Article

Potentiometric MIP-Modified Screen-Printed Cell for Phenoxy Herbicides Detection

1
Department of Chemistry, University of Pavia, Via Taramelli 12, 27100 Pavia, Italy
2
Unità di Ricerca di Pavia, INSTM, Via G. Giusti 9, 50121 Firenze, Italy
3
Departamento de Química, Facultad de Ciencias, Universidad de Burgos, Plaza de Misael Bañuelos s/n, 09001 Burgos, Spain
*
Author to whom correspondence should be addressed.
Int. J. Environ. Res. Public Health 2022, 19(24), 16488; https://doi.org/10.3390/ijerph192416488
Submission received: 26 October 2022 / Revised: 30 November 2022 / Accepted: 7 December 2022 / Published: 8 December 2022

Abstract

:
In this study, a molecularly imprinted polymer (MIP)-based screen-printed cell is developed for detecting phenoxy herbicides using 2-methyl-4-chlorophenoxyacetic acid (MCPA) as the template. MCPA is a phenoxy herbicide widely used since 1945 to control broadleaf weeds via growth regulation, primarily in pasture and cereal crops. The potentiometric cell consists of a silver/silver chloride pseudo-reference electrode and a graphite working electrode coated with a MIP film. The polymeric layer is thermally formed after drop-coating of a pre-polymeric mixture composed of the reagents at the following molar ratio: 1 MCPA: 15 MAA (methacrylic acid): 7 EGDMA (ethylene glycol dimethacrylate). After template removal, the recognition cavities function as the ionophore of a classical ion selective electrode (ISE) membrane. The detected ion is the deprotonated MCPA specie, negatively charged, so the measurements were performed in phosphate buffer at pH 5.5. A linear decrease of the potential with MCPA concentration, ranging from 4 × 10−8 to 1 × 10−6 mol L−1, was obtained. The detection limit and the limit of quantification were, respectively, 10 nmol L−1 and 40 nmol L−1. A Nernstian slope of about −59 mV/dec was achieved. The method has precision and LOD required for MCPA determination in contaminated environmental samples.

1. Introduction

In the last few years, the use of pesticides has grown due to the demand of the market that claims more and more abundant and leafy products. The first synthetic herbicides were produced during the Second World War to remove crop weeds. The use of herbicides can be helpful for the agricultural field, but they may degrade and contaminate the environment. These pollutants can cause human diseases, so detecting and quantifying them is crucial [1].
MCPA (2-methyl-4-chlorophenoxyacetic acid) was the first synthetic herbicide synthesized in England around 1941. It is a selective phenoxy herbicide that controls broadleaf weeds in arable and cereal crops, acting on growth hormones. MCPA can mimic auxin, encouraging uncontrolled growth and subsequent death of certain plants, such as dicotyledons. It is synthesized by exposing 4-chloro-2-methylphenol to chloroacetic acid in diluted alkaline solutions [2]. MCPA has a soil half-life of 24 days and a water-octanol partition coefficient of 0.2–1 mL/g, but some environmental pH and temperature values may facilitate the sorption and the degradation [3]. Interestingly but alarmingly, MCPA and similar phenoxy herbicides can form stable complexes with toxic metal ions, such as lead and cadmium, in environmental pH ranges, increasing their bioavailability; for this reason, it is fundamental to detect these pesticides in the environment [4].
As far as toxicity is concerned, these herbicides, and in particular the most common MCPA, are approved for use in all EU countries (except Cyprus), the United States, China, Canada, Australia, and New Zealand [5] since their toxicity depends on the species affected and the method of exposure to the pesticide. In fact, for example, MCPA has a very low bioconcentration factor (BCF) 1–14 [6], which means that the bioaccumulation rate is low, and if excessive exposure to the poison kills an organism, it is not dangerous for other organisms [7]. The human symptoms due to MCPA poisoning are fatigue, weakness, anoxia, nausea, vomiting, diarrhea, lowering of blood pressure, body temperature disturbance, progressive hypotension, ataxia, neuromuscular inability, and convulsion [8], and the human lethal oral concentration is around 814 mg/Kg [9].
Moving to phenoxy herbicides detection, the literature reports many chromatographic [10,11,12,13,14] and voltammetric [15,16,17,18,19] methods, but no potentiometric ones are described. Potentiometric and, more in general, all electrochemical techniques have several advantages compared to the classic ones: the instrumentation is less expensive, the time of analysis is low, the apparatus may be portable, the sensors can be used for in situ measurements, and the surface of the electrodes can be modified in order to obtain more specific and sensitive detections. In particular, the electrodic surface can be activated with different kinds of modifiers such as multi-Walled carbon nano tubes (MWCNTs) [20,21,22,23,24,25], metal nano particles (NPs) [26,27,28,29,30,31], graphene [32,33,34,35,36,37], metal organics frameworks (MOFs) [38,39,40,41,42,43] in order to enhance the sensitivity, aptamers, antibodies [44,45,46,47,48,49], enzymes [50,51,52,53,54,55], or molecularly imprinted polymers (MIPs) [56,57,58,59,60,61] to improve the selectivity.
Among these receptors for electrode surface modification, MIPs represent an emerging, interesting topic in polymeric chemistry that has recently caught on since these polymers can be used as synthetic receptors, replacing biomolecules such as enzymes, aptamers, and others [62]. MIPs are characterized by the presence of selective cavities meant to sorb a target molecule; specific monomers must be chosen to obtain these cavities since they have to interact through weak bonds with a template molecule, i.e., the specific analyte. Once the polymerization is completed, the template molecule must be removed to obtain a selective cavity with the shape of the analyte [63]. Figure 1 reports the schematic procedure to obtain a molecularly imprinted polymer:
MIPs offer several advantages: they can be designed selectively for the target molecule using different monomers, the monomers used are usually low-cost, reusable, and easily storable, and conductive polymers can also be obtained [64]. Conversely, enzymes and other biological receptors are costly, must be stored in certain conditions, and, generally, biosensors have a short lifetime [65].
In this context, the present study aims to develop a MIP-based screen-printed potentiometric cell to detect phenoxy herbicides using MCPA as a template molecule. The sensor is prepared according to a previously developed procedure [56] schematized in Figure 2. The MIP-modified working electrode acts as an ion-selective electrode, so the solution pH required for the measurements has to be selected in a range in which MCPA is negatively charged. The pH value chosen is 5.5 according to the MCPA pKa of 3.07 [66]. The method is tested on synthetic solutions and applied to fortified natural water samples.

2. Materials and Methods

2.1. Reagents and Materials

Methacrylic acid (MAA) (Sigma-Aldrich, Milan, Italy) and ethylene glycol dimethacrylate (EGDMA) (Sigma-Aldrich) were filtered by using a column of aluminum oxide (Sigma-Aldrich) in order to remove stabilizers. 2,2-azobisisobutyronitrile (AIBN), MCPA, Mecoprop, Dichloroprop, 2,4-D Pestanal, acetic acid, and ethanol were used as received from Sigma-Aldrich (Milan, Italy). Phosphate buffer solutions (PBS buffer) were prepared in ultrapure water, adjusting the pH with hydrochloric acid or sodium hydroxide (Sigma-Aldrich, Milan, Italy). Solutions for electrode surface characterization were prepared from sodium chloride, potassium chloride, and potassium hexacyanoferrate(III) (Sigma-Aldrich, Milan, Italy). Tap water samples were obtained from the lab sink (Department of Chemistry, University of Pavia). Three-electrodes screen-printed cells with a graphite-ink working electrode, a graphite-ink counter electrode, and Ag/AgCl-ink pseudo-reference electrode were obtained by Topflight Italia SPA (Vidigulfo, Pavia, Italy).

2.2. Instruments

The ultrasonic bath AU-32 Argo Lab (Tecno-Lab, Milan, Italy), with ultrasound power 120 W, was employed for the pre-polymeric mixture preparation.
pH measurements were performed with a pH-meter Mettler Toledo mod. SevenMulti, equipped with a combined glass electrode InLab Pro (Mettler Toledo, Milan, Italy).
Potentiometric analyses and impedance spectroscopy measures were performed with the potentiostat/galvanostat EmStat4s-PalmSens BV (Houten-The Netherlands https://www.palmsens.com/product/emstat4s/ (accessed on 7 June 2022)).
Scanning electron microscopy (SEM) images were acquired with EVO MA10 scanning electron microscope. The measurements were performed under an ultra-high vacuum with an electron generation voltage of 20 kV and a working distance of 8.5 mm.

2.3. MIPs Pre-Polymeric Mixture

The pre-polymeric mixture was prepared by dissolving 28 mg of MCPA (0.139 mmol) with 175 μL of MAA (2.07 mmol), 175 μL of EGDMA (0.927 mmol), AIBN 35 mg, and 200 μL of toluene. The so obtained pre-polymeric mixture (with a molar ratio of 1:15:7 = MCPA:MAA:EGDMA) was sonicated for 30 min and then deaerated for 3 min with N2.
The pre-polymeric mixture for the non-imprinted polymer (NIP) was prepared following the same procedure but without adding the template MCPA.

2.4. Working Electrode Modification

Each screen-printed cell (SPC) was washed with ethanol and dried with N2. Then, 7 μL of pre-polymeric mixture was drop-coated on the working electrode (WE) surface. The SPC was then placed in an oven at 70 °C for 12 h for thermo-induced polymerization. After polymerization, the SPC was washed, immersing the SPC in 10 mL of a mixture of ethanol:acetic acid = 4:1 under gentle stirring. Five cycles of 30 min each to remove the template and the unreacted monomers as checked by analyzing the UV–vis spectrum of the mother waters at each washing step. Before use, the sensor was placed in ultrapure water for 10 min.
Screen-printed electrodes are disposable, so they do not require particular maintenance procedures. After modification with the MIP/NIP layer, no more than two experiments (calibration or sample analysis) can be performed with the same electrode without compromising the measurements.

2.5. Electrochemical Impedance Spectroscopy (EIS)

For the electrodic surface characterization, electrochemical impedance spectroscopy (EIS) measurements were performed using the potentiostat/galvanostat EmStat4s (EmStat4s-PalmSens BV (Houten-The Netherlands), using as electrochemical probe a solution of 5 mmol L−1 K4Fe(CN)6/0.1 mol L−1 KCl at pH 7.2 and the following instrumental parameters: frequency range from 100 kHz to 10 mHz and signal amplitude of 50 mV. EIS measurements were performed on the washed MIP-modified electrodes, NIP-modified electrodesm and the MIP-modified electrode after immersion in 10−7 mol L−1 MCPA solution.

2.6. Potentiometric Measurements

Potentiometric measurements were performed by dipping the washed MIP-modified screen-printed cell in 15 mL of PBS buffer at pH 5.5 and adding portions of 20 μL of MCPA standard solution 25 nmol L−1 under gentle stirring. At each addition of MCPA, the potential was recorded, and the average of the values sampled in the last 60 s after waiting 3 min for the steady-state was considered the response signal (stability criterium ΔE/Δt = 0.02 mV/s, see Figure S1 in Supplementary Materials). Figure S2 shows the experimental setup (see Supplementary Materials).

3. Results

3.1. Working Electrode Modification by a MIP Layer and Characterization Using EIS

The surface of the working electrode was functionalized by a MIP layer to obtain a selective sensor. The pre-polymeric mixture molar ratio was defined according to previous studies [56]. The volume selected for drop-coating was 7 µL, i.e., a quantity enough to cover the working electrode and keep the pseudo-reference and the counter electrodes unaltered. MAA was chosen as the functional monomer to obtain the best imprinting features since its carboxylic groups can interact with the template molecule through hydrogen bonds. EGDMA was employed as the crosslinker, toluene as the porogenic solvent, and AIBN as the radical initiator. The formation of the MIP cavities was verified by SEM measurement on MIP and NIP powders obtained by polymerization in bulk (see Supplementary Materials, Figure S3). For the MIP material, filamentous and compact particles of 10/20 microns appeared; conversely, only larger and uniform particles can be observed for NIP.
Electrochemical impedance spectroscopy (EIS) is a useful technique for characterizing the surface of a modified electrode [67]; indeed, EIS can help to understand the changing of the conducting properties of a modified electrodic surface, schematizing the phenomena that are taking place at the interface electrode/electrolyte. Moreover, the results of EIS measurements can be helpful as a check of the successful formation of the imprinting polymer film. The data obtained by the electrochemical impedance spectroscopy can be summarized in the Nyquist plot, and they may be schematized using a Randles equivalent circuit [68].
For the present study, the measurements were performed using 5 mmol L−1 K4Fe(CN)6/0.1 mol L−1 KCl at pH 7.2 as the electrochemical probe and supporting electrolyte. The Nyquist plot (Figure 3) of the modified electrodes by respectively NIP, re-charged MIP (after equilibration with 10−7 mol L−1 MCPA solution), and washed MIP showed a semicircle; this is a typical pattern of the surface-modified electrodes when the electron transfer is the limiting step. It can be observed that the resistance to the electron transfer at the electrode/electrolyte interface increases passing from the washed MIP to the NIP; this behavior means that the non-imprinted polymer hinders the flow of charges to the electrode surface more than the washed imprinted polymer. This assumption is confirmed by the impedance behavior of the not-washed MIP, which showed the highest resistance value; probably, the cavities occupied by the template prevented the flow of charges through the membrane. The diameter of the semicircle reported in the Nyquist plot corresponds to the resistance of the electrode, confirming the non-conductive nature of the polymer used to modify the working electrode surface.
The results represented by the Nyquist plot were schematized using an equivalent circuit helpful to separate any contribution to the total system impedance (Randles equivalent circuit, insert in Figure 3). The equivalent circuit used for fitting the experimental points is characterized by the presence of R1, which corresponds to the solution resistance; R2, which corresponds to the resistance of the electrode/electrolyte interface; and C1, which corresponds to the double layer capacitance.

3.2. Potentiometric Measurements by MIP-Modified Screen-Printed Cell

After the screen-printed cell modification with the MIP layer and after the removal of the template by the washing procedure previously described, it can be used to perform potentiometric measurements. For obtaining a potentiometric response, the analyte must be charged. For this reason, given that the MCPA pKa = 3.07, measurements were performed in PBS buffer pH 5.5 to ensure that the predominant species at this pH is the negatively charged deprotonated form [66]. Figure 4a shows the anion structure, and Figure 4b the graph of species distribution in function of the pH for MCPA.
Potentiometric measurements were performed by dipping the modified screen-printed cell in 15 mL of PBS buffer at pH 5.5 for 3 min under gentle stirring after reaching the potential stability (stability criterium ΔE/Δt = 0.02 mV/min). The calibration curve was obtained by adding portions of 20 µL of MCPA standard 25 µmol L−1 to the buffer and plotting the potential (E, mV) vs. the logarithm of MCPA concentration (log[MCPA]). Figure 5 shows an example of the calibration curves obtained for MIP and NIP functionalized electrodes.
Calibrations were performed with three different MIP or NIP-modified electrodes. Each electrode was used twice, so the standard deviation is calculated on the slopes obtained from 6 different calibrations using the three different electrodes two times each. As an example, one of these calibrations is shown in Figure 6. The slope of the calibration plots (expressed as the average value of the six experiments) for the electrodes modified with MIP is −59(1) mV/dec, very similar to the theoretical −59.2 mV/dec, expected from the Nernst equation. Conversely, the slope of the NIP-modified electrodes is much lower (of about −9(11) mV/dec); this means that these electrodes are lightly sensitive to the presence of MCPA in solution due to the unspecific interactions between the analyte and the polymeric film.
The detection limit (LOD) and the quantification limit (LOQ) for the potentiometric measurements with the MIP-modified electrodes can be obtained from the following relations (Equations (1) and (2)):
LOD = 3.3 s y / x slope
LOQ = 10 s y / x slope  
by considering the linearized Nernst equation 10E/slope vs. MCPA (mol L−1). sy/x is the standard deviation of y-residuals (i.e., the random errors in the y-direction); sy/x can be considered not significantly different from the standard deviation of replicate measurements of blank solutions [69].
The LOD and the LOQ obtained are, respectively, 13 nmol L−1 and 40 nmol L−1, demonstrating high sensitivity, although it is a potentiometric method.
For comparison in Table 1, the detection limits for MCPA analysis obtained from the present and other electroanalytical methods are reported.

3.3. Interferences

In order to study the selectivity of the sensor, tests were performed, investigating the pesticides shown in Figure 6, three structurally analogues (Mecroprop, Dichloroprop and 2,4-D Pestanal) and one of a different structure of MCPA (Bentazone).
Table 2 summarizes the slope of the calibration curves (E/mV vs. log[pesticide]) obtained by analyzing the interferents without any MCPA.
As expected, the phenoxy-based molecules, similar to MCPA, show a slope very close to the theoretical one of −59.2 mV/dec, while the MIP-functionalized screen-printed cell is not sensitive to the presence of bentazone; indeed, in this case, the slope is about −1 mV/dec.
To corroborate these results, other measurements were conducted in solutions containing MCPA at a constant concentration (0.5 µmol L−1) and increasing quantities of each interferent. As foreseeable, the sensor is sensitive to the presence of both MCPA and another phenoxy herbicide when they are in solution together, while it is sensitive to only MCPA in solutions containing bentazone.
Table 3 reports the slope of the calibrations performed. For the phenoxy herbicides, in the x-axis of the calibration plots, the logarithm of the sum of both herbicides’ concentrations is reported; conversely, the logarithm of the MCPA concentration was considered for the calibration plot obtained in the presence of bentazone. For this last pesticide, it was seen that the sensor is sensitive only to the presence of MCPA. Moreover, as further confirmation, another experiment was performed by adding increasing quantities of bentazone (ranging from 0.5 mmol L−1 to 2.5 mmol L−1) to a solution containing MCPA 0.5 mmol L−1: In this experiment, the potential does not change after each addition of Bentazone (see Figure S4, Supplementary Materials).
It is interesting to highlight that the MIP-base electrode is selective only to phenoxy herbicides with similar structure and chemical behavior to the template MCPA so the sensor could be applied to determining the total degree of contamination from similar substances.

3.4. Real Samples Analyses

In order to verify the applicability of the MIP-based potentiometric screen-printed cell to phenoxy herbicides detection in environmental samples, some experiments were performed testing tap water samples spiked with MCPA; the quantification was performed by the standard addition technique. The tap water samples’ pH was adjusted to 5.5 and spiked with different concentrations of MCPA. The % recovery and the RDS were calculated and reported in Table 4.
The standard addition method is used to determine the concentration of MCPA in the spiked samples as the usual procedure for electrochemical analyses of real samples. The straight lines obtained during the measurements have slopes ranging from −57(3) mV/dec and −62(2) mV/dec, so the sample’s matrix does not affect the sensitivity and the LOD of the electrode
As can be observed in Table 4, a good % recovery, ranging from 98.5% and 103.6%, was obtained; the precision was quite good since the highest value of the % RDS is 4.6.
These results were promising for practical applications of the developed potentiometric sensor for environmental analyses of samples contaminated by phenoxy herbicides.

4. Conclusions

This paper presents a potentiometric screen-printed cell modified with MIP for phenoxy herbicides detection using MCPA as the template molecule. The molecularly imprinted polymer (MIP) film over the working electrode surface recognizes these herbicides thanks to the specific cavities interacting with the analyte through weak bonds. It acts like an ionophore of a classical ion-selective electrode when the target analyte is present as an anion in aqueous solutions.
Electrochemical impedance spectroscopy (EIS) was applied to characterize the surface of the modified working electrode, confirming the non-conductive nature of the acrylic polymer film coverage.
Calibration studies were performed, and the slope of the linear plot E/mV vs. log[MCPA] was Nernstian, being equal to −59 mV/dec.
The interferents tests highlighted similar sensing properties of the MIP-modified screen-printed electrode to MCPA for Mecoprop, Dichloroprop, and 2,4-D Pestanal, i.e., phenoxy pesticides structurally similar to the target one, confirming that the sensor could be applied to determine the total degree of contamination coming from similar substances. Conversely, the MIP-based electrode does not respond to molecules of very different structures, such as the Bentazone examined here.
The low detection limit (about 10 nM) and the good results obtained from recovery tests on spiked tap water samples are promising for future sensor applications. It can be highlighted that the sensors can directly detect phenoxy herbicides present in environmental waters at concentrations compared to the World Health Organization (WHO) guideline value (2 µg L−1 = 10 nmol L−1) [7]. It is not sensitive enough to quantify, without a preconcentration step, the herbicides at concentration levels required by the European Commission Drinking Water Directive 98/83/EC, concerning the quality of water for human consumption, i.e., 0.1 μg L−1 = 0.5 nmol L−1) [7].

Supplementary Materials

The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/ijerph192416488/s1, Figure S1: Potentiometric response of a MIP-based screen-printed cell at different MCPA concentrations in PBS buffer at pH = 5.5. Steady-state reached in 3 min. Cell potential registered in the last 60 s. Figure S2: (a) Screen-printed xell; (b) Experimental setup: the electrode is dipped in the buffer solution and connected to (c) EmStat 4s for potential acquisition; Figure S3: SEM images of (a) MCPA molecularly imprinted polymer and (b) not-imprinted polymer at the same magnification; Figure S4: Potentiometric response of the MIP-modified electrode in MCPA 0.5 µmol L−1 solution after the addition of bentazone of 0.5 µmol L−1, 1.0 µmol L−1 and 2.5 µmol L−1.

Author Contributions

Conceptualization, C.Z. and G.A.; methodology, C.Z. and G.A.; formal analysis, C.Z. and G.A.; investigation, C.Z. and S.S.; data curation, C.Z. and R.B.; writing—original draft preparation, C.Z. and G.A.; writing—review and editing, L.R.M., S.S., M.G.-G. and R.B. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Data Availability Statement

Not applicable.

Acknowledgments

We thank MIUR for funding Camilla Zanoni’s Ph.D. grants, Ministerio de Universidades-European Union in the frame of NextGeneration EU RD 289/2021 for funding Marta Guembe-Garcia’s Post-Doc, and Topflight Italia (S.P.A.) for providing us with screen-printed cells free of charge.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Aharonson, N. Potential contamination of ground water by pesticides. Pure Appl. Chem. 1987, 59, 1419. [Google Scholar] [CrossRef]
  2. Moszczyński, W.; Białek, A. Ecological Production Technology of Phenoxyacetic Herbicides MCPA and 2,4-D in the Highest World Standard. In Herbicides-Properties, Synthesis and Control of Weeds; InTechOpen: Rijeka, Croatia, 2012; pp. 347–362. [Google Scholar]
  3. Paszko, T. Degradation of MCPA in Soil Horizons of Polish Agricultural Soils. Polish J. Environ. Stud. 2009, 18, 1083–1091. [Google Scholar]
  4. Bala, T.; Prasad, B.L.V.; Sastry, M.; Kahaly, M.U.; Waghmare, U.V. Interaction of different metal ions with carboxylic acid group: A quantitative study. J. Phys. Chem. A 2007, 111, 6183–6190. [Google Scholar] [CrossRef] [PubMed]
  5. Agricuture and Environment Research Unit. Pesticide properties database MCPA. AERU, University of Hertfordshire. 2016. Available online: http://sitem.herts.ac.uk/aeru/ppdb/en/Reports/427.htm (accessed on 18 October 2022).
  6. Mackay, D.; Shiu, W.Y.; Ma, K.-C.; Lee, S.C. Nitrogen and sulfur containing compounds and pesticides. In Handbook of Physical–Chemical Properties and Environmental Fate for Organic Chemicals, 2nd ed.; Taylor and Francis Group: Boca Raton, FL, USA, 2006; Volume 4, pp. 3457–3710. [Google Scholar]
  7. Morton, P.A.; Fennell, C.; Cassidy, R.; Doody, D.; Fenton, O.; Mellander, P.; Jordan, P. A review of the pesticide MCPA in the land-water environment and emerging research needs. WIREs Water 2020, 7, e1402. [Google Scholar] [CrossRef] [Green Version]
  8. Government of Canada. Guidelines for Canadian Drinking Water Quality: Guideline Technical Document—4-Chloro-2-methylphenoxyacetic Acid (MCPA). 2022; ISBN 978-0-660-42035-6. Available online: https://www.canada.ca/en/health-canada/services/publications/healthy-living/guidelines-canadian-drinking-water-quality-guideline-technical-document-2-methyl-4-chlorophenoxyacetic-acid-mcpa.html (accessed on 20 October 2022).
  9. Roberts, D.M.; Seneviratne, R.; Mohammed, F.; Patel, R.; Senarathna, L.; Hittarage, A.; Buckley, N.A.; Dawson, A.H.; Eddleston, M. Intentional self-poisoning with the chlorophenoxy herbicide 4-chloro-2-methylphenoxyacetic acid (MCPA). Ann. Emerg. Med. 2005, 46, 275–284. [Google Scholar] [CrossRef] [Green Version]
  10. Grande Martínez, Á.; Arrebola Liébanas, F.J.; Santiago Valverde, R.; Hernández Torres, M.E.; Ramírez Casinello, J.; Garrido Frenich, A. Multifamily Determination of Phytohormones and Acidic Herbicides in Fruits and Vegetables by Liquid Chromatography–Tandem Mass Spectrometry under Accredited Conditions. Foods 2020, 9, 906. [Google Scholar] [CrossRef]
  11. Seebunrueng, K.; Phosiri, P.; Apitanagotinon, R.; Srijaranai, S. A new environment-friendly supramolecular solvent-based liquid phase microextraction coupled to high performance liquid chromatography for simultaneous determination of six phenoxy acid herbicides in water and rice samples. Microchem. J. 2020, 152, 104418. [Google Scholar] [CrossRef]
  12. Wintersteiger, R.; Goger, B.; Krautgartner, H. Quantitation of chlorophenoxy acid herbicides by high-performance liquid chromatography with coulometric detection. J. Chromatogr. A 1999, 846, 349–357. [Google Scholar] [CrossRef]
  13. Meier, M.; Hamann, R.; Kettrup, A. Determination of phenoxy acid herbicides by high-performance liquid chromatography and on-line enrichment. Anal. Bioanal. Chem. 1989, 334, 235–237. [Google Scholar] [CrossRef]
  14. Bruner, F.; Berloni, A.; Palma, P. Determination of acidic chlorinated pesticides in water: Comparison of official EPA method 515.1 and liquid-solid extraction HPLC/UV and HPLC/PBMS analyses. Chromatographia 1996, 43, 279–284. [Google Scholar] [CrossRef]
  15. Rahemi, V.; Garrido, J.M.P.J.; Borges, F.; Brett, C.M.A.; Garrido, E.M.P.J. Electrochemical sensor for simultaneous determination of herbicide mcpa and its metabolite 4-chloro-2-methylphenol. Application to photodegradation environmental monitoring. Environ. Sci. Pollut. Res. 2015, 22, 4491–4499. [Google Scholar] [CrossRef]
  16. Rahemi, V.; Vandamme, J.J.; Garrido, J.M.P.J.; Borges, F.; Brett, C.M.A.; Garrido, E.M.P.J. Enhanced host-guest electrochemical recognition of herbicide mcpa using a β-cyclodextrin carbon nanotube sensor. Talanta 2012, 99, 288–293. [Google Scholar] [CrossRef] [Green Version]
  17. Wong, A.; Silva, T.; Caetano, F.; Bergamini, M.; Marcolino-Junior, L.; Fatibello-Filho, O.; Janegitz, B. An Overview of Pesticide Monitoring at Environmental Samples Using Carbon Nanotubes-Based Electrochemical Sensors. C 2017, 3, 8. [Google Scholar] [CrossRef] [Green Version]
  18. Bialek, A.; Skrzypczynska, K.; Kusmierek, K.; Swiatkowski, A. Voltammetric determination of MCPA, 4-chloro-o-cresol and o-cresol in water using a modified carbon paste electrode. Int. J. Electrochem. Sci. 2019, 14, 228–237. [Google Scholar] [CrossRef]
  19. Yu, T.; Fenelon, O.; Herdman, K.M.; Breslin, C.B. The Electrochemical Detection of 4-chloro-2-methylphenoxyacetic Acid (MCPA) Using a Simple Activated Glassy Carbon Electrode. J. Electrochem. Soc. 2022, 169, 037514. [Google Scholar] [CrossRef]
  20. Chen, P.Y.; Vittal, R.; Nien, P.C.; Ho, K.C. Enhancing dopamine detection using a glassy carbon electrode modified with MWCNTs, quercetin, and Nafion®. Biosens. Bioelectron. 2009, 24, 3504–3509. [Google Scholar] [CrossRef] [PubMed]
  21. Akhoundian, M.; Alizadeh, T.; Ganjali, M.R.; Rafiei, F. A new carbon paste electrode modified with MWCNTs and nano-structured molecularly imprinted polymer for ultratrace determination of trimipramine: The crucial effect of electrode components mixing on its performance. Biosens. Bioelectron. 2018, 111, 27–33. [Google Scholar] [CrossRef] [PubMed]
  22. Kim, S.K.; Jeong, Y.N.; Ahmed, M.S.; You, J.M.; Choi, H.C.; Jeon, S. Electrocatalytic determination of hydrazine by a glassy carbon electrode modified with PEDOP/MWCNTs-Pd nanoparticles. Sensors Actuators B Chem. 2011, 153, 246–251. [Google Scholar] [CrossRef]
  23. Alizadeh, T.; Hamidi, N.; Ganjali, M.R.; Rafiei, F. Determination of subnanomolar levels of mercury (II) by using a graphite paste electrode modified with MWCNTs and Hg(II)-imprinted polymer nanoparticles. Microchim. Acta 2018, 185, 16. [Google Scholar] [CrossRef]
  24. Sirivisoot, S.; Webster, T.J. Multiwalled carbon nanotubes enhance electrochemical properties of titanium to determine in situ bone formation. Nanotechnology 2008, 19, 295101. [Google Scholar] [CrossRef]
  25. Brandão, A.T.S.C.; Rosoiu, S.; Costa, R.; Lazar, O.A.; Silva, A.F.; Anicai, L.; Pereira, C.M.; Enachescu, M. Characterization and electrochemical studies of MWCNTs decorated with Ag nanoparticles through pulse reversed current electrodeposition using a deep eutectic solvent for energy storage applications. J. Mater. Res. Technol. 2021, 15, 342–359. [Google Scholar] [CrossRef]
  26. Li, Y.; Keller, A.L.; Cryan, M.T.; Ross, A.E. Metal Nanoparticle Modified Carbon-Fiber Microelectrodes Enhance Adenosine Triphosphate Surface Interactions with Fast-Scan Cyclic Voltammetry. ACS Meas. Sci. Au 2022, 2, 96–105. [Google Scholar] [CrossRef]
  27. Baig, N.; Sajid, M.; Saleh, T.A. Recent trends in nanomaterial-modified electrodes for electroanalytical applications. TrAC-Trends Anal. Chem. 2019, 111, 47–61. [Google Scholar] [CrossRef]
  28. Tayeb, I.A.; Abdul Razak, K. Development of Gold Nanoparticles Modified Electrodes for the Detection of Heavy Metal Ions. J. Phys. Conf. Ser. 2018, 1083, 012044. [Google Scholar] [CrossRef]
  29. Antuña-Jiménez, D.; González-García, M.B.; Hernández-Santos, D.; Fanjul-Bolado, P. Screen-printed electrodes modified with metal nanoparticles for small molecule sensing. Biosensors 2020, 10, 9. [Google Scholar] [CrossRef] [Green Version]
  30. Yakupova, E.; Ziyatdinova, G. Electrode Modified with Tin(IV) Oxide Nanoparticles and Surfactants as Sensitive Sensor for Hesperidin. Chem. Proc. 2021, 5, 54. [Google Scholar]
  31. Liu, F.; Huang, L.; Duan, X.; Li, Y.Y.; Hu, J.Q.; Li, B.H.; Lu, J. A facile method to prepare noble metal nanoparticles modified Self-Assembly (SAM) electrode. J. Exp. Nanosci. 2018, 13, 1–10. [Google Scholar] [CrossRef] [Green Version]
  32. Wang, Y.; Li, Y.; Tang, L.; Lu, J.; Li, J. Application of graphene-modified electrode for selective detection of dopamine. Electrochem. Commun. 2009, 11, 889–892. [Google Scholar] [CrossRef]
  33. Kim, Y.R.; Bong, S.; Kang, Y.J.; Yang, Y.; Mahajan, R.K.; Kim, J.S.; Kim, H. Electrochemical detection of dopamine in the presence of ascorbic acid using graphene modified electrodes. Biosens. Bioelectron. 2010, 25, 2366–2369. [Google Scholar] [CrossRef]
  34. Arvinte, A.; Mahosenaho, M.; Pinteala, M.; Sesay, A.M.; Virtanen, V. Electrochemical oxidation of p-nitrophenol using graphene-modified electrodes, and a comparison to the performance of MWNT-based electrodes. Microchim. Acta 2011, 174, 337–343. [Google Scholar] [CrossRef]
  35. Wu, J.F.; Xu, M.Q.; Zhao, G.C. Graphene-based modified electrode for the direct electron transfer of Cytochrome c and biosensing. Electrochem. Commun. 2010, 12, 175–177. [Google Scholar] [CrossRef]
  36. Brownson, D.A.C.; Foster, C.W.; Banks, C.E. The electrochemical performance of graphene modified electrodes: An analytical perspective. Analyst 2012, 137, 1815–1823. [Google Scholar] [CrossRef] [PubMed]
  37. Fan, Y.; Liu, J.H.; Lu, H.T.; Zhang, Q. Electrochemical behavior and voltammetric determination of paracetamol on Nafion/TiO2-graphene modified glassy carbon electrode. Colloids Surf. B 2011, 85, 289–292. [Google Scholar] [CrossRef] [PubMed]
  38. Chen, Q.; Li, X.; Min, X.; Cheng, D.; Zhou, J.; Li, Y.; Xie, Z.; Liu, P.; Cai, W.; Zhang, C. Determination of catechol and hydroquinone with high sensitivity using MOF-graphene composites modified electrode. J. Electroanal. Chem. 2017, 789, 114–122. [Google Scholar] [CrossRef]
  39. Dang, W.; Sun, Y.; Jiao, H.; Xu, L.; Lin, M. AuNPs-NH2/Cu-MOF modified glassy carbon electrode as enzyme-free electrochemical sensor detecting H2O2. J. Electroanal. Chem. 2020, 856, 113592. [Google Scholar] [CrossRef]
  40. Wen, Y.; Meng, W.; Li, C.; Dai, L.; He, Z.; Wang, L.; Li, M.; Zhu, J. Enhanced glucose sensing based on a novel composite CoII-MOF/Acb modified electrode. Dalt. Trans. 2018, 47, 3872–3879. [Google Scholar] [CrossRef]
  41. Meng, W.; Xu, S.; Dai, L.; Li, Y.; Zhu, J.; Wang, L. An enhanced sensitivity towards H2O2 reduction based on a novel Cu metal–organic framework and acetylene black modified electrode. Electrochim. Acta 2017, 230, 324–332. [Google Scholar] [CrossRef]
  42. Li, X.; Sun, X.; Li, M. Detection of sudan I in foods by a mof-5/mwcnt modified electrode. ChemistrySelect 2020, 5, 12777–12784. [Google Scholar] [CrossRef]
  43. Huang, T.Y.; Kung, C.W.; Liao, Y.T.; Kao, S.Y.; Cheng, M.; Chang, T.H.; Henzie, J.; Alamri, H.R.; Alothman, Z.A.; Yamauchi, Y.; et al. Enhanced Charge Collection in MOF-525–PEDOT Nanotube Composites Enable Highly Sensitive Biosensing. Adv. Sci. 2017, 4, 1700261. [Google Scholar] [CrossRef]
  44. Wang, L.; Xu, M.; Han, L.; Zhou, M.; Zhu, C.; Dong, S. Graphene enhanced electron transfer at aptamer modified electrode and its application in biosensing. Anal. Chem. 2012, 84, 7301–7307. [Google Scholar] [CrossRef]
  45. Liu, Y.; Kwa, T.; Revzin, A. Simultaneous detection of cell-secreted TNF-α and IFN-γ using micropatterned aptamer-modified electrodes. Biomaterials 2012, 33, 7347–7355. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  46. Xu, D.; Xu, D.; Yu, X.; Liu, Z.; He, W.; Ma, Z. Label-free electrochemical detection for aptamer-based array electrodes. Anal. Chem. 2005, 77, 5107–5113. [Google Scholar] [CrossRef] [PubMed]
  47. Álvarez-Martos, I.; Campos, R.; Ferapontova, E.E. Surface state of the dopamine RNA aptamer affects specific recognition and binding of dopamine by the aptamer-modified electrodes. Analyst 2015, 140, 4089–4096. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  48. Villalonga, A.; Pérez-Calabuig, A.M.; Villalonga, R. Electrochemical biosensors based on nucleic acid aptamers. Anal. Bioanal. Chem. 2020, 412, 55–72. [Google Scholar] [CrossRef] [PubMed]
  49. Liu, X.; Cheng, Z.; Fan, H.; Ai, S.; Han, R. Electrochemical detection of avian influenza virus H5N1 gene sequence using a DNA aptamer immobilised onto a hybrid nanomaterial-modified electrode. Electrochim. Acta 2011, 56, 6266–6270. [Google Scholar] [CrossRef]
  50. Ruzgas, T.; Csöregi, E.; Emnéus, J.; Gorton, L.; Marko-Varga, G. Peroxidase-modified electrodes: Fundamentals and application. Anal. Chim. Acta 1996, 330, 123–138. [Google Scholar] [CrossRef]
  51. Iannlello, R.M.; Yacynych, A.M. Immobilized Enzyme Chemically Modified Electrode as an Amperometric Sensor. Anal. Chem. 1981, 53, 2090–2095. [Google Scholar] [CrossRef]
  52. Yao, T. A chemically-modified enzyme membrane electrode as an amperometric glucose sensor. Anal. Chim. Acta 1983, 148, 27–33. [Google Scholar] [CrossRef]
  53. Trojanowicz, M.; Matuszewski, W.; Podsiadła, M. Enzyme entrapped polypyrrole modified electrode for flow-injection determination of glucose. Biosens. Bioelectron. 1990, 5, 149–156. [Google Scholar] [CrossRef]
  54. Zhang, X.; Liao, Q.; Chu, M.; Liu, S.; Zhang, Y. Structure effect on graphene-modified enzyme electrode glucose sensors. Biosens. Bioelectron. 2014, 52, 281–287. [Google Scholar] [CrossRef]
  55. Eftekhari, A. Electropolymerization of aniline onto passivated substrate and its application for preparation of enzyme-modified electrode. Synth. Met. 2004, 145, 211–216. [Google Scholar] [CrossRef]
  56. Alberti, G.; Zanoni, C.; Spina, S.; Magnaghi, L.R.; Biesuz, R. MIP-Based Screen-Printed Potentiometric Cell for Atrazine Sensing. Chemosensors 2022, 10, 339. [Google Scholar] [CrossRef]
  57. Ata, S.; Feroz, M.; Bibi, I.; Mohsin, I.-U.; Alwadai, N.; Iqbal, M. Investigation of electrochemical reduction and monitoring of p-nitrophenol on imprinted polymer modified electrode. Synth. Met. 2022, 287, 117083. [Google Scholar] [CrossRef]
  58. Bu, L.; Chen, X.; Song, Q.; Jiang, D.; Shan, X.; Wang, W.; Chen, Z. Supersensitive detection of chloramphenicol with an EIS method based on molecularly imprinted polypyrrole at UiO-66 and CDs modified electrode. Microchem. J. 2022, 179, 107459. [Google Scholar] [CrossRef]
  59. Preda, D.; David, I.G.; Popa, D.E.; Buleandra, M.; Radu, G.L. Recent Trends in the Development of Carbon-Based Electrodes Modified with Molecularly Imprinted Polymers for Antibiotic Electroanalysis. Chemosensors 2022, 10, 243. [Google Scholar] [CrossRef]
  60. Das, D.; Nag, S.; Adaval, A.; Hazarika, A.K.; Sabhapondit, S.; Bhattacharyya, A.R.; Tudu, B.; Bandopadhyay, R.; Banerjee Roy, R. Amine Functionalized MWCNTs Modified MIP-Based Electrode for Detection of Epicatechin in Tea. IEEE Sens. J. 2022, 22, 10323–10330. [Google Scholar] [CrossRef]
  61. Bohlooli, S.; Kia, S.; Bohlooli, S.; Sariri, R. Development of molecularly imprinted polymer on ferric oxide nanoparticles modified electrode as electrochemical sensor for detection of human growth hormone. Mon. Chem. 2022, 153, 39–48. [Google Scholar] [CrossRef]
  62. Pohanka, M. Sensors based on molecularly imprinted polymers. Int. J. Electrochem. Sci. 2017, 12, 8082–8094. [Google Scholar] [CrossRef] [Green Version]
  63. Belbruno, J.J. Molecularly Imprinted Polymers. Chem. Rev. 2019, 119, 94–119. [Google Scholar] [CrossRef]
  64. Uzun, L.; Turner, A.P.F. Molecularly-imprinted polymer sensors: Realising their potential. Biosens. Bioelectron. 2016, 76, 131–144. [Google Scholar] [CrossRef]
  65. Rakesh Kumar, R.K.; Shaikh, M.O.; Chuang, C.H. A review of recent advances in non-enzymatic electrochemical creatinine biosensing. Anal. Chim. Acta 2021, 1183, 338748. [Google Scholar] [CrossRef] [PubMed]
  66. Teixeira, M.C.; Sá-Correia, I. Saccharomyces cerevisiae resistance to chlorinated phenoxyacetic acid herbicides involves Pdr1p-mediated transcriptional activation of TPO1 and PDR5 genes. Biochem. Biophys. Res. Commun. 2002, 292, 530–537. [Google Scholar] [CrossRef] [PubMed]
  67. Velásquez, P.; Leinen, D.; Pascual, J.; Ramos-Barrado, J.R.; Cordova, R.; Gómez, H.; Schrebler, R. SEM, EDX and EIS study of an electrochemically modified electrode surface of natural enargite (Cu3AsS4). J. Electroanal. Chem. 2000, 494, 87–95. [Google Scholar] [CrossRef]
  68. Randles, J.E.B. Kinetics of rapid electrode reactions. Discuss. Faraday Soc. 1947, 1, 11–19. [Google Scholar] [CrossRef]
  69. Miller, J.N.; Miller, J.C. Calibration methods in instrumental analysis: Regression and correlation. In Statistics and Chemometrics for Analytical Chemistry, 6th ed.; Pearson Education Limited: Harlow Essex, UK, 2010; pp. 124–126. [Google Scholar]
  70. Lu, Y.; Bao, C.; Zou, J.; Xiao, J.; Zhong, W.; Gao, Y. Highly Sensitive Electrochemical Sensor for Sunset Yellow Based on Electrochemically Activated Glassy Carbon Electrode. Molecules 2022, 27, 5221. [Google Scholar] [CrossRef]
Figure 1. Schematic representation of molecularly imprinting procedure to form a MIP.
Figure 1. Schematic representation of molecularly imprinting procedure to form a MIP.
Ijerph 19 16488 g001
Figure 2. Schematic representation of MIP-modified screen-printed cell assembly.
Figure 2. Schematic representation of MIP-modified screen-printed cell assembly.
Ijerph 19 16488 g002
Figure 3. Nyquist plot of electrode modified with NIP (blue), washed MIP (purple) and MIP re-charged with MCPA 10−7 mol L−1 (pink). Measurements were performed in 5 mmol L−1 K4Fe(CN)6/0.1 mol L−1 KCl at pH 7.2. The insert in the figure is the Randles equivalent circuit of the trends reported in the Nyquist plot for all three modified electrodes.
Figure 3. Nyquist plot of electrode modified with NIP (blue), washed MIP (purple) and MIP re-charged with MCPA 10−7 mol L−1 (pink). Measurements were performed in 5 mmol L−1 K4Fe(CN)6/0.1 mol L−1 KCl at pH 7.2. The insert in the figure is the Randles equivalent circuit of the trends reported in the Nyquist plot for all three modified electrodes.
Ijerph 19 16488 g003
Figure 4. (a) Negatively charged deprotonated form of MCPA. (b) Graph of species distribution in function of the pH for MCPA.
Figure 4. (a) Negatively charged deprotonated form of MCPA. (b) Graph of species distribution in function of the pH for MCPA.
Ijerph 19 16488 g004
Figure 5. Calibration curve of MCPA of screen-printed cell modified with MIP (purple), with NIP (orange), and theoretical Nernstian slope. Experimental conditions: 15 mL PBS pH 5.5, additions of 0.02 mL MCPA std 25 µmol L−1.
Figure 5. Calibration curve of MCPA of screen-printed cell modified with MIP (purple), with NIP (orange), and theoretical Nernstian slope. Experimental conditions: 15 mL PBS pH 5.5, additions of 0.02 mL MCPA std 25 µmol L−1.
Ijerph 19 16488 g005
Figure 6. Molecular structure of interferents.
Figure 6. Molecular structure of interferents.
Ijerph 19 16488 g006
Table 1. Detection limits of different electroanalytical methods for MCPA analyses.
Table 1. Detection limits of different electroanalytical methods for MCPA analyses.
Electrode, MethodLOD (nmol L−1)Reference
Glassy carbon, DPV 18[19]
PANI-β-CD/fMWCNT/GCE, CV 2990[16]
Activated glassy carbon, SWV 31.6[70]
PANI-β-CD/fMWCNT/GCE, CV 21100–1600[15]
Silica-modified carbon paste electrode, DPV 11300[18]
MIP-modified SPC 4, potentiometry13This work
1 DPV, differential pulsed voltammetry; 2 CV, cyclic voltammetry; 3 SWV, square wave voltammetry; 4 SPC, screen-printed cell.
Table 2. Slopes of the calibration curves for each pesticide (average values obtained for three calibration curves for each pesticide). Numbers in parentheses are the standard deviation on the last digit.
Table 2. Slopes of the calibration curves for each pesticide (average values obtained for three calibration curves for each pesticide). Numbers in parentheses are the standard deviation on the last digit.
PesticideSlope (mV/dec)
Mecoprop−54 (3)
Dichloroprop−57 (5)
2,4-D Pestanal−55 (3)
Bentazone−1.3 (2)
MCPA−59 (1)
Table 3. Slope of the calibration curves performed in solutions containing 0.5 µmol L−1 of MCPA and increasing quantities of each interferent. For the phenoxy herbicides, in the x-axis of the calibration plots, the logarithm of the sum of both herbicides’ concentrations is reported; conversely, the logarithm of the MCPA concentration was considered for the calibration plot obtained in the presence of bentazone.
Table 3. Slope of the calibration curves performed in solutions containing 0.5 µmol L−1 of MCPA and increasing quantities of each interferent. For the phenoxy herbicides, in the x-axis of the calibration plots, the logarithm of the sum of both herbicides’ concentrations is reported; conversely, the logarithm of the MCPA concentration was considered for the calibration plot obtained in the presence of bentazone.
PesticideSlope (mV/dec)
MCPA + Mecoprop−57 (2)
MCPA + Dichloroprop−55 (3)
MCPA + 2,4-D Pestanal−57 (4)
MCPA+ Bentazone−60 (2)
MCPA−59 (1)
Table 4. Recovery experiments. Tap water samples adjusted to pH 5.5 with PBS and spiked with different MCPA concentrations.
Table 4. Recovery experiments. Tap water samples adjusted to pH 5.5 with PBS and spiked with different MCPA concentrations.
Spike Concentration (μmol L−1)Found (μmol L−1, x ± CI 1)% RDS% Recovery
0.1690.166 ± 0.042.598.5
0.1010.101 ± 0.054.6103.6
0.5310.5338 ± 0.020.5100.5
0.9180.928 ± 0.031.1101.1
1 CI = 95% confidence interval. For 3 replicates, t = 4.30.
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Zanoni, C.; Spina, S.; Magnaghi, L.R.; Guembe-Garcia, M.; Biesuz, R.; Alberti, G. Potentiometric MIP-Modified Screen-Printed Cell for Phenoxy Herbicides Detection. Int. J. Environ. Res. Public Health 2022, 19, 16488. https://doi.org/10.3390/ijerph192416488

AMA Style

Zanoni C, Spina S, Magnaghi LR, Guembe-Garcia M, Biesuz R, Alberti G. Potentiometric MIP-Modified Screen-Printed Cell for Phenoxy Herbicides Detection. International Journal of Environmental Research and Public Health. 2022; 19(24):16488. https://doi.org/10.3390/ijerph192416488

Chicago/Turabian Style

Zanoni, Camilla, Stefano Spina, Lisa Rita Magnaghi, Marta Guembe-Garcia, Raffaela Biesuz, and Giancarla Alberti. 2022. "Potentiometric MIP-Modified Screen-Printed Cell for Phenoxy Herbicides Detection" International Journal of Environmental Research and Public Health 19, no. 24: 16488. https://doi.org/10.3390/ijerph192416488

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop