Article

Evaluating a 3D Ultrasound Imaging Resolution of Single Transmitter/Receiver with Coding Mask by Extracting Phase Information

Graduate School of Systems Design, Tokyo Metropolitan University, 6-6 Asahigaoka, Hino 191-0065, Tokyo, Japan
*
Author to whom correspondence should be addressed.
Sensors 2024, 24(5), 1496; https://doi.org/10.3390/s24051496
Submission received: 5 December 2023 / Revised: 15 February 2024 / Accepted: 21 February 2024 / Published: 25 February 2024
(This article belongs to the Special Issue Advances in 3D Imaging and Multimodal Sensing Applications)

Abstract

:
We are currently investigating the ultrasound imaging of a sensor that consists of a randomized encoding mask attached to a single lead zirconate titanate (PZT) oscillator for a puncture microscope application. The proposed model was conducted using a finite element method (FEM) simulator. To increase the number of measurements required by a single element system that affects its resolution, the transducer was rotated at different angles. The image was constructed by solving a linear equation of the image model resulting in a poor quality. In a previous work, the phase information was extracted from the echo signal to improve the image quality. This study proposes a strategy by integrating the weighted frequency subbands compound and a super-resolution technique to enhance the resolution in range and lateral direction. The image performance with different methods was also evaluated using the experimental data. The results indicate that better image resolution and speckle suppression were obtained by applying the proposed method.

1. Introduction

Ultrasound imaging has been widely applied to medical diagnosis due to its superiority via non-ionizing radiation and its relatively low-cost system. At present, it is performed by using an array system, which requires a number of sensors to be properly arranged and applies a general beamforming method. For accurate diagnosis, it should have a high resolution and high signal-to-noise ratio (SNR), and therefore researchers have intensively studied many methods to obtain a better image quality [1,2,3,4,5,6,7,8,9,10].
In the current pathological examination, the tissue diagnosis takes a certain time for its performance. It is started by obtaining the tissue sample of human body by biopsy process which gives a burden to the patient and performs the observation under a standard optical microscope. It can be solved by developing an ultrasonic puncture microscopy which is possible to capture the tissue area of human body directly with easy operation and while being comfortable for the patient. However, such a device should have a simple compact structure system and is not suitable for implementation by a sensor array due to the significant amount of wiring and complex electronic circuit needed to generate the ultrasound image [11,12,13,14].
Recently, structured ultrasound microscopy (SUM) has been proposed by using a single transmitter/receiver circuit [15,16] while placing the coding mask [17]. However, single transducers with a single time-series echo signal required several additional measurements compared to array transducers consisting of numerous elements to construct a 3D image. Therefore, it was integrated by attaching an acoustic mask with irregular abberations to establish spatial coding. Randomizing the thickness of the coding mask leads to a different delay time of the transmission signal traveling inside the mask material and finally to a randomized wave propagation in the target medium. The reflection signals from the scatterers with different delays are also received by the same transducer in a single accumulated echo signal. By rotating the transducer, variations of the measurement have been obtained to increase the spatial coding variation of the region of interest.
Currently, many researchers are developing techniques for improving the ultrasound image resolution based on array transducers [18,19,20,21,22]. The minimum variance beamformer method on the received echo has been proposed to an array system enabling the improvement of image resolution by using adaptive weight [23,24,25]. Instead of that, the imaging methods utilizing a convolution of the transmitted pulse wave and multiple scatterer reflectances to accurately detect scatterer position have also been introduced via Time-Reversal MUltiple SIgnal Classification (TR-MUSIC) [26,27]. It utilizes the principle that echoes received at each element converge on a scatterer when they are time-reversed and retransmitted. The TR-MUSIC detects the received phase of a real scatterer at each element and achieves lateral super-resolution using the MUSIC algorithm. Another technique using a super-resolution FM-chirp correlation method (SCM) to achieve range-MUSIC which uses the noise subspace by subtracting the signal subspace from the total function space of the echo. The fundamental of this method is based on eigenvalue analysis [28].
To improve the image quality, the above methods were adopted and implemented in a single-element system. The minimum variance method with weight-frequency compound was implemented to a single-element system with a single scatterer placed in the region of interest by decomposing the base wide band of echo signal into several narrow subbands by [29]. The method was effective to improve the image resolution in the range direction. However, the resolution in the lateral direction was not significantly improved. Instead of that, the SCM performance showed a better lateral direction resolution. The SCM with a short pulse transmission wave was also implemented adaptively by constructing the image using a linear equation of the image model [30]. Due to the advantages separately, in this study, we proposed a method by integrating both techniques in order to achieve a better image quality. The main contributions of this study can be clearly listed as follows:
  • to develop ultrasound imaging based on single element with coding mask by numerically solving the image model equation,
  • to apply the proposed method by integrating the frequency subband compound and SCM, and
  • to evaluate the image performance by comparing it with other methods by varying the single scatterer position in the region of interest.

2. Methods

2.1. Image Model

In this study, the measured data in radio frequency (RF) signal form was converted to an in-phase quadrature (IQ) signal for baseband processing. The image model was constructed using the following model
y = D x ,
where y is a vector that consists of the time sampling series of the measured IQ echo signal, and the M-column vectors of the matrix D are the time sampling series of the IQ reflectance signals from each grid in the region of interest (ROI). D is assumed to be known a priori. The vector x consists of the complex reflectance coefficients corresponding to each grid and is obtained by numerically solving the image model.

2.2. Proposed Method: Super-Resolution Weighted Frequency Subbands Compound

Many researchers have studied methods for improving the ultrasound imaging based on sensor array systems. In a previous study, the weighted frequency subbands compound and super-resolution methods have improved the image resolution based on a single element instead of an array system. In this study, both methods have been integrated to be proposed to enhance the image quality as shown in Figure 1.
In this method, the weighted frequency subband compound and the super-resolution method were performed separately. We define the measurement data at M-different angles of rotation by y = [ y 1 , y 2 , , y M ] and D = [ D 1 , D 2 , , D M ] . The wide base-band of the measured signal was separated into N narrow-subbands and M-angle of rotations. After applying the IQ detector, the solution-x was computed numerically for each subband and angle of rotation, resulting in M × N -images (tensor data X). The weighted frequency subband compound was applied to the second-order tensor data X by following the minimum-variance distortion-free response (MVDR) method [31]. The details of MVDR can be clearly seen in Appendix A. The method is started by determining the variance-covariance matrix (R) from the snapshot vector- p i ( t ) . In the sensor array, the variance-covariance matrix was obtained by averaging over number of elements. However, in the single-element system, it was achieved by averaging over a number of sensor rotations. The appropriate weight of the subbands compound was adaptively determined based on the variance-covariance matrix and was common for all angles of rotation. The final image of this method for each angle of rotations after feeding back to the previous tensor-X is Y M V D R as shown in the figure.
In the present study, SCM was applied to a sensor array system employing a standard beamforming method for image construction. Details of the procedures for applying SCM can be studied in Appendix B. In this study, SCM was applied adaptively to a single element. However, the SCM application based on the single-element system in the previous study was modified by applying the compression process to the matrix-D resulting in the equation D H y = D H D x where H indicates Hermitian transpose. Thus, the SCM profile was calculated based on the modified model and multiplied by the MVDR output, Y M V D R . The final image was obtained by applying a simple averaging or coherence factor (CF) [32] to integrate over the number of rotations.

2.3. Previous Methods for Image Construction Were Based on a Single Transducer

In this simulation, several methods were performed to evaluate the image quality and compared with the proposed method above. The details of these methods are described below:
  • Method-A: In this method, the image was constructed by numerically solving the linear equation (LE) of the image model, y = D x , for each angle of rotations [15,16].
  • Method-B: This method uses weighted frequency subband compounds to obtain images for each rotation. The original wide band of the measured signal was decomposed into several narrow subbands with their different center of frequency. The subband decomposition was applied to vector-y and matrix-D for all rotation angles. The images were constructed by numerically solving a linear equation of the image model with different subbands and angles of rotation. The weight of each subband was calculated and summed up for each angle of rotation using the minimum variance distortion-free response (MVDR) method [29].
  • Method-C: Based on the image model, y = D x , the SCM profile was extracted from the measured data for each angle of rotation. The final image at different rotation angles was obtained by multiplying the SCM profile with the image-x computed by numerically solving the LE [30].
  • Method-D: The image model was modified by applying the D-compression to the original image model resulting in D H y = D H D x . The SCM profile was determined based on the modified image model. On the other hand, the solution-x of this equation was computed by analytically solving x = ( D H D ) 1 D H y . The final image of each rotation was achieved by properly multiplying the image-x and the SCM profile.
The final image was obtained by simple averaging or CF to integrate the image over the number of rotations.

3. Simulation and Experiment Results

3.1. Simulation Model

The proposed model was developed using a finite element method (FEM) with OnScale—a simulator for ultrasound propagation and piezoelectric analysis. Figure 2a,b show the simulation model, which consists of a backing layer, a single PZT element, and an encoding mask with different scatterer number positions using water as the medium for propagating the ultrasonic wave. In reality, a physical device is a circular disk transducer that can be implemented to generate 3D images. However, due to the complexity of the computational process, a 2D simulation was performed in this study. Table 1 shows details of the physical parameters used in the simulation.
The important part of the device is a coding mask attached to the surface of the transducer as shown in Figure 2c. It consists of a number of patches with the same width. However, the height of each patch is different with a randomized thickness with a minimum thickness of 0.25 λ and a maximum thickness of 1 λ that creates a different local wave propagation with different time delays. The coding mask has a function to generate a spatial coding and a focusing transmitted waveform exceeding that of the broken transducer.In this simulation, the transducer was fired by a short pulse with center of frequency ( 7 MHz ) (see Figure 3). The snapshot of the FEM simulation in Figure 2c,d clearly shows the effect of the coding mask on the outspreading transmitted wave.
To construct the image based on the model equation, the reflected signal from the scatterers placed in the imaging area represented by the vector-y and the grid reflected signal at the ROI of the matrix-D were measured. The transducer was rotated at 25 rotation angles to increase the number of the measurements that affect the improvement of the image resolution. The reflected signal of D at the grid position in the ROI was calculated as follows. First, the sound field in the ROI produced by pulse transmission was measured. The ROI for the imaging area was set by 2 mm × 2 mm . A one-sided acoustic transfer function was calculated from the sound pressure waveforms obtained at each point in the ROI and the transmitted pulse. This was then multiplied by the sound pressure waveform at each point to obtain the reflected signal.

3.2. Simulation Result

In this subsection, the simulation results based on the proposed method are presented. After performing the model simulation and measuring data required for solving the image model process, the post-processing and constructed image were performed by MATLAB R2022a. Currently, the image target which is the vector-x was obtained by numerically solving Equation (1) using a least squares method (LSQR), which is available in the MATLAB library, with an iteration residual of 0.1 set in this simulation.

3.2.1. Basic Method: Numerically Solving a Linear Equation of Image Model

In the previous ultrasound wave simulation, the data consisting of the vector y and the matrix D were collected at different angles of rotation. Figure 4 shows the measured echo signal in RF waveform at different angles of rotation. The RF signal was converted to an analytical signal (IQ) before solving the LE. In this method, the image x was constructed by numerically solving the image model for each angle of rotation. The image with a single scatterer at the depth of about 4.5 mm was constructed with a different number of angles to evaluate the effect on the image quality as shown in Figure 5a–c. In this simulation, two scenarios were applied to assemble all the images at different rotation angles. The top images were constructed by simple averaging over the number of rotations. While the bottom images were constructed by applying the CF. The images with CF show better speckle suppression compared to the simple averaging method. At present, the image with 25 rotations shows a better image resolution and a good suppression of unwanted signal. However, the image with 15 rotations does not look much different compared to 25 rotations. To confirm this result, the signal intensity distribution crossing the line through the scatterer position in the range and lateral direction of the CF image was also plotted as shown in Figure 5d,e.

3.2.2. Frequency Subbands Compound

In this method, the original baseband echo was decomposed into 35 subbands with a bandwidth of 6 MHz . Some decomposed subband signals after applying a Hanning window filter can be seen in Figure 6. It shows that the center frequency of the base band ( 7 MHz ) has been changed to a specific frequency for each subband. The decomposition process was performed on the vector y and the matrix D for all rotation angles. The image model for each subband and rotation angle was solved numerically, resulting in 35 × 25 images. Figure 7 shows the performance of B-mode images in different subbands at the same number of rotation angle. The image in sub-25 and sub-35 shows poor quality due to its subband being far from the original center of frequency. The image quality for each subband shows a poor resolution and therefore subband merging must be performed to improve it. In a previous study by [29], the narrow bandwidth of subbands showed the better resolution. The narrow bandwidth of the subband shows better resolution in comparison.
In this method, to compound over the number of subbands, we proposed two methods which are simple averaging subbands and weighted frequency subband compound. In the simple averaging method, the image with 35 subbands was averaged for each rotation angle. While in a weighted frequency subband compound, the weight for subbands was calculated and summed up to obtain the image for each rotation angle. The final image was obtained by simple averaging or CF to compound over the number of rotations as in the previous method. Figure 8 shows the B-mode image by simple averaging over the number of subbands and applying weighted frequency subband compounding. The B-mode image by averaging over the number of subbands has better resolution in the range direction compared to the image without subband processing. However, the unwanted signal in the background image is almost the same as before. The image with the CF method is not much different with simple averaging, although it has a better image resolution in the range direction.
The image constructed by a weighted frequency subband averaging shows a better image quality. By simply averaging the compound over the number of rotations, the image resolution in the range direction was improved. It also suppressed more speckle compared to the averaging subbands method. The image quality became better when the CF method was used to average the image at different angles of rotation. The effect of increasing the number of rotation angles also affected the image quality as shown in Figure 9 with simple averaging for rotation compounding. The image with five rotations shows poor resolution compared to the other images. The images with 15 rotations and 25 rotations are not much different to suppress the unwanted signal. In general, the image with the weighted frequency subband combination shows a better image resolution, especially in the range direction, and has a better speckle suppression. Figure 10 shows the effect of the subbands width to the image resolution. The narrow subbands (2 MHz) represent the better image resolution.

3.2.3. Super-Resolution Method (SCM)

In our previous study, the SCM was applied to a sensor array system with a chirp transmitted waveform signal and employed a famous beamforming technique to construct the image. It is effective to improve the image resolution by an eigenvalue analysis approach. In this study, the method was adaptively implemented on a single-element system with a short-pulse transmitted waveform signal, and the image was constructed based on the linear equation of the model.
The SCM was applied to the original image model, y = D x , for all rotation angles. A variance-covariance matrix was first determined by decomposing the echo signal (y) into 35 subbands with a bandwidth of 6 MHz . The eigenvalue and eigenvector were extracted from the matrix for eigenvalue analysis, and the eigenvalue was sorted in descending order to detect the signal and noise subspaces. Figure 11 shows the eigenvalue analysis by extracting an eigenvalue profile. From the profile, the eigenvector corresponding to the first eigenvector with a higher value spans a signal subspace. While the eigenvectors corresponding to the next eigenvalues consist of a noise subspace. By assigning each column of the matrix D in the image model ( y = D x ) as a steering vector, the noise for each grid position was evaluated for 25 rotation angles, resulting in an SCM profile. The final image was obtained by multiplying the profile with the previously numerically solved image.
To avoid the numerical computation of image-x, the original image model was modified by applying the compression process to matrix-D, resulting in
D H y = D H D x ,
y = E x ,
where y = D H y and E = D H D . Based on Equation (2), the SCM profile was computed for all rotation angles as in the previous strategy. Figure 12a,b show the SCM profile at a given rotation angle with different approximations. In the case of a single scatterer, the C value was set to 1 to compute the SCM profile. The varying C value affected the SCM profile with the appearance of an unwanted signal as shown in Figure 12c. The SCM profile was plotted in the range direction on a line crossing the scatterer position. B-mode images of a single scatterer (C = 1) are shown in Figure 13 with simple averaging and CF composite.

3.2.4. Proposed Method

From the results presented before, the image constructed by applying the weighted subband frequency compound has better resolution only in the range direction. While by applying a super-resolution method, the image quality has better improvement in the lateral direction. In the current super-resolution method, the image was obtained by multiplying the image x by the computed SCM profile for each angle of rotation. Therefore, we integrated both methods to obtain better image quality as our proposed method. The SCM profile was multiplied by the image obtained by applying the weighted frequency subbands method. Figure 14 shows B-mode images with simple averaging and CF compound, respectively. The B-mode image obtained by varying the bandwidth of subbands can be found in Figure 15 for comparison. To show the qualitative performance of the described methods, Figure 16 shows the amplitude profiles in range and lateral resolution. The cross-sectional profile was obtained by picking up a normalized intensity along the line passing through the target position in both directions of the image with CF compound. The label ‘LE’ represents the image obtained by numerically solving the linear equation of the image model. The method for combining weighted frequency subbands is referred to as ‘Freq. subbands’. The super-resolution method without compression is labeled as ‘SCM’, while the method with compression is labeled as ‘SCMcompr.’. The label ‘Proposed’ indicates the integration of frequency subbands compound and super-resolution methods.
To assess the effectiveness of the methods, we placed a single scatterer at various positions and performed a quantitative analysis and evaluation. A single target with the same properties and size was placed at nine positions in the imaging region. Three parameters were used to evaluate the image characteristics: signal-to-noise ratio (SNR), speckle level, and full width at half-maximum (FWHM). SNR in dB was calculated by dividing the averaged signal by the standard deviation of the noise. Speckle level in dB was determined by averaging the image noise. FWHM was calculated at 3 dB of the amplitude profile intersecting the single scatterer of the image with CF compound. Figure 17 shows the SNR, speckle level and FWHM associated with the error bar, respectively. The error bar was calculated by averaging the standard deviation over the number of samples (nine positions).

3.2.5. Evaluate the Methods with Multiple Scatterers

This study also investigated the imaging characteristics by placing multiple targets, specifically five scatterers, within the region of interest (ROI) instead of a single scatterer. The targets were identical in size and properties as the single scatterer. To compute the SCM profile for five scatterers, the C-value in Equation (A15) was set to 5. Figure 18 displays the B-mode images with different methods applied for multiple scatterers. The image obtained by the proposed method with integration between weighted frequency subbands compound and SCM with compression shows better resolution. The evaluation parameters for the case of mutiple scatterers can be found in Figure 19, consisting of the SNR, the speckle level and the FWHM.

3.3. Experimental Condition

To confirm the simulation results, we demonstrated the proposed method using experimental data. An array transducer system modeled the single element with a coding mask in this experiment. We used an experimental platform for medical ultrasound applications (RSYS0003, Microsonic Co., Ltd., Tokyo, Japan) to perform the transmit and receive sequence. The sampling frequency was 31.25 MHz, and the maximum number of samples was 1984. The ultrasound transducer used was a linear array probe (T0-1599, Nihon Dempa Koygo Co., Ltd., Tokyo, Japan) with 64 elements (0.2 mm thick and 0.115 mm spacing) transmitting a short pulse signal with a center frequency of 5.21 MHz. Following the ultrasound measurement, signal processing was conducted using MATLAB R2022a.
Figure 20 shows the experimental platform used in this work. We present the experimental results obtained using a soft tissue mimicking phantom (Kyoto Kagaku US-2 multipurpose phantom N-365), with a speed of sound of 1432 m/s (25 ° C) and an attenuation of 0.59 dB/cm/MHz. As illustrated in Figure 20, the phantom contains four string wires with the diameter of 0.1 mm. For high computational efficiency, we evaluated the single scatterer located approximately 21 mm from the probe.

3.4. Experimental Results

Each element of the array transducer separately transmitted a short pulse waveform which was received by all 64 elements. To model the coding mask, a set of random delay times was applied to shift each of the received echo signals based on the element order. Based on the image model, the vector-y was represented by a single echo signal that was obtained by summing up all the received echo signals. We provided 40 sets of random delay times that represent the number of rotations in the real single coding mask transducer. The pure received signal of single element consisting of a short pulse transmitted wave and backscattered wave is shown in Figure 21a. Several echo signals at different sets of random delay times can be seen in Figure 21b–d.
The matrix D, with a size of 171 × 51, was defined in the region of the desired scatterer position and for computational purposes. It was started by calculating the delay time between the element and each grid point of the ROI and it was added by a random delay time indicating the time interval within the coding mask of a certain thickness attached to the surface of the single transducer in real-time. The total delay time was used to shift the reference signal which is a pulse transmission waveform. The procedure was repeated for all elements and finally summed up to yield a single signal before being multiplied by a transfer function using frequency domain processing. After obtaining the vector t y and the matrix D for all sets, the image construction can be performed.
Based on the experimental data, the image was constructed with different methods as in the previous simulation. The residual coefficient for the numerical iteration solution of the image model equation was set to 0.01. A 15-subband with a subband width of 6 MHz was used to compute the SCM and the weighted frequency subband compound. Figure 22 shows the B-modes with simple averaging and CF compound over the 40 angles of rotation.

4. Discussion

Generally, ultrasound imaging has been performed by employing a sensor array, which results in more measurement data from each element to receive the backscattered signal. However, in the single transmitter/receiver system, it has a finite measurement limit. Therefore, the device was rotated at several different angles to collect more measurement data. As the device rotates, the coding mask also rotates along it, so that the new interference transmitted signal flows to the medium and collects the other information echo signal from the same scatterer. To obtain more varying new information, the angle of rotation is highly dependent on a randomized structure of the coding mask. Figure 4 shows a view of several measured echo signals at different rotation angles. More measurements affect the unique pixel characteristic of the image as shown in Figure 5.
Based on the results, the original image, which is constructed by solving the linear equation at each angle of rotation without any processing, has been improved. The weighted frequency subband composite shows the best image resolution in the range direction and higher speckle suppression compared to other methods. To achieve better spatial coding, the number of measurements in the single-element system was increased by increasing the number of rotations. Speckle suppression and spatial resolution improved as the number of rotations increased, as depicted in Figure 9 with simple averaging.
The image produced by applying the SCM was effective in terms of range direction resolution. The value of C is determined based on the number of scatterers. Higher values result in more noise, as shown in Figure 12c. Furthermore, Figure 13 confirms the effect of the D compression, which improved the image resolution in both directions compared to the SCM without compression, resulting in better speckle suppression performance. Therefore, we integrated both methods to enhance image quality. On the other hand, the subband width affects image resolution through the weight frequency compound, which is demonstrated by a narrow band. However, in the proposed method, the subband width has a minimal effect on image resolution, as shown in Figure 15. Therefore, we chose a subband width of 6 MHz for this study. For the final comparison of the methods proposed in this study, Figure 17 presents a quantitative evaluation of the image performance on the single scatterer.
The proposed method demonstrates superior image quality when applied to the five targets, as shown in Figure 18. The integrated method between weighted frequency subband compounds and super-resolution methods shows better image resolution by simple averaging or CF to compound over the number of rotations. The frequency subbands compound and the proposed method show almost the same level in speckle suppression. Figure 19 displays the evaluation based on SNR, level of speckle and FWHM of the multiple scatterers case. The SNR with multiple scatterers shows the same trend as the single scatterer case. In terms of the degree of speckle suppression, the frequency subband shows a slightly higher value compared to the proposed method. In addition, the proposed method shows the highest resolution in both directions as shown by FWHM. It is worth noting that the presence of multiple targets and the resulting reflected signals between scatterers can affect the evaluation parameters, as compared to the previous single target evaluation.
Instead of the simulation, we also demonstrated the proposed methods using the experimental data by modelling a single coding mask transducer with the available array transducer system. The constructed B-mode images in Figure 22 confirm the qualitative evaluation of the proposed method, which is effective in improving the image resolution. In the future, the quantitative evaluation of the proposed method using the experimental data obtained from the real single coding mask transducer system can also be achieved as the simulation results show. Furthermore, the current experimental results provide evidence that the proposed method can be implemented in the real single-coding mask system in the next stage of our research.

5. Conclusions

The study constructed an ultrasonic image of an encoding mask using a single oscillator. The imaging resolution was evaluated using several methods, including weighted frequency subbands, the super-resolution technique, and our proposed method by integrating both. The final image was obtained by simple averaging or using the coherence factor across 25 different angles of rotation. The highest resolution image was achieved through weighted frequency subband in the range direction. Meanwhile, a super-resolution method demonstrated the highest lateral resolution. Our proposed method combines weighted subband frequency compounding and super-resolution techniques, resulting in the highest quality image. The quality of image was assessed by calculating FWHM, SNR, and speckle level. In the future, experiments will be carried out with a real single-coding mask transducer system to confirm the current results with a qualitative and quantitative evaluation using the proposed method.

Author Contributions

Authors contributed as follows: performed the simulation and experiment: M.S. and E.N.; provided overall guidance: N.T.; provided other support: M.Y. All authors have read and agreed to the published version of the manuscript.

Funding

This work was supported by Tokyo Metropolitan University.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The raw data supporting the conclusions of this article will be made available by the authors on request.

Conflicts of Interest

The authors declare no conflict of interest.

Appendix A. Frequency Subband Compound

Frequency compounding is a common method for improving the resolution of ultrasound images. The original wide band of the measured IQ signal at different angles was subdivided into M narrow subbands as shown in Figure A1. Using the Equation (1), the vector x is calculated for each subband for all angles of rotation, resulting in the following tensor data X .
X = x 1 , 1 x 1 , 2 x 1 , N x 2 , 1 x 2 , 2 x 2 , N x M , 1 x M , 2 x M , N
where x i , j denotes the pixel value at subband-i and angle-j. Following [31], the frequency subbands compound is started by computing the variance covariance matrix-R that requires the following vector s j ,
s j = 1 N 1 l = 1 , l j N ( x i j ) l ,
and matrix R is given by
R = 1 N j = 1 N s j s j H + ϵ I ,
where H indicates Hermitian transpose, ϵ is diagonal loading parameter, and I represents an identify matrix. The number of rotation angles N is required for averaging matrix-R as well as for averaging the frequency subbands. The weight of frequency compound each angle is determined by
w j F = R 1 1 L 1 L R 1 1 L , 1 L = [ 1 , , 1 ] .
The frequency contraction is computed by:
y j = 1 M i = 1 M ( w j F ) H ( x i j ) .
This process is applied to all grids corresponding to the pixels of the ROI. Equation (A5) is computed for all angles of rotations for each grid.
Figure A1. Subbands compound; (a) original band with f c as a center of frequency; (b) extracted subbands with a different center of frequency.
Figure A1. Subbands compound; (a) original band with f c as a center of frequency; (b) extracted subbands with a different center of frequency.
Sensors 24 01496 g0a1

Appendix B. Super-Resolution Method (SCM)

Multiple echo signals y ( t ) from scatterers are received with transmitting signal s ( t ) . It is expressed in term of the convolution of the transmitted signal and impulse response h ( t ) of the channel as
y ( t ) = h ( τ ) u ( t τ ) d τ .
By assuming the received echo signals compress number of echo signals from C-scatterers with different delay times, hence an impulse response h ( t ) is defined by
h ( t ) = i = l C h i δ ( t τ i ) ,
where h i is a set of complex amplitudes, τ i represents a set of propagation delay times to be estimated and δ ( · ) is the Dirac delta function.
Considering multipath propagation, the correlator output z ( ζ ) is the superposition of scaled copies of the chirp signals auto correlation. The cross-correlation of echo with transmission signal s ( t ) is
v ( ζ ) = i = l C h i r ( ζ τ i ) + η ( ζ ) .
Therein, r ( ζ ) is the auto-correlation function of the chirp signal and η ( ζ ) is the cross correlation of the chirp signal and the noise process η ( ζ ) . The received signal z ( ζ ) is described in the base band as follows
z ( ζ ) = i = l C h i e j ω 0 τ i r ( ζ τ i ) + μ ( ζ ) .
Next, a delay profile vector is defined by z ( ζ ) = [ z ( ζ 1 ) z ( ζ 2 ) z ( ζ M ) ] T . With steering vector r ( τ i ) = [ r ( ζ 1 τ i ) r ( ζ 2 τ i ) r ( ζ M τ i ) ] T , thus z ( ζ ) might be expressed as
z ( ζ ) = i = l C h i e j ω 0 τ i r ( τ i ) + μ .
The steering vector r ( τ ) might calculated for any τ since the auto-correlation function r ( · ) is known. In terms of the array manifold matrix Γ = [ r ( τ 1 ) r ( τ 2 ) r ( τ C ) ] T and the gain vector, g = [ h 1 e j ω 0 τ 1 h 2 e j ω 0 τ 2 h C e j ω 0 τ C ] T , the delay profile vector takes the form
z = Γ g + μ .
The covariance matrix R = E { z z H } can also be written as
R = i , j E { h i h j * } e j ω 0 ( τ j τ i ) r ( τ i ) r ( τ j ) H + E { v v H } .
The following form of R will be used:
R = Γ G Γ + R n ,
where G is covariance matrix G = E { g g H } . The noise correlation matrix R n can be expressed as R n = E { μ μ H } = τ 2 R 0 , where R 0 is a Hermitian matrix whose k l th element is r ( ζ k ζ l ) . The SCM relies on a generalized eigendecomposition of the following form:
R e i = λ i R 0 e i : i = 1 , 2 , , M ,
and the validaty of the following assumption: The number of samples at the correlator output is greater than the number of paths (i.e., M > C ). Furthermore, the column of Γ are lineraly independent, G is nonsingular. Under these conditions, the signal covariance matrix R R n will have rank C, and R will have C generalized eigenvalues greater than σ 2 and M C generalized eigenvalue equal to σ 2 . The set of C eigenvector { e i } i = 0 C corresponding to the C largest generalized eigenvalues span the signal subspace, and the set of M C smallest generalized eigenvector { e i } i = C + 1 M corresponding to the C largest generalized eigenvalues span the noise subspace. It can be shown that the noise subspace is orthogonal to the columns of matrix Γ , i.e., the noise subspace is orthogonal to the steering vectors evaluated at the true delays. In order to estimate the delay, we use a measure of the ortyhoganality of the steering vector to { e i } i = C + 1 M . Accordingly, a super-resolution delay profile S ( τ ) based on the MUSIC algorithm might be defined as
S C M ( τ ) = r ( τ ) H R 0 1 r ( τ ) i = C + 1 M | r ( τ ) H e i | 2 .

Appendix C. Coherence Factor Beamformer

In this study, instead of simple averaging to compound the image over the number of rotation, we used coherence factor (CF). CF is an effective beamformer for removing the background signal [32]. The CF is defined by the following equation
C F = | i = 1 N x i | 2 N i = 1 N | x i | 2 ,
thus,
Y F = C F × i = 1 N x i .

References

  1. Synnevag, J.F.; Austeng, A.; Holm, S. Adaptive beamforming applied to medical ultrasound imaging. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2007, 54, 1606–1613. [Google Scholar] [CrossRef]
  2. Hasegawa, H. Improvement of range spatial resolution of medical ultrasound imaging by element-domain signal processing. Jpn. J. Appl. Phys. 2017, 56, 07JF02. [Google Scholar] [CrossRef]
  3. Danilouchkine, M.G.; van Neer, P.L.M.J.; Verweij, M.D.; Matte, G.M.; Vletter, W.B.; van der Steen, A.F.W.; de Jong, N. Single pulse frequency compounding protocol for superharmonic imaging. Phys. Med. Biol. 2013, 58, 4791–4805. [Google Scholar] [CrossRef]
  4. Youn, J.; Luijten, B.; Bo Stuart, M.; Eldar, Y.C.; van Sloun, R.J.G.; Arendt Jensen, J. Deep learning models for fast ultrasound localization microscopy. In Proceedings of the IEEE International Ultrasonics Symposium, Las Vegas, NV, USA, 7–11 September 2020; pp. 1–4. [Google Scholar]
  5. Matrone, G.; Savoia, A.S.; Caliano, G.; Magenes, G. The delay multiply and sum beamforming algorithm in ultrasound B-mode medical imaging. IEEE Trans. Med. Imaging 2015, 34, 940–949. [Google Scholar] [CrossRef] [PubMed]
  6. Sannou, F.; Nagaoka, R.; Hasegawa, H. Estimation of speed of sound using coherence factor and signal-to-noise ratio for improvement of performance of ultrasonic beamformer. Jpn. J. Appl. Phys. 2020, 59, SKKE14. [Google Scholar] [CrossRef]
  7. Camacho, J.; Parrilla, M.; Fritsch, C. Phase coherence imaging. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2009, 56, 958–974. [Google Scholar] [CrossRef]
  8. Misaridis, T.; Jensen, J.A. Use of modulated excitation signals in medical ultrasound. Part II: Design and performance for medical imaging applications. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2005, 52, 192–207. [Google Scholar] [CrossRef]
  9. Lan, Z.; Jin, L.; Feng, S.; Zheng, C.; Han, Z.; Peng, H. Joint generalized coherence factor and minimum variance beamformer for synthetic aperture ultrasound imaging. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2021, 68, 1167–1183. [Google Scholar] [CrossRef] [PubMed]
  10. Yan, J.; Wang, B.; Riemer, K.; Shearer, J.H.; Lerendegui, M.; Toulemonde, M.; Rowlands, C.J.; Weinberg, P.D.; Tang, M.-X. Fast 3D super-resolution ultrasound with adaptive weight-based beamforming. IEEE Trans. Biomed. Eng. 2023, 70, 2752–2761. [Google Scholar] [CrossRef]
  11. Tanter, M.; Fink, M. Ultrafast imaging in biomedical ultrasound. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2014, 61, 102–119. [Google Scholar] [CrossRef]
  12. Nguyen, N.Q.; Prager, R.W. A spatial coherence approach to minimum variance beamforming for plane-wave compounding. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2018, 65, 522–534. [Google Scholar] [CrossRef]
  13. Ni, P.; Lee, H.N. High-resolution ultrasound imaging using random interference. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2020, 67, 1785–1799. [Google Scholar] [CrossRef] [PubMed]
  14. Sternini, S.; Pau, A.; Di Scalea, F.L. Minimum-variance imaging in plates using guided-wave-mode beamforming. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2019, 66, 1906–1919. [Google Scholar] [CrossRef]
  15. Janjic, J.; Kruizinga, P.; van der Meulen, P.; Springeling, G.; Mastik, F.; Leus, G.; Bosch, J.G.; van der Steen, A.F.W.; van Soest, G. Structured ultrasound microscopy. Appl. Phys. Lett. 2018, 112, 251901. [Google Scholar] [CrossRef]
  16. Kruizinga, P.; van der Meulen, P.; Fedjajevs, A.; Mastik, F.; Springeling, G.; de Jong, N.; Bosch, J.G.; Leus, G. Compressive 3D ultrasound imaging using a single sensor. Sci. Adv. 2017, 3, e170142. [Google Scholar] [CrossRef] [PubMed]
  17. Sheng, J.; Cai, H.; Wang, Y.; Chen, X.; Xu, Y. Improved Exponential Phase Mask for Generating Defocus Invariance of Wavefront Coding Systems. Appl. Sci. 2022, 12, 5290. [Google Scholar] [CrossRef]
  18. Luijten, B.; Cohen, R.; de Bruijn, F.J.; Schmeitz, H.A.W.; Mischi, M.; Eldar, Y.C.; van Sloun, R.J.G. Adaptive ultrasound beamforming using deep learning. IEEE Trans. Med. Imaging 2020, 39, 3967–3978. [Google Scholar] [CrossRef]
  19. Oelze, M.L. Bandwidth and resolution enhancement through pulse compression. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2007, 54, 768–781. [Google Scholar] [CrossRef] [PubMed]
  20. Prieur, F.; Rindal, O.M.H.; Austeng, A. Signal coherence and image amplitude with the filtered delay multiply and sum beamformer. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2018, 65, 1133–1140. [Google Scholar] [CrossRef]
  21. Nilsen, C.C.; Holm, S. Wiener beamforming and the coherence factor in ultrasound imaging. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2010, 57, 1329–1346. [Google Scholar] [CrossRef]
  22. Shen, C.-C.; Xing, Y.-Q.; Jeng, G. Autocorrelation-based generalized coherence factor for low-complexity adaptive beamforming. Ultrasonics 2016, 72, 177–183. [Google Scholar] [CrossRef]
  23. Capon, J. High-resolution frequency-wavenumber spectrum analysis. Proc. IEEE 1969, 57, 1408–1418. [Google Scholar] [CrossRef]
  24. Vignom, F.; Burcher, M.R. Capon beamforming in medical ultrasound imaging with focused beams. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2008, 55, 619–628. [Google Scholar] [CrossRef]
  25. Asl, B.M.; Mahloojifar, A. A low-complexity adaptive beamformer for ultrasound imaging using structured covariance matrix. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2012, 59, 660–667. [Google Scholar] [CrossRef]
  26. Labyed, Y.; Huang, L. Ultrasound time-reversal MUSIC imaging with diffraction and attenuation compensation. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2012, 59, 2186–2200. [Google Scholar] [CrossRef] [PubMed]
  27. Labyed, Y.; Huang, L. Super-resolution ultrasound imaging using a phase-coherent MUSIC method with compensation for the phase response of transducer elements. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2013, 60, 1048–1060. [Google Scholar] [CrossRef] [PubMed]
  28. Fujiwara, M.; Okubo, K.; Tagawa, N. Novel technique for high resolution ultrasound super resolution FM-Chirp Correlation Method (SCM). In Proceedings of the IEEE International Ultrasonics Symposium, Rome, Italy, 20–23 September 2009; pp. 2390–2393. [Google Scholar]
  29. Syaryadhi, M.; Tagawa, N.; Yang, M. Weighted frequency subband compounding in ultrasonic imaging sensor consisting of a single transducer and a random coding mask. In Proceedings of the 2023 IEEE Applied Sensing Conference (APSCON), Bangalore, India, 23–25 January 2023; pp. 1–3. [Google Scholar]
  30. Syaryadhi, M.; Zheng, J.; Tagawa, N.; Yang, M. Super-resolution based on frequency dependence of echo phase rotation in 3D ultrasound imaging by spatially encoded transmit/receive. In Proceedings of the 2023 International Congress on Ultrasonics, Beijing, China, 18–21 September 2023. [Google Scholar]
  31. Zheng, J.; Tagawa, N.; Yoshizawa, M.; Irie, T. Plane wave beamforming with adaptively weighted frequency compound using bandpass filtering. Jpn. J. Appl. Phys. 2021, 60, SDDB08. [Google Scholar] [CrossRef]
  32. Li, P.C.; Li, M.L. Adaptive imaging using the generalized coherence factor. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2003, 50, 128–141. [Google Scholar]
Figure 1. Proposed method by integrating the weighted frequency subband compound and super-resolution method (SCM).
Figure 1. Proposed method by integrating the weighted frequency subband compound and super-resolution method (SCM).
Sensors 24 01496 g001
Figure 2. Proposed simulation model with (a) single scatterer, (b) multiple scatterers, (c) side view of coding mask attached to top of transducer, screenshot of FEM simulation (d) without coding mask and (e) with coding mask of single scatterer.
Figure 2. Proposed simulation model with (a) single scatterer, (b) multiple scatterers, (c) side view of coding mask attached to top of transducer, screenshot of FEM simulation (d) without coding mask and (e) with coding mask of single scatterer.
Sensors 24 01496 g002
Figure 3. (a) Short pulse transmission waveform and (b) its center of frequency at 7 MHz.
Figure 3. (a) Short pulse transmission waveform and (b) its center of frequency at 7 MHz.
Sensors 24 01496 g003
Figure 4. Several echo signals at different angles of rotation; (a) angle#1; (b) angle#2; (c) angle#3; and (d) angle#4.
Figure 4. Several echo signals at different angles of rotation; (a) angle#1; (b) angle#2; (c) angle#3; and (d) angle#4.
Sensors 24 01496 g004
Figure 5. B-mode images constructed by numerically solving the linear equation (LE) with different numbers of rotations, simple averaging (top) and CF composite (bottom); (a) five-rotations, (b) 15-rotations, and (c) 25-rotations. Amplitude profile crossing the scatterer position in (d) range and (e) lateral direction.
Figure 5. B-mode images constructed by numerically solving the linear equation (LE) with different numbers of rotations, simple averaging (top) and CF composite (bottom); (a) five-rotations, (b) 15-rotations, and (c) 25-rotations. Amplitude profile crossing the scatterer position in (d) range and (e) lateral direction.
Sensors 24 01496 g005
Figure 6. Subband decomposition of a wideband base signal (a) into several narrow subband signals in (bf).
Figure 6. Subband decomposition of a wideband base signal (a) into several narrow subband signals in (bf).
Sensors 24 01496 g006
Figure 7. B-modes on different subbands with the same rotation angle; (a) sub-1, (b) sub-8, (c) sub-17, (d) sub-25, (e) sub-35.
Figure 7. B-modes on different subbands with the same rotation angle; (a) sub-1, (b) sub-8, (c) sub-17, (d) sub-25, (e) sub-35.
Sensors 24 01496 g007
Figure 8. B-mode images by applying the frequency subbands compound, simple averaging (up) and CF compound (down); (a) averaged subbands and (b) weighted frequency subbands compound.
Figure 8. B-mode images by applying the frequency subbands compound, simple averaging (up) and CF compound (down); (a) averaged subbands and (b) weighted frequency subbands compound.
Sensors 24 01496 g008
Figure 9. B-mode images by weighted frequency subband method with simple averaging for (a) five-rotations, (b) 15-rotations, and (c) 25-rotations.
Figure 9. B-mode images by weighted frequency subband method with simple averaging for (a) five-rotations, (b) 15-rotations, and (c) 25-rotations.
Sensors 24 01496 g009
Figure 10. Amplitude profile of B-mode images with different bandwidth of subband; (a) range and (b) lateral direction.
Figure 10. Amplitude profile of B-mode images with different bandwidth of subband; (a) range and (b) lateral direction.
Sensors 24 01496 g010
Figure 11. Eigenvalue analysis; (a) eigenvalue distribution in descending order with (b) the first eigenvector and (c) the second eigenvector corresponding to its eigenvalue, respectively.
Figure 11. Eigenvalue analysis; (a) eigenvalue distribution in descending order with (b) the first eigenvector and (c) the second eigenvector corresponding to its eigenvalue, respectively.
Sensors 24 01496 g011
Figure 12. SCM profile (a) without compression and (b) with D-compression. Effect of different C-values on (c) SCM profile in range direction.
Figure 12. SCM profile (a) without compression and (b) with D-compression. Effect of different C-values on (c) SCM profile in range direction.
Sensors 24 01496 g012
Figure 13. B-mode images by applying the super-resolution method, simple averaging (top) and CF compound (bottom), (a) without compression; (b) with the compression process.
Figure 13. B-mode images by applying the super-resolution method, simple averaging (top) and CF compound (bottom), (a) without compression; (b) with the compression process.
Sensors 24 01496 g013
Figure 14. B-mode images by integrating weighted frequency subbands and SCM with (a) simple averaging; (b) CF compound.
Figure 14. B-mode images by integrating weighted frequency subbands and SCM with (a) simple averaging; (b) CF compound.
Sensors 24 01496 g014
Figure 15. B-mode images constructed by the proposed method with the different width of the subbands; (a) 4 MHz; (b) 6 MHz; (c) 8 MHz with simple averaging (top) and CF (bottom).
Figure 15. B-mode images constructed by the proposed method with the different width of the subbands; (a) 4 MHz; (b) 6 MHz; (c) 8 MHz with simple averaging (top) and CF (bottom).
Sensors 24 01496 g015
Figure 16. Amplitude profile of B-mode image of CF compound in (a) range and (b) lateral direction.
Figure 16. Amplitude profile of B-mode image of CF compound in (a) range and (b) lateral direction.
Sensors 24 01496 g016
Figure 17. Evaluation parameters with the error bar; (a) signal-to-noise ratio (SNR); (b) speckle level; and (c) FWHM of a single scatterer case placed at nine different positions.
Figure 17. Evaluation parameters with the error bar; (a) signal-to-noise ratio (SNR); (b) speckle level; and (c) FWHM of a single scatterer case placed at nine different positions.
Sensors 24 01496 g017
Figure 18. B-mode images with multiple scatterers with simple averaging (top) and CF compund (bottom), (a) solving the original linear equation model, (b) weighted frequency subband compound; (c) applying the super-resolution method; (d) super-resolution method with compression; (e) proposed method.
Figure 18. B-mode images with multiple scatterers with simple averaging (top) and CF compund (bottom), (a) solving the original linear equation model, (b) weighted frequency subband compound; (c) applying the super-resolution method; (d) super-resolution method with compression; (e) proposed method.
Sensors 24 01496 g018
Figure 19. Evaluation parameters for multiple scatterer case; (a) signal-to-noise ratio (SNR); (b) speckle level; and (c) FWHM.
Figure 19. Evaluation parameters for multiple scatterer case; (a) signal-to-noise ratio (SNR); (b) speckle level; and (c) FWHM.
Sensors 24 01496 g019
Figure 20. Experimental setting for the soft tissue-mimicking phantom.
Figure 20. Experimental setting for the soft tissue-mimicking phantom.
Sensors 24 01496 g020
Figure 21. Echo signals of the experiment; (a) pure received echo signal of a single element consisting of a short pulse waveform; sum of received echo signals for all elements with (b) set#1; (c) set#2 and (d) set#3.
Figure 21. Echo signals of the experiment; (a) pure received echo signal of a single element consisting of a short pulse waveform; sum of received echo signals for all elements with (b) set#1; (c) set#2 and (d) set#3.
Sensors 24 01496 g021aSensors 24 01496 g021b
Figure 22. B-mode images using experimental data with simple averaging (top) and CF compund (bottom), (a) numerical linear equation solution method, (b) weighted frequency subband compound; (c) super-resolution method; (d) super-resolution method with compression; (e) proposed method.
Figure 22. B-mode images using experimental data with simple averaging (top) and CF compund (bottom), (a) numerical linear equation solution method, (b) weighted frequency subband compound; (c) super-resolution method; (d) super-resolution method with compression; (e) proposed method.
Sensors 24 01496 g022
Table 1. Physical parameters used in the simulation.
Table 1. Physical parameters used in the simulation.
ParameterValue
Short transmission pulse voltage50 Volts
Device length5 mm
Center of frequency7 MHz
Backing thickness1.25 mm
PZT transducer:
- thickness0.165 mm
- density7500 kg/m3
- dielectric constant1700
Coding mask:
- materialPlastic
- density1060 kg/m3
- bulk velocity2340 m/s
- number of patches30
- randomized thickness0.083–0.335 mm/(0.25 λ –1.00 λ )
Scatterer radius0.1 mm
Distance from the scatterer to the surface of the transducer2.5 mm
Region of interest (ROI) size2 mm × 2 mm
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Syaryadhi, M.; Nakazawa, E.; Tagawa, N.; Yang, M. Evaluating a 3D Ultrasound Imaging Resolution of Single Transmitter/Receiver with Coding Mask by Extracting Phase Information. Sensors 2024, 24, 1496. https://doi.org/10.3390/s24051496

AMA Style

Syaryadhi M, Nakazawa E, Tagawa N, Yang M. Evaluating a 3D Ultrasound Imaging Resolution of Single Transmitter/Receiver with Coding Mask by Extracting Phase Information. Sensors. 2024; 24(5):1496. https://doi.org/10.3390/s24051496

Chicago/Turabian Style

Syaryadhi, Mohammad, Eiko Nakazawa, Norio Tagawa, and Ming Yang. 2024. "Evaluating a 3D Ultrasound Imaging Resolution of Single Transmitter/Receiver with Coding Mask by Extracting Phase Information" Sensors 24, no. 5: 1496. https://doi.org/10.3390/s24051496