Next Article in Journal
Immunogenicity of BNT162b2 Vaccine in Patients with Inflammatory Bowel Disease on Infliximab Combination Therapy: A Multicenter Prospective Study
Next Article in Special Issue
Glenoid Component Position Does Not Affect Short-Term Clinical and Radiologic Outcomes in Total Shoulder Arthroplasty
Previous Article in Journal
Expert Consensus on the Contraindications and Cautions of Foam Rolling—An International Delphi Study
Previous Article in Special Issue
Primary Stability in Hip Revision Arthroplasty: Comparison of the Stability of Cementless Fixed Augments on a Modular Acetabular Cage System with and without Cranial Straps
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Does Additive Pressurized Carbon Dioxide Lavage Improve Cement Penetration and Bond Strength in Cemented Arthroplasty?

1
Department of Orthopedic Surgery, Heidelberg University, 69118 Heidelberg, Germany
2
Laboratory of Biomechanics and Implant Research, Clinic for Orthopedics and Trauma Surgery, Heidelberg University, 69118 Heidelberg, Germany
3
Department of Orthopedic and Trauma Surgery, Linz University, 4020 Linz, Austria
*
Author to whom correspondence should be addressed.
J. Clin. Med. 2021, 10(22), 5361; https://doi.org/10.3390/jcm10225361
Submission received: 13 October 2021 / Revised: 10 November 2021 / Accepted: 12 November 2021 / Published: 17 November 2021
(This article belongs to the Special Issue State-of-the-Art and Research on Joint Arthroplasties)

Abstract

:
The modern cementing technique in cemented arthroplasty is a highly standardized and, therefore, safe procedure. Nevertheless, aseptic loosening is still the main reason for revision after cemented total knee or cemented total hip arthroplasty. To investigate whether an additional carbon dioxide lavage after a high-pressure pulsatile saline lavage has a positive effect on the bone–cement interface or cement penetration, we set up a standardized laboratory experiment with 28 human femoral heads. After a standardized cleaning procedure, the test implants were cemented onto the cancellous bone. Subsequently, the maximum failure load of the bone–cement interface was determined using a material testing machine to pull off the implant, and the cement penetration was determined using computed tomography. Neither the maximum failure load nor cement penetration into the cancellous bone revealed significant differences between the groups. In conclusion, according to our experiments, the additive use of the carbon dioxide lavage after the high-pressure pulsatile lavage has no additional benefit for the cleaning of the cancellous bone and, therefore, cannot be recommended without restrictions.

1. Introduction

During the last few decades, cementless fixation in total hip arthroplasty in Germany increased [1]. This is most likely the result of very good long-term outcomes [2,3,4]. Nevertheless, cemented endoprosthesis is still important in patients with osteoporosis, complex anatomy, hemiarthroplasty, or revision cases. In total knee arthroplasties, more than 90% are still implanted using bone cement. Excellent long-term results can be achieved using cemented implant fixation in total joint arthroplasty [5,6,7,8]. National registries demonstrate slight differences in the surgical technique for total knee arthroplasty. In the Netherlands, New Zealand, and Germany, 95% or more of the total knee arthroplasties (TKAs) are performed using the cemented technique [9,10,11] compared with only 92% in Sweden in 2019 [12]. In total hip arthroplasty (THA), those differences are even more obvious. In Sweden, a total of 76% of THAs are cemented, whereas, in Denmark, only 29% are implanted in this way [13]. Even though there are regional differences, the implantation procedures and technique are standardized. High-pressure pulsatile saline lavage irrigation, drying the bone bed, drilling holes in the tibia, and applying vacuum-mixed-cement to the implant and bone are mandatory steps in the implanting of TKAs [14]. The third generation cementing technique is performed in THA. The necessary steps are: a properly prepared bone bed through aggressive rasping, using high-pressure pulsatile saline lavage irrigation, using a distal cement restrictor, applying vacuum-mixed-cement in a retrograde technique into the femur via a cement gun, pressurizing the cement, and inserting the stem with a distal centralizer [15]. However, even though the cementing technique has evolved over the decades, aseptic loosening is still the main reason for the revision after the cemented TKA and THA [9,16,17].
The adequate cleaning of the cancellous bone prior to the cement application increases the implant stability. In bone with reduced bone density, a pulsed lavage can have a protective effect on the implant stability [18]. The integrity of the bone–cement interface can even be reduced by the presence of blood between the bone and cement [19]. Knowing this, the removal of bone marrow containing residual bone material, blood, and fat is obligatory. The depth of penetration of the cement and bone–cement interface is directly related to prior cleaning. Furthermore, 3–5 mm is the desired cement thickness for stable anchorage [20,21]. Not only does polymerizing cement not have the harmful thermal effects described for a cement thickness of more than 5 mm [21,22]; more importantly, the implant failure rates are significantly lower compared with a cement thickness of less than 2 mm [23]. In THA, the use of a high-pressure pulsatile lavage increased from 13% to 78% between 1998 and 2010 in Germany [24]. This shows that using a high-pressure pulsatile lavage quickly established itself as a standard application in the field of cemented arthroplasty.
A well observed and described procedure of cleaning the cancellous bone is using a high-pressure pulsatile lavage and drying the bone bed with suction and an abdominal cloth. However, fat or blood may remain, which prevents the cement from penetrating the cancellous bone. Using a pulsatile lavage has been shown to be more efficient than manual rinse cleaning by a bladder syringe alone [25,26,27,28,29]. Using a high-pressure pulsatile lavage before cementing showed less radiolucent lines in TKA compared to the reference group [30]. This could also be shown in unicompartmental knee arthroplasty (UKA), where the pulsed lavage group only showed radiolucent lines in 4% of all four of the zones compared to the reference group with 22% [25]. A high-pressure carbon dioxide lavage is available for better cleaning of the cancellous bone and simultaneous drying of the bone [31]. It is recommended to be additionally used after conducting a high-pressure pulsatile saline lavage. In 2019, Gapinski et al. used sterile compressed CO2 in addition to a pulsatile lavage in 303 total knee endoprostheses. According to the Knee Society Radiographic Evaluation System, a significant deeper cleaning could be shown in three of the seven zones [32]. A significant deeper bone cement penetration could be shown in a cadaveric study using a combination of pressurized carbon dioxide and common pulsatile saline lavage [33]. In addition, in an experimental study using human radii, the bone–cement interface was stronger when the syringe lavage was compared to the carbon dioxide lavage [34]. In a standardized laboratory setup using a standardized carbon foam model, a deeper cleaning by using a compressed carbon dioxide lavage in addition to a saline high-pressure pulsatile lavage was shown. Nevertheless, the effect was not significant [35]. In contrast, embolic events were reported using a carbon dioxide lavage in intramedullary cleaning during cemented THA [36,37].
The aim of this experimental cadaver study was to investigate two different methods of bone cleaning and their effect on cement penetration and bone–cement interface stability. For this purpose, a conventional pulsatile lavage was compared with a combination of a pulsatile lavage and an additional compressed carbon dioxide lavage.

2. Materials and Methods

The study was divided into two separate investigations. The cement penetration was analyzed in the first part, and the bone–cement interface stability was analyzed in the second part. The study was conducted according to the guidelines of the Declaration of Helsinki and approved by the local ethics committee.

2.1. Test Setup and Preparations

We investigated implant stability and cement penetration in 14 pairs of fresh frozen human femoral heads (28 in total). Femoral heads provided sufficient cancellous bone and made the test setup independent of different implant designs in TKA and THA. Seven pairs each were randomly assigned to Groups 1 (maximum failure load) and 2 (cement penetration). The specimens were randomly allocated with regard to right and left hip to Group A (high-pressure pulsatile lavage) or Group B (high-pressure pulsatile lavage with additional carbon dioxide lavage) by means of a computer-generated list (Randlist 1.2; Datinf GmbH, Tübingen, Germany) (Figure 1). In Group 1, the mean donor data showed an age of 74 ± 12.5 years, a height of 179 ± 5.9 cm, a weight of 107.3 ± 22.1 kg, and a body mass index of 33.2 ± 5.6 kg/m2. All donors in Group 1 were male. In Group 2, the mean donor data showed an age of 66.6 ± 18.3 years, a height of 170 ± 9.6 cm, a weight of 92.2 ± 37.2 kg, and a body mass index of 31.2 ± 11.4 kg/m2. Four of the donors in Group 2 were male and three were female. The bone mineral density was evaluated using standard hip parameters for dual-energy X-ray absorption (DXA) (Hologic QDR-2000, Marlborough, MA, USA). This allowed us to compare the groups and to assess bone quality.
The specimens were casted in a mold using synthetic resin (Rencast FC 53, Huntsman Advanced Materials GmbH, Bergkamen, Germany) in order to be able to integrate them into the testing setup. All femoral heads were trimmed slightly above the equator to remove the fovea capitis femoris and to preserve the maximum volume of cancellous bone (Figure 2). In the right–left comparison, the identical resection heights were selected. By measuring the sliced fragments, the adherence to the resection heights was checked again. Since the cement–bone interface was investigated in this experimental study, we manufactured aluminum test implants as part of the test standardization following the peg design of the femoral component in TKA (Figure 3). In addition, a central hole for the peg with a diameter of 9 mm and a depth of 9 mm was drilled in the center of the fixed and resected femoral head (Figure 2). The hole diameter in relation to the peg diameter was chosen to allow a homogeneous cement mantle and to avoid a press fit.

2.2. Cancellous Bone Cleaning

The paired heads were divided into Group 1 (determination of the maximum failure load) and Group 2 (determination of cement penetration). This resulted in Group 1-A and Group 1-B to determine the maximum failure load and in Group 2-A and 2-B to determine the cement penetration of the different cancellous bone cleaning techniques, as described in the following section (Figure 1).
In Group 1-A and 2-A, femoral heads were cleaned by high-pressure pulsatile saline lavage and dried by a compress. High-pressure pulsatile saline lavage was performed using the InterPulse with its bone cleaning tip (REF 0210-010-00; Stryker, Kalamazoo, MI, USA) (Figure 4 left). A flushing volume of one liter was used. The distance to the surface was determined by the splash-shield of the bone cleaning tip. While flushing, the lavage was continuously moved over the surface of the specimen.
In Group 1-B and 2-B, after cleaning the cancellous bone by high-pressure pulsatile saline lavage, as done in Group 1-A and 2-A, pressurized carbon dioxide lavage was performed additionally using the device CarboJet (Kinamed, Camarillo, CA, USA) with the supplied wide-angle-knee-nozzle (Figure 4 right) to dry the bone and to test for additional cleaning effects. According to manufacturer’s information, we performed this procedure for exactly 30 s with the nozzle slightly touching the cancellous bone in an angle of 45° to the plan surface of the cancellous bone.

2.3. Cementing Technique and Implantation of Test Implants

Immediately after cleaning and drying of the cancellous bone, the cementation process was started. The bone cement (PALACOS R+G, Heraeus, Hanau, Germany) was mixed under vacuum (Optivac®, Zimmer Biomet Holdings, Warsaw, IN, USA) and applied with a cement gun. Early cement timing for vacuum mixed cement at a room temperature of 25 ± 1 °C was used. We applied the still sticky bone cement to the undersurface of the implant 50 s after the start of the mixing process. In the next step, the cement was applied to the bone at 90 s using the cement gun. The impaction of the implant was performed 150 s after the start of the mixing process. After impaction with manual compression, the ligament tension force was applied 180 s after the start of the mixing by a clamping system with fine thread (Figure 5). The ligament tension force was 180 N, as reported by Clarius et al., for unicompartmental knee arthroplasty [38,39]. The applied force was controlled during cement curing by means of a force measurement system. After reaching the end of the final curing phase (10 min after start of mixing), the ligament tension force was removed. The described procedure of cementation was performed identically in all used specimens regardless of which group they belonged to.

2.4. Determination of Maximum Failure Load

In accordance with ISO 5833, the maximum load to failure of the bone–cement interface was determined 24 ± 2 h after cementation using a pull-out test. The specimens were stored in a refrigerator at +4 °C and kept at room temperature for 2 h before the start of the experiment. The test was performed for Group 1-A and 1-B using a material testing machine (Zwick Roell, Z005, Ulm, Germany) under displacement control. The implant was pulled out at a traverse speed of 2 mm/min. The resulting force was recorded, and the maximum force values were defined as failure load (Figure 6).

2.5. Cement Analysis/Cement Penetration

For analyzing cement penetration, we used an implant design without undercuts (Figure 1, bottom right). In addition, we coated the Group 2-A and 2-B implants with a wax-based release agent spray (goessl pfaff®, Karlskron, Germany). This enabled us to remove the implant after complete hardening of the bone cement without causing defects to the cement mantle. Subsequently, we obtained CT images (SOMATOM Emotion, Siemens Healthcare GmbH, Erlangen, Germany) of all 14 femoral heads of Group 2-A and 2-B with a slice thickness of 0.75 mm without metal artifacts. In the first step, the cement was segmented from the CT scans using the ITK-SNAP software [40]. Then, the cement volume and cement penetration were analyzed with the Geomagic Studio Program (Raindrop Geomagic, NC, USA). In addition, with regard to cement penetration depth, four different points in the plane and two further points under the peg were measured. Only the cement that penetrated the bone was analyzed (Figure 7). The cement outside the test implants and on top of the surface was not included in the calculation as it did not penetrate the cancellous structure.

2.6. Statistics

Prior to the start of the experimental study, a sample size calculation was performed using G*Power 3.1 (Kiel University, Kiel, Germany) [41]. This was based on the reported data by Kalteis et al. [29]. Input parameters to compute the required sample size were tails: two, effect size d: 2.031, α err prob: 0.05, and power (1-β err prob): 0.95. This results in the output parameters sample size 6 for each group and an actual power of 0.97. The data were evaluated descriptively using the arithmetic mean standard deviation, minimum, and maximum. Pre-analysis, the normal distribution of the data was evaluated using a Shapiro–Wilk-test, and the homogeneity of variance was verified using the Levene test. We conducted a two tailed t-test for independent samples to assess effects between the two groups on the parameters of bone mineral density, maximum failure load, cement penetration, and cement volume. All data were analyzed using SPSS V.25 (IBM Corp. Released 2017. IBM SPSS Statistics for Windows, Version 25.0. IBM Corp., Armonk, NY, USA) with a significance level of p < 0.05.

3. Results

3.1. Bone Mineral Density

The t-test revealed no statistically significant difference in the bone mineral density between Group 1-A 0.758 ± 0.10 g/cm2 and Group 1-B 0.751 ± 0.12 g/cm2, t(6) = 0.467, p = 0.657, (95% CI [−0.028, 0.041]). It also showed no statistically significant difference between Group 2-A 0.998 ± 0.21 g/cm2 and Group 2-B 0.976 ± 0.24 g/cm2, t(5) = −1.575, p = 0.176, (95% CI [−0.093, 0.022]).

3.2. Group 1-Bond Strength of the Bone–Cement Interface

The t-test revealed no statistically significant difference for the additional use of a carbon dioxide lavage prior to cement application. The needed force to pull off the implant was 3.7 ± 0.9 N/mm2 in Group 1-A (saline lavage only) and 4.0 ± 0.5 N/mm2 in Group 1-B (saline lavage + carbon dioxide lavage). This difference of 0.3 N/mm2 showed no significance −t(6) = −1.029, p = 0.343 (95% CI [−1.206, 0.492]) (Figure 8).

3.3. Group 2-Cement Penetration

The t-test also revealed no statistically significant difference for the additional use of a carbon dioxide lavage prior to cement application in terms of cement penetration. In Group 2-A, in which only a saline high-pressure lavage was used, the mean cement volume that penetrated into the cancellous bone was 1526.5 ± 217 mm3 according to the 3D analysis via itk-Snap [40]. In Group 2-B, where an additional carbon dioxide lavage was used, the mean cement volume was 1431.1 ± 963 mm3. The difference of 95.4 mm3 was not significant −t(6) = 1.815, p = 0.119 (95% CI [−33.19, 223.94]) (Figure 9).
No significant difference could be found with regard to the depth measurement carried out at all six representative locations (four measurement points in the plane area t(6) = 0.91, p = 0.40 (95% CI [−0.314, 0.685]; two measurement points beneath the peg t(6) = −1.04, p = 0.34 (95% CI [−1.729, 0.700]) (Figure 7, Figure 10 and Figure 11).

4. Discussion

Since cementless THA has very good long-term results, the number of non-cemented total joint replacements rose during the last few decades [1,2,3,4]. In Germany, for example, 77.6% of the primary THAs were implanted without cement in 2020 [42]. Nevertheless, cemented endoprosthesis is still important in patients with osteoporosis, complex anatomy, hemiarthroplasty, or revision cases. In Germany in 2020, less than 2% of all the TKAs were implanted using no bone cement [42]. During the last three decades, cemented arthroplasty proved to be an excellent procedure in total joint replacement. Additionally, even though there are regional differences in whether a cemented or non-cemented technique is performed, a worldwide state of the art cementing technique has been established [14,15]. The main part of the third generation cementing technique is the adequate cleaning of the cancellous bone via a pulsatile saline high-pressure lavage. This crucial step in the process of cementing improves the interdigitation and sustainability of the bone–cement interface [18]. In a standardized laboratory setup, we compared two different lavage procedures in 28 paired human femoral heads. Half of them were cleaned by a high-pressure pulsatile lavage and a compress. For the other half, a high-pressure pulsatile lavage and additional carbon dioxide lavage was used. In two separate experiments, we tested for the bond-strength of the bone–cement interface and cement penetration in the cancellous bone. Our data showed no significant difference between the different cleaning procedures. Since even the presence of blood between the bone and cement can reduce the integrity of the bone–cement interface [19], we would have expected higher bond-strength in the group that has been edited additionally by a carbon dioxide lavage (Group 1-B) and not only by a compress (Group 1-A). In addition, knowing that cleaning is directly related to the cement penetration into the cancellous bone, we would have also expected higher volumes of cement penetration in the carbon dioxide cleaned group (Group 2-B), but there was also no significant difference to the group that had been dried only by a compress (Group 2-A). In our standardized laboratory setup, we used frozen human femoral heads, which, after defrosting, contained the same bone marrow and blood as live bones. Since a carbon dioxide lavage is recommended to be used after a saline pulsatile lavage, our data suggest that this procedure is not better than using a compress. Perhaps, fluids that remain in the cancellous bone after the lavage are not that obstructive for bone cement to penetrate into the bone [19]. It is even possible that the compress dried as much or more of the flushing medium than the carbon dioxide lavage was able to do. An experimental study using eight human radii showed that, in using a syringed saline or carbon dioxide lavage cleaning, there was a significantly stronger bone–cement interface in the four radii that were prepared by carbon dioxide lavage alone [34]. Since this study was not performed according to the manufacturer’s information in which a carbon dioxide lavage is recommended to be added to the high-pressure pulsatile lavage and a relatively low case number was used, the obtained data of this experiment cannot be applied to our study.
In addition, our data suggest that a carbon dioxide lavage had no additional cleaning effect, i.e., no bone marrow or remnants were harvested that had not been purified by a saline lavage. Knowing that using a system that requires additional devices and materials always comes at a financial cost, it seems debatable. Even additional waste must be considered if orthopedic surgery is to be more sustainable [43,44,45]. The usage of an additional device also takes time. Additional operation time is related to a higher possibility of infections and postoperative complications [46,47,48,49,50]. Therefore, an extensive duration of surgeries has to be avoided if possible. Since our greatest good is patient safety and knowing about reported cases of embolisms using a carbon dioxide lavage, the added value of this procedure has to be carefully tested [36,37]. Using a simplified cadaver laboratory setup, it is uncertain how the different procedures perform in humans if used in different anatomical regions in cemented arthroplasty. We used the cancellous structure of paired femoral heads to investigate the bone–cement interface and to determine the cement penetration. The orientation of the trabeculae differs from other regions. However, the paired femoral heads allowed us a high degree of standardization. Stability in the cemented joint replacement is achieved via the interdigitation of the bone cement into the cancellous bone. Since this is in contrast with uncemented arthroplasty, in which stability is gained through a process of growing bone to the prosthesis called osteoconduction [51,52,53,54], this should not be a significant factor.
In conclusion, the additive use of a carbon dioxide lavage after a high-pressure pulsatile saline lavage has no additional benefit in cleaning the cancellous bone in a laboratory setup and, therefore, cannot be recommended without restrictions.

Author Contributions

Conceptualization, K.K. and S.J.; Data curation, K.K., M.S., C.S. and S.J.; Formal analysis, K.K., S.J. and C.S.; Methodology, K.K. and S.J.; Project administration, K.K., S.J. and M.M.I.; Writing—original draft, K.K. and S.J.; Writing—review & editing, K.K., C.S, M.M.I., M.S., T.G., T.R. and S.J. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

The study was conducted according to the guidelines of the Declaration of Helsinki and approved by the Ethics Committee of Heidelberg University (protocol code: S-013/2021, date of approval: 4 February 2021).

Informed Consent Statement

Not applicable.

Data Availability Statement

The data presented in this study are available within this article.

Conflicts of Interest

The authors declare no pertinent conflict of interest. S.J. reports grants from B Braun Aesculap, Johnson & Johnson Depuy Synthes, Heraeus Medical, Waldemar Link, Peter Brehm and Zimmer Biomet that are not related to the current study. T.G. reports personal fees paid to T.G. during the conduct of the study from Zimmer Biomet, Europe, and from Depuy Synthes Orthopaedie GmbH, Peter Brehm GmbH, ImplanTec GmbH outside the submitted work. Research grants paid to the institution during the conduct of the study from Zimmer Biomet, Europe, Mathys AG Switzerland, Anika Therapeutics outside the submitted work that are not related to the current study. T.R. reports research funding/travel expenses and/or paid speaking engagements by the German Federal Ministry of Education and Research (BMBF), The German Federal Ministry for Economic Cooperation and Development (BMZ), Otto Bock Foundation, Stiftung Oskar-Helene-Heim Berlin, DePuy Int, Zimmer, Aesculap/B. Braun, the Vielberth Foundation, German Society of Orthopedics and Traumatology (DGOU), and the German Association of Orthopedics and Orthopedic Surgery (DGOOC). T.R. is associate editor of “Der Orthopäde” and “Der Unfallchirurg” (Springer Heidelberg, Berlin, New York) and member of the International Advisory Board of the Journal of the American Academy of Orthopedic Surgeons (AAOS).

References

  1. Klug, A.; Gramlich, Y.; Hoffmann, R.; Pfeil, J.; Drees, P.; Kutzner, K.P. Trends in Total Hip Arthroplasty in Germany from 2007 to 2016: What Has Changed and Where Are We Now? Z. Orthop. Unf. 2021, 159, 173–180. [Google Scholar] [CrossRef]
  2. Kim, Y.H.; Park, J.W.; Jang, Y.S. Long-Term Survival (up to 34 Years) of Retained Cementless Anatomic Femoral Stem in Patients < 50 Years Old. J. Arthroplast. 2021, 36, 1388–1392. [Google Scholar] [CrossRef]
  3. Schmolders, J.; Amvrazis, G.; Pennekamp, P.H.; Strauss, A.C.; Friedrich, M.J.; Wimmer, M.D.; Rommelspacher, Y.; Wirtz, D.C.; Wallny, T. Thirteen year follow-up of a cementless femoral stem and a threaded acetabular cup in patients younger than fifty years of age. Int. Orthop. 2017, 41, 39–45. [Google Scholar] [CrossRef]
  4. Kawamura, H.; Mishima, H.; Sugaya, H.; Nishino, T.; Shimizu, Y.; Miyakawa, S. The 21- to 27-year results of the Harris-Galante cementless total hip arthroplasty. J. Orthop. Sci. 2016, 21, 342–347. [Google Scholar] [CrossRef] [PubMed]
  5. Dalury, D.F. Cementless total knee arthroplasty: Current concepts review. Bone Jt. J. 2016, 98, 867–873. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  6. Ritter, M.A.; Keating, E.M.; Sueyoshi, T.; Davis, K.E.; Barrington, J.W.; Emerson, R.H. Twenty-Five-Years and Greater, Results After Nonmodular Cemented Total Knee Arthroplasty. J. Arthroplast. 2016, 31, 2199–2202. [Google Scholar] [CrossRef] [PubMed]
  7. Herberts, P.; Malchau, H. Long-term registration has improved the quality of hip replacement: A review of the Swedish THR Register comparing 160,000 cases. Acta Orthop. Scand. 2000, 71, 111–121. [Google Scholar] [CrossRef] [Green Version]
  8. Malchau, H.; Herberts, P.; Eisler, T.; Garellick, G.; Soderman, P. The Swedish Total Hip Replacement Register. J. Bone Jt. Surg. Am. 2002, 84, 2–20. [Google Scholar] [CrossRef]
  9. German Arthroplasty Registry (EPRD)-2019 Annual Report. 2019. Available online: https://www.eprd.de/fileadmin/user_upload/Dateien/Publikationen/Berichte/EPRD_Jahresbericht_2019_EN_doppelseitig_F_Web-Stand-251120.pdf (accessed on 17 November 2021). [CrossRef]
  10. The New Zealand Joint Registry—Fifteen-year report January 1999 to December 2013. 2014, Volume 15. Available online: https://www.nzoa.org.nz/sites/default/files/NZJR2014Report.pdf (accessed on 17 November 2021).
  11. Gademan, M.G.J.; Van Steenbergen, L.N.; Cannegieter, S.C.; Nelissen, R.; Marang-Van De Mheen, P.J. Population-based 10-year cumulative revision risks after hip and knee arthroplasty for osteoarthritis to inform patients in clinical practice: A competing risk analysis from the Dutch Arthroplasty Register. Acta Orthop. 2021, 92, 1–5. [Google Scholar] [CrossRef] [PubMed]
  12. The Swedish Knee Arthroplasty Register—Anual Report 2020. 2020. Available online: http://myknee.se/pdf/SVK_2020_Eng_1.0.pdf (accessed on 17 November 2021).
  13. Bunyoz, K.I.; Malchau, E.; Malchau, H.; Troelsen, A. Has the Use of Fixation Techniques in THA Changed in This Decade? The Uncemented Paradox Revisited. Clin. Orthop. Relat. Res. 2020, 478, 697–704. [Google Scholar] [CrossRef]
  14. Refsum, A.M.; Nguyen, U.V.; Gjertsen, J.E.; Espehaug, B.; Fenstad, A.M.; Lein, R.K.; Ellison, P.; Hol, P.J.; Furnes, O. Cementing technique for primary knee arthroplasty: A scoping review. Acta Orthop. 2019, 90, 582–589. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  15. Niculescu, M.; Solomon, B.L.; Viscopoleanu, G.; Antoniac, I.V. Evolution of Cementation Techniques and Bone Cements in Hip Arthroplasty. Handb. Bioceram. Biocompos. 2016, 859–899. [Google Scholar] [CrossRef]
  16. Khan, M.; Osman, K.; Green, G.; Haddad, F.S. The epidemiology of failure in total knee arthroplasty: Avoiding your next revision. Bone Jt. J. 2016, 98, 105–112. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  17. Kochbati, R.; Rbai, H.; Jlailia, M.; Makhlouf, H.; Bouguira, A.; Daghfous, M.S. Predictive factors of aseptic loosening of cemented total hip prostheses. Pan Afr. Med. J. 2016, 24, 260. [Google Scholar] [CrossRef]
  18. Jaeger, S.; Rieger, J.S.; Bruckner, T.; Kretzer, J.P.; Clarius, M.; Bitsch, R.G. The protective effect of pulsed lavage against implant subsidence and micromotion for cemented tibial unicompartmental knee components: An experimental cadaver study. J. Arthroplast. 2014, 29, 727–732. [Google Scholar] [CrossRef]
  19. Benjamin, J.B.; Gie, G.A.; Lee, A.J.; Ling, R.S.; Volz, R.G. Cementing technique and the effects of bleeding. J. Bone Jt. Surg. Br. 1987, 69, 620–624. [Google Scholar] [CrossRef]
  20. Vanlommel, J.; Luyckx, J.P.; Labey, L.; Innocenti, B.; De Corte, R.; Bellemans, J. Cementing the tibial component in total knee arthroplasty: Which technique is the best? J. Arthroplast. 2011, 26, 492–496. [Google Scholar] [CrossRef]
  21. Cawley, D.T.; Kelly, N.; McGarry, J.P.; Shannon, F.J. Cementing techniques for the tibial component in primary total knee replacement. Bone Jt. J. 2013, 95, 295–300. [Google Scholar] [CrossRef]
  22. Sih, G.C.; Connelly, G.M.; Berman, A.T. The effect of thickness and pressure on the curing of PMMA bone cement for the total hip joint replacement. J. Biomech. 1980, 13, 347–352. [Google Scholar] [CrossRef]
  23. Hampton, C.B.; Berliner, Z.P.; Nguyen, J.T.; Mendez, L.; Smith, S.S.; Joseph, A.D.; Padgett, D.E.; Rodriguez, J.A. Aseptic Loosening at the Tibia in Total Knee Arthroplasty: A Function of Cement Mantle Quality? J. Arthroplast. 2020, 35, S190–S196. [Google Scholar] [CrossRef] [PubMed]
  24. Fischer, C.A.; Kaszap, B.; Drexler, C.; Lehner, B.; Clarius, M. Cemented total hip arthroplasty in Germany—Update 2010. Z Orthop. Unf. 2012, 150, 309–317. [Google Scholar] [CrossRef] [PubMed]
  25. Clarius, M.; Hauck, C.; Seeger, J.B.; James, A.; Murray, D.W.; Aldinger, P.R. Pulsed lavage reduces the incidence of radiolucent lines under the tibial tray of Oxford unicompartmental knee arthroplasty: Pulsed lavage versus syringe lavage. Int. Orthop. 2009, 33, 1585–1590. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  26. Breusch, S.J.; Norman, T.L.; Schneider, U.; Reitzel, T.; Blaha, J.D.; Lukoschek, M. Lavage technique in total hip arthroplasty: Jet lavage produces better cement penetration than syringe lavage in the proximal femur. J. Arthroplast. 2000, 15, 921–927. [Google Scholar] [CrossRef] [PubMed]
  27. Seeger, J.B.; Jaeger, S.; Bitsch, R.G.; Mohr, G.; Rohner, E.; Clarius, M. The effect of bone lavage on femoral cement penetration and interface temperature during Oxford unicompartmental knee arthroplasty with cement. J. Bone Jt. Surg. Am. 2013, 95, 48–53. [Google Scholar] [CrossRef] [PubMed]
  28. Helwig, P.; Konstantinidis, L.; Hirschmuller, A.; Miltenberger, V.; Kuminack, K.; Sudkamp, N.P.; Hauschild, O. Tibial cleaning method for cemented total knee arthroplasty: An experimental study. Indian J. Orthop. 2013, 47, 18–22. [Google Scholar] [CrossRef]
  29. Kalteis, T.; Pforringer, D.; Herold, T.; Handel, M.; Renkawitz, T.; Plitz, W. An experimental comparison of different devices for pulsatile high-pressure lavage and their relevance to cement intrusion into cancellous bone. Arch. Orthop. Trauma Surg. 2007, 127, 873–877. [Google Scholar] [CrossRef] [PubMed]
  30. Ritter, M.A.; Herbst, S.A.; Keating, E.M.; Faris, P.M. Radiolucency at the bone-cement interface in total knee replacement. The effects of bone-surface preparation and cement technique. J. Bone Jt. Surg. Am. 1994, 76, 60–65. [Google Scholar] [CrossRef]
  31. Goldstein, W.M.; Gordon, A.; Goldstein, J.M.; Berland, K.; Branson, J.; Sarin, V.K. Improfement of Cement mantle thickness with pressurized carbon dioxide lavage. In Proceedings of the 20th Annual Meeting of the International Society for Technology in Arthroplasty, Paris, France, 3–6 October 2007. [Google Scholar]
  32. Gapinski, Z.A.; Yee, E.J.; Kraus, K.R.; Deckard, E.R.; Meneghini, R.M. The Effect of Tourniquet Use and Sterile Carbon Dioxide Gas Bone Preparation on Cement Penetration in Primary Total Knee Arthroplasty. J. Arthroplast. 2019, 34, 1634–1639. [Google Scholar] [CrossRef]
  33. Boontanapibul, K.; Ruangsomboon, P.; Charoencholvanich, K.; Pornrattanamaneewong, C. Effectiveness Testing of Combined Innovative Pressurized Carbon Dioxide Lavage and Pulsatile Normal Saline Irrigation to Enhance Bone Cement Penetration in Total Knee Replacement: A Cadaveric Study. J. Med. Assoc. Thai. 2016, 99, 1198–1202. [Google Scholar]
  34. Ravenscroft, M.J.; Charalambous, C.P.; Mills, S.P.; Woodruff, M.J.; Stanley, J.K. Bone-cement interface strength in distal radii using two medullary canal preparation techniques: Carbon dioxide jet cleaning versus syringed saline. Hand Surg. 2010, 15, 95–98. [Google Scholar] [CrossRef]
  35. Knappe, K.; Stadler, C.; Innmann, M.; Schonhoff, M.; Gotterbarm, T.; Renkawitz, T.; Jaeger, S. Does Pressurized Carbon Dioxide Lavage Improve Bone Cleaning in Cemented Arthroplasty? Appl. Sci. 2021, 11, 6103. [Google Scholar] [CrossRef]
  36. Martinez, M.R.G.; Ruiz, G.M.V.; Perez, J.M.R.; Perez, R.L. Air embolism after cleaning the intramedullary canal with a high pressure CO2 wash in hip fracture. Rev. Esp. Anestesiol. Reanim. 2013, 60, 353–354. [Google Scholar] [CrossRef]
  37. Lax-Prez, R.; Ferrero-Manzanal, F.; Marin-Penia, O.; Murcia-Asensio, A. Gas embolism after the use of the CarboJet lavage system during hip hemiarthroplasty. Clinical case and literature review. Acta Ortop. Mex. 2014, 28, 45–48. [Google Scholar]
  38. Clarius, M.; Seeger, J.B.; Jaeger, S.; Mohr, G.; Bitsch, R.G. The importance of pulsed lavage on interface temperature and ligament tension force in cemented unicompartmental knee arthroplasty. Clin. Biomech. 2012, 27, 372–376. [Google Scholar] [CrossRef] [PubMed]
  39. Schwarze, M.; Schonhoff, M.; Beckmann, N.A.; Eckert, J.A.; Bitsch, R.G.; Jager, S. Femoral Cementation in Knee Arthroplasty—A Comparison of Three Cementing Techniques in a Sawbone Model Using the ATTUNE Knee. J. Knee Surg. 2021, 34, 258–266. [Google Scholar] [CrossRef] [PubMed]
  40. Yushkevich, P.A.; Piven, J.; Hazlett, H.C.; Smith, R.G.; Ho, S.; Gee, J.C.; Gerig, G. User-guided 3D active contour segmentation of anatomical structures: Significantly improved efficiency and reliability. Neuroimage 2006, 31, 1116–1128. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  41. Faul, F.; Erdfelder, E.; Lang, A.G.; Buchner, A. G * Power 3: A flexible statistical power analysis program for the social, behavioral, and biomedical sciences. Behav. Res. Methods 2007, 39, 175–191. [Google Scholar] [CrossRef]
  42. German Arthroplasty Registry (EPRD)—2021 Annual Report. 2021. Available online: https://www.eprd.de/fileadmin/user_upload/Dateien/Publikationen/Berichte/Jahresbericht2021_2021-10-25_F.pdf (accessed on 17 November 2021). [CrossRef]
  43. Watts, N.; Adger, W.N.; Ayeb-Karlsson, S.; Bai, Y.; Byass, P.; Campbell-Lendrum, D.; Colbourn, T.; Cox, P.; Davies, M.; Depledge, M.; et al. The Lancet Countdown: Tracking progress on health and climate change. Lancet 2017, 389, 1151–1164. [Google Scholar] [CrossRef] [Green Version]
  44. Thiel, C.L.; Eckelman, M.; Guido, R.; Huddleston, M.; Landis, A.E.; Sherman, J.; Shrake, S.O.; Copley-Woods, N.; Bilec, M.M. Environmental impacts of surgical procedures: Life cycle assessment of hysterectomy in the United States. Environ. Sci. Technol. 2015, 49, 1779–1786. [Google Scholar] [CrossRef]
  45. Rizan, C.; Steinbach, I.; Nicholson, R.; Lillywhite, R.; Reed, M.; Bhutta, M.F. The Carbon Footprint of Surgical Operations: A Systematic Review. Ann. Surg. 2020, 272, 986–995. [Google Scholar] [CrossRef]
  46. Hanada, M.; Hotta, K.; Furuhashi, H.; Matsuyama, Y. Intraoperative bacterial contamination in total hip and knee arthroplasty is associated with operative duration and peeling of the iodine-containing drape from skin. Eur. J. Orthop. Surg. Traumatol. 2020, 30, 917–921. [Google Scholar] [CrossRef] [PubMed]
  47. Ravi, B.; Jenkinson, R.; O’Heireamhoin, S.; Austin, P.C.; Aktar, S.; Leroux, T.S.; Paterson, M.; Redelmeier, D.A. Surgical duration is associated with an increased risk of periprosthetic infection following total knee arthroplasty: A population-based retrospective cohort study. EClinicalMedicine 2019, 16, 74–80. [Google Scholar] [CrossRef] [PubMed]
  48. Teo, B.J.X.; Yeo, W.; Chong, H.C.; Tan, A.H.C. Surgical site infection after primary total knee arthroplasty is associated with a longer duration of surgery. J. Orthop. Surg. (Hong Kong) 2018, 26, 2309499018785647. [Google Scholar] [CrossRef] [Green Version]
  49. Held, M.B.; Boddapati, V.; Sarpong, N.O.; Cooper, H.J.; Shah, R.P.; Geller, J.A. Operative Duration and Short-Term Postoperative Complications after Unicompartmental Knee Arthroplasty. J. Arthroplast. 2021, 36, 905–909. [Google Scholar] [CrossRef] [PubMed]
  50. Orland, M.D.; Lee, R.Y.; Naami, E.E.; Patetta, M.J.; Hussain, A.K.; Gonzalez, M.H. Surgical Duration Implicated in Major Postoperative Complications in Total Hip and Total Knee Arthroplasty: A Retrospective Cohort Study. J. Am. Acad. Orthop. Surg. Glob. Res. Rev. 2020, 4, e20 00043. [Google Scholar] [CrossRef] [PubMed]
  51. Zweymuller, K.A. Bony ongrowth on the surface of HA-coated femoral implants: An X-ray analysis. Z. Orthop. Unf. 2012, 150, 27–31. [Google Scholar] [CrossRef]
  52. Robertson, D.M.; Pierre, L.; Chahal, R. Preliminary observations of bone ingrowth into porous materials. J. Biomed. Mater. Res. 1976, 10, 335–344. [Google Scholar] [CrossRef]
  53. Albrektsson, T.; Johansson, C. Osteoinduction, osteoconduction and osseointegration. Eur. Spine J. 2001, 10, S96–S101. [Google Scholar] [CrossRef] [Green Version]
  54. Hu, C.; Ashok, D.; Nisbet, D.R.; Gautam, V. Bioinspired surface modification of orthopedic implants for bone tissue engineering. Biomaterials 2019, 219, 119366. [Google Scholar] [CrossRef]
Figure 1. Distribution of the 28 femoral heads to the different cleaning procedures and different test implants for either determination of maximum failure load or cement penetration.
Figure 1. Distribution of the 28 femoral heads to the different cleaning procedures and different test implants for either determination of maximum failure load or cement penetration.
Jcm 10 05361 g001
Figure 2. Embedded femoral head with drilled hole for the peg, prior to cleaning procedure.
Figure 2. Embedded femoral head with drilled hole for the peg, prior to cleaning procedure.
Jcm 10 05361 g002
Figure 3. Technical drawing in mm of the test implant that was used in Group 1 (with undercuts/ cement pocket).
Figure 3. Technical drawing in mm of the test implant that was used in Group 1 (with undercuts/ cement pocket).
Jcm 10 05361 g003
Figure 4. Pulsatile high-pressure saline lavage InterPulse with its bone cleaning tip (left) and carbon dioxide lavage CarboJet with its wide-angle-knee-nozzle (right).
Figure 4. Pulsatile high-pressure saline lavage InterPulse with its bone cleaning tip (left) and carbon dioxide lavage CarboJet with its wide-angle-knee-nozzle (right).
Jcm 10 05361 g004
Figure 5. Femoral head (bottom) with round test implant cemented to the bone surface and the force measurement system (top).
Figure 5. Femoral head (bottom) with round test implant cemented to the bone surface and the force measurement system (top).
Jcm 10 05361 g005
Figure 6. Already implanted test implants with cement after removal of all organic components.
Figure 6. Already implanted test implants with cement after removal of all organic components.
Jcm 10 05361 g006
Figure 7. Defined measurement points to determine the cement penetration in Group 2-A and 2-B.
Figure 7. Defined measurement points to determine the cement penetration in Group 2-A and 2-B.
Jcm 10 05361 g007
Figure 8. Force needed to pull off test implants (Group 1-A (left); 1-B (right)).
Figure 8. Force needed to pull off test implants (Group 1-A (left); 1-B (right)).
Jcm 10 05361 g008
Figure 9. Cement volume penetrating the cancellous bone (Group 2-A (left); 2-B (right)).
Figure 9. Cement volume penetrating the cancellous bone (Group 2-A (left); 2-B (right)).
Jcm 10 05361 g009
Figure 10. Mean cement penetration measurement point in the plane of cementation.
Figure 10. Mean cement penetration measurement point in the plane of cementation.
Jcm 10 05361 g010
Figure 11. Mean cement penetration measurement points beneath the peg.
Figure 11. Mean cement penetration measurement points beneath the peg.
Jcm 10 05361 g011
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Knappe, K.; Stadler, C.; Innmann, M.M.; Schonhoff, M.; Gotterbarm, T.; Renkawitz, T.; Jaeger, S. Does Additive Pressurized Carbon Dioxide Lavage Improve Cement Penetration and Bond Strength in Cemented Arthroplasty? J. Clin. Med. 2021, 10, 5361. https://doi.org/10.3390/jcm10225361

AMA Style

Knappe K, Stadler C, Innmann MM, Schonhoff M, Gotterbarm T, Renkawitz T, Jaeger S. Does Additive Pressurized Carbon Dioxide Lavage Improve Cement Penetration and Bond Strength in Cemented Arthroplasty? Journal of Clinical Medicine. 2021; 10(22):5361. https://doi.org/10.3390/jcm10225361

Chicago/Turabian Style

Knappe, Kevin, Christian Stadler, Moritz M. Innmann, Mareike Schonhoff, Tobias Gotterbarm, Tobias Renkawitz, and Sebastian Jaeger. 2021. "Does Additive Pressurized Carbon Dioxide Lavage Improve Cement Penetration and Bond Strength in Cemented Arthroplasty?" Journal of Clinical Medicine 10, no. 22: 5361. https://doi.org/10.3390/jcm10225361

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop