Article

Performance Evaluation of Miniature Integrated Electrochemical Cells Fabricated Using LTCC Technology †

Faculty of Microsystem Electronics and Photonics, Wrocław University of Technology, Z. Janiszewskiego 11/17, 50-372 Wrocław, Poland
*
Author to whom correspondence should be addressed.
This paper is an extended version of the conference paper: Influence of the Geometry on the LTCC Integrated Electrochemical Cells Performance. In Proceedings of the EUROSENSORS 2018, Graz, Austria, 9–12 September 2018.
Sensors 2019, 19(6), 1314; https://doi.org/10.3390/s19061314
Submission received: 31 January 2019 / Revised: 11 March 2019 / Accepted: 12 March 2019 / Published: 15 March 2019
(This article belongs to the Special Issue Eurosensors 2018 Selected Papers)

Abstract

:
Miniaturized and integrated analytical devices, including chemical sensors, are at the forefront of modern analytical chemistry. The construction of novel analytical tools takes advantage of contemporary micro- and nanotechnologies, as well as materials science and technology. Two electrochemical techniques were used in experiments: electrochemical impedance spectroscopy and cyclic voltammetry. The goal of this study was to investigate electron transfer resistance in a model solution containing Fe 2 + / 3 + ions and protein adsorption using integrated electrochemical cells with different geometry. Tests performed at various Fe 2 + / 3 + concentration allowed to verify that these cells work properly. The influence of bovine serum albumin adsorbing to the surface of the integrated electrochemical cells was investigated. In electrochemical impedance spectroscopy, the value of R c t increased with protein adsorption and the relative change of R c t was in range 21% to 55%. In cyclic voltammetry the decreasing amperometric response of the working electrode was used as evidence of protein adsorption on the electrode.

1. Introduction

Over the years, many methods were invented or used in the investigation of protein layers immobilized on the surface. The most commonly used are the techniques based on immunological and fluorescent tests. Presently the most popular method is Enzyme-Linked Immunosorbent Assay (ELISA) [1]. ELISA is a versatile and sensitive technique that can be used for qualitative or quantitative determinations of antigen or antibody [2,3]. The addition of substrate giving colored, fluorescent or luminescent reaction products makes it possible to determine the concentrations of the reactants at low levels [1]. The second type of methods are marker—free techniques which include optical (ellipsometry [4]), microscopic (atomic force microscopy (AFM)) [5], transmission electron microscopy (TEM) [6]) and electrical (potentiometric, voltammetric [7], impedance [7]).
Electrochemical impedance spectroscopy (EIS) and cyclic voltammetry (CV) are electrical techniques in which the electrochemical cells are used. They consist of three individual electrodes (working electrode—WE, reference electrode—RE and counter electrode—CE) placed inside the vessel. In EIS the electrical response of the investigated system to the small amplitude periodic small amplitude alternating voltage (AC) signal is measured [8]. The analysis of the system response provides information about the interface and reactions occurring in it. This technique is used in the investigation of material properties [7,9] and measurement of biological and biochemical layers, e.g., bacteria and proteins [10]. In cyclic voltammetry the current is measured as the potential is changed linearly with time, a current-potential curve is recorded. The resulting curve is called a linear potential sweep voltammogram or peak polarogram [11]. CV is proven to be very effective for the study of the following areas: the redox process on gold electrode [12], characterization of biomolecule-modified electrode surfaces and the analysis of the alteration in the interfacial properties originating from biomolecular recognition events [13].
Integration of the electrochemical cell electrodes on the surface of the common substrate allows for the cell miniaturization and facilitates the increase of the number of simultaneously conducted experiments in multiplexed measurement system. Integrated electrochemical cells can be fabricated on various substrates such as silicon [14], Low Temperature Co-fired Ceramic (LTCC) [15] and Printed Circuit Board (PCB) [16].
Integrated electrochemical cells (IEC) used in our measurements were fabricated using LTCC technology. The devices made using LTCC technology can be characterized by their chemical inactivity, hermeticity, high reliability and high temperature stability. This technology allows fabrication of microfluidic systems such as flow sensors, micropumps, microvalves, micromixers, microreactors and polymerase chain reaction (PCR) devices [17,18,19].
Electrochemical cells fabricated using LTCC technology were used in many chemical and biological measurements such as devices for the detection of cortisol [20], heavy metal detection system in biomedical fluids [21] and enzyme gas sensor [22]. For our experiments a multi-channel measurement system was created. It allows the detection of biological structures on eight LTCC electrochemical cells at the same time.
In this work, the preliminary experiments with the integrated electrochemical cells fabricated in the LTCC technology were conducted. The results of the test measurements with model solution containing Fe 2 + / 3 + ions and the protein adsorption monitoring obtained using electrochemical impedance spectroscopy and cyclic voltammetry were shown.

2. Materials and Methods

IEC were fabricated using LTCC technology. Schematic drawing depicting the LTCC manufacturing process steps is shown in Figure 1. In the first step of the manufacturing process the green tape was blanked and registration holes were made. In the next step, the conductors and passive components were deposited by a screen printing method. Finished sheets were stacked together and laminated. After this process the structures were co-fired in a furnace. Before or after the co-firing process, the LTCC structures were singulated using dicing saw, ultrasonic cutting or laser cutting.
The IECs used in the experiments presented in this paper were made on rectangular substrate (20 × 3.5 mm) and were formed using three layers of green tape (DP 951, DuPont—Stevenage, United Kingdom). Nine types of IECs were fabricated with electrodes varying in size (Figure 2) which dimensions are shown in Table 1. Sensors 1–4 differed in both the WE and CE length and the distance between the electrodes while in sensors 5–9 only the length of WE and CE changed. The width of the RE was 0.5 mm in all cases. Both gold working (WE) and counter electrodes (CE) as well as silver reference electrode (RE) and contact pads were screen-printed with 325 mesh stainless steel screen (conductive Au paste ESL 8880-H and Ag paste ESL 903A were used). After the lamination and shaping of structures, they were co-fired in chamber furnace with standard firing profile at maximum temperature 875 °C. The surface of fabricated IECs were investigated using optical and Scanning Electron Microscope (SEM). An Ag/AgCl reference electrode was formed using two methods: electrolysis [23] and immersion of an RE in a highly concentrated NaCl solution [24]. In EIS and CV experiments, one sensor of each type was used.
Fabricated IECs are meant to be used in the protein adsorption measurements done in the presence of the Fe 2 + / 3 + ions. Those ions provided the occurence of the redox reaction on the electrodes surface [6]. Results presented in this work were obtained using model solution of potassium hexacyanoferrate (III) and potassium hexacyanoferrate (II) in either physiological saline or Phosphate Buffered Saline (PBS). Role of the latter was to maintain a constant pH. Providing constant measurement conditions is important because protein adsorption is a process with high sensitivity to the influence of the environment. The protein used in the experiment was Bovine Serum Albumin (BSA). BSA is a rigid, neutral, stable, moderately non-reactive protein [6,25]. It adsorbs to any type of surface and is highly stable [26].

2.1. Multichannel Electrochemical System EIS-STM32

The eight-channel potentiostat of our own design was used in the measurements. It was designed as the EIS extension of the IMP-STM32 impedance analyzer [27]. Its simplified schematic is shown in Figure 3. The impedance analyzer is responsible for generating AC component of the excitation voltage and for WE current measurement (both AC and DC). The fast low-frequency impedance measurement algorithms implemented in the device allow to shorten the duration of the EIS spectra acquisition [28]. The General Purpose Input–Outtput (GPIO) port of the impedance analyzer interfaces with the potentiostat controlling its functions: channel switching using the analogue multiplexer (MUX), CE DC bias voltage generation and the separation of the RE voltage AC component.
Our system was designed to be fully compatible with the integrated electrochemical cells. IECs dimensions and contact pads placements were designed in such way that they fit into typical micro-USB connectors. The measurement head connected to the multichannel potentiostat has eight such connectors and may be placed on top of the titrate plate with sensors positioned vertically in eight of 24 wells. Impedance analyzer IMP-STM32 was used in the measurements and the range of the frequency sweep was from 1 Hz to 100 kHz. The measurement system is shown in the Figure 4. This system was controlled by a software which provided a multi-channel EIS and CV measurement. As a result EIS spectra and CV voltammograms were obtained.

2.2. Electric Equivalent Circuit Modeling

The EIS spectra were analyzed using the Electrical Equivalent Circuit (EEC) method with the circuit shown in Figure 5. The R s is the electrolyte resistance, C P E d l is the electrical double layer capacitance, R c t is the electron transfer resistance. A constant phase element (CPE) is used in modeling the behaviour of the imperfect RLC elements. The admittance of CPE is given by:
Y = Q ( j ω ) n
where the Q (expressed in S s n ) and n (dimensionless) are the parameters and ω is a radial frequency [8]. The influence of other factors such as electrodes and connections resistance on the R s was negligible. The preliminary tests revealed that they are smaller than 3 Ω which was less than 1 % of R s . Scribner ZView 3.2c software (Southern Pines, NC, USA) was used to obtain the values of EEC parameters from fitting of real spectra.

3. Results

3.1. Scanning Electron Microscope

The SEM images of the IEC electrodes are shown in Figure 6. As the electrodes were fabricated using LTCC technology the surface roughness was clearly visible. As the result the actual area of the WE may be larger than it would result from its planar dimensions. The dark spots in the SEM images were most probably the glassy remains of the paste used to screen-print the electrodes left after the sintering.

3.2. Reference Electrode Forming

The preliminary tests of the Ag/AgCl reference electrode on LTCC ceramic formed using both highly concentrated NaCl solution [24] and electrolysis [23] methods were carried out. Two samples were prepared with each method and marked as NaCl-1, NaCl-2 and E-1, E-2, respectively. After their forming the tested electrodes were used as a RE in traditional electrochemical cells and compared with commercial Ag/AgCl RE. On the Bode plots (Figure 7) the impedance spectra obtained using tested electrodes were compared. Measurements were performed twice: directly after RE forming and after keeping them in the distilled water for 20 h. Results obtained using both methods were similar. Immediately after RE forming they were not satisfactory indicating the necessity of their stabilization after which their performance was similar to the commercial RE. Simpler concentrated NaCl solution method was used for RE forming in further experiments.
The Bode and Nyquist graphs (Figure 8—dots) show an example of the impedance spectra for complete IECs no 1 and 5 in a solution containing BSA. The spectra were approximated with electrical equivalent circuit (Figure 5). Results of the approximation are shown in Figure 8 as lines. The values of Chi-squared for this sensors were below 0.01.

3.3. Influence of the IEC Geometry on EIS Results

All types of the IECs were tested using EIS in the experiments which were carried out in the PBS solution with the presence of various concentrations of Fe 2 + / 3 + ions. Obtained R c t was inversely proportional to the c F e 2 + / 3 + (Figure 9). It can be concluded that integrated sensors worked correctly. The obtained value of R c t was not greater than 1 M Ω at c F e 2 + / 3 + = 1.17 mmol dm 3 , which is in the range which IMP-STM impedance analyzer is capable of precise measurement. It means that even the IEC with smallest electrodes may be easily used in further experiments conducted at similar c F e 2 + / 3 + . Additionally it was concluded that further IECs miniaturization is still possible.
The influence of the electrode geometry on the components of the Equivalent Electrical Circuit (electrolyte resistance, electron transfer resistance and electrical double layer capacitance) is shown in Figure 10. Values of the parameters were constant for sensors 5–9 because the distance between the electrodes has not changed. For other sensors the value of the parameters varied with electrodes distance. IECs cell constants were calculated as well. Cell constant changed in the same way as the EEC parameters, cell constant varied with electrodes distance (1–5 Table 1) and remained about 2.7 cm - 1 for other sensors.

3.4. EIS Monitoring of Protein Adsorption

Protein adsorption to the WE surface was investigated using EIS for five type of sensors (Table 1 no 1, 3, 5–7). The experiment was divided into two steps. In the first step the impedance sensor was placed in 2 mL PBS solution with the presence of Fe 2 + / 3 + ions ( c F e 2 + / 3 + = 5 mmol dm 3 ) and stabilized for 60 min. Afterwards 10% solution was replaced by bovine albumin dissolved in buffer in the proportions of 0.1 mg BSA in 5 mL PBS. The time between the consecutive impedance spectra measurements was less than 2 min. Protein adsorption modified a solution-electrode system causing the change of EIS spectra and EEC parameters. The value of R c t gradually increased with protein adsorption on WE surface. The process lasted 70 min for sensor no 1 and 20–30 min for other sensors. At the end, the system was stabilized—parameters of EEC had a constant value. The values of the R s , Q d l and n d l as a function of time are shown in Figure 11a–c.
It was observed that the obtained value of the electrical double layer capacitance Q d l did not correlate directly with the planar area of the electrodes. It was concluded that it was the effect of the WE surface roughness. As a result the attempt to normalize the R c t of each sensor to its planar dimensions would introduce significant error. Therefore to allow for easy comparison of the R c t between the sensors their R c t was scaled by the factor k calculated as:
k = Q d l Q d l 1 m m 2
where the Q d l is the value obtained for the sensor and Q d l 1 m m 2 is the value for the sensor with 1 mm 2 planar WE area. The normalized R c t as a function of time are shown in Figure 11d. The relative change of R c t after BSA adsorption is shown in Table 2.

3.5. CV Monitoring of Protein Adsorption

The CV experiments were conducted in two steps. In the first step the cyclic voltammograms for five type of sensors (Table 1 no 1, 3, 5–7) were measured for bare gold WE in PBS solution with the presence of Fe 2 + / 3 + ions ( c F e 2 + / 3 + = 5 mmol dm 3 ). In the second step the cyclic voltammograms were carried out after BSA adsorption on WE. The complete sweep time was 5 min and range of potential sweep was −0.6 V to 0.6 V.
Cyclic voltammetry is a simple and easy means of showing the changes of electrode behavior after each assemble step and CV experiments further confirmed that the BSA was successfully adsorbed on gold WE surface. The results of experiments of Fe 2 + / 3 + ions at a bare gold WE (curve a) and WE after BSA adsorption (curve b) are shown in the Figure 12. Characteristics of diffusion-limited redox processes were observed at the bare WE. The decreasing amperometric response of the WE was used as evidence of protein adsorption on the electrode.
The normalized cyclic voltammograms of Fe 2 + / 3 + ions at a bare gold WE and WE after BSA adsorption for all types of sensors are shown in the Figure 13. Measurement results were scaled by the factor k for the same reason as mentioned in Section 3.4. Decreasing the area of WE cause a reduction in the current value. Additional peaks at the sensors 6 and 7 may result from contamination or insufficient working electrode surface quality.

4. Conclusions

Miniature integrated electrochemical cells fabricated using LTCC technology were used in measurement with model solution containing Fe 2 + / 3 + ions and for investigation of protein adsorption on working electrode. Tests performed at various Fe 2 + / 3 + concentrations confirmed their proper operation. Electrodes geometry influenced cell constant and charge transfer resistance. Cell constant was in range of 2.5 to 5.1 cm - 1 . R c t determined for Fe 2 + / 3 + concentration of 1.17 mmol dm 3 was not greater than 1 M Ω . The influence of organic substances (BSA) adsorbing to the surface of the IECs was investigated by EIS and CV. The value of R c t increased with protein adsorption and the relative change of R c t was in the range of 21–55%. The largest relative R c t change after BSA adsorption (55%) was observed for the IEC with the smallest surface of working electrode (0.0625 mm 2 ).
The results of the presented preliminary work confirmed that the IECs were used in the measurement with the presence of Fe 2 + / 3 + ions and seem suitable for multichannel EIS systems for biological layer measurements. In the future, such sensors may also be integrated in complete microfluidic systems fabricated using LTCC technology.

Author Contributions

T.P. and P.S. conceived the ICE idea and design. D.N. fabricated the IECs using LTCC technology. P.S. performed and analyzed the experiments and wrote paper under supervision of T.P. D.N. contributed with LTCC technology description.

Funding

This work was funded by Wrocław University of Science and Technology statutory grant number 0401/0138/18.

Acknowledgments

This work was supported by Wrocław University of Science and Technology statutory grant.

Conflicts of Interest

The authors declare no conflict of interests.

Abbreviations

The following abbreviations are used in this manuscript:
AFMatomic force microscopy
BSABovine Serum Albumin
CEcounter electrode
CPEconstant phase element
CVcyclic voltammetry
EECElectrical Equivalent Circuit
EISelectrochemical impedance spectroscopy
ELISAEnzyme-Linked Immunosorbent Assay
IECintegrated electrochemical cell
LTCCLow Temperature Cofired Ceramic
PBSPhosphate Buffered Saline
PCRpolymerase chain reaction
REreference electrode
SEMscanning electron microscope
TEMtransmission electron microscopy
WEworking electrode

References

  1. Celis, J.E. Cell Biology; Academic Press: Cambridge, MA, USA, 2005. [Google Scholar]
  2. Amon, A.; Simchen, G.; Amorim, I.; Sunkel, C.E.; Reed, S.I.; Zachariae, W.; Ciosk, R.; Nasmyth, K.; Lottspeich, F.; Nasmyth, K.; et al. Common Structure of Soluble Amyloid Oligomers Implies Common Mechanism of Pathogenesis. Science 2003, 300, 486–490. [Google Scholar]
  3. Zhao, F.; Tian, Y.; Shen, Q.; Liu, R.; Shi, R.; Wang, H.; Yang, Z. A novel nanobody and mimotope based immunoassay for rapid analysis of aflatoxin B1. Talanta 2019, 195, 55–61. [Google Scholar] [CrossRef] [PubMed]
  4. Seitz, R.; Brings, R.; Geiger, R. Protein adsorption on solid–liquid interfaces monitored by laser-ellipsometry. Appl. Surf. Sci. 2005, 252, 154–157. [Google Scholar] [CrossRef]
  5. Drelich, J.; Mittal, K.L. Atomic Force Microscopy in Adhesion Studies; CRC Press: Boca Raton, FL, USA, 2005. [Google Scholar]
  6. Du, J.; Li, B.; Zhang, P.; Wang, Y. Cationized bovine serum albumin as gene carrier: Influence of specific secondary structure on DNA complexibility and gene transfection. Colloids Surf. B Biointerfaces 2016, 143, 37–46. [Google Scholar] [CrossRef] [PubMed]
  7. Zhang, S.; Wang, N.; Yu, H.; Niu, Y.; Sun, C. Covalent attachment of glucose oxidase to an Au electrode modified with gold nanoparticles for use as glucose biosensor. Bioelectrochemistry 2005, 67, 15–22. [Google Scholar] [CrossRef] [PubMed]
  8. Barsoukov, E.; Macdonald, J. Impedance Spectroscopy Theory, Experiment, and Applications; Wiley: Hoboken, NJ, USA, 2005. [Google Scholar]
  9. Liu, W.; Zhong, D.; Dai, Z.; Liu, Y.; Wang, J.; Wang, Z.; Pan, J. Synergetic utilization of photoabsorption and surface facet in crystalline/amorphous contacted BiOCl-Bi2S3 composite for photocatalytic degradation. J. Alloys Compd. 2019, 780, 907–916. [Google Scholar] [CrossRef]
  10. Sun, G.; Kang, K.; Qiu, L.; Guo, X.; Zhu, M. Electrochemical performance and microbial community analysis in air cathode microbial fuel cells fuelled with pyroligneous liquor. Bioelectrochemistry 2019, 126, 12–19. [Google Scholar] [CrossRef] [PubMed]
  11. Rieger, P.H. Electrochemistry; Springer Science + Business Media: Dordrecht, The Netherlands, 1994; Volume 91, pp. 399–404. [Google Scholar]
  12. Ahmad, A.; Moore, E. Electrochemical immunosensor modified with self-assembled monolayer of 11-mercaptoundecanoic acid on gold electrodes for detection of benzo[a]pyrene in water. Analyst 2012, 137, 5839–5844. [Google Scholar] [CrossRef] [PubMed]
  13. Haji-Hashemi, H.; Habibi, M.M.; Safarnejad, M.R.; Norouzi, P.; Ganjali, M.R. Label-free electrochemical immunosensor based on electrodeposited Prussian blue and gold nanoparticles for sensitive detection of citrus bacterial canker disease. Sens. Actuators B Chem. 2018, 275, 61–68. [Google Scholar] [CrossRef]
  14. Ziółkowski, R.; Górski, Ł.; Prokaryn, P.; Zaborowski, M.; Kutyła-Olesiuk, A.; Ciosek, P.; Wróblewski, W.; Malinowska, E. Development of silicon-based electrochemical transducers. Anal. Methods 2013, 5, 5464–5470. [Google Scholar]
  15. Manjakkal, L.; Synkiewicz, B.; Zaraska, K.; Cvejin, K.; Kulawik, J.; Szwagierczak, D. Development and characterization of miniaturized LTCC pH sensors with RuO2 based sensing electrodes. Sens. Actuators B Chem. 2016, 223, 641–649. [Google Scholar] [CrossRef]
  16. Evans, D.; Papadimitriou, K.I.; Vasilakis, N.; Pantelidis, P.; Kelleher, P.; Morgan, H.; Prodromakis, T. A Novel Microfluidic Point-of-Care Biosensor System on Printed Circuit Board for Cytokine Detection. Sensors 2018, 18, 4011. [Google Scholar] [CrossRef] [PubMed]
  17. Khanna, P.K.; Hornbostel, B.; Burgard, M.; Schäfer, W.; Dorner, J. Studies on three-dimensional moulding, bonding and assembling of low-temperature-cofired ceramics for MEMS and MST applications. Mater. Chem. Phys. 2005, 89, 72–79. [Google Scholar] [CrossRef]
  18. Malecha, K.; Golonka, L.J. Microchannel fabrication process using LTCC ceramics. Microelectron. Reliab. 2008, 48, 866–871. [Google Scholar] [CrossRef]
  19. Golonka, L.; Bembnowicz, P.; Jurków, D.; Malecha, K.; Roguszczak, H.; Tadaszak, R. Low temperature co-fired ceramics ( LTCC ) microsystems. Optica Appl. 2011, 41, 383–388. [Google Scholar]
  20. Bhansali, S.; Vasudev, A.; Norena, N.; Tomizawa, Y.; Kaushik, A. An LTCC-based microfluidic system for label-free, electrochemical detection of cortisol. Sens. Actuators B Chem. 2013, 182, 139–146. [Google Scholar]
  21. Angnes, L.; Richter, E.M.; Gongora-Rubio, M.R.; da Rocha, Z.M.; Fontes, M.B. LTCC manifold for heavy metal detection system in biomedical and environmental fluids. Sens. Actuators B Chem. 2004, 103, 468–473. [Google Scholar]
  22. Achmann, S.; Hämmerle, M.; Kita, J.; Moos, R. Miniaturized low temperature co-fired ceramics (LTCC) biosensor for amperometric gas sensing. Sens. Actuators B Chem. 2008, 135, 89–95. [Google Scholar] [CrossRef]
  23. Wang, T.; Gao, H.; Jin, X.; Chen, H.; Peng, J.; Chen, G.Z. Electrolysis of solid metal sulfide to metal and sulfur in molten NaCl–KCl. Electrochem. Commun. 2011, 13, 1492–1495. [Google Scholar] [CrossRef]
  24. Jin, M.; Jiang, L.; Xu, J.; Chu, H.; Tao, D.; Bai, S.; Jia, Y. Electrochemical Characterization of Solid Ag/AgCl Reference Electrode with Different Electrolytes for Corrosion Monitoring of Steel in Concrete. Electrochemistry 2016, 84, 383–389. [Google Scholar] [CrossRef] [Green Version]
  25. AL-Jawad, S.M.; Taha, A.A.; Al-Halbosiy, M.M.; AL-Barram, L.F. Synthesis and characterization of small-sized gold nanoparticles coated by bovine serum albumin (BSA) for cancer photothermal therapy. Photodiagn. Photodyn. Ther. 2018, 21, 201–210. [Google Scholar] [CrossRef] [PubMed]
  26. Ignat, T.; Miu, M.; Kleps, I.; Bragaru, A.; Simion, M.; Danila, M. Electrochemical characterization of BSA/11-mercaptoundecanoic acid on Au electrode. Mater. Sci. Eng. B Solid-State Mater. Adv. Technol. 2010, 169, 55–61. [Google Scholar] [CrossRef]
  27. Piasecki, T.; Chabowski, K.; Nitsch, K. Design, calibration and tests of versatile low frequency impedance analyser based on ARM microcontroller. Measurement 2016, 91, 155–161. [Google Scholar] [CrossRef]
  28. Piasecki, T. Fast impedance measurements at very low frequencies using curve fitting algorithms. Meas. Sci. Technol. 2015, 26. [Google Scholar] [CrossRef]
Figure 1. Schematic drawing of LTCC manufacturing process steps.
Figure 1. Schematic drawing of LTCC manufacturing process steps.
Sensors 19 01314 g001
Figure 2. (a) Design of integrated electrochemical cell fabricated using LTCC technology; (b) Photograph of all types of integrated electrochemical cell fabricated using LTCC technology; (c) Schematic drawing presenting the integrated electrochemical cell dimensions.
Figure 2. (a) Design of integrated electrochemical cell fabricated using LTCC technology; (b) Photograph of all types of integrated electrochemical cell fabricated using LTCC technology; (c) Schematic drawing presenting the integrated electrochemical cell dimensions.
Sensors 19 01314 g002
Figure 3. Schematic of eight-channel, compact, digitally controlled potentiostat dedicated for impedance analyzer IMP-STM32.
Figure 3. Schematic of eight-channel, compact, digitally controlled potentiostat dedicated for impedance analyzer IMP-STM32.
Sensors 19 01314 g003
Figure 4. Measurement system: (a) eight-channel measurement head with sensors placed vertically in 24-well titrate plate and (b) eight-channel potentiostat.
Figure 4. Measurement system: (a) eight-channel measurement head with sensors placed vertically in 24-well titrate plate and (b) eight-channel potentiostat.
Sensors 19 01314 g004
Figure 5. Electrical Equivalent Circuit used in experiment.
Figure 5. Electrical Equivalent Circuit used in experiment.
Sensors 19 01314 g005
Figure 6. Electrodes surface of the integrated electrochemical cell on LTCC from SEM: (a) working electrode—gold and (b) reference electrode—silver.
Figure 6. Electrodes surface of the integrated electrochemical cell on LTCC from SEM: (a) working electrode—gold and (b) reference electrode—silver.
Sensors 19 01314 g006
Figure 7. Bode plots for samples used as RE after (a) modification in NaCl solution and (b) electrolysis, where RE—commercial Ag/AgCl RE, NaCl-1 and NaCl-2—sample of Ag/AgCl RE formed in highly concentrated NaCl, E-1 and E-2—sample of Ag/AgCl RE formed using electrolysis.
Figure 7. Bode plots for samples used as RE after (a) modification in NaCl solution and (b) electrolysis, where RE—commercial Ag/AgCl RE, NaCl-1 and NaCl-2—sample of Ag/AgCl RE formed in highly concentrated NaCl, E-1 and E-2—sample of Ag/AgCl RE formed using electrolysis.
Sensors 19 01314 g007
Figure 8. Impedance spectra and its approximation for sensors no 1 and 5 in PBS solution with the presence of Fe 2 + / 3 + ions and BSA: (a) Z = f ( Z ) , (b) | Z | = f ( f ) and Φ = f ( f ) .
Figure 8. Impedance spectra and its approximation for sensors no 1 and 5 in PBS solution with the presence of Fe 2 + / 3 + ions and BSA: (a) Z = f ( Z ) , (b) | Z | = f ( f ) and Φ = f ( f ) .
Sensors 19 01314 g008
Figure 9. Charge transfer resistance as a function of the concentration of Fe 2 + / 3 + ions for integrated electrochemical cell.
Figure 9. Charge transfer resistance as a function of the concentration of Fe 2 + / 3 + ions for integrated electrochemical cell.
Sensors 19 01314 g009
Figure 10. (a) Electrolyte resistance, (b) charge exchange resistance and (c) double layer capacitance as a function of the surface of the working electrode.
Figure 10. (a) Electrolyte resistance, (b) charge exchange resistance and (c) double layer capacitance as a function of the surface of the working electrode.
Sensors 19 01314 g010
Figure 11. Parameters of electrical equivalent circuit as a function of time: (a) Q d l , (b) R s , (c) n d l and (d) normalized R c t .
Figure 11. Parameters of electrical equivalent circuit as a function of time: (a) Q d l , (b) R s , (c) n d l and (d) normalized R c t .
Sensors 19 01314 g011
Figure 12. Cyclic voltammograms recorded in a 0.5 mM Fe 2 + / 3 + + PBS solution after different step of modification, (a) bare Au WE, (b) BSA adsorption on WE electrode for sensors no (a) 5 ( S W E , C E = 1 mm 2 ), (b) 3 ( S W E , C E = 0.5 mm 2 ), (c) 1 ( S W E , C E = 0.0625 mm 2 ), (d) 6 ( S W E , C E = 1.25 mm 2 ) and (e) 7 ( S W E , C E = 1.5 mm 2 ).
Figure 12. Cyclic voltammograms recorded in a 0.5 mM Fe 2 + / 3 + + PBS solution after different step of modification, (a) bare Au WE, (b) BSA adsorption on WE electrode for sensors no (a) 5 ( S W E , C E = 1 mm 2 ), (b) 3 ( S W E , C E = 0.5 mm 2 ), (c) 1 ( S W E , C E = 0.0625 mm 2 ), (d) 6 ( S W E , C E = 1.25 mm 2 ) and (e) 7 ( S W E , C E = 1.5 mm 2 ).
Sensors 19 01314 g012aSensors 19 01314 g012b
Figure 13. Normalized cyclic voltammograms recorded in a 0.5 mM Fe 2 + / 3 + + PBS solution after different step of modification for every type of sensor, (a) bare Au WE, (b) BSA adsorption on WE electrode.
Figure 13. Normalized cyclic voltammograms recorded in a 0.5 mM Fe 2 + / 3 + + PBS solution after different step of modification for every type of sensor, (a) bare Au WE, (b) BSA adsorption on WE electrode.
Sensors 19 01314 g013
Table 1. Dimensions of the integrated electrochemical sensors and calculated values of R c t for c F e 2 + / 3 +  = 1.17 mmol dm 3 and cell constant, where h—height of WE and CE, w—width of WE and CE, d—distance between WE and CE, A W E , C E —area of WE and CE, A R E —area of RE, R c t —charge transfer resistance, κ —cell constant.
Table 1. Dimensions of the integrated electrochemical sensors and calculated values of R c t for c F e 2 + / 3 +  = 1.17 mmol dm 3 and cell constant, where h—height of WE and CE, w—width of WE and CE, d—distance between WE and CE, A W E , C E —area of WE and CE, A R E —area of RE, R c t —charge transfer resistance, κ —cell constant.
Noh (mm)w (mm)d (mm) A WE , CE ( mm 2 ) A RE ( mm 2 ) R ct ( k Ω ) κ ( cm - 1 )
10.1250.520.06250.125594.25.1
20.50.51.50.250.5702.54.8
310.51.50.50.5442.13.9
40.75110.750.5291.73.1
511110.5346.73.5
61.25111.250.5321.42.7
71.5111.50.5308.52.7
81.75111.750.5197.52.6
921120.5110.82.5
Table 2. Relative change of R c t for every type of sensor.
Table 2. Relative change of R c t for every type of sensor.
No S WE (mm 2 )Relative Change of R c t
10.062555%
30.521%
5138%
61.2526%
71.521%

Share and Cite

MDPI and ACS Style

Szymanowska, P.; Nowak, D.; Piasecki, T. Performance Evaluation of Miniature Integrated Electrochemical Cells Fabricated Using LTCC Technology. Sensors 2019, 19, 1314. https://doi.org/10.3390/s19061314

AMA Style