Article

Virtual Non-Contrast Spectral CT in Renal Masses: Is It Time to Discard Conventional Unenhanced Phase?

1
Department of Biomedical Sciences and Morphological and Functional Imaging, University of Messina, 98122 Messina, Italy
2
Postgraduation School in Radiodiagnostics, Università degli Studi di Milano, Via Festa del Perdono 7, 20122 Milan, Italy
3
Diagnostic and Interventional Radiology Unit, ASST Settelaghi, Insubria University, 21100 Varese, Italy
4
Division of Experimental Imaging, Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, 60590 Frankfurt am Main, Germany
5
Department of Radiology and Nuclear Medicine, Erasmus MC, 3015 GD Rotterdam, The Netherlands
*
Author to whom correspondence should be addressed.
J. Clin. Med. 2023, 12(14), 4718; https://doi.org/10.3390/jcm12144718
Submission received: 29 May 2023 / Revised: 8 July 2023 / Accepted: 14 July 2023 / Published: 17 July 2023

Abstract

:
Dual-layer Dual-Energy CT (dl-DECT) allows one to create virtual non-contrast (VNC) reconstructions from contrast-enhanced CT scans, with a consequent decrease of the radiation dose. This study aims to assess the reliability of VNC for the diagnostic evaluation of renal masses in comparison with true non-contrast (TNC) images. The study cohort included 100 renal masses in 40 patients who underwent dl-DECT between June and December 2021. Attenuation values and standard deviations were assessed through the drawing of regions of interest on TNC and VNC images reconstructed from corticomedullary and nephrographic phases. A Wilcoxon signed-rank test was performed in order to assess equivalence of data and Spearman’s Rho correlation coefficient to evaluate correlations between each parameter. The diagnostic accuracy of VNC was estimated through the performance of receiver operating characteristic (ROC) curve analysis. Differences between attenuation values were, respectively, 74%, 18%, 5% and 3% (TNC-VNCcort), and 74%, 15%, 9% and 2% (TNC-VNCneph). The Wilcoxon signed-rank test demonstrated the equivalence of attenuation values between the TNC and VNC images. The diagnostic performance of VNC images in the depiction of kidney simple cysts remains high compared to TNC (VNCcort-AUC: 0.896; VNCneph-AUC: 0.901, TNC-AUC: 0.903). In conclusion, quantitative analysis of attenuation values showed a strong agreement between VNC and TNC images in the evaluation of renal masses.

1. Introduction

As is well known, the unquestionable advantages of CT-scan, such as a high spatial and contrast resolution as well as rapid scan times, are counterbalanced by potential health risks related to radiation exposure [1,2,3,4].
In order to overcome this critical issue, over the last decades, specific reconstruction algorithms, for instance iterative or deep-learning-based ones, have been developed [5,6,7,8].
However, it should be kept in mind that radiation dose can be easily decreased by just avoiding unnecessary scans when multiphase CT protocols are used.
Dual-energy computed tomography (DECT) is an already established imaging technology that exploits images’ raw data obtained at different energy levels in order to achieve material decomposition [9,10].
This concept includes the differentiation of body-tissue components according to the atomic number, with the possibility of highlighting or minimizing their effect on the final image using dedicated post-processing algorithms.
Among these, virtual non-contrast (VNC) aims at removing iodine from contrast-enhanced images [11]. Since iodine is the main element of intravenous contrast medium, the result is an unenhanced set of images obtained without the need for a real acquisition. The consequent advantages in terms of the spareness of radiation dose have been made evident.
In the abdominal field, the use of VNC has shown encouraging results, particularly in the characterization of hepatic and pancreatic lesions [12,13].
Conventional non-contrast imaging plays a crucial role in the comprehensive evaluation and characterization of renal masses, providing accurate identification of common lesions with confidence. It provides essential information for classifying cystic kidney masses based on their attenuation values; for instance, a homogeneous mass measuring 70 HU or higher at the non-contrast phase indicates a high probability of representing a high-attenuation renal cyst [14]. Additionally, unenhanced CT is highly effective in detecting calcification or macroscopic fat tissue, contributing to accurate diagnoses. In fact, the presence of macroscopic fat within a solid renal mass allows for differentiation between classical angiomyolipomas and other lesions, such as renal cell carcinoma [15].
Thus, the purpose of this study is to evaluate the reliability of attenuation values of renal masses in VNC images, compared to true non-contrast (TNC) images, using a dual-layer (dl) DECT scanner.

2. Materials and Methods

2.1. Study Population

A retrospective single-center study was conducted.
All contrast-enhanced abdominal dl-DECT scans performed with a multiphasic scan protocol in patients with a clinical suspicion of renal masses between June 2021 and December 2021, present on our Picture archiving and communication system (PACS), were reviewed. No preselection regarding patients’ weight, age, sex, or other characteristics was made.
Only patients with renal lesions larger than 5 mm at a maximum diameter have been included in the final patient cohort (Figure 1).

2.2. DECT Scan Protocol

All scans were obtained using a clinical dl-DECT scanner (IQon, Philips Healthcare, Best, the Netherlands). Patients were examined in supine position with a field of scan ranging from the diaphragm to the pubic symphysis. Image acquisition was performed in the craniocaudal direction during inspiratory breath hold.
The study protocol consisted of unenhanced, cortico-medullary and nephrographic phases.
A bolus tracking ROI was positioned on the descending aorta, and the cortico-medullary phase started 15 s after reaching the 100 HU trigger threshold. Nephrograpic phase scans was subsequently acquired at a time delay of 70–80 s after bolus tracking. The settings used for dl-DECT imaging were: tube voltage 120 kVp, tube current modulation activated (DoseRight 3D-DOM, Philips Healthcare), gantry rotation time 0.33 s, pitch 0.9, collimation 64 × 0.625 mm, rotation time 0.5, and volume CT dose index (CTDIvol) of 2.9 milligrays (mGy).
The intravenous contrast agent (Iomeron 400 mgI/mL, Bracco) was injected at a dose of 1.2 mL per kilogram body weight and at a flow rate of 3–4 mL/s and followed by a bolus of 40 mL of saline. No oral contrast material was administered.
Technical parameters are summarized in Table 1.

2.3. Image Reconstruction

Axial conventional images with a 2.00 mm slice thickness were reconstructed from contributions of both detector layers using the proprietary iDose4 reconstruction algorithm.
Spectral database images (SBI) were automatically generated in order to obtain post-processing reconstructions through a dedicated dl-DECT WorkStation (IntelliSpace Portal (vv. 8.0.2), Philips Healthcare, Best, Netherlands).
SBI data were exploited to create VNC image series starting from the cortico-medullary and nephrographic phases (Figure 2).

2.4. Objective Image Evaluation

Two radiologists, with 3 and 5 years of experience, respectively, in abdominal radiology, evaluated the renal masses, drawing a circular region of interest (ROI within kidney lesions, covering the largest area at the level of the maximum diameter). The ROIs’ size areas ranged between 17.41–4318.32 mm2 and were kept constant across the different image series of the same patient.
Attenuation values measured on TNC images served as the reference standard.
Mean values of measurements were used for the statistical analysis.

2.5. Statistical Analysis

Statistical analysis was performed using dedicated software (R version 4.2.0, R Foundation for Statistical Computing, Vienna, Austria; Matlab, Matworks v. R2022a, Natick, MA, USA; Excel 365, Microsoft, Redmond, WA, USA).
The non-normality distribution of data (CT values and SD) was evaluated using both Shapiro–Wilk and Kolmogorov–Smirnov tests. A Wilcoxon signed-rank test was performed using a TOST (two one-sided test) equivalence test to reject the null hypothesis. The null hypothesis argued that the difference between the median of the attenuation values in TNCs and VNCs was greater than 3/5/10 HU. The statistical test was conducted while considering overall lesions and then dividing lesions into three groups (renal masses and cystic lesions, the latter further classified as simple or complex).
The statistically significant difference was indicated by a p value under 0.05.
Moreover, Spearman’s Rank correlation coefficient was assessed between CT values and standard deviations (HU) measured on TNC and VNC of the corticomedullary phase (VNCcort) and those on TNC and VNC of the nephrographic phase (VNCneph).
A Bland–Altmann plot was drawn to assess the agreement of VNC attenuation values in comparison to the standard conventional imaging attenuation values obtained from TNC [16].
The difference between attenuation values on TNC and VNCcort images was calculated while evaluating how many times it was under 5 HU, between 5–10 HU and 10–20 HU, or over 20 HU. The same was performed between TNC and VNCneph.
The receiver operating characteristic curves (ROC) and the associated area under the curve (AUC) with a 95% confidence interval were assessed to compare the sensitivity and sensibility of VNC images and unenhanced images. The difference between the curves was evaluated by using the DeLong test for correlated curves.

3. Results

3.1. Population Characteristics

From a total of 66 patients, the following patients were excluded: 23 (33%) because of the lack of SBI data storage on PACS, 2 (3%) because the lesion’s maximum diameter was smaller than 5 mm, and 1 (1%) due to absence of renal masses.
The final study cohort consisted of 40 patients (30 males, mean age: 68.19 years; 36–87 years; 10 females, mean age: 61.60, range 35–81).
A total of 100 masses were included in the study, consisting of 30 renal solid masses, 13 complex cysts, and 57 simple cysts.

3.2. Quantitative Assessment

The median values of data obtained from TNC, VNCcort, and VNCneph are 10.65 HU, 9.70 HU, and 8.05 HU, respectively. The maximum attenuation value is 75.10 HU in TNC images, compared with 72.20 and 69.40 HU on VNCcort and VNCneph, respectively; meanwhile, the minimum attenuation value is −40.00 HU on TNC series, −26.40 HU on VNCcort series, and −23.4 HU on VNCneph series (Figure 3).
The means of all CT values of renal masses obtained from TNC and VNC reconstructions are summarized in Table 2.
Overall, the difference in attenuation values on TNC and VNCcort images was lower than 5 HU in 74% of cases, between 5–10 HU in 18% of cases, between 10–20 HU in 5% of cases, and greater than 20 HU in 3% of cases.
The attenuation values obtained by the TNC and VNCneph differ by 5 HU in 74% of cases, while the difference is between 5–10 HU, between 10–20 HU, and greater than 20 HU, respectively, in 15%, 9%, and 2% of cases (Figure 4A–D).
Considering each group individually (kidney masses, simple and complex cysts), the results obtained are similar to those obtained when evaluating all the lesions.
Spearman’s Rank correlation coefficient (Spearman rho) between TNC and VNCcort was 0.87, while it was 0.89 between TNC and VNCneph.
The results of the Wilcoxon signed-rank test based on TOST rejected the null hypothesis, indicating that the differences between all the median values are between the ranges of −3, +3 HU, −5, +5 HU, and −10, +10 HU, so that the attenuation values of TNC and VNC are equivalent (Table 3). The rejection of the null hypothesis has confirmed that the attenuation values obtained from the TNC and VNC image series were equivalent to the threshold assessed.
This is valid in evaluating overall lesions and also each group (kidney masses, simple and complex cysts). Only in the interval −3; +3 does the test not show statistical significance in the depiction of cystic lesions (p > 0.05) (Table 3).
The Bland−Altman plot (Figure 5) provides the difference between the attenuation values on single TNC and VNC measurements and their mean values in order to identify the agreement between the measured values in TNC and VNC.
The ROC curves evaluated for simple cystic lesions show a mild difference in diagnostic performance between TNC (AUC: 0.903), considered as a standard reference, VNCcort (AUC: 0.896), and VNCneph (AUC: 0.901). Moreover, the comparison of ROC curves does not show a statistical difference between TNC, VNC corticomedullary, and nephrographic reconstructions (p: 0.817 and p: 0.944, respectively), confirming the interchangeability of the true and virtual image series (Figure 6).

4. Discussion

CT scan plays a key role in the evaluation of renal masses [17]. The main purposes of their characterization are represented by the differentiation of cystic from solid lesions and, within those groups, the distinction of benign from malignant [18].
However, the increasingly widespread use of CT scan over the last decades has raised the issue of radiation dose delivery to the patient, leading radiologists to individualize each examination to the patient in order to minimize potentially harmful effects [19].
In this context, the exploitation of the dl-DECT potential through VNC reconstructions may avoid the performance of a true unenhanced scan and therefore reduce the final radiation exposure.
In this study, the reliability of attenuation values within renal masses obtained on VNC images reconstructed from cortico-medullary and nephrographic phases was assessed and compared to those of TNC.
Our results support the interchangeability of TNC with VNC series. The differences in attenuation values obtained from the VNCcort and TNC series are less than 10 HU in 92% of cases, and only 3% of cases show more than 20 HU. VNC obtained from the nephrographic phase shows attenuation values under 10 HU in 89% of cases but with only 2% of cases having more than 20 HU.
Moreover, the comparison of TNC and VNC attenuation values with the TOST Wilcoxon signed-rank test does not demonstrate a significant difference between the variables (p < 0.05), supporting the fact that TNC and VNC images are closely related.
ROC curves showed a high diagnostic accuracy for the detection of kidney simple cysts on the VNC series (VNCcort-AUC: 0.896; VNCneph-AUC: 0.901; TNC images AUC: 0.903) (Figure 6).
In the evaluation of renal cell carcinoma recurrence, the differences in attenuation values and image noise obtained between TNC and VNC reconstructions are essentially equivalent, demonstrating a difference lower than 5 HU (Figure 7).
Even for complex cystic lesions (Bosniak II), we found a strong agreement between attenuation values and a slightly lower noise level (Figure 8).
On the contrary, in the evaluation of a native CT hyperdense cystic lesion (Bosniak II), we obtained discordant attenuation values in VNC images, which would have led to an incorrect diagnosis and unnecessary further diagnostic investigations (Figure 9).
Up to now, several in vivo and in vitro studies were performed to assess the reliability of VNC instead of TNC images.
In 2018, Andreas P. Sauter et al. investigated the reliability of VNC in a study with 62 patients. They assessed the attenuation values of TNC and VNC images obtained from the corticomedullary and portal venous phases of different tissues (aorta, fat, muscle, liver kidney, fluid, bone). Their results showed a difference of 1.1 ± 6.7 HU between TNC and VNC for liver, kidney, muscle, and fluid combined tissue, concluding that VNC could represent a promising tool for daily clinical routine [20].
Jasmin A. Holz et al. evaluated the relationship between TNC and dl-DECT VNC using an abdomen phantom with seven different tissue types and compared their attenuation and SD. Their results showed that the VNC error was −1.4 ± 6.1 and independent of dose ranges, kernel, and denoising setting [21].
Meyer et al. demonstrated a close agreement between the attenuation values of real and virtual images. Nonetheless, the mean error rate for the enhancement assessment was 12.1% (8.7% false positives and 3.4% false negatives). Therefore, they concluded that this bias may be accepted in the evaluation of incidental renal masses, but not for CT studies performed to characterize renal masses [22].
Moreover, Zhang et al. evaluated whether VNC images, obtained from the excretory phase, could replace the TNC series in the evaluation of patients with renal cell carcinoma [15]. They found an optimal agreement between VNC and TNC attenuation values, concluding that there was a possible interchangeability of VNC and TNC series.
However, despite the strong agreement between true and virtual non-contrast imaging, we have to keep in mind that VNC reconstructions cannot yet completely replace TNC images, especially when the latter become necessary in differential diagnosis. This could be relevant in cases of cystic lesions, as our study showed statistical significance in the equivalence test within the range of ±5 and ±10 HU, but not within the narrower range of ±3 HU (p > 0.05), in both corticomedullary and nephrographic phases.
Nevertheless, in patients who underwent many follow-up imaging examinations throughout their life, especially when young, and in patients from the emergency department with specific clinical questions, VNC reconstructions may be helpful in order to characterize an incidental lesion [20,22], with significant advantages coming from a decrease of radiation exposure [23,24,25].
The limitations of this study must be acknowledged. Firstly, the retrospective study design could have influenced our results. Secondly, we consider only the most frequent renal masses; other kinds of lesions will have to be assessed in future studies because they could show different results. Thirdly, we used just one dual-layer DECT scanner; the validity of VNC reconstructions with other DECT technology may be different, and this has to be evaluated in further studies.

5. Conclusions

Our study has demonstrated a strong agreement between VNC and TNC images when evaluating renal lesions. In particular, we observed a difference in attenuation values of less than 10 HU between the VNCcort and TNC series and between the VNCneph and TNC series in 92% and 89% of cases, respectively.
Thus, although the small discrepancies observed in a minority of cases emphasize the diagnostic value of TNC images, especially at the initial characterization of indeterminate renal masses, the VNC algorithm could be reliably exploited in following examinations as a substitute for an unenhanced scan, with relevant benefits for the patient in terms of dose protection.

Author Contributions

Conceptualization, G.A. and G.M.B.; Methodology, M.V.; Software, A.C.; Validation, T.D., C.B. and T.J.V.; Resources, F.M.A. and V.A.; Data curation, M.A.M.; Writing—original draft, G.M.B.; Writing—review & editing, G.C. and S.B.; Visualization, F.F. and F.P.; Supervision, G.A. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Ethical approval was not required due to the retrospective nature of the study, as stated by our institutional review board.

Informed Consent Statement

Written informed consent has been obtained from the patients to publish this paper.

Data Availability Statement

Not applicable.

Conflicts of Interest

C.B. has received speaking fees from Siemens Healthineers. T.D. has received speaking fees from Bracco and Philips. No support or funding was received for this project. The other authors have no conflict of interest to disclose.

Abbreviations

dl-DECTdual layer-Dual Energy CT
HUHounsfield Unit
ROIRegion of Interest
SBISpectral Basis Images
TNCTrue Non-Contrast
VNCVirtual Non-Contrast

References

  1. Meulepas, J.M.; Ronckers, C.M.; Smets, A.; Nievelstein, R.A.J.; Gradowska, P.; Lee, C.; Jahnen, A.; van Straten, M.; de Wit, M.Y.; Zonnenberg, B.; et al. Radiation Exposure From Pediatric CT Scans and Subsequent Cancer Risk in the Netherlands. J. Natl. Cancer Inst. 2019, 111, 256–263. [Google Scholar] [CrossRef] [PubMed]
  2. Berdahl, C.T.; Vermeulen, M.J.; Larson, D.B.; Schull, M.J. Emergency department computed tomography utilization in the United States and Canada. Ann. Emerg. Med. 2013, 62, 486–494.e3. [Google Scholar] [CrossRef]
  3. Hu, Y.; Ma, M.; Yin, H.; Ren, P.; Tian, X.; Zheng, Z.; Zhong, Z.; Wang, Z.; Yang, Z.; Chen, H. Assessment of cumulative cancer risk attributable to diagnostic X-ray radiation: A large cohort study. Eur. Radiol. 2022, 33, 1769–1778. [Google Scholar] [CrossRef]
  4. Langevin, C.; Normandeau, L.; Bouin, M. Diagnostic Radiation Exposure in Patients with Inflammatory Bowel Disease. Can. J. Gastroenterol. Hepatol. 2019, 2019, 2030735. [Google Scholar] [CrossRef] [Green Version]
  5. Solomon, J.; Marin, D.; Roy Choudhury, K.; Patel, B.; Samei, E. Effect of Radiation Dose Reduction and Reconstruction Algorithm on Image Noise, Contrast, Resolution, and Detectability of Subtle Hypoattenuating Liver Lesions at Multidetector CT: Filtered Back Projection versus a Commercial Model-based Iterative Reconstruction Algorithm. Radiology 2017, 284, 777–787. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  6. Nakamoto, A.; Yamamoto, K.; Sakane, M.; Nakai, G.; Higashiyama, A.; Juri, H.; Yoshikawa, S.; Narumi, Y. Reduction of the radiation dose and the amount of contrast material in hepatic dynamic CT using low tube voltage and adaptive iterative dose reduction 3-dimensional. Medicine 2018, 97, e11857. [Google Scholar] [CrossRef] [PubMed]
  7. Greffier, J.; Dabli, D.; Hamard, A.; Belaouni, A.; Akessoul, P.; Frandon, J.; Beregi, J.P. Effect of a new deep learning image reconstruction algorithm for abdominal computed tomography imaging on image quality and dose reduction compared with two iterative reconstruction algorithms: A phantom study. Quant. Imaging Med. Surg. 2022, 12, 229–243. [Google Scholar] [CrossRef]
  8. Nagayama, Y.; Oda, S.; Nakaura, T.; Tsuji, A.; Urata, J.; Furusawa, M.; Utsunomiya, D.; Funama, Y.; Kidoh, M.; Yamashita, Y. Radiation Dose Reduction at Pediatric CT: Use of Low Tube Voltage and Iterative Reconstruction. Radiographics 2018, 38, 1421–1440. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  9. Siegel, M.J.; Bhalla, S.; Cullinane, M. Dual-Energy CT Material Decomposition in Pediatric Thoracic Oncology. Radiol. Imaging Cancer 2021, 3, e200097. [Google Scholar] [CrossRef]
  10. Tatsugami, F.; Higaki, T.; Nakamura, Y.; Honda, Y.; Awai, K. Dual-energy CT: Minimal essentials for radiologists. Jpn. J. Radiol. 2022, 40, 547–559. [Google Scholar] [CrossRef]
  11. Rassouli, N.; Etesami, M.; Dhanantwari, A.; Rajiah, P. Detector-based spectral CT with a novel dual-layer technology: Principles and applications. Insights Imaging 2017, 8, 589–598. [Google Scholar] [CrossRef] [Green Version]
  12. Zhang, L.J.; Peng, J.; Wu, S.Y.; Wang, Z.J.; Wu, X.S.; Zhou, C.S.; Ji, X.M.; Lu, G.M. Liver virtual non-enhanced CT with dual-source, dual-energy CT: A preliminary study. Eur. Radiol. 2010, 20, 2257–2264. [Google Scholar] [CrossRef] [PubMed]
  13. Mileto, A.; Mazziotti, S.; Gaeta, M.; Bottari, A.; Zimbaro, F.; Giardina, C.; Ascenti, G. Pancreatic dual-source dual-energy CT: Is it time to discard unenhanced imaging? Clin. Radiol. 2012, 67, 334–339. [Google Scholar] [CrossRef] [PubMed]
  14. Silverman, S.G.; Pedrosa, I.; Ellis, J.H.; Hindman, N.M.; Schieda, N.; Smith, A.D.; Remer, E.M.; Shinagare, A.B.; Curci, N.E.; Raman, S.S.; et al. Bosniak Classification of Cystic Renal Masses, Version 2019: An Update Proposal and Needs Assessment. Radiology 2019, 292, 475–488. [Google Scholar] [CrossRef] [Green Version]
  15. Israel, G.M.; Bosniak, M.A. Pitfalls in renal mass evaluation and how to avoid them. Radiographics 2008, 28, 1325–1338. [Google Scholar] [CrossRef]
  16. Giavarina, D. Understanding Bland Altman analysis. Biochem. Med. 2015, 25, 141–151. [Google Scholar] [CrossRef] [Green Version]
  17. Pallwein-Prettner, L.; Flory, D.; Rotter, C.R.; Pogner, K.; Syre, G.; Fellner, C.; Frauscher, F.; Aigner, F.; Krause, F.S.; Fellner, F. Assessment and characterisation of common renal masses with CT and MRI. Insights Imaging 2011, 2, 543–556. [Google Scholar] [CrossRef] [Green Version]
  18. Bazzocchi, M.V.; Zilioli, C.; Gallone, V.I.; Commisso, C.; Bertolotti, L.; Pagnini, F.; Ziglioli, F.; Maestroni, U.; Aliprandi, A.; Buti, S.; et al. The Role of CT Imaging in Characterization of Small Renal Masses. Diagnostics 2023, 13, 334. [Google Scholar] [CrossRef]
  19. Parakh, A.; Kortesniemi, M.; Schindera, S.T. CT Radiation Dose Management: A Comprehensive Optimization Process for Improving Patient Safety. Radiology 2016, 280, 663–673. [Google Scholar] [CrossRef] [PubMed]
  20. Sauter, A.P.; Muenzel, D.; Dangelmaier, J.; Braren, R.; Pfeiffer, F.; Rummeny, E.J.; Noel, P.B.; Fingerle, A.A. Dual-layer spectral computed tomography: Virtual non-contrast in comparison to true non-contrast images. Eur. J. Radiol. 2018, 104, 108–114. [Google Scholar] [CrossRef] [PubMed]
  21. Holz, J.A.; Alkadhi, H.; Laukamp, K.R.; Lennartz, S.; Heneweer, C.; Pusken, M.; Persigehl, T.; Maintz, D.; Grosse Hokamp, N. Quantitative accuracy of virtual non-contrast images derived from spectral detector computed tomography: An abdominal phantom study. Sci. Rep. 2020, 10, 21575. [Google Scholar] [CrossRef] [PubMed]
  22. Meyer, M.; Nelson, R.C.; Vernuccio, F.; Gonzalez, F.; Farjat, A.E.; Patel, B.N.; Samei, E.; Henzler, T.; Schoenberg, S.O.; Marin, D. Virtual Unenhanced Images at Dual-Energy CT: Influence on Renal Lesion Characterization. Radiology 2019, 291, 381–390. [Google Scholar] [CrossRef] [PubMed]
  23. Zhang, X.; Zhang, G.; Xu, L.; Bai, X.; Lu, X.; Yu, S.; Sun, H.; Jin, Z. Utilisation of virtual non-contrast images and virtual mono-energetic images acquired from dual-layer spectral CT for renal cell carcinoma: Image quality and radiation dose. Insights Imaging 2022, 13, 12. [Google Scholar] [CrossRef] [PubMed]
  24. Si-Mohamed, S.; Dupuis, N.; Tatard-Leitman, V.; Rotzinger, D.; Boccalini, S.; Dion, M.; Vlassenbroek, A.; Coulon, P.; Yagil, Y.; Shapira, N.; et al. Virtual versus true non-contrast dual-energy CT imaging for the diagnosis of aortic intramural hematoma. Eur. Radiol. 2019, 29, 6762–6771. [Google Scholar] [CrossRef]
  25. Jamali, S.; Michoux, N.; Coche, E.; Dragean, C.A. Virtual unenhanced phase with spectral dual-energy CT: Is it an alternative to conventional true unenhanced phase for abdominal tissues? Diagn. Interv. Imaging 2019, 100, 503–511. [Google Scholar] [CrossRef]
Figure 1. Flowchart of patient inclusion and exclusion criteria.
Figure 1. Flowchart of patient inclusion and exclusion criteria.
Jcm 12 04718 g001
Figure 2. Flowchart of study design.
Figure 2. Flowchart of study design.
Jcm 12 04718 g002
Figure 3. The boxplot shows a similar distribution of TNC and VNC attenuation values. Asterisks (*) represent outlier values. TNC—True non-contrast; VNCcort—Virtual non-contrast of corticomedullary phase; VNCneph—VNC of nephrographic phase.
Figure 3. The boxplot shows a similar distribution of TNC and VNC attenuation values. Asterisks (*) represent outlier values. TNC—True non-contrast; VNCcort—Virtual non-contrast of corticomedullary phase; VNCneph—VNC of nephrographic phase.
Jcm 12 04718 g003
Figure 4. The graph shows how frequently the differences between TNC and VNC attenuation values (HU) are <5, in the range of 5–10, of 10–20, and >20 for (A) overall kidney masses, (B) kidney masses, (C) complex cysts, and (D) simple cysts. TNC—True non-contrast; VNCcort—VNC of corticomedullary phase; VNCneph—VNC of nephrographic phase.
Figure 4. The graph shows how frequently the differences between TNC and VNC attenuation values (HU) are <5, in the range of 5–10, of 10–20, and >20 for (A) overall kidney masses, (B) kidney masses, (C) complex cysts, and (D) simple cysts. TNC—True non-contrast; VNCcort—VNC of corticomedullary phase; VNCneph—VNC of nephrographic phase.
Jcm 12 04718 g004
Figure 5. Bland−Altman plot demonstrates the agreement between the measurements using the virtual non-contrast (VNC) attenuation values and unenhanced attenuation values. The horizontal dashed lines represent the mean (1.7 HU) and the limits of the 95% confidence interval (6.6 HU, −9.9 HU) for the difference. VNCcort—VNC of corticomedullary phase; VNCneph—VNC of nephrographic phase.
Figure 5. Bland−Altman plot demonstrates the agreement between the measurements using the virtual non-contrast (VNC) attenuation values and unenhanced attenuation values. The horizontal dashed lines represent the mean (1.7 HU) and the limits of the 95% confidence interval (6.6 HU, −9.9 HU) for the difference. VNCcort—VNC of corticomedullary phase; VNCneph—VNC of nephrographic phase.
Jcm 12 04718 g005
Figure 6. The comparison of ROC curves shows the diagnostic performance of VNC series compared to TNC images. TNC—True non-contrast; AUC—Area under the curve; VNCcort—VNC of corticomedullary phase; VNCneph—VNC of nephrographic phase.
Figure 6. The comparison of ROC curves shows the diagnostic performance of VNC series compared to TNC images. TNC—True non-contrast; AUC—Area under the curve; VNCcort—VNC of corticomedullary phase; VNCneph—VNC of nephrographic phase.
Jcm 12 04718 g006
Figure 7. A 78-year-old male patient with recurrent renal cell carcinoma. Equal ROIs were drawn in all three phases: (a) TNC, (b) VNC obtained from the arterial phase, (c) VNC obtained from the venous phase. The attenuation value (HU) and the standard deviations were, respectively, 41.8 ± 13.5, 46.2 ± 16.8, and 39.8 ± 13.7. The difference in attenuation between the real and virtual images is less than 5 HU, and the images are qualitatively comparable, also in terms of the noise level. TNC—True non-contrast; VNC—Virtual non-contrast. Asterisks (*) indicate the HU of VNC images.
Figure 7. A 78-year-old male patient with recurrent renal cell carcinoma. Equal ROIs were drawn in all three phases: (a) TNC, (b) VNC obtained from the arterial phase, (c) VNC obtained from the venous phase. The attenuation value (HU) and the standard deviations were, respectively, 41.8 ± 13.5, 46.2 ± 16.8, and 39.8 ± 13.7. The difference in attenuation between the real and virtual images is less than 5 HU, and the images are qualitatively comparable, also in terms of the noise level. TNC—True non-contrast; VNC—Virtual non-contrast. Asterisks (*) indicate the HU of VNC images.
Jcm 12 04718 g007
Figure 8. A 71-year-old male patient with a complex cystic lesion (Bosniak type II F). The attenuation values and SD of (a) baseline image, (b) virtual non-contrast (VNC) obtained from the arterial phase, and (c) VNC obtained from the venous phase are, respectively, 6.8 ± 20.4 HU, 7.6 ± 11.6 HU, and 8.1 ± 10.9 HU. The difference between the attenuations is less than 5 HU, and the VNC images have a lower noise level, resulting in a lower SD value. Asterisks (*) indicate the HU of VNC images.
Figure 8. A 71-year-old male patient with a complex cystic lesion (Bosniak type II F). The attenuation values and SD of (a) baseline image, (b) virtual non-contrast (VNC) obtained from the arterial phase, and (c) VNC obtained from the venous phase are, respectively, 6.8 ± 20.4 HU, 7.6 ± 11.6 HU, and 8.1 ± 10.9 HU. The difference between the attenuations is less than 5 HU, and the VNC images have a lower noise level, resulting in a lower SD value. Asterisks (*) indicate the HU of VNC images.
Jcm 12 04718 g008
Figure 9. (a) Hyperdense cystic lesion in TNC image with an attenuation value of 75.1 HU represents a Bosniak 2 lesion, not to be further investigated. However, in the VNC images obtained from the (b) arterial and (c) portal-venous phases, the lesion’s attenuation values are, respectively, 48.9 and 50.4 HU. VNC images, in this case, do not allow an accurate and correct diagnosis. TNC—True non-contrast; VNC—Virtual non-contrast. Asterisks (*) indicate the HU of VNC images.
Figure 9. (a) Hyperdense cystic lesion in TNC image with an attenuation value of 75.1 HU represents a Bosniak 2 lesion, not to be further investigated. However, in the VNC images obtained from the (b) arterial and (c) portal-venous phases, the lesion’s attenuation values are, respectively, 48.9 and 50.4 HU. VNC images, in this case, do not allow an accurate and correct diagnosis. TNC—True non-contrast; VNC—Virtual non-contrast. Asterisks (*) indicate the HU of VNC images.
Jcm 12 04718 g009
Table 1. Acquisition parameters.
Table 1. Acquisition parameters.
ParametersDECT
Collimation (mm)64 × 0.625
Tube voltage (kv)120
Rotation time (s)0.33
Tube current (ma)89
Pitch0.9
Duration of tube current (s)0.5
CTDI (mgy)2.9
DECT—Dual-Energy CT; CTDI—Computed tomography dose index.
Table 2. The means of all CT values of renal masses obtained from TNC and VNC reconstructions.
Table 2. The means of all CT values of renal masses obtained from TNC and VNC reconstructions.
Type of LesionPhaseMean Value (HU)SD
All lesionsTNC17.616.5
VNCcort15.713.2
VNCneph15.612.7
Simple CystsTNC10.715.6
VNCcort7.812.2
VNCneph7.511.9
Complex CystsTNC15.318.6
VNCcort12.713.6
VNCneph13.114.4
Renal MassesTNC34.316.9
VNCcort34.115.1
VNCneph33.913.4
TNC—True non-contrast; VNCcort—Virtual non-contrast of corticomedullary phase; VNCneph—VNC of nephrographic phase.
Table 3. Results of Wilcoxon signed-rank test based on TOST between the ranges −3, +3 HU, −5, +5 HU, and −10, +10 HU.
Table 3. Results of Wilcoxon signed-rank test based on TOST between the ranges −3, +3 HU, −5, +5 HU, and −10, +10 HU.
Corticomedullar PhaseNephrographic Phase
p ValueCI 95%p ValueCI 95%
Range (HU)±3±5±10 ±3±5±10
All lesions0.0034.68 × 10−98.42 × 10−15[0.8, 2.4]0.0022.19 × 10−82.03 × 10−15[1, 2.54]
Renal masses0.0014.54 × 10−56.27 × 10−6[−1.3, 1.5]0.0012.63 × 10−51.5 × 10−6[−0.8, 2]
Cystic lesions0.16 *2.38 × 10−51.61 × 10−10[1.4, 3.25]0.07 *6.26 × 10−52.47 × 10−10[1.4, 3.3]
* Values with asterisk (*) do not show statistical significance (p value > 0.05). CI—Confidence Interval.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Bucolo, G.M.; Ascenti, V.; Barbera, S.; Fontana, F.; Aricò, F.M.; Piacentino, F.; Coppola, A.; Cicero, G.; Marino, M.A.; Booz, C.; et al. Virtual Non-Contrast Spectral CT in Renal Masses: Is It Time to Discard Conventional Unenhanced Phase? J. Clin. Med. 2023, 12, 4718. https://doi.org/10.3390/jcm12144718

AMA Style

Bucolo GM, Ascenti V, Barbera S, Fontana F, Aricò FM, Piacentino F, Coppola A, Cicero G, Marino MA, Booz C, et al. Virtual Non-Contrast Spectral CT in Renal Masses: Is It Time to Discard Conventional Unenhanced Phase? Journal of Clinical Medicine. 2023; 12(14):4718. https://doi.org/10.3390/jcm12144718

Chicago/Turabian Style

Bucolo, Giuseppe M., Velio Ascenti, Simone Barbera, Federico Fontana, Francesco M. Aricò, Filippo Piacentino, Andrea Coppola, Giuseppe Cicero, Maria Adele Marino, Christian Booz, and et al. 2023. "Virtual Non-Contrast Spectral CT in Renal Masses: Is It Time to Discard Conventional Unenhanced Phase?" Journal of Clinical Medicine 12, no. 14: 4718. https://doi.org/10.3390/jcm12144718

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop