Review

Effectiveness and Safety of Xen Gel Stent in Glaucoma Surgery: A Systematic Review of the Literature

1
Eye Clinic, IRCCS San Martino Polyclinic Hospital, 16132 Genoa, Italy
2
Department of Neurosciences, Rehabilitation, Ophthalmology, Genetics, Maternal and Child Health (DiNOGMI), University of Genoa, 16126 Genoa, Italy
3
Centro Italiano Glaucoma, 20124 Milan, Italy
4
Department of Surgical Sciences, University of Turin, 10122 Turin, Italy
5
Department of Surgical, Medical and Molecular Pathology and Critical Care Medicine, University of Pisa, 56126 Pisa, Italy
6
AbbVie S.r.l., 04011 Campoverde, LT, Italy
7
AbbVie Inc., North Chicago, IL 60064, USA
8
Department of Medicine and Surgery, University of Parma, 43121 Parma, Italy
*
Author to whom correspondence should be addressed.
J. Clin. Med. 2023, 12(16), 5339; https://doi.org/10.3390/jcm12165339
Submission received: 13 July 2023 / Revised: 8 August 2023 / Accepted: 10 August 2023 / Published: 16 August 2023
(This article belongs to the Special Issue Advances in Glaucoma Surgery)

Abstract

:
Although topical medical therapy and selective-laser-trabeculoplasty represent the treatments of choice to reduce intraocular pressure, many patients do not achieve adequate glaucoma control; therefore, they require further options and eventually surgery. Trabeculectomy is still considered the gold standard, but the surgical management of glaucoma has undergone continuous advances in recent years, XEN-gel-stent has been introduced as a safer and less traumatic means of lowering intraocular pressure (IOP) in patients with open-angle glaucoma (OAG). This study aimed to review the effectiveness and safety of clinical data on XEN-stent in OAG patients with a Synthesis-Without-Meta-analysis (SWiM) methodology. A total of 339 studies were identified following a literature search adhering to PRISMA guidelines and, after evaluation, 96 studies are discussed. XEN63 and XEN45 device data were collected both short and long term. In addition, this document has evaluated different aspects related to the XEN implant, including: its role compared to trabeculectomy; the impact of mitomycin-C dose on clinical outcomes; postoperative management of the device; and the identification of potential factors that might predict its clinical outcomes. Finally, current challenges and future perspectives of XEN stent, such as its use in fragile or high myopia patients, were discussed.

1. Introduction

The term open-angle glaucoma covers a wide range of chronic and progressive optic neuropathies which have the loss of retinal ganglion cells and their axons in common, as well as the subsequent loss of the visual field [1].
Lowering intraocular pressure (IOP) is currently considered as the main known modifiable risk factor [2]. Topical hypotensive medication and selective laser trabeculoplasty are currently considered as the first treatment approaches in most patients [3]. However, some patients do not achieve adequate glaucoma control; therefore, they require further therapies and eventually surgery [3,4,5], such as trabeculectomy [6], which unfortunately may lead to potential vision-threatening complications [7].
Glaucoma surgery has experienced important advances over the last several years.
One of the most important advances in glaucoma surgery in recent years was the development of the minimally or microinvasive glaucoma surgery (MIGS) devices [8]. They have been developed as safer and less traumatic means of lowering IOP in patients with glaucoma [8,9].
The definition of the term MIGS has been evolving since its introduction [9,10], and the generally accepted definition of MIGS has been changing over the years [11].
Among the different MIGS devices, XEN gel stents obtained the CE mark in December 2015 and were approved by the Food and Drug Administration (FDA) in November 2016 [12].
According to the classical definition, XEN is not defined as a MIGS because it is a bleb-forming device [9]; therefore, minimally invasive or micro-incisional filtration surgery have been suggested as more appropriate terms.
The aim of the current paper is to review the effectiveness and safety clinical data of XEN stent in open-angle glaucoma (OAG) patients.

XEN Gel Stent

The XEN gel stent (AbbVie Inc., Chicago, IL, USA) was originally developed as an ab interno procedure that reduces IOP by draining aqueous fluid from the anterior chamber into the subconjunctival space [13,14,15]. Unlike other MIGS devices that target Schlemm’s canal and the supraciliary space to lower IOP, XEN gel stent was the first micro-incisional filtration surgical procedure to drain aqueous to subconjunctival space [14,15].
The stent is a hydrophilic tube that is 6 mm long, and it is composed of porcine gelatine crosslinked with glutaraldehyde to prevent degradation when implanted [13,16].
The XEN device is based on the Hagen–Poiseuille law of laminar flow, where the length and the inner diameter of the tube determine the flow resistance and, consequently, the flow rate. Three different devices with different inner diameters, namely 45, 63, and 140 μm, were investigated [13].
The 140 μm XEN device has not been commercialized to date and the evidence is limited to a single paper [17]. The evidence evaluating the clinical outcomes of the XEN63 device is very limited [18,19,20,21,22,23] and most studies were performed with an earlier version of the device injector that was never marketed [18,19,20,21]. The new XEN63 device uses the same needle injector as the XEN45 for preventing early sideflow and hypotony [22].
Although the XEN45 device was originally designed for ab interno implantation [13,14,15], surgeons have been gaining experience with the device, and different changes aiming to provide better clinical outcomes have been introduced in the implantation technique [24,25,26].
Systematic reviews are currently considered as an essential source of evidence when making decisions in the clinical management of patients [27]. However, there are some limitations in relation to the method used (Preferred Reporting Items for Systematic Reviews and Meta-Analyses [PRISMA]; http://www.prisma-statement.org/, accessed on 1 July 2023) [28,29]. Pooling data is not possible in all cases due to high levels of heterogeneity or lack of data [12,29]. In addition, there is a growing clinical demand to respond to complex questions or situations that incorporate various data sources due to their characteristics [30,31]. These facts have opened the door to a growing number of narrative syntheses of quantitative data. However, a major concern about such papers is lack of transparency and the consequent introduction of bias [32,33].
Synthesis Without Meta-analysis (SWiM) has emerged as a systematic review to address questions that meta-analysis may not be able to provide an adequate answer for [34].
This SWiM aims to examine the currently available scientific evidence to answer different clinical questions about the XEN device.

2. Materials and Methods

2.1. Information Sources and Search Strategy

This SWiM was carried out according to the guidelines of the of the PRISMA statement [35], although the current systematic review was not registered.
A group of Italian glaucoma specialists convened to review the currently available evidence about the efficacy and safety of XEN devices in patients with glaucoma. Searches of MEDLINE, the Cochrane Database, EMBASE, and Google Scholar were conducted using the search terms “Glaucoma”, “Open-angle glaucoma”, “XEN”, “MIGS”, “Combined surgery”, and “Gel implant”. References cited in selected articles were also reviewed to identify additional relevant reports.
In addition, abstracts from the American Glaucoma Society, American Academy of Ophthalmology, European Glaucoma Society, and Association of Research in Vision and Ophthalmology were manually searched for relevant publications.

2.2. Study Selection, Data Extraction, and Synthesis Method

The authors independently generated the queries for the literature search and selected the articles fulfilling the criteria established for each subject and solved any disagreement through discussion and consensus.
Limits were set for articles written in English, French, Spanish, Portuguese, and Italian with human subjects. The studies were published between August 2014 and April 2023.
Additionally, the following exclusion criteria were applied: (1) the study did not examine clinical outcomes or response to treatment of XEN device (any device); (2) the study was on subjects other than human adults; (3) the publication was a review article, an editorial, or an opinion piece.
To synthesize the results, we applied the SWiM guidelines [34].

3. Results

3.1. Results of the PRISMA Procedure

The steps of the literature search are summarized in the PRISMA 2009 flow diagram (Figure 1). A total of 339 records were identified after the initial search. After the removal of duplicates, 187 scientific papers remained. After careful review, 96 papers met all the requirements of the inclusion/exclusion criteria and were included for qualitative synthesis (Figure 1).

3.2. XEN63

The currently available scientific evidence evaluating the clinical outcomes of XEN63 is very limited [18,19,20,21,22,23] (See Table 1), and most of the evidence was generated from a former device that was never commercialized; therefore, we will not go into detail analysing the results of those articles [18,19,20,21].
To date, there are two papers evaluating the efficacy and safety of the XEN63 device. Fea et al. [22] evaluated the new XEN63 device and found a mean (95% confidence interval, 95% CI) IOP lowering effect of −14.8 (−20.1 to −9.5) mmHg, p < 0.0001 at month-3, reporting that the mean IOP achieved with XEN63 was consistently lower than that obtained with XEN45. This may be an important point because the larger calibre of the Xen 63 device could by clinically meaningful for patients with high preoperative IOP.
Additionally, Fea et al. [23] evaluated the effectiveness and safety of the new XEN63 device over a follow-up period of 18 months in patients with glaucoma in a real clinical setting. They reported significant IOP lowering and a reduction in the number of ocular hypotensive drugs during a follow-up period of 18 months. Moreover, the incidence rate of adverse events was relatively low and most of them were mild in severity [23].
Table 1 summarizes the main outcomes of XEN63 according to the current evidence.

3.3. XEN45

3.3.1. Effect on Intraocular Pressure Lowering and Reduction in the Number of IOP-Lowering Medications

Multiple studies have evaluated the IOP lowering effect of the XEN device, either alone or in combination with cataract surgery, in patients with glaucoma [36,37,38,39,40,41,42,43,44,45,46,47,48,49,50,51,52,53,54,55,56,57,58,59,60,61,62,63,64,65,66,67,68,69,70,71,72,73,74,75,76,77,78,79,80,81,82,83,84,85,86,87,88,89,90,91,92,93,94,95,96,97,98,99,100,101]. Although most of the studies were performed in patients with primary open-angle glaucoma (POAG), other studies were performed in patients with pseudoexfoliative glaucoma (PXG) [40,42,54,72,98] or glaucoma secondary to uveitis [47].
The results of a systematic review and meta-analysis published recently, which used a pooled analysis with a random effects model, have shown a mean (95% CI) IOP lowering from baseline of −7.8 (−7.4 to −8.2) mmHg and −8.4 (−6.9 to −9.8) mmHg in the eyes of patients who underwent XEN-solo and XEN + Phaco, respectively. All patients were treated and followed up as routine clinical practice between May 2013 and February 2020. The mean sample size was 79 ± 67 and the average follow-up time was 17.0 ± 8.1 months [12].
Similarly, Yang et al. [102] did not find significant differences in IOP lowering between XEN-solo and XEN + Phaco (standardized mean difference: −0.01, 95% CI −0.09 to 0.08, p value 0.894). Moreover, Panarelli et al. [103] reported that, on average, successful gel stent surgery achieved a postoperative IOP of approximately 14.0 mm Hg and reduction to fewer than 1 ocular hypotensive medication.
Regarding the number of ocular hypotensive medications (See Table 2, Table 3 and Table 4), Chen et al. [12] reported a significant reduction in the number of ocular hypotensive drugs in both XEN-solo procedures (Mean: −1.97 drugs; 95% CI: −2.19 to −1.75 drugs, p < 0.001) and in the XEN + Phaco ones (Mean: −1.86 drugs; 95% CI: −2.11 to −1.60 drugs, p < 0.001).
Similar results have been published by Yan et al. [102], who found a statistically significant reduction in ocular hypotensive medications (standardized mean difference: 2.11, 95% CI 1.84 to 2.38, p value <  0.001). Furthermore, after adjusting for different covariates, the reduction in the number of ocular hypotensive drugs was significantly lower in the XEN + Phaco group than in the XEN-solo group (Risk ratio: 1.45, 95% CI 1.06 to 1.99, p value 0.019) [102].

IOP Lowering at 12 Months

Thirty-four studies have evaluated the IOP lowering effect of the XEN device over 12 months of follow-up (see Table 2). On average, the results of the different studies pointed in the same general direction, mostly indicating that XEN45 provided IOPs below 15 mmHg after one year of follow-up (Table 2), which is consistent with the most recent meta-analysis published [12,101,102].

IOP Lowering at 24 Months

Twelve studies have evaluated the IOP lowering effect of XEN45 after 24 months of follow-up (See Table 3). The results of the different studies included in this SWiM paper have shown that, after 2 years of follow-up, XEN45 provided a good IOP lowering, with final IOPs ≤15 mmHg and an IOP lowering that ranged between −5 and −10 mmHg (Table 3).

IOP Lowering at 36-Months

Six papers have evaluated the longer-term efficacy of XEN45 in terms of IOP lowering and reducing the number of ocular hypotensive medications, (See Table 4). With the exception of one paper (n = 23 eyes), which reported an IOP of 19.6 ± 2.1 mmHg at 36 months [78], currently available evidence shows good hypotensive effects for XEN45, with IOPs at 36–48 months in the range of 13–14 mmHg, as well as a significant reduction in the number of ocular hypotensive medications [45,61,81,87,88,94,101]. In addition, compared to preoperative values, the reduction in IOP was greater than 30% (Table 4).
Current evidence has demonstrated the efficacy and safety of the XEN45 device in the short, medium, and long term.

3.3.2. Can XEN45 Be Safely and Effectively Implanted in Myopic Patients?

Evidence surrounding the use of XEN45 in myopic OAG patients is limited [52,64,104,105,106]. In fact, there are currently only two studies that specifically evaluated the efficacy and safety of XEN45 in patients with high myopia.
Laborda-Guirao et al. did not find significant differences between OAG eyes with or without high myopia in intraocular pressure (IOP) lowering, success rate, reduction in the number of ocular hypotensive medications, or postoperative complications, which clearly suggested that XEN45 may be safely and effectively used in glaucomatous eyes with high myopia [52].
Chao et al. [64], have retrospectively evaluated the effectiveness and safety of the XEN45 in East Asian patients with primary open angle glaucoma (POAG). Although this study did not specifically evaluate the efficacy and safety of XEN45 in patients with high myopia, the mean spherical equivalent was −5.13 ± 4.44 diopters, with a range of −13.63 to −2.88 diopters and mean axial length of 26.67 ± 1.65 mm (up to 29.34 mm). According to the results of this study, axial length was not significantly associated with success, either complete (odds ratio: 0.082, p = 0.559) or qualified (odds ratio: 0.659; p = 0.186); need of subsequent intervention (odds ratio: 0.959, p = 0.803); or need of additional surgery (odds ratio: 1.382, p = 0.382). Additionally, no major complications were observed [64]. These results suggested that axial length did not have any influence on XEN45 outcomes.
Sacchi et al. [105], in a retrospective study (which included seven eyes followed for 2 years) evaluated the effectiveness and safety of Xen 45 in patients with medically uncontrolled glaucoma and concomitant high myopia (>6 Diopters). Preoperative mean IOP was significantly lowered from 22.1± 4.9 mmHg to 14.8 ± 4.0 mmHg at month-24, p < 0.0001. Regarding safety, two eyes had hypotony (without maculopathy) and one eye had choroidal detachment [106]. According to the results of this study, the XEN implant had a better safety profile than trabeculectomy, with a similar hypotensive profile [105].
Additionally, Fea et al. assessed the effectiveness and safety of XEN in 31 glaucomatous eyes with a refractive error higher than −6 D and an axial length ≥26 mm [106]. Mean preoperative IOP (95% CI) was significantly lowered from 23.5 (20.5–26.4) mm Hg to 13.0 (12.2–13.8) mm Hg, p < 0.0001. Regarding safety, hypotony (an IOP <6 mm Hg) was reported in eight eyes (28.6%) during the first postoperative day and remained for a week [106].
Conversely, one publication described the clinical outcomes of XEN45 in a high myopic eye. The patient required a XEN removal to control IOP, which resulted in loss of the remaining visual field in the eye [104].
Currently available scientific evidence suggests that XEN45 may be effectively and safely implanted in myopic eyes. However, there is a need for a prospective study specifically evaluating the clinical outcomes of XEN45 in eyes with OAG and high myopia.

3.3.3. XEN-Solo Versus XEN + Phaco: Is There a Difference in Terms of IOP Lowering?

This question has brought an increasing interest among glaucoma specialists.
Fifteen papers have compared the efficacy of XEN45 solo or in combination with cataract surgery (phacoemulsification). After reviewing the literature, the most plausible conclusion is that there is no agreement regarding the superiority of the solo procedure over the combined procedure with cataract surgery (See Table 5).
According to the results of Chen et al. [12], both XEN alone (mean difference: −7.8 mmHg; 95% CI: −8.21 to −7.38 mmHg, p < 0.001) and XEN + Phaco (mean difference: −8.35 mmHg; 95% CI: −9.82 to −6.88 mmHg, p < 0.001) significantly lowered the IOP.
Similarly, Yang et al. [103], in a systematic review and metanalysis, did not find significant differences between XEN and phaco-XEN surgery in terms of IOP after surgery (standardized mean difference: −0.01, 95% CI −0.09 to 0.08, p value 0.894). Nevertheless, the reduction in the number of ocular hypotensive medications was greater in the XEN solo group (p = 0.019) [102].
However, another systematic review and meta-analysis published by Wang et al. [107] has shown different results. They found better results for XEN alone compared to XEN + Phaco procedures in IOP lowering, but not in reducing ocular hypotensive medications [107].
In light of the available scientific evidence and based on our own experience, the XEN45 device, either alone or in combination with cataract surgery (phacoemulsification), significantly lowers IOP and reduces the number of ocular hypotensive medications. Similar to trabeculectomy, there are no data to support the superiority of the combined surgery over the solo surgery, and vice versa.

3.3.4. XEN Versus Trabeculectomy: Is There a Difference in Terms of IOP Lowering?

Due mainly to its well-stablished IOP lowering effect, trabeculectomy is currently considered as the gold standard in glaucoma surgery [5]. However, it may lead to potential vision-threatening complications [6].
The Gold-Standard Pathway Study (GPS) [108] was the first multicenter, prospective, and randomized study comparing the effectiveness and safety of XEN45 versus trabeculectomy in glaucoma that was poorly controlled with topical IOP-lowering therapy. The results of this study showed that the mean IOP change from baseline was significantly greater at month 12 in the trabeculectomy group (0.024), although both treatments significantly lowered IOP from preoperative values (p < 0.001). XEN and trabeculectomy significantly reduced the preoperative mean number of ocular hypotensive medications, without significant differences between them (p = 0.068).
In addition, XEN was noninferior to trabeculectomy, regarding the prespecified primary endpoint (62.1% and 68.2% achieved the primary endpoint, respectively; p = 0.487), namely the percentage of patients achieving ≥20% IOP reduction from baseline at Month 12 without medication increase, clinical hypotony, vision loss to counting fingers, or secondary surgical intervention. Finally, XEN resulted in less need for in-office postoperative interventions (p = 0.024 after excluding laser suture lysis), faster visual recovery (p < 0.05), and greater 6-month improvements in visual function problems (p ≤ 0.022) [108].
Similarly, Marcos-Parra et al. [48] compared the IOP lowering effect and the reduction in ocular hypotensive medications between the XEN device and trabeculectomy in OAG patients. They reported a significantly greater IOP lowering in the trabeculectomy group than in the XEN group (p = 0.001), with a similar reduction in the ocular hypotensive drugs [48]. However, this difference was mainly due to the combined surgery (glaucoma + cataract). When comparing the IOP lowering effect between XEN + Phaco and Trabeculectomy + Phaco, IOP lowering from preoperative values was found to be significantly greater in the Trabeculectomy + Phaco group at day 1, week 1, and months 1 and 3. While comparing XEN alone versus trabeculectomy alone, the only significant differences were observed in month 6. Moreover, in terms of success rates, there were no differences in the proportion of eyes who achieved a final IOP ≥6 and ≤16 mmHg (p = 0.1317), although it was slightly lower in the XEN (66.2%) than in the trabeculectomy group (78.6%) [48].
Additionally, Schlenker et al. [109] observed similar rates of complete and qualified success for both interventions.
Wang et al. [102], in a systematic review and meta-analysis, reported no significant differences between XEN gel implant and trabeculectomy on lowering IOP, although the analysis showed a high heterogeneity (I2:60%). However, after a sensitive analysis that excluded the Schenkler et al. study [109] and reduced the heterogeneity (I2:0%), the results showed that trabeculectomy was more effective in lowering IOP, without significant differences in terms of the reduction in ocular hypotensive medications [102].
Table 6 shows a comparison of the clinical outcomes of XEN45 and Trabeculectomy.
Compared to trabeculectomy, XEN45 has been associated with several advantages, including less conjunctival manipulation, less postoperative inflammation, and lower incidence of postoperative adverse events. Moreover, the rates of postoperative interventions seem to be lower with XEN45 than with trabeculectomy. Therefore, XEN45 could be considered as the first-choice surgery in certain types of patients.

3.3.5. XEN Implant: What Dose of Mitomycin-c Is the Most Effective?

Mitomycin-c (MMC) and 5-fluorouracil (5-FU) have been commonly used in traditional glaucoma filtration surgery, with good evidence suggesting a significant increase in surgery success; however, there is also an increase in the risk of complications [112].
Although MMC was not initially used in all studies [17,19,21], the use of intraoperative MMC has become a common practice of the XEN implant surgical technique [36,37,38,39,40,41,42,43,44,45,46,47,48,49,50,51,52,53,54,55,56,57,58,59,60,61,62,63,64,65,66,67,68,69,70,71,72,73,74,75,76,77,78,79,80,81,82,83,84,85,86,87,88,89,90,91,92,93,94,95,96,97,98,99,100,101].
Regarding XEN45, MMC concentration may influence clinical outcomes. It has been suggested that the success rate may be related to the MMC dose, although other studies have shown a lack of correlation between MMC dosage and the surgical outcomes [113]. In general terms, the use of MMC seemed to increase the therapeutic success rate after XEN45 gel stent implantation, although the ideal dose has not been stablished yet.
Selection of MMC concentration is based on a patient-tailored approach. This may entail an important bias because surgeons will use higher concentrations of MMC in more complicated cases in which they anticipate that the possibility of surgical failure is greater.
A study comparing the efficacy and safety of two MMC doses (0.01% versus 0.02%) in eyes who underwent a XEN45 device implant, either alone or in combination with phacoemulsification, was recently published [95]. The results of this study found that IOP lowering, number of hypotensive medication reduction, or incidence of adverse events were not related to MMC concentration [95]. The lack of significant differences, in terms of IOP lowering or reducing hypotensive medications, between MMC 0.01% and MMC 0.02% raises the possibility of using lower doses of MMC. However, it should be mentioned that MMC 0.01% was not associated with lower incidence rates of adverse events [95].
So far, there is no evidence to recommend the use of a specific MMC concentration. It seems that MMC dose does not significantly impact either the IOP lowering or the reduction in the number of ocular hypotensive medications, so it would seem prudent to recommend “the lowest dose of MMC that, in the surgeon’s opinion, may be effective in that patient”. However, new studies will be necessary to clarify this issue.

3.3.6. XEN 45 Device Implant: Postoperative Bleb Management

Bleb fibrosis is a common complication that may occur after a XEN implant, with rates as high as 45% [114]. Needling is a minimally invasive procedure that is commonly used for restoring the functionality of failed filtering blebs [115].
The most frequently reported postoperative intervention in eyes who underwent a XEN device is needling of the conjunctival bleb, which ranges from 5% to 62% [12] (Table 2, Table 3 and Table 4 show the needling rates reported in the different studies).
Needling is not currently considered as an additional procedure, but rather as part of the normal bleb management in XEN implant surgery (similarly to laser suture lysis in trabeculectomy). In most cases, needling is required within the first month postoperatively.
However, in most studies, needling has been used as a rescue strategy once bleb failure has occurred.
Primary needling, at the time of ab interno XEN implantation, has been recently proposed as a technique that may reduce the number of postoperative interventions [96,116].
In a retrospective study, Kerr et al. [116] reported that primary needling at the time of XEN insertion was associated with a significant reduction in the number of bleb interventions (p = 0.003) and, consequently, the subsequent postoperative visits (p = 0.043). Additionally, compared to preoperative values, this technique has provided a significant reduction in both IOP and ocular hypotensive medication [116].
More recently, Buenasmañanas-Maeso et al. [96] performed a retrospective study that assessed the efficacy and safety of primary needling in eyes who underwent a XEN45 implant, either alone or in combination with phacoemulsification. According to the results of this study, primary needling was associated with fewer postoperative interventions. Nevertheless, these eyes were not associated with greater IOP lowering or a greater reduction in ocular hypotensive therapy [96].
Regarding the use of primary needling, there is not enough evidence to recommend its use on a routine basis. Therefore, from a clinical point of view, its use was reserved for those cases in which the XEN implant was twisted, trapped, or did not appear free and mobile in the subconjunctival space. Additionally, it may eventually be used in eyes with previously failed glaucoma surgery or in eyes with pathologies that may be associated with an increased risk of conjunctival fibrosis.

3.3.7. Can XEN45 Device Be Considered a Safe Procedure?

Although XEN has emerged as a safer and less traumatic approach for lowering IOP in patients with glaucoma, it is not free of complications.
According to the currently available scientific evidence, transient hypotony (defined as IOP <6 mmHg) is the most commonly reported complication of XEN45, with an incidence rate of 9.59%. In the vast majority of patients, hypotony is successfully resolved without additional surgery interventions and the rate of chronic hypotony is very low [12].
The second most common adverse event is hyphema (5.53%). Most of patients have grade I hyphema (less than 1/3 of anterior chamber), which had resolved spontaneously by the first week after surgery [12]. The third most commonly reported adverse event is the incidence of transient IOP spikes, which have been reported in 2.11% of eyes (in most cases associated with hyphema) [12].
Figure 2 shows an overview of the incidence of different adverse events reported after XEN implantation. Regarding the number of corneal endothelial cells, trabeculectomy and glaucoma drainage device implantation can damage corneal endothelial cells [117,118,119,120]. With regards to the impact of XEN on the corneal endothelial cells’ loss, there were no significant differences between XEN + phacoemulsification and phacoemulsification alone [121,122,123]. Additionally, the corneal endothelial cell density reduction after XEN implantation as a solo procedure was low [122].
The results of a prospective, cross-sectional, and non-randomized clinical trial suggested that, in eyes who underwent a XEN standalone procedure, there were no significant changes in endothelial cell counts over the follow-up period (5 years). However, a statistically significant reduction in the central endothelial cell count was observed in the eyes who underwent a combined procedure (XEN + phacoemulsification) at different timepoints. This clearly suggests that the loss of corneal endothelial cells seems to be related with the cataract surgery, rather than with the XEN implant [123].
Both the experience of the panel and the scientific evidence have shown an acceptable safety profile of the XEN implant. The rate of serious complications (endophthalmitis, hypotonus maculopathy, corneal or macular edema) is low. Moreover, compared to trabeculectomy, the XEN implant has provided a better safety profile [93].

3.3.8. XEN45 Device: Is It Possible to Predict Its Clinical Outcomes?

Different studies have evaluated potential predictive factors for failure [19,42,50,70,98] or for success [52,64,116].
Three studies [42,50,70] found that none of the evaluated factors were significantly associated with surgery failure, whereas two studies reported that the risk of failure is greater in men than in women [19,98]. However, Gabbay et al. [87] reported that women were more likely to fail. Gillmann et al. [98] observed that eyes with a primary OAG diagnosis and those requiring needling are more likely to fail.
Fea et al. [70] suggested that patients with lower IOP at day 1 and month 1 had a higher chance of success. The same was true if the difference between month 1 and week 1 IOP was lower than 6 mmHg.
Regarding surgery success, each mmHg increase in preoperative IOP was positively associated with surgery success (Odds ratio: 1.33; 95% CI: 1.13 to 1.55, p = 0.0004) [52].
Chao et al. [64] reported that surgical success was more likely for eyes with a better preoperative visual field mean deviation and lower IOP at day 1, week 2, and month 1.
Conversely, Ibáñez-Muñoz et al. [116] found that none of the analysed factors were statistically associated with success.
The question of whether patient ethnicity impacts the XEN clinical outcomes has not been fully elucidated. Success rates appear to be lower in Black and Afro-Latino patients than in Caucasian populations [12,51,56]. Similarly, it has been observed that the ethnic Chinese group presents a reoperation rate as high as 45.9% [64]. However, other studies have suggested that there is no relationship between the ethnic origin of the patients and the clinical outcomes of XEN [39,62].

4. Discussion

The objective of this article was to address, in the most practical way possible, different aspects related to the XEN device that may generate doubts, mainly for specialists who are beginning to use this technique.
Based on currently available evidence, the XEN45 device, either alone or in combination with cataract surgery, significantly lowers IOP and reduces the number of ocular hypotensive medications in patients with glaucoma [12,36,37,38,39,40,41,42,43,44,45,46,47,48,49,50,51,52,53,54,55,56,57,58,59,60,61,62,63,64,65,66,67,68,69,70,71,72,73,74,75,76,77,78,79,80,81,82,83,84,85,86,87,88,89,90,91,92,93,94,95,96,97,98,99,100,101,108,110,111,116]. There are promising results with the new XEN63, but the evidence is very limited [22,23].
Compared to trabeculectomy [12,48,58,65,72,84,108,109,110,111], XEN seems not to be comparable in terms of IOP lowering, although XEN has similar efficacy in reducing the number of ocular hypotensive medications and provided a better safety profile.
Regarding its implantation technique, XEN has usually been delivered using an ab interno approach through a corneal incision [36,37,38,39,40,41,42,43,44,45,46,47,48,49,50,51,52,53,54,55,56,57,58,59,60,61,62,63,64,65,66,67,68,69,70,71,72,73,74,75,76,77,78,79,80,81,82,83,84,85,86,87,88,89,90,91,92,93,94,95,96,97,98,99,100,101]. However, as surgeons have been gaining experience with the device, different changes that aim to provide better clinical outcomes have been introduced in the implantation technique [24,25,26].
High myopia represents a challenging scenario in glaucoma patients who need surgical treatment. There are two studies that assessed the efficacy and safety of XEN in patients with high myopia. The results of both studies have pointed in the same direction, clearly indicating that the XEN device is an effective and safe option in patients with high myopia [105,106].
Trabeculectomy is still considered as the gold standard in glaucoma surgery [5]. It effectively lowers IOP and reduced the need of ocular hypotensive medications [6]. However, it requires a close postoperative follow-up to prevent potential complications that may lead to severe vision loss [7,124,125,126].
Nevertheless, XEN has been associated with lower conjunctival manipulation and incidence of postoperative complications and adverse effects [12,48,93,100].
For precisely this reason, this need for postoperative conjunctival manipulation has been one of the main points of discussion of XEN. The vast majority of the studies show a rate of needling which ranges from approximately 20 to 40% (See Table 2 and Table 3).
Primary needling (at the time of ab interno XEN implantation) has emerged as a valuable option to reduce the number of postoperative interventions [96,116].
Regarding the existence of predictive factors of the XEN clinical outcomes, current evidence does not show conclusive results.
Finally, regarding the impact of the XEN device on the corneal endothelium, current evidence suggests that XEN does not have a significant impact on corneal endothelial cell loss [121,122,123], unlike what occurs with trabeculectomy or other drainage devices [117,118,119,120].

5. Conclusions

According to published evidence, the XEN45 device lowers IOP by approximately 35% from preoperative values, obtaining a mean IOP value of ≤15 mmHg, as far as 4 years after surgery. In addition, XEN significantly reduced the need for ocular hypotensive medication, with a mean number of postoperative hypotensive medications ≤ 1 drug.
Based on the evidence and panel’s opinion, the XEN device may be considered as the first surgical option in patients who require a target IOP in the mid to low teens. Even though trabeculectomy seems to have a better IOP-lowering effect, the XEN device has been shown to have a better safety profile.
Although XEN device implantation is a relatively new procedure, a large number of studies have been published in recent years, pointing to its long-term potential in the treatment of glaucoma.
Several issues related to XEN clinical outcomes remain to be clarified, such as the role of ethnicity, factors influencing the outcomes, needling rates and the role of primary needling, the impact of high myopia, and the use of XEN in different types of glaucoma, including narrow-angle and angle-closure glaucoma.
Data from randomized and multicentre clinical trials will help surgeons develop patient-tailored management strategies.

Author Contributions

Conceptualization, C.E.T. and S.G.; methodology, R.G.C. and A.M.F.; software, C.A. and B.P.; validation, M.F. and A.M.F.; formal analysis, C.A.; investigation, Not Applicable; resources, R.G.C. and S.G.; data curation, V.V.; writing—original draft preparation, V.V. and M.F.; writing—review and editing, S.G. and C.E.T.; visualization, M.F.; supervision, M.F.; project administration, C.A. and B.P.; funding acquisition, C.E.T. and M.F. All authors have read and agreed to the published version of the manuscript.

Funding

The medical writer and editorial assistance for this manuscript was supported by AbbVie with no input into the preparation, review, approval, and writing of the manuscript. The authors maintained complete control over the manuscript content, and it reflects their opinions.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Acknowledgments

The authors wish to thank Antonio Martinez from Ciencia y Deporte SL, for medical writing and editorial assistance.

Conflicts of Interest

This manuscript was funded by AbbVie S.r.l. AbbVie participated in writing, reviewing, and approving the publication. No honoraria or payments were made for authorship. C. Astarita and V. Vera are AbbVie employees and may own AbbVie stocks/options. B. Piergentili was a former AbbVie employee. She is now affiliated with Alnylam Italia Srl. She has nothing else to declare. A.M. Fea is consultant for Glaukos, EliosVision, Ivantis, Oculus, EyeD, iSTAR, Santen, Abbvie. S. Gandolfi research contracts and/or unrestricted research Grants from Ivantis, Glaukos, AERIE and Allergan/AbbVie, Advisory Board of Santen, Novartis, Baush and Lomb, Allergan/AbbVie, Fidia, Visufarma and AERIE, Speakers’ bureau of Allergan/AbbVie, Santen, Omikron, Visufarma, Baush and Lomb and Novartis. C. E. Traverso: support to the Department from Novartis, Bausch and Lomb, Alcon, Santen. Occasional travel support/speaker fee from Abbvie, Alcon, Santen, Omikron. R. Carassa and M. Figus have no conflicts of interest to declare.

References

  1. Weinreb, R.N.; Leung, C.K.; Crowston, J.G.; Medeiros, F.A.; Friedman, D.S.; Wiggs, J.L.; Martin, K.R. Primary open-angle glaucoma. Nat. Rev. Dis. Prim. 2016, 2, 16067. [Google Scholar] [CrossRef] [PubMed]
  2. Heijl, A.; Leske, M.C.; Bengtsson, B.; Hyman, L.; Bengtsson, B.; Hussein, M.; Early Manifest Glaucoma Trial Group. Reduction of intraocular pressure and glaucoma progression: Results from the Early Manifest Glaucoma Trial. Arch. Ophthalmol. 2002, 120, 1268–1279. [Google Scholar] [CrossRef] [PubMed]
  3. European Glaucoma Society Terminology and Guidelines for Glaucoma, 5th Edition. Br. J. Ophthalmol. 2021, 105 (Suppl. S1), 1–169. [CrossRef]
  4. Lichter, P.R.; Musch, D.C.; Gillespie, B.W.; Guire, K.E.; Janz, N.K.; Wren, P.A.; CIGTS Study Group. Interim clinical outcomes in the Collaborative Initial Glaucoma Treatment Study comparing initial treatment randomized to medications or surgery. Ophthalmology 2001, 108, 1943–1953. [Google Scholar] [CrossRef] [PubMed]
  5. Newman-Casey, P.A.; Robin, A.L.; Blachley, T.; Farris, K.; Heisler, M.; Resnicow, K.; Lee, P.P. The Most Common Barriers to Glaucoma Medication Adherence: A Cross-Sectional Survey. Ophthalmology 2015, 122, 1308–1316. [Google Scholar] [CrossRef]
  6. Landers, J.; Martin, K.; Sarkies, N.; Bourne, R.; Watson, P. A twenty-year follow-up study of trabeculectomy: Risk factors and outcomes. Ophthalmology 2012, 119, 694–702. [Google Scholar] [CrossRef]
  7. Jampel, H.D.; Musch, D.C.; Gillespie, B.W.; Lichter, P.R.; Wright, M.M.; Guire, K.E.; Collaborative Initial Glaucoma Treatment Study Group. Perioperative complications of trabeculectomy in the collaborative initial glaucoma treatment study (CIGTS). Am. J. Ophthalmol. 2005, 140, 16–22. [Google Scholar] [CrossRef]
  8. Bar-David, L.; Blumenthal, E.Z. Evolution of Glaucoma Surgery in the Last 25 Years. Rambam Maimonides Med. J. 2018, 9, e0024. [Google Scholar] [CrossRef]
  9. Saheb, H.; Ahmed, I.I. Micro-invasive glaucoma surgery: Current perspectives and future directions. Curr. Opin. Ophthalmol. 2012, 23, 96–104. [Google Scholar] [CrossRef]
  10. Ahmed, I.I. MIGS and the FDA: What’s in a Name? Ophthalmology 2015, 122, 1737–1739. [Google Scholar] [CrossRef]
  11. Caprioli, J.; Kim, J.H.; Friedman, D.S.; Kiang, T.; Moster, M.R.; Parrish, R.K., 2nd; Rorer, E.M.; Samuelson, T.; Tarver, M.E.; Singh, K.; et al. Special Commentary: Supporting Innovation for Safe and Effective Minimally Invasive Glaucoma Surgery: Summary of a Joint Meeting of the American Glaucoma Society and the Food and Drug Administration, Washington, DC, February 26, 2014. Ophthalmology 2015, 122, 1795–1801. [Google Scholar] [CrossRef]
  12. Chen, X.Z.; Liang, Z.Q.; Yang, K.Y.; Lv, K.; Ma, Y.; Li, M.Y.; Li, M.Y.; Wu, H.J. The Outcomes of XEN Gel Stent Implantation: A Systematic Review and Meta-Analysis. Front. Med. 2022, 9, 804847. [Google Scholar] [CrossRef] [PubMed]
  13. Lewis, R.A. Ab interno approach to the subconjunctival space using a collagen glaucoma stent. J. Cataract. Refract. Surg. 2014, 40, 1301–1306. [Google Scholar] [CrossRef]
  14. Lavia, C.; Dallorto, L.; Maule, M.; Ceccarelli, M.; Fea, A.M. Minimally-invasive glaucoma surgeries (MIGS) for open angle glaucoma: A systematic review and meta-analysis. PLoS ONE 2017, 12, e0183142. [Google Scholar] [CrossRef] [PubMed]
  15. Ansari, E. An Update on Implants for Minimally Invasive Glaucoma Surgery (MIGS). Ophthalmol. Ther. 2017, 6, 233–241. [Google Scholar] [CrossRef]
  16. Shute, T.S.; Dietrich, U.M.; Baker, J.F.; Carmichael, K.P.; Wustenberg, W.; Ahmed, I.I.; Sheybani, A. Biocompatibility of a Novel Microfistula Implant in Nonprimate Mammals for the Surgical Treatment of Glaucoma. Investig. Opthalmol. Vis. Sci. 2016, 57, 3594–3600. [Google Scholar] [CrossRef]
  17. Sheybani, A.; Dick, H.B.; Ahmed, I.I. Early Clinical Results of a Novel Ab Interno Gel Stent for the Surgical Treatment of Open-angle Glaucoma. J. Glaucoma 2016, 25, e691–e696. [Google Scholar] [CrossRef]
  18. Sheybani, A.; Lenzhofer, M.; Hohensinn, M.; Reitsamer, H.; Ahmed, I.I. Phacoemulsification combined with a new ab interno gel stent to treat open-angle glaucoma: Pilot study. J. Cataract. Refract. Surg. 2015, 41, 1905–1909. [Google Scholar] [CrossRef] [PubMed]
  19. Lenzhofer, M.; Kersten-Gomez, I.; Sheybani, A.; Gulamhusein, H.; Strohmaier, C.; Hohensinn, M.; Burkhard Dick, H.; Hitzl, W.; Eisenkopf, L.; Sedarous, F.; et al. Four-year results of a minimally invasive transscleral glaucoma gel stent implantation in a prospective multi-centre study. Clin. Exp. Ophthalmol. 2019, 47, 581–587. [Google Scholar] [CrossRef]
  20. Fernández-García, A.; Zhou, Y.; García-Alonso, M.; Andrango, H.D.; Poyales, F.; Garzón, N. Comparing Medium-Term Clinical Outcomes following XEN® 45 and XEN® 63 Device Implantation. J. Ophthalmol. 2020, 2020, 4796548. [Google Scholar] [CrossRef] [PubMed]
  21. Lavin-Dapena, C.; Cordero-Ros, R.; D’Anna, O.; Mogollón, I. XEN 63 gel stent device in glaucoma surgery: A 5-years follow-up prospective study. Eur. J. Ophthalmol. 2021, 31, 1829–1835. [Google Scholar] [CrossRef]
  22. Fea, A.M.; Menchini, M.; Rossi, A.; Posarelli, C.; Malinverni, L.; Figus, M. Early Experience with the New XEN63 Implant in Primary Open-Angle Glaucoma Patients: Clinical Outcomes. J. Clin. Med. 2021, 10, 1628. [Google Scholar] [CrossRef]
  23. Fea, A.M.; Menchini, M.; Rossi, A.; Posarelli, C.; Malinverni, L.; Figus, M. Outcomes of XEN 63 Device at 18-Month Follow-Up in Glaucoma Patients: A Two-Center Retrospective Study. J. Clin. Med. 2022, 11, 3801. [Google Scholar] [CrossRef]
  24. Panarelli, J.F.; Yan, D.B.; Francis, B.; Craven, E.R. XEN Gel Stent Open Conjunctiva Technique: A Practical Approach Paper. Adv. Ther. 2020, 37, 2538–2549. [Google Scholar] [CrossRef]
  25. Vera, V.; Gagne, S.; Myers, J.S.; Ahmed, I.I.K. Surgical Approaches for Implanting Xen Gel Stent without Conjunctival Dissection. Clin. Ophthalmol. 2020, 14, 2361–2371. [Google Scholar] [CrossRef] [PubMed]
  26. Tan, N.E.; Tracer, N.; Terraciano, A.; Parikh, H.A.; Panarelli, J.F.; Radcliffe, N.M. Comparison of Safety and Efficacy Between Ab Interno and Ab Externo Approaches to XEN Gel Stent Placement. Clin. Ophthalmol. 2021, 15, 299–305. [Google Scholar] [CrossRef] [PubMed]
  27. Donnelly, C.A.; Boyd, I.; Campbell, P.; Craig, C.; Vallance, P.; Walport, M.; Whitty, C.J.M.; Woods, E.; Wormald, C. Four principles to make evidence synthesis more useful for policy. Nature 2018, 558, 361–364. [Google Scholar] [CrossRef]
  28. Liberati, A.; Altman, D.G.; Tetzlaff, J.; Mulrow, C.; Gøtzsche, P.C.; Ioannidis, J.P.; Clarke, M.; Devereaux, P.J.; Kleijnen, J.; Moher, D. The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate healthcare interventions: Explanation and elaboration. BMJ 2009, 339, b2700. [Google Scholar] [CrossRef] [PubMed]
  29. Petticrew, M.; Anderson, L.; Elder, R.; Grimshaw, J.; Hopkins, D.; Hahn, R.; Krause, L.; Kristjansson, E.; Mercer, S.; Sipe, T.; et al. Complex interventions and their implications for systematic reviews: A pragmatic approach. J. Clin. Epidemiol. 2013, 66, 1209–1214. [Google Scholar] [CrossRef]
  30. Cochrane Collaboration. Cochrane Strategy to 2020. 2017. Available online: https://community.cochrane.org/sites/default/files/uploads/00%20S2020%202017%20Targets%20End%20of%20Year%20Report.pdf (accessed on 21 April 2023).
  31. Smith, E.A.; Cooper, N.J.; Sutton, A.J.; Abrams, K.R.; Hubbard, S.J. A review of the quantitative effectiveness evidence synthesis methods used in public health intervention guidelines. BMC Public Health 2021, 21, 278. [Google Scholar] [CrossRef]
  32. Valentine, J.C.; Wilson, S.J.; Rindskopf, D.; Lau, T.S.; Tanner-Smith, E.E.; Yeide, M.; LaSota, R.; Foster, L. Synthesizing Evidence in Public Policy Contexts: The Challenge of Synthesis When There Are Only a Few Studies. Eval. Rev. 2017, 41, 3–26. [Google Scholar] [CrossRef] [PubMed]
  33. Campbell, M.; Katikireddi, S.V.; Sowden, A.; Thomson, H. Lack of transparency in reporting narrative synthesis of quantitative data: A methodological assessment of systematic reviews. J. Clin. Epidemiol. 2019, 105, 1–9. [Google Scholar] [CrossRef]
  34. Campbell, M.; McKenzie, J.E.; Sowden, A.; Katikireddi, S.V.; Brennan, S.E.; Ellis, S.; Hartmann-Boyce, J.; Ryan, R.; Shepperd, S.; Thomas, J.; et al. Synthesis without meta-analysis (SWiM) in systematic reviews: Reporting guideline. BMJ 2020, 368, l6890. [Google Scholar] [CrossRef] [PubMed]
  35. Moher, D.; Liberati, A.; Tetzlaff, J.; Altman, D.G.; PRISMA Group. Preferred reporting items for systematic reviews and meta-analyses: The PRISMA statement. PLoS Med. 2009, 6, e1000097. [Google Scholar] [CrossRef] [PubMed]
  36. Pérez-Torregrosa, V.T.; Olate-Pérez, Á.; Cerdà-Ibáñez, M.; Gargallo-Benedicto, A.; Osorio-Alayo, V.; Barreiro-Rego, A.; Duch-Samper, A. Combined phacoemulsification and XEN45 surgery from a temporal approach and 2 incisions. Arch Soc. Esp. Oftalmol. 2016, 91, 415–421. [Google Scholar] [CrossRef] [PubMed]
  37. Ozal, S.A.; Kaplaner, O.; Basar, B.B.; Guclu, H.; Ozal, E. An innovation in glaucoma surgery: XEN45 gel stent implantation. Arq. Bras. Oftalmol. 2017, 80, 382–385. [Google Scholar] [CrossRef] [PubMed]
  38. Galal, A.; Bilgic, A.; Eltanamly, R.; Osman, A. XEN Glaucoma Implant with Mitomycin C 1-Year Follow-Up: Result and Complications. J. Ophthalmol. 2017, 2017, 5457246. [Google Scholar] [CrossRef]
  39. Grover, D.S.; Flynn, W.J.; Bashford, K.P.; Lewis, R.A.; Duh, Y.J.; Nangia, R.S.; Niksch, B. Performance and Safety of a New Ab Interno Gelatin Stent in Refractory Glaucoma at 12 Months. Am. J. Ophthalmol. 2017, 183, 25–36. [Google Scholar] [CrossRef]
  40. Hengerer, F.H.; Kohnen, T.; Mueller, M.; Conrad-Hengerer, I. Ab Interno Gel Implant for the Treatment of Glaucoma Patients with or Without Prior Glaucoma Surgery: 1-Year Results. J. Glaucoma 2017, 26, 1130–1136. [Google Scholar] [CrossRef]
  41. De Gregorio, A.; Pedrotti, E.; Russo, L.; Morselli, S. Minimally invasive combined glaucoma and cataract surgery: Clinical results of the smallest ab interno gel stent. Int. Ophthalmol. 2018, 38, 1129–1134. [Google Scholar] [CrossRef]
  42. Mansouri, K.; Guidotti, J.; Rao, H.L.; Ouabas, A.; D’Alessandro, E.; Roy, S.; Mermoud, A. Prospective Evaluation of Standalone XEN Gel Implant and Combined Phacoemulsification-XEN Gel Implant Surgery: 1-Year Results. J. Glaucoma 2018, 27, 140–147. [Google Scholar] [CrossRef] [PubMed]
  43. Tan, S.Z.; Walkden, A.; Au, L. One-year result of XEN45 implant for glaucoma: Efficacy, safety, and postoperative management. Eye 2018, 32, 324–332. [Google Scholar] [CrossRef]
  44. Widder, R.A.; Dietlein, T.S.; Dinslage, S.; Kühnrich, P.; Rennings, C.; Rössler, G. The XEN45 Gel Stent as a minimally invasive procedure in glaucoma surgery: Success rates, risk profile, and rates of re-surgery after 261 surgeries. Graefes Arch. Clin. Exp. Ophthalmol. 2018, 256, 765–771. [Google Scholar] [CrossRef]
  45. Reitsamer, H.; Sng, C.; Vera, V.; Lenzhofer, M.; Barton, K.; Stalmans, I.; Apex Study Group. Two-year results of a multicenter study of the ab interno gelatin implant in medically uncontrolled primary open-angle glaucoma. Graefes Arch. Clin. Exp. Ophthalmol. 2019, 257, 983–996. [Google Scholar] [CrossRef] [PubMed]
  46. Qureshi, A.; Jones, N.P.; Au, L. Urgent Management of Secondary Glaucoma in Uveitis Using the Xen-45 Gel Stent. J. Glaucoma 2019, 28, 1061–1066. [Google Scholar] [CrossRef] [PubMed]
  47. Kalina, A.G.; Kalina, P.H.; Brown, M.M. XEN® Gel Stent in Medically Refractory Open-Angle Glaucoma: Results and Observations After One Year of Use in the United States. Ophthalmol. Ther. 2019, 8, 435–446. [Google Scholar] [CrossRef]
  48. Parra, M.T.M.; López, J.A.S.; Grau, N.S.L.; Ceausescu, A.M.; Santonja, J.J.P. XEN implant device versus trabeculectomy, either alone or in combination with phacoemulsification, in open-angle glaucoma patients. Graefes Arch. Clin. Exp. Ophthalmol. 2019, 257, 1741–1750. [Google Scholar] [CrossRef]
  49. Lenzhofer, M.; Strohmaier, C.; Hohensinn, M.; Hitzl, W.; Steiner, V.; Baca, B.; Moussa, S.; Motloch, K.; Reitsamer, H.A. Change in visual acuity 12 and 24 months after transscleral ab interno glaucoma gel stent implantation with adjunctive Mitomycin C. Graefes Arch. Clin. Exp. Ophthalmol. 2019, 257, 2707–2715. [Google Scholar] [CrossRef]
  50. Sng, C.C.A.; Chew, P.T.K.; Htoon, H.M.; Lun, K.; Jeyabal, P.; Ang, M. Case Series of Combined XEN Implantation and Phacoemulsification in Chinese Eyes: One-Year Outcomes. Adv. Ther. 2019, 36, 3519–3529. [Google Scholar] [CrossRef]
  51. Gabbay, I.E.; Allen, F.; Morley, C.; Pearsall, T.; Bowes, O.M.; Ruben, S. Efficacy and safety data for the XEN45 implant at 2 years: A retrospective analysis. Br. J. Ophthalmol. 2020, 104, 1125–1130. [Google Scholar] [CrossRef]
  52. Laborda-Guirao, T.; Cubero-Parra, J.M.; Hidalgo-Torres, A. Efficacy and safety of XEN 45 gel stent alone or in combination with phacoemulsification in advanced open angle glaucoma patients: 1-year retrospective study. Int. J. Ophthalmol. 2020, 13, 1250–1256. [Google Scholar] [CrossRef] [PubMed]
  53. Heidinger, A.; Schwab, C.; Lindner, E.; Riedl, R.; Mossböck, G. A Retrospective Study of 199 Xen45 Stent Implantations from 2014 to 2016. J. Glaucoma 2019, 28, 75–79. [Google Scholar] [CrossRef] [PubMed]
  54. Ibáñez-Muñoz, A.; Soto-Biforcos, V.S.; Chacón-González, M.; Rúa-Galisteo, O.; Arrieta-Los Santos, A.; Lizuain-Abadía, M.E.; Del Río Mayor, J.L. One-year follow-up of the XEN® implant with mitomycin-C in pseudoexfoliative glaucoma patients. Eur. J. Ophthalmol. 2019, 29, 309–314. [Google Scholar] [CrossRef] [PubMed]
  55. Karimi, A.; Lindfield, D.; Turnbull, A.; Dimitriou, C.; Bhatia, B.; Radwan, M.; Gouws, P.; Hanifudin, A.; Amerasinghe, N.; Jacob, A. A multi-centre interventional case series of 259 ab-interno Xen gel implants for glaucoma, with and without combined cataract surgery. Eye 2019, 33, 469–477. [Google Scholar] [CrossRef]
  56. Laroche, D.; Nkrumah, G.; Ng, C. Real-World Retrospective Consecutive Study of Ab Interno XEN 45 Gel Stent Implant with Mitomycin C in Black and Afro-Latino Patients with Glaucoma: 40% Required Secondary Glaucoma Surgery at 1 Year. Middle East Afr. J. Ophthalmol. 2020, 26, 229–234. [Google Scholar] [CrossRef] [PubMed]
  57. Smith, M.; Charles, R.; Abdel-Hay, A.; Shah, B.; Byles, D.; Lim, L.A.; Rossiter, J.; Kuo, C.H.; Chapman, P.; Robertson, S. 1-year outcomes of the Xen45 glaucoma implant. Eye 2019, 33, 761–766. [Google Scholar] [CrossRef]
  58. Teus, M.A.; Moreno-Arrones, J.P.; Castaño, B.; Castejon, M.A.; Bolivar, G. Optical coherence tomography analysis of filtering blebs after long-term, functioning trabeculectomy and XEN® stent implant. Graefes Arch. Clin. Exp. Ophthalmol. 2019, 257, 1005–1011. [Google Scholar] [CrossRef]
  59. Mansouri, K.; Bravetti, G.E.; Gillmann, K.; Rao, H.L.; Ch’ng, T.W.; Mermoud, A. Two-Year Outcomes of XEN Gel Stent Surgery in Patients with Open-Angle Glaucoma. Ophthalmol. Glaucoma 2019, 2, 309–318. [Google Scholar] [CrossRef]
  60. Gillmann, K.; Bravetti, G.E.; Rao, H.L.; Mermoud, A.; Mansouri, K. Combined and stand-alone XEN 45 gel stent implantation: 3-year outcomes and success predictors. Acta Ophthalmol. 2021, 99, e531–e539. [Google Scholar] [CrossRef]
  61. Post, M.; Lubiński, W.; Śliwiak, D.; Podborączyńska-Jodko, K.; Mularczyk, M. XEN Gel Stent in the management of primary open-angle glaucoma. Doc. Ophthalmol. 2020, 141, 65–76. [Google Scholar] [CrossRef]
  62. Bravetti, G.E.; Mansouri, K.; Gillmann, K.; Rao, H.L.; Mermoud, A. XEN-augmented Baerveldt drainage device implantation in refractory glaucoma: 1-year outcomes. Graefes Arch. Clin. Exp. Ophthalmol. 2020, 258, 1787–1794. [Google Scholar] [CrossRef] [PubMed]
  63. Scheres, L.M.J.; Kujovic-Aleksov, S.; Ramdas, W.D.; de Crom, R.M.P.C.; Roelofs, L.C.G.; Berendschot, T.T.J.M.; Webers, C.A.B.; Beckers, H.J.M. XEN® Gel Stent compared to PRESERFLO™ MicroShunt implantation for primary open-angle glaucoma: Two-year results. Acta Ophthalmol. 2021, 99, e433–e440. [Google Scholar] [CrossRef]
  64. Chao, Y.J.; Ko, Y.C.; Chen, M.J.; Lo, K.J.; Chang, Y.F.; Liu, C.J. XEN45 Gel Stent implantation in eyes with primary open angle glaucoma: A study from a single hospital in Taiwan. J. Chin. Med. Assoc. 2021, 84, 108–113. [Google Scholar] [CrossRef] [PubMed]
  65. Olgun, A.; Aktas, Z.; Ucgul, A.Y. XEN gel implant versus gonioscopy-assisted transluminal trabeculotomy for the treatment of open-angle glaucoma. Int. Ophthalmol. 2020, 40, 1085–1093. [Google Scholar] [CrossRef] [PubMed]
  66. Schargus, M.; Theilig, T.; Rehak, M.; Busch, C.; Bormann, C.; Unterlauft, J.D. Outcome of a single XEN microstent implant for glaucoma patients with different types of glaucoma. BMC Ophthalmol. 2020, 20, 490. [Google Scholar] [CrossRef]
  67. Theilig, T.; Rehak, M.; Busch, C.; Bormann, C.; Schargus, M.; Unterlauft, J.D. Comparing the efficacy of trabeculectomy and XEN gel microstent implantation for the treatment of primary open-angle glaucoma: A retrospective monocentric comparative cohort study. Sci. Rep. 2020, 10, 19337. [Google Scholar] [CrossRef]
  68. Theillac, V.; Blumen-Ohana, E.; Akesbi, J.; Hamard, P.; Sellam, A.; Brasnu, E.; Baudouin, C.; Labbe, A.; Nordmann, J.P. Cataract and glaucoma combined surgery: XEN® gel stent versus nonpenetrating deep sclerectomy, a pilot study. BMC Ophthalmol. 2020, 20, 231. [Google Scholar] [CrossRef]
  69. Subaşı, S.; Yüksel, N.; Özer, F.; Tugan, B.Y.; Pirhan, D. A Retrospective Analysis of Safety and Efficacy of XEN 45 Microstent Combined Cataract Surgery in Open-Angle Glaucoma over 24 Months. Turk. J. Ophthalmol. 2021, 51, 139–145. [Google Scholar] [CrossRef]
  70. Fea, A.M.; Bron, A.M.; Economou, M.A.; Laffi, G.; Martini, E.; Figus, M.; Oddone, F. European study of the efficacy of a cross-linked gel stent for the treatment of glaucoma. J. Cataract. Refract. Surg. 2020, 46, 441–450. [Google Scholar] [CrossRef]
  71. Rauchegger, T.; Angermann, R.; Willeit, P.; Schmid, E.; Teuchner, B. Two-year outcomes of minimally invasive XEN Gel Stent implantation in primary open-angle and pseudoexfoliation glaucoma. Acta Ophthalmol. 2021, 99, 369–375. [Google Scholar] [CrossRef] [PubMed]
  72. Wagner, F.M.; Schuster, A.K.; Emmerich, J.; Chronopoulos, P.; Hoffmann, E.M. Efficacy and safety of XEN®–Implantation vs. trabeculectomy: Data of a “real-world” setting. PLoS ONE 2020, 15, e0231614. [Google Scholar] [CrossRef] [PubMed]
  73. Busch, T.; Skiljic, D.; Rudolph, T.; Bergström, A.; Zetterberg, M. Learning Curve and One-Year Outcome of XEN 45 Gel Stent Implantation in a Swedish Population. Clin. Ophthalmol. 2020, 14, 3719–3733. [Google Scholar] [CrossRef] [PubMed]
  74. Barão, R.C.; José, P.; Teixeira, F.J.; Ferreira, N.P.; Sens, P.; Pinto, L.A. Automated Gonioscopy Assessment of XEN45 Gel Stent Angle Location After Isolated XEN or Combined Phaco-XEN Procedures: Clinical Implications. J. Glaucoma 2020, 29, 932–940. [Google Scholar] [CrossRef] [PubMed]
  75. Ucar, F.; Cetinkaya, S. Xen implantation in patients with primary open-angle glaucoma: Comparison of two different techniques. Int. Ophthalmol. 2020, 40, 2487–2494. [Google Scholar] [CrossRef] [PubMed]
  76. Navero-Rodríguez, J.M.; Espinosa-Barberi, G.; Morilla-Grasa, A.; Anton, A. Efficacy of the Ologen collagen matrix in combination with the XEN gel stent implantation in the treatment of open-angle glaucoma: A case-control study. Clin. Exp. Ophthalmol. 2020, 48, 1003–1005. [Google Scholar] [CrossRef]
  77. Stoner, A.M.; Young, C.E.C.; SooHoo, J.R.; Pantcheva, M.B.; Patnaik, J.L.; Kahook, M.Y.; Seibold, L.K. A Comparison of Clinical Outcomes After XEN Gel Stent and EX-PRESS Glaucoma Drainage Device Implantation. J. Glaucoma 2021, 30, 481–488. [Google Scholar] [CrossRef] [PubMed]
  78. Düzgün, E.; Olgun, A.; Karapapak, M.; Alkan, A.A.; Ustaoğlu, M. Outcomes of XEN Gel Stent Implantation in the Inferonasal Quadrant after Failed Trabeculectomy. J. Curr. Glaucoma Pract. 2021, 15, 64–69. [Google Scholar] [CrossRef] [PubMed]
  79. Schargus, M.; Busch, C.; Rehak, M.; Meng, J.; Schmidt, M.; Bormann, C.; Unterlauft, J.D. Functional Monitoring after Trabeculectomy or XEN Microstent Implantation Using Spectral Domain Optical Coherence Tomography and Visual Field Indices-A Retrospective Comparative Cohort Study. Biology 2021, 10, 273. [Google Scholar] [CrossRef]
  80. Nuzzi, R.; Gremmo, G.; Toja, F.; Marolo, P. A Retrospective Comparison of Trabeculectomy, Baerveldt Glaucoma Implant, and Microinvasive Glaucoma Surgeries in a Three-Year Follow-Up. Semin. Ophthalmol. 2021, 36, 839–849. [Google Scholar] [CrossRef]
  81. Eraslan, M.; Özcan, A.A.; Dericioğlu, V.; Çiloğlu, E. Multicenter case series of standalone XEN implant vs. combination with phacoemulsification in Turkish patients. Int. Ophthalmol. 2021, 41, 3371–3379. [Google Scholar] [CrossRef]
  82. Bormann, C.; Schmidt, M.; Busch, C.; Rehak, M.; Scharenberg, C.T.; Unterlauft, J.D. Implantation of XEN After Failed Trabeculectomy: An Efficient Therapy? Klin. Monbl. Augenheilkd. 2022, 239, 86–93. [Google Scholar] [CrossRef]
  83. Lewczuk, K.; Konopińska, J.; Jabłońska, J.; Rudowicz, J.; Laszewicz, P.; Dmuchowska, D.A.; Mariak, Z.; Rękas, M. XEN Glaucoma Implant for the Management of Glaucoma in Naïve Patients versus Patients with Previous Glaucoma Surgery. J. Clin. Med. 2021, 10, 4417. [Google Scholar] [CrossRef] [PubMed]
  84. Wanichwecharungruang, B.; Ratprasatporn, N. 24-month outcomes of XEN45 gel implant versus trabeculectomy in primary glaucoma. PLoS ONE 2021, 16, e0256362. [Google Scholar] [CrossRef]
  85. Lewczuk, K.; Konopińska, J.; Jabłońska, J.; Rudowicz, J.; Laszewicz, P.; Mariak, Z.; Rękas, M. XEN Glaucoma Implant for the Management of Operated Uncontrolled Glaucoma: Results and Complications during a Long-Term Follow-Up. J. Ophthalmol. 2021, 2021, 2321922. [Google Scholar] [CrossRef] [PubMed]
  86. Reitsamer, H.; Vera, V.; Ruben, S.; Au, L.; Vila-Arteaga, J.; Teus, M.; Lenzhofer, M.; Shirlaw, A.; Bai, Z.; Balaram, M.; et al. Three-year effectiveness and safety of the XEN gel stent as a solo procedure or in combination with phacoemulsification in open-angle glaucoma: A multicentre study. Acta Ophthalmol. 2022, 100, e233–e245. [Google Scholar] [CrossRef]
  87. Gabbay, I.E.; Goldberg, M.; Allen, F.; Lin, Z.; Morley, C.; Pearsall, T.; Muraleedharan, V.; Ruben, S. Efficacy and safety data for the Ab interno XEN45 gel stent implant at 3 Years: A retrospective analysis. Eur. J. Ophthalmol. 2021, 32, 11206721211014381. [Google Scholar] [CrossRef]
  88. José, P.; Teixeira, F.J.; Barão, R.C.; Sens, P.; Pinto, L.A. Needling after XEN gel implant: What’s the efficacy? A 1-year analysis. Eur. J. Ophthalmol. 2021, 31, 3087–3092. [Google Scholar] [CrossRef] [PubMed]
  89. Nicolaou, S.; Khatib, T.Z.; Lin, Z.; Sheth, T.; Ogbonna, G.; Hamidovic, L.; Zaheer, A.; Dimitriou, C. A retrospective review of XEN implant surgery: Efficacy, safety and the effect of combined cataract surgery. Int. Ophthalmol. 2022, 42, 881–889. [Google Scholar] [CrossRef]
  90. Olsen, L.P.; Ruhlmann, P.B.; Vestergaard, A.H. Implantation of the XEN® 45 Gel Stent in patients with glaucoma at a University Hospital—A retrospective quality control study. Acta Ophthalmol. 2021, 99, e968–e969. [Google Scholar] [CrossRef]
  91. Franco, F.G.S.; Branchetti, M.; Spagnuolo, V.; Piergentili, M.; Serino, F.; De Vitto, M.L.; Bertelli, E.; Virgili, G.; Giasanti, F. Efficacy and safety of Ab interno XEN gel implant after a failed filtering surgery. Romnian J. Ophthalmol. 2021, 65, 365–370. [Google Scholar] [CrossRef]
  92. Steiner, S.; Resch, H.; Kiss, B.; Buda, D.; Vass, C. Needling and open filtering bleb revision after XEN-45 implantation–A retrospective outcome comparison. Graefes Arch. Clin. Exp. Ophthalmol. 2021, 259, 2761–2770. [Google Scholar] [CrossRef] [PubMed]
  93. Cappelli, F.; Cutolo, C.A.; Olivari, S.; Testa, V.; Sindaco, D.; Pizzorno, C.; Ciccione, S.; Traaverso, C.E.; Iester, M. Trabeculectomy versus Xen gel implant for the treatment of open-angle glaucoma: A 3-year retrospective analysis. BMJ Open Ophthalmol. 2022, 7, e000830. [Google Scholar] [CrossRef] [PubMed]
  94. Szigiato, A.A.; Touma, S.; Jabbour, S.; Lord, F.; Agoumi, Y.; Singh, H. Efficacy of ab-interno gelatin microstent implantation in primary and refractory glaucoma. Can. J. Ophthalmol. 2022, 58, 328–337. [Google Scholar] [CrossRef] [PubMed]
  95. Monja-Alarcón, N.; Perucho-Martínez, S.; Buenasmañanas-Maeso, M.; Toledano-Fernández, N. Does mitomycin-C concentration have any influence on XEN45 gel stent outcomes in a real-world setting? Graefes Arch. Clin. Exp. Ophthalmol. 2022, 260, 2649–2661. [Google Scholar] [CrossRef]
  96. Buenasmañanas-Maeso, M.; Perucho-Martínez, S.; Monja-Alarcón, N.; Toledano-Fernández, N. Impact of Primary Needling on the XEN Implant Clinical Outcomes: A Real-Life Retrospective Study. Clin. Ophthalmol. 2022, 16, 935–946. [Google Scholar] [CrossRef]
  97. Hüppi, R.; Wagels, B.; Todorova, M. Two-Year Outcome of Surgery in Glaucoma Patients. Klin. Monbl. Augenheilkd. 2022, 239, 435–442. [Google Scholar] [CrossRef]
  98. Gillmann, K.; Bravetti, G.E.; Mermoud, A.; Rao, H.L.; Mansouri, K. XEN Gel Stent in Pseudoexfoliative Glaucoma: 2-Year Results of a Prospective Evaluation. J. Glaucoma 2019, 28, 676–684. [Google Scholar] [CrossRef]
  99. Almendral-Gómez, J.; Perucho-Martínez, S.; Martín-Giral, E.; Fernández-Escámez, C.; Buenasmañanas-Maeso, M.; Monja-Alarcón, N.; Toledano-Fernández, N. XEN Gel Stent Versus Non-penetrating Deep Sclerectomy in Ocular-hypertension and Open-angle Glaucoma Patients. J. Glaucoma 2023, 32, 511–519. [Google Scholar] [CrossRef]
  100. Marcos-Parra, M.T.; Salinas-López, J.A.; Mateos-Marcos, C.; Moreno-Castro, L.; Mendoza-Moreira, A.L.; Pérez-Santonja, J.J. Long-Term Effectiveness of XEN 45 Gel-Stent in Open-Angle Glaucoma Patients. Clin. Ophthalmol. 2023, 17, 1223–1232. [Google Scholar] [CrossRef]
  101. Vukmirovic, A.; Ong, J.; Mukhtar, A.; Yu, D.Y.; Morgan, W.H. Outcomes of 45 μm gelatin stent surgery over 24-monthfollow-up. Clin. Exp. Ophthalmol. 2023, 51, 19–30. [Google Scholar] [CrossRef]
  102. Yang, X.; Zhao, Y.; Zhong, Y.; Duan, X. The efficacy of XEN gel stent implantation in glaucoma: A systematic review and meta-analysis. BMC Ophthalmol. 2022, 22, 305. [Google Scholar] [CrossRef]
  103. Panarelli, J.F.; Vera, V.; Sheybani, A.; Radcliffe, N.; Fiscella, R.; Francis, B.A.; Smith, O.U.; Noecker, R.J. Intraocular Pressure and Medication Changes Associated with Xen Gel Stent: A Systematic Review of the Literature. Clin. Ophthalmol. 2023, 17, 25–46. [Google Scholar] [CrossRef] [PubMed]
  104. Huth, A.; Viestenz, A. Hohe Myopie bei vitrektomiertem Auge: Kontraindikation für minimal-invasive glaukomchirurgische Implantate? Ophthalmologe 2020, 117, 461–466. [Google Scholar] [CrossRef]
  105. Sacchi, M.; Fea, A.M.; Monsellato, G.; Tagliabue, E.; Villani, E.; Ranno, S.; Nucci, P. Safety and Efficacy of Ab Interno XEN 45 Gel Stent in Patients with Glaucoma and High Myopia. J. Clin. Med. 2023, 12, 2477. [Google Scholar] [CrossRef]
  106. Fea, A.; Sacchi, M.; Franco, F.; Laffi, G.L.; Oddone, F.; Costa, G.; Serino, F.; Giansanti, F. Effectiveness and Safety of XEN45 in Eyes With High Myopia and Open Angle Glaucoma. J. Glaucoma 2023, 32, 178–185. [Google Scholar] [CrossRef] [PubMed]
  107. Wang, B.; Leng, X.; An, X.; Zhang, X.; Liu, X.; Lu, X. XEN gel implant with or without phacoemulsification for glaucoma: A systematic review and meta-analysis. Ann. Transl. Med. 2020, 8, 1309. [Google Scholar] [CrossRef]
  108. Sheybani, A.; Vera, V.; Grover, D.S.; Vold, S.D.; Cotter, F.; Bedrood, S.; Sawhney, G.; Piette, S.D.; Simonyi, S.; Gu, X.; et al. Gel Stent vs Trabeculectomy: The Randomized, Multicenter, Gold Standard Pathway Study (GPS) of Effectiveness and Safety at 12 Months: Gel Stent vs Trabeculectomy: A Prospective Randomized Study. Am. J. Ophthalmol. 2023, 252, 306–325. [Google Scholar] [CrossRef] [PubMed]
  109. Schlenker, M.B.; Gulamhusein, H.; Conrad-Hengerer, I.; Somers, A.; Lenzhofer, M.; Stalmans, I.; Reitsamer, H.; Hengerer, F.H.; Ahmed, I.I.K. Efficacy, Safety, and Risk Factors for Failure of Standalone Ab Interno Gelatin Microstent Implantation versus Standalone Trabeculectomy. Ophthalmology 2017, 124, 1579–1588. [Google Scholar] [CrossRef]
  110. Sacchi, M.; Agnifili, L.; Brescia, L.; Oddone, F.; Villani, E.; Nucci, P.; Mastropasqua, L. Structural imaging of conjunctival filtering blebs in XEN gel implantation and trabeculectomy: A confocal and anterior segment optical coherence tomography study. Graefes Arch. Clin. Exp. Ophthalmol. 2020, 258, 1763–1770. [Google Scholar] [CrossRef]
  111. Olgun, A.; Duzgun, E.; Yildiz, A.M.; Atmaca, F.; Yildiz, A.A.; Sendul, S.Y. XEN Gel Stent versus trabeculectomy: Short-term effects on corneal endothelial cells. Eur. J. Ophthalmol. 2021, 31, 346–353. [Google Scholar] [CrossRef]
  112. Cabourne, E.; Clarke, J.C.; Schlottmann, P.G.; Evans, J.R. Mitomycin C versus 5-Fluorouracil for wound healing in glaucoma surgery. Cochrane Database Syst. Rev. 2015, 2015, CD006259. [Google Scholar] [CrossRef] [PubMed]
  113. Bell, K.; Bezerra, B.d.P.S.; Mofokeng, M.; Montesano, G.; Nongpiur, M.E.; Marti, M.V.; Lawlor, M. Learning from the past: Mitomycin C use in trabeculectomy and its application in bleb-forming minimally invasive glaucoma surgery. Surv. Ophthalmol. 2021, 66, 109–123. [Google Scholar] [CrossRef] [PubMed]
  114. Midha, N.; Gillmann, K.; Chaudhary, A.; Mermoud, A.; Mansouri, K. Efficacy of Needling Revision After XEN Gel Stent Implantation: A Prospective Study. J. Glaucoma 2020, 29, 11–14. [Google Scholar] [CrossRef] [PubMed]
  115. Feldman, R.M.; Tabet, R.R. Needle revision of filtering blebs. J. Glaucoma 2008, 17, 594–600. [Google Scholar] [CrossRef]
  116. Kerr, N.M.; Lim, S.; Simos, M.; Ward, T. Primary Needling of the Ab Interno Gelatin Microstent Reduces Postoperative Needling and Follow-up Requirements. Ophthalmol. Glaucoma 2021, 4, 581–588. [Google Scholar] [CrossRef]
  117. Gedde, S.J.; Herndon, L.W.; Brandt, J.D.; Budenz, D.L.; Feuer, W.J.; Schiffman, J.C. Surgical complications in the Tube Versus Trabeculectomy Study during the first year of follow-up. Am. J. Ophthalmol. 2007, 143, 23–31.e2. [Google Scholar] [CrossRef] [PubMed]
  118. Kim, M.S.; Kim, K.N.; Kim, C.S. Changes in Corneal Endothelial Cell after Ahmed Glaucoma Valve Implantation and Trabeculectomy: 1-Year Follow-up. Korean J. Ophthalmol. 2016, 30, 416–425. [Google Scholar] [CrossRef]
  119. Shaheer, M.; Amjad, A.; Ahmed, N. Comparison of Mean Corneal Endothelial Cell Loss after Trabeculectomy with and without Mitomycin C. J. Coll. Physicians Surg. Pak. 2018, 28, 301–303. [Google Scholar] [CrossRef]
  120. Iwasaki, K.; Arimura, S.; Takihara, Y.; Takamura, Y.; Inatani, M. Prospective cohort study of corneal endothelial cell loss after Baerveldt glaucoma implantation. PLoS ONE 2018, 13, e0201342. [Google Scholar] [CrossRef]
  121. Gillmann, K.; Bravetti, G.E.; Rao, H.L.; Mermoud, A.; Mansouri, K. Impact of Phacoemulsification Combined with XEN Gel Stent Implantation on Corneal Endothelial Cell Density: 2-Year Results. J. Glaucoma 2020, 29, 155–160. [Google Scholar] [CrossRef]
  122. Oddone, F.; Roberti, G.; Posarelli, C.; Agnifili, L.; Mastropasqua, L.; Carnevale, C.; Micelli Ferrari, T.; Pace, V.; Sacchi, M.; Cremonesi, E.; et al. Endothelial Cell Density After XEN Implant Surgery: Short-term Data from the Italian XEN Glaucoma Treatment Registry (XEN-GTR). J. Glaucoma 2021, 30, 559–565. [Google Scholar] [CrossRef] [PubMed]
  123. Lenzhofer, M.; Motaabbed, A.; Colvin, H.P.; Hohensinn, M.; Steiner, V.; Hitzl, W.; Runge, C.; Moussa, S.; Reitsamer, H.A. Five-year follow-up of corneal endothelial cell density after transscleral ab interno glaucoma gel stent implantation. Graefes Arch. Clin. Exp. Ophthalmol. 2023, 261, 1073–1082. [Google Scholar] [CrossRef] [PubMed]
  124. Gedde, S.J.; Herndon, L.W.; Brandt, J.D.; Budenz, D.L.; Feuer, W.J.; Schiffman, J.C.; Tube Versus Trabeculectomy Study Group. Postoperative complications in the Tube Versus Trabeculectomy (TVT) study during five years of follow-up. Am. J. Ophthalmol. 2012, 153, 804–814.e1. [Google Scholar] [CrossRef] [PubMed]
  125. Kirwan, J.F.; Lockwood, A.J.; Shah, P.; Macleod, A.; Broadway, D.C.; King, A.J.; Trabeculectomy Outcomes Group Audit Study Group. Trabeculectomy in the 21st century: A multicenter analysis. Ophthalmology 2013, 120, 2532–2539. [Google Scholar] [CrossRef]
  126. Gedde, S.J.; Feuer, W.J.; Lim, K.S.; Barton, K.; Goyal, S.; Ahmed, I.I.K.; Primary Tube Versus Trabeculectomy Study Group. Treatment Outcomes in the Primary Tube Versus Trabeculectomy Study after 3 Years of Follow-up. Ophthalmology 2020, 127, 333–345. [Google Scholar] [CrossRef]
Figure 1. PRISMA 2009 flow diagram.
Figure 1. PRISMA 2009 flow diagram.
Jcm 12 05339 g001
Figure 2. Incidence rate of adverse events reported after XEN implantation (n = 4410). Adapted from Chen et al. [12]. * ≥2 Snellen lines vison loss lasting >1 month. IOP: intraocular pressure; AC: anterior chamber; VA: visual acuity.
Figure 2. Incidence rate of adverse events reported after XEN implantation (n = 4410). Adapted from Chen et al. [12]. * ≥2 Snellen lines vison loss lasting >1 month. IOP: intraocular pressure; AC: anterior chamber; VA: visual acuity.
Jcm 12 05339 g002
Table 1. An overview of currently available scientific evidence with the XEN63 device.
Table 1. An overview of currently available scientific evidence with the XEN63 device.
StudyFollow-UpPreop IOP, mmHgFinal IOP, mmHgIOP Lowering Mean Preoperative NOHMMean NOHM, Last VisitSuccess Rates, (%) 1
Fea et al. [22]3 months27.0 ± 7.8 *12.2 ± 3.4 *−14.8 (−20.1 to −9.5) **2.3 ± 0.9 *0.1 ± 0.4 *69.6
Fea et al. [23]18 months27.0 ± 7.8 *14.1 ± 3.4 *−12.9 (−16.9 to −8.9) **2.3 ± 0.9 *1.0 ± 1.4 *77.8
* Mean (Standard deviation). ** Mean (95% confidence interval). 1 Complete success. Preop: preoperative; IOP: intraocular pressure; NOHM: number of ocular hypotensive medications.
Table 2. A comparison of the clinical outcomes of XEN45 after 12 months of follow-up.
Table 2. A comparison of the clinical outcomes of XEN45 after 12 months of follow-up.
StudyType of StudyNPreop IOP, mm HgM12 IOP, mm HgM12 IOP Lowering Mean Preoperative NOHMMean NOHM, Last VisitNeedling Rates at Last Follow-Up Visit, n (%)
Ozal et al. [37] 1Retrospective1536.1 ± 3.7 *16.7 ± 3.6 *−19.3 ± 5.0 *3.6 ± 0.5 *0.3 ± 0.9 *Not reported
Galal et al. [38]Prospective1316.0 ± 4.0 *12.0 ± 3.023 ⁑,†1.9 ± 1.0 *0.13 ± 0.11 *4 (30.7)
Grover et al. [39]Prospective6525.1 ± 3.7 *15.9 ± 5.2 *−9.1 (−10.7 to 7.5) **3.5 ± 1.0 *1.7 21 (32.3)
Hengerer et al. [40]Retrospective24232.2 ± 9.1 *14.2 ± 4.032.2 ⁑,†3.1 ± 1.0 *0.3 ± 0.7 *67 (27.7) ***
De Gregorio et al. [41]Prospective3322.5 ± 3.7 *13.1 ± 2.4 *−9.4 ± 3.1 *2.5 ± 0.9 *0.4 ± 0.8 *Not reported
Mansouri et al. [42]Prospective14920.0 ± 7.1 *13.9 ± 4.3−31 ⁑,†1.9 ± 1.3 *0.5 ± 0.8 *55 (36.9)
Widder et al. [44]Retrospective26124.3 ± 6.6 *16.8 ± 7.6 *−7.5 ± 7.1 *2.6 ± 1.1 *0.2 ± 0.7 *80 (34)
Reitsamer et al. [45]Prospective20221.4 ± 3.6 *14.9 ± 4.5 *−6.5 ± 5.3 *2.7 ± 0.9 *1.1 ± 1.2 *83 (41.1)
Kalina et al. [47]Prospective4722.3 ± 7.3 *13.4 ± 3.6 *−8.9 ± 5.8 *3.0 ± 1.2 *0.8 ± 1.3 *14 (29.8)
Marcos-Parra et al. [48]Retrospective6519.1 ± 5.4 *N.A.−6.7 (−12.9 to −0.5) **2.5 ± 0.8 *0.2 ± 0.6 *13 (20.0)
Sng et al. [50]Prospective3115.6 ± 2.7 *12.1 ± 2.6 *−3.5 ± 2.7 *1.4 ± 0.6 *0.1 ± 0.4 *Not reported
Gabbay et al. [51]Retrospective15122.1 ± 6.5 *15.4 ± 5.9 *−6.7 ± 6.2 *2.77 ± 1.1 *0.5 ± 1.0 *57 (37.7)
Laborda-Guirao et al. [52]Retrospective8021.0 ± 5.2 *14.7 (13.9 to 15.4) **−6.3 (−8.8 to −4.4) **2.8 (2.7 to 3.0) **1.1 (0.8 to 1.3) **7 (8.8)
Heidinger et al. [53]Retrospective19922.8 ± 6.9 *17.1 ± 5.9 *−5.7 ± 6.6 *2.9 ± 1.0 *1.8 ± 1.4 *44 (22.1)
Ibáñez-Muñoz et al. [54]Retrospective2122.3 (21.0−23.5) **15.3 (14.3−16.3) **−7.3 (−9.7 to −5.0) **3.0 ± 1.0 *1.2 ± 1.2 *19 (26.0)
Karimi et al. [55]Retrospective25919.3 ± 6.0 *14.3 ± 4.4 *−5.1 ± 5.6 *2.6 ± 0.1 *1.6 ± 0.5 *106 (40.9) 2
Laroche et al. [56]Retrospective2015.3 ± 6.2 *12.9 ± 4.5 *−2.4 ± 5.4 *3.6 ± 0.7 *1.8 ± 1.50 (0.0)
Smith et al. [57]Retrospective6822.1 ± 6.4 *14.8 ± 5.1 *−7.3 ± 5.8 *2.9 ± 0.8 *1.1 ± 1.1 *30 (44.1)
Mansouri et al. [59]Prospective10920.0 ± 7.5 *N.A.N.A.2.0 ± 1.3 *0.6 ± 0.9 *58 (45)
Post et al. [61]Prospective2021.6 ± 2.3 *17.7 ± 2.1 *−3.9 ± 2.2 *3.2 ± 0.8 *1.6 ± 1.0 *5 (25.0)
Chao et al. [64]Retrospective3721.7 ± 7.7 *15.0 ± 2.0 *−6.7 ± 5.6 *3.4 ± 0.9 *1.3 ± 1.5 *17 (45.9)
Schargus et al. [66]Retrospective15323.9 ± 7.4 *15.4 ± 5.1 *−8.5 ± 6.4 *2.6 ± 1.2 *0.8 ± 1.3 *64 (35.3)
Theilig et al. [67]Retrospective10024.5 ± 6.7 *16.6  ±  4.8 *N.A.3.0 ± 1.1 *1.4  ±  1.5 *42 (42.0)
Subaşı et al. [69]Retrospective3020.4 ± 4.8 *15.0 ± 1.9 *−6.2 ± 0.9 *3.1 ± 1.0 *0.9 ± 1.1 *13 (43.3)
Fea et al. [70]Prospective17123.9 ± 7.6 *15.5 ± 3.9 *−7.4 ± 7.93.0 ± 1.0 *0.5 ± 1.0 *79 (46.2)
Rauchegger et al. [71]Retrospective7923.4 ± 7.9 *14.6 ± 3.6 *31(20−42) ⁑,⁋2.7 ± 1.1 *1.0 ± 1.2 *37 (62)
Eraslam et al. [81]Retrospective2623.9 ± 6.8 *16.4 ± 2.6 *−7.5 ± 5.2 *2.9 ± 0.70.9 ± 0.921 (36.2)
Lewczuk et al. [83] 3Retrospective8625.0 ± 7.5 *16.8 ± 5.1 *−8.2 ± 6.4 *Not reportedNot reported30 (69.8)
Wanichwecharungruang and Ratprasatporn [84]Retrospective5721.6 ± 4.015 −30.6 ⁑,†2.1 ± 1.4 *0.5 ± 0.7 *10 (17.5)
Reitsamer et al. [86]Retrospective21220.7 ± 5.114.8 −5.6 2.5 ± 1.0 *0.7 ± 1.0 *78 (36.8)
Nicolau et al. [89]Retrospective18618.1 ± 5.1 *13.7 ± 5.6 *−4.4 ± 5.3 *2.5 ± 1.1 *0.8 ± 1.025 (13%)
Monja-Alarcón et al. [95]Retrospective6317.6 (0.7) 12.6 (0.3) 21.9 ± 27.6 2.1 (1.9 to 2.3) 0.2 (0.04 to 0.3) 19 (30.2)
Buenasmañanas-Maeso et al. [96]Retrospective6317.6 ± 5.3 *12.6 (12.0 to 13.3) −21.9 ± 27.4 2.1 (1.9 to 2.3) 0.2 (0.04 to 0.3) 8 (12.7)
Almendral et al. [99]Retrospective6317.6 ± 5.3 *12.6 ± 2.6 *−22.0 ± 27.5 2.1 ± 0.7 *0.2 ± 0.5 *7 (11.1)
Data about standard deviation was not provided. * Mean (Standard deviation). ** Mean (95% confidence interval). Percentage. Mean (Standard error). 95% Confidence interval. *** All the needling procedures were done between week 1 and month 3. 1 Data calculated from Table 2 of the paper. 2 Postoperative bleb needling or antimetabolite injection. 3 Data from naïve patients. Abbreviations: IOP: intraocular pressure; N: number of eyes; NOHM: number of ocular hypotensive medications.
Table 3. A comparison of the clinical outcomes of XEN45 after 24 months of follow-up.
Table 3. A comparison of the clinical outcomes of XEN45 after 24 months of follow-up.
StudyType of StudyNPreop IOP, mm HgM 24 IOP, mm HgM 24 IOP Lowering, mm HgMean Preoperative NOHMMean NOHM, Last VisitNeedling Rates at Last Follow-Up Visit, n (%)
Reitsamer et al. [45]Prospective20221.4 (3.6) *15.2 ± 4.2 *−6.2 ± 4.9 *2.7 ± 0.9 *1.1 ± 1.2 *83 (41.1)
Gabbay et al. [51]Retrospective15122.1 ± 6.5 *14.5 ± 3.3 *−7.6 ± 5.2 *2.77 ± 1.1 *0.5 ± 1.0 *57 (37.7)
Karimi et al. [55]Retrospective25919.3 ± 6.0 *13.5 ± 3.3 *,1−5.8 ± 4.8 *2.6 ± 0.1 *1.1 ± 1.3 *,1106 (40.9) 2
Mansouri et al. [59]Prospective11320.0 ± 7.5 *14.1 ± 3.7 *−6.4 ± 5.9 *2.0 ± 1.3 *0.6 ± 0.9 *58 (45)
Scheres et al. [63]Retrospective4119.2 ± 4.4 *13.8 ± 3.8 *−5.4 ± 4.1 *2.5 ± 1.4 *0.9 ± 1.2 *8 (20)
Subaşı et al. [69]Retrospective3020.4 ± 4.8 *14.8 ± 1.9 *−6.4 ± 1.2 *3.1 ± 1.0 *0.9 ± 1.1 *13 (43.3)
Rauchegger et al. [71]Retrospective7923.4 ± 7.9 *14.8 ± 4.429(30–41) 2.7 ± 1.1 *1.0 ± 1.2 *37 (62)
Wanichwecharungruang and Ratprasatporn [84]Retrospective7721.6 ± 4.014.6 ± 3.5 *32.4 ⁑,†2.1 ± 1.40.5 ± 0.7 *10 (17.5)
Lewczuk et al. [85]Retrospective7224.8 ± 8.0 *17.5 ± 5.8 *,3−7.3 ± 7.0 *Not reportedNot reported43 (67.2)
Nicolau et al. [89]Retrospective18618.1 ± 5.1 *12.6 ± 3.1 *−5.5 ± 4.2 *2.5 ± 1.1 *1.7 ± 1.7 *25 (13%)
Szigiato et al. [94]Retrospective14123.3 ± 7.0 *13.3 ± 4.7 *−10.0 ± 6.0 *3.4 ± 0.8 *1.9 ± 1.5 *54 (38.3)
Gillmann et al. [98]
POAG
PEX
Prospective57
53
19.8 ± 5.8 *
19.8 ± 8.2 *
14.5 ± 3.6 *
14.2 ± 3.8 *
−5.3 ± 4.8 *
−5.6 ± 6.4 *
1.9 ± 1.6 *
2.0 ± 1.3 *
0.6 ± 0.9 *
0.4 ± 0.7 *
42.8
43.2
Vukmirovic et al. [101]Retrsopective26220.40  ±  6.31 *Not reportedNot reported2.70 ± 1.01 *0.6 ± 0.9111 (42.4)
1 Month-18 IOP. 2 Postoperative bleb needling or antimetabolite injection. 3 IOP at the last study visit (mean follow-up period was 26.87  ±  15.33 months). Data about standard deviation was not provided. * Mean ± Standard deviation. Percentage. Abbreviations: IOP: intraocular pressure; N: number of eyes; NOHM: number of ocular hypotensive medications; PEX: Pseudoexfoliative glaucoma.
Table 4. A comparison of the long-term follow-up clinical outcomes of XEN45.
Table 4. A comparison of the long-term follow-up clinical outcomes of XEN45.
StudyType of StudyNLength of Study (months)Preoperative
IOP (mmHg)
Final IOPIOP Lowering (%)Mean Reduction
in Ocular
Hypotensive Medication
Lenzhofer et al. [19]Prospective344822.5 ± 4.2 *13.4 ± 3.1 *40.41.2
Gillmann et al. [60]
XEN alone
XEN + Phaco
Prospective
Prospective
26
76
36
36
21.0 ± 7.4 *
20.0 ± 6.9 *
12.9 ± 2.9 *
12.9 ± 3.4 *
38.6
35.5
2.1
1.4
Nuzzi et al. [80]Retrospective233624.9 ± 6.1 *19.6 ± 2.1 *21.3Not reported
Reitsamer et al. [86]Retrospective763620.7 ± 5.1 *13.9 ± 4.3 *32.91.4
Gabbay et al. [87]Retrospective2053622.6 ± 7.0 *14.0 ± 2.9 *38.12.0
Capelli et al. [93]Retrospective343623 (19–28) **Not reportedNot reportedNot reported
Marcos-Parra et al. [100]Retrospective633619.1 ± 5.0 *14.9 ± 3.8 *−4.2 ± 4.7 *−1.9 ± 0.8 *
Abbreviations: IOP: intraocular pressure; N: number of eyes. * Mean ± Standard deviation. ** Median (Interquartile range).
Table 5. A comparison of the clinical outcomes of XEN45 solo and in combination with cataract surgery (Phacoemulsification).
Table 5. A comparison of the clinical outcomes of XEN45 solo and in combination with cataract surgery (Phacoemulsification).
StudyNFollow-Up (months)Preop IOP, mm HgFinal IOP, mm HgFinal IOP Lowering Mean Preoperative NOHMMean NOHM, Last Visit
Ozal et al. [37] 1
XEN
XEN + Phaco
9
6
1236.7 ± 4.1 *
35.2 ± 3.2 *
17.0 ± 4.2 *
15.5 ± 2.3 *
−19.7 ± 4.2 *
−19.7 ± 2.8 *
3.7 ± 0.5 *
Not estimable
0.3 ± 0.9 *
Not estimable
Lenzhofer et al. [19]
XEN
XEN + Phaco
35
29
4822.5 ± 6.5 *
23.4 ± 6.3 *
13.2 ± 5.2 *
12.7 ± 6.9 *
−9.5 ± 5.9 *
−13.7 ± 6.6 *
3.0 ± 0.9 *
1.4 ± 0.6
0.8 ± 0.9 *
0.1 ± 0.4
Reitsamer et al. [45]
XEN
XEN + Phaco
106
79
2421.7 ± 3.8 *
21.0 ± 3.4 *
15.4 ± 4.2 *
14.9 ± 4.5 *
−6.3 ± 4.0 *
−6.1 ± 4.0 *
2.7 ± 0.9 *
2.9 ± 1.0
1.2 ± 1.2 *
1.4 ± 1.3 *
Kalina et al. [47]
XEN
XEN + Phaco
20
27
1224.2 ± 8.2 *
21.0 ± 6.5 *
13.0 ± 4.5 *
13.6 ± 2.9 *
−11.2 ± 6.6 *
−7.4 ± 5.0 *
Not reportedNot reported
Parra et al. [48]
XEN
XEN + Phaco
17
48
1222.2 ± 6.8 *
18.0 ± 4.5 *
N.A.−6.7 (−10.4 to −3.0) **
−3.5 (−5.0 to −2.0) **
2.5 ± 0.8 *
2.1 ± 0.9 *
0.2 ± 0.6 *
0.1 ± 0.3 *
Laborda-Guirao et al. [52]
XEN
XEN + Phaco
40
40
1221.8 ± 5.3 *
20.1 ± 5.1 *
−14.4 (−15.7 to −13.2) **
−14.9 (−15.8 to −14.1) **
−6.8 ± 0.9)
−5.9 ± 0.9)
Not reportedNot reported
Karimi et al. [55]
XEN
XEN + Phaco
187
72
1819.6 (18.8−20.5) **
18.3 (17.0−19.7) **
13.5 (12.3−14.7) **
14.0
Not reportedNot reportedNot reported
Gillmann et al. [60]
XEN
XEN + Phaco
26
66
3621.0 ± 7.4 *
20.0 ± 6.9 *
12.9 ± 2.9 *
12.9 ± 3.4 *
−8.1 ± 5.6 *
−7.1 ± 5.4 *
2.4 ± 1.5 *
1.9 ± 1.2 *
0.3 ± 0.8 *
0.5 ± 0.9 *
Olgun et al. [65]
XEN
XEN + Phaco
51
45
2424.4 ± 4.3 *
24.8 ± 3.5 *
14.2 ± 2.2 *
13.4 ± 1.4 *
−10.2 ± 3.5 *
−11.4 ± 2.7 *
3.4 ± 0.5 *
3.4 ± 0.4 *
2.2 ± 2.0 *
1.8 ± 1.7 *
Theilig et al. [67]
XEN
XEN + Phaco
48
52
1224.4 ± 6.6 *
24.8 ± 6.9 *
16.9 ± 5.9 *
16.4 ± 4.2 *
−7.5 ± 6.2 *
−8.6 ± 5.3 *
3.1 ± 1.2 *
2.8 ± 1.1 *
1.5  ±  1.4 *
0.9 ± 1.4 *
Fea et al. [70]
XEN
XEN + Phaco
115
56
1225.0
21.4
15.8
15.4
−8.8 ± 7.5 *
−4.5 ± 8.4 *
3.0 ± 1.0 *
2.5 ± 1.0 *
0.5
0.5
Schargus et al. [79]
XEN
XEN + Phaco
38
42
2424.1 ± 4.7 *
25.4 ± 5.6 *
15.7 ± 3.0 *
14.7 ± 3.2 *
−8.4 ± 3.9 *
−10.7 ± 4.6 *
3.3 ± 0.8 *
2.9 ± 0.6 *
1.2 ± 0.8 *
1.0 ± 0.4 *
Eraslam et al. [81]
XEN
XEN + Phaco
26
32
1223.7 ± 6.0 *
24.4 ± 7.2 *
16.3 ± 3.0 *
16.4 ± 2.3 *
−7.4 ± 4.7 *
−8.0 ± 5.3 *
2.9 ± 0.7 *
2.9 ± 0.6 *
1.0 ± 0.9 *
0.8 ± 0.8 *
Reitsamer et al. [86]
XEN
XEN + Phaco
98
76
3620.4 ± 4.7 *
20.4 ± 5.5 *
14.2
13.6
−6.4
−6.7
2.6 ± 1.0 *
2.4 ± 1.0 *
1.1 ± 1.3 *
1.2 ± 1.1 *
Marcos-Parra et al. [100]
XEN
XEN + Phaco
37
117
3621.2 ± 6.2 *
18.4 ± 4.3 *
14.5 ± 4.0 *
14.9 ± 3.1 *
−3.1 ± 0.8
−3.8 ± 0.4
2.6 ± 0.7 *
1.9 ± 0.8 *
0.4 ± 0.8 *,1
0.2 ± 0.5 *
Data about standard deviation was not provided. * Mean ± Standard deviation. ** Mean (95% confidence interval). Mean ± Standard error of the mean. 1 Greater reduction in the XEN-alone than in the XEN + Phaco group (mean difference: 0.4 drugs; 95% CI: 0.1 to 0.7; p = 0.0134). Abbreviations: IOP: intraocular pressure; NOHM: number of ocular hypotensive medications.
Table 6. A comparison of the clinical outcomes of XEN45 and trabeculectomy. Adapted from Chen et al. [12], Wang et al. [107], and Sheybani et al. [108].
Table 6. A comparison of the clinical outcomes of XEN45 and trabeculectomy. Adapted from Chen et al. [12], Wang et al. [107], and Sheybani et al. [108].
StudyXENTrabeculectomyMD (95% CI) between Surgeries 1
MDSDMDSD
Parra et al. [48]−4.349.3−7.738.970.37 (0.01 to 0.73)
Teus et al. [58]
Chen et al. [12]−7.29.2−10.59.2−3.30 (−6.08 to −0.52)
Wang et al. [107]−8.85.2−8.55.3−0.06 (−0.86 to 0.75)
Olgun et al. [65]−116.4−169.70.63 (0.17 to 1.09)
Wagner et al. [72]−8.55.3−8.85.2−0.30 (−4.51 to 3.91)
Wanichwecharungruang and Ratprasatporn [84]−73.8−105−3.0 (−4.65 to −1.35)
Sheybani et al. [108]−8.7 *5−10.84.7−2.1 (−3.9 to −0.3)
Schlenker et al. [109]−110.74−117.290.0 (−0.21 to 0.21)
Sacchi et al. [110]−13.382.76−15.173.20.59 (0.04 to 1.15)
Olgun et al. [111]−116.4−169.7−5.00 (−8.86 to −1.14)
1 Random effect model. * Extracted directly from the study. NP: not provided.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Traverso, C.E.; Carassa, R.G.; Fea, A.M.; Figus, M.; Astarita, C.; Piergentili, B.; Vera, V.; Gandolfi, S. Effectiveness and Safety of Xen Gel Stent in Glaucoma Surgery: A Systematic Review of the Literature. J. Clin. Med. 2023, 12, 5339. https://doi.org/10.3390/jcm12165339

AMA Style

Traverso CE, Carassa RG, Fea AM, Figus M, Astarita C, Piergentili B, Vera V, Gandolfi S. Effectiveness and Safety of Xen Gel Stent in Glaucoma Surgery: A Systematic Review of the Literature. Journal of Clinical Medicine. 2023; 12(16):5339. https://doi.org/10.3390/jcm12165339

Chicago/Turabian Style

Traverso, Carlo Enrico, Roberto G. Carassa, Antonio Maria Fea, Michele Figus, Carlo Astarita, Benedetta Piergentili, Vanessa Vera, and Stefano Gandolfi. 2023. "Effectiveness and Safety of Xen Gel Stent in Glaucoma Surgery: A Systematic Review of the Literature" Journal of Clinical Medicine 12, no. 16: 5339. https://doi.org/10.3390/jcm12165339

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop