Mechanical and Tribological Characterization of a Dental Ceromer

by
Mariana Santos
1,*,
Ana Sofia Coelho
1,2,3,4,
Anabela Baptista Paula
1,2,3,4,
Carlos Miguel Marto
1,2,3,4,5,
Inês Amaro
1,
José Saraiva
1,
Manuel Marques Ferreira
2,3,4,6,
Pedro Antunes
7,8 and
Eunice Carrilho
1,2,3,4
1
Institute of Integrated Clinical Practice, Faculty of Medicine, University of Coimbra, 3000-075 Coimbra, Portugal
2
Institute for Clinical and Biomedical Research (iCBR), Area of Environment, Genetics and Oncobiology (CIMAGO), Faculty of Medicine, University of Coimbra, 3000-548 Coimbra, Portugal
3
CNC.IBILI Consortium, Faculty of Medicine, University of Coimbra, 3000-548 Coimbra, Portugal
4
Centre for Innovative Biomedicine and Biotechnology (CIBB), University of Coimbra, 3000-548 Coimbra, Portugal
5
Institute of Experimental Pathology, Faculty of Medicine, University of Coimbra, 3000-548 Coimbra, Portugal
6
Institute of Endodontics, Faculty of Medicine, University of Coimbra, 3000-075 Coimbra, Portugal
7
Department of Mechanical Engineering, Centre for Mechanical Engineering, Materials and Processes (CEMMPRE), University of Coimbra, 3030-788 Coimbra, Portugal
8
Itecons—Institute of Research and Technological Development in Construction, Energy, Environment and Sustainability, Rua Pedro Hispano, 3030-289 Coimbra, Portugal
*
Author to whom correspondence should be addressed.
J. Funct. Biomater. 2020, 11(1), 11; https://doi.org/10.3390/jfb11010011
Submission received: 1 October 2019 / Revised: 16 February 2020 / Accepted: 18 February 2020 / Published: 21 February 2020

Abstract

:
Background: Indirect restorations using composites with ceramic fillings can be an alternative to ceramic veneering and direct composite restorations for the treatment of posterior teeth. The aim of this study was the evaluation of the mechanical and tribological properties of a ceromer. Materials and Methods: Sixty specimens were produced and divided into two groups: one control group not submitted to thermocycling (n = 20) and one test group submitted to 5000 cycles of thermocycling (n = 40). The studied parameters were microhardness, surface roughness and the coefficient of friction (scratch test). Results: The ceromer exhibits a reduction of polymerization shrinkage, higher wear, and fracture resistance than the composite resins. The studied ceromer presented good mechanical properties, even after being submitted to thermocycling. Roughness was the property most affected, increasing 25.8%, microhardness decreased by 10.5% and the coefficient of friction increased by 4.2%. Conclusions: In certain situations, ceromers can be an alternative to composite resins and ceramics, providing an aesthetic, conservative and longevity option.

1. Introduction

The continuous evolution of materials and techniques used in dentistry, especially adhesive systems, and restorative materials, aims to provide materials with excellent aesthetic and mechanical properties, capable of restoring function, shape, contour, and color, as well as restoring tooth resistance [1,2]. The application of nanotechnology in the field of biomaterials has driven the development of dental materials, namely composite resins, improving their mechanical and biological properties [3,4].
Both teeth and restorative materials are subject to oral cavity conditions such as humidity, load cycles, temperature variation, bacteria, external agents and contact geometry [5], which makes it challenging to find a material that can withstand the mechanical, thermal and chemical stresses [6] without damaging the structures of the oral cavity while maintaining high longevity. The degradation and erosion of resin composites occur more easily when incomplete polymerizated and/or when the teeth isolation during treatments is not achieved, due to degradation processes caused by oxidation [7]. Material loss due to wear may be the result of physiological factors, such as friction during chewing, pathological factors such as bruxism, and prophylactic factors, such as abrasion which occurs during brushing [8]. Restoration wear is one of the main causes of failure of composite resins in posterior teeth [5].
However, the main disadvantage of direct composite resin rehabilitation is the high polymerization contraction of these materials [9,10], which may be from 2% to up to 5% [2]. Marginal infiltration is the most common clinical consequence in situations where there is decompensation of forces at the tooth-restoration interface, exceeding the adhesion force. When forces related to contraction stress are inferior to the adhesive force, cusp flexion and postoperative sensitivity may occur [2,10,11]. Since polymerization shrinkage remains the main problem of composite resins, the use of these materials should be limited to small cavities to increase their clinical success [10,12].
Ceramic rehabilitation is a treatment that is often offered to patients. This treatment has many advantages, such as the aesthetic factor (enamel-like translucency), the possibility of characterizing both the internal and external surfaces, and the high clinical success rate [11]. However, these materials have a very high modulus of elasticity, which can lead to fractures [13], due to the difficulty in absorbing occlusal forces. This can result in the transfer of stress to the interface of the adhesion material [14]. Furthermore, their high wear resistance can lead to antagonist tooth deterioration [15].
To overcome all the disadvantages of direct composite resins and ceramics, resinous materials with ceramic charge, known as ceromers, appeared about 15 years ago [11]. Ceromers contain inorganic micro-hybrid particles, surrounded by a photopolymerizable organic matrix [16,17]. These materials have a monomer conversion rate in the order of 98% [11]. This factor improves the mechanical properties, particularly the marginal integrity and the color of the restoration-tooth interface [10]. Additionally, its lower modulus of elasticity, more similar to the modulus of elasticity of dentin (when compared to ceramic), allows a better distribution of forces when in function, providing a deformation very similar to that of the natural tooth when subjected to intra-oral stress [17]. Further advantages are now described for the use of ceramic-reinforced resin materials such as high wear and abrasion resistance, good aesthetics, low fracture potential, high flexural strength, reduced antagonist tooth wear, maintenance of marginal integrity and easy intra-oral repair [2,15,16,18,19,20]. Ku et al. [21] reported that the main advantage of using this material is the possibility of fabrication of crowns with enough mechanical strength to resist occlusal forces, maintaining aesthetics and marginal integrity [21].
However, some disadvantages are associated with these materials, such as having a lower hardness than enamel and long-term persistent roughness, which may result in restoration staining, marginal staining, bacterial plaque retention and aesthetic alteration [2,15,16,18,19,20].
This study aimed to evaluate the mechanical and tribological properties of a ceromer, before and after thermocycling. According to the manufacturer, this material contains approximately 80% (by weight) nanoceramic particles bound to the resin matrix. Ceramic particles are composed of three different types of ceramic fillers that reinforce the highly cross-linked polymer matrix. The charges are composed of non-aggregated particles of silica (20 nanometres) and zirconia (4 to 11 nanometres) and aggregate particles of zirconia (4 to 11 nanometres) and silica (20 nanometres).

2. Materials and Methods

Sixty specimens of resinous restorative material with ceramic charge, 3M ESPE Lava Ultimate CAD/CAM Restorative® (3M ESPE, Neuss, Germany), were molded in the following dimensions in millimeters: 4 × 6 × 12 (thickness × width × length). The superficies of the specimens were treated with silicon carbide (SiC) sandpaper with increasing particle size (P1000, P2000), with continuous water flow, for 20 s, using the LaboPol-5 (Struers A/S, Ballerup, Denmark) machine to reduce roughness. Subsequently, the specimens were subjected to an ultrasound bath for 10 min to remove any impurity that might have arisen during polishing and then were numbered and measured with a Mitutoyo Co. digital calliper (Kawasaki, Japan). Each specimen was further weighed on a precision A&D Semi-Micro Analytical Balance GH 202 scale (Tokyo 170-0013, Tokyo, Japan).
The specimens were divided into two groups: one control group (n = 20), not subjected to thermocycling, and one test group (n = 40) submitted to 5000 cycles of thermocycling for 60 s with a temperature variation of 5 to 55 °C. The control group was only used for the scratch test.
The mechanical properties of the material were studied before and after the thermocycling cycles. The Vickers microhardness test was carried out and surface roughness was also evaluated. The tribological sliding indentation test was chosen. The scratch specimens produced by the tribological test were observed in a scanning electronic microscope (SEM).

2.1. Surface Microhardness Analysis

Surface microhardness was determined by the Vickers test method; a Struers Duramin durometer (Struers, Balleruo, Denmark) was used according to standard ASTM E384-10 [22]. The Vickers microhardness value, HV, was determined, after measuring the diagonals left by the indenter in the samples. Due to the low loads of the Vickers hardness measurements, several tests were done with different indentation loads in order to ensure that the indentation appearance wasn’t changed, nor an indentation size effect appeared. Thus, for the selected load the microhardness values were independent of indentation size (Kick’s law [23]).
A 200 g load (1.962 N) was applied due to the viscoelastic nature of the material for 40 s and ten indentations were made on the surface of each block. Measurements of each indentation were made according to the standard test method for micro-indentation material hardness.

2.2. Surface Roughness

Surface roughness was determined using a surface roughness evaluation machine, Mitutoyo Surftest 402 Surface Tester (Mitutoyo Co., Kawasaki, Japan), according to ISO 4288 [24] standard pattern. The analyzed parameters are presented in Table 1.
The arithmetic mean of the evaluation profile (Ra) corresponds to the arithmetic average of the absolute values of the profile heights over the evaluation length, where c is the base length. The maximum height of the profile (Rz) corresponds to the average of the maximum height of the profile peaks, Zp, with the maximum depth of the profile values, Zv, included in the sample length, lr. The Rmax corresponds to the highest-profile roughness value during the measurement path, which is the highest value of Rz. The mean square deviation of the evaluated profile (Rq) corresponds to the root mean square average of the profile heights over the evaluation length, where c is the base length.
The maximum height of the roughness profile (Rt) is defined as the sum of the highest of the peak height of the profile, Zp, with the highest of the valley depths of the profile, Zv, within the evaluation length, ln. The maximum height of the profile peaks (Rp) corresponds to the distance from the maximum peak to the midline.
Five measurements were performed on each specimen, uniformly and at a constant speed, along the surface. The specimens were placed on a metal support, with the same orientation of their length. The extent of surface measurement was from 0.8 to 4 mm, as indicated in Table 2, for a Ra between 0.1 and 2 μm.

2.3. Tribological Characterization

The study of the material wear resistance was performed through the sliding indentation test, which allows the applied load to be varied and, consequently, the frictional force. The characteristics of this test allow unidirectional contact, and the possibility of measuring, with precision, the normal and tangential loads along the test; ability to observe the scars generated during the test allowing to associate to each point the contacting loads, in both normal and tangential directions.
To determine the coefficient of friction, a scratch test was performed using a computed numerically controlled machine (CNC) in order to allow a precise displacement and load control, (Mikron, Agno, Switzerland), according to ASTM G 171-13 norm [25]. This equipment allows precise control of the force applied and the deformation of the specimen. A rounded, 50 μm, conical (60°) tungsten carbide tip, horizontally scribed the blocks at a constant velocity of 0.5 mm/s, with a linearly increasing charge overtime of 0 to 50 N, producing scratches of 5 mm in length. When advancing through the specimen, the indenter creates a compression zone in the material and a traction zone as it travels the predetermined length. The roughness will be the first to come into contact with the indenter, increasing the friction and, consequently, separation of particles from the matrix may occur.
The friction force increases linearly with the normal force applied by the indenter, as exemplified in Figure 1.
The slope of the line represents the coefficient of friction, μ, which is obtained by representing the frictional force as a function of the normal force. The surface observations of the sample were performed with a Philips XL 30 (Eindhoven, The Netherlands). Despite the calculation of the coefficient of friction, the wear surface analysis was only qualitative.

2.4. Statistical Analysis

The results were initially evaluated in a descriptive way, for which adequate statistics were calculated. For the description of the quantitative variables, the mean and the standard deviation were chosen. In addition to the descriptive analysis, we used parametric and non-parametric statistical tests, according to the verification of the assumptions. For the quantitative variables, the normality assumption was verified through the Shapiro–Wilk test. The Wilcoxon–Mann–Whitney test was used to evaluate differences between the means of two groups dependent on quantitative variables. The level of significance adopted throughout the statistical analysis was 5%. The software used in data processing was IBM®SPSS® v.24.0 (IBM Corporation, Armonk, NY, USA). The whole process of statistical analysis was blind.

3. Results

3.1. Weight

No differences were identified between the weight of the specimens evaluated before (0.4467 ± 0.0385 g) and after (0.4480 ± 0.0385 g) 5000 cycles of thermocycling (Z = −1.923; p = 0.054).

3.2. Microhardness

The mean initial and final microhardness values and respective standard deviation for each specimen are shown in Figure 2.
In Figure 3 the initial and final microhardness values (after 5000 cycles of thermocycling) of the evaluated ceromer are shown. There was a 10.5% decrease in the microhardness of the material after 5000 cycles of thermocycling. The average initial microhardness was 107.66 ± 4.84 HV and the mean final microhardness value was 96.34 ± 3.22 (Z = −5.511, p < 0.001).

3.3. Roughness

The roughness results were represented by the different variables considered in the study: Ra, Rz, Rmax, Rq, Rt, and Rp.
Statistically significant differences were found between the initial and final evaluations for the Rz (Z = −2.742; p = 0.005), Rmax (Z = −3.549; p < 0.001), Rt (Z = −3.381; p = 0.001) and Rp (Z = −2.267; p = 0.025). No significant differences were found for the values of Ra (Z = −1.540; p = 0.125) and Rq (Z = −1.806; p = 0.071).
Figure 4 shows the initial and final Ra, Rz, Rmax, Rq, Rt and Rp values (after 5000 thermocycling cycles).
There was a 19.3% increase in the value of Ra. The mean value of the initial Ra was 0.22 ± 0.07 μm and the mean value of the final Ra was 0.26 ± 0.15 μm. There was a 27.7% increase in the value of Rz. The mean value of the initial Rz was 1.61 ± 0.39 μm and the mean value of the final Rz was 2.05 ± 0.61 μm. There was a 32.4% increase in the Rmax value. The mean value of initial the Rmax was 2.08 ± 0.49 μm and the mean value of final the Rmax was 2.75 ± 1.08 μm. There was a 20.3% increase in the value of Rq. The mean value of the initial Rq was 0.29 ± 0.10 μm and the mean value of the final Rq was 0.35 ± 0.18 μm. There was a 33.7% increase in the value of Rt. The average value of the initial Rt was 2.21 ± 0.54 μm and the mean value of the final Rt was 2.95 ± 1.16 μm. There was a 21.4% increase in the value of Rp. The mean value of the initial Rp was 0.69 ± 0.17 μm and the mean value of the final Rp was 0.84 ± 0.39 μm.
Figure 5 shows a comparison of the arithmetic averages of roughness parameters analyzed. Overall, there was an increase in roughness of 25.8%.

3.4. Sliding Indentation Test

The results related to the friction coefficient of the studied material, determined by the Scratch Test, are shown in Table 3.
Figure 6 shows the relationship between the normal force and the frictional force, with its coefficient of friction. Between the control group and the test group, there was an increase of 4.2%.
Figure 7 is the graphical representation of the coefficient of friction comparison and the frictional force between the control group and the test group. There was a greater oscillation in the friction force of the control group than in the test group.
The presented analysis concerning the evolution of friction force and the comparison of the coefficient of friction is corroborated by the images obtained from both optical and electronic microscopes. In Figure 8, wear marks are shown on a 200 μm scale.
Figure 9 shows the wear scars that the indenter left on specimens from both groups. The wear marks represented correlate with the graphs of the friction force (Figure 10). The control group had larger and more irregular fractures. For the same initial indenter displacement, the surface of the specimen showed greater damage in the control group than in the test group. The control group had larger and more irregular fractures.
Figure 10 shows the wear scars along with the friction force and normal force plot that the indenter left on both specimens. Relating these graphics to the wear scar produced by each indenter on the antagonist allows us to establish a relationship between the observed morphology features and the values of the loads that occurred in those specific places. From the analysis of the previous graphs assembled along with the SEM pictures scar surface of the ceromer, it was possible to find the exact location of the fractures produced by the increasing loading of the indenter. This gives the perception of the severity of the damage. Figure 11 shows the various details of the scratch test for control and test groups.

4. Discussion

In this experimental study, several mechanical parameters of the 3M ESPE Lava Ultimate CAD/CAM Restorative® ceramics (3M ESPE, Neuss, Germany) were tested. The initial microhardness of the material was 107 Vickers (1049 MPa), which after 5000 cycles of thermocycling reduced for 96 Vickers (941.5 MPa). These values are closer to the values of microhardness of dentin, which is around 1000 MPa, than the microhardness of enamel, which is around 4000 MPa [26]. The microhardness values of the non-reinforced composite resins are around 90 Vickers (883 MPa) [27]. Reinforcement with ceramic particles positively influences the mechanical properties of the composite resins and, even after aging, these properties are better than the properties of the non-reinforced composite resins.
The degradation that takes place in the restorations, whether these are composite resins or composite resins reinforced with ceramic particles, is because, when in contact with water (and, in the case of the oral cavity, with saliva), the diffusion of water molecules occurs in the polymer matrix, which degrades the siloxane bonds, compromising the physical and mechanical properties of the material. This produces a decrease in the hardness and modulus of elasticity of the restorations over time [6,28].
Roughness has been indicated as the main disadvantage resulting from the use of ceromers, as this increases over time (13,18). In the present study, the roughness parameters that varied significantly after 5000 thermocycling cycles were Rz, Rmax, Rt, and Rp. The maximum roughness parameter (Rmax) provides data on the deterioration of the vertical surface of the specimens. Thus, its increase is associated with an increase in peaks and roughness on the surface of the restoring material [6]. The total roughness depth parameter (Rt) measures the distance between the highest peak and the deepest valley, within the total evaluation length, which also supports a significant increase in Rmax. The mean square root of the height of profiles (Rq) and the arithmetic mean of the absolute values of the profile heights (Ra) do not define the shape of the irregularities [29], which may explain the fact that there were no statistically significant differences in these parameters. In general, in the present study, the roughness of the specimens increased by 25.8% after the aging of the specimens; this corroborates the results of previous studies, which refer to the significant increase in the surface roughness of the ceromers [10,11,13,18]. However, the present study would benefit from an increase in the number of specimens, since the standard deviation of the initial measurements intersected the arithmetic means of the final measurements.
The coefficient of friction increased after thermocycling, and a variation of the frictional force between the test group and the control group was identified. This has been observed in materials that have hard reinforcing particles [6] and therefore resist the advancement of the indenter, causing a decrease in their roughness. The larger the size of the reinforcing particles, the higher the coefficient of friction and consequently the higher the contact forces, which can increase the volume of material removed by wear [6].
The damages that were presented in the direction transversal to the displacement occurred via subsurface fractures, due to the increase of the coefficient of friction (or the friction force), when reaching high local values (state of higher tension), taking into account that the voltage distribution obeys the Hertz theory [30] and that the tangential component of the applied load increases the local voltage in the subsurface. Thus, with the increase of the frictional force and consequently of the stress in the subsurface of the material, this voltage distribution induced surface faults below the contact area through a fatigue process and, consequently, the propagation of this failure to the surface occurred. This explains the larger fractures in the test group compared to the control group. However, the matrix of this material is resinous and, in addition to the behavior of the reinforcing material, it is also important to consider the possible wear that the resin matrix may suffer, compromising its behavior.
Increased roughness appears to indicate that thermocycling affects the matrix, not the particles. Surface roughness influence on the wear behavior can be explained by the materials strength which is determined by its surface roughness and the inner structure of the material. An increase in the surface roughness may cause a larger stress concentration than the one produced by the surface roughness in combination with the surface flaws [31]. If a material has low roughness, when in contact with other materials is less prone for cracks to appear. This is possible to observe (Figure 9), that the initial scar path on the thermocycled specimens, the ones with higher values of roughness parameters, present a larger wear scar then the control material group, for the same test conditions. The smother the material surface more difficult is to the cracks to appear, and therefore if the material remains smoother the restoration will be long lasting. The matrix can both retract with respect to the particles and become softer [5], which justifies the greater resistance of the test group to scratches, with more subsurface fractures, which were expected to occur in the zones of higher wear, and oscillations of the force of friction. The more premature appearance of subsurface fractures and their greater extend indicates that the material in the control group is more fragile than the test group, which was also confirmed by microhardness analysis. For the control group, which had a higher reinforcement load and lower toughness, the indenter encountered higher resistance, changing the friction force in relation to the normal force. In the test group, as toughness increased, the slopes of the friction force and normal force were similar.

5. Conclusions

Ceromers can be an aesthetic alternative to restorations of teeth with great destruction, showing clinical longevity, with a low fracture rate and low incidence of caries recurrence. The fracture limit of ceromers is superior to the occlusal forces exerted, which, together with t its elasticity modulus similar to dentin. These results support the possibility that these materials can be used in patients with parafunctional habits, however, future clinical studies must be performed to determine the clinical performance of these materials. In areas where aesthetics is fundamental (anterior sector), ceramics may be a more viable option.
The studied ceromer, Lava Ultimate CAD/CAM Restorative® ceramics, proved to be a good material for complex restorations since it maintains excellent mechanical properties even after aging.

Author Contributions

Data curation, M.S., A.S.C., C.M.M., I.A., J.S., P.A. and E.C.; formal analysis, M.S., A.S.C., A.B.P. and P.A.; funding acquisition, M.M.F.; investigation, M.S., A.S.C., C.M.M. and P.A.; methodology, M.S., A.B.P., J.S. and P.A.; project administration, I.A., M.M.F., P.A. and E.C.; resources, A.S.C., A.B.P., M.M.F., P.A. and E.C.; software, A.S.C., C.M.M., I.A. and J.S.; supervision, A.S.C., C.M.M., M.M.F. and E.C.; validation, A.S.C. and A.B.P.; visualization, M.S., J.S. and E.C.; writing—original draft, M.S. and A.B.P.; writing—review and editing, A.S.C., C.M.M., I.A., M.M.F., P.A. and E.C. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Terry, D.A.; Leinfelder, K.E.; Maragos, C. Developing an intracoronal restoration with indirect composite resin. Part 2. Dent. Today 2005, 24, 84–89. [Google Scholar] [PubMed]
  2. Bouschlicher, M.R.; Cobb, D.S.; Vargas, M.A. Effect of two abrasive systems on resin bonding to laboratory-processed indirect resin composite restorations. J. Esthet. Dent. 1999, 11, 185–196. [Google Scholar] [CrossRef] [PubMed]
  3. Zafar, M.S.; Khurshid, Z.; Najeeb, S.; Zohaib, S.; Rehman, I.U. Therapeutic Applications of Nanotechnology in Dentistry. In Nanostructures for Oral Medicine; Elsevier: Amsterdam, The Netherlands, 2017; pp. 833–862. [Google Scholar]
  4. Khurshid, Z.; Zafar, M.; Qasim, S.; Shahab, S.; Naseem, M.; AbuReqaiba, A. Advances in Nanotechnology for Restorative Dentistry. Materials 2015, 8, 717–731. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  5. Antunes, P.A.A.V. Physical-Mechanical Characterization of Direct Restorative Dental Materials. Ph.D. Thesis, University of Coimbra, Coimbra, Portugal, 2012. [Google Scholar]
  6. Carreira, M.; Antunes, P.V.; Ramalho, A.; Paula, A.; Carrilho, E. Thermocycling effect on mechanical and tribological characterization of two indirect dental restorative materials. J. Braz. Soc. Mech. Sci. Eng. 2017, 39, 1–17. [Google Scholar] [CrossRef]
  7. Ortengren, U.; Wellendorf, H.; Karlsson, S.; Ruyter, I.E. Water sorption and solubility of dental composites and identification of monomers released in an aqueous environment. J. Oral Rehabil. 2001, 28, 1106–1115. [Google Scholar] [CrossRef] [PubMed]
  8. Heintze, S.D.; Zellweger, G.; Zappini, G. The relationship between physical parameters and wear of dental composites. Wear 2007, 263, 1138–1146. [Google Scholar] [CrossRef]
  9. Ortega, V.L.; Pegoraro, L.F.; Conti, P.C.R.; Valle, A.L.; Bonfante, G. Evaluation of fracture resistance of endodontically treated maxillary premolars, restored with ceromer or heat-pressed ceramic inlays and fixed with dual-resin cements. J. Oral Rehabil. 2004, 31, 393–397. [Google Scholar] [CrossRef]
  10. Monaco, C.; Baldissara, P.; Dall’Orologio, G.D.; Scotti, R. Short-term clinical evaluation of inlay and onlay restorations made with a ceromer. Int. J. Prosthodont. 2001, 14, 81–86. [Google Scholar]
  11. Khairallah, C.; Sabbagh, J.; Hokayem, A. Clinical study comparing at 5 years a ceramic and a ceromer used for making esthetic inlays. Odontostomatol. Trop. 2009, 32, 21–28. [Google Scholar]
  12. GEMALMAZ, D.; KUKRER, D. In vivo and in vitro evaluation of marginal fit of class II ceromer inlays. J. Oral Rehabil. 2006, 33, 436–442. [Google Scholar] [CrossRef]
  13. Kükrer, D.; Gemalmaz, D.; Kuybulu, E.O.; Bozkurt, F.O. A prospective clinical study of ceromer inlays: Results up to 53 months. Int. J. Prosthodont. 2004, 17, 17–23. [Google Scholar] [PubMed]
  14. Cesar, P.F.; Miranda, W.G.; Braga, R.R. Influence of shade and storage time on the flexural strength, flexural modulus, and hardness of composites used for indirect restorations. J. Prosthet. Dent. 2001, 86, 289–296. [Google Scholar] [CrossRef]
  15. Cho, L.; Song, H.; Koak, J.; Heo, S. Marginal accuracy and fracture strength of ceromer/fiber-reinforced composite crowns: Effect of variations in preparation design. J. Prosthet. Dent. 2002, 88, 388–395. [Google Scholar] [CrossRef] [PubMed]
  16. Cho, L.; Choi, J.; Jin Yi, Y.; Jin Park, C. Effect of finish line variants on marginal accuracy and fracture strength of ceramic optimized polymer/fiber-reinforced composite crowns. J. Prosthet. Dent. 2004, 91, 554–560. [Google Scholar] [CrossRef] [PubMed]
  17. Göehring, T.N.; Peters, O.A.; Lutz, F. Marginal adaptation of inlay-retained adhesive fixed partial dentures after mechanical and thermal stress: An in vitro study. J. Prosthet. Dent. 2001, 86, 81–92. [Google Scholar] [CrossRef] [PubMed]
  18. CHO, L.-R.; YI, Y.-J.; HEO, S.-J. Effect of tooth brushing and thermal cycling on a surface change of ceromers finished with different methods. J. Oral Rehabil. 2002, 29, 816–822. [Google Scholar] [CrossRef]
  19. Iglesia-Puig, M.A.; Arellano-Cabornero, A. Inlay fixed partial denture as a conservative approach for restoring posterior missing teeth: A clinical report. J. Prosthet. Dent. 2003, 89, 443–445. [Google Scholar] [CrossRef]
  20. Stappert, C.F.J.; Chitmongkolsuk, S.; Silva, N.R.F.A.; Att, W.; Strub, J.R. Effect of mouth-motion fatigue and thermal cycling on the marginal accuracy of partial coverage restorations made of various dental materials. Dent. Mater. 2008, 24, 1248–1257. [Google Scholar] [CrossRef]
  21. Ku, C.-W.; Park, S.-W.; Yang, H.-S. Comparison of the fracture strengths of metal-ceramic crowns and three ceromer crowns. J. Prosthet. Dent. 2002, 88, 170–175. [Google Scholar] [CrossRef]
  22. Standard, A.S.T.M. E384 Standard Test Method for Microindentation Hardness of Materials. West Conshohocken PA ASTM Int. 2000, 1–40. [Google Scholar]
  23. Cai, X.; Yang, X.; Zhou, P. Dependence of Vickers microhardness on applied load in indium. J. Mater. Sci. Lett. 1997, 16, 741–742. [Google Scholar] [CrossRef]
  24. EN ISO-4287. Geometrical Product Specifications (GPS)—Surface Texture: Profile method—Terms, Definitions and Surface Texture Parameters; International Organization for Standardization (ISO): Geneva, Switzerland, 1997; pp. 1–25. [Google Scholar]
  25. ASTM, A. G171 Standard test method for scratch hardness of materials using a diamond stylus. ASTM Stand 2009, 3, 1–7. [Google Scholar]
  26. Zaytsev, D.; Panfilov, P. Deformation behavior of human enamel and dentin–enamel junction under compression. Mater. Sci. Eng. C 2014, 34, 15–21. [Google Scholar] [CrossRef] [PubMed]
  27. Moharam, L.; El-Hoshy, A.; Abou-Elenein, K. The effect of different insertion techniques on the depth of cure and vickers surface micro-hardness of two bulk-fill resin composite materials. J. Clin. Exp. Dent. 2017, 9, e266. [Google Scholar] [CrossRef]
  28. Gul, E.B.; Atala, M.H.; Eser, B.; Polat, N.T.; Asilturk, M.; Gultek, A. Effects of coating with different ceromers on the impact strength, transverse strength and elastic modulus of polymethyl methacrylate. Dent. Mater. J. 2015, 34, 379–387. [Google Scholar] [CrossRef] [Green Version]
  29. Neto, J. Metrologia e Controle Dimensional, 1st ed.; Elsevier: Amsterdam, The Netherlands, 2012. [Google Scholar]
  30. Peixoto, D.F.C. Estudo do Contacto Roda/Carril: Análise de Tensões e Fadiga; Universidade do Porto: Porto, Portugal, 2008. [Google Scholar]
  31. Rashid, H. The effect of surface roughness on ceramics used in dentistry: A review of literature. Eur. J. Dent. 2014, 8, 571–579. [Google Scholar] [CrossRef]
Figure 1. Friction and normal forces due to the displacement.
Figure 1. Friction and normal forces due to the displacement.
Jfb 11 00011 g001
Figure 2. Average initial and final microhardness values and respective standard deviations of the specimens.
Figure 2. Average initial and final microhardness values and respective standard deviations of the specimens.
Jfb 11 00011 g002
Figure 3. Comparison of the arithmetic means of the group before and after thermocycling.
Figure 3. Comparison of the arithmetic means of the group before and after thermocycling.
Jfb 11 00011 g003
Figure 4. Comparison of the arithmetic means of the roughness (initial and final) for Ra (a), Rz (b), Rmax (c), Rq (d), Rt (e), Rp (f).
Figure 4. Comparison of the arithmetic means of the roughness (initial and final) for Ra (a), Rz (b), Rmax (c), Rq (d), Rt (e), Rp (f).
Jfb 11 00011 g004
Figure 5. Comparison of arithmetic means for all roughness parameters.
Figure 5. Comparison of arithmetic means for all roughness parameters.
Jfb 11 00011 g005
Figure 6. Relation between the normal force and the friction force of the control group (a) and test group (b), with the coefficient of friction of 0.72 and 0.75 respectively.
Figure 6. Relation between the normal force and the friction force of the control group (a) and test group (b), with the coefficient of friction of 0.72 and 0.75 respectively.
Jfb 11 00011 g006
Figure 7. Comparison between the normal force and the friction force of control and test groups (a). Comparison of the friction force between the control and test groups (b).
Figure 7. Comparison between the normal force and the friction force of control and test groups (a). Comparison of the friction force between the control and test groups (b).
Jfb 11 00011 g007
Figure 8. Wear observation on non-polarized lens. (A) Control group; (B) Test group, and SEM imagens for, (C) Control group; (D) Test group.
Figure 8. Wear observation on non-polarized lens. (A) Control group; (B) Test group, and SEM imagens for, (C) Control group; (D) Test group.
Jfb 11 00011 g008
Figure 9. Observation of wear suffered by a specimen of the control group (1) and test group (2): (A) Non-polarized lens; (B) Polarized lens.
Figure 9. Observation of wear suffered by a specimen of the control group (1) and test group (2): (A) Non-polarized lens; (B) Polarized lens.
Jfb 11 00011 g009
Figure 10. Wear observation obtained for control and test groups compared to normal force and frictional force. (A) Control group; (B) Test group.
Figure 10. Wear observation obtained for control and test groups compared to normal force and frictional force. (A) Control group; (B) Test group.
Jfb 11 00011 g010
Figure 11. Detailed wear images observation obtained for: (A) Control group (500× top image and 2000× bottom image); (B) Test group (500× top image and 2000× bottom image).
Figure 11. Detailed wear images observation obtained for: (A) Control group (500× top image and 2000× bottom image); (B) Test group (500× top image and 2000× bottom image).
Jfb 11 00011 g011
Table 1. Roughness parameters analyzed.
Table 1. Roughness parameters analyzed.
Roughness Parameters
RaRoughness Average
RzAverage Maximum Height of the Profile
RmaxMaximum Roughness Depth
RqRMS Roughness
RtMaximum Height of the Profile
RpMaximum Profile Peak Height
Table 2. Tabulated values according to standard pattern EM ISO 4288 for sampling length selection.
Table 2. Tabulated values according to standard pattern EM ISO 4288 for sampling length selection.
Periodic ProfilesNon-Periodic ProfilesCut-offLength of Sample Evaluation
Rsm (mm)Rz (µm)Ra (µm)λc (mm)lr/ln (mm)
>0.013 and <0.04<0.1<0.020.080.08/0.4
>0.04 and <0.13>0.1 and <0.5>0.02 and <0.10.250.25/1.25
>0.13 and <0.4>0.5 and <10>0.1 and <20.80.8/4
>0.4 and <1.3>10 and <50>2 and <102.52.5/12.5
>1.3 and <4>50 and <200>10 and <8088/40
Table 3. Average value of the coefficient of friction.
Table 3. Average value of the coefficient of friction.
Control GroupTest Group
µ0.720.75

Share and Cite

MDPI and ACS Style

Santos, M.; Coelho, A.S.; Paula, A.B.; Marto, C.M.; Amaro, I.; Saraiva, J.; Marques Ferreira, M.; Antunes, P.; Carrilho, E. Mechanical and Tribological Characterization of a Dental Ceromer. J. Funct. Biomater. 2020, 11, 11. https://doi.org/10.3390/jfb11010011

AMA Style

Santos M, Coelho AS, Paula AB, Marto CM, Amaro I, Saraiva J, Marques Ferreira M, Antunes P, Carrilho E. Mechanical and Tribological Characterization of a Dental Ceromer. Journal of Functional Biomaterials. 2020; 11(1):11. https://doi.org/10.3390/jfb11010011

Chicago/Turabian Style

Santos, Mariana, Ana Sofia Coelho, Anabela Baptista Paula, Carlos Miguel Marto, Inês Amaro, José Saraiva, Manuel Marques Ferreira, Pedro Antunes, and Eunice Carrilho. 2020. "Mechanical and Tribological Characterization of a Dental Ceromer" Journal of Functional Biomaterials 11, no. 1: 11. https://doi.org/10.3390/jfb11010011

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop