Fetal Arrhythmia Detection Based on Labeling Considering Heartbeat Interval

1
Graduate School of Science and Technology, Keio University, Yokohama 223-8522, Kanagawa, Japan
2
Department of Information and Computer Science, Keio University, Yokohama 223-8522, Kanagawa, Japan
*
Author to whom correspondence should be addressed.
Bioengineering 2023, 10(1), 48; https://doi.org/10.3390/bioengineering10010048
Submission received: 25 November 2022 / Revised: 20 December 2022 / Accepted: 26 December 2022 / Published: 30 December 2022
(This article belongs to the Special Issue Monitoring and Analysis of Human Biosignals)

Abstract

:
Arrhythmia is one of the causes of sudden infant death, and it is very important to detect fetal arrhythmia for fetal well-being. Fetal electrocardiogram (FECG) is one of the methods to detect a heartbeat. Fetal arrhythmia can be detected based on the heartbeat detection results from FECG signals such as heartbeat intervals. However, the accuracy of arrhythmia detection easily degrades depending on the accuracy of heartbeat detection. In this paper, we propose a deep learning-based fetal arrhythmia detection method using FECG signals. Recently, arrhythmia detection methods using adult ECG signals have achieved a high arrhythmia detection accuracy based on deep learning. Motivated by this fact, in the proposed method, the acquired FECG signals are segmented, and the segments are input into a deep learning model that classifies them into normal or arrhythmia ones. Based on the classification results of multiple segments, a subject is judged as a healthy or arrhythmia subject. Each segment of the training data is divided into three categories based on the estimated heartbeat interval: (i) normal, (ii) arrhythmia, and (iii) a segment that could be both normal and arrhythmic. Only segments labeled as normal or arrhythmia are used for training a deep learning model to achieve a higher classification accuracy of the model. Through these procedures, the proposed method detects fetal arrhythmia with fewer effects of heartbeat detection results. The experimental results show that the proposed method achieves 96.2% accuracy, 100% specificity, and 100% recall, improving the values of conventional methods based on heartbeat detection and feature detection.

Graphical Abstract

1. Introduction

Fetal heart rate (FHR) is an important indicator for accessing fetal health status. The risk of unexplained, unexpected fetal death is particularly high during the prenatal and perinatal periods. Fetal arrhythmia is one of the causes of sudden infant death [1,2]. Fetal arrhythmia is an abnormality of the fetal heartbeat, and is generally diagnosed as tachycardia if it is greater than 180 beats per minute and bradycardia if it is less than 100 beats per minute [3]. Fetal arrhythmias can be treated with antiarrhythmic drugs if diagnosed promptly [4]. Therefore, there has been great interest in the efficient diagnosis of fetal arrhythmia, and various fetal arrhythmia detection methods have been investigated. Ultrasonography, echocardiography, electrocardiography, and magnetocardiography are methods to collect fetal biological signals for this purpose [5]. Among these methods, fetal electrocardiogram (FECG) acquisition can be divided into invasive and non-invasive types. The invasive type involves a large physical burden on the mother and fetus [6]. In contrast, a non-invasive fetal electrocardiogram (FECG) can be used to obtain information on the electrical activity of the fetal heart from electrodes attached to the maternal abdomen [7]. Thanks to the low price of the used device, FECG has been widely used to detect the fetal heartbeat. The most conventional fetal arrhythmia detection methods are based on fetal heartbeat detection [8,9,10]. Specifically, in the conventional methods [8,10], the heartbeat is firstly detected, and then features are extracted from FECG signals based on the heartbeat detection results. Using the features, a subject is classified into healthy or arrhythmia. In addition, the conventional method [9] estimates FHR, extracts features from the variability of the FHR and classifies a subject into healthy or arrhythmia. The arrhythmia detection accuracy of these conventional methods depends on the accuracy of heartbeat detection. However, there is a problem that the fetal ECG signal is easily distorted by external disturbances such as maternal heartbeats and fetal body movements, and the accuracy of heartbeat detection is easily degraded [11]. In other words, as the accuracy of heartbeat detection deteriorates, the accuracy of fetal arrhythmia detection also deteriorates. Hori et al. have proposed a method for arrhythmia detection based on deep learning using adult ECG signals on data from 48 adult arrhythmia patients in the MIT-BIH arrhythmia database [12,13]. Since ECG patterns vary from subject to subject, developing an arrhythmia detection method that can accommodate different waveform patterns will enable the detection of abnormalities that could not be detected by the conventional methods. In this method, a versatile ECG abnormality determination method is proposed using an autoencoder and a convolutional neural network (CNN). Specifically, the autoencoder learns only normal waveforms that are easy to collect and acquires features of normal waveforms. The CNN is then used to compare the features obtained from the ECG waveforms targeted for arrhythmia detection with the features of the normal waveforms of healthy subjects. The results are used to determine whether the input ECG waveform is a normal waveform or an abnormal one containing arrhythmia based on a threshold value. This makes it possible to take into account the differences in waveforms for each subject. Gyohten et al. have also proposed a method to detect any arrhythmia by building a normal ECG model using deep learning in 42 adult arrhythmia patients in the MIT-BIH arrhythmia database [12,14]. This method uses CNN and long short-term memory (LSTM) to build a model of the normal ECG signal. This model takes a normal ECG signal as input and learns to predict subsequent normal ECG signals. When an abnormal ECG signal is input, the model can predict the subsequent ECG signal, which is far from the actual ECG signal, and can determine whether the input signal is abnormal or not. This means that the method can determine any arrhythmia because it does not require prior knowledge of annotation. Next, we present other literature that applies deep learning models to FECG signals. In [15], a multichannel signal quality classifier for FECG waveforms is presented. Each recording was labeled as a whole and assigned a quality class label of good or bad. Preference was given to recordings with consistent perceived signal quality to ensure that the entire recording was labeled correctly. Recordings with inconsistent signal conditions were discarded. Labels were assigned by the study’s data engineers based on FECG visibility and perceived SNR of the abdominal channel (hospital clinicians were consulted for the decision process). A deep learning model was developed in the literature [16] to automatically identify maternal heart rate (MHR) and more generally false signals (FS) in FHR recordings. The model can be used to preprocess FHR data prior to automated analysis or as a clinical alert system to assist practitioners. The training data set included data annotated by experts on the observed records. These two references also use deep learning models to classify FECG signals. It is important to note, however, that these two references are concerned with signal accuracy. Arrhythmia is not only a disturbance of the signal waveform, but also includes abnormalities in the heartbeat interval. Therefore, arrhythmias may not be correctly classified as arrhythmias using these methods [15,16].
In this paper, we propose a fetal arrhythmia detection based on deep learning using FECG signals. A fetal arrhythmia detection method based on deep learning has been proposed in our paper [17]. In the proposed method, the acquired FECG signals are first segmented to include several heartbeats. The FECG segment is fed into a deep learning model that classifies each segment as normal or arrhythmia. The classification results of the model are used to determine whether each subject is a normal or arrhythmia subject. Each segment of training data is labeled as a normal segment, an arrhythmia segment, or a moderate segment that may be associated with both normal and arrhythmia ones, based on the estimated heartbeat intervals. To improve the classification accuracy of the model, only normal and arrhythmia segments are used for training; moderate segments are not used. This deep learning-based arrhythmia detection is expected to prevent arrhythmia detection accuracy from being degraded by errors in heartbeat detection, since the ECG signal waveform itself is segmented as input data for the model. To evaluate the performance of our proposed method, we carried out the performance evaluation based on the dataset containing 12 arrhythmia subjects and 14 normal subjects [18]. As a result, the proposed method achieved at most 96.2% of accuracy, 100% of recall, and 100% of specificity in the binary classification of healthy and arrhythmia subjects. In addition, the experimental results showed that our method could outperform the other existing methods [8,9,10] in arrhythmia detection accuracy. Here, it is worth mentioning that there are two major differences between this paper and [17]. First, this paper describes the parameter setting and segment labeling methods in the proposed method and clarifies the rationale for these methods through more detailed evaluation and discussion. Specifically, we explain the impact of different parameter settings on the proposed method and provide the rationale for the parameter settings. We also give the results of characterization using additional metrics for different parameter settings and emphasize the justification for the values we set. Furthermore, a more detailed discussion of the experimental results will describe specific issues with the method and possible improvements. Second, this paper contains a new comparison and discussion with the characterization results of a new method that takes into account the problems of the proposed method. Specifically, the segments are labeled taking into account the differences in the mean and standard deviation of the heartbeat intervals for each subject. The subjects are then binarized into healthy subjects and arrhythmia subjects based on the results of the three-value classification of the segments.
The rest of this paper is organized as follows: In Section 2, we describe related research on fetal arrhythmia detection. In Section 3, we explain the proposed method. In Section 4, we evaluate the performance of our method. Finally, we conclude this paper in Section 5.

2. Related Work

In this section, we describe previous research related to fetal arrhythmia detection via FECG. FECG can be derived from AECG [19,20]. To detect fetal heartbeat based on abdominal ECG (AECG), the conventional method [19] detects the maternal QRS wave from the AECG by using the Pan–Tompkins algorithm [21], creates a template of the maternal ECG (MECG) based on the detected QRS wave, and subtracts the template from the AECG to extract the FECG. FHR is estimated by detecting peaks due to fetal heartbeat over the extracted FECG. However, the extracted FECG contains not only fetal heartbeat components but also noise that could cause wrong peak detection. To improve the accuracy of FHR extraction, Niida et al. have proposed a fetal heart rate estimation method using the first derivative of the FECG signal and multiple weighting functions [20]. In this method, the first derivative of the FECG signal is calculated to acquire R-peak candidates. To emphasize the actual R-peaks, the amplitudes of the peaks are weighted based on the assumption that the RR intervals follow a Gaussian distribution.
With these FHR estimation methods, some researchers have proposed fetal arrhythmia detection [8,9,10]. All of these conventional methods use PhysioNet’s public database [18]. The conventional method [8] firstly detects the characteristic points of the FECG signal, namely Q, R, S, and T points, and calculates the average values of RR, SS, QQ, ST, and TT intervals. Furthermore, the QRS interval reflects the electrical activity of the heart during ventricular contraction and is the most important segment of the ECG signal [22]. Therefore, the average amplitude and width of the QRS interval for each ECG are calculated and used as features as well. In addition to that, the nonlinear operator of energy tracking Teager Energy Operator (TEO), which includes the nonlinear behavior of the RR interval, is also extracted as a feature [23]. Using these features, a support vector machine (SVM) classifier is trained to classify a subject as healthy or arrhythmia. The Leave One Out (LOO) cross-validation method is used to evaluate the learning model. The experimental results have shown that this method achieves a specificity of 91.7%, a recall of 75.0%, and an accuracy of 83.3%. In the conventional method [9], independent component analysis (ICA) and singular value decomposition (SVD) are used to estimate the FHR by removing the maternal ECG components, and the FHR is segmented. Next, the entropy-based features, which represent the similarity between adjacent segments, are calculated from the FHR variation. A threshold is set for the entropy-based features, and a subject is classified as healthy or arrhythmia. A specificity and a recall of this method are 100% and 91.0%, respectively. The conventional method [10] also uses ICA to remove the maternal ECG component to obtain the FECG. Features are then extracted using a peak detection algorithm to identify the extracted FECG signal. Specifically, the peak detection algorithm using the state machine logic is used to extract features such as the RR interval, ST interval, the width of the QRS wave, and the amplitudes of QRS and T waves from the FECG. Binary classification of the subjects with healthy and arrhythmia is performed using a naive Bayes classifier. The naive Bayes classifier has the advantage of requiring fewer inputs of feature values. Through the experiments, this method has been shown to achieve a specificity of 96.3%, a recall of 74.8%, and an accuracy of 93.7%. However, the fetal heartbeat components can be easily distorted by maternal heartbeat components and fetal body motion, which tends to degrade the accuracy of fetal heartbeat detection. Nevertheless, these conventional fetal arrhythmia methods [8,9,10] are based on fetal heartbeat detection from the FECG, and the accuracy of arrhythmia detection depends on the heartbeat detection accuracy. Therefore, it is necessary to develop a fetal arrhythmia detection method that is less dependent on fetal heartbeat detection.

3. Proposed Method

In this section, we explain the proposed fetal arrhythmia detection using deep learning.

3.1. Framework of Proposed Arrhythmia Detection

First, FECG is extracted from AECG, and the RRI (RR-Interval) is estimated using the fetal heartbeat detection method [20]. In this method, since the output of the A/D converter is contaminated by fetal body motion, the preprocessing is performed to remove saturated data and invalid data from the raw AECG. Next, the maternal R-peaks are extracted from the preprocessed AECG. After the locations of maternal R-peaks are detected, the maternal cycles are removed from the AECG, and FECG is extracted. In addition, to detect the fetal heart rate, the candidates of the R-peak are generated. By designing a weighting function, we can detect the fetal R-peaks from the candidates. Figure 1 and Figure 2 show examples of the AECG and the extracted FECG, respectively. The acquired FECG signal is then down-sampled to 500 Hz. The FECG signal is then segmented by a 3 s-time window that contains several heartbeats. In addition, the step size of the window is set as 1 s.
The segments of FECG are classified as normal or arrhythmia segments based on deep learning. Here, note that it is still challenging to detect arrhythmia by using only RRIs, since the RRIs have some estimation errors. Therefore, for more accurate arrhythmia detection, we utilize the deep learning technique that performs the classification using the FECG waveform itself as input data. As a deep learning model, we use CNN which has been successfully applied for the classification task [24,25]. Figure 3 shows the structure of the CNN model used in the proposed method. The convolution layer with the activation function ReLU highlights the features of the input FECG signal, and then the pooling layer reduces the feature dimension. The combination of the convolution and pooling layers is repeated twice, which is followed by the affine layer. Finally, the input FECG segment is classified into normal or arrhythmia ones.
After the classification, a subject is judged based on the multiple results of the classification. Specifically, a threshold t h is set for the ratio of the segments classified as arrhythmia by the CNN. A subject is judged as an arrhythmia one, when the ratio is greater than t h ; otherwise, a subject is judged as a healthy one.

3.2. Training Dataset

Arrhythmia is mainly determined by abnormal heartbeat intervals in ECG signals. However, since arrhythmia does not always occur, normal waveforms are considered to be included in the ECG signal data of arrhythmia subjects. Therefore, to train the CNN, the training data are labeled based on the estimated RRI. Figure 4 shows the distributions of the estimated RRIs for all the arrhythmia and normal subjects in the database used in this study. As can be seen from Figure 4, the distribution of the estimated RRIs of the healthy subjects is clustered around 400 ms, whereas those of the arrhythmia subjects are relatively widely distributed. The standard deviations of the respective estimated RRIs are 33.94 ms for the healthy subjects and 71.73 ms for the arrhythmia subjects. In the proposed method, the estimated RRIs are classified into three categories, i.e., (i) normal, (ii) moderate, and (iii) arrhythmic RRIs, by a 25 ms-RRI range as shown in Figure 4. When the RRI range has more RRIs of healthy subjects than those of arrhythmia subjects, the RRIs within the RRI ranges are defined as normal RRIs. When the ratio of the number of healthy subjects’ RRI to that of the arrhythmia one exceeds x, the RRIs within the RRI ranges are defined as moderate RRIs. Here, it is worth mentioning that moderate RRIs might be associated with both healthy and arrhythmia subjects. When such a ratio is lower than or equal to x, the RRIs within the RRI ranges are defined as arrhythmic RRIs.
The FECG segments are then labeled based on the three types of RRIs. Specifically, each segment is labeled as a normal segment, when the segment contains only normal RRIs. In addition, each segment is labeled as a moderate segment, when the segment contains at least one moderate RRI without arrhythmic RRIs. Each segment is labeled as an arrhythmia segment, when the segment contains at least one arrhythmic RRIs. Figure 5 shows examples of the three types of the FECG segments. To improve the classification accuracy of the CNN, moderate segments that could be associated with both normal or arrhythmic RRIs are not used to train the CNN model, meaning that only normal and arrhythmia segments are used for the training. The arrhythmia segment generated from a normal subject is then treated as a moderate segment.

4. Performance Evaluation

Evaluation Setup

In this study, we used the AECG database containing fetal arrhythmias provided on the PhysioNet website [18]. Table 1 lists the experimental specification. Each of the subjects’ AECG recordings consists of four or five channels, and one maternal thoracic channel. The database contains about 10 min of AECG signals from 12 arrhythmia and 14 healthy subjects. The sampling frequency of the AECG signal is 500 Hz or 1 kHz. The data set is also labeled as healthy or arrhythmic for each subject. The loss function for the network is binary cross-entropy, and the optimizer is Adam with l r = 0.0003 . The model was implemented using Keras v1.1.2 on TensorFlow v2.4.0.
To demonstrate the accuracy of the proposed method in detecting fetal arrhythmia, we evaluated the binary classification accuracy of healthy and arrhythmia subjects to the proposed and conventional methods [8,9,10]. In the conventional method [8], the subject is classified using SVM based on features extracted from the FECG. In the conventional method [9], the subject is classified by using threshold values based on the entropy features extracted from fetal heart rate variability. In the conventional method [10], the features are extracted from FECG based on the peak detection algorithm using the state machine logic, and the subject is classified by using a naive Bayes classifier. As the performance metrics, we calculated a specificity, a recall, and an accuracy using the following equations:
s p e c i f i c i t y = T N T N + F P ,
r e c a l l = T P T P + F N ,
a c c u r a c y = T P + T N T P + F P + T N + F N ,
where T N represents the number of healthy subjects that are correctly classified as healthy subjects. F N represents that of arrhythmia subjects that are incorrectly classified as healthy subjects. F P represents that of healthy subjects that are incorrectly classified as arrhythmia subjects. T P represents that of arrhythmia subjects that are correctly classified as arrhythmia subjects. As the experimental parameters, the threshold x for RRI classification was set as 0.63 based on the results of the search, and the threshold t h for the ratio of arrhythmia segments was varied from 0.00 to 1.00 in increments of 0.05.

5. Results

Table 2 shows the ratio of the number of RRIs of healthy subjects to that of RRIs of arrhythmic subjects in Figure 4, and the classification results of each RRI range when each value of the threshold x for RRI classification is applied to it. Each column of Table 2 is described below. “ R R I r a n g e ” indicates the value of RRI representing each range in Figure 4. “ R R I r a t i o ” indicates the ratio of the number of healthy subjects’ RRI to that of arrhythmia one in each range in Figure 4. “The classification result” indicates the classification result for each range when the threshold x for the R R I r a t i o is varied. “A” represents arrhythmic RRI, “M” represents moderate RRI, and “N” represents normal RRI. When the R R I r a t i o is lower than or equal to x, the RRIs within the RRI ranges are defined as arrhythmic RRIs. When the R R I r a t i o exceeds x, the RRIs within the RRI ranges are defined as moderate RRIs. When the RRI range has more RRIs of healthy subjects than those of arrhythmia subjects, the RRIs within the RRI ranges are defined as normal RRIs. The comparison is made for x = 0.50 , 0.60 , 0.63 , 0.65 , 0.70 . The values of x are set so that the three classification groups of arrhythmic RRI, moderate RRI, and normal RRI, which are assigned to each range of the histogram shown in Figure 4, are different from each other. From Table 2, it can be seen that the number of RRI ranges classified as moderate RRIs and arrhythmic RRIs changes when the threshold x is varied. The more RRIs classified as Moderate RRIs, the more FECG segments are labeled as moderate segments and the fewer segments are used to train the deep learning model. Figure 6, Figure 7 and Figure 8 respectively show the recall, the specificity, and the accuracy of the subject classification when segments are labeled by varying the threshold x used for RRI classification. A subject is judged as an arrhythmia one, when the ratio is greater than t h ; otherwise, a subject is judged as a healthy one.
Figure 6 shows that, for x = 0.63 , 0.65 , 0.70 , t h 0.75 gives the highest recall compared to other threshold x. The lower the threshold x used for RRI classification, the more RRIs are classified as moderate RRIs. Correspondingly, the more segments are tentatively labeled as moderate segments and the number of arrhythmia segments is reduced. Based on this, the number of segments used to train and test the deep learning model is reduced, and segments needed for subject classification may also be excluded. Therefore, the accuracy of arrhythmia subject classification is considered to have deteriorated. Figure 7 shows that the highest specificity is obtained at x = 0.63 for almost all the thresholds t h . When the threshold x used for RRI classification is high, the number of RRI data classified as moderate RRI decreases and that of RRI data classified as arrhythmic RRI increases. Based on this, the number of moderate segments would decrease and segments that could be both normal and arrhythmia could be used as arrhythmia segments. Therefore, the classification accuracy of the normal segments would be degraded, resulting in a degradation of the classification results for normal subjects. From Figure 8, it can be seen that, at x = 0.60 , 0.63 , 0.65 , 0.70 , 0.6 t h 0.7 , the highest accuracy is obtained compared to other threshold values of x. It can also be seen that, for x = 0.63 , high accuracy is obtained in the case of t h 0.35 compared to the other x values. This may be due to the same reason as for recall and specificity. Based on these results, x = 0.63 was considered optimal for the proposed method for the data set used in this study.
Figure 9, Figure 10 and Figure 11 respectively show the recall, the specificity, and the accuracy of the proposed method for the threshold x = 0.63 used for RRI classification and conventional methods [8,9,10]. In addition, for a better comparison, these figures include the result of our method with the simple labeling algorithm: the FECG segments of healthy and arrhythmia subjects are labeled as normal and arrhythmia ones, respectively. First, we compare the recall of the proposed method with those of the other methods.
From Figure 9, it can be seen that the proposed method outperforms the conventional ones [8,9] and provides almost the same recall as the conventional ones [10] in the range where 0.40 t h 0.70 . In these conventional methods, the arrhythmia subject is detected based on heart rate estimation and feature detection of the FECG signal. In contrast, our method classifies healthy and arrhythmia subjects based on the FECG signal waveforms using deep learning. Therefore, our method reduces the effect of errors induced by the heart rate estimation and the feature detection. In addition, our method achieves high recall, compared to the conventional one [10] in the range where t h 0.40 . This is because the arrhythmia is determined when the ratio of the segments classified as arrhythmia exceeds the threshold t h , and thus the smaller t h makes it easier to judge that the subject has arrhythmia. Furthermore, our proposed method achieves a higher recall than our method with the simple labeling algorithm, which is brought by labeling the training data based on the RRI distribution.
In terms of specificity, as can be seen from Figure 10, the proposed method improves the specificity of the conventional ones [8,9], and achieves 100% specificity in the range where t h 0.6 . As aforementioned, these results are brought by the arrhythmia detection that does not depend on heart rate estimation and the feature detection of the FECG signal. In addition, the larger t h makes it easier to determine that the subject is healthy, since the subject is judged as normal when the ratio of the segments classified as arrhythmia is lower than the threshold t h . In addition, as well as the result of the recall, we can see that our proposed method based on the RRI distribution outperforms our method with the simple labeling algorithm that labels all segments of arrhythmia subjects as arrhythmia segments and all segments of healthy subjects as normal segments. Therefore, it can be said that labeling segments based on the RRI resulted in higher classification accuracy for both normal and arrhythmia subjects. This may be because the effect of the estimation error of the RRI was reduced.
Finally, from Figure 11, it can be seen that the proposed method improves the accuracy of the conventional methods [8,9] by 16.7% and 2.44%, respectively, in the range where 0.6 t h 0.7 . Furthermore, our method achieves a higher accuracy than our method with the simple labeling algorithm for almost all the threshold values. In our method with the simple labeling algorithm, all the segments of arrhythmia subjects are labeled as arrhythmia one. However, arrhythmia does not necessarily happen all the time, meaning that some segments of arrhythmia subjects are not arrhythmia ones but normal ones. In contrast, our proposed method performs segment labeling based on the RRI distribution, and does not use the segments that might be the normal and arrhythmia ones. Therefore, our proposed method provides a high classification accuracy of the CNN, compared to our method with the simple labeling algorithm. Note that the proposed method requires threshold setting.
We focus on the classification results for the segments of healthy subjects. Table 3 shows the number of arrhythmia segments and normal segments when the segments of three healthy subjects in the database are labeled based on the estimated RRI and classified using CNN, and the ratio of segments classified as arrhythmia out of all segments. From Table 3, it can be seen that even segments from healthy subjects can be misclassified as arrhythmia segments. Therefore, if the threshold t h used for subject classification for this percentage of arrhythmia segments is too low, the accuracy and the specificity deteriorate.
Based on this fact, in our future work, it is necessary to make our method robust to the threshold setting. In particular, since the lower t h makes it easier to detect arrhythmia subjects, we will improve the accuracy for lower t h .
Next, we consider how to take into account the differences of the mean and standard deviation of the RRI among subjects in order to improve the accuracy of a subject classification and increase versatility. Figure 12 shows a scatter plot of the mean and standard deviation of the RRI for each subject. From Figure 12, it seems that the mean and standard deviation of RRI vary widely among normal subjects. In the proposed method, RRIs were classified and segments were labeled for all normal subjects and all arrhythmia subjects collectively based on the histogram shown in Figure 4. In this method, the differences in the mean and standard deviation of RRIs for each subject could not be taken into account. Therefore, we will compare the proposed method with the method that takes into account differences in the mean and standard deviation of RRIs per subject in the following.

5.1. Method Based on 3-Value Classification Considering Differences in Mean and Standard Deviation of RRI

Each arrhythmia subject has a different mean and standard deviation of RRIs. The proposed method of labeling based on the RRI distributions of arrhythmia and healthy subjects compared above does not take this into account. Therefore, we describe a method that takes into account the differences in the mean and standard deviation of the RRI for each subject.

5.1.1. Segment Labeling

Based on the mean and standard deviation of the RRI for each subject, the RRI is classified according to the following definition for arrhythmia subjects:
m o d e r a t e _ m i n = M e a n _ R R I y × R R I _ s t d m o d e r a t e _ m a x = M e a n _ R R I + y × R R I _ s t d
RRIs between m o d e r a t e _ m i n and m o d e r a t e _ m a x are defined as moderate RRIs; all others are defined as arrhythmic RRIs. M e a n _ R R I is the mean value of RRI for each subject, and R R I _ s t d is the standard deviation of RRI for each subject. The percentage of data classified as moderate RRI among the RRI data changes depending on the value of y. Specifically, the larger y is, the greater the percentage of RRI data are classified as moderate RRI, and the smaller y is, the smaller the percentage becomes. The value of y is set so that the percentage of data classified as medium RRI is 50, 60, 70, 80, or 90%. After classifying the RRI data as moderate RRI or arrhythmic RRI for each arrhythmia subject based on Equation (4), we label the segments using the same definition as in the proposed method. Specifically, each segment is labeled as a moderate segment, when the segment contains at least one moderate RRI without arrhythmic RRIs. Each segment is labeled as an arrhythmia segment, when the segment contains at least one arrhythmic RRIs. Label all segments of normal subjects as normal segments.

5.1.2. Classification of the Subject

Based on the results of the three-value classification of the segments, the subjects are binary classified as either arrhythmia subjects or normal subjects. The segments of arrhythmia subjects are labeled considering the mean and standard deviation of the RRI for each subject. All segments of normal subjects are labeled as normal segments. These segments are classified into three values, arrhythmia segment, medium segment, and normal segment, using a deep learning model as in the proposed method. The subject is then classified based on the ratio of arrhythmia segments to the total number of normal and arrhythmia segments among the classification results. This ratio is calculated based on the following equation:
A R R _ N R _ r a t i o = A R R _ s e g m e n t s A R R _ s e g m e n t s + N R _ s e g m e n t s
A R R _ s e g m e n t s represents the number of segments classified as arrhythmia, and N R _ s e g m e n t s represents the number of segments classified as normal. A R R _ N R _ r a t i o represents the ratio of arrhythmia segments to the total number of normal and arrhythmia segments. For this A R R _ N R _ r a t i o , a threshold t h is set as in the proposed method: if A R R _ N R _ r a t i o is greater than t h , the subject is determined to be arrhythmic; otherwise, the subject is determined to be normal.

5.1.3. Comparison of the Proposed Method with the Method Based on Three-Value Classification

Figure 13, Figure 14 and Figure 15 show the recall, specificity, and accuracy calculated by the method based on the three-value classification and the proposed method, respectively. However, in the method based on the three-value classification, the percentage of RRI data classified as moderate RRI is varied to 50, 60, 70, 80, and 90%. Figure 13 shows that the proposed method achieves the highest recall at almost all threshold t h . Comparing the methods based on three-value classification, the highest recall is achieved when the percentage of classification to moderate RRI is 50%. However, looking at the overall value trends, it is not necessarily the case that the recall improves with a lower percentage of classification to moderate RRI. Next, Figure 14 shows that the method based on three-value classification achieves greater specificity than the proposed method for 0.2 t h 0.55 , when 90% of the methods are classified as moderate RRI. Similarly, when the ratio is 80%, the method based on three-value classification achieves greater specificity than the proposed method when 0.35 t h 0.55 . Consider this: the greater the percentage of data classified as moderate RRI, the fewer the number of data classified as arrhythmic RRI. In other words, fewer segments are labeled as arrhythmia segments. This is expected to reduce the number of segments learned as arrhythmia segments in the deep learning model, and therefore the number of segments classified as arrhythmia segments will also decrease. Therefore, it is believed that a subject is more likely to be determined to be a normal subject when the percentage of segments classified as moderate RRI is increased. Finally, Figure 15 shows that the proposed method achieves the highest accuracy at almost all thresholds. It also seems that the highest accuracy is achieved for 0.35 t h 0.4 when 50% of the methods are classified as moderate RRI among the methods based on three-value classification. However, looking at the overall trend of values, as in the discussion for recall, it can be considered that specificity does not necessarily improve with a lower percentage of classification to moderate RRI. This result suggests that there is no significant relationship between the percentage of subjects classified as moderate RRI and the accuracy of their classification.
Based on the above results, we believe that the proposed method is the most capable of correctly classifying subjects.

6. Conclusions

In this paper, we proposed the fetal arrhythmia detection method based on deep learning using FECG signals. The proposed method uses the CNN that classifies FECG signal segments as normal or arrhythmia ones. Since the classification is based on the FECG signal itself, it could be possible to reduce the effects of errors in FHR estimation and feature detection. In addition, by removing the FECG signal segment that might be associated with both the normal and arrhythmia ones, our proposed method improves the classification accuracy of the CNN. The experimental results show that the proposed method achieves 96.2% accuracy, 100% specificity, and 100% recall, improving the values of conventional methods based on heartbeat detection and feature detection. In our future work, we will consider improving the sensitivity to the threshold used in our method for more practical arrhythmia detection.

Author Contributions

Conceptualization, S.N., K.Y. and T.O.; methodology, S.N.; formal analysis, S.N.; investigation, S.N. and K.Y.; writing—original draft preparation, S.N.; writing—review and editing, K.Y. and T.O.; visualization, S.N.; supervision, T.O.; project administration, T.O. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The data used in this study are open database provided on the PhisioNet website. Available: https://physionet.org/content/nifeadb/1.0.0/ (accessed on 29 December 2022).

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Krapp, M.; Kohl, T.; Simpson, J.M.; Sharland, G.K.; Katalinic, A.; Gembruch, U. Review of diagnosis treatment and outcome of fetal atrial flutter compared with supraventricular tachycardia. Heart 2003, 89, 913–917. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  2. Schade, R.P.; Stoutenbeek, P.; de Vries, L.S.; Meijboom, E.J. Neurological morbidity after fetal supraventricular tachyarrhythmia. Ultrasound Obstet. Gynecol. 1999, 13, 43–47. [Google Scholar] [CrossRef] [PubMed]
  3. Strasburger, J.F. Fetal arrhythmias. Prog. Pediatr. Cardiol. 2000, 11, 1–17. [Google Scholar] [CrossRef] [PubMed]
  4. Wacker-Gussmann, A.; Strasburger, J.F.; Cuneo, B.F.; Wakai, R.T. Diagnosis and Treatment of Fetal Arrhythmia. Am. J. Perinatol. 2014, 31, 617–628. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  5. Wacker-Gussmann, A.; Wakai, R.T.; Strasburger, J.F. Importance of fetal arrhythmias to the neonatologist and pediatrician. Neoreviews 2016, 17, e568–e578. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  6. Peters, C.; Vullings, R.; Bergmans, J.; Oei, G.; Wijn, P. Heart Rate Detection in Low Amplitude Non-Invasive Fetal ECG Recordings. In Proceedings of the International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), New York, NY, USA, 30 August–3 September 2006; pp. 6092–6094. [Google Scholar]
  7. Behar, J.; Zhu, T.; Oster, J.; Niksch, A.; Mah, D.Y.; Chun, T. Evaluation of the fetal QT interval using non-invasive fetal ECG technology. Physiol. Meas. 2016, 37, 1392. [Google Scholar] [CrossRef] [PubMed]
  8. Pavel, M.S.R.; Islam, M.R.; Siddiqee, A.M. Fetal Arrhythmia Detection Using Fetal ECG Signal. In Proceedings of the 2019 IEEE International Conference on Telecommunications and Photonics (ICTP), Dhaka, Bangladesh, 28–30 December 2019; pp. 1–4. [Google Scholar]
  9. Keenan, E.; Udhayakumar, R.K.; Karmakar, C.K.; Brownfoot, F.C.; Palaniswami, M. Entropy Profiling for Detection of Fetal Arrhythmias in Short Length Fetal Heart Rate Recordings. In Proceedings of the 2020 42nd Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC), Montreal, QC, Canada, 20–24 July 2020; pp. 621–624. [Google Scholar]
  10. Apsana, S.; Suresh, M.G.; Aneesh, R.P. A novel algorithm for early detection of fetal arrhythmia using ICA. In Proceedings of the 2017 International Conference on Intelligent Computing, Instrumentation and Control Technologies (ICICICT), Kannur, Kerala, India, 6–7 July 2017; pp. 1277–1283. [Google Scholar]
  11. Zarzoso, V.; Nandi, A.K. Noninvasive fetal electrocardiogram extraction: Blind separation versus adaptive noise cancellation. IEEE Trans. Biomed. Eng. 2001, 48, 12–18. [Google Scholar] [CrossRef] [PubMed]
  12. MIT-BIH Arrhythmia Database. Available online: https://physionet.org/content/mitdb/1.0.0/ (accessed on 13 December 2022).
  13. Hori, S.; Shouo, T.; Gyohten, K.; Ohki, H.; Takami, T.; Sato, N. Arrhythmia Detection Based on Patient-Specific Normal ECGs Using Deep Learning. In Proceedings of the 2020 Computing in Cardiology, Rimini, Italy, 13–16 September 2020; pp. 1–4. [Google Scholar]
  14. Gyohten, K.; Hori, S.; Ohki, H.; Takami, T.; Sato, N. Building Normal ECG Models to Detect Any Arrhythmias Using Deep Learning. In Proceedings of the 2020 Computing in Cardiology, Rimini, Italy, 13–16 September 2020; pp. 1–4. [Google Scholar]
  15. Mertes, G.; Long, Y.; Liu, Z.; Li, Y.; Yang, Y.; Clifton, D.A. A Deep Learning Approach for the Assessment of Signal Quality of Non-Invasive Foetal Electrocardiography. Sensors 2022, 22, 3303. [Google Scholar] [CrossRef] [PubMed]
  16. Boudet, S.; Houzé de l’Aulnoit, A.; Peyrodie, L.; Demailly, R.; Houzé de l’Aulnoit, D. Use of Deep Learning to Detect the Maternal Heart Rate and False Signals on Fetal Heart Rate Recordings. Biosensors 2022, 12, 691. [Google Scholar] [CrossRef] [PubMed]
  17. Sara, N.; Kohei, Y.; Tomoaki, O. Fetal Arrhythmia Dtection based Deep Learning using Fetal ECG Signals. In Proceedings of the IEEE GLOBECOM, Rio de Janeiro, Brazil, 4–8 December 2022. manuscript in preparation. [Google Scholar]
  18. Behar, J. Non-Invasive Fetal ECG Arrhythmia Database. PhysioNet. Available online: https://physionet.org/content/nifeadb/1.0.0/ (accessed on 13 December 2022).
  19. Karim Khondakar, M.F.; Sarowar, M.H.; Hossain, Q.D. Extraction of Fetal ECG in a Non-invasive Way and Estimation of Fetal Heart Rate. In Proceedings of the 2020 IEEE Region 10 Symposium (TENSYMP), Dhaka, Bangladesh, 5–7 June 2020; pp. 358–361. [Google Scholar]
  20. Natsuho, N.; Lu, W.; Tomoaki, O.; Kazunari, O.; Naoki, H.; Hayato, H. Fetal Heart Rate Detection Using First Derivative of ECG Waveform and Multiple Weighting Functions. In Proceedings of the 2021 43rd Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC), Virtual, 1–5 November 2021; pp. 434–438. [Google Scholar]
  21. Pan, J.; Tompkins, W.J. A real-time QRS detection algorithm. IEEE Trans. Biomed. Eng. 1985, 32, 230–236. [Google Scholar] [CrossRef] [PubMed]
  22. Isin, A.; Ozdalili, S. Cardiac arrhythmia detection using deep learning. Procedia Comput. Sci. 2017, 120, 268–275. [Google Scholar] [CrossRef]
  23. Jabloun, F.; Cetin, A.E.; Erzin, E. Teager energy based feature parameters for speech recognition in car noise. IEEE Signal Process. Lett. 1999, 6, 259–261. [Google Scholar] [CrossRef]
  24. Sarvan, Ç.; Özkurt, N. ECG Beat Arrhythmia Classification by using 1D CNN in case of Class Imbalance. In Proceedings of the 2019 Medical Technologies Congress (TIPTEKNO), Izmir, Turkey, 3–5 October 2019; pp. 1–4. [Google Scholar]
  25. Xiaolin, L.; Panicker, R.C.; Cardiff, B.; John, D. Multistage Pruning of CNN Based ECG Classifiers for Edge Devices. In Proceedings of the 2021 43rd Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC), Virtual, 1–5 November 2021; pp. 1965–1968. [Google Scholar]
Figure 1. The part of AECG waveform when signal accuracy is relatively good.
Figure 1. The part of AECG waveform when signal accuracy is relatively good.
Bioengineering 10 00048 g001
Figure 2. The part of the extracted FECG from AECG when the signal accuracy is relatively good.
Figure 2. The part of the extracted FECG from AECG when the signal accuracy is relatively good.
Bioengineering 10 00048 g002
Figure 3. The structure of CNN model in the proposed method including layers, each parameter value, and input/output data.
Figure 3. The structure of CNN model in the proposed method including layers, each parameter value, and input/output data.
Bioengineering 10 00048 g003
Figure 4. The distributions of the estimated RRIs for all the arrhythmia and normal subjects in the database used in this paper.
Figure 4. The distributions of the estimated RRIs for all the arrhythmia and normal subjects in the database used in this paper.
Bioengineering 10 00048 g004
Figure 5. The labeled segments. (a) normal FECG segment; (b) moderate FECG segment; and (c) arrhythmia FECG segment.
Figure 5. The labeled segments. (a) normal FECG segment; (b) moderate FECG segment; and (c) arrhythmia FECG segment.
Bioengineering 10 00048 g005
Figure 6. The recall of the proposed method when the threshold x used for RRI classification is varied.
Figure 6. The recall of the proposed method when the threshold x used for RRI classification is varied.
Bioengineering 10 00048 g006
Figure 7. The specificity of the proposed method when the threshold x used for RRI classification is varied.
Figure 7. The specificity of the proposed method when the threshold x used for RRI classification is varied.
Bioengineering 10 00048 g007
Figure 8. The accuracy of the proposed method when the threshold x used for RRI classification is varied.
Figure 8. The accuracy of the proposed method when the threshold x used for RRI classification is varied.
Bioengineering 10 00048 g008
Figure 9. The recall of the conventional methods [8,9,10], the proposed method with the simple labeling algorithm, and the proposed method.
Figure 9. The recall of the conventional methods [8,9,10], the proposed method with the simple labeling algorithm, and the proposed method.
Bioengineering 10 00048 g009
Figure 10. The specificity of the conventional methods [8,9,10], the proposed method with the simple labeling algorithm, and the proposed method.
Figure 10. The specificity of the conventional methods [8,9,10], the proposed method with the simple labeling algorithm, and the proposed method.
Bioengineering 10 00048 g010
Figure 11. The accuracy of the conventional methods [8,9], the proposed method with the simple labeling algorithm, and the proposed method.
Figure 11. The accuracy of the conventional methods [8,9], the proposed method with the simple labeling algorithm, and the proposed method.
Bioengineering 10 00048 g011
Figure 12. The scatter plot of the mean and standard deviation of the RRI for each of the subjects.
Figure 12. The scatter plot of the mean and standard deviation of the RRI for each of the subjects.
Bioengineering 10 00048 g012
Figure 13. The recall of the method based on three-value classification (the percentage classified as moderate RRI = 50, 60, 70, 80, 90%) and the proposed method.
Figure 13. The recall of the method based on three-value classification (the percentage classified as moderate RRI = 50, 60, 70, 80, 90%) and the proposed method.
Bioengineering 10 00048 g013
Figure 14. The specificity of the method based on three-value classification (the percentage classified as moderate RRI = 50, 60, 70, 80, 90%) and the proposed method.
Figure 14. The specificity of the method based on three-value classification (the percentage classified as moderate RRI = 50, 60, 70, 80, 90%) and the proposed method.
Bioengineering 10 00048 g014
Figure 15. The accuracy of the method based on three-value classification (the percentage classified as moderate RRI = 50, 60, 70, 80, 90%) and the proposed method.
Figure 15. The accuracy of the method based on three-value classification (the percentage classified as moderate RRI = 50, 60, 70, 80, 90%) and the proposed method.
Bioengineering 10 00048 g015
Table 1. Experimental specification.
Table 1. Experimental specification.
ItemValue
The number of AECG4 or 5
channels for each subject
Sampling frequency of AECG recording500 Hz or 1000 Hz
Measurement timeAbout 10 min
The number of subjectsHealthy subjects: 14
Arrhythmia subjects: 12
Threshold x used to categorize RRI ranges.0.63
Threshold t h 0.00, 0.05, 0.10,..., 1.00
Table 2. The classification results of each RRI range.
Table 2. The classification results of each RRI range.
RRI Range [ms] RRI Ratio *The Classification Result
x = 0 . 50 x = 0 . 60 x = 0 . 63 x = 0 . 65 x = 0 . 70
100–1250.2000AAAAA
125–1500.4000AAAAA
150–1750.3846AAAAA
175–2000.0870AAAAA
200–2250.2000AAAAA
225–2500.2113AAAAA
250–2750.2541AAAAA
275–3000.1312AAAAA
300–3250.0427AAAAA
325–3500.1233AAAAA
350–3750.6093MMAAA
375–4004.2561NNNNN
400–4253.3120NNNNN
425–4501.1848NNNNN
450–4750.6470MMMAA
475–5000.2834AAAAA
500–5250.1530AAAAA
525–5500.1506AAAAA
550–5750.1387AAAAA
575–6000.2712AAAAA
600–6250.1389AAAAA
625–6500.3571AAAAA
650–6750.2105AAAAA
675–7000.2353AAAAA
700–7250.1111AAAAA
725–7500.6667MMMMA
750–7750.0000AAAAA
775–8000.0000AAAAA
800–8250.0000AAAAA
825–8500.0000AAAAA
850–8750.5000MAAAA
*: The ratio of the number of healthy subjects’ RRI to the number of arrhythmia one.
Table 3. The results of labeling and classification of segments of healthy subjects.
Table 3. The results of labeling and classification of segments of healthy subjects.
LabelingClassification by CNNThe Ratio of Arrhythmia Segments
ArrhythmiaNormalArrhythmiaNormal
subject105212003210.384
subject208132490.395
subject3013642940.309
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Nakatani, S.; Yamamoto, K.; Ohtsuki, T. Fetal Arrhythmia Detection Based on Labeling Considering Heartbeat Interval. Bioengineering 2023, 10, 48. https://doi.org/10.3390/bioengineering10010048

AMA Style

Nakatani S, Yamamoto K, Ohtsuki T. Fetal Arrhythmia Detection Based on Labeling Considering Heartbeat Interval. Bioengineering. 2023; 10(1):48. https://doi.org/10.3390/bioengineering10010048

Chicago/Turabian Style

Nakatani, Sara, Kohei Yamamoto, and Tomoaki Ohtsuki. 2023. "Fetal Arrhythmia Detection Based on Labeling Considering Heartbeat Interval" Bioengineering 10, no. 1: 48. https://doi.org/10.3390/bioengineering10010048