Article

Two-Channel Detecting Sensor with Signal Cross-Correlation for FTIR Instruments

by
Krzysztof Achtenberg
*,
Janusz Mikołajczyk
and
Zbigniew Bielecki
Institute of Optoelectronics, Military University of Technology, 00-908 Warsaw, Poland
*
Author to whom correspondence should be addressed.
Sensors 2022, 22(22), 8919; https://doi.org/10.3390/s22228919
Submission received: 14 October 2022 / Revised: 13 November 2022 / Accepted: 16 November 2022 / Published: 18 November 2022

Abstract

:
This paper’s purpose was to demonstrate a performance of a novel approach in a low-noise optical sensor for an FTIR spectrometer. Methods: Compared to the standard FTIR detection setup, our sensor ensures a higher signal-to-noise ratio (SNR) and lower signal standard deviation by reducing the uncorrelated noise components (e.g., thermal and 1/f noises of the detection module). Its construction is based on two-channel detection modules and a processing unit with implemented cross-correlation signal analyses. Each module was built of LWIR HgCdTe photodiodes and low-noise transimpedance amplifiers. Results: the experiments demonstrated a decrease in a signal standard deviation of about 1.7 times with a 10 dB-improvement in the SNR. Conclusion: this result indicates our sensor’s main benefit, especially in registered “weak” and noisy interferograms.

1. Introduction

Fourier Transform Infrared spectrometers (FTIR) are widely used for chemical or optical compound analyses. It is a universal research tool to identify pure substances, mixtures, impurities, or composites [1]. More practical FTIR applications include the analysis of pharmaceuticals, food, semiconductors, polymers, plastics, biological specimens, the environment, and hazardous materials [2,3,4,5,6]. The measurement virtues depend mainly on the performances of the applied FTIR instrument, e.g., resolution, precision, and floor (background) noise.
In practice, many noise sources essentially affect both spectral and power domains of the registered spectra. Usually, the quality of the spectral analysis is determined by the ratio of signal-to-noise (SNR) [7,8,9]. Its value depends on operation scenarios and FTIR instrument configuration and can be expressed by [10]:
S N R = U V · Θ · Δ υ · t 1 2 · ξ N E P
where Θ—throughput, ξ—efficiency, Δν—resolution, t—measurement time, NEP—noise equivalent power of detection module, and UV—detector output signal. Equation (1) describes the main spectroscopic trading rule specifying SNR, resolution, and measurement–time relationships.
If detection module noise is predominant, spectral SNR increases proportionally with the FTIR throughput. This parameter depends on the area of the projected mirror and spectral resolving power (ratio of the maximum analysed wavenumber and spectral resolution). It can also be related to a collimating system’s divergence angle, aperture, and focal length.
In an ideal setup, the total noise is substantially limited by the noise of the used detection module regardless of the signal level. The simplest and most common method to increase the SNR is to perform a series of measurements and average the results. For the coherent signals and the random noise, the SNR increases with the square root of the number of averaged interferograms. However, if the noise level of the detection module is below the resolution of the used analogue-to-digital converter (ADC) (below the least significant bit of the ADC), this procedure does not improve the SNR, and quantisation noise in the spectrum is noticed. Applying a gain-ranging amplifier with ADC minimises this noise’s impact by setting the interferogram’s dynamic range. In this configuration, the gain is switched in the time interval between sample points. Hirschfeld [11] applied gain ranging as a series of steps approximating the interferogram envelope. For the registered interferogram, the SNRI is given by [12]:
S N R F T I R I ( T N I ) 1 / 2 ,
where T is the measurement time and NI is the number of interferogram points determined by NI = xmax/ds (xmax—the maximum mirror retardation distance, ds—the sampling period).
The output signal of the detection module depends on parameters individually defined for a specific model of FTIR spectrometer, e.g., transmission and geometry of optics, detector’s characteristics, the speed of the moving mirror, and signal bandwidth. The average radiation power registered by FTIR’s detector equals:
P I = V G R ,
where V is the interferogram wing’s mean signal, G is the gain of the detection module’s amplifier, and R is the photodetector responsivity. In the best practice, the SNR value of FTIR is optimised to provide a level defined by the minimum noise level of the detection module:
S N R d e t e c t i o n I = P I N d e t e c i i o n ,
where Ndetection is the power spectral density (PSD) of the detection module noise (both photodetector and amplifier).
During conversion to a spectral area, the spectral signal-to-noise ratio (SNRS) increases by:
S N R S = S N R I × N I 1 / 2 .
The value of SNRS can also be defined by:
S N R S = S N R I × N I 1 / 2 × ρ ( υ ) ρ t o t a l ,
where ρ(υ) and ρtotal are the spectral power density at the analysed wavenumbers and the total spectral power, respectively.
The FTIR spectrometer operation is accepted if the spectral SNR is not below 105. We can determine the spectral noise N(ν) during the ratio measurement of two spectra (Ta and Tb) by calculating:
N ( υ ) = 1 T a ( υ ) T b ( υ ) .
Its level corresponds to the peak-to-peak value (Np-p—difference between the maximum and minimum values) or the root–mean–square value (Nrms—a standard deviation of all data points) (Figure 1). N(ν) equals 0 in the tested wavelength range for a noiseless device. Usually, the FTIR manufacturer’s datasheet specifies the noise feature by a 100% line defined by 100 × Ta/Tb. Considering values of the spectral noise Nrms, the spectral SNRS can also be described by:
S N R s = 1 N r m s .
A more practical method of SNR determination analyses the shape of the registered transmission spectrum with the peak of spectral transmission and the floor noise. In this case, the noise level can be defined at the base of this spectrum. In this situation, the SNR equals:
S N R = µ ( υ ) r m s [ T ( υ ) µ ( υ ) ] ,
where µ is the ideal mean (“noiseless”) transmission spectrum and T is the measured one. This equation can be transformed into:
S N R = µ ( υ ) µ ( υ ) r m s ( T ( υ ) μ ( υ ) 1 ) = 1 r m s ( T ( υ ) μ ( υ ) 1 ) .
We can visualise the detection noise impact on spectral measurements with an FTIR spectrometer based on the transmission characteristics presented in Figure 1. The detection limit and the noise floor define the detection resolution and the spread of noise amplitude, respectively. The same noise parameters are a critical issue in the measurements of transmission characteristics and absorption lines.
Usually, the primary noise sources of detection are a photodetector, its power supply, and a preamplifier [13]. Currently, much work is being carried out to obtain ultra-low-noise photoreceivers. We ensured the maximum SNR detection value based on photodetectors’ noise models and read-out electronics. Finally, it allows the selection of the detector’s operating point and electronics configuration, determining the actual performance of the detection module, e.g., operating temperature, biasing voltage, signal gain, and bandwidth. It is intended that the noise of the detection module is determined solely by the detector one [14]. However, detector technology development poses a significant challenge for designing “noiseless” read-out electronics. In papers [15,16,17,18], we analysed some advanced development methods to minimise the noise influence of such electronics.
However, the presented paper describes a novel approach in this field of knowledge because it defines new possibilities for FTIR spectroscopy with a unique low-noise optical sensor. The literature analysis showed that many methods of increasing the SNR ratio have already been developed [12,19,20,21,22]. For the detection module, a common way was to select a low-noise detector and reduce its operating temperature [23]. A more complex method applies a dual-beam FTIR spectrometer with a single detection module [23,24,25]. It minimises the common background and decreases the dynamic range of the measured signal. The obtained reduction of the dual-beam interferogram amplitude decreases quantisation noise contribution by effectively using the ADC dynamic range. For example, the SNR of the dual-beam systems described in [23,25] were a factor of 9 and 6 better than that of the single-beam one.
However, none of the referenced papers has considered the possibility of using the described configuration of the detection module and procedure of signal analyses. To the best of the authors’ knowledge, this work demonstrates, for the first time, that:
  • A two-channel detection module with cross-correlation can be used as an optical detection unit in an FTIR setup;
  • The sensor with signal cross-correlation improves the signal-to-noise ratio of the FTIR spectrometer and is more effective than spectra averaging.

2. Materials and Methods

2.1. Experimental Design

The proposed method is based on a two-channel optical detection with a signal cross-correlation approach. Although this is a known method, it has not yet been applied to decrease floor noise in an FTIR spectrometer. The core of this method is that the measurements are performed simultaneously with two similar detection channels [26]. Based on the statistical character of signals, spectrum extraction of the valid signal from one noise is possible [27]. Figure 2 presents a scheme and a photo of an FTIR spectrometer with an optical sensor consisting of two photodetectors, read-out electronics, and a signal processing unit with implemented cross-correlation algorithm (to calculate a cross-power spectrum density—CPSD).
The cross-correlation technique is based on the simultaneous measurement of two signals, V1(t) and V2(t) [28,29]. These signals contain the common signal Vs(t) and uncorrelated noise components of Vn1(t) and Vn2(t):
V 1 ( t ) = V s ( t ) + V n 1 ( t ) ,
V 2 ( t ) = V s ( t ) + V n 2 ( t ) .
The cross-correlated spectrum S12(f) is calculated by multiplying the Fourier transforms of the first signal F[V1(t)] by a complex conjugate of the second one F[V2(t)]:
S 12 ( f ) = F [ V s ( t ) ] × F [ V 2 ( t ) ] * .
The procedure is averaged, and the uncorrelated components of these signals are reduced:
S 12 ( f ) a v g = n = 1 N S ( f ) N .  
However, there is a need to perform more averages to obtain smooth spectra compared to FTIR standard averaging operated with a single detector.
The result of these calculations has the form of the frequency spectrum. This spectrum is transformed into the wavenumber domain in FTIR spectrometers by applying a reference laser with a unique wavelength. The laser beam travels the same optical path through the interferometer generating an interferogram as a sinusoidal signal. This signal determines the interferometer displacement and defines the wavenumber [30]. If there is no possibility to apply reference laser signals, the common practice is to use some special calibration samples with a well-defined spectral position of absorption lines [31].
The described sensor is mounted into the radiation path of the FTIR spectrometer, and the FTIR output beam is focused on its photodetectors using an off-axis parabolic mirror. A high-resolution data acquisition card and PC software digitise the photodetectors output signals and perform calculations.
The designed FTIR setup requires the application of two photodetectors with similar spectral characteristics of relative responsivity. Any differences in these characteristics cause significant distortions in the results of the cross-correlation procedure. These differences can generate a high-level spread of cross-spectrum products in some wavelength ranges, leading to adulterations in the final results.
In a classical FTIR detection setup, the parabolic mirror focuses optical radiation on an active area of a single detector. In the case of the described sensor, irradiation of an active surface of two separate detectors causes a decrease in the registered optical signal. Theoretically, better technology can use a two-segment detector in which registered optical power depends on the chip dimensions. Each segment detects a part power of FTIR light.

2.2. Sensor Design

There are some critical issues in analysing a sensor construction. For example, the applied detectors should have high spectral detectivity and broad signal bandwidth. These parameters define, e.g., the SNR level and the limitation of radiation analyses in the high wavenumber region, respectively. The noise level also depends on a detector’s read-out electronics, and some aspects of its low-noise design were described in the literature [32,33].

2.3. Photodetectors

The applied HgCdTe heterostructure photodiodes were manufactured during the same technology process. These photodiodes with an optical area of 0.5 × 0.5 mm2 (TO-8 package) operated at room temperature with no biasing voltage [34,35]. Their similar relative responsivity has a peak value at a wavelength of λpeak = 7 µm (Figure 3).

2.4. Read-Out Electronics

The constructed read-out electronics consist of two identical signal channels based on the typical opamps configurations. The photocurrent signal from the detector is converted by a transimpedance amplifier (TIA) and gained by a voltage amplifier (VA). Figure 4 presents a schematic of the single signal channel. Its key parameters are a transimpedance of 15 × 105 V/A and −3 dB bandwidth in the 0.1 Hz–5 kHz range.
The channel output signals were connected to a PC-mounted data acquisition card from the National Instruments, model PCI-4462. It has four simultaneous sampling inputs with built-in anti-aliasing filters providing a 24-bit resolution and a 204.8 kS/s sampling frequency. The digitised interferograms are stored in the PC and transformed into the spectrum domain (power spectrum density and cross-power spectrum density). Finally, the spectrum can be expressed in the wavenumbers or wavelengths domain using, e.g., calibrated absorbers.

3. Experiment Results

3.1. FTIR Spectrometer with a Classical Detection Setup

In the first step of the work, the noise power spectral density of two signal channels (CH1 and CH2) was measured separately. It allows determining the fundamental limit considering floor noise for each of them. Figure 5 presents the measured noise characteristics in which 1/f and thermal noise regions can be defined. In comparison, the noise limit of the CH1 detection channel is a little worse than CH2’s, resulting from different values of the detectors’ resistances. The 3.2 × 10−21 A2/Hz and 5 × 10−21 A2/Hz PSD values of thermal noise at 1 kHz are consistent with the photodiodes resistances.
Both channels were placed into an optical beam of an FTIR spectrometer, model Interspec 402-X [36]. The position of the photodiodes was adjusted to have the same amplitude of the registered signals. Figure 6 presents an example of interferograms and their power spectral density. The shape of the recorded spectra is determined by the spectral characteristics of the FTIR radiation source, attenuation of an optical path (optics and air transmission), and the spectral responsivity of HgCdTe photodiodes (without any absorbing sample).
The registered interferograms were characterised by a low amplitude difference and a phase shift (time delay) generated by photodetectors and their read-out electronics (different signal phase-frequency response). A difference in the noise levels of the used detectors was also observed. In practice, it is impossible to manufacture two detecting structures with perfectly even parameters, e.g., responsivity and noise.

3.2. FTIR Spectrometer with the Designed Sensor

The preliminary experiment defined the performance of cross-power spectrum analyses in an FTIR spectrometer. The influence of some key factors (the FFT algorithm parameters and averaging time) on signal conditioning, acquisition, or transformation was also defined. These analyses are based on signals registered using a classical FTIR setup (with one detector) and the designed two-channel system with the CPSD procedure. The classical setup uses autocorrelation of the one detector output signal to determine its PSD (S11). In contrast, the CPSD is simultaneously based on signals registered by both detectors (S12).
To validate the performance of the measurement procedure, the influence of the phase shift and amplitude differences of the two detectors’ signals on the CPSD procedure was analysed. Figure 7a,b illustrate interferograms and their cross-spectra for two values of phase shifts (0° and 180°) using an analogue phase shifter and different amplitude ratios (1/1, 1/2, 1/3) setting the optical beam alignment. The phase shift was provided at the amplifier output using an opamp-based analogue circuit described in [37].
The interferograms’ phase shift (time delay) does not influence the cross-spectra. A significant change in cross-spectra was observed based on the results for different amplitude ratios. The highest level of CPSD was for the interferograms with the same amplitudes.
The cross spectra were also determined for different averaging times (10 s, 60 s, 600 s, and 3600 s). These spectra were compared with averaged PSD for one detector’s setup (NO CORR.—green line) (Figure 8a). We observed a noticeable decrease in the floor noise level with increasing averaging time. However, there is a slight reduction in the spread of noise amplitude. Noise influence can be spectrally filtrated by the FFT procedure, defined by its resolution bandwidth (RBW). The resolution is determined by the number of data points and sampling frequency. Figure 8b presents CPSD values for different RBW and averaging times. The better RBW corresponds to a high resolution in the spectrum, while for limited stored data, it also limits the maximal number of averaging. There are also observed differences in the theoretical analyses of RBW’s influence on noise reduction with the experimental results. The measured noise level is higher because of “real” signal parameters (e.g., correlated components of signal) and “non-ideal” applied devices (e.g., transimpedance gain characteristics).
Figure 9 presents the influence of a linear averaging of segmented data points (1-h data) on registered spectra. However, this procedure “smooths” the noise spectrum and increases the floor noise. It is due to the shorter calculated time of each cross-spectrum.
It was shown that the application of data segmentation (averaging many cross-spectra) compromises the reduction of floor noise or standard deviation σ. However, the proposed averaging provides a better standard deviation and floor noise than the single detector results. Using this method, the σ was about 1.7 times lower, with a 10 dB improvement of the SNR.

3.3. Practical Validation of the Designed FTIR Setup

The practical advantages of the designed sensor in FTIR spectra measurement were defined during the transmission investigation of a foil sample (Polyvinyl Chloride containing a plasticiser). The transmission was determined using the same FTIR spectrometer (model Interspec 402-X, Interspectrum OU, Tartumaa, Estonia) equipped with the sensor but using auto-correlation of the signal registered by one channel and cross-correlation with both channels. The RBW was set to 1.53 Hz by fs = 25 kS/s and 214 FFT points during that research. Figure 10a presents a determined transmission of the sample with the designed sensor. These characteristics were drawn in the frequency domain, comparing the results obtained in the two-detector (CROSS-CORR.) and one-detector (NO CORR.) configuration of FTIR detection to simplify analyses. The HgCdTe detection modules were adjusted into the optical path to ensure the amplitude difference of interferograms below 1%.
The sensor collected data for 1 h. The shape of the foil transmission characteristics is similar, but the most remarkable differences are in the low-frequency range. We observe a large spread of transmission values resulting from the division of two low signals corresponding to the noise level after the cross-correlation operation and an expansion analysis range with the 1/f noise decrease of the read-out electronics. Low transmission in the 650 Hz–700 Hz range shows a better detection limit of the “cross-correlation sensor”. It is an essential virtue in spectral absorption line measurement in which both ultra-low level and ultra-low spread of floor noise are required. These parameters directly influence the detection limit and the resolution of spectral analyses. That is why further analyses concern the foil absorbance. The results were compared with the data determined by the FTIR spectrometer equipped with its standard detection module with a pyroelectric photodetector (PYR–FTIR). Figure 10b compares the spectral absorbances of the foil registered by the PYR detection module and the designed sensor (with two detection channels—CROSS-CORR., and a single one—NO CORR.). The application of the designed sensor in FTIR spectrometry provides better dynamic sensitivity compared to the standard setup with the PYR–FTIR module. We can distinguish absorption lines more clearly using our sensor. This virtue is essential, e.g., during analyses of the spectral blocking filters or optical high-attenuated compounds.

4. Conclusions

The new approach to analysing and processing signals in FTIR instruments was described in this paper. The developed two-channel sensor can be used mutually with others dedicated to improving SNR, e.g., detector cooling. Its operation is based on signal cross-correlation that minimises thermal noise and reduces the 1/f one, which is typically problematic in the standard configuration when uncooled or biased detectors are used. The proposed sensor can recover the interferogram signals from unexpected noise. Most noise sources in the described sensor configuration are uncorrelated, so cross-correlation is a perfect algorithm to reduce them. The paper also analysed the advantages and disadvantages of cross-correlation operation considering photodetector characteristics and a two-channel detection setup.
A direct comparison of results from the same averaging time speaks for the cross-correlation procedure. The obtained spectra were characterised by a standard deviation about 1.7 times lower, with a 10 dB improvement of the SNR value comparing a “single-channel” classical setup. For FTIR systems, the designed sensor ensures a wider wavelength span and improves measurement precision in the case of noisy or low-amplitude interferograms. It becomes a practical tool for detecting and selecting the absorption lines of the analysed compounds.

Author Contributions

Conceptualisation, K.A. and J.M.; methodology, K.A. and J.M.; software, K.A.; validation, J.M. and Z.B.; formal analysis, J.M. and Z.B.; investigation, K.A.; data curation, K.A.; writing—original draft preparation, K.A.; writing—review and editing, K.A., J.M. and Z.B.; supervision, Z.B. All authors participated in the discussion and proofread the work. All authors have read and agreed to the published version of the manuscript.

Funding

The experimental part was carried out at the Institute of Optoelectronics MUT Laboratory and supported in the frame of grant No. UGB-22/732.

Institutional Review Board Statement

No applicable.

Informed Consent Statement

No applicable.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Kowalczuk, D.; Pitucha, M. Application of FTIR Method for the Assessment of Immobilisation of Active Substances in the Matrix of Biomedical Materials. Materials 2019, 12, 2972. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  2. Rachah, A.; Reksen, O.; Tafintseva, V.; Stehr, F.J.M.; Rukke, E.-O.; Prestløkken, E.; Martin, A.; Kohler, A.; Afseth, N.K. Exploring Dry-Film FTIR Spectroscopy to Characterise Milk Composition and Subclinical Ketosis throughout a Cow’s Lactation. Foods 2021, 10, 2033. [Google Scholar] [CrossRef] [PubMed]
  3. Brusač, E.; Jeličić, M.-L.; Cvetnić, M.; Amidžić Klarić, D.; Nigović, B.; Mornar, A. A Comprehensive Approach to Compatibility Testing Using Chromatographic, Thermal and Spectroscopic Techniques: Evaluation of Potential for a Monolayer Fixed-Dose Combination of 6-Mercaptopurine and Folic Acid. Pharmaceuticals 2021, 14, 274. [Google Scholar] [CrossRef] [PubMed]
  4. Santos, D.I.; Neiva Correia, M.J.; Mateus, M.M.; Saraiva, J.A.; Vicente, A.A.; Moldão, M. Fourier Transform Infrared (FT-IR) Spectroscopy as a Possible Rapid Tool to Evaluate Abiotic Stress Effects on Pineapple By-Products. Appl. Sci. 2019, 9, 4141. [Google Scholar] [CrossRef] [Green Version]
  5. Fadlelmoula, A.; Pinho, D.; Carvalho, V.H.; Catarino, S.O.; Minas, G. Fourier Transform Infrared (FTIR) Spectroscopy to Analyse Human Blood over the Last 20 Years: A Review towards Lab-on-a-Chip Devices. Micromachines 2022, 13, 187. [Google Scholar] [CrossRef]
  6. Wojtas, J.; Szala, M. Thermally Enhanced FTIR Spectroscopy Applied to Study of Explosives Stability. Measurement 2021, 184, 110000. [Google Scholar] [CrossRef]
  7. Treffers, R.R. Signal-to-Noise Ratio in Fourier Spectroscopy. Appl. Opt. 1977, 16, 3103–3106. [Google Scholar] [CrossRef]
  8. Xia, X.-G. A Quantitative Analysis of SNR in the Short-Time Fourier Transform Domain for Multicomponent Signals. IEEE Trans. Signal Process. 1998, 46, 200–203. [Google Scholar] [CrossRef]
  9. Yu, H.-G.; Lee, J.-H.; Park, D.-J.; Nam, H.-W.; Park, B.-H. I confirm Algorithms in the FTIR Spectrometer. In Proceedings of the Algorithms and Technologies for Multispectral, Hyperspectral, and Ultraspectral Imagery XXIV, Orlando, FL, USA, 17–19 April 2018; Volume 10644, pp. 491–496. [Google Scholar]
  10. Griffiths, P.R. Fourier Transform Infrared Spectrometry. Science 1983, 222, 297–302. [Google Scholar] [CrossRef] [Green Version]
  11. Hirschfeld, T. Quantitative FT-IR: A Detailed Look at the Problems Involved. Fourier Transform. Infrared Spectrosc. 1979, 2, 193–241. [Google Scholar]
  12. Saptari, V. Fourier Transform Spectroscopy Instrumentation Engineering; SPIE Optical Engineering Press: Bellingham, WA, USA, 2003. [Google Scholar]
  13. Decoster, D.; Harari, J. Optoelectronic Sensors; John Wiley & Sons: Hoboken, NJ, USA, 2013. [Google Scholar]
  14. Biclecki, Z.; Kolosowski, W.; Sedek, E.; Borejko, M. Analysis of Signal-to-Noise Ratio in Optical Receivers. In Proceedings of the 6th International Conference on Telecommunications in Modern Satellite, Cable and Broadcasting Service, Nis, Yugoslavia, 1–3 October 2003; IEEE: Manhattan, NY, USA, 2003; Volume 1, pp. 71–74, TELSIKS 2003. [Google Scholar]
  15. Achtenberg, K.; Mikołajczyk, J.; Ciofi, C.; Scandurra, G.; Michalczewski, K.; Bielecki, Z. Low-Frequency Noise Measurements of IR Photodetectors with Voltage Cross Correlation System. Measurement 2021, 183, 109867. [Google Scholar] [CrossRef]
  16. Scandurra, G.; Achtenberg, K.; Bielecki, Z.; Mikołajczyk, J.; Ciofi, C. On the Use of Supercapacitors for DC Blocking in Transformer-Coupled Voltage Amplifiers for Low-Frequency Noise Measurements. Electronics 2022, 11, 2011. [Google Scholar] [CrossRef]
  17. Achtenberg, K.; Mikołajczyk, J.; Bielecki, Z. Application of Cross-Correlation-Based Transimpedance Amplifier in InAs and InAsSb IR Detectors Noise Measurements. Opto-Electron. Rev. 2022, 30, e141126. [Google Scholar]
  18. Achtenberg, K.; Mikołajczyk, J.; Ciofi, C.; Scandurra, G.; Bielecki, Z. Transformer-Based Low Frequency Noise Measurement System for the Investigation of Infrared Detectors’ Noise. Measurement 2022, 190, 110657. [Google Scholar] [CrossRef]
  19. Kempfert, K.D.; Jiang, E.Y.; Oas, S.; Coffin, J. Detectors for Fourier Transform Spectroscopy. AN-00125; Thermo Nicolet Spectroscopy: Madison, WI, USA, 2001; Available online: https://mmrc.caltech.edu/FTIR/Nicolet/Nicolet%20Tech%20Notes/DetectorsforFTIR1204.pdf (accessed on 15 November 2022).
  20. Lindon, J.C.; Ferrige, A.G. Digitisation and Data Processing in Fourier Transform NMR. Prog. Nucl. Magn. Reson. Spectrosc. 1980, 14, 27–66. [Google Scholar] [CrossRef]
  21. Kang, Y. Analysis of Sampling Error in FTIR; Tan, H., Ed.; SPIE: Bellingham, WA, USA, 2013; pp. 122–127. [Google Scholar]
  22. Min, X.; Liu, R.; Hu, Y.; Fu, B.; Xu, K. Double-Beam near-Infrared Spectroscopy to Correct Light Source Drift in Aqueous Glucose Solution Experiments. Anal. Methods 2014, 6, 9831–9840. [Google Scholar] [CrossRef]
  23. Kuehl, D.; Griffiths, P.R. Dual-Beam Fourier Transform Infrared Spectrometer. Anal. Chem. 1978, 50, 418–422. [Google Scholar] [CrossRef]
  24. Anderson, V.E.; Fox, N.P.; Nettleton, D.H. Highly Stable, Monochromatic and Tunable Optical Radiation Source and Its Application to High Accuracy Spectrophotometry. Appl. Opt. 1992, 31, 536–545. [Google Scholar] [CrossRef]
  25. Jensen, P.S.; Bak, J.; Andersen, P.E.; Andersson-Engels, S. Fourier Transform Infrared Spectroscopy of Aqueous Solutions Using Optical Substraction. In Optical Diagnostics and Sensing of Biological Fluids and Glucose and Cholesterol Monitoring II; Priezzhev, A.V., Cote, G.L., Eds.; SPIE: Bellingham, WA, USA, 2002; pp. 150–159. [Google Scholar]
  26. SQUID, D.C. Noise Reduction by Parallel Cross-Correlation Measurements; Zurich Instruments AG: Zurich, Switzerland, 2020; Available online: https://www.zhinst.com/sites/default/files/zi_appnote_mfli_cross_correlation.pdf (accessed on 15 November 2022).
  27. Rubiola, E.; Vernotte, F. The Cross-Spectrum Experimental Method. arXiv 2010, arXiv:1003.0113. [Google Scholar]
  28. Sampietro, M.; Fasoli, L.; Ferrari, G. Spectrum Analyzer with Noise Reduction by Cross-Correlation Technique on Two Channels. Rev. Sci. Instrum. 1999, 70, 2520–2525. [Google Scholar] [CrossRef]
  29. Venables, W.N.; Ripley, B.D. Modern Applied Statistics with S-PLUS; Statistics and Computing; Springer: New York, NY, USA, 2002; ISBN 978-1-4419-3008-8. [Google Scholar]
  30. Giechaskiel, B.; Clairotte, M. Fourier Transform Infrared (FTIR) Spectroscopy for Measurements of Vehicle Exhaust Emissions: A Review. Appl. Sci. 2021, 11, 7416. [Google Scholar] [CrossRef]
  31. Lowry, S.; McCarthy, B.; Hyatt, J. Verification of Wavelength Accuracy in an FT-NIR Spectrometer. Sci. Appl. Note 2000, 54, 50772. [Google Scholar]
  32. Bielecki, Z. Some Problems with Optimization of Signal-to-Noise Ratio in Infrared Radiation Receivers. In Photonics Applications in Astronomy, Communications, Industry, and High-Energy Physics Experiments; Romaniuk, R.S., Pozniak, K.T., Eds.; SPIE: Bellingham, WA, USA, 2003; pp. 238–245. [Google Scholar]
  33. Bielecki, Z. Maximisation of Signal-to-Noise Ratio in Infrared Radiation Receivers. Opto-Electron. Rev. 2002, 10, 209–216. [Google Scholar]
  34. Rogalski, A. HgCdTe Infrared Detector Material: History, Status and Outlook. Rep. Prog. Phys. 2005, 68, 2267–2336. [Google Scholar] [CrossRef]
  35. Reine, M.B. Photovoltaic Detectors in MCT. In Infrared Detectors and Emitters: Materials and Devices; Springer: Boston, MA, USA, 2001; pp. 313–376. [Google Scholar]
  36. FTIR/FTNIR MODULE INTERSPEC 402-X. Available online: https://interspectrum.ee/ftir-module/ftirftnir-module-interspec-402-x/ (accessed on 15 November 2022).
  37. Semiconductor, D. Digitally-Controlled Phase Shift Using the DS1669. Available online: https://pdfserv.maximintegrated.com/en/an/AN184.pdf (accessed on 10 November 2022).
Figure 1. Noise influence analyses in transmission characteristics.
Figure 1. Noise influence analyses in transmission characteristics.
Sensors 22 08919 g001
Figure 2. Scheme (a) and photo (b) of an FTIR spectrometer with a designed sensor.
Figure 2. Scheme (a) and photo (b) of an FTIR spectrometer with a designed sensor.
Sensors 22 08919 g002
Figure 3. The relative responsivity of the applied HgCdTe photodiodes at 300 K.
Figure 3. The relative responsivity of the applied HgCdTe photodiodes at 300 K.
Sensors 22 08919 g003
Figure 4. Scheme of a single channel of the designed read-out electronics.
Figure 4. Scheme of a single channel of the designed read-out electronics.
Sensors 22 08919 g004
Figure 5. Spectral density of noise power of the designed channels.
Figure 5. Spectral density of noise power of the designed channels.
Sensors 22 08919 g005
Figure 6. The interferograms registered separately by CH1 channel and CH2 channel (a) and their PSD (b) without any absorbing sample placed in FTIR optical path.
Figure 6. The interferograms registered separately by CH1 channel and CH2 channel (a) and their PSD (b) without any absorbing sample placed in FTIR optical path.
Sensors 22 08919 g006
Figure 7. Cross-spectra of two interferograms with different phase shifts (a) and amplitudes (b).
Figure 7. Cross-spectra of two interferograms with different phase shifts (a) and amplitudes (b).
Sensors 22 08919 g007
Figure 8. Influence of averaging time on SNR values for different RBWs (a) and noise spread (standard deviation) (b).
Figure 8. Influence of averaging time on SNR values for different RBWs (a) and noise spread (standard deviation) (b).
Sensors 22 08919 g008
Figure 9. The spectra after cross-spectra averaging.
Figure 9. The spectra after cross-spectra averaging.
Sensors 22 08919 g009
Figure 10. Comparison of the foil spectral transmission registered with one- (NO CORR.) and two-(CROSS-CORR.) detection channels. (a) The foil absorbance determined with the standard pyroelectric detection module (PYR–FTIR) and designed sensor in a two-channel (CROSS-CORR.) and a single-channel (NO CORR.) configuration (b).
Figure 10. Comparison of the foil spectral transmission registered with one- (NO CORR.) and two-(CROSS-CORR.) detection channels. (a) The foil absorbance determined with the standard pyroelectric detection module (PYR–FTIR) and designed sensor in a two-channel (CROSS-CORR.) and a single-channel (NO CORR.) configuration (b).
Sensors 22 08919 g010
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Achtenberg, K.; Mikołajczyk, J.; Bielecki, Z. Two-Channel Detecting Sensor with Signal Cross-Correlation for FTIR Instruments. Sensors 2022, 22, 8919. https://doi.org/10.3390/s22228919

AMA Style

Achtenberg K, Mikołajczyk J, Bielecki Z. Two-Channel Detecting Sensor with Signal Cross-Correlation for FTIR Instruments. Sensors. 2022; 22(22):8919. https://doi.org/10.3390/s22228919

Chicago/Turabian Style

Achtenberg, Krzysztof, Janusz Mikołajczyk, and Zbigniew Bielecki. 2022. "Two-Channel Detecting Sensor with Signal Cross-Correlation for FTIR Instruments" Sensors 22, no. 22: 8919. https://doi.org/10.3390/s22228919

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop