Next Article in Journal
Contribution of Pore-Connectivity to Permeation Performance of Silicalite-1 Membrane; Part II, Diffusivity of C6 Hydrocarbon in Micropore
Next Article in Special Issue
Electron Microscopic Confirmation of Anisotropic Pore Characteristics for ECMO Membranes Theoretically Validating the Risk of SARS-CoV-2 Permeation
Previous Article in Journal
Radiation-Grafted Anion-Exchange Membrane for Fuel Cell and Electrolyzer Applications: A Mini Review
Previous Article in Special Issue
Diabetic Pathophysiology Enhances Inflammation during Extracorporeal Membrane Oxygenation in a Rat Model
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Comparison of Circular and Parallel-Plated Membrane Lungs for Extracorporeal Carbon Dioxide Elimination

1
Department of Internal Medicine V–Pneumology, Allergology and Intensive Care Medicine, University Hospital of Saarland, 66424 Homburg, Germany
2
Department of Anaesthesiology and Critical Care, Campus Kassel of the University of Southampton, 34125 Kassel, Germany
3
Department of Internal Medicine I, University Hospital of Saarland, 66424 Homburg, Germany
*
Author to whom correspondence should be addressed.
These authors contributed equally. L.S.S. and A.M.J.
Membranes 2021, 11(6), 398; https://doi.org/10.3390/membranes11060398
Submission received: 19 April 2021 / Revised: 20 May 2021 / Accepted: 21 May 2021 / Published: 27 May 2021
(This article belongs to the Special Issue Advances in Extracorporeal Membrane Oxygenation)

Abstract

:
Extracorporeal carbon dioxide removal (ECCO2R) is an important technique to treat critical lung diseases such as exacerbated chronic obstructive pulmonary disease (COPD) and mild or moderate acute respiratory distress syndrome (ARDS). This study applies our previously presented ECCO2R mock circuit to compare the CO2 removal capacity of circular versus parallel-plated membrane lungs at different sweep gas flow rates (0.5, 2, 4, 6 L/min) and blood flow rates (0.3 L/min, 0.9 L/min). For both designs, two low-flow polypropylene membrane lungs (Medos Hilte 1000, Quadrox-i Neonatal) and two mid-flow polymethylpentene membrane lungs (Novalung Minilung, Quadrox-iD Pediatric) were compared. While the parallel-plated Quadrox-iD Pediatric achieved the overall highest CO2 removal rates under medium and high sweep gas flow rates, the two circular membrane lungs performed relatively better at the lowest gas flow rate of 0.5 L/min. The low-flow Hilite 1000, although overall better than the Quadrox i-Neonatal, had the most significant advantage at a gas flow of 0.5 L/min. Moreover, the circular Minilung, despite being significantly less efficient than the Quadrox-iD Pediatric at medium and high sweep gas flow rates, did not show a significantly worse CO2 removal rate at a gas flow of 0.5 L/min but rather a slight advantage. We suggest that circular membrane lungs have an advantage at low sweep gas flow rates due to reduced shunting as a result of their fiber orientation. Efficiency for such low gas flow scenarios might be relevant for possible future portable ECCO2R devices.
Keywords:
ECCO2R; COPD; ARDS

1. Introduction

Extracorporeal lung support is a promising tool to treat critical lung diseases such as exacerbated COPD, or mild or moderate ARDS, where gas exchange is restricted [1,2]. Two major methods: Extra corporeal membrane oxygenation (ECMO), used for oxygenation and CO2 removal of patients in hypoxic and hypercapnic respiratory failure and extracorporeal carbon dioxide removal (ECCO2R) only applied in hypercapnic respiratory failure to remove surplus carbon dioxide [3].
In lung failure, the main goal of applying those extracorporeal therapies is, to ensure a sufficient gas exchange and lung protective ventilation to prevent the lung from further damage by invasive ventilation. The Berlin criteria for ARDS state, that the therapy of ARDS requires low tidal volume ventilation with 6 mL/kg body weight and a plateau pressure of 30 mmHg or less [4,5]. Ultra-lung-protective ventilation that uses tidal volumes <6 mL/kg body weight and a plateau pressure of <30 mmHG, in combination with the application of ECCO2R, has been proposed by several studies to further improve the outcome [6,7]. The main advantage of the application of ECCO2R is the prevention of hypercapnia and respiratory acidosis induced by the reduced ventilation [2,3,6]. Carbon dioxide removal is mainly determined by the sweep gas flow rate and independent from high blood flow rates [3,8,9]. Therefore, in comparison to ECMO, smaller cannulas and membrane lungs with a lower surface area can be used, reducing the risk of bleeding and infections [3,9,10,11].
In general, for ECCO2R, there are mainly two predominant types of membrane lungs (MLs): Parallel-plated and circular MLs both promise a sufficient CO2 removal at low blood flow rates. In this work, we use an already established mock circuit to compare the two types of MLs and their efficiency on CO2 removal at low and ultralow blood flow rates [12].

2. Materials and Methods

2.1. Standard Protocol

In this work, we used our previously presented ECCO2R mock circuit to compare the CO2 removal capacity of circular and parallel-plated membrane lungs. The setup consisted of two circuits, the primary and the test circuit.
The primary circuit simulated the human vena cava and was responsible for creating a venous environment by CO2 enrichment and O2 depletion. In the test circuit, which simulated the ECCO2R device connected to the vena cava, the CO2 removal rate of four different membrane lungs was tested. The whole setup was filled with fresh porcine blood as test fluid. The primary circuit consisted of a Getinge Quadrox PLS (Permanent Life Support) membrane lung built into a loop circuit with a Rotaflow centrifugal pump and 3/8″polyvinyl chloride (PVC) tubings. To create a venous environment, N2 and CO2 were applied as sweep gases to the Quadrox PLS at rates of 7.5 L/min N2 and 0.55 L/min CO2 to create a CO2 partial pressure in blood pCO2(blood) between 43 mmHg ± 2 mmHg and a venous oxygen saturation of 65 ± 5% (data not shown). A blood flow of 5 L/min was chosen to operate the primary circuit according to the human cardiac output. The system was tempered to 37 °C with a Maquet HU 35 via the Quadrox PLS membrane lung.
Via Luer Lock connectors, the test circuit with the membrane lung (ML) to be tested was connected according to Figure 1. The blood flow through these Luer Lock connectors was called cannula flow in this work. Pure oxygen was applied as sweep gas to the test membrane lung. At the sweep gas outlet of the test ML, a mass flow sensor (TSI41403) and a side-flow capnometer (Philips Intellivue) were installed.
CO2 removal rates QCO2 (N = 6 for each group) were calculated from the sweep gas flow rate Qsweep at the sweep outlet, the partial pressure of CO2 in the sweep gas pCO2(gas) and the atmospheric pressure patm according to the following formula:
QCO2 = Qsweep ∗ pCO2(gas)/patm

2.2. Monitoring of the Test Fluid

The setup was filled with fresh porcine blood, added with 10.000 IE Heparin to avoid clotting and 1g Meropenem to avoid bacterial growth during the measurement. Blood gas analysis was performed to monitor the quality of the blood. Throughout the whole measurement series blood gas analysis confirmed that hemoglobin was constantly at 15 g/dl. PH was in range of 7.40 ± 0.20, bicarbonate in the rage from 23–26 mmol/L, venous saturation of oxygen in the range of 65% ± 5 %. Lactate values of up to 10 mmol/L at the beginning of the measurement were tolerated.

2.3. Comparison of Membrane Lungs

Of the four test membrane lungs, two were parallel-plated MLs (Maquet Quadrox-i Neonatal, Maquet Quadrox-iD Pediatric) and two were circular (Medos Hilite 1000, Novalung MiniLung).
In general, two sessions, measuring all 4 MLs, were performed. Each session consisted of 2 experiments (Figure 2). For each ML, blood flow (BF) rates were set at 0.3 L/min and 0.9 L/min. Moreover, for each blood flow rate, the sweep gas flow (Qsweep) was set at 6, 4, 2 and 0.5 L/min. In each session, a total of 6 repeated measurements were performed for each BF and Qsweep configuration. For each session, new MLs were chosen, blood samples were changed for each experiment. In session I, experiment I, the low-flow MLs, the Quadrox i-Neonatal with a surface area (SA) of 0.38 m² and the Hilite 1000 oxygenator with a surface area of 0.39 m², both consisting of polypropylene (PP) fibers, were compared. In session I, experiment II, the mid-flow MLs, Quadrox-iD Pediatric with a SA of 0.8 m² and the Novalung Minilung ML with a SA of 0.65 m², both consisting of polymethylpentene (PMP) fibers, were compared. In session II, performed on a separate day, the two experiments were repeated with new MLs. Each session had four measuring slots per experiment, as each ML was measured twice. The membrane lung that was tested in the first slot, would also be tested in the last slot. In session II, the measurements were repeated with new blood and MLs. The order was also modified, so that a membrane design, that had been tested in the first and last slot in session I, was tested in the second and third slot.

2.4. Statistics

Statistical analysis was performed with GraphPad Prism 5.02 (GraphPad Software, Inc., La Jolla, CA). Data were presented as means ± SEM. The Kolmogorov–Smirnov test was used to evaluate that all groups showed Gaussian distribution. Differences between groups were tested with the unpaired t-test. P-values <0.05 (*) and <0.01 (**) were considered significant. P-values <0.001 (***) were considered highly significant.

3. Results

The lowest CO2 removal rate (20.4 ± 0.2 mL/min) was found for the membrane lung with the smallest surface area, the Quadrox-i Neonatal, at the lowest blood (0.3 L/min) and sweep gas flow rates (0.5 L/min), that were tested (Figure 3). The best CO2 removal performance (105.0 ± 1.3 mL/min) was observed for the membrane lung with the highest SA, the Quadrox-iD Pediatric at the highest blood flow (0.9 L/min) and sweep gas flow (6 L/min). The QCO2 increased in a nonlinear rate both with an increased sweep gas (Figure 3, Figure 4, Figure 5 and Figure 6) for both the circular and parallel-plated membrane lungs. For the higher blood flow of 0.9 L/min, increased CO2 removal rates were observed. However, the increase was not proportional (Figure 4 and Figure 6).
In the comparison of the two low-flow PP fiber membrane lungs, the circular Hilite 1000 demonstrated an overall better performance than the parallel-plated Quadrox-i Neonatal for all sweep gas flow rates and for both evaluated blood flow rates of 0.3 L/min (Figure 3) and 0.9 L/min (Figure 4). The most significant advantages in CO2 removal of the Hilite 1000 over the Quadrox-i Neonatal were observed at the low sweep gas rate of 0.5 L/min. At higher sweep gas flow rates (4, 6 L/min), the differences of the two low-flow MLs were not always significantly different.
In the comparison of the mid-flow PMP fiber membrane lungs, the parallel-plated Quadrox-iD Pediatric showed significantly higher CO2 removal rates than the respective circular Minilung for both blood flow rates and the sweep gas flow rates of 2, 4, and 6 L/min (Figure 5 and Figure 6). For the Qsweep of 0.5 L/min, the advantage in the CO2 removal of the Quadrox-iD Pediatric cannot be observed anymore. Rather, a tendency of a benefit of the Minilung with a blood flow rate of 0.9 L/min can be shown in the second session (Figure 6).
Moreover, the pCO2(gas) vs the sweep gas flow rate was displayed in Figure 7 for each ML. It showed, that for all MLs, at low sweep gas flow rates of 0.5 L/min, pCO2(gas) approximates the pCO2(blood), which was 43 ± 2 mmHg in our measurements. The greater Qsweep, the lower pCO2(gas) is measured at the sweep gas outlet of the MLs.

4. Discussion

In this work, the efficiency in CO2 removal of parallel-plated oxygenators and circular oxygenators was evaluated with our already established mock circuit. The CO2 removal rate of the Quadrox-iD Pediatric, the standard oxygenator for ECCO2R in our unit, was in range with earlier results [12,13].
Few comparisons of different MLs can be found in the literature [10,14,15,16,17]. However, most of these works only focus on the efficiency of the compared MLs. Surface area, fiber structure and construction of the oxygenator were not completely respected [14,15,16]. Rambaud et al. compared a parallel-plated oxygenator to a circular oxygenator on the CO2 removal capacity in newborn patients and demonstrated a benefit of the circular oxygenator [17]. However, the surface area of the compared MLs was different, as the study compared the mid-flow membrane lung Quadrox-iD Pediatric (0.8 m2) to the low-flow membrane lung Medos Hilite 800LT (0.32 m2) [17]. Although the CO2 removal rates were normalized to the surface area, it remains debatable whether there is a linear relation over such a large range for the SA. In fact, Karagiannidis et al. showed a strong, but not linear, impact of the membrane SA on CO2 removal in his study with pigs [10]. However, this study used only parallel-plated MLs, and the effect of an increased membrane area varied for different blood flow rates. Hence, to make a reasonable comparison of two MLs, surface area, fiber structure, blood flow and sweep gas flow need to be equal.
Therefore, in our study we did not use normalization but rather tried to minimize the difference in membrane area of the compared membrane lungs.
Hence, the Novalung Minilung, a rebranded Medos Hilite 2400LT, with a surface area of 0.65 m2 rather than a Medos Hilite 800LT was used as the circular comparison to the Quadrox-iD Pediatric, with 0.80 m2 SA. Although the match of the surface area is not perfect, it was the closest we could get for parallel-plated and circular membrane lungs in the mid-flow range.
For the low-flow MLs, we used polypropylene MLs because, according to our knowledge, no commercially available low-flow parallel plated polymethylpentene MLs are produced. In this category, the two MLs, Quadrox-i-Neonatal (0.38 m2) and Medos Hilite 1000 (0.39 m2) were closer in surface area than their mid-flow counterparts.
The overall higher CO2 removal rates of the circular oxygenator Hilite 1000 in the low-flow contest and the Quadrox-iD Pediatric in the mid-flow contest were unsurprising, as both membranes surpassed their counterpart in terms of SA. More interesting was, however, how the relative performance differences shifted, when different sweep gas flow rates were used. For the lowest tested gas flow of 0.5 L/min, the parallel-plated Quadrox-iD Pediatric lost its lead, while the circular Hilite 1000 even increased its relative advantage.
All in all, we see a slight advantage of circular MLs for low sweep gas flow rates. We contribute this to the principal design of the membrane lungs. While the parallel-plated MLs have a blood flow perpendicular to the gas fiber orientation, the circular MLs have a blood flow that is almost antiparallel to the gas fibers. When the sweep gas travels along the gas fibers, it accumulates CO2. At low sweep gas flow rates, the CO2 partial pressure at the distant end of the gas fibers is close to the CO2 partial pressure in blood (Figure 7). As a result, this part of the membrane lung has an inefficient CO2 removal due to a small diffusion gradient.
In case of the circular membrane lung, the lower part of the oxygenator (according to the arrangement in Figure 8), where the blood enters, is affected by the adjustment of the CO2 partial pressure in gas and in blood. For the parallel-plated membrane lung this effect is more complex, as it has alternating layers of perpendicular orientated gas fibers. Therefore, the upper part of the membrane lung (according to the arrangement in Figure 8) has sweep gas with low CO2 content in both fiber orientations. The left and right parts of the membrane lung have sweep gas with low CO2 content in one and sweep gas with high CO2 content in the other fiber orientation, while the lower part of the membrane lung has sweep gas with high CO2 content in both fiber orientations. As a consequence, a blood fraction that passes the parallel-plated membrane lung in the lower part, will shunt to the blood outlet with little to no CO2 removal. As shown in Figure 7, for a sweep gas flow of 0.5 L/min, the CO2 partial pressure in sweep gas at the sweep gas outlet of the membrane lung gets close enough to the CO2 partial pressure in blood for this effect to be relevant. However, in the circular membrane lung, due to the antiparallel arrangement, all blood must pass the membrane area with inefficient CO2 removal first, but after that also passes a membrane area with efficient CO2 removal. Therefore, in this membrane lung design, no blood can shunt, independent of the sweep gas flow rate.
The most notable limitation of this study is the fact that there are no perfectly matched MLs for the circular and the parallel-plated design. Although we tried to use as similar membrane lungs as possible, the surface area remains a strong unmatched confounder.
Another critical factor for CO2 removal is the freshness of the used blood fluid itself. Ideally, for each session, both compared membranes would have to be tested at the same time with the same blood sample. As this is not possible to perform, we designed a symmetrical measuring sequence as shown in Figure 2 that minimizes the possible disadvantage of the ML measured second due to a less fresh or different blood sample. In our setup, for each experiment, each ML is tested two times. To overcome the disadvantage, the ML that is measured in the first slot will be measured in the last slot again. Moreover, in the second session, where the experiment is repeated with new MLs and a new blood sample, the measurement order is changed. That means that the ML that was measured in the first and in the last slot in the first session, was measured in the two middle slots of the second session and vice versa. Because of this arrangement, we have high confidence in the comparability of the measurements.
However, the performance comparison was only tested for new membrane lungs. We did not observe CO2 removal rates over longer time frames of several days. The longer a ML runs, the more clotting occurs within the ML, reducing the CO2 removal performance. It is possible that these deterioration processes are faster in some ML designs than in others. Therefore, the CO2 removal performance over longer time frames must be addressed in future experiments.
All in all, we conclude that circular MLs are more efficient in CO2 removal than parallel-plated MLs at low sweep gas flow rates. Such low sweep gas flow rates might not be relevant for everyday clinical practice of patients in acute respiratory distress syndrome and are only observed in the process of weaning a patient from ECMO or ECCO2R [18]. However, we think that our results might be relevant in the design of specialized ECMO devices, such as a portable ECCO2R, that needs to be extremely efficient with its limited sweep gas supply.

Author Contributions

Conceptualization, A.J.O.; investigation, L.S.S., A.M.J., N.S., S.C., P.M.L., and A.J.O.; supervision, P.M.L. and A.J.O.; writing—original draft, L.S.S., A.M.J., N.S., and A.J.O.; writing—review and editing, S.C., F.S., R.M.M., M.B., Q.T.D., R.B., and P.M.L. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Data can be provided on request addressed to the corresponding author. All data sharing statements are subject to conformity with German data protection legislation and rules (Datenschutzgrundverordnung—DGSVO).

Acknowledgments

Butcher’s shop “Peter Braun” in Homburg for fresh porcine blood free of charge.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Cooley, L.A.; Bausch, D.G.; Stojkovic, M.; Hosch, W.; Junghanss, T.; Stojkovic, M.; Hosch, W.; Junghanss, T.; Fletcher, N.; Pines, J.M.; et al. Extracorporeal Lung Support. In Encyclopedia of Intensive Care Medicine; Vincent, J.-L., Hall, J.B., Eds.; Springer: Berlin/Heidelberg, Germany, 2012; pp. 910–918. ISBN 978-3-642-00417-9. [Google Scholar]
  2. May, A.G.; Jeffries, R.G.; Frankowski, B.J.; Burgreen, G.W.; Federspiel, W.J. Bench Validation of a Compact Low-Flow CO2 Removal Device. Intensive Care Med. Exp. 2018, 6, 34. [Google Scholar] [CrossRef] [PubMed]
  3. Jeffries, R.G.; Lund, L.; Frankowski, B.; Federspiel, W.J. An Extracorporeal Carbon Dioxide Removal (ECCO2R) Device Operating at Hemodialysis Blood Flow Rates. Intensive Care Med. Exp. 2017, 5, 41. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  4. Ohshimo, S. Oxygen Administration for Patients with ARDS. J. Intensive Care 2021, 9, 17. [Google Scholar] [CrossRef] [PubMed]
  5. May, A.G.; Sen, A.; Cove, M.E.; Kellum, J.A.; Federspiel, W.J. Extracorporeal CO2 Removal by Hemodialysis: In Vitro Model and Feasibility. ICMx 2017, 5, 20. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  6. Fanelli, V.; Ranieri, M.V.; Mancebo, J.; Moerer, O.; Quintel, M.; Morley, S.; Moran, I.; Parrilla, F.; Costamagna, A.; Gaudiosi, M.; et al. Feasibility and Safety of Low-Flow Extracorporeal Carbon Dioxide Removal to Facilitate Ultra-Protective Ventilation in Patients with Moderate Acute Respiratory Distress Sindrome. Crit Care 2016, 20, 36. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  7. Terragni, P.P.; Del Sorbo, L.; Mascia, L.; Urbino, R.; Martin, E.L.; Birocco, A.; Faggiano, C.; Quintel, M.; Gattinoni, L.; Ranieri, V.M. Tidal Volume Lower than 6 Ml/Kg Enhances Lung Protection: Role of Extracorporeal Carbon Dioxide Removal. Anesthesiology 2009, 111, 826–835. [Google Scholar] [CrossRef] [Green Version]
  8. Barrett, N.A.; Hart, N.; Camporota, L. In-Vitro Performance of a Low Flow Extracorporeal Carbon Dioxide Removal Circuit. Perfusion 2019, 267659119865115. [Google Scholar] [CrossRef] [PubMed]
  9. Madhani, S.P.; May, A.G.; Frankowski, B.J.; Burgreen, G.W.; Federspiel, W.J. Blood Recirculation Enhances Oxygenation Efficiency of Artificial Lungs. ASAIO J. 2019. [Google Scholar] [CrossRef]
  10. Karagiannidis, C.; Strassmann, S.; Brodie, D.; Ritter, P.; Larsson, A.; Borchardt, R.; Windisch, W. Impact of Membrane Lung Surface Area and Blood Flow on Extracorporeal CO2 Removal during Severe Respiratory Acidosis. Intensive Care Med. Exp. 2017, 5, 34. [Google Scholar] [CrossRef] [Green Version]
  11. Baker, A.; Richardson, D.; Craig, G. Extracorporeal Carbon Dioxide Removal (ECCO2 R) in Respiratory Failure: An Overview, and Where Next? J. Intensive Care Soc. 2012, 13, 232–237. [Google Scholar] [CrossRef]
  12. Schwärzel, L.S.; Jungmann, A.M.; Schmoll, N.; Seiler, F.; Muellenbach, R.M.; Schenk, J.; Dinh, Q.T.; Bals, R.; Lepper, P.M.; Omlor, A.J. A Mock Circulation Loop to Test Extracorporeal CO2 Elimination Setups. Intensive Care Med. Exp. 2020, 8, 52. [Google Scholar] [CrossRef] [PubMed]
  13. Seiler, F.; Trudzinski, F.C.; Hennemann, K.; Niermeyer, T.; Schmoll, C.; Kamp, A.; Bals, R.; Muellenbach, R.M.; Haake, H.; Lepper, P.M. The Homburg Lung: Efficacy and Safety of a Minimal-Invasive Pump-Driven Device for Veno-Venous Extracorporeal Carbon Dioxide Removal. ASAIO J. 2017, 63, 659–665. [Google Scholar] [CrossRef] [PubMed]
  14. Sun, L.; Kaesler, A.; Fernando, P.; Thompson, A.J.; Toomasian, J.M.; Bartlett, R.H. CO2 Clearance by Membrane Lungs. Perfusion 2018, 33, 249–253. [Google Scholar] [CrossRef] [PubMed]
  15. Hospach, I.; Goldstein, J.; Harenski, K.; Laffey, J.G.; Pouchoulin, D.; Raible, M.; Votteler, S.; Storr, M. In Vitro Characterization of PrismaLung+: A Novel ECCO2R Device. Intensive Care Med. Exp. 2020, 8, 14. [Google Scholar] [CrossRef]
  16. Khan, S.; Vasavada, R.; Qiu, F.; Kunselman, A.; Undar, A. Extracorporeal Life Support Systems: Alternative vs. Conventional Circuits. Perfusion 2011, 26, 191–198. [Google Scholar] [CrossRef] [PubMed]
  17. Rambaud, J.; Guilbert, J.; Guellec, I.; Renolleau, S. A Pilot Study Comparing Two Polymethylpentene Extracorporeal Membrane Oxygenators. Perfusion 2013, 28, 14–20. [Google Scholar] [CrossRef] [PubMed]
  18. Grant, A.A.; Hart, V.J.; Lineen, E.B.; Badiye, A.; Byers, P.M.; Patel, A.; Vianna, R.; Koerner, M.M.; El Banayosy, A.; Loebe, M.; et al. A Weaning Protocol for Venovenous Extracorporeal Membrane Oxygenation with a Review of the Literature: VV ECMO WEANING PROTOCOL. Artif. Organs 2018, 42, 605–610. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Photographic and schematic view of our mock setup. The mock setup was filled with heparinized fresh porcine blood. The primary circuit simulated the human vena cava and generated venous blood by applying a sweep gas of N2 and CO2 to a Quadrox PLS membrane lung. The test circuit simulated the actual ECCO2R setup connected to the vena cava. In this experiment, comparable circular and parallel-plated membrane lungs were sequentially built into the test setup as test membrane lung (ML). The CO2 removal rate, which was calculated from gas flow rate and CO2 content at the outlet of the membrane lungs, was compared between circular and parallel-plated membrane lungs.
Figure 1. Photographic and schematic view of our mock setup. The mock setup was filled with heparinized fresh porcine blood. The primary circuit simulated the human vena cava and generated venous blood by applying a sweep gas of N2 and CO2 to a Quadrox PLS membrane lung. The test circuit simulated the actual ECCO2R setup connected to the vena cava. In this experiment, comparable circular and parallel-plated membrane lungs were sequentially built into the test setup as test membrane lung (ML). The CO2 removal rate, which was calculated from gas flow rate and CO2 content at the outlet of the membrane lungs, was compared between circular and parallel-plated membrane lungs.
Membranes 11 00398 g001
Figure 2. Measurement order of the MLs in session I and II: In each session, two experiments were conducted. In experiment I, the two low-flow membrane lungs were compared: The parallel-plated Quadrox-i Neonatal and the circular Hilite 1000. In experiment II, the two mid-flow membrane lungs were compared: The parallel-plated Quadrox-iD Pediatric and the circular Minilung. Each session had four symmetrical measuring slots to prevent an advantage of one ML due to fresher blood. Therefore, the membrane was either measured in slot 1 and slot 4 or in slot 2 and slot 3. In session II, the order was swapped.
Figure 2. Measurement order of the MLs in session I and II: In each session, two experiments were conducted. In experiment I, the two low-flow membrane lungs were compared: The parallel-plated Quadrox-i Neonatal and the circular Hilite 1000. In experiment II, the two mid-flow membrane lungs were compared: The parallel-plated Quadrox-iD Pediatric and the circular Minilung. Each session had four symmetrical measuring slots to prevent an advantage of one ML due to fresher blood. Therefore, the membrane was either measured in slot 1 and slot 4 or in slot 2 and slot 3. In session II, the order was swapped.
Membranes 11 00398 g002
Figure 3. Comparison of the CO2 removal performance of the Quadrox-i Neonatal (Neo) and the Hilite 1000 (H1000) at a blood flow of 0.3 L/min. (a): Session I, (b): Session II, (a)+(b): Combination of session I and II. Both sessions were performed on separate days with new blood samples and MLs.
Figure 3. Comparison of the CO2 removal performance of the Quadrox-i Neonatal (Neo) and the Hilite 1000 (H1000) at a blood flow of 0.3 L/min. (a): Session I, (b): Session II, (a)+(b): Combination of session I and II. Both sessions were performed on separate days with new blood samples and MLs.
Membranes 11 00398 g003
Figure 4. Comparison of the CO2 removal performance of the Quadrox-i Neonatal (Neo) and the Hilite 1000 (H1000) at a blood flow of 0.9 L/min. (a): Session I, (b): Session II, (a)+(b): Combination of session I and II. Both sessions were performed on separate days with new blood samples and MLs.
Figure 4. Comparison of the CO2 removal performance of the Quadrox-i Neonatal (Neo) and the Hilite 1000 (H1000) at a blood flow of 0.9 L/min. (a): Session I, (b): Session II, (a)+(b): Combination of session I and II. Both sessions were performed on separate days with new blood samples and MLs.
Membranes 11 00398 g004
Figure 5. Comparison of the CO2 removal performance of the Quadrox-iD Pediatric (Ped) and the Minilung (Mini) at a blood flow of 0.3 L/min. (a): Session I, (b): Session II, (a)+(b): Combination of session I and II. Both sessions were performed on separate days with new blood samples and MLs.
Figure 5. Comparison of the CO2 removal performance of the Quadrox-iD Pediatric (Ped) and the Minilung (Mini) at a blood flow of 0.3 L/min. (a): Session I, (b): Session II, (a)+(b): Combination of session I and II. Both sessions were performed on separate days with new blood samples and MLs.
Membranes 11 00398 g005
Figure 6. Comparison of the CO2 removal performance of the Quadrox-iD Pediatric (Ped) and the Minilung (Mini) at a blood flow of 0.9 L/min. (a): Session I, (b): Session II, (a)+(b): Combination of session I and II. Both sessions were performed on separate days with new blood samples and MLs.
Figure 6. Comparison of the CO2 removal performance of the Quadrox-iD Pediatric (Ped) and the Minilung (Mini) at a blood flow of 0.9 L/min. (a): Session I, (b): Session II, (a)+(b): Combination of session I and II. Both sessions were performed on separate days with new blood samples and MLs.
Membranes 11 00398 g006
Figure 7. Comparison of the pCO2(gas) of the parallel-plated and circular ML at Qsweep configurations of 0.5, 2, 4, 6 L/min and a BF of 0.3, 0.9 L/min. A: Comparison of the pCO2(gas) of the Quadrox-i Neonatal and the Hilite 1000. B: Comparison of the pCO2(gas) of the Quadrox-iD Pediatric and the Novalung Minilung.
Figure 7. Comparison of the pCO2(gas) of the parallel-plated and circular ML at Qsweep configurations of 0.5, 2, 4, 6 L/min and a BF of 0.3, 0.9 L/min. A: Comparison of the pCO2(gas) of the Quadrox-i Neonatal and the Hilite 1000. B: Comparison of the pCO2(gas) of the Quadrox-iD Pediatric and the Novalung Minilung.
Membranes 11 00398 g007
Figure 8. Fiber arrangement of parallel-plated and the circular MLs. (a): For the circular ML, blood flow is almost antiparallel to the gas fibers and the gas flow. At the lower part of the ML, where blood enters, it is in contact with the sweep gas with the highest amount of CO2. As diffusion of CO2 through the ML depends on the diffusion gradient between gas and blood, which is small in this part of the ML, the CO2 removal in this area is inefficient. (b): For the parallel-plated MLs, blood flow is perpendicular to the gas fibers. At the lower part of the ML where gas with highest CO2 content is in contact with blood, CO2 removal is inefficient.
Figure 8. Fiber arrangement of parallel-plated and the circular MLs. (a): For the circular ML, blood flow is almost antiparallel to the gas fibers and the gas flow. At the lower part of the ML, where blood enters, it is in contact with the sweep gas with the highest amount of CO2. As diffusion of CO2 through the ML depends on the diffusion gradient between gas and blood, which is small in this part of the ML, the CO2 removal in this area is inefficient. (b): For the parallel-plated MLs, blood flow is perpendicular to the gas fibers. At the lower part of the ML where gas with highest CO2 content is in contact with blood, CO2 removal is inefficient.
Membranes 11 00398 g008
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Schwärzel, L.S.; Jungmann, A.M.; Schmoll, N.; Caspari, S.; Seiler, F.; Muellenbach, R.M.; Bewarder, M.; Dinh, Q.T.; Bals, R.; Lepper, P.M.; et al. Comparison of Circular and Parallel-Plated Membrane Lungs for Extracorporeal Carbon Dioxide Elimination. Membranes 2021, 11, 398. https://doi.org/10.3390/membranes11060398

AMA Style

Schwärzel LS, Jungmann AM, Schmoll N, Caspari S, Seiler F, Muellenbach RM, Bewarder M, Dinh QT, Bals R, Lepper PM, et al. Comparison of Circular and Parallel-Plated Membrane Lungs for Extracorporeal Carbon Dioxide Elimination. Membranes. 2021; 11(6):398. https://doi.org/10.3390/membranes11060398

Chicago/Turabian Style

Schwärzel, Leonie S., Anna M. Jungmann, Nicole Schmoll, Stefan Caspari, Frederik Seiler, Ralf M. Muellenbach, Moritz Bewarder, Quoc Thai Dinh, Robert Bals, Philipp M. Lepper, and et al. 2021. "Comparison of Circular and Parallel-Plated Membrane Lungs for Extracorporeal Carbon Dioxide Elimination" Membranes 11, no. 6: 398. https://doi.org/10.3390/membranes11060398

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop