBI. A significant proportion of all TBI cases are actually mild TBI (concussion, 70–85%), which is considerably more difficult to diagnose due to the absence of apparent symptoms. Current clinical practice of diagnosing mTBI largely resides on the patients’ history, clinical aspects, and CT and MRI neuroimaging observations. The latter methods are costly, time-consuming, and not amenable for decentralized or accident site measurements. As an alternative (and/or complementary), mTBI diagnostics can be performed by detection of mTBI biomarkers from patients’ blood. Herein, we proposed two strategies for the detection of three mTBI-relevant biomarkers (GFAP, h-FABP, and S100β), in standard solutions and in human serum samples by using an electrochemiluminescence (ECL) immunoassay on (i) a commercial ECL platform in 96-well plate format, and (ii) a “POC-friendly” platform with disposable screen-printed carbon electrodes (SPCE) and a portable ECL reader. We further demonstrated a proof-of-concept for integrating three individually developed mTBI assays (“singleplex”) into a three-plex (“multiplex”) assay on a single SPCE using a spatially resolved ECL approach. The presented methodology demonstrates feasibility and a first step towards the development of a rapid POC multiplex diagnostic system for the detection of a mTBI biomarker panel on a single SPCE.

Graphical Abstract

1. Introduction

Traumatic brain injuries (TBI) are physical injuries that can lead to brain function alterations of temporary or permanent nature [1]. In the absence of routine diagnostic screening tests for TBI, the number of truly affected patients is difficult to assess and is probably significantly under-reported [2]. According to the CDC statistics, 1.5 million people suffer from TBI each year, while the most frequent causes of TBI are motor vehicle crashes, falls, and violence [3]. The Glasgow Coma Scale (GCS) developed in the seventies of the 20th century, is an assessment tool used to determine the consciousness level of the patient after a TBI, based on the ability to open eyes, be oriented, respond to questions, and obey motor commands [4]. Based on the extent of injury, TBIs can be classified as mild (GCS 13–15), moderate (GCS 9–12), and severe (GCS 3–8). A vast majority of TBI cases can be attributed to concussion, i.e., mild TBI [5]. Post-injury, there is often an absence of symptoms or primarily the presence of non-specific symptoms (e.g., headaches, fatigue, depression, visual and/or sleep disturbances, seizures, etc.) [6].
Diagnosis of mTBI can be rather challenging. The GCS scale is dependent on the assessment skills of the observer, and it can be inaccurate in distinguishing between mild and moderate TBIs [7]. Alternatives are neuroimaging tools, such as MRI and CT scans, which can be quite costly, expose patients to harmful radiation, and are not available on-site (accident sites, sports fields, remote and underdeveloped areas, etc.). Also, approximately 90% of all mTBI cases will not have any evidence of structural abnormalities visible on a CT scan [8]. Therefore, there is a growing need to develop additional diagnostic screening tools for aiding diagnosis and enabling an accurate, inexpensive and fast triage of patients with mTBI [8,9,10,11,12].
Biomarkers may have different intended uses, such as diagnostic, prognostic, predictive, pharmacodynamic, or efficacy response relevance. As for mTBI, the preferred ones are brain protein biomarkers that can pass the blood—brain barrier (BBB) into circulation. They can be found in human serum and/or plasma, and the analytical tools should be able to accurately detect them in the lower picogram range (pg mL−1) [13]. The Scandinavian guideline proposes the use of the biomarker S100 calcium-binding protein β (S100β) for mTBI patients who are admitted to the hospital within 6 h after the injury [14]. However, S100β can be increased in multiple extra-cerebral tissues, for example, after extracranial injuries [15], peripheral lesions [16], and physical exercise [17].
Recently, several publications indicated other mTBI biomarkers that could be more suitable. For example, glial fibrillary acidic protein (GFAP) is a protein found in the cytoskeleton of glial cells [18] and has been used for the detection of acute intracranial injuries following a TBI [19]. The sensitivity of GFAP for the detection of intracranial lesions on CT scans has been reported to be 67% to 100%, with specificity between 0% and 100% [20]. GFAP could be useful for differentiating focal and diffuse brain injury; however, its ability to enter into the bloodstream depends on the BBB damage [21]. Furthermore, heart fatty-acid binding protein (h-FABP) is a cytosolic trafficking protein that can predict intracranial pathologies linked with brain injuries [22,23].
TBI biomarkers have the potential to significantly advance the way TBI is assessed and treated today, beyond the evaluation of the need for a head CT scan [24]. However, given the complexity of the TBI, it is unlikely that any single biomarker would be sufficiently sensitive and specific for use as a clinical diagnostic test. Posti et al. and Lagerstedt et al. reported on panels of protein biomarkers that perform better in discriminating CT positive from CT negative patients with mTBI than individual biomarkers [25,26]. As early as the beginning of 2018, a central laboratory test developed by Banyan Biomarkers was approved by the FDA [27]. The test is based on a chemiluminescent ELISA format for determining two TBI biomarkers, GFAP and ubiquitin carboxyl-terminal hydrolase isozyme L1 (UCH-L1), with a sensitivity of 97.5% for CT positive cases and a negative predictive value of 99.6% [27]. Practically the latter means that in more than 33% of the cases, the patients being suspected of brain injury can be ruled out prior to a CT scan [27]. Just recently, in January 2021, Abbott launched the POC diagnostic device i-STATTM AlinityTM that measures amperometrically UCH-L1 and GFAP [28], while the company NanoDx™ is in the progress of developing an ultrasensitive nanowire technology to resistively measure the S100β and GFAP biomarkers [29].
Despite the recent developments in the mTBI diagnostic field, there is still a growing need to expand the biomarker panel with inflammatory and brain damage biomarkers to guarantee adequate diagnostic specificity and sensitivity and to allow decentralized mTBI diagnostics. To address these limitations, the conception of a minimally invasive multiplexed detection device would allow rapid and reliable detection of mTBI biomarkers at the POC level, which could revolutionize mTBI-patients management compared to conventional methods used. Many different electrochemical (EC) (bio)sensors approaches [30] and few ECL biosensor approaches have been proposed for detection of single, individual mTBI biomarkers (h-FABP [31,32]), but to the best of our knowledge, none using a mTBI biomarker panel. Electrochemiluminescence immunoassays (ECLIA) show great promise for the detection of low-level concentration compounds, with a wide linear range, high sensitivity, and low background noise [33]. In this work, we have developed a sandwich-type ECL assay for detecting and quantifying three mTBI-relevant biomarkers (GFAP, h-FABP and S100β, Table 1), in standard solutions and in human serum samples (HS), using a 96-well plate commercial “benchtop” platform from MesoScale Discovery (MSD) (Scheme 1b). These three individual “singleplex” assays were then translated into a miniaturized platform based on screen-printed carbon electrodes (SPCE) coupled with a portable ECL reader (µSTAT-ECL DropSens Metrohm, Scheme 1c). Furthermore, we showed a proof-of-concept for integration of three individually developed mTBI biomarker assays (“singleplex”) into a three-plex “multiplex” assay on a single SPCE using a spatially resolved ECL approach (Scheme 1d), demonstrating a first step towards the development of a POC diagnostic prototype instrument based on ECL detection of a mTBI biomarker panel (Scheme 1e).

2. Materials and Methods

2.1. Materials

GFAP assay: Antigen GFAP human recombinant (ref. 8G45, HyTest Ltd., Turku, Finland); monoclonal mouse anti-human glial fibrillary acidic protein (ref. 4G25, HyTest Ltd., Turku, Finland) clone 83cc and clone 81cc were employed as capture and detection antibody, respectively.
h-FABP assay: Antigen FABP human (ref. 8F65, HyTest Ltd., Turku, Finland); monoclonal mouse anti-human fatty acid-binding protein (ref. 4F29, HyTest Ltd., Turku, Finland) clone 22 and clone 28cc were employed as capture and detection antibody, respectively.
S100β assay: Antigen S100BB homodimer and S100A1B heterodimer human (ref. 8S9h, HyTest Ltd., Turku, Finland); monoclonal mouse anti-human S100 proteins (ref. 4S37, HyTest Ltd., Turku, Finland) clone 8B10cc and clone 6G1cc were employed as capture and detection antibody, respectively.
Serum samples: Human serum from human male AB plasma, US origin, sterile filtered (ref. H4522, Sigma Aldrich, St. Louis, MO, USA) was diluted 2× with appropriate assay diluents and used for the recovery studies.
Other reagents: All chemicals were used as received without further purification. All aqueous solutions were prepared with MQ water. Gold SULFO-Tag NHS-Ester lyophilized (ref. RA19AO), read buffer T 4X (ref. R92TC), conjugation buffer (ref. R60AJ), conjugation storage buffer (ref. R60AC) and QuickPlex 96-well plate (ref. L55XA) were all purchased from Meso Scale Discovery (MSD). Other materials included: Zeba Spin desalting columns 40K MWCO 0.5 mL (87766, ThermoFisher, Waltham, MA, USA), Millex-GV Filter 0.22 µm (SLGV004SL, Sigma-Aldrich, MO, USA), syringe 1 mL BD Luer-Lok tip (309628, BD, New York, NJ, USA), casein sodium salt from bovine milk (C8654-500G, Sigma-Aldrich, MO, USA), bovine serum albumin fraction V (ref. 10735078001, Roche Diagnostics, Rotkreuz, Switzerland), PBS 10× pH 7.4 phosphate saline buffer (ref. 7011-044, Gibco, Billings, MT, USA), Tween-20 (ref. P1379-100 mL, Sigma-Aldrich, MO, USA), CaCl2·2H2O (ref. 223506, Fluka, Buchs, Switzerland). Technical grade ethanol was used for cleaning SPCE electrodes.

2.2. Apparatus

QuickPlex SQ120 (Meso Scale Discovery MSD): The MSD platform is based on ECL detection of a SULFO-Tag labelled detection antibody that emits light upon electrochemical stimulation. QuickPlex SQ120 was employed as an ECL plate reader for QuickPlex 96-well plates. The detection process is initiated on carbon electrodes located at the bottom of the wells, and only labels in the electrode proximity can be detected. The ECL mechanism of the co-reactant system ruthenium tris(bipyridine)-tripropylamine (read buffer) has been previously described [40].
Disposable screen-printed electrodes (SPEs, Metrohm DropSens): The SPEs incorporate a three-electrode setup, printed on ceramic substrates (size 33.0 mm × 10.0 mm). Both working (WE; disk-shaped 4-mm diameter) and counter-electrodes are fabricated from carbon or carbon-based inks (for electrodes with modified WE), while pseudo-reference electrodes and electrical contact pads are fabricated from silver ink. An insulating layer is printed over the three-electrode system, leaving the electric contacts and a working area with an actual volume of 50 μL. SPEs with different types of working electrodes were tested as the solid-state support throughout this work: screen-printed carbon electrodes (SPCE, ref. DRP-110), screen-printed carbon electrodes modified with gold nanoparticles (SPCE-GNP, ref. DRP-110-GNP), screen-printed carbon electrodes modified with carbon nanotubes and gold nanoparticles (SPCE-CNT-GNP, ref. DRP-110CNT-GNP) and screen-printed carbon electrodes modified with quantum dots (SPCE-QD, ref. DRP-110QD).
Homemade incubation cell for SPEs: Customized incubation cell for SPEs was fabricated from Teflon at the HES-SO Valais-Wallis mechanical workshop (Supplementary Figure S6). The cell was designed using the SolidWorks CAD package and was intended to fit 12 SPEs, leaving only the WE area to be exposed to the reagents during various incubation/mixing/washing steps of the immunoassay protocol. When the immunoassay protocol was finished, the SPEs were taken out of the incubation cell, washed with wash buffer solution, and transferred into µSTAT-ECL cell for the read-out.
µSTAT Bipotentiostat with ECL Cell (Metrohm DropSens): The device is composed of a potentiostat/galvanostat (±4 V DC potential range, ±40 mA maximum measurable current) (size 127.8 mm × 124.1 mm × 34.1 mm) combined with a detector that is integrated in the ECL cell (size 75.0 mm × 88.4 mm × 40.0 mm). The detector is a low-noise photosensor composed of a silicon photodiode with a spectral response range of 340–1100 nm, and a maximum sensitivity of 0.62 V/nW at 960 nm [41]. The device uses the DropView 8400 software for signal acquisition and can be employed as a portable, battery-operated ECL reader.
Multiplex electrode spotting with nano-spotter device: A S3 contactless nano-spotting device from Scienion AG (Berlin, Germany) equipped with a Piezo Dispense Capillary (PDC-70, coating type 3, p/n: P-2030 S-6051) was used to dispense drops of 300 pL on pre-defined positions of Metrohm DropSens electrodes.
Multiplex assay measurement imager: For ECL read-out from multiplex assay developed on SPE electrodes, a Vilber Fusion FX6 EDGE imager (Vilber Smart Imaging) was employed in combination with a µSTAT Bipotentiostat. The imager is equipped with an eVo6 scientific grade CCD camera (6.3 MPx, −30 °C cooling, f/0.7) to acquire signals generated from SPE. ImageJ software (v 1.53) was used for image processing.

2.3. Methods

Detection antibody conjugation with Ru(bpy)32+-NHS ester: Detection antibody labeling with Gold SULFO-Tag NHS-Ester was performed following the protocol provided by MesoScale Discovery (MSD GOLD SULFO-TAG Conjugation Quick Guide). Challenge ratio of 50:1 was used for all detection antibodies, while the labeling incorporation ratio was calculated based on the OD455 measured for each labelled antibody using the NanoDrop OnceC Microvolume UV-VIS Spectrophotometer (Thermo-Fisher Scientific, Waltham, MA, USA). The calculated label ratio was 19:1, 14:1 and 21:1 for GFAP, h-FABP and S100β detection antibodies, respectively.
Singleplex assay on MSD platform: 30 µL of capture antibody (at desired concentration, diluted in the appropriate coating diluent) was added in the well of QuickPlex 96-well plate. After the incubation step (1.5 h, room temperature RT, 700 rpm) the plate was washed using a dedicated wash buffer (3 × 300 µL), and 100 µL of blocking agent was added (45 min, RT, 700 rpm). After the second washing step, 30 µL of antigen or blank was added in the well and incubated (1 h, RT, 700 rpm). After the third washing step, 30 µL of detection antibody was added at desired concentration (diluted in corresponding diluent) and incubated (1 h, RT, 700 rpm). After the final washing step, 150 µL of MSD read buffer 2× was added in each well and ECL signal was recorded using MSD QuickPlex SQ120 plate reader.
Singleplex assay on SPCEs: SPCEs were firstly washed using the mixture of ethanol/water (2:1) for 20 min, then rinsed with DI water, dried with nitrogen, and placed inside the “homemade” incubation cell. The immunoassay protocol was identical as described in the paragraph above for MSD platform, with the difference that all incubation steps were performed at 500 rpm. After the last incubation/washing step, the SPCEs were taken out from the incubation cell and placed inside the ECL cell of µSTAT-ECL instrument with 50 µL of MSD read buffer 2× to perform the read-out.
Multiplex assay on SPCEs: Electrodes were cleaned using the same protocol as described for the singleplex assay. The multiplex assay was performed using the assay conditions established for S100β biomarker (Table 2). The capture antibodies (50 µg mL−1) were deposited on the working electrode using the automated nano-spotter device (S3, Scienion, Berlin, Germany) by collocated spotting of 30 drops of 300 pL (±10 pL), to form spots of 250 µm (±50 µm) diameter. The source plate temperature was set at RT and the relative humidity in the spotting area at 60%. After deposition, SPCEs were let in the spotting area during 30 min before blocking with 2% BSA for 1 h at RT and washing with wash buffer. Incubation with antigen and detection antibodies was carried out in homemade incubation cells for SPEs. The read-out was performed with 150 µL of MSD read buffer 2× using Vilber Fusion FX6 EDGE imager (Vilber Smart Imaging) combined with µSTAT Bipotentiostat.

3. Results

3.1. Development of ECL “Singleplex” Assays for mTBI Biomarkers on MSD Platform

MSD QuickPlex SQ120 is a versatile and robust platform that can be very useful tool for the detection of different types of analytes in 96-well plate format, and it was employed for development of ECL sandwich immunoassays for each of the individual mTBI biomarkers (GFAP, h-FABP and S100β) (Scheme 1b).
A design of experiment (DoE) approach was used to determine the optimal settings and conditions for the major controllable factors in the assay. The following conditions were jointly assessed for development/optimization of each individual mTBI biomarker assay:
  • Coating diluent (PBS 1× pH 7.4 or TRIS 50 mM pH 8.6, with addition of 0.1–5 mM CaCl2 for S100β assay)
  • Blocking agent (0.1–2% of BSA or 0.1–2% casein in PBS 1× pH 7.4 or TRIS 50 mM pH 8.6, with/without addition of 0.1% Tween-20);
  • Wash buffer (PBS 1× pH 7.4 or TRIS 50 mM pH 8.6, with 0.05–0.4% Tween-20);
  • Detection antibody diluent (PBS 1× pH 7.4 or TRIS 50 mM pH 8.6, with/without addition of 0.1–1% of blocking agent and/or 0.06% Tween-20);
  • Capture antibody (cAb) and detection antibody concentration (dAb) (cAb concentration range: 1–25 μg mL−1; dAb concentration range: 0.5–10 μg mL−1).
Table 2 summarizes the pre-optimized assay conditions obtained for each individual mTBI biomarker. In overall, the optimized assay conditions for GFAP and h-FABP assay were very similar. The optimal blocking agent was 1% BSA and all diluents were based on PBS 1×. In the case of the S100β assay, all diluents were based on 50 mM TRIS buffer (pH 8.6) with the addition of CaCl2, due to the fact that the S100 protein is a dimeric member of the EF-hand calcium-binding protein superfamily, and its calcium-binding properties influencing the antibody recognition [42,43]. Several studies indicated that such interaction happens through a calcium-induced conformational change, which leads to the exposure of a hydrophobic protein region [42].
In some publications, the authors reported that the addition of Ca2+ in the antibody diluents had as a consequence, an improvement of the recognition activity [44,45], while others reported that calcium-chelators improved the antigen recognition (immunoassays with Sangtec antibodies) [46]. In the present study, we noticed that the addition of Ca2+ in the coating, assay and detection antibody diluent had a positive impact on the assay performance. This is also supported by the fact that the antibody provider recommends add it the ion of Ca2+ in the antibody diluents. Concentrations of capture and detection antibody were optimized for each biomarker using the mean matrix heatmap format. The results are presented in Figure 1 (left figures) as a heat map showing the signal intensities (GFAP assay) and S/B ratios (h-FABP and S100β assay) from low (red) to high (green). The capture and detection antibody concentrations were determined based on the maximum S/B results, apart from GFAP assay, where the antibody concentrations were chosen based on the highest signal intensities.
To evaluate the analytical performance of each singleplex biomarker assay, buffer solutions containing each individual biomarker concentration ranging from 10 pg mL−1 to 10 ng mL−1 were analyzed. Based on the obtained results, a calibration curve (Figure 1, right figures) was established for each biomarker using the pre-optimized conditions from Table 2. Data were analyzed by assuming that the ECL intensity was proportional to the biomarker concentration through a four-parameter dose-response regression function (4PL) model with 1/Y2 weighting (OriginPro software). To fit the data, the following equation was used (Equation (1)):
y = A 1 A 2 1 + ( x x 0 ) p + A 2
where x denotes the concentration of the biomarker, and A1, A2, x0, and p are the four parameters. The A1 and A2 parameters correspond to the upper and lower asymptotes for the function, respectively, while the p (Hill slope) and x0 parameters correspond to the slope.
As Figure 1 indicates, the developed singleplex assays had good dynamic ranges, background levels, and sensitivities. The LOD values were calculated by interpolating the curve using the average value of the blank plus three times the standard deviation of the blank. Obtained LOD values of 6.94 pg mL−1 (R2 = 0.9999), 1.35 pg mL−1 (R2 = 0.9999) and 15.73 pg mL−1 (R2 = 0.9999) were achieved for GFAP, h-FABP and S1ooβ biomarker, respectively.
Once the suitability for mTBI biomarker panel detection has been established on the MSD platform, the assays were translated to SPE-based ECL test set-up platform.

3.2. Development of ECL “Singleplex” Assays for mTBI Biomarkers on SPE

3.2.1. Choice of Electrode Material

The use of SPE in the context of highly sensitive ECL analytical and electrochemical diagnostic applications requires several different electrode characteristics such as: (a) fast electron transfer; (b) reproducible electrode surfaces that can improve the assay accuracy and precision; (c) large electroactive electrode areas to boost signal intensities; (d) broad potential window; and (e) a hydrophobic electrode surface to facilitate TPA oxidation on the electrode surface [47,48,49]. Hence, four commercially available SPEs were investigated for an application in ECL detection of mTBI biomarkers: SPCEs (DRP-110), SPCEs modified with gold nanoparticles SPCE-GNP (DRP-110-GNP), SPCEs modified with carbon nanotubes and gold nanoparticles SPCE-CNT-GNP (DRP-110CNT-GNP), and SPCEs modified with quantum dots SPCE-QD (DRP-110QD).
Firstly, the electrochemical properties of each electrode were tested using a standard redox couple (ferri/ferrocyanide) (Table 3). All electrodes showed quasi-reversible electrochemical behavior with the peak-to-peak separation (ΔE) higher than anticipated for the one-electron transfer process (>59 mV). The values were consistent with the ones reported by Banks et al. [50] and Fanjul-Bolado et al. [51] stating that it arises from a combination of the electrode properties (electrode material, composition of the paste used for the fabrication, the curing temperature, the hydrophilic characteristics of the electrode surface) [49]. The electrode electro-active area (A) was calculated using the Randles–Sevcik equation (Equation (2)) by studying the scan rates of 5 mM K4[Fe(CN)6]/K3[Fe(CN)6] in 0.1 M PBS) [52]:
ip = (2.69 × 105) · n3/2 · A · C · D1/2 · ν1/2
ip is the peak current (A), n is the number of electrons involved in the redox reaction, A is the active electrode area (cm2), C is the concentration of redox molecule (mol/cm3), D is the diffusion coefficient (cm2/s), and ν is the scan rate (V/s).
The representative cyclic voltammograms showing the scan rate studies with ferri/ferrocyanide couple on different SPEs are shown in Figure S2a. SPCE and SPCE-GNP electrodes showed the highest active electrode area (A) and the highest conductivity (the lowest peak-to-peak separation, ΔE). Electrochemical properties of each electrode material were also tested with [Ru(bpy)3]2+ in PBS 1×. All CVs showed a reversible oxidation peak at ~0.5 V vs. Ag and a corresponding reduction peak at ~0.3 V vs. Ag pseudo-reference electrode (illustrated in Figure S2b).
Furthermore, to be able to compare the SPEs in terms of relative ECL intensities, a commercial solution containing [Ru(bpy)3]2+ and TPA (MSD Free Tag 15,000) was used to generate ECL signal upon applying the potential on SPEs (Figure 2). In that context, SPCE electrodes showed the highest ECL intensities, followed in decreasing order by SPCE-GNP > SPCE-CNT-GNP > SPCE-QD.
Since SPCEs showed the best electrochemical performance and exhibited the highest ECL intensity (Figure 2), they were selected as solid-state support for development of mTBI assays. SEM images of SPCE electrodes are shown in Figure S3.

3.2.2. Optimization of SPCE-Based ECL Sensor

The ECL signal of the co-reactant couple Ru(bpy)32+–TPA is produced via the reaction between the deprotonated TPA radical (TPA) and electrogenerated Ru(bpy)33+, forming a [Ru(bpy)32+] * radical that generates light emission when returning to the ground state (Equations (3)–(6)). TPA can be created via catalytic oxidation by electrogenerated Ru(bpy)33+ (Equation (4a)) and direct electrode oxidation (Equation (4b)) [53]. Figure 3a shows an exemplary linear sweep voltammetric curve of SPCE (black line) and generated ECL curve (red line) obtained for h-FABP assay indicating a strong ECL peak at ~1.3 V vs. Ag electrode.
Ru(bpy)32+ -> Ru(bpy)33+ + e
Ru(bpy)33+ + TPA -> Ru(bpy)32+ + TPA
TPA -> TPA + e
Ru(bpy)33+ + TPA -> Ru(bpy)32+ *
Ru(bpy)32+ * -> Ru(bpy)32+ + hν (610 nm)
Furthermore, two options were evaluated for the ECL signal trigger from SPCE: for (i) linear sweep voltammetry (LSV) (Figure 3a), versus (ii) constant potential chronoamperometry (CPA) (Figure 3b). For LSV, the scan rate of 200 mV/s for LSV has been selected considering that higher scan rates led to broadening and shift of ECL peak towards positive potentials, which consequently led to water oxidation, causing the formation of bubbles on the electrode surface that had a negative impact on detected ECL signals. For CPA three different potentials were tested: 1.4, 1.5, 1.55, and 1.6 V.
ECL signals generated by LSV were ~800 a.u. (Figure 3a), while for CPA it was found that the ECL signal intensity increased as the applied potentials became more positive, showing the maximum value of ~2250 a.u. at 1.55 V vs. Ag electrode (Figure 3b—inset plot). Based on these results it was decided to use CPA for the further experiments and development of mTBI biomarker assay on SPCE electrodes.

3.2.3. Optimization of the Assay Conditions

The pre-optimized assay conditions obtained for each mTBI biomarker on the MSD platform (Table 2) were translated to the SPCE and µSTAT-ECL platform, with the only difference that the concentrations of capture and detection antibodies were re-optimized using the checkerboard assay due to the differences in the WE area and morphology between MSD and SPCE electrodes (cAb-dAb ratio was kept constant, Table 4). The results are presented in Figure 4 (left figures) as a heat map showing the S/B ratios from low (red) to high (green). The optimal capture and detection antibody concentrations were determined based on the maximum S/B results and summarized in Table 4.
Analytical performance of the developed singleplex mTBI assays on the SPCEs was evaluated using the buffer samples containing each individual biomarker concentration ranging from 1 ng mL−1 to 50 ng mL−1 using pre-optimized conditions from Table 4. Based on the measured values, a calibration curve (Figure 4) was modeled by the linear regression equations for each mTBI biomarker (n = 3) (Figure S4). The LOD values were calculated using the average value of the blank and adding to it three times the standard deviation of the blank. LOD values of 0.59 ng mL−1 (R2 = 0.9777), 0.44 ng mL−1 (R2 = 0.9931) and 1.34 ng mL−1 (R2 = 0.9984) were achieved for GFAP, h-FABP and S100β, respectively.

3.3. Standard Recovery Test

The applicability and reliability of developed mTBI biomarker assays, on both ECL instrument platforms, were evaluated in a complex physiological matrix by standard addition method. Different concentrations of biomarkers were added in human serum diluted 2× with respective assay diluents (defined as “spiked” concentration). Based on the obtained ECL signal intensity, the corresponding biomarker concentrations (defined as “detected” concentration) were calculated according to the regression equation performed using the same matrix (4PL dose-response curve). The recoveries were calculated by the ratio between the “detected” and “spiked”, and the obtained results are summarized in Table 5 (Figure S5). It could be seen that the recoveries ranged from 79% to 128%, which seems acceptable at this development stage. These results suggest that with further development and optimization, these assays have a potential for future mTBI clinical diagnostic applications.

3.4. Multiplex ECL Assay for mTBI Biomarker Panel on SPCE

Once the singleplex assays have been developed on SPCE (GFAP, h-FABP, S100β) they were integrated into a multiplexed assay with the goal to enable high-throughput simultaneous detection of multiple mTBI biomarkers on a single electrode.
SPCEs were spotted with capture antibodies of each biomarker (14 nL/spot, diameter ~500 µm) and with BSA protein labelled with SULFOTAG (alignment spots on the electrodes) (Figure 5), and the assay protocol was performed as described in Section 3.3 (S100β diluents were used as common diluents for all three biomarkers). Different electrode patterns were prepared, as described in Figure 5a–e. Obtained results indicated no cross-reactivity between the antibodies, providing a proof-of-concept that the present methodology based on spatially resolved approach can serve as a foundation for simultaneous, single electrode, detection of a mTBI biomarker panel, and other multi-biomarker panels (e.g., cardiac).

4. Discussion

The purpose of this study was to evaluate the feasibility of developing a multiplex ECL assay for mTBI biomarkers targeted for a POC diagnostic format and to identify the most important steps that need to be conquered on the way.
Firstly, ECL assays for three mTBI biomarkers (GFAP, h-FABP, S100β) have been developed on a reference benchtop ECL platform from MesoScale Discovery (96-well plate format). The analytical performances of these singleplex assays were impressive, reaching LODs of 6.94 pg mL−1, 1.35 pg mL−1 and 15.73 pg mL−1 for GFAP, h-FABP and S100β biomarker, respectively. The obtained LODs were close or below 1/10 of the cut-off values reported for these three biomarkers (22 pg mL−1, 2.62 ng mL−1 and 42 pg mL−1, Table 1), indicating that the developed assays have analytical sensitivities to distinguish a “mTBI condition” from “normal”, physiological concentration of the specific biomarker.
When comparing the obtained results with other published ECL-based methodologies, it is worth mentioning the work of Button et al., who reported a sandwich immunoassay for detection of GFAP biomarker on MSD platform in mouse plasma samples, with a LOD of 9 pg mL−1 [54]. Roche Diagnostics provides ECL-based immunoassay for the in vitro diagnostic quantitative determination of S100β in human serum based on Elecsys® technology on Cobas instruments, with reported LOD of 15 pg mL−1 [55]. Gan et al. reported an ECL immunosensor for detection of h-FABP biomarker using luminophore coupled with 2D metal-organic framework (LOD of 44.5 fg mL−1) [31]. Regarding the methodologies reported for detection of mTBI biomarkers suitable for POC settings, it is worth mentioning the lateral flow immunoassay (LFIA) approaches reported by Natarajan and Joseph [56] for rapid detection of GFAP biomarker (time-resolved fluorescence read-out, 25 min detection time, LOD 10 pg mL−1), the approach of Savin et al. [57] for detection of h-FABP biomarker (CdTe quantum dots labelled detection antibodies, LOD 221 pg mL−1), and the approach developed by Gao et al. [58] for detection of S100β (surface-enhanced Raman spectroscopy read-out, LOD 5 pg mL−1). A detailed list of other methodologies published so far in the context of detection of GFAP, h-FABP and S100β biomarkers is given in Table S1.
Even though the MSD platform allows facile assay development and optimization, the instrument is not suitable for POC diagnostics and particularly not for on-site accident applications due to the size, weight, and cost limitations. Thus, the ECL assays for mTBI biomarkers have been translated to screen-printed electrodes (SPEs) coupled with a (trans-)portable µSTAT-ECL reader from Metrohm DropSens (see Section 2.1). SPEs are an attractive choice for realizing POC-oriented devices due to their low cost, mass fabrication, small sample volume requirements, and ability to be easily integrated and miniaturized [59]. SPEs can be easily combined with ECL detection to achieve a cost-effective, simplified POC diagnostic device.
One of the most important electrode aspects in the context of ECL detection is the electrode material. Our results have shown that non-modified SPCE exhibit better performances in the ECL assays than all other tested electrodes from Metrohm DropSens (SPCE-GNP, SPCE-CNT-GNP, or SPCE-QD). A similar observation was also reported (excluding SPCE-QD) in the work of Kerr et al. [49]. Unmodified carbon is often chosen for WE since it is a cheap, widely commercially available material that exhibits rapid and efficient oxidation of TPA, is relatively hydrophobic, allowing high concentrations of TPA on the electrode surface, and has low rates of surface oxide formation compared to noble metal electrodes [49,52]. Furthermore, two different electrochemical techniques were employed for ECL signal trigger, linear sweep voltammetry (LSV) and constant potential chronoamperometry (CPA). The obtained results showed that CPA at 1.55 V increased ~3× the ECL signal intensities in the assay compared to the LSV (Figure 3b).
Even though in the literature there are plenty of ECL biosensors based on screen-printed electrodes [60], to the best of our knowledge, this is the first study showing the development of ECL assays on SPCE combined with the miniaturized µSTAT-ECL reader from Metrohm DropSens [41]. The LOD values of 0.59, 0.44, and 1.34 ng mL−1 were achieved for GFAP, h-FABP and S100β biomarker, respectively. The evident loss of sensitivity (compared to the MSD platform) could likely be attributed to the less performant μSTAT-ECL photodiode versus MSD CCD detector systems. On the other hand, new, highly sensitive and compact detectors in development will likely soon unfold possibilities for low pg mL−1 level LODs in POC diagnostic assays. Alternatively, for SPCEs to be applicable for detection of a mTBI-relevant biomarker panel at the clinically relevant concentration ranges (lower pg mL−1 range), further developments with ECL signal amplification strategies (e.g., exploring nanomaterials that have a higher capacity for loading of luminophores—e.g., solid-state luminophores) would be highly beneficial.
In terms of the simultaneous detection of multiple analytes, a variety of analytical technologies have been exploited, including fluorescence [61,62], electrochemistry [63,64], surface plasmon resonance [65], and chemiluminescence [66]. However, some techniques need expensive detectors or light sources, while at the same time may suffer insufficient sensitivity for measuring clinical samples. Electrochemiluminescence has shown an excellent potential thanks to the good selectivity and the fact that signal generation can be adjusted by the alternation of the trigger potential, allowing control over emission location and thus enabling multi-analyte detection from the sample, by employing single or multiple electrodes [67]. The methodologies for ECL-based multiplex assays can be mainly categorized as spatial-resolved, potential-resolved, spectrum-resolved, and other miscellaneous strategies [67]. Herein the spatially resolved approach on a single SPCE has been applied for the first time for multiplex ECL detection of three mTBI biomarkers. Translation of three individual singleplex assays into multiplex assay has been successfully done, indicating that such an approach could be further explored for detection of mTBI biomarker panel from limited sample volume (e.g., capillary blood sample, cerebrospinal fluid, etc.), and with the option to extend the biomarker panel and improve the clinical specificity of the test (e.g., fewer false positives).
Due to the differences in mTBI biomarker release kinetics, a multi-point detection at different time intervals after the injury would be an important requirement. For example, GFAP is present in serum samples at detectable concentrations (>30 pg mL−1) even within 1 h of injury [68], while for S100β only 12–36 h after trauma. The methodology could certainly benefit both clinicians and patients by being easily extended to other multi-target panels (cancer, cardiac disease, etc.). The spatially resolved approach evidently brings the requirement for the use of a CCD camera and for POC diagnostic applications, it would be necessary to employ a miniaturized and performant detector device and a cartridge unit that would handle all the sample processing steps.

5. Conclusions

Herein, electrochemiluminescence assays for detection of mTBI biomarkers (GFAP, h-FABP, S100β) have been developed and optimized on the MesoScale Discovery (MSD) platform in 96-well plate format, and on SPCE electrodes combined with µSTAT-ECL from Metrohm DropSens. A “proof-of-concept” for the development of a multiplex ECL assay on a single electrode has been shown. Obtained data could serve as a steppingstone towards the development of an ECL-based POC diagnostic device for multiplex detection of mTBI biomarkers, which can be employed as a diagnostic screening test, but also aid the preclinical evaluation of currently investigated biomarkers and in the establishment of an ‘ideal’ biomarker panel for TBI diagnostics (or adapted for the detection of other biomarker panels in different clinical body fluids).

Supplementary Materials

The following supporting information are available online at https://www.mdpi.com/article/10.3390/bios12030172/s1, Table S1: Summary of GFAP, h-FABP and S100β sensors developed in the last 10 years and their analytical performances; Figure S1: Four-parameter (4PL) dose-response nonlinear regression model for MSD calibration curves; Figure S2: Electrochemical characterization of commercially available screen-printed electrodes (SPE) for ECL applications; Figure S3: SEM images of commercially available Screen-Printed Cabon Electrodes (SPCEs) for ECL applications; Figure S4: Linear regression model for SPCE µSTAT-ECL calibration curves; Figure S5: Results of the recovery studies; Figure S6: CAD drawing of incubation cells used for SPEs. References [27,30,31,32,54,55,56,57,58,69,70,71,72,73,74,75,76,77,78,79,80,81,82,83,84,85,86,87,88,89,90,91,92,93,94,95,96,97,98,99,100,101,102,103,104,105,106,107,108,109,110,111,112,113,114,115,116,117,118,119] are cited in the supplementary materials.

Author Contributions

Conceptualization, M.J., D.P. and M.E.P.; methodology, M.J., E.S. and D.P.; writing—original draft preparation, M.J. and E.S.; writing—review and editing, D.P. and M.E.P.; supervision, M.E.P.; project administration, M.E.P.; funding acquisition, M.E.P. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by BRIDGE (joint programme conducted by the Swiss National Science Foundation (SNSF) and Innosuisse—the Swiss Innovation Agency), grant No. 40B2-0_181013.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The experimental data is contained within the article.

Acknowledgments

The authors would like to thank Alexandre Nicollier (HES-SO Valais-Wallis), Joana Soares da Silva (HES-SO Valais-Wallis), Léa Loiseau (HES-SO Valais-Wallis), and Aurélie Cino (Université de Genève) for their technical support with the execution of immunoassay experiments. The authors would also like to thank the mechanical workshop of HES-SO Valais-Wallis for the fabrication of customized incubation cell for SPEs.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Traumatic Brain Injury. Fact Sheets and Policy Brief. Available online: https://www.center-tbi.eu/ (accessed on 24 September 2020).
  2. Mondello, S.; Muller, U.; Jeromin, A.; Streeter, J.; Hayes, R.L.; Wang, K.K.W. Blood-Based Diagnostics of Traumatic Brain Injuries. Expert Rev. Mol. Diagn. 2011, 11, 65–78. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  3. Report to Congress: Traumatic Brain Injury in the United States|Concussion|Traumatic Brain Injury|CDC Injury Center. Available online: https://www.cdc.gov/traumaticbraininjury/pubs/tbi_report_to_congress.html (accessed on 15 April 2021).
  4. Mehta, R.; Trainee1, G.P.; Chinthapalli, K.; Neurologist2, C. Glasgow Coma Scale Explained. BMJ 2019, 365, l1296. [Google Scholar] [CrossRef] [PubMed]
  5. Dewan, M.C.; Rattani, A.; Gupta, S.; Baticulon, R.E.; Hung, Y.-C.; Punchak, M.; Agrawal, A.; Adeleye, A.O.; Shrime, M.G.; Rubiano, A.M.; et al. Estimating the Global Incidence of Traumatic Brain Injury. J. Neurosurg. 2018, 130, 1080–1097. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  6. Cnossen, M.C.; Winkler, E.A.; Yue, J.K.; Okonkwo, D.O.; Valadka, A.B.; Steyerberg, E.W.; Lingsma, H.F.; Manley, G.T. Development of a Prediction Model for Post-Concussive Symptoms Following Mild Traumatic Brain Injury: A TRACK-TBI Pilot Study. J. Neurotrauma 2017, 34, 2396–2409. [Google Scholar] [CrossRef] [PubMed]
  7. Powell, J.M.; Ferraro, J.V.; Dikmen, S.S.; Temkin, N.R.; Bell, K.R. Accuracy of Mild Traumatic Brain Injury Diagnosis. Arch. Phys. Med. Rehabil. 2008, 89, 1550–1555. [Google Scholar] [CrossRef] [PubMed]
  8. Martinez, B.I.; Martinez, B.I.; Stabenfeldt, S.E. Current Trends in Biomarker Discovery and Analysis Tools for Traumatic Brain Injury. J. Biol. Eng. 2019, 16, 1–12. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  9. Yokobori, S.; Hosein, K.; Burks, S.; Sharma, I.; Gajavelli, S.; Bullock, R. Biomarkers for the Clinical Differential Diagnosis in Traumatic Brain Injury-a Systematic Review. CNS Neurosci. Ther. 2013, 19, 556–565. [Google Scholar] [CrossRef]
  10. Tomar, G.S.; Singh, G.P.; Lahkar, D.; Sengar, K.; Nigam, R.; Mohan, M.; Anindya, R. New Biomarkers in Brain Trauma. Clin. Chim. Acta 2018, 487, 325–329. [Google Scholar] [CrossRef]
  11. Sharma, R.; Rosenberg, A.; Bennett, E.R.; Laskowitz, D.T.; Acheson, S.K. A Blood-Based Biomarker Panel to Risk-Stratify Mild Traumatic Brain Injury. PLoS ONE 2017, 12, e0173798. [Google Scholar] [CrossRef] [Green Version]
  12. Blennow, K.; Brody, D.L.; Kochanek, P.M.; Levin, H.; McKee, A.; Ribbers, G.M.; Yaffe, K.; Zetterberg, H. Traumatic Brain Injuries. Nat. Rev. Dis. Primers 2016, 2, 1–19. [Google Scholar] [CrossRef]
  13. Krausz, A.D.; Korley, F.K.; Burns, M.A. The Current State of Traumatic Brain Injury Biomarker Measurement Methods. Biosensors 2021, 11, 319. [Google Scholar] [CrossRef] [PubMed]
  14. Undén, L.; Calcagnile, O.; Undén, J.; Reinstrup, P.; Bazarian, J. Validation of the Scandinavian Guidelines for Initial Management of Minimal, Mild and Moderate Traumatic Brain Injury in Adults. BMC Med. 2015, 13, 292–301. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  15. Ohrt-Nissen, S.; Friis-Hansen, L.; Dahl, B.; Stensballe, J.; Romner, B.; Rasmussen, L.S. How Does Extracerebral Trauma Affect the Clinical Value of S100B Measurements? Emerg. Med. J. 2011, 28, 941–944. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  16. Kahouadji, S.; Salamin, P.; Praz, L.; Coiffier, J.; Frochaux, V.; Durif, J.; Pereira, B.; Arlettaz, L.; Oris, C.; Sapin, V.; et al. S100B Blood Level Determination for Early Management of Ski-Related Mild Traumatic Brain Injury: A Pilot Study. Front. Neurol. 2020, 11, 856. [Google Scholar] [CrossRef]
  17. Hasselblatt, M.; Mooren, F.C.; von Ahsen, N.; Keyvani, K.; Fromme, A.; Schwarze-Eicker, K.; Senner, V.; Paulus, W. Serum S100beta Increases in Marathon Runners Reflect Extracranial Release Rather than Glial Damage. Neurology 2004, 62, 1634–1636. [Google Scholar] [CrossRef] [PubMed]
  18. Missler, U.; Wiesmann, M.; Wittmann, G.; Magerkurth, O.; Hagenström, H. Measurement of Glial Fibrillary Acidic Protein in Human Blood: Analytical Method and Preliminary Clinical Results. Clin. Chem. 1999, 45, 138–141. [Google Scholar] [CrossRef] [Green Version]
  19. Luoto, T.M.; Raj, R.; Posti, J.P.; Gardner, A.J.; Panenka, W.J.; Iverson, G.L. A Systematic Review of the Usefulness of Glial Fibrillary Acidic Protein for Predicting Acute Intracranial Lesions Following Head Trauma. Front. Neurol. 2017, 8, 652. [Google Scholar] [CrossRef] [Green Version]
  20. Mondello, S.; Sorinola, A.; Czeiter, E.; Vámos, Z.; Amrein, K.; Synnot, A.; Donoghue, E.; Sandor, J.; Wang, K.K.; Diaz-Arrastia, R. Blood-Based Protein Biomarkers for the Management of Traumatic Brain Injuries in Adults Presenting to Emergency Departments with Mild Brain Injury: A Living Systematic Review and Meta-Analysis. J. Neurotrauma 2021, 38, 1086–1106. [Google Scholar] [CrossRef] [Green Version]
  21. Di Pietro, V.; Yakoub, K.M.; Scarpa, U.; Di Pietro, C.; Belli, A. MicroRNA Signature of Traumatic Brain Injury: From the Biomarker Discovery to the Point-of-Care. Front. Neurol. 2018, 9, 429. [Google Scholar] [CrossRef] [Green Version]
  22. Furuhashi, M.; Hotamisligil, G.S. Fatty Acid-Binding Proteins: Role in Metabolic Diseases and Potential as Drug Targets. Nat. Rev. Drug Discov. 2008, 7, 489–503. [Google Scholar] [CrossRef] [Green Version]
  23. Lagerstedt, L.; Egea-Guerrero, J.J.; Bustamante, A.; Montaner, J.; Rodriguez-Rodriguez, A.; El Rahal, A.; Turck, N.; Quintana, M.; Garcia-Armengol, R.; Prica, C.M.; et al. H-FABP: A New Biomarker to Differentiate between CT-Positive and CT-Negative Patients with Mild Traumatic Brain Injury. PLoS ONE 2017, 12, e0175572. [Google Scholar] [CrossRef] [PubMed]
  24. Bogoslovsky, T.; Gill, J.; Jeromin, A.; Davis, C.; Diaz-Arrastia, R. Fluid Biomarkers of Traumatic Brain Injury and Intended Context of Use. Diagnostics 2016, 6, 37. [Google Scholar] [CrossRef] [PubMed]
  25. Posti, J.P.; Takala, R.S.K.; Lagerstedt, L.; Dickens, A.M.; Hossain, I.; Mohammadian, M.; Ala-Seppälä, H.; Frantzén, J.; van Gils, M.; Hutchinson, P.J.; et al. Correlation of Blood Biomarkers and Biomarker Panels with Traumatic Findings on Computed Tomography after Traumatic Brain Injury. J. Neurotrauma 2019, 36, 2178–2189. [Google Scholar] [CrossRef] [PubMed]
  26. Lagerstedt, L.; Egea-Guerrero, J.J.; Bustamante, A.; Rodríguez-Rodríguez, A.; Rahal, A.E.; Quintana-Diaz, M.; García-Armengol, R.; Prica, C.M.; Andereggen, E.; Rinaldi, L.; et al. Combining H-FABP and GFAP Increases the Capacity to Differentiate between CT-Positive and CT-Negative Patients with Mild Traumatic Brain Injury. PLoS ONE 2018, 13, e0200394. [Google Scholar] [CrossRef] [PubMed]
  27. Bazarian, J.J.; Biberthaler, P.; Welch, R.D.; Lewis, L.M.; Barzo, P.; Bogner-Flatz, V.; Brolinson, P.G.; Büki, A.; Chen, J.Y.; Christenson, R.H.; et al. Serum GFAP and UCH-L1 for Prediction of Absence of Intracranial Injuries on Head CT (ALERT-TBI): A Multicentre Observational Study. Lancet Neurol. 2018, 17, 782–789. [Google Scholar] [CrossRef]
  28. Abbott Receives FDA 510(k) Clearance for the First Rapid Handheld Blood Test for Concussions. Available online: https://abbott.mediaroom.com/2021-01-11-Abbott-Receives-FDA-510-k-Clearance-for-the-First-Rapid-Handheld-Blood-Test-for-Concussions (accessed on 24 March 2021).
  29. Nanodiagnostics Nanosensor Technology|Our Solution. Available online: https://nanodiagnostics.com/our-solution/nanosensor-technology/ (accessed on 29 March 2021).
  30. Pankratova, N.; Jović, M.; Pfeifer, M.E. Electrochemical Sensing of Blood Proteins for Mild Traumatic Brain Injury (MTBI) Diagnostics and Prognostics: Towards a Point-of-Care Application. RSC Adv. 2021, 11, 17301–17319. [Google Scholar] [CrossRef] [PubMed]
  31. Gan, X.; Han, D.; Wang, J.; Liu, P.; Li, X.; Zheng, Q.; Yan, Y. A Highly Sensitive Electrochemiluminescence Immunosensor for H-FABP Determination Based on Self-Enhanced Luminophore Coupled with Ultrathin 2D Nickel Metal-Organic Framework Nanosheets. Biosens. Bioelectron. 2021, 171, 112735. [Google Scholar] [CrossRef] [PubMed]
  32. Qin, X.; Zhang, X.; Wang, M.; Dong, Y.; Liu, J.; Zhu, Z.; Li, M.; Yang, D.; Shao, Y. Fabrication of Tris(Bipyridine)Ruthenium(II)-Functionalized Metal–Organic Framework Thin Films by Electrochemically Assisted Self-Assembly Technique for Electrochemiluminescent Immunoassay. Anal. Chem. 2018, 90, 11622–11628. [Google Scholar] [CrossRef]
  33. Fu, Z.; Wei, W.; Li, C.; Wang, Z. Electrochemiluminescent Immunosensing and Its Application in Biological and Pharmaceutical Analysis. Sci. Sin.-Chim 2011, 41, 773–784. [Google Scholar] [CrossRef]
  34. Hoffmann, U.; Espeter, F.; Weiss, C.; Ahmad-Nejad, P.; Lang, S.; Brueckmann, M.; Akin, I.; Neumaier, M.; Borggrefe, M.; Behnes, M. Ischemic Biomarker Heart-Type Fatty Acid Binding Protein (HFABP) in Acute Heart Failure—Diagnostic and Prognostic Insights Compared to NT-ProBNP and Troponin I. BMC Cardiovasc. Disord. 2015, 15, 50. [Google Scholar] [CrossRef] [Green Version]
  35. Abir, J.; Sondes, S.; Rania, E.; Latifa, K.; Mokhles, B.D.; Nedia, B.; Hadj, B.; Manel, M.; Souhir, K.; Hejer, G.; et al. Serum Heart Type Fatty Acid Binding Protein (H-FABP) Levels in Metabolic Syndrome. Int. J. Pharm. Sci. Res. 2017, 8, 1441–1448. [Google Scholar]
  36. Wiesmann, M.; Missler, U.; Gottmann, D.; Gehring, S. Plasma S-100b Protein Concentration in Healthy Adults Is Age-and Sex-Independent. Clin. Chem. 1998, 44, 1056–1058. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  37. Adrian, H.; Marten, K.; Salla, N.; Lasse, V. Biomarkers of Traumatic Brain Injury: Temporal Changes in Body Fluids. eNeuro 2016, 3, 1–13. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  38. Gordillo-Escobar, E.; Egea-Guerrero, J.J.; Rodríguez-Rodríguez, A.; Murillo-Cabezas, F. Usefulness of Biomarkers in the Prognosis of Severe Head Injuries. Med. Intensiva 2016, 40, 105–112. [Google Scholar] [CrossRef] [PubMed]
  39. Lewis, L.M.; Schloemann, D.T.; Lindburg, M.; Papa, L.; Fucetola, R.P.; Bazarian, J.; Welch, R.D. Utility of Serum Biomarkers in the Diagnosis and Stratification of Mild Traumatic Brain Injury. Acad. Emerg. Med. 2017, 24, 710–720. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  40. MSD Technology Platform. Available online: https://www.mesoscale.com/~/media/files/brochures/techbrochure.pdf (accessed on 24 September 2020).
  41. Neves, M.M.P.S.; Bobes-Limenes, P.; Pérez-Junquera, A.; González-García, M.B.; Hernández-Santos, D.; Fanjul-Bolado, P. Miniaturized Analytical Instrumentation for Electrochemiluminescence Assays: A Spectrometer and a Photodiode-Based Device. Anal. Bioanal. Chem. 2016, 408, 7121–7127. [Google Scholar] [CrossRef] [PubMed]
  42. Smith, S.P.; Shaw, G.S. A Novel Calcium-Sensitive Switch Revealed by the Structure of Human S100B in the Calcium-Bound Form. Structure 1998, 6, 211–222. [Google Scholar] [CrossRef] [Green Version]
  43. Gonçalves, D.S.; Lenz, G.; Karl, J.; Gonçalves, C.A.; Rodnight, R. Extracellular S100B Protein Modulates ERK in Astrocyte Cultures. Neuroreport 2000, 11, 807–809. [Google Scholar] [CrossRef] [Green Version]
  44. Green, A.J.; Keir, G.; Thompson, E.J. A Specific and Sensitive ELISA for Measuring S-100b in Cerebrospinal Fluid. J. Immunol. Methods 1997, 205, 35–41. [Google Scholar] [CrossRef]
  45. Tramontina, F.; Karl, J.; Gottfried, C.; Mendez, A.; Gonçalves, D.; Portela, L.V.; Gonçalves, C.-A. Digitonin-Permeabilization of Astrocytes in Culture Monitored by Trypan Blue Exclusion and Loss of S100B by ELISA. Brain Res. Protoc. 2000, 6, 86–90. [Google Scholar] [CrossRef]
  46. Tort, A.B.L.; Dietrich, M.O.; Gonçalves, C.A.; Souza, D.O.; Portela, L.V.C. Influence of Anticoagulants on the Measurement of S100B Protein in Blood. Clin. Biochem. 2003, 36, 629–632. [Google Scholar] [CrossRef]
  47. Valenti, G.; Fiorani, A.; Li, H.; Sojic, N.; Paolucci, F. Essential Role of Electrode Materials in Electrochemiluminescence Applications. ChemElectroChem 2016, 3, 1990–1997. [Google Scholar] [CrossRef]
  48. Workman, S.; Richter, M.M. The Effects of Nonionic Surfactants on the Tris(2,2‘-Bipyridyl)Ruthenium(II)–Tripropylamine Electrochemiluminescence System. Anal. Chem. 2000, 72, 5556–5561. [Google Scholar] [CrossRef] [PubMed]
  49. Kerr, E.; Alexander, R.; Francis, P.S.; Guijt, R.M.; Barbante, G.J.; Doeven, E.H. A Comparison of Commercially Available Screen-Printed Electrodes for Electrogenerated Chemiluminescence Applications. Front. Chem. 2021, 8, 628483. [Google Scholar] [CrossRef] [PubMed]
  50. Banks, C.E.; Davies, T.J.; Wildgoose, G.G.; Compton, R.G. Electrocatalysis at Graphite and Carbon Nanotube Modified Electrodes: Edge-Plane Sites and Tube Ends Are the Reactive Sites. Chem. Commun. 2005, 829–841. [Google Scholar] [CrossRef] [PubMed]
  51. Fanjul-Bolado, P.; Hernández-Santos, D.; Lamas-Ardisana, P.J.; Martín-Pernía, A.; Costa-García, A. Electrochemical Characterization of Screen-Printed and Conventional Carbon Paste Electrodes. Electrochim. Acta 2008, 53, 3635–3642. [Google Scholar] [CrossRef]
  52. Bard, A.J.; Faulkner, L.R. Electrochemical Methods: Fundamentals and Applications, 2nd ed.; John Wiley & Sons: New York, NY, USA, 2001. [Google Scholar]
  53. Zu, Y.; Bard, A.J. Electrogenerated Chemiluminescence. 67. Dependence of Light Emission of the Tris(2,2‘)Bipyridylruthenium(II)/Tripropylamine System on Electrode Surface Hydrophobicity. Anal. Chem. 2001, 73, 3960–3964. [Google Scholar] [CrossRef] [PubMed]
  54. Button, E.B.; Cheng, W.H.; Barron, C.; Cheung, H.; Bashir, A.; Cooper, J.; Gill, J.; Stukas, S.; Baron, D.C.; Robert, J.; et al. Development of a Novel, Sensitive Translational Immunoassay to Detect Plasma Glial Fibrillary Acidic Protein (GFAP) after Murine Traumatic Brain Injury. Alz. Res. Ther. 2021, 13, 58. [Google Scholar] [CrossRef]
  55. Elecsys® S100. Available online: https://diagnostics.roche.com/global/en/products/params/elecsys-s100.html (accessed on 27 August 2021).
  56. Natarajan, S.; Joseph, J. A Novel Time-Resolved Fluorescent Lateral Flow Immunoassay for Quantitative Detection of the Trauma Brain Injury Biomarker-Glial Fibrillary Acidic Protein. Sens. Diagn. 2022, 1, 193–197. [Google Scholar] [CrossRef]
  57. Savin, M.; Mihailescu, C.-M.; Matei, I.; Stan, D.; Moldovan, C.A.; Ion, M.; Baciu, I. A Quantum Dot-Based Lateral Flow Immunoassay for the Sensitive Detection of Human Heart Fatty Acid Binding Protein (HFABP) in Human Serum. Talanta 2018, 178, 910–915. [Google Scholar] [CrossRef]
  58. Gao, X.; Boryczka, J.; Zheng, P.; Kasani, S.; Yang, F.; Engler-Chiurazzi, E.B.; Simpkins, J.W.; Wigginton, J.G.; Wu, N. A “Hot Spot”-Enhanced Paper Lateral Flow Assay for Ultrasensitive Detection of Traumatic Brain Injury Biomarker S-100β in Blood Plasma. Biosens. Bioelectron. 2021, 177, 112967. [Google Scholar] [CrossRef] [PubMed]
  59. Mincu, N.-B.; Lazar, V.; Stan, D.; Mihailescu, C.M.; Iosub, R.; Mateescu, A.L. Screen-Printed Electrodes (SPE) for In Vitro Diagnostic Purpose. Diagnostics 2020, 10, 517. [Google Scholar] [CrossRef] [PubMed]
  60. Martínez-Periñán, E.; Gutiérrez-Sánchez, C.; García-Mendiola, T.; Lorenzo, E. Electrochemiluminescence Biosensors Using Screen-Printed Electrodes. Biosensors 2020, 10, 118. [Google Scholar] [CrossRef] [PubMed]
  61. Cao, Z.; Li, H.; Lau, C.; Zhang, Y. Cross-Talk-Free Simultaneous Fluoroimmunoassay of Two Biomarkers Based on Dual-Color Quantum Dots. Anal. Chim. Acta 2011, 698, 44–50. [Google Scholar] [CrossRef] [PubMed]
  62. Wang, S.; Zheng, L.; Cai, G.; Liu, N.; Liao, M.; Li, Y.; Zhang, X.; Lin, J. A Microfluidic Biosensor for Online and Sensitive Detection of Salmonella Typhimurium Using Fluorescence Labeling and Smartphone Video Processing. Biosens. Bioelectron. 2019, 140, 111333. [Google Scholar] [CrossRef]
  63. Wu, Y.; Xue, P.; Kang, Y.; Hui, K.M. Paper-Based Microfluidic Electrochemical Immunodevice Integrated with Nanobioprobes onto Graphene Film for Ultrasensitive Multiplexed Detection of Cancer Biomarkers. Anal. Chem. 2013, 85, 8661–8668. [Google Scholar] [CrossRef] [PubMed]
  64. Liu, Z.; Rong, Q.; Ma, Z.; Han, H. One-Step Synthesis of Redox-Active Polymer/AU Nanocomposites for Electrochemical Immunoassay of Multiplexed Tumor Markers. Biosens. Bioelectron. 2015, 65, 307–313. [Google Scholar] [CrossRef]
  65. Otsuki, S.; Ishikawa, M. Wavelength-Scanning Surface Plasmon Resonance Imaging for Label-Free Multiplexed Protein Microarray Assay. Biosens. Bioelectron. 2010, 26, 202–206. [Google Scholar] [CrossRef]
  66. Yang, Z.; Zong, C.; Yan, F.; Ju, H. Automated Chemiluminescent Dual-Analyte Immunoassay Based on Resolved Immunosensing Channels. Talanta 2010, 82, 1462–1467. [Google Scholar] [CrossRef]
  67. Lv, W.; Ye, H.; Yuan, Z.; Liu, X.; Chen, X.; Yang, W. Recent Advances in Electrochemiluminescence-Based Simultaneous Detection of Multiple Targets. TrAC Trends Anal. Chem. 2020, 123, 115767. [Google Scholar] [CrossRef]
  68. Papa, L.; Brophy, G.M.; Welch, R.D.; Lewis, L.M.; Braga, C.F.; Tan, C.N.; Ameli, N.J.; Lopez, M.A.; Haeussler, C.A.; Giordano, D.I.M. Time Course and Diagnostic Accuracy of Glial and Neuronal Blood Biomarkers GFAP and UCH-L1 in a Large Cohort of Trauma Patients with and without Mild Traumatic Brain Injury. JAMA Neurol. 2016, 73, 551–560. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  69. Arya, S.K.; Pui, T.S.; Wong, C.C.; Kumar, S.; Rahman, A.R.A. Effects of the Electrode Size and Modification Protocol on a Label-Free Electrochemical Biosensor. Langmuir 2013, 29, 6770–6777. [Google Scholar] [CrossRef] [PubMed]
  70. Huang, W.; Besar, K.; LeCover, R.; Dulloor, P.; Sinha, J.; Martínez Hardigree, J.F.; Pick, C.; Swavola, J.; Everett, A.D.; Frechette, J.; et al. Label-Free Brain Injury Biomarker Detection Based on Highly Sensitive Large Area Organic Thin Film Transistor with Hybrid Coupling Layer. Chem. Sci. 2014, 5, 416–426. [Google Scholar] [CrossRef]
  71. Wang, T.; Fang, Y.; He, Z. Electrochemical Quantitative Detection of Glial Fibrillary Acidic Protein Based on Molecularly Imprinted Polymer Sensor. Int. J. Electrochem. Sci. 2017, 12, 7341–7350. [Google Scholar] [CrossRef]
  72. Song, J.; Dailey, J.; Li, H.; Jang, H.-J.; Zhang, P.; Wang, J.T.-H.; Everett, A.D.; Katz, H.E. Extended Solution Gate OFET-Based Biosensor for Label-Free Glial Fibrillary Acidic Protein Detection with Polyethylene Glycol-Containing Bioreceptor Layer. Adv. Funct. Mater. 2017, 27, 1606506. [Google Scholar] [CrossRef] [PubMed]
  73. Khetani, S.; Ozhukil Kollath, V.; Kundra, V.; Nguyen, M.D.; Debert, C.; Sen, A.; Karan, K.; Sanati-Nezhad, A. Polyethylenimine Modified Graphene-Oxide Electrochemical Immunosensor for the Detection of Glial Fibrillary Acidic Protein in Central Nervous System Injury. ACS Sens. 2018, 3, 844–851. [Google Scholar] [CrossRef] [PubMed]
  74. Ma, Y.; Xu, G.; Wei, F.; Cen, Y.; Song, Y.; Ma, Y.; Xu, X.; Shi, M.; Sohail, M.; Hu, Q. Carbon Dots Based Immunosorbent Assay for the Determination of GFAP in Human Serum. Nanotechnology 2018, 29, 145501. [Google Scholar] [CrossRef]
  75. Cardinell, B.A.; Addington, C.P.; Stabenfeldt, S.E.; La Belle, J.T. Multi-Biomarker Detection Following Traumatic Brain Injury. Crit. Rev. Biomed. Eng. 2019, 47, 193–206. [Google Scholar] [CrossRef] [PubMed]
  76. Agostini, M.; Amato, F.; Vieri, M.L.; Greco, G.; Tonazzini, I.; Baroncelli, L.; Caleo, M.; Vannini, E.; Santi, M.; Signore, G.; et al. Glial-Fibrillary-Acidic-Protein (GFAP) Biomarker Detection in Serum-Matrix: Functionalization Strategies and Detection by an Ultra-High-Frequency Surface-Acoustic-Wave (UHF-SAW) Lab-on-Chip. Biosens. Bioelectron. 2021, 172, 112774. [Google Scholar] [CrossRef]
  77. Krausz, A.D.; Korley, F.K.; Burns, M.A. A Variable Height Microfluidic Device for Multiplexed Immunoassay Analysis of Traumatic Brain Injury Biomarkers. Biosensors 2021, 11, 320. [Google Scholar] [CrossRef] [PubMed]
  78. GFAP ELISA Kit|Sigma-Aldrich. Available online: https://www.sigmaaldrich.com/HK/zh/life-science/supelco (accessed on 30 August 2021).
  79. R-PLEX Human GFAP Antibody Set|Meso Scale Discovery. Available online: https://www.mesoscale.com/en/products/r-plex-human-gfap-antibody-set-f211m/ (accessed on 31 August 2021).
  80. Available online: https://www.Banyanbio.Com/Assets/Files/000175_vA-IFU-Banyan-BTI.Pdf (accessed on 27 April 2020).
  81. Simoa Technology. Available online: https://www.Quanterix.Com/Technology (accessed on 27 April 2020).
  82. TBI Plasma Cartridge|Abbott Point of Care. Available online: https://www.pointofcare.abbott/us/en/offerings/istat/istat-test-cartridges/TBI-Plasma (accessed on 30 August 2021).
  83. Okonkwo, D.O.; Puffer, R.C.; Puccio, A.M.; Yuh, E.L.; Yue, J.K.; Diaz-Arrastia, R.; Korley, F.K.; Wang, K.K.W.; Sun, X.; Taylor, S.R.; et al. Point-of-Care Platform Blood Biomarker Testing of Glial Fibrillary Acidic Protein versus S100 Calcium-Binding Protein B for Prediction of Traumatic Brain Injuries: A Transforming Research and Clinical Knowledge in Traumatic Brain Injury Study. J. Neurotrauma 2020, 37, 2460–2467. [Google Scholar] [CrossRef] [PubMed]
  84. Korley, F.K.; Datwyler, S.A.; Jain, S.; Sun, X.; Beligere, G.; Chandran, R.; Marino, J.A.; McQuiston, B.; Zhang, H.; Caudle, K.L.; et al. Comparison of GFAP and UCH-L1 Measurements from Two Prototype Assays: The Abbott i-STAT and ARCHITECT Assays. Neurotrauma Rep. 2021, 2, 193–199. [Google Scholar] [CrossRef] [PubMed]
  85. Miller, C.J.; Hu, J. Magnetic Immunosensor and Method of Use. U.S. Patent 9,958,440, 2018. [Google Scholar]
  86. The NanoDxTM System for TBI|Our Solution. Available online: https://nanodiagnostics.com/our-solution/the-nanodx-system-for-tbi/ (accessed on 30 August 2021).
  87. Feng, L.-N.; Bian, Z.-P.; Peng, J.; Jiang, F.; Yang, G.-H.; Zhu, Y.-D.; Yang, D.; Jiang, L.-P.; Zhu, J.-J. Ultrasensitive Multianalyte Electrochemical Immunoassay Based on Metal Ion Functionalized Titanium Phosphate Nanospheres. Anal. Chem. 2012, 84, 7810–7815. [Google Scholar] [CrossRef] [PubMed]
  88. Stan, D.; Mihailescu, C.-M.; Iosub, R.; Moldovan, C.; Savin, M.; Baciu, I. Electrochemical Studies of Homogeneous Self-Assembled Monolayers versus Mixed Self-Assembled Monolayers on Gold Electrode for “Label Free” Detection of Heart Fatty Acid Binding Protein. Thin Solid Film. 2012, 526, 143–149. [Google Scholar] [CrossRef]
  89. Carless, D.R.; Wnęk, M.; Knox, C.; Harrison, K.R.; Calder, N.; Hall, A.S.; Barth, J.H. Clinical and Analytical Evaluation of an Immunoturbidimetric Heart-Type Fatty Acid-Binding Protein Assay. Scand. J. Clin. Lab. Investig. 2013, 73, 48–53. [Google Scholar] [CrossRef] [PubMed]
  90. Mihailescu, C.-M.; Stan, D.; Iosub, R.; Moldovan, C.; Savin, M. A Sensitive Capacitive Immunosensor for Direct Detection of Human Heart Fatty Acid-Binding Protein (h-FABP). Talanta 2015, 132, 37–43. [Google Scholar] [CrossRef] [PubMed]
  91. Qin, X.; Xu, A.; Liu, L.; Sui, Y.; Li, Y.; Tan, Y.; Chen, C.; Xie, Q. Selective Staining of CdS on ZnO Biolabel for Ultrasensitive Sandwich-Type Amperometric Immunoassay of Human Heart-Type Fatty-Acid-Binding Protein and Immunoglobulin G. Biosens. Bioelectron. 2017, 91, 321–327. [Google Scholar] [CrossRef] [PubMed]
  92. Crapnell, R.D.; Canfarotta, F.; Czulak, J.; Johnson, R.; Betlem, K.; Mecozzi, F.; Down, M.P.; Eersels, K.; van Grinsven, B.; Cleij, T.J.; et al. Thermal Detection of Cardiac Biomarkers Heart-Fatty Acid Binding Protein and ST2 Using a Molecularly Imprinted Nanoparticle-Based Multiplex Sensor Platform. ACS Sens. 2019, 4, 2838–2845. [Google Scholar] [CrossRef] [PubMed]
  93. Li, F.; Guo, L.; Hu, Y.; Li, Z.; Liu, J.; He, J.; Cui, H. Multiplexed Chemiluminescence Determination of Three Acute Myocardial Infarction Biomarkers Based on Microfluidic Paper-Based Immunodevice Dual Amplified by Multifunctionalized Gold Nanoparticles. Talanta 2020, 207, 120346. [Google Scholar] [CrossRef] [PubMed]
  94. TBICheck—The Fastest Diagnostic Test for Traumatic Brain Injury. Available online: https://tbicheck.com/home (accessed on 30 August 2021).
  95. Human H-FABP ELISA Kit (Ab243682)|Abcam. Available online: https://www.abcam.com/Human-H-FABP-ELISA-Kit-ab243682.html?gclsrc=aw.ds|aw.ds&gclid=Cj0KCQjwg7KJBhDyARIsAHrAXaH7SK6R4I_1DbOvLN1pxlnqW42hdrT0Nk4Yjew7t-ynxwT-aoeUn4EaAuTJEALw_wcB (accessed on 30 August 2021).
  96. R-PLEX Human FABP3/H-FABP Antibody Set|Meso Scale Discovery. Available online: https://www.mesoscale.com/en/products/r-plex-human-fabp3-h-fabp-antibody-set-f214t/ (accessed on 31 August 2021).
  97. Missler, U.; Wiesmann, M.; Friedrich, C.; Kaps, M. S-100 Protein and Neuron-Specific Enolase Concentrations in Blood as Indicators of Infarction Volume and Prognosis in Acute Ischemic Stroke. Stroke 1997, 28, 1956–1960. [Google Scholar] [CrossRef] [PubMed]
  98. PII: S0022-1759(97)00050-1|Elsevier Enhanced Reader. Available online: https://reader.elsevier.com/reader/sd/pii/S0022175997000501?token=A464853B1ED8E1A9E89C9B8F041BB4E0DF5188385648EF4C35723E513BFC3977CFC76D86BA77D9A85AFD7B71FF48711E&originRegion=eu-west-1&originCreation=20210827152016 (accessed on 27 August 2021).
  99. Leite, M.C.; Galland, F.; Brolese, G.; Guerra, M.C.; Bortolotto, J.W.; Freitas, R.; de Almeida, L.M.V.; Gottfried, C.; Gonçalves, C.-A. A Simple, Sensitive and Widely Applicable ELISA for S100B: Methodological Features of the Measurement of This Glial Protein. J. Neurosci. Methods 2008, 169, 93–99. [Google Scholar] [CrossRef] [PubMed]
  100. Liu, Y.; Wang, H.; Chen, J.; Liu, C.; Li, W.; Kong, J.; Yang, P.; Liu, B. A Sensitive Microchip-Based Immunosensor for Electrochemical Detection of Low-Level Biomarker S100B. Electroanalysis 2013, 25, 1050–1055. [Google Scholar] [CrossRef]
  101. Wang, X.; Li, H.; Li, X.; Chen, Y.; Yin, Y.; Li, G. Electrochemical Assay of Melanoma Biomarker in Human Blood. Electrochem. Commun. 2014, 39, 12–14. [Google Scholar] [CrossRef]
  102. Mikuła, E.; Wysłouch-Cieszyńska, A.; Zhukova, L.; Puchalska, M.; Verwilst, P.; Dehaen, W.; Radecki, J.; Radecka, H. Voltammetric Detection of S100B Protein Using His-Tagged Receptor Domains for Advanced Glycation End Products (RAGE) Immobilized onto a Gold Electrode Surface. Sensors 2014, 14, 10650–10663. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  103. Kim, C.; Searson, P.C. Magnetic Bead-Quantum Dot Assay for Detection of a Biomarker for Traumatic Brain Injury. Nanoscale 2015, 7, 17820–17826. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  104. Kurzątkowska, K.; Jankowska, A.; Wysłouch-Cieszyńska, A.; Zhukova, L.; Puchalska, M.; Dehaen, W.; Radecka, H.; Radecki, J. Voltammetric Detection of the S100B Protein Using His-Tagged RAGE Domain Immobilized onto a Gold Electrode Modified with a Dipyrromethene—Cu(II) Complex and Different Diluents. J. Electroanal. Chem. 2016, 767, 76–83. [Google Scholar] [CrossRef]
  105. Khetani, S.; Aburashed, R.; Singh, A.; Sen, A.; Sanati-Nezhad, A. Immunosensing of S100β Biomarker for Diagnosis of Spinal Cord Injuries (SCI). Sens. Actuators B Chem. 2017, 247, 163–169. [Google Scholar] [CrossRef]
  106. Kuo, Y.-C.; Lee, C.-K.; Lin, C.-T. Improving Sensitivity of a Miniaturized Label-Free Electrochemical Biosensor Using Zigzag Electrodes. Biosens. Bioelectron. 2018, 103, 130–137. [Google Scholar] [CrossRef] [PubMed]
  107. Wang, Y.; Zhao, P.; Mao, L.; Hou, Y.; Li, D. Determination of Brain Injury Biomarkers by Surface-Enhanced Raman Scattering Using Hollow Gold Nanospheres. RSC Adv. 2018, 8, 3143–3150. [Google Scholar] [CrossRef] [Green Version]
  108. Mathew, A.S.; Shi, X.; Yau, S.-T. Detection of a Traumatic Brain Injury Biomarker at the 10 Fg/ML Level. Mol. Diagn. Ther. 2018, 22, 729–735. [Google Scholar] [CrossRef]
  109. Sun, J.; Zhao, Y.; Hou, Y.; Li, H.; Yang, M.; Wang, Y.; Sun, B. Multiplexed Electrochemical and SERS Dual-Mode Detection of Stroke Biomarkers: Rapid Screening with High Sensitivity. New J. Chem. 2019, 43, 13381–13387. [Google Scholar] [CrossRef]
  110. Tabrizi, M.A.; Ferré-Borrull, J.; Kapruwan, P.; Marsal, L.F. A Photoelectrochemical Sandwich Immunoassay for Protein S100β, a Biomarker for Alzheimer’s Disease, Using an ITO Electrode Modified with a Reduced Graphene Oxide-Gold Conjugate and CdS-Labeled Secondary Antibody. Microchim. Acta 2019, 186, 117. [Google Scholar] [CrossRef] [PubMed]
  111. Harpaz, D.; Koh, B.; Marks, R.S.; Seet, R.C.S.; Abdulhalim, I.; Tok, A.I.Y. Point-of-Care Surface Plasmon Resonance Biosensor for Stroke Biomarkers NT-ProBNP and S100β Using a Functionalized Gold Chip with Specific Antibody. Sensors 2019, 19, 2533. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  112. Han, L.; Ding, C.; Guo, Y.; Wang, Y.; Ding, Y. Sensitively Detecting MTBI Biomarker S100B by Using Peptide-Modified Ratiometric Fluorescent C/AuNCs Nanoprobe. Anal Bioanal Chem 2020, 412, 3695–3702. [Google Scholar] [CrossRef] [PubMed]
  113. Hassanain, W.A.; Sivanesan, A.; Ayoko, G.A.; Izake, E.L. Rapid Electrochemical Nanosensing of S100ß in Blood. J. Electrochem. Soc. 2020, 167, 067518. [Google Scholar] [CrossRef]
  114. Delefortrie, Q.; Lejeune, F.; Kerzmann, B.; Levy, R.; Adam, J.-F.; Sottiaux, T.; Grimmelprez, A.; Vankerkhoven, P.; Hachimi-Idrissi, S. Evaluation of the Roche® Elecsys and the Diasorin® Liaison S100 Kits in the Management of Mild Head Injury in the Emergency Room. Clin. Biochem. 2018, 52, 123–130. [Google Scholar] [CrossRef] [PubMed]
  115. CanAg® S100 EIA|Fujirebio. Available online: https://www.fujirebio.com/en/products-solutions/canag-s100-eia (accessed on 27 August 2021).
  116. Human S100B ELISA, EZHS100B-33K This Human S100B ELISA Kit Is to Be Used for the Quantification of Human S100B in Cerebrospinal Fluid, Serum & Plasma.|Sigma-Aldrich. Available online: http://www.sigmaaldrich.com/ (accessed on 30 August 2021).
  117. S100B ELISA Kit (Ab234573)|Abcam. Available online: https://www.abcam.com/s100b-elisa-kit-ab234573.html (accessed on 30 August 2021).
  118. Human S100B ELISA Kit by Bioss, Cat. No. BSKH1066|Lucerna-Chem AG. Available online: https://lucerna-chem.ch/shop/3438089/human-s100b-elisa-kit (accessed on 30 August 2021).
  119. S100b ELISA Kit (Rabbit) (GWB-KBBVZ1)|Quantitative Sandwich ELISA. Available online: https://www.genwaybio.com/s100b-elisa-kit-rabbit-gwb-kbbvz1 (accessed on 30 August 2021).
Scheme 1. Schematic representation of a strategy proposed for the detection of blood proteins for mild traumatic brain injury (mTBI) diagnostics: (a) Illustration of the human body suffering from an mTBI injury and collection of blood sample potentially containing mTBI-relevant protein biomarkers. (b) Development of ECL-based sandwich immunoassay for each individual mTBI-relevant biomarker (“singleplex”) on a benchtop MesoScale Discovery (MSD) platform in 96-well plate format: (c) Translation of developed ECL detection strategy for each individual mTBI-relevant biomarker (“singleplex”) from benchtop MSD platform into a miniaturized “apropos Point-Of-Care (POC)” ECL platform based on screen-printed carbon electrodes (SPCE) and µSTAT-ECL reader from DropSens Metrohm. (d) Proof-of-concept for a multiplex ECL detection of three different mTBI-relevant biomarkers (i.e., GFAP, h-FABP and S100β) on a single SPCE using a spatially resolved approach and a charge-couple device (CCD) camera as ECL detector. (e) Illustration of the envisioned POC diagnostic instrument in the context of medical emergencies allowing rapid treatment of patients following the accident. Steps (bd) are achieved in the context of the present publication, while the step (e) represents the “ultimate” goal of the proposed methodology being the integration of the developed assays into the POC prototype for decentralized mTBI diagnostics (undeveloped areas, emergency rooms, battlefield, sports facilities, car accident sites, etc.).
Scheme 1. Schematic representation of a strategy proposed for the detection of blood proteins for mild traumatic brain injury (mTBI) diagnostics: (a) Illustration of the human body suffering from an mTBI injury and collection of blood sample potentially containing mTBI-relevant protein biomarkers. (b) Development of ECL-based sandwich immunoassay for each individual mTBI-relevant biomarker (“singleplex”) on a benchtop MesoScale Discovery (MSD) platform in 96-well plate format: (c) Translation of developed ECL detection strategy for each individual mTBI-relevant biomarker (“singleplex”) from benchtop MSD platform into a miniaturized “apropos Point-Of-Care (POC)” ECL platform based on screen-printed carbon electrodes (SPCE) and µSTAT-ECL reader from DropSens Metrohm. (d) Proof-of-concept for a multiplex ECL detection of three different mTBI-relevant biomarkers (i.e., GFAP, h-FABP and S100β) on a single SPCE using a spatially resolved approach and a charge-couple device (CCD) camera as ECL detector. (e) Illustration of the envisioned POC diagnostic instrument in the context of medical emergencies allowing rapid treatment of patients following the accident. Steps (bd) are achieved in the context of the present publication, while the step (e) represents the “ultimate” goal of the proposed methodology being the integration of the developed assays into the POC prototype for decentralized mTBI diagnostics (undeveloped areas, emergency rooms, battlefield, sports facilities, car accident sites, etc.).
Biosensors 12 00172 sch001
Figure 1. ECL sandwich immunoassays for mTBI-relevant biomarkers on the MSD platform in a 96-well plate format: (a) GFAP, (b) h-FABP, and (c) S100β. Left figures—Optimization of antibody concentrations in the sandwich assay: checkerboard optimization assay for the antibody pairs with the x-axis representing the capture antibody concentrations and the y-axis the detection antibody concentrations. The results are visualized as a heat map plot with reported signal intensities or S/B ratios (100 ng mL−1/blank). Right figures—ECL calibration curves established using the conditions listed in Table 2. The error bars represent the standard deviation from three replicates; bars are smaller than the data symbol employed. CO—cutoff value for the biomarker.
Figure 1. ECL sandwich immunoassays for mTBI-relevant biomarkers on the MSD platform in a 96-well plate format: (a) GFAP, (b) h-FABP, and (c) S100β. Left figures—Optimization of antibody concentrations in the sandwich assay: checkerboard optimization assay for the antibody pairs with the x-axis representing the capture antibody concentrations and the y-axis the detection antibody concentrations. The results are visualized as a heat map plot with reported signal intensities or S/B ratios (100 ng mL−1/blank). Right figures—ECL calibration curves established using the conditions listed in Table 2. The error bars represent the standard deviation from three replicates; bars are smaller than the data symbol employed. CO—cutoff value for the biomarker.
Biosensors 12 00172 g001
Figure 2. (a) Representation of ECL images obtained from four different commercial screen-printed electrodes (SPEs) using Vilber Fusion FX6 EDGE (Vilber Smart Imaging) (MSD Free Tag solution 15,000 a.u., 100 µL, LSV 200 mV/s, 0.1–1.8 V); (b) ECL intensity (n = 3) calculated by integration of light spot using ImageJ software, errors bars correspond to STDEV. Electrodes tested: SPCE (DRP-110), SPCE-GNP (DRP-110-GNP), SPCE-CNT-GNP (DRP-110CNT-GNP), and SPCE-QD (DRP-110QD).
Figure 2. (a) Representation of ECL images obtained from four different commercial screen-printed electrodes (SPEs) using Vilber Fusion FX6 EDGE (Vilber Smart Imaging) (MSD Free Tag solution 15,000 a.u., 100 µL, LSV 200 mV/s, 0.1–1.8 V); (b) ECL intensity (n = 3) calculated by integration of light spot using ImageJ software, errors bars correspond to STDEV. Electrodes tested: SPCE (DRP-110), SPCE-GNP (DRP-110-GNP), SPCE-CNT-GNP (DRP-110CNT-GNP), and SPCE-QD (DRP-110QD).
Biosensors 12 00172 g002
Figure 3. (a) Linear sweep voltammogram of SPCE for h-FABP biomarker assay (scan rate 200 mV/s) (black curve) and generated ECL signal (red curve). (b) ECL intensity curves from h-FABP biomarker assay on SPCE generated by chronoamperometry at different potentials: 1.4; 1.5; 1.55 and 1.6 V. Acquisition time 10 s. h-FABP assay conditions in Table 4 below (antigen concentration 100 ng mL−1).
Figure 3. (a) Linear sweep voltammogram of SPCE for h-FABP biomarker assay (scan rate 200 mV/s) (black curve) and generated ECL signal (red curve). (b) ECL intensity curves from h-FABP biomarker assay on SPCE generated by chronoamperometry at different potentials: 1.4; 1.5; 1.55 and 1.6 V. Acquisition time 10 s. h-FABP assay conditions in Table 4 below (antigen concentration 100 ng mL−1).
Biosensors 12 00172 g003
Figure 4. ECL immunoassays for mTBI-relevant biomarkers on SPCE µSTAT-ECL platform: (a) GFAP, (b) h-FABP, and, (c) S100β. Left figures–Optimization of antibody concentrations in the assay: checkerboard optimization assay for the antibody pairs with the x-axis containing the capture antibody concentrations and the y-axis the detection antibody concentrations (capture/detection antibody ratio 10/1). The results are presented as a heat map plot with reported S/B ratios (100 ng mL−1/blank). Right figures—ECL calibration curve established using the conditions listed in Table 4. The error bars represent the standard deviation from three replicates (n = 3); bars are smaller than the data symbol employed.
Figure 4. ECL immunoassays for mTBI-relevant biomarkers on SPCE µSTAT-ECL platform: (a) GFAP, (b) h-FABP, and, (c) S100β. Left figures–Optimization of antibody concentrations in the assay: checkerboard optimization assay for the antibody pairs with the x-axis containing the capture antibody concentrations and the y-axis the detection antibody concentrations (capture/detection antibody ratio 10/1). The results are presented as a heat map plot with reported S/B ratios (100 ng mL−1/blank). Right figures—ECL calibration curve established using the conditions listed in Table 4. The error bars represent the standard deviation from three replicates (n = 3); bars are smaller than the data symbol employed.
Biosensors 12 00172 g004
Figure 5. Multiplex ECL detection of mTBI biomarker panel established on a single SPCE. Top panel figures indicate the electrode patterns that contain alignment spots (“control”) made with BSA protein labelled with SULFO-TAG (white spots); spots for GFAP biomarker (blue spots with letter “G” indicate the position of GFAP capture antibodies); spots for h-FABP biomarker (yellow spots with letter “H” indicate the position of h-FABP capture antibodies); and spots for S100β biomarker (red spots with letter “S” indicate the position of S100β capture antibodies). Dashed letter spots indicate that no antigen was added (blank), while full-colored letter spots indicate that respective antigen was added in the assay. Middle panel figures respectively show real ECL images obtained from the electrode pattern indicated on top panel images, and bottom panel figures show the concept of the signal read-out displayed for the future envisioned mTBI POC diagnostic device: (a) biomarkers not detected in the sample (only three “control” spots are visible); (b) h-FABP biomarker detected; (c) GFAP biomarker detected; (d) S100β biomarker detected; (e) all three biomarkers detected.
Figure 5. Multiplex ECL detection of mTBI biomarker panel established on a single SPCE. Top panel figures indicate the electrode patterns that contain alignment spots (“control”) made with BSA protein labelled with SULFO-TAG (white spots); spots for GFAP biomarker (blue spots with letter “G” indicate the position of GFAP capture antibodies); spots for h-FABP biomarker (yellow spots with letter “H” indicate the position of h-FABP capture antibodies); and spots for S100β biomarker (red spots with letter “S” indicate the position of S100β capture antibodies). Dashed letter spots indicate that no antigen was added (blank), while full-colored letter spots indicate that respective antigen was added in the assay. Middle panel figures respectively show real ECL images obtained from the electrode pattern indicated on top panel images, and bottom panel figures show the concept of the signal read-out displayed for the future envisioned mTBI POC diagnostic device: (a) biomarkers not detected in the sample (only three “control” spots are visible); (b) h-FABP biomarker detected; (c) GFAP biomarker detected; (d) S100β biomarker detected; (e) all three biomarkers detected.
Biosensors 12 00172 g005
Table 1. mTBI protein biomarkers used in this work and their clinically relevant concentration ranges.
Table 1. mTBI protein biomarkers used in this work and their clinically relevant concentration ranges.
BiomarkerPhysiological Concentration 1
AbbreviationFull NameNormalMild TBI
GFAPGlial fibrillary acidic proteinRG: 2–49 pg mL−1 [18]
MD: 4 pg mL−1 [18]
≥33 pg mL−1 [18]
CO: 22 pg mL−1 [27]
h-FABPHeart-fatty acidic binding protein<5.5 ng mL−1 [34]
MN: 3.78 ng mL−1 [35]
CO: 2.62 ng mL−1 (HS/HP) [23]
S100βS100β calcium-binding proteinMD: 50 pg mL−1 (HP) [36]
<0.11 pg mL−1 [37]
≥100 pg mL−1 (HP) [38]
>75 pg mL−1 [39]
CO: 42 pg mL−1 (HS/HP) [23]
1 Physiological concentration is indicated for human serum unless otherwise specified. Values reported in samples other than blood/serum/plasma (e.g., sweat, urine, muscle-on-tissue, etc.) are not considered. Abbreviations: CO—cutoff value; HP—human plasma; HS—human serum; MN—mean value; MD—median value; RG—range.
Table 2. Preliminary conditions evaluated on MSD platform for each individual mTBI biomarker ECL sandwich immunoassay (singleplex).
Table 2. Preliminary conditions evaluated on MSD platform for each individual mTBI biomarker ECL sandwich immunoassay (singleplex).
mTBI
Biomarker
cAb
Diluent
Blocking AgentAssay
Diluent
Wash BufferdAb
Diluent
cAb Conc. (μg mL−1)dAb Conc. (μg mL−1)LOD
(pg mL−1)
Range
(pg mL−1)
GFAPPBS 1×1% BSA; 0.06% Tween-20;
PBS 1×
MQ
water
0.06% Tween-20;
PBS 1×
1% BSA;
PBS 1×
25.002.506.94
[6.9–10.0; CI:95%]
0–10,000
h-FABPPBS 1×1% BSA; 0.06% Tween-20;
PBS 1×
PBS 1×0.06% Tween-20;
PBS 1×
1% BSA;
PBS 1×
5.000.501.35
[0–4.4; CI:95%]
0–10,000
S100β5 mM CaCl2;
50 mM TRIS
2% BSA;
1 mM CaCl2;
PBS 1×
0.1% BSA;
1 mM CaCl2;
50 mM TRIS
0.06% Tween-20;
PBS 1×
0.1% BSA;
1 mM
CaCl2;
50 mM
TRIS
10.005.0015.73
[11.0–20.5; CI:95%]
0–10,000
Table 3. Electrochemical properties of tested commercially available SPEs.
Table 3. Electrochemical properties of tested commercially available SPEs.
SPEProduct NumberA (cm2) *ΔE (mV)Ia/Ic
SPCE DRP-1100.15401501.04
SPCE-GNPDRP-110-GNP0.15801801.00
SPCE-CNT-GNPDRP-110CNT-GNP0.14702851.03
SPCE-QDDRP-110QD0.02702701.00
* Calculated using Equation (2), (average n = 3).
Table 4. Preliminary conditions evaluated on SPCE using µSTAT-ECL platform for each individual mTBI biomarker ECL sandwich immunoassay (singleplex).
Table 4. Preliminary conditions evaluated on SPCE using µSTAT-ECL platform for each individual mTBI biomarker ECL sandwich immunoassay (singleplex).
mTBI
Biomarker
cAb
Diluent
Blocking AgentAssay DiluentWash BufferdAb
Diluent
cAb Concentration
(μg mL−1)
dAb Concentration
(μg mL−1)
LODDynamic Range
GFAPPBS 1×1% BSA; 0.06% Tween-20;
PBS 1×
MQ
water
0.06% Tween-20;
PBS 1×
1% BSA; PBS 1×25.002.500.59
ng mL−1
0–50
ng mL−1
h-FABPPBS 1×1% BSA; 0.06% Tween-20;
PBS 1×
PBS 1×0.06% Tween-20;
PBS 1×
1% BSA; PBS 1×5.000.500.44
ng mL−1
0–50
ng mL−1
S100β5 mM CaCl2;
50 mM TRIS
2% BSA;
1 mM CaCl2;
PBS 1×
1 mM CaCl2;
50 mM TRIS
0.06% Tween-20;
PBS 1×
0.1% BSA;
1 mM
CaCl2;
50 mM
TRIS
25.0012.501.34
ng mL−1
0–50
ng mL−1
Table 5. Results of the recovery test for mTBI biomarkers in human serum (HS, n = 2) samples diluted 2× with respective assay diluent (details of the assay conditions are listed in Table 2).
Table 5. Results of the recovery test for mTBI biomarkers in human serum (HS, n = 2) samples diluted 2× with respective assay diluent (details of the assay conditions are listed in Table 2).
SampleAdded
Biomarker
Spiked
Concentration
(ng mL−1)
Detected
Concentration by MSD
(ng mL−1)
Recovery (%)Detected
Concentration by µSTAT-ECL
(ng mL−1)
Recovery (%)
Human
serum
h-FABP00/0/
2.502.3092%1.9979%
5.004.3787%5.33106%
GFAP00/0/
2.502.2389%2.62105%
5.005.30106%4.3487%
S100β00/0/
2.502.4396%2.1184%
5.006.41128%5.92118%
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Jović, M.; Prim, D.; Saini, E.; Pfeifer, M.E. Towards a Point-of-Care (POC) Diagnostic Platform for the Multiplex Electrochemiluminescent (ECL) Sensing of Mild Traumatic Brain Injury (mTBI) Biomarkers. Biosensors 2022, 12, 172. https://doi.org/10.3390/bios12030172

AMA Style

Jović M, Prim D, Saini E, Pfeifer ME. Towards a Point-of-Care (POC) Diagnostic Platform for the Multiplex Electrochemiluminescent (ECL) Sensing of Mild Traumatic Brain Injury (mTBI) Biomarkers. Biosensors. 2022; 12(3):172. https://doi.org/10.3390/bios12030172

Chicago/Turabian Style

Jović, Milica, Denis Prim, Edis Saini, and Marc Emil Pfeifer. 2022. "Towards a Point-of-Care (POC) Diagnostic Platform for the Multiplex Electrochemiluminescent (ECL) Sensing of Mild Traumatic Brain Injury (mTBI) Biomarkers" Biosensors 12, no. 3: 172. https://doi.org/10.3390/bios12030172

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop