Abstract

Sample Preparation and qPCR Detection of Tuberculosis on a Centrifugal Microfluidic Cartridge Enabling Molecular Downstream Resistance Profiling †

1
Hahn-Schickard, 79110 Freiburg, Germany
2
WHO Supranational Tuberculosis Reference Laboratory, IML Red, 82131 Gauting, Germany
3
Laboratory for MEMS Applications, IMTEK University of Freiburg, 79110 Freiburg, Germany
4
SYNLAB Gauting, SYNLAB MVZ Dachau GmbH, 82131 Gauting, Germany
*
Author to whom correspondence should be addressed.
Presented at the XXXV EUROSENSORS Conference, Lecce, Italy, 10–13 September 2023.
Proceedings 2024, 97(1), 208; https://doi.org/10.3390/proceedings2024097208
Published: 26 April 2024

Abstract

:
Tuberculosis (TB) is still one of the world’s deadliest infections. Fast detection of the pathogen M. tuberculosis (MTB) and its genetic resistance markers substantially improves treatment success and outcome. A key element for rapid genetic diagnostics is the efficient extraction of DNA from sputum for qPCR detection at the point of care. We present the fully automated sample preparation of MTB DNA from 3 mL of liquefied sputum and qPCR detection of MTB on a centrifugal microfluidic cartridge. Our method achieves a limit of detection (LoD) between 17 and 57 CFU/mL and provides a purified DNA solution for molecular downstream testing, such as targeted NGS.

1. Introduction

Only a few fully automated point-of-care solutions are available for TB diagnostics [1,2], and none of them provide extracted DNA for subsequent comprehensive antibiotic resistance testing from the same sample. Here, we present a new sample preparation of MTB DNA using centrifugal microfluidics, a technology dedicated to implementing complex workflows on compact cartridges [3]. However, to date, fully automated centrifugal microfluidic sample-to-answer approaches have only been demonstrated for small sample volumes [4] or nasal swabs [5]. The combination of a liquefaction and inactivation (LI) reagent with a filter and lysis (FL) module allows us to increase sample volumes 15-fold from 200 µL to 3000 µL. It enables the detection of MTB on the cartridge and the provision of the DNA solution in a detachable DNA tube for subsequent resistance profiling.

2. Materials and Methods

The cavity for the FL module is integrated into an injection-molded cartridge (Figure 1A,B). A polyethylene frit (PE3510, SPC Technologies, diameter of 9 mm) and a glass microfiber filter (GMF150, Whatman, diameter of 10 mm) are fixed with a press-fit O-ring (8 × 1.25 mm, FKM, IR-Dichtungstechnik). The custom-developed LI reagent consists of 45% isopropanol, 3% NaOH, 0.5% N-lauroyl-sarcosine, and 0.2% Tween. 1 mL of sputum is mixed with 2 mL of LI reagent and incubated for 20 min.

3. Results and Discussion

For the automation of sample preparation and qPCR detection of MTB from sputum on a centrifugal microfluidic cartridge, the FL module is the key element. It enables the efficient provision of MTB DNA from sputum and consists of a filter frit and a bacteria enrichment filter (Figure 1A). The externally inactivated, liquefied sputum sample is added to the cartridge and filtered, retaining MTB in the bacteria enrichment filter (Figure 2A). A washing step is performed to reduce interfering sputum and LI reagent residues. This is followed by thermal lysis to release the MTB DNA. With this method, large sample volumes of up to 3 mL can be automatically processed and analyzed. Spiking dilutions of H37Rv in sputum identified a LoD between 17 and 57 CFU/mL (Figure 2B). One important achievement is the inactivation and structural weakening of the hard-to-lyse MTB during the LI step while not losing the DNA. MTB DNA can be efficiently released by thermal lysis and subsequently used for qPCR-based MTB detection on the cartridge. Resistance profiles can be generated externally downstream using the DNA solution obtained via the DNA tube interface, as depicted in Figure 1C.

Author Contributions

Conceptualization, J.S.; methodology, J.S.; formal analysis, J.S.; investigation, J.S. and M.B.; writing—original draft, J.S.; visualization, J.S.; project administration, J.S. and M.B.; writing—review and editing, M.B., J.L., H.H. and N.P.; supervision, J.L., H.H. and N.P.; funding acquisition, H.H. and N.P. All authors have read and agreed to the published version of the manuscript.

Funding

Financial support by the Federal Ministry of Education and Research, Germany, within the project BT-SeqDisK (No. 16GWO154) is gratefully acknowledged.

Institutional Review Board Statement

The study was approved by the Nepal Health Research Council and the Ethical committee of the medical faculty of the University of Munic (project ID 17-761).

Informed Consent Statement

Informed consent was obtained from all subjects involved in the study.

Data Availability Statement

Data are partially contained within the original research article Schlanderer et al. (DOI: 10.1039/d3lc00783a).

Conflicts of Interest

We declare that the three unit operation of the integrated workflow is partially protected by patents. The rehydration of the pre-stored reagents is protected by DE102013220257 and family. The switching after the rehydration is patented by DE102016207845 and family. The aliquoting is protected by DE102008003979 and family. All patents are owned by Hahn-Schickard. On the first two, Nils Paust is a co-inventor.

References

  1. Helb, D.; Jones, M.; Story, E.; Boehme, C.; Wallace, E.; Ho, K.; Kop, J.; Owens, M.R.; Rodgers, R.; Banada, P.; et al. Rapid Detection of Mycobacterium tuberculosis and Rifampin Resistance by Use of On-Demand, Near-Patient Technology. J. Clin. Microbiol. 2010, 48, 229–237. [Google Scholar] [CrossRef] [PubMed]
  2. Lu, H.-W.; Sakamuri, R.; Kumar, P.; Ferguson, T.M.; Doebler, R.W.; Herrington, K.D.; Talbot, R.P.; Weigel, K.M.; Nguyen, F.K.; Cangelosi, G.A.; et al. Integrated nucleic acid testing system to enable TB diagnosis in peripheral settings. Lab Chip 2020, 20, 4071–4081. [Google Scholar] [CrossRef] [PubMed]
  3. Strohmeier, O.; Keller, M.; Schwemmer, F.; Zehnle, S.; Mark, D.; von Stetten, F.; Zengerle, R.; Paust, N. Centrifugal microfluidic platforms: Advanced unit operations and applications. Chem. Soc. Rev. 2015, 44, 6187–6229. [Google Scholar] [CrossRef] [PubMed]
  4. Czilwik, G.; Messinger, T.; Strohmeier, O.; Wadle, S.; Von Stetten, F.; Paust, N.; Roth, G.; Zengerle, R.; Saarinen, P.; Niittymäki, J.; et al. Rapid and fully automated bacterial pathogen detection on a centrifugal-microfluidic LabDisk using highly sensitive nested PCR with integrated sample preparation. Lab Chip 2015, 15, 3749–3759. [Google Scholar] [CrossRef] [PubMed]
  5. Spindiag. Rhonda SARS-CoV-2 Disk—Spindiag. 2023. Available online: https://www.spindiag.de/rhonda-sars-cov-2-disk/ (accessed on 18 March 2023).
Figure 1. (A) Exploded view of the FL module. (B) Microfluidic cartridge. (C) Top-side view of the DNA tube.
Figure 1. (A) Exploded view of the FL module. (B) Microfluidic cartridge. (C) Top-side view of the DNA tube.
Proceedings 97 00208 g001
Figure 2. (A) Sample preparation steps automated by the cartridge. (B) LoD achieved with the automated sample preparation and qPCR detection of the microfluidic cartridge with H37Rv spiked in sputum.
Figure 2. (A) Sample preparation steps automated by the cartridge. (B) LoD achieved with the automated sample preparation and qPCR detection of the microfluidic cartridge with H37Rv spiked in sputum.
Proceedings 97 00208 g002
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Schlanderer, J.; Beutler, M.; Lüddecke, J.; Hoffmann, H.; Paust, N. Sample Preparation and qPCR Detection of Tuberculosis on a Centrifugal Microfluidic Cartridge Enabling Molecular Downstream Resistance Profiling. Proceedings 2024, 97, 208. https://doi.org/10.3390/proceedings2024097208

AMA Style

Schlanderer J, Beutler M, Lüddecke J, Hoffmann H, Paust N. Sample Preparation and qPCR Detection of Tuberculosis on a Centrifugal Microfluidic Cartridge Enabling Molecular Downstream Resistance Profiling. Proceedings. 2024; 97(1):208. https://doi.org/10.3390/proceedings2024097208

Chicago/Turabian Style

Schlanderer, Judith, Markus Beutler, Jan Lüddecke, Harald Hoffmann, and Nils Paust. 2024. "Sample Preparation and qPCR Detection of Tuberculosis on a Centrifugal Microfluidic Cartridge Enabling Molecular Downstream Resistance Profiling" Proceedings 97, no. 1: 208. https://doi.org/10.3390/proceedings2024097208

Article Metrics

Back to TopTop