Scientia Pharmaceutica is published by MDPI from Volume 84 Issue 3 (2016). Previous articles were published by another publisher in Open Access under a CC-BY (or CC-BY-NC-ND) licence, and they are hosted by MDPI on mdpi.com as a courtesy and upon agreement with Austrian Pharmaceutical Society (Österreichische Pharmazeutische Gesellschaft, ÖPhG).
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Flow Injection Analysis With Trimetazidine Dihydrochloride Ion-Selective Electrode

by
Yousry M. Issa
*,
Nour T. Abdel-Ghani
and
Howayda M. Ahmed
Chemistry Department, Faculty of Science, Cairo University, Cairo, Egypt
*
Author to whom correspondence should be addressed.
Sci. Pharm. 2005, 73(4), 173-193; https://doi.org/10.3797/scipharm.aut-05-14
Submission received: 28 August 2005 / Accepted: 22 September 2005 / Published: 30 December 2005

Abstract

New Plastic membrane ion-selective electrode for trimetazidine dihydrochloride based on trimetazidinium-tetraphenylborate was prepared. The electrode exhibited mean calibration graph slope of 29.5 mV per decade of (TrimCl2) concentration at 25°C. The electrode can be used within the concentration range 3.2×10-5-10-2 M (TrimCl2) at pH ranges of 1.5-3.8 and 4.5-7.5. The standard electrode potentials were determined at different temperatures and used to calculate the isothermal coefficient of the electrode, which was (0.00088 V°C-1). The electrode showed a very good selectivity for (TrimCl2) with respect to a number of inorganic cations, sugars and amino acids. The electrode was applied to the potentiometric determination of the trimetazidine ion and its pharmaceutical preparation under batch and flow injection conditions. Graphite coated wire was prepared and characterized as sensor for the drug under investigation. Also, trimetazidine was determined by conductimetric titrations.
Keywords: Trimetazidine ion-selective electrode; coated graphite-pencil rod; conductimetric titrations; flow injection analysis Trimetazidine ion-selective electrode; coated graphite-pencil rod; conductimetric titrations; flow injection analysis

Share and Cite

MDPI and ACS Style

Issa, Y.M.; Abdel-Ghani, N.T.; Ahmed, H.M. Flow Injection Analysis With Trimetazidine Dihydrochloride Ion-Selective Electrode. Sci. Pharm. 2005, 73, 173-193. https://doi.org/10.3797/scipharm.aut-05-14

AMA Style

Issa YM, Abdel-Ghani NT, Ahmed HM. Flow Injection Analysis With Trimetazidine Dihydrochloride Ion-Selective Electrode. Scientia Pharmaceutica. 2005; 73(4):173-193. https://doi.org/10.3797/scipharm.aut-05-14

Chicago/Turabian Style

Issa, Yousry M., Nour T. Abdel-Ghani, and Howayda M. Ahmed. 2005. "Flow Injection Analysis With Trimetazidine Dihydrochloride Ion-Selective Electrode" Scientia Pharmaceutica 73, no. 4: 173-193. https://doi.org/10.3797/scipharm.aut-05-14