Next Article in Journal
MWCNTs-GNPs Reinforced TPU Composites with Thermal and Electrical Conductivity: Low-Temperature Controlled DIW Forming
Next Article in Special Issue
A Lab-on-a-Tube Biosensor Combining Recombinase-Aided Amplification and CRISPR-Cas12a with Rotated Magnetic Extraction for Salmonella Detection
Previous Article in Journal
Spatial Distribution of Flagellated Microalgae Chlamydomonas reinhardtii in a Quasi-Two-Dimensional Space
Previous Article in Special Issue
Determining Spatial Variability of Elastic Properties for Biological Samples Using AFM
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Quantification of Blood Viscoelasticity under Microcapillary Blood Flow

Department of Mechanical Engineering, Chosun University, 309 Pilmun-daero, Dong-gu, Gwangju 61452, Republic of Korea
Micromachines 2023, 14(4), 814; https://doi.org/10.3390/mi14040814
Submission received: 10 March 2023 / Revised: 31 March 2023 / Accepted: 2 April 2023 / Published: 3 April 2023
(This article belongs to the Special Issue Advanced Biomanufacturing for Biomedical Engineering Applications)

Abstract

:
Blood elasticity is quantified using a single compliance model by analyzing pulsatile blood flow. However, one compliance coefficient is influenced substantially by the microfluidic system (i.e., soft microfluidic channels and flexible tubing). The novelty of the present method comes from the assessment of two distinct compliance coefficients, one for the sample and one for the microfluidic system. With two compliance coefficients, the viscoelasticity measurement can be disentangled from the influence of the measurement device. In this study, a coflowing microfluidic channel was used to estimate blood viscoelasticity. Two compliance coefficients were suggested to denote the effects of the polydimethylsiloxane (PDMS) channel and flexible tubing (C1), as well as those of the RBC (red blood cell) elasticity (C2), in a microfluidic system. On the basis of the fluidic circuit modeling technique, a governing equation for the interface in the coflowing was derived, and its analytical solution was obtained by solving the second-order differential equation. Using the analytic solution, two compliance coefficients were obtained via a nonlinear curve fitting technique. According to the experimental results, C2/C1 is estimated to be approximately 10.9–20.4 with respect to channel depth (h = 4, 10, and 20 µm). The PDMS channel depth contributed simultaneously to the increase in the two compliance coefficients, whereas the outlet tubing caused a decrease in C1. The two compliance coefficients and blood viscosity varied substantially with respect to homogeneous hardened RBCs or heterogeneous hardened RBCs. In conclusion, the proposed method can be used to effectively detect changes in blood or microfluidic systems. In future studies, the present method can contribute to the detection of subpopulations of RBCs in the patient’s blood.

1. Introduction

The rheological properties of blood are significantly influenced by highly flexible red blood cells (RBCs) and aqueous plasma. Physiological disorders (i.e., hypertension [1,2], diabetes [3,4], sickle cell anemia [5,6], and malaria [7,8]) contribute to altered hemorheological properties (i.e., hematocrit (Hct) [9,10], RBC deformability, RBC aggregation, and viscosity) [11,12]. The mechanical properties of blood are regarded as potential biomarkers and have been widely used in clinical settings [13,14]. Among the mechanical properties of blood, fluid viscosity is commonly used to monitor substantial variations in blood. Blood behaves as a viscoelastic fluid (i.e., viscosity or elasticity) [15]. Blood viscoelasticity is generally obtained with respect to sinusoidally varying angular velocities. In other words, a conventional rheometer (i.e., cone-and-plate or plate-and-plate), as gold standard, can provide quantitative information about blood viscoelasticity [16,17]. However, the conventional method requires a large blood volume (on the order of milliliters) to fill the space between the two plates, tedious cleaning procedures per experiment, and an expert for confirming reliable and consistent data, in addition to having a high operating cost. Additionally, owing to its bulky size and precision requirements, the conventional rheometer is available in limited research environments [18].
A microfluidic platform is a potential solution for resolving the problems associated with the conventional rheometer. Microfluidic devices offer distinctive merits, including low blood volume consumption, short measurement, and high sensitivity. Most importantly, the microfluidic device embodies a bio-mimicking interface or environment (i.e., pressure-driven flow, similar dimensions, and configuration) when compared with in vivo capillary blood flow. Therefore, microfluidic devices have been suggested and adapted to obtain rheological properties [19,20]. First, to measure the blood viscosity under a constant shearing flow, the flow rate is set to a constant value using a syringe pump or pressure source. Using the Hagen–Poiseuille equation (i.e., pressure drop = fluidic resistance × flow rate), the blood viscosity is obtained by quantifying the pressure drop or fluidic resistance under constant shearing blood flow. According to previous studies, several quantification methods, including droplet velocity [21], droplet length [22,23], blood advancing velocity [24,25,26], digital flow compartment [27,28], and parallel flow [29,30], have been demonstrated to effectively measure fluidic resistance (or viscosity) in microfluidic channels. According to the parallel flow method [29,31], the reference fluid (i.e., 1× phosphate-buffered saline [PBS]) and test fluid (i.e., blood) are supplied into each inlet at the same flow rate, and the viscosity of the reference fluid is specified in advance. As the interface is determined by the viscosity ratio of the two fluids, the blood viscosity is obtained by detecting the interface without calibration. Next, to obtain blood viscoelasticity, the flow rate is set to a periodic on/off or sinusoidal pattern. Under transient blood flow, blood velocity [32] or image intensity [33] gradually decreases over time. On the basis of the transient behaviors of physical parameters (i.e., velocity and image intensity [34]), the regression formula was assumed to be I (t) = I0 + I1 exp (−t/κ) [33]. Using temporal variations in image intensity, time constant κ was obtained by conducting nonlinear regression analysis. Using the linear Maxwell model (i.e., time constant = viscosity/elasticity) [13], blood elasticity is obtained by dividing viscosity by the time constant (i.e., elasticity = viscosity/time constant) [35]. Here, the blood viscosity should be specified in advance to extract blood elasticity from the time constant. The Maxwell model was suggested to represent the viscoelasticity of a single RBC under Couette flow conditions. However, for blood flow in a microfluidic channel, a new mathematical model is required to effectively evaluate the blood elasticity effect. Recently, our group suggested that a single compliance element should be adopted to extract blood elasticity under transient blood flows [36]. Compliance has a reciprocal relationship with elasticity (i.e., compliance ∝ 1/elasticity) [37]. Although it is aimed at obtaining blood elasticity, a single compliance element includes the combined effects of several factors including RBC elasticity, flexible microfluidic channels, and polyethylene tubing. Therefore, it is necessary to update the previous mathematical model. Therefore, instead of a single compliance element, two compliance elements are suggested to effectively separate the contributions of RBC elasticity from the other two factors.
In this study, two compliance elements are added to effectively evaluate the contribution of the RBC elastic effect. As the coflowing method gives accurate blood viscosity without calibration, the present study adopts a coflowing microfluidic channel. On the basis of the discrete fluidic circuit modeling technique, a new governing equation for the microfluidic system is derived as a second-order differential equation. The interfacial location is selected as the dependent variable and used for obtaining blood viscosity and two compliances. According to the governing equation, the blood viscosity is obtained using the steady values of the interface under steady blood flow. Next, nonlinear regression analysis is conducted to obtain two compliance elements after transient variations of the interface are obtained under transient blood flow. The two compliance elements exhibit individual contributions of RBC elasticity, microfluidic device, and tubing.
Compared with previous methods [35,36], the elastic effect of the microfluidic system is modeled as two compliance elements rather than a single compliance element. The governing equation of the microfluidic system is changed from the first-order differential equation to a second-order differential equation. The two compliance elements are effectively used to evaluate the elastic effect of RBCs, flexible device, and tubing. As a demonstration, blood viscosity and two compliance coefficients are used to detect homogeneous and heterogeneous hardened RBCs.

2. Materials and Methods

2.1. Microfluidic Device and Experimental Procedures

To measure the blood viscosity and two compliance coefficients, a microfluidic device was designed with two inlets (a and b), an outlet, and three microfluidic channels (i.e., reference, blood, and coflowing channels), as shown in Figure 1(A-i). The corresponding lengths of the blood and coflowing channels were set to l1 = 7500 µm and l2 = 4800 µm. The three channels were set to have the same width (w = 250 µm). However, to evaluate the effect of the channel depth on the compliance effect, the three channel depths were set to h = 4, 10, and 20 μm.
A four-inch silicon master mold was fabricated using conventional microelectromechanical system fabrication techniques (i.e., photolithography and deep reactive ion etching). To perform soft lithography on the silicon master mold, polydimethylsiloxane (PDMS) (Sylgard 184, Dow Corning, Midland, Michigan, USA) was mixed with a curing agent in a mass ratio of 10:1 (i.e., PMDS to curing agent). The PDMS mixture was poured onto a silicon master mold adhered to a petri dish. Air bubbles in the PDMS were removed using a vacuum pump for 1 h. After curing the PDMS mixture in a convection oven at 70 °C for 1 h, the PDMS block was peeled off from the silicon master mold. The PDMS device was then cut using a razor blade, and three ports (two inlets and one outlet) were punched with a biopsy punch (outer diameter = 0.75 mm). After oxygen plasma treatment (CUTE-MPR, Femto Science Co., Gyeonggi-do, Korea), a microfluidic device was fabricated by bonding a PDMS device on a glass substrate. To ensure strong bonding between the two surfaces, the device was placed on a hot plate (120 °C) for 10 min.
Two types of inlet tubing (inner diameter = 250 µm, Lin = 300 mm) were attached to each inlet port (a, b). The outlet tubing (inner diameter = 250 µm, Lout = 200–400 mm) was connected to the outlet port. To remove the initial air in the microfluidic channels and avoid nonspecific binding of plasma protein to the surface of the microfluidic channels, bovine serum albumin (2 mg/mL) was loaded from the outlet. After 10 min, blood (approximately 0.3 mL) and reference fluid (approximately 0.3 mL) were loaded into two disposable syringes (1 mL). The needle in each syringe was connected to the end of each inlet tube. Thereafter, two syringes were installed in each syringe pump (NeMESYS; Cetoni GmbH, Germany). Blood was supplied to inlet (a) in a periodic on/off pattern (amplitude: Q0 and period: T = 240 s). The reference fluid was supplied to inlet (b) at a constant flow rate (Q0).
As shown in Figure 1(A-ii), the microfluidic device was positioned on an optical microscope (BX51, Olympus, Japan) equipped with a 10× objective lens (NA = 0.25). A high-speed camera (FASTCAM MINI, Photron, London, United Kingdom) was used to capture the blood flow images in the microfluidic channels. To clearly visualize blood flow, the frame rate was set to 5000 fps (frame per second). Using a function generator (WF1944B, NF Corporation, Yokohama, Japan), a pulse signal with a period of 0.5 s triggered the high-speed camera. Two microscopic images were captured sequentially at intervals of 0.5 s.

2.2. Quantification of Interface in the Coflowing Channel

In Figure 1(A-iii), Qr and Qb represent the flow rate of the reference fluid in the reference channel and blood flow rate in the blood channel, respectively. To obtain the interface in the coflowing channel, grayscale images were converted into binary-scale images using Otsu method [38]. A specific region of interest (ROI, 250 µm × 330 µm) was selected within the coflowing channel. By averaging the blood-filled widths distributed within the ROI, the average blood-filled width was calculated and denoted by wb. Dividing the averaged blood-filled width (wb) by the channel width (w), the dimensionless form of the interface could be expressed as αb = wb/w. The variations in the interface were then obtained at intervals of 0.5 s.

2.3. Quantification of Blood Velocity in the Blood Channel

To monitor blood flow in the blood channel supplied from a syringe pump (Qb), a specific ROI (250 µm × 330 µm) was selected at a far distance from the junction (‘j’). Blood velocity fields were obtained with time-resolved microparticle image velocimetry [39]. The interrogation window was set to 32 × 32 pixels. One pixel corresponded to 1.67 µm. The window overlap set to 50%. The blood velocity (Ub) was then obtained by arithmetically averaging the velocity fields distributed over the ROI.

2.4. Mathematical Representation of the Present Microfluidic System

To quantify the blood viscosity and two compliance coefficients, the blood flow rate was set to a periodic on/off pattern. By analyzing the interface (αb) in the coflowing channel, the viscosity and two compliance coefficients were obtained at constant blood flow and transient blood flow, respectively. Three assumptions were made to derive the governing differential equation of blood flow in the microfluidic system. First, by designing a microfluidic channel with a low aspect ratio (AR) (AR = h/w < 0.1), the fluid flow in the rectangular channel was approximated as a two-dimensional distribution. Second, by supplying it at a sufficiently high shear rate of γ ˙ > 103 s−1, blood was assumed to be a Newtonian fluid. Lastly, referring to a previous study [35], the blood viscosity remained constant within a specific interface (i.e., 0.1 < αb < 0.9). Therefore, the present study did not consider viscosity reduction owing to the cell-free layer (i.e., Fåhraeus effect). According to previous studies [35,36,40], the contribution of blood elasticity is represented by a single compliance element. By suddenly stopping the blood flow, the characteristic time was quantified by analyzing the transient behaviors of the cell-to-liquid interface in the coflowing channel. Next, according to the linear Maxwell model (i.e., characteristic time = viscosity/elasticity), the blood elasticity was obtained by dividing the viscosity by the characteristic time. However, as a limitation of the previous study, the single compliance element represented the combined effects of blood elasticity, tubing effect, and PDMS device effect. Two compliance elements were added to the fluidic circuit model to separate the individual contributions of the three components. As shown in Figure 1B, the first compliance element (C1) represents the compliance of the inlet tubing and PDMS channel, while the second compliance element (C2) denotes that of blood elasticity. As expected, blood exhibited a higher value of compliance than the tubing or device; therefore, it was assumed that C1 is smaller than C2 (C1 < C2). Owing to the two compliance elements, the order of the governing differential equation increased from first-order to second-order. As a result, the governing equation became more complex. Under a transient blood flow, the analytical solution of αb was derived by solving the governing equation. Thereafter, two compliance coefficients were obtained by conducting a nonlinear regression analysis.
As shown in Figure 1B, the microfluidic system (i.e., two fluids, a flexible microfluidic device, and polyethylene tubing) was modeled as discrete fluidic elements, including flow rate elements (Qr and Qb), fluidic resistance elements (Rr1, Rr2, Rb1, and Rb2), and compliance elements (C1 and C2). The subscripts ‘r’ and ‘b’ represent the reference fluid and blood, respectively; symbol ‘►’ denotes the zero value of pressure (P = 0), called ‘GND’. To model the coflowing channel filled with blood and reference fluid, the interface was assumed to be a virtual wall. Two fluid streams were modeled as two fluidic resistances (Rr2 and Rb2) connected in parallel. Subsequently, a correction factor (CF) was used to compensate for the difference between the physical model and simple mathematical model [35,41,42]. According to previous studies, the correction factor can be expressed as an interface (i.e., CF = CF [αb]). The fluidic resistance equations of two fluid streams (i.e., width of the reference fluid stream = [1 − αb] × w, and width of blood stream = αb × w) in the coflowing channel were derived as
R r 2 = 12 μ r l 2 1 - α b w h 3 ,
R b 2 = 12 μ b l 2 C F α b w h 3 ,
where μ r   and   μ b represent the viscosities of the reference fluid and blood, respectively. Next, the fluidic resistance equation for the upper channel filled with blood (i.e., channel length = l1) was derived as
R b 1 = 12 μ b l 1 w h 3 .
As a function of the junction point (‘j’) where both channels were joined, mass conservation of the reference fluid resulted in the following expression:
Q r = P j R r 2 .
In addition, as a function of each point (i.e., ‘a’ and ’j’), the mass conservation of blood yields the following equations:
Q b = C 1 d d t P a + P a - P j R b 1 ,
P a - P j R b 1 = C 2 d d t P j + P j R b 2 .
When Equations (1)–(4) and (6) were substituted into Equation (5), the differential equation of the interface was derived as
R b 1 R r 3 C 1 C 2 d 2 d t 2 1 1 - α b + C 1 R r 3 + C 2 R r 3 d d t 1 1 - α b + R r 3 C 1 l 1 l 2 d d t α b C F 1 - α b + μ r μ b α b C F 1 - α b = Q b Q r ,
where R r 3 = 12 μ r l 2 w h 3 . Under steady blood flow conditions, the equation for blood viscosity was derived as
μ b = μ r × α b C F 1 - α b × Q r Q b .
Next, by setting the blood flow rate to zero (Qb = 0), the right-hand term of Equation (7) becomes equal to zero (Qb/Qr = 0). However, the governing equation includes the nonlinear term α b C F ( α b ) 1 - α b . Therefore, it is necessary to convert the nonlinear term into a linear term to obtain an approximate solution for the transient blood flow. CF (αb) was assumed as a weighting function in the expression of α b C F ( α b ) 1 - α b . That is, the CF (αb) was varied continuously within the specific range of the interface. For convenience, it was assumed that CF (αb) remained the constant over interface. Using an arithmetic averaging procedure, the constant CF0 was obtained by averaging CF (αb) × αb/(1 − αb) within a specific value of interface. According to the approximation procedure reported in a previous study [43],
α b C F α b 1 - α b α b 1 - α b × C F 0 ,
where the term α b C F ( α b ) 1 - α b is simplified as
α b C F ( α b ) 1 - α b α b C F 0 1 - α b .
Under transient blood flow conditions, Equation (7) becomes the following second-order linear differential equation:
a d 2 d t 2 β b + b d d t β b + c β b = c ,
where β b can be expressed as β b = ( 1 - α b ) - 1 . The three constants are given as
a = R b 1 R r 3 C 1 C 2 , b = R r 3 C 1 + C 2 + C 1 × l 1 l 2 × C F 0 , c = μ r μ b × C F 0 .
The general solution to the differential Equation (11) was derived as
β b = d 1 exp - λ 1 t + d 2 exp - λ 2 t + 1 ,
where the two eigenvalues (λ1 and λ2) satisfy the following relations:
λ 1 + λ 2 = C 1 + C 2 + C 1 × l 1 l 2 × C F 0 R b 1 C 1 C 2 ,
λ 1 × λ 2 = 1 R b 1 R r 3 C 1 C 2 × μ r μ b × C F 0 .
In this study, the two eigenvalues (λ1 and λ2) were obtained by conducting a nonlinear regression analysis of the temporal variations of β b = ( 1 - α b ) - 1 . As a preliminary demonstration (shown in Figure 1(C-i)), variations in the interface (αb) were obtained under periodic on/off blood flow. Control blood (Hct = 50%) was prepared by adding normal RBCs to the autologous plasma. The blood flow rate was set to a pulsatile pattern (Q0 = 1 mL/h and T = 240 s), and the flow rate of the reference fluid was set at Qr = 1 mL/h. After turning on a syringe pump, interface arrived at constant value after an elapse of 70 s. To obtain blood viscosity under constant shearing blood flow, it was necessary to guarantee enough interval of constant interface over time. According to a transient response, the turn-on time of the syringe pump was set to 120 s, while the turn-off time of the syringe pump was fixed at 120 s. As shown in Figure 1(C-ii), using Equation (8), the blood viscosity (µb) could be obtained with steady values of αb. Next, as shown in Figure 1(C-iii), the temporal variations in αb were converted into βb = (1 − αb)−1. Subsequently, the two eigenvalues (λ1 and λ2) were calculated as λ1 = 0.1945 s−1 and λ2 = 0.0127 s−1 by conducting a nonlinear regression of βb (i.e., βb = d1 exp [−λ1t] + d2 exp [−λ2t] +1). The two compliance coefficients (C1 and C2) were obtained as C1 = 53.3 µm3/mPa and C2 = 1398.2 µm3/mPa by simultaneously solving Equations (13) and (14).

2.5. Blood Preparation for Validating the Present Method

This study was conducted in accordance with the principles of the Declaration of Helsinki. The protocol was approved by the ethics committee of Chosun University under the reference code (2-1041055-AB-N-01-2022-47). Packed red blood cell (RBC) bags (approximately 320 mL) and fresh frozen plasma (FFP) bags (approximately 320 mL) were purchased from the Gwangju–Chonnam blood bank (Gwangju, Korea). Before the blood test, RBCs and FFP were stored at 4 °C and −20 °C, respectively, in a refrigerator. To collect normal RBCs from concentrated RBCs, suspended blood was prepared by adding concentrated RBCs to 1× PBS (pH 7.4, Gibco, Life Technologies, California, USA). After operation of the centrifugal separator for approximately 10 min, the suspended blood was separated into two layers (i.e., upper and lower layers). Normal RBCs in the lower layer were retained by removing the upper layer (i.e., buffy layer and 1× PBS) from the suspended blood. The washing procedure was repeated thrice. After FFP was melted at a constant temperature of 25 °C, autologous plasma was prepared by removing debris from FFP using a syringe filter (mesh size = 5 μm, Minisart, Sartorius, Göttingen, Germany). To simulate various values of blood viscoelasticity, suspended blood was prepared by adding normal RBCs or hardened RBCs to specific diluents (i.e., 1× PBS and autologous plasma). To sufficiently rigidify normal RBCs, five different types of diluted glutaraldehyde (GA) (CGA = 2, 4, 6, 8, and 10 µL/mL) were prepared by adding GA (Grade II, 25% in H2O, Sigma-Aldrich, St. Louis, MO, USA) to 1× PBS. Normal RBCs were hardened by mixing them with each concentration of the GA solution for 10 min. After preparation of the suspended blood, specific experiments were conducted immediately to obtain consistent results.

3. Results and Discussion

3.1. Contribution of Channel Depth to Blood Viscosity under Steady Flow

First, to measure the blood viscosity using Equation (8), it was necessary to obtain a correction factor within a specific range of the interface. Figure 2(A-i) shows a microscopic image of the interface of the two fluids (i.e., test fluid: glycerin [30%], reference fluid: 1× PBS) in the coflowing channel at the same flow rate (i.e., Qr = Qt = 1 mL/h); αt denotes the interface in the coflowing channel. Various values of αt were obtained by adjusting Qr or Qt using a syringe pump. By manipulating Equation (8), the correction factor (CF = CF [αt]) could be expressed as
C F = μ t μ r × 1 - α t α t × Q t Q r ,
where the viscosity of glycerin (30%) and 1× PBS was taken as µt = 3 cP and µr = 1 cP [44]. Using three channel depths (h = 4, 10, and 20 µm), variations in the interface were obtained by adjusting either Qr or Qt with a syringe pump. By substituting five parameters (µt, µr, αt, Qb, and Qr) into Equation (15), the correction factor was obtained as a function of the interface. Figure 2(A-ii) shows the variations in CF with respect to the channel depth (h) and interface. According to the linear regression analysis, the corresponding correction factor for each channel depth was obtained as a polynomial expression: (i) CF = 7.5698αt3 − 7.669αt2 + 4.1983αt − 0.1289 (R2 = 0.9858) for h = 4 µm; (ii) CF = 0.6931αt2 + 0.9378αt + 0.38 (R2 = 0.9525) for h = 10 µm; (iii) CF = 27.95αt5 − 54.452αt4 + 36.954αt3 − 11.062αt2 + 2.5117αt + 0.4163 (R2 = 0.9354) for h = 20 µm.
Next, to approximate the nonlinear term ( α b C F [ α b ] 1 - α b ) in Equation (7), the constant correction factor (CF0) was obtained using Equation (9). The corresponding CF0 for each channel depth was obtained as (i) CF0 = 1.637 for h = 4 µm, (ii) CF0 = 1.41 for h = 10 µm, and (iii) CF0 = 1.183 for h = 20 µm. Figure 2B shows the difference between the original expression ( α t C F [ α b ] 1 - α t ) and approximation expression ( α t C F 0 1 - α t ) with respect to the interface and channel depth (h = 4, 10, and 20 µm). The lower value of the channel depth exhibited a slight difference when compared with a high channel depth. For a higher channel depth (h = 20 µm), the difference between the original expression and approximation expression was much smaller with respect to the specific interface.
Lastly, the blood viscosity was measured with respect to the channel depth. Figure 2(C-i) shows the microscopic images of the interface with respect to h = 4 and 20 µm. Blood (Hct = 50%) was prepared by adding normal RBCs to 1× PBS. The corresponding interface for each channel was obtained as αb = 0.56 ± 0.01 for h = 4 µm and αb = 0.61 ± 0.01 for h = 20 µm. Figure 2(C-ii) shows the effect of channel depth on µb with respect to shear rate. According to shear rate formula in the rectangular channel (i.e., γ ˙ = 6 Q b w h 2 ), channel depth contributed to decreasing shear rate. Shear rate decreased substantially at higher values of channel depth. The corresponding blood viscosity for each channel depth was obtained as µb = 1.63 ± 0.04 cP for h = 4 µm, µb = 1.73 ± 0.11 cP for h = 10 µm, and µb = 1.86 ± 0.06 cP for h = 20 µm. The inset shows variations of µb with respect to h. The channel depth substantially contributed to the increase in blood viscosity. The results are in agreement with those of a previous study [35]. Furthermore, variations in blood viscosity were obtained with respect to shear rate. The channel depth was set as h = 20 µm, and three types of blood (i.e., plasma, normal RBCs in plasma, and normal RBCs in 1× PBS) were prepared to evaluate variations in blood viscosity with respect to shear rate. The flow rates of both the fluids (i.e., blood and reference fluid) were set to the same flow rate (i.e., Qb = Qr). The flow rate of syringe pump was set to Qb = 0.025, 0.05, 0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.7, 0.8, 0.9, and 1 mL/h. According to the equation for the shear rate in a rectangular blood channel, the shear rate was estimated to be γ ˙ = 679–27,742 s−1. Figure 2(C-iii) shows the variations in µb with respect to γ ˙ . As expected, the plasma viscosity remained constant with respect to the shear rate (µb = 1.58 ± 0.04 cP), and the remaining two blood samples exhibited non-Newtonian behavior. Plasma contributed substantially to the increase in blood viscosity compared with 1× PBS. For values greater than Qb = 0.2 mL/h (i.e., γ ˙ > 4000 s−1), the blood viscosity remained constant with respect to the shear rate.
According to the experimental results, the correction factor varied substantially with respect to the channel depth. The blood viscosity increased significantly at higher channel depths. Furthermore, according to the approximation procedure of the nonlinear term in the governing equation, the constant correction factor (CF0) was obtained as CF0 = 1.183 for a channel depth of h = 20 µm. However, the approximation of the nonlinear term did not show a substantial difference when compared with the original nonlinear expression.

3.2. Contribution of Channel Depth and Tubing Length to Two Compliance Coefficients

Before testing the blood, glycerin (30%) and 1× PBS were supplied to the microfluidic channel. As shown in Figure 1(A-i), Q0 was set to 1 mL/h. Figure 3A shows the temporal variations in αt with respect to channel depth (h = 4, 10, and 20 µm). Under transient flow (i.e., t > 180 s), the interface decreased substantially at higher channel depths. Using Equation (11), αt was replaced by βt = (1 − αt)−1 for calculating the two eigenvalues (i.e., λ1 and λ2). Among nonlinear regression models in Matlab, the selected exponential regression model consisted of two exponent terms. To conduct regression analysis with Matlab, on the basis of Equation (12), the regression formula was assumed to be ( β b - 1 ) = d 1 exp - λ 1 t + d 2 exp - λ 2 t .
As shown in Figure 3B, temporal variations of ( β b - 1 ) were replotted with respect to h. Using a curve-fitting toolbox in Matlab (2022a, Mathworks, Natick, MA, USA), values of four unknown constants (i.e., d1, d2, λ1, and λ2) were obtained with respect to h. Figure 3C shows the variations of four unknown constants with respect to h; a1 and a2 did not exhibit consistent trends with respect to h. However, the two eigenvalues (λ1 and λ2) tended to increase substantially with respect to h. The two compliance coefficients (C1 and C2) were obtained by simultaneously solving Equations (13) and (14). Figure 3D shows the variations in C1, C2, and C2/C1 with respect to h. As expected, C2 was considerably greater than C1. However, both C1 and C2 tended to increase substantially with respect to h. The ratio of the two compliance coefficients (C2/C1) did not exhibit a notable trend with respect to h. According to the results, the channel depth was observed to contribute substantially to the increase in the two compliance coefficients.
Instead of glycerin (30%), blood (i.e., normal RBCs in plasma, Hct = 50%) was supplied into the microfluidic channel as the test fluid. The blood viscosity and two compliance coefficients were obtained with respect to the channel depth. To monitor blood flow controlled by a syringe pump, blood velocity (Ub) in the blood channel was obtained using microparticle image velocimetry [31,39,45]. Figure 4(A-i) shows the temporal variations in Ub with respect to h; Ub tended to increase substantially at lower channel depths. Here, during the turn-on blood flow interval, Ub did not exhibit consistent variations with respect to channel depth. According to a previous study [46], blood velocity obtained using the micro-PIV technique is influenced by several factors (i.e., hematocrit, diluent, and flow rate). In addition, as shown in Figure 2(C-ii), blood viscosity tended to increase at higher channel depths. Considering that hematocrit contributes to varying blood viscosity, it was estimated that hematocrit might have an influence on the quantification of blood velocity with respect to channel depth.
Figure 4(A-ii) shows the temporal variations in the interface (αb) with respect to h. After turning on a syringe pump, it took a long time to attain steady blood flow. The αb increased at higher channel depths. Before turning off the syringe pump, at least 30–50 data of interface were selected to calculate blood viscosity. Next, after turning off the syringe pump, two compliance coefficients were obtained by analyzing transient values of αb. Figure 4(A-iii) shows the variations in µb with respect to h; blood viscosity tended to increase gradually with respect to h. Compared with Figure 2(C-ii), plasma resulted in a larger increase in blood viscosity than 1× PBS. Figure 4(A-iv) shows variations in the two compliance coefficients (C1 and C2) with respect to h. Both compliance coefficients tended to increase substantially with respect to the channel depth. The results showed that the channel depth substantially increased the blood viscosity and two compliance coefficients.
The outlet tubing connected to the outlet port was expected to have an influence on compliance. A previous study reported that tubing length contributed to the varying compliance coefficient [37]. In subsequent experiments, channel depth was fixed at 20 µm unless otherwise specified. As shown in Figure 4(B-i), the inlet and outlet tubing was connected to the inlet and outlet, respectively, with Lin and Lout representing the lengths of the inlet and outlet tubing, respectively. The length of the inlet tube was fixed at Lin = 300 mm, and that of the outlet tubing was set to Lout = 200, 300, and 400 mm. Blood (RBCs in plasma, Hct = 50%) was supplied in a periodic on/off pattern (Q0 = 1 mL/h, T = 240 s). Figure 4(B-ii) shows the temporal variations in αb with respect to h. At a steady value of αb, a minor difference in αb with respect to Lout was observed. The blood viscosity was consistently obtained as µb = 2.46 ± 0.07 cP. Using transient variations in αb, two compliance coefficients were obtained with respect to Lout. As shown in Figure 4(B-iii), C1 tended to decrease gradually with respect to Lout; however, C2 remained constant with respect to Lout.
According to the experimental results, the outlet tubing length was found to decrease C1. However, this did not substantially influence C2.

3.3. Contribution of Hematocrit and Diluted Plasma to Two Compliance Coefficients

To quantify the contribution of blood to the two compliance coefficients, several types of blood were prepared by varying the hematocrit or plasma concentration. Subsequently, blood viscosity and two compliance coefficients were obtained with respect to the hematocrit and plasma concentration.
Firstly, to quantify the effect of hematocrit on the two compliance coefficients, the hematocrit of suspended blood was adjusted to Hct = 20%, 30%, 40%, and 50% by adding normal RBCs into autologous plasma. As shown in Figure 2B, because the difference between the nonlinear and approximation terms decreased substantially at higher values of channel depth, the channel depth was set to h = 20 µm in subsequent blood tests. Figure 5(A-i) shows the variations in µb with respect to Hct. Blood viscosity tended to increase significantly with respect to hematocrit. In addition, as shown in Figure 5(A-ii), the two compliance coefficients were obtained with respect to hematocrit. Both compliance coefficients exhibited a distinctive pattern (fluctuation) with respect to hematocrit. According to the results, the blood viscosity was a better indicator than the compliance coefficient for detecting the contribution of hematocrit in suspended blood. Considering that the compliance coefficient was influenced by hematocrit (or blood viscosity), the hematocrit was fixed at Hct = 50% for consistent measurements.
Secondly, to verify the contribution of plasma concentration to blood viscosity and compliance coefficients, autologous plasma was diluted to Cplasma values of 0%, 25%, 50%, 75%, and 100%; Cplasma = 0% and Cplasma = 100% denote pure 1× PBS and autologous plasma, respectively. Suspended blood (Hct = 50%) was prepared by adding normal RBCs to the diluted plasma. Figure 5(B-i) shows the variations in µb with respect to Cplasma. As expected, blood viscosity increased at higher plasma concentrations. Autologous plasma contributed to an increase in the blood viscosity compared with 1× PBS. Figure 5(B-ii) shows the variations in the two compliance coefficients (C1 and C2) with respect to Cplasma; C1 tended to decrease substantially above Cplasma = 50%, and C2 tended to gradually decrease above Cplasma = 25%. In other words, both compliance coefficients decreased substantially above a specific plasma concentration.
On the basis of these results, blood viscosity and compliance coefficients could be used to detect differences in blood (i.e., hematocrit or diluent). To obtain consistent results, it was necessary to fix the hematocrit concentration. Furthermore, for detecting changes in the hematocrit or diluent, the blood viscosity was a better indicator than the two compliance coefficients.

3.4. Detection of Homogenous Hardened RBCs and Heterogeneous RBCs

Furthermore, the present method was employed to detect homogeneous hardened RBCs and heterogeneous RBCs in terms of blood viscosity and two compliance coefficients.
According to a previous study [40], normal RBCs were rigidified with specific concentrations of GA solution (CGA = 0, 2, 4, 6, 8, and 10 µL/mL); RBC rigidity increased substantially when normal RBCs were exposed to higher concentrations of the GA solution. Homogeneous hardened blood (Hct = 50%) was prepared by adding homogeneous hardened RBCs to 1× PBS. As shown in Figure 1(A-i), the flow rates of both the fluids were set to Q0 = 1 mL/h. Figure 6(A-i) shows the variations in µb with respect to CGA; the blood viscosity increased significantly above CGA = 4 µL/mL. These results indicate that rigidified RBCs contributed to the increased blood viscosity. The blood viscosity had the maximum value at the highest concentration of CGA (10 µL/mL). Subsequently, two compliance coefficients were obtained with respect to the degree of RBC rigidity (i.e., the concentration of the GA solution). Figure 6(A-ii) shows the variations in the two compliance coefficients (C1 and C2) with respect to CGA; C1 decreased gradually from CGA = 0 µL/mL to 8 µL/mL, while C2 decreased significantly from CGA = 4 µL/mL to CGA = 8 µL/mL. C1 and C2 remained constant between CGA = 8 µL/mL and CGA = 10 µL/mL. These results indicate that RBC rigidity had a maximum value at a concentration of CGA = 8 µL/mL. The results of the present method are comparable to those of a previous study [40]. Furthermore, the proposed method provided consistent compliance coefficient values (i.e., a smaller standard deviation).
Lastly, the present method was used to detect heterogeneous blood samples. Instead of heterogeneous RBC distributions in the whole blood, this study used the homogeneous normal RBCs and worst deformable RBCs (CGA = 8 µL/mL) to prepare heterogeneous blood. Each RBC sample was loaded into the same diluent of 1× PBS. Two blood suspensions were partially mixed in terms of a volume fraction of each suspension. The volume fraction of hardened blood (ψh) was defined as the ratio of hardened blood volume to total blood volume (i.e., ψh = Vh/[Vh + Vn], where Vh is the hardened blood volume, and Vn is the normal blood volume); it was selected as ψh = 0%, 10%, 20%, 30%, and 100%. Figure 6(B-i) shows the variations in µb with respect to ψh; the blood viscosity tended to increase gradually with respect to ψh. When the volume fraction of the hardened RBCs increased within a certain amount of blood, the blood viscosity tended to increase. Furthermore, two compliance coefficients were obtained with respect to ψh. Figure 6(B-ii) shows the variations in the two compliance coefficients (C1 and C2) with respect to ψh; C1 tended to decrease gradually above ψh = 20%, while C2 tended to decrease substantially between ψh = 0% and ψh = 20%, remaining constant above ψh = 20%. The results indicate that blood viscosity and C2 can be regarded as effective parameters for detecting minor subpopulations in blood.
From the experimental results, it can be concluded that the blood viscosity and two compliance coefficients can be used to effectively detect homogeneous or heterogeneous RBCs.

4. Conclusions

In this study, the blood viscosity and compliance coefficients were quantified simultaneously by analyzing the interfacial location of blood flow in a coflowing channel. Two compliance coefficients (i.e., C1 and C2) were newly suggested to separate the effect of RBC elasticity within the microfluidic system (i.e., soft PDMS device and flexible tubing). Herein, C1 denoted the combined effect of the PDMS tubing and PDMS device, while C2 represented the elastic effect of RBCs. C1 was assumed to be much smaller than C2. Using a discrete fluidic circuit modeling technique, a new governing equation for the microfluidic system was derived and expressed as a second-order nonlinear differential equation. Following approximation procedures, the governing equation became a linear differential equation, and the analytical solution for the governing equation was derived. According to the governing equation, the blood viscosity was obtained under steady blood flow. Next, under transient blood flow, nonlinear regression analysis was conducted to obtain two eigenvalues. Two compliance coefficients were obtained by solving two nonlinear equations. Firstly, according to the linear approximation procedure for the nonlinear term in the governing equation, the constant correction factor (CF0) was obtained as CF0 = 1.183 for a channel depth of h = 20 µm. The linear approximation term was in good agreement with the nonlinear term. Secondly, for glycerin solution (30%) as the test fluid, C2/C1 was estimated to be approximately 10.9–20.3 with respect to channel depth (h = 4, 10, and 20 µm). The channel depth caused an increase in the two compliance coefficients and blood viscosity. However, the outlet tubing length contributed to a decrease in C1. Thirdly, for detecting changes in blood (i.e., hematocrit or diluent), blood viscosity was a better indicator than the two compliance coefficients. Lastly, the present method was adopted to detect homogeneously and heterogeneously hardened RBCs. Consequently, the blood viscosity and two compliance coefficients exhibited substantially different trends with respect to the blood suspensions. In conclusion, the proposed method can be used to effectively detect changes in blood or microfluidic systems. The present method was used to detect artificial heterogeneous RBCs, composed of normal RBCs and partially hardened RBCs. However, in a clinical setting, it is necessary to detect subpopulations, deformability, and density in patient blood [47,48]. In future studies, the present method can contribute to detecting the heterogeneity of RBCs in patient blood.

Funding

This work was supported by the Basic Science Research Program through the NRF funded by the Ministry of Education (NRF-2021R1I1A3040338).

Conflicts of Interest

The author declares no conflict of interest.

References

  1. Spencer, C.; Lip, G. Haemorheological factors in hypertension. J. Hum. Hypertens. 2000, 14, 291–293. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  2. Fu, G.X.; Ji, M.; Han, L.Z.; Xu, C.C.; Pan, F.F.; Hu, T.J.; Zhong, Y. Erythrocyte rheological properties but not whole blood and plasma viscosity are associated with severity of hypertension in older people. Z. Gerontol. Geriatr. 2017, 50, 233–238. [Google Scholar] [CrossRef] [PubMed]
  3. Lee, H.; Na, W.; Lee, S.B.; Ahn, C.W.; Moon, J.S.; Won, K.C.; Shin, S. Potential Diagnostic Hemorheological Indexes for Chronic Kidney Disease in Patients With Type 2 Diabetes. Front. Physiol. 2019, 10, 1062. [Google Scholar] [CrossRef] [PubMed]
  4. Agrawal, R.; Smart, T.; Nobre-Cardoso, J.; Richards, C.; Bhatnagar, R.; Tufail, A.; Shima, D.; Jones, P.H.; Pavesio, C. Assessment of red blood cell deformability in type 2 diabetes mellitus and diabetic retinopathy by dual optical tweezers stretching technique. Sci. Rep. 2016, 6, 15873. [Google Scholar] [CrossRef] [Green Version]
  5. Barabino, G.A.; Platt, M.O.; Kaul, D.K. Sickle cell biomechanics. Annu. Rev. Biomed. Eng. 2010, 12, 345–367. [Google Scholar] [CrossRef]
  6. Proença-Ferreira, R.; Brugnerotto, A.F.V.; Garrido, V.T.; Dominical, V.M.; Vital, D.M.; Ribeiro, M.d.F.t.R.; Santos, M.E.d.; Traina, F.o.; Olalla-Saad, S.T.; Costa, F.F.; et al. Endothelial activation by platelets from sickle cell anemia patients. PLoS ONE 2014, 9, e89012. [Google Scholar] [CrossRef]
  7. Kang, Y.J.; Ha, Y.-R.; Lee, S.-J. High-throughput and label-free blood-on-a-chip for malaria diagnosis. Anal. Chem. 2016, 88, 2912–2922. [Google Scholar] [CrossRef] [Green Version]
  8. Guo, Q.; Duffy, S.P.; Matthews, K.; Deng, X.; Santoso, A.T.; Islamzada, E.; Ma, D.H. Deformability based sorting of red blood cells improves diagnostic sensitivity for malaria caused by Plasmodium falciparum. Lab Chip 2016, 16, 645–654. [Google Scholar] [CrossRef]
  9. Jalal, U.M.; Kim, S.C.; Shim, J.S. Histogram analysis for smartphone-based rapid hematocrit determination. Biomed. Opt. Express 2017, 8, 3317–3328. [Google Scholar] [CrossRef] [Green Version]
  10. Zhbanov, A.; Yang, S. Electrochemical impedance spectroscopy of blood for sensitive detection of blood hematocrit, sedimentation and dielectric properties. Anal. Methods 2017, 9, 3302–3313. [Google Scholar] [CrossRef]
  11. Baskurt, O.K.; Meiselman, H.J. Blood rheology and hemodynamics. Semin. Thromb. Hemost. 2003, 29, 435–450. [Google Scholar]
  12. Tomaiuolo, G. Biomechanical properties of red blood cells in health and disease towards microfluidics. Biomicrofluidics 2014, 8, 051501. [Google Scholar] [CrossRef] [Green Version]
  13. Guzman-Sepulveda, J.R.; Batarseh, M.; Wu, R.; Decampli, W.M.; Dogariu, A. Passive high-frequency microrheology of blood. Soft Matter 2022, 18, 2452–2461. [Google Scholar] [CrossRef]
  14. Salipante, P.F. Microfluidic techniques for mechanical measurements of biological samples. Biophys. Rev. 2023, 4, 011303. [Google Scholar] [CrossRef]
  15. Prado, G.; Farutin, A.; Misbah, C.; Bureau, L. Viscoelastic transient of confined red blood cells. Biophys. J. 2015, 108, 2126–2136. [Google Scholar] [CrossRef] [Green Version]
  16. Tomaiuolo, G.; Carciati, A.; Caserta, S.; Guido, S. Blood linear viscoelasticity by small amplitude oscillatory flow. Rheol. Acta 2016, 55, 485–495. [Google Scholar] [CrossRef] [Green Version]
  17. Campo-Deano, L.; Dullens, R.P.A.; Aarts, D.G.A.L.; Pinho, F.T.; Oliveira, M.S.N. Viscoelasticity of blood and viscoelastic blood analogues for use in polydymethylsiloxane in vitro models of the circulatory system. Biomicrofluidics 2013, 7, 034102. [Google Scholar] [CrossRef] [Green Version]
  18. Giudice, F.D. A review of microfluidic devices for rheological characterisation. Micromachines 2022, 13, 167. [Google Scholar] [CrossRef]
  19. Rodrigues, T.; Mota, R.; Gales, L.; Campo-Deaño, L. Understanding the complex rheology of human blood plasma. J. Rheol. 2022, 66, 761–774. [Google Scholar] [CrossRef]
  20. Torrisi, F.; Stella, G.; Guarino, F.M.; Bucolo, M. Cell counting and velocity algorithms for hydrodynamic study of unsteady biological flows in micro-channels. Biomicrofluidics 2023, 17, 014105. [Google Scholar] [CrossRef]
  21. Chen, L.; Li, D.; Liu, X.; Xie, Y.; Shan, J.; Huang, H.; Yu, X.; Chen, Y.; Zheng, W.; Li, Z. Point-of-Care Blood Coagulation Assay Based on Dynamic Monitoring of Blood Viscosity Using Droplet Microfluidics. ACS Sens. 2022, 7, 2170–2177. [Google Scholar] [CrossRef]
  22. Li, Y.; Ward, K.R.; Burns, M.A. Viscosity measurements using microfluidic droplet length. Anal. Chem. 2017, 89, 3996–4006. [Google Scholar] [CrossRef]
  23. Mena, S.E.; Li, Y.; McCormick, J.; McCracken, B.; Colmenero, C.; Ward, K.; Burns, M.A. A droplet-based microfluidic viscometer for the measurement of blood coagulation. Biomicrofluidics 2020, 14, 014109. [Google Scholar] [CrossRef]
  24. Khnouf, R.; Karasneh, D.; Abdulhay, E.; Abdelhay, A.; Sheng, W.; Fan, Z.H. Microfluidics-based device for the measurement of blood viscosity and its modeling based on shear rate, temperature, and heparin concentration. Biomed. Microdevices 2019, 21, 80. [Google Scholar] [CrossRef]
  25. Oh, S.; Kim, B.; Lee, J.K.; Choi, S. 3D-printed capillary circuits for rapid, low-cost, portable analysis of blood viscosity. Sens. Actuator B-Chem. 2018, 259, 106–113. [Google Scholar] [CrossRef]
  26. Solomon, D.E.; Abdel-Raziq, A.; Vanapalli, S.A. A stress-controlled microfluidic shear viscometer based on smartphone imaging. Rheol. Acta 2016, 55, 727–738. [Google Scholar] [CrossRef]
  27. Kim, B.J.; Lee, Y.S.; Zhbanov, A.; Yang, S. A physiometer for simultaneous measurement of whole blood viscosity and its determinants: Hematocrit and red blood cell deformability. Analyst 2019, 144, 3144–3157. [Google Scholar] [CrossRef]
  28. Kang, Y.J.; Yoon, S.Y.; Lee, K.-H.; Yang, S. A highly accurate and consistent microfluidic viscometer for continuous blood viscosity measurement. Artif. Organs 2010, 34, 944–949. [Google Scholar] [CrossRef]
  29. Hintermüller, M.A.; Offenzeller, C.; Jakoby, B. A microfluidic viscometer with capacitive readout using screen-printed electrodes. IEEE Sens. J. 2021, 21, 2565–2572. [Google Scholar] [CrossRef]
  30. Kang, Y.J.; Ryu, J.; Lee, S.-J. Label-free viscosity measurement of complex fluids using reversal flow switching manipulation in a microfluidic channel. Biomicrofluidics 2013, 7, 044106. [Google Scholar] [CrossRef] [Green Version]
  31. Kang, Y.J. Assessment of blood biophysical properties using pressure sensing with micropump and microfluidic comparator. Micromachines 2022, 13, 483. [Google Scholar] [CrossRef] [PubMed]
  32. Tomaiuolo, G.; Barra, M.; Preziosi, V.; Cassinese, A.; Rotoli, B.; Guido, S. Microfluidics analysis of red blood cell membrane viscoelasticity. Lab Chip 2011, 11, 449–454. [Google Scholar] [CrossRef] [PubMed]
  33. Kang, Y.J.; Lee, S.-J. Blood viscoelasticity measurement using steady and transient flow controls of blood in a microfluidic analogue of Wheastone-bridge channel. Biomicrofluidics 2013, 7, 054122. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  34. Isiksacan, Z.; Serhatlioglu, M.; Elbuken, C. In vitro analysis of multiple blood flow determinants using red blood cell dynamics under oscillatory flow. Analyst 2020, 145, 5996–6005. [Google Scholar] [CrossRef]
  35. Kang, Y.J. Continuous and simultaneous measurement of the biophysical properties of blood in a microfluidic environment. Analyst 2016, 141, 6583–6597. [Google Scholar] [CrossRef]
  36. Kang, Y.J. Biosensing of Haemorheological Properties Using Microblood Flow Manipulation and Quantification. Sensors 2022, 23, 408. [Google Scholar] [CrossRef]
  37. Hébert, M.; Huissoon, J.; Ren, C.L. A quantitative study of the dynamic response of compliant microfluidic chips in a microfluidics context. J. Micromech. Microeng. 2022, 32, 085004. [Google Scholar] [CrossRef]
  38. OTSU, N. A tlreshold selection method from gray-level histograms. IEEE Trans. Syst. Man. Cybern. 1979, 9, 62–66. [Google Scholar] [CrossRef] [Green Version]
  39. Thielicke, W.; Stamhuis, E.J. PIVlab–Towards user-friendly, affordable and accurate digital particle image velocimetry in MATLAB. J. Open Res. Softw. 2014, 2, e30. [Google Scholar] [CrossRef] [Green Version]
  40. Kang, Y.J. Simultaneous measurement of erythrocyte deformability and blood viscoelasticity using micropillars and co-flowing streams under pulsatile blood flows. Biomicrofluidics 2017, 11, 014102. [Google Scholar] [CrossRef] [Green Version]
  41. Kang, Y.J. Periodic and simultaneous quantification of blood viscosity and red blood cell aggregation using a microfluidic platform under in-vitro closed-loop circulation. Biomicrofluidics 2018, 12, 024116. [Google Scholar] [CrossRef]
  42. Kang, Y.J. Quantitative monitoring of dynamic blood flows using coflowing laminar streams in a sensorless approach. App. Sci.-Basel 2021, 11, 7260. [Google Scholar] [CrossRef]
  43. Kang, Y.J. Blood viscoelasticity measurement using interface variations in coflowing streams under pulsatile blood flows. Micromachines 2020, 11, 245. [Google Scholar] [CrossRef] [Green Version]
  44. Cheng, N.-S. formula for the viscosity of a glycerin-water mixture. Ind. Eng. Chem. Res. 2008, 47, 3285–3288. [Google Scholar] [CrossRef]
  45. Pitts, K.L.; Mehri, R.; Mavriplis, C.; Fenech, M. Micro-particle image velocimetry measurement of blood flow: Validation and analysis of data pre-processing and processing methods. Meas. Sci. Technol. 2012, 23, 105302. [Google Scholar] [CrossRef]
  46. Kang, Y.J. Microfluidic-based biosensor for blood viscosity and erythrocyte sedimentation rate using disposable fluid delivery system. Micromachines 2020, 11, 215. [Google Scholar] [CrossRef] [Green Version]
  47. Huisjes, R.; Makhro, A.; Llaudet-Planas, E.; Hertz, L.; Petkova-Kirova, P.; Verhagen, L.P.; Pignatelli, S.; Rab, M.A.; Schiffelers, R.M.; Seiler, E.; et al. Density, heterogeneity and deformability of red cells as markers of clinical severity in hereditary spherocytosis. Haematologica 2020, 105, 338–347. [Google Scholar] [CrossRef]
  48. Bogdanova, A.; Kaestner, L.; Simionato, G.; Makhro, A. Heterogeneity of red blood cells: Causes and consequences. Front. Physiol. 2020, 11, 392. [Google Scholar] [CrossRef]
Figure 1. Proposed method for measurement of blood viscoelasticity under microcapillary blood flow. (A) Schematic diagram of experimental setup, including a microfluidic device, two syringe pumps, and image acquisition system. (i) Microfluidic device consisting of two inlets (a, b), an outlet, and microfluidic channels (i.e., reference channel, blood channel, and coflowing channel). Using two syringe pumps, blood flow rate (Qb) set to a periodic on/off profile (i.e., amplitude: Q0, and period: T). Flow rate of reference fluid set to a constant flow rate (i.e., Qr = Q0). The corresponding length of each polyethylene tubing is denoted as Lin (for inlet tubing) and Lout (for outlet tubing). (ii) Image acquisition system for capturing blood flow in the microfluidic channels. With high-speed camera set to 5000 fps, microscopic images are sequentially captured at an interval of 0.5 s. (iii) Microscopic image showing interface (αb) in the coflowing channel and blood velocity (Ub) in the blood channel. (B) Mathematical representation with discrete fluidic circuit elements for estimating viscoelasticity. The microfluidic system (i.e., two fluids, flexible microfluidic device, and polyethylene tubing) is modeled as discrete fluidic elements, including flow rate (Qr and Qb), fluidic resistance (Rr1, Rr2, Rb1, and Rb2), and compliance (C1 and C2). Here, ‘►’ denotes the zero value of pressure (P = 0). (C) As a preliminary demonstration, control blood (Hct = 50%) was prepared by adding normal RBCs to autologous plasma, with flow rate of blood set to on/off pattern (i.e., Q0 = 1 mL/h, T = 240 s), and flow rate of reference fluid set to Qr = 1 mL/h. (i) Temporal variations of αb for up to 1000 s. (ii) Quantification of blood viscosity using steady variations of αb after turning on syringe pump. (iii) Temporal variations of αb and βb = (1 − αb)−1 after turning off syringe pump. Two compliances were obtained by conducting nonlinear regression of transient behaviors of βb (i.e., βb = d1 exp [−λ1t] + d2 exp [−λ2t] +1).
Figure 1. Proposed method for measurement of blood viscoelasticity under microcapillary blood flow. (A) Schematic diagram of experimental setup, including a microfluidic device, two syringe pumps, and image acquisition system. (i) Microfluidic device consisting of two inlets (a, b), an outlet, and microfluidic channels (i.e., reference channel, blood channel, and coflowing channel). Using two syringe pumps, blood flow rate (Qb) set to a periodic on/off profile (i.e., amplitude: Q0, and period: T). Flow rate of reference fluid set to a constant flow rate (i.e., Qr = Q0). The corresponding length of each polyethylene tubing is denoted as Lin (for inlet tubing) and Lout (for outlet tubing). (ii) Image acquisition system for capturing blood flow in the microfluidic channels. With high-speed camera set to 5000 fps, microscopic images are sequentially captured at an interval of 0.5 s. (iii) Microscopic image showing interface (αb) in the coflowing channel and blood velocity (Ub) in the blood channel. (B) Mathematical representation with discrete fluidic circuit elements for estimating viscoelasticity. The microfluidic system (i.e., two fluids, flexible microfluidic device, and polyethylene tubing) is modeled as discrete fluidic elements, including flow rate (Qr and Qb), fluidic resistance (Rr1, Rr2, Rb1, and Rb2), and compliance (C1 and C2). Here, ‘►’ denotes the zero value of pressure (P = 0). (C) As a preliminary demonstration, control blood (Hct = 50%) was prepared by adding normal RBCs to autologous plasma, with flow rate of blood set to on/off pattern (i.e., Q0 = 1 mL/h, T = 240 s), and flow rate of reference fluid set to Qr = 1 mL/h. (i) Temporal variations of αb for up to 1000 s. (ii) Quantification of blood viscosity using steady variations of αb after turning on syringe pump. (iii) Temporal variations of αb and βb = (1 − αb)−1 after turning off syringe pump. Two compliances were obtained by conducting nonlinear regression of transient behaviors of βb (i.e., βb = d1 exp [−λ1t] + d2 exp [−λ2t] +1).
Micromachines 14 00814 g001
Figure 2. Contribution of channel depth (h) to blood viscosity under steady flow. (A) Quantification of correction factor with respect to channel depth. (i) Microscopic image showing interface of two fluids (i.e., test fluid: glycerin solution [30%], reference fluid: 1× PBS) in the coflowing channel, at the same flow rate (i.e., Qr = Qb = 1 mL/h); αt denotes the interface. (ii) Variations of correction factor (CF) with respect to channel depth (h = 4, 10, and 20 µm) and interface (αt). (B) Approximation of αt/(1 − αt) × CF (αt) as αt/(1 − αt) × CF0 with respect to channel depth (h) ([i] CF0 = 1.637 for h = 4 µm, [ii] CF0 = 1.41 for h = 10 µm, and [iii] CF0 = 1.183 for h = 20 µm). (C) Variations of blood viscosity (µb) with respect to channel width. (i) Microscopic images showing interface with respect to h = 4 and 20 µm. (ii) Contribution to channel depth to µb. Blood (Hct = 50%) was prepared by adding normal RBCs into 1× PBS. The inset shows variations of µb with respect to h. (iii) Variations in viscosity of blood (plasma, normal RBCs in plasma, and normal RBCs in 1× PBS) with respect to shear rate. Here, channel depth was set to 20 µm. Hematocrit was fixed to 50%.
Figure 2. Contribution of channel depth (h) to blood viscosity under steady flow. (A) Quantification of correction factor with respect to channel depth. (i) Microscopic image showing interface of two fluids (i.e., test fluid: glycerin solution [30%], reference fluid: 1× PBS) in the coflowing channel, at the same flow rate (i.e., Qr = Qb = 1 mL/h); αt denotes the interface. (ii) Variations of correction factor (CF) with respect to channel depth (h = 4, 10, and 20 µm) and interface (αt). (B) Approximation of αt/(1 − αt) × CF (αt) as αt/(1 − αt) × CF0 with respect to channel depth (h) ([i] CF0 = 1.637 for h = 4 µm, [ii] CF0 = 1.41 for h = 10 µm, and [iii] CF0 = 1.183 for h = 20 µm). (C) Variations of blood viscosity (µb) with respect to channel width. (i) Microscopic images showing interface with respect to h = 4 and 20 µm. (ii) Contribution to channel depth to µb. Blood (Hct = 50%) was prepared by adding normal RBCs into 1× PBS. The inset shows variations of µb with respect to h. (iii) Variations in viscosity of blood (plasma, normal RBCs in plasma, and normal RBCs in 1× PBS) with respect to shear rate. Here, channel depth was set to 20 µm. Hematocrit was fixed to 50%.
Micromachines 14 00814 g002
Figure 3. Contribution of channel depth to compliance. (A) Temporal variations of αt with respect to channel depth (h) (h = 4, 10, and 20 µm). (B) Variation of (βt − 1) with respect to h. On the basis of the relationship βt = (1 − αt)−1, the nonlinear regression formula was assumed as (βt − 1) = d1 exp (–λ1t) + d2 exp (–λ2t). (C) Variations of four constants (i.e., d1, d2, λ1, and λ2) with respect to h. (D) Variations of C1, C2, and C2/C1 with respect to h.
Figure 3. Contribution of channel depth to compliance. (A) Temporal variations of αt with respect to channel depth (h) (h = 4, 10, and 20 µm). (B) Variation of (βt − 1) with respect to h. On the basis of the relationship βt = (1 − αt)−1, the nonlinear regression formula was assumed as (βt − 1) = d1 exp (–λ1t) + d2 exp (–λ2t). (C) Variations of four constants (i.e., d1, d2, λ1, and λ2) with respect to h. (D) Variations of C1, C2, and C2/C1 with respect to h.
Micromachines 14 00814 g003
Figure 4. Contributions of channel depth (h) and outlet tubing length (Lout) to blood viscosity (µb) and two compliances (C1 and C2). Blood (Hct = 50%) was prepared by adding normal RBCs into autologous plasma. (A) Contribution of channel depth (h) to two compliances (C1 and C2). (i) Temporal variations of blood velocity (Ub) with respect to channel depth (h) (h = 4, 10, and 20 µm). (ii) Temporal variations of interface (αb) with respect to h. (iii) Variations of blood viscosity (µb) with respect to h. (iv) Variations of two compliances (C1 and C2) with respect to h. (B) Contribution of outlet tubing length (Lout) to two compliances (C1 and C2). (i) Schematic of a microfluidic device connected with three kinds of tubing. Lin and Lout represent the length of inlet tubing and length of outlet tubing. The length of inlet tubing was fixed at Lin = 300 mm. The length of outlet tubing set to Lout = 200, 300, and 400 mm. (ii) Temporal variations of interface (αb) with respect to h. (iii) Variations of two compliances (C1 and C2) with respect to Lout.
Figure 4. Contributions of channel depth (h) and outlet tubing length (Lout) to blood viscosity (µb) and two compliances (C1 and C2). Blood (Hct = 50%) was prepared by adding normal RBCs into autologous plasma. (A) Contribution of channel depth (h) to two compliances (C1 and C2). (i) Temporal variations of blood velocity (Ub) with respect to channel depth (h) (h = 4, 10, and 20 µm). (ii) Temporal variations of interface (αb) with respect to h. (iii) Variations of blood viscosity (µb) with respect to h. (iv) Variations of two compliances (C1 and C2) with respect to h. (B) Contribution of outlet tubing length (Lout) to two compliances (C1 and C2). (i) Schematic of a microfluidic device connected with three kinds of tubing. Lin and Lout represent the length of inlet tubing and length of outlet tubing. The length of inlet tubing was fixed at Lin = 300 mm. The length of outlet tubing set to Lout = 200, 300, and 400 mm. (ii) Temporal variations of interface (αb) with respect to h. (iii) Variations of two compliances (C1 and C2) with respect to Lout.
Micromachines 14 00814 g004
Figure 5. Contributions of hematocrit and diluted plasma to blood viscosity (µb) and two compliances (C1 and C2). Hematocrit of suspended blood was adjusted by adding normal RBCs into autologous plasma. Autologous plasma was diluted with 1× PBS. Channel depth was fixed at h = 20 µm. (A) Contributions of hematocrit (Hct = 20%, 30%, 40%, and 50%) to blood viscosity (µb) and two compliances (C1 and C2). (i) Variations of µb with respect to Hct. (ii) Variations of two compliances (C1 and C2) with respect to Hct. (B) Contributions of diluted plasma to blood viscosity (µb) and two compliances (C1 and C2). (i) Variations of µb with respect to plasma concentration (Cplasma) (Cplasma = 0%, 25%, 50%, 75%, and 100%). Cplasma = 0% and Cplasma = 100% denote pure 1× PBS and autologous plasma, respectively. (ii) Variations of two compliances (C1 and C2) with respect to Cplasma.
Figure 5. Contributions of hematocrit and diluted plasma to blood viscosity (µb) and two compliances (C1 and C2). Hematocrit of suspended blood was adjusted by adding normal RBCs into autologous plasma. Autologous plasma was diluted with 1× PBS. Channel depth was fixed at h = 20 µm. (A) Contributions of hematocrit (Hct = 20%, 30%, 40%, and 50%) to blood viscosity (µb) and two compliances (C1 and C2). (i) Variations of µb with respect to Hct. (ii) Variations of two compliances (C1 and C2) with respect to Hct. (B) Contributions of diluted plasma to blood viscosity (µb) and two compliances (C1 and C2). (i) Variations of µb with respect to plasma concentration (Cplasma) (Cplasma = 0%, 25%, 50%, 75%, and 100%). Cplasma = 0% and Cplasma = 100% denote pure 1× PBS and autologous plasma, respectively. (ii) Variations of two compliances (C1 and C2) with respect to Cplasma.
Micromachines 14 00814 g005
Figure 6. Detection of homogeneous hardened RBCs and heterogeneous hardened RBCs with blood viscosity and two compliances. (A) Detection of homogeneous hardened blood composed of RBCs hardened with the same concentrations of GA solution. Normal RBCs were hardened using the specific concentrations of GA solution. After that, hematocrit of blood was adjusted to 50% by adding hardened RBCs into 1× PBS. (i) Variations of µb with respect to concentrations of GA solution (CGA) (CGA = 0, 2, 4, 6, 8, and 10 µL/mL). (ii) Variations of two compliances (C1 and C2) with respect to CGA. (B) Detection of heterogeneous blood composed of normal blood and hardened blood. Hardened blood was prepared by adding hardened RBCs (CGA = 8 µL/mL) into PBS solution. Specifically, the volume fraction of hardened blood (ψh) was defined as the ratio of hardened blood volume to total blood volume (i.e., ψh = Vh/[Vh + Vn], where Vh is the hardened blood volume, and Vn is the normal blood volume). (i) Variations of µb with respect to ψh = 0%, 10%, 20%, 30%, and 100%. (ii) Variations of two compliances (C1 and C2) with respect to ψh.
Figure 6. Detection of homogeneous hardened RBCs and heterogeneous hardened RBCs with blood viscosity and two compliances. (A) Detection of homogeneous hardened blood composed of RBCs hardened with the same concentrations of GA solution. Normal RBCs were hardened using the specific concentrations of GA solution. After that, hematocrit of blood was adjusted to 50% by adding hardened RBCs into 1× PBS. (i) Variations of µb with respect to concentrations of GA solution (CGA) (CGA = 0, 2, 4, 6, 8, and 10 µL/mL). (ii) Variations of two compliances (C1 and C2) with respect to CGA. (B) Detection of heterogeneous blood composed of normal blood and hardened blood. Hardened blood was prepared by adding hardened RBCs (CGA = 8 µL/mL) into PBS solution. Specifically, the volume fraction of hardened blood (ψh) was defined as the ratio of hardened blood volume to total blood volume (i.e., ψh = Vh/[Vh + Vn], where Vh is the hardened blood volume, and Vn is the normal blood volume). (i) Variations of µb with respect to ψh = 0%, 10%, 20%, 30%, and 100%. (ii) Variations of two compliances (C1 and C2) with respect to ψh.
Micromachines 14 00814 g006
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Kang, Y.J. Quantification of Blood Viscoelasticity under Microcapillary Blood Flow. Micromachines 2023, 14, 814. https://doi.org/10.3390/mi14040814

AMA Style

Kang YJ. Quantification of Blood Viscoelasticity under Microcapillary Blood Flow. Micromachines. 2023; 14(4):814. https://doi.org/10.3390/mi14040814

Chicago/Turabian Style

Kang, Yang Jun. 2023. "Quantification of Blood Viscoelasticity under Microcapillary Blood Flow" Micromachines 14, no. 4: 814. https://doi.org/10.3390/mi14040814

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop