Reconstruction of the Physiological Behavior of Real and Synthetic Vessels in Controlled Conditions

by
Andrzej Polanczyk
1,*,†,
Aleksandra Piechota-Polanczyk
2,†,
Agnieszka W. Piastowska-Ciesielska
2,
Ihor Huk
3,
Christoph Neumayer
3,
Julia Balcer
4 and
Michal Strzelecki
4
1
The Faculty of Safety Engineering and Civil Protection, Fire University, 01-629 Warsaw, Poland
2
Department of Cell Cultures and Genomic Analysis, Medical University of Lodz, 90-752 Lodz, Poland
3
Department of Surgery, Division of Vascular Surgery, Medical University of Vienna, Spitalgasse 23, 1090 Vienna, Austria
4
Institute of Electronics, Lodz University of Technology, 93-590 Lodz, Poland
*
Author to whom correspondence should be addressed.
These authors contributed equally to this work.
Appl. Sci. 2024, 14(6), 2600; https://doi.org/10.3390/app14062600
Submission received: 17 February 2024 / Revised: 15 March 2024 / Accepted: 16 March 2024 / Published: 20 March 2024
(This article belongs to the Special Issue Recent Advances in Biomedical Engineering and Biomaterials)

Abstract

:
The objective of this study is to assess the ability of an Artificial Circulatory Phantom (ACP) to verify its accuracy in simulating the movement of artificial vessels vs. real vessels under changing cardiovascular parameters such as heartbeat, ejection fraction, and total peripheral resistance. The experiments were conducted with blood-like fluid that flows through two types of vessels: iliac arteries and different types of ePTFE vascular prostheses. Parameters such as diameter and tortuosity were measured and analyzed. The flow characteristics included a pulsating pattern with a frequency of 60–120 min−1 and ejection volumes ranging from 70 to 115 mL. The results showed a predominantly positive correlation between wall displacement (Wd) and tortuosity index (Ti) for the iliac artery (R2 = 0.981), as well as between Wd and mean tortuosity index (MTi) (R2 = 0.994). Similarly, positive correlations between Wd and Ti (R2 = 0.942) and Wd and MTi (R2 = 0.922) were computed for the ePTFE vascular prosthesis. The ACP introduced in this study is a valuable tool for evaluating various vessel types and the spatial configurations of vascular prostheses under diverse hemodynamic conditions. These findings are promising for the advancement of novel approaches to the testing and design of vascular grafts, ultimately enhancing their patency rates in future applications.

1. Introduction

Human organs are complex structures composed of specialized cells that collaboratively execute unique functions that are crucial for survival [1]. Ethical considerations necessitate the utilization of bio-tissue and mechanical systems that can replicate the physiological and pathophysiological aspects of organs [2]. Such systems hold the potential to provide enhanced predictive capabilities for human tissue responses compared to existing testing methodologies. Substantial progress has been made in microfluidic technologies, particularly in the context of organ-on-a-chip devices [3]. Each vascular product requires an individualized bioreactor [4]. Many important considerations should be included in the bioreactor’s construction [5]. Bioreactors for vascular products, as well as vascular grafts, commonly represent a pulsatile system [6]. Different parameters that are important for patients may be monitored in vitro, e.g., perfusion for different hemodynamic conditions, as well as the graft’s mechanical properties [7]. Such an approach minimizes the risk of failure during surgery.
In the realm of medical diagnostics, the integration of various techniques, such as ultrasonography [8], 3D ultrasound [9], 4D ultrasound strain (4D-US) [10], and echocardiography, is gaining popularity [11]. Combining these methods with engineering approaches, such as computational fluid dynamics (CFD), has the potential to enhance analyses and contribute to patient diagnosis [12,13,14]. These approaches are commonly employed for predicting therapeutic outcomes in cases of aortic-wall-associated injuries, such as aneurysms or wall ruptures [15]. However, limited efforts have been directed toward leveraging ex vivo systems to faithfully replicate blood flow parameters and investigate flow and pressure conditions in an artificial environment [16]. Moreover, most investigated models primarily reconstruct laminar flow or focus on graft preservation solutions [17,18]. Conversely, the pulsatile nature of blood hemodynamics in the cardiac system indicates the significant stretching of vessel walls, potentially leading to wall damage, especially in the presence of pre-existing conditions like atherosclerosis [19]. Furthermore, existing systems in the literature often only reconstruct laminar flow through artificial vessels or grafts [20], meaning that the relationship between elastic storage and the viscous dissipation of energy transmitted to vessel walls during systole in the cardiac cycle is not fully understood [21].
This study aims to verify its accuracy in simulating the movement of artificial vessels vs. real vessels under changing cardiovascular parameters such as heartbeat, ejection fraction, and total peripheral resistance. The experiments were conducted with blood-like fluid that flows through two types of vessels: iliac arteries and different types of ePTFE vascular prostheses.
The remainder of the manuscript is organized as follows: Section 2, Materials and Methods, outlines the experimental setup, the initial and boundary conditions, and the description of statistical analyses. Section 3, Results, presents the obtained results; Section 4, Discussion, contextualizes the findings within the current state-of-the-art; and Section 5, Conclusions, summarizes the paper.

2. Materials and Methods

2.1. Experimental Set-Up

An Artificial Circulatory Phantom (ACP) (Figure 1), designed to simulate the functioning of a segment of the human cardiovascular system, was constructed explicitly for this project and comprehensively detailed in our previous work [22]. In essence, at the core of the setup, a transparent rectangular vessel reactor, constructed from Plexiglas and equipped to maintain a constant temperature (temperature was measured with an installed temperature sensors), was positioned. The reactor was filled with a non-commercial fluid designed to mimic the rheological properties of blood. This fluid comprised 60% distilled water and 40% glycerol, presenting a density of 1.2 g cm−3 and a viscosity of 4.8 × 10−3 Pa·s. The fluid was supplied from a fluid chamber via a pump. The pulsating character of the flow was achieved by self-made artificial heart. This device manifested as a moving plunger housed in a stainless-steel corpus and driven by an electric motor, controlled discharge pressure, discharge volume, and the pulsation of prediction. These parameters were controlled by an electrical engine connected to the computer. Each time, a specified amount of the medium supplies the analyzed vessels, as well as ePTFE vascular prosthesis.
The vision system (VS) incorporated nine cameras positioned on an aluminum circular profile parallel to the floor, enabling horizontal movement [19]. The size of a pixel in the captured images of the real object was 0.35 mm. To increase the accuracy of vision acquisition, cameras were able to move 10 deg. Steps were included to both sides in the range of 40 deg. The entire system was controlled by a portable workstation (Dell Precision M6400 (Dell, Round Rock, TX, USA)) equipped with a four-core Intel CPU (2.4 GHz), 4 GB RAM (1333 MHz), and a 500 GB SSD HD. A self-developed acquisition and control software (ACS) facilitated the operation of the system.

2.2. Material

Two types of materials were investigated: iliac arteries (Figure 2a) and expanded polytetrafluoroethylene arterial (ePTFE) vascular prostheses (Figure 2b). Iliac arteries, with a diameter of 8 mm, thickness of 1 mm, and length of 100 mm, were procured from six male organ donors (mean age 50 ± 10 years) at the Medical University of Vienna. These arteries were promptly mounted in the experimental set-up, and all experiments were conducted within 24 h of their harvest from the six donor organs.
For comparison, six ePTFE vascular prostheses, commonly employed materials in vascular surgeries, were utilized. These prostheses had a diameter of 8 mm, thickness of 0.1 mm, and length of 100 mm. In total, six iliac arteries and six ePTFE vascular prostheses underwent investigation in this study.

2.3. Experimental Conditions

Changes in the diameter and displacement of the walls of the iliac arteries were investigated and compared with vascular prostheses made of expanded polytetrafluoroethylene (ePTFE) with a specified length of 100 mm, diameter of 8 mm, and wall thickness of 1 mm. The analysis encompassed iliac arteries sharing identical dimensions (length of 100 mm and diameter of 8 mm). The study received approval from the local Institutional Review Board (2069/2012) of the Medical University of Vienna. To validate the accuracy of the experimental setup, vision data were juxtaposed with clinical data obtained using the ultrasound 2D- Speckle tracking technique (2DSTT) (GE Vivid 7, GE Healthcare, Pittsburgh, PA, USA), equipped with software for strain analysis (EchoPac PC, GE Medical System, Pittsburgh, PA, USA, https://www.gehealthcare.com/products/ultrasound/vivid/echopac, accessed on 16 February 2024).
To verify the area change percentage (ACP) and vessel shape (VS), flow construction was performed with various parameters: pulsatile nature heart rates (FP)—60, 75, 90, 105, and 120 min−1—and different stroke volumes (volumes (EV): 70, 85, 100 and 115 mL) using non-commercial fluid for one cycle. The movement of the vessel wall was analyzed using 2DSTT, and it was ascertained whether the changes recorded with VS replicated the wall changes recorded with 2DSTT. Furthermore, current studies indicate that pulsatile flow plays a crucial role in the evaluation of pathological states, e.g., aneurysms [23].
For standardization, the changes in vessel diameter recorded by the ACP were compared with clinical data collected during patient examinations using the 2DSTT approach. Six ACP iliac arteries were compared with data from six healthy patients diagnosed with 2DSTT using the same type of arteries and hemodynamic conditions as in the ACP.
To further standardize the results for ACP patients and healthy subjects, tests were conducted under the conditions of heart rates (75 min−1 and 90 min−1) and stroke volume (EV = 70 mL). Medical data were then compared with the ACP results obtained under the same hemodynamic conditions. Ultimately, 120 cases were studied for iliac arteries and 120 cases for the ePTFE vascular prostheses.
The diameter change (Dc) (Equation (1)) and wall displacement (Wd) (Equation (2)) were calculated using the following equations:
Dc = Ddynamic(x,y,z) − Dstatic(x,y,z)
where:
  • Dc—vessel’s diameter change [mm].
  • Dstatic(x,y,z)—vessel’s diameter, recorded under static conditions [mm].
  • Ddynamic(x,y,z)—the maximal vessel’s diameter under dynamic conditions [mm].
Wd = Wdynamic (xp’,yp’,zp’) − Wstatic (xp,yp,zp)
where:
  • Wd—vessel’s wall displacement [mm].
  • Wstatic (xp,yp,zp)—spatial configuration of vessel’s wall in relation to its central axis, recorded under static conditions [mm].
  • Wdynamic (xp’,yp’,zp’)—spatial configuration of vessel’s wall in relation to its central axis, recorded under dynamic conditions [mm].
Moreover, tortuosity (Tortuosity index Ti (Equation (3)); Tortuosity (T) (Equation (4)) was applied to verify the correct operation of the WD parameter measurement.
Ti = L/C
where:
  • Ti—tortuosity index [-].
  • L—the linear distance of the vessel’s segment [mm].
  • C—the distance between vessel’s start and endpoint [mm].
T = 1 − d/l
where:
  • T—tortuosity [-].
  • d—the linear distance of the centerline [-].
  • L—the distance along the centerline [-].
Furthermore, a modification of the tortuosity (modified tortuosity index (Mti) (Equation (5))) presented in the literature was proposed.
MTi = L/C + 1/R
where:
  • MTi—modified tortuosity index [mm−1].
  • L, C—as defined in Equation (3).
  • R—the reciprocal of the radius of curvature at a particular point on the centerline of the analyzed vessel.

2.4. Statistical Analysis

Statistica 12.0 software (Tulsa, OK, USA) was used for statistical calculations. The results are presented as mean ± SD. The Bland–Altman method was used to assess the compatibility between the experimental setup and the 2DSTT technique. Additionally, a simple linear regression analysis was conducted to examine the correlation. Comparisons between the analyzed groups were performed using the Kolmogorov–Smirnov test after confirming normality and variance. Statistical significance was considered when p < 0.05, unless stated otherwise.

3. Results

3.1. Diameter Dilatation

The mean measured Dc (for FP = 75 min−1) for the iliac arteries was 0.94 ± 0.02 mm when calculated with the set-up and 0.92 ± 0.01 mm when calculated with 2DSTT (p = 0.069). The mean Dc (for FP = 75 min−1) for the ePTFE vascular prosthesis was 0.44 ± 0.01 mm when calculated with the set-up and 0.43 ± 0.01 mm when calculated with 2DSTT (p = 0.022). Moreover, the mean measured Dc (for FP = 90 min−1) for the iliac arteries was 1.04 ± 0.01 mm when calculated with the set-up and 1.02 ± 0.02 mm when calculated with 2DSTT (p = 0.022). The mean Dc (for FP = 90 min−1) for the ePTFE vascular prosthesis was 0.49 ± 0.01 mm when calculated with the set-up and 0.48 ± 0.01 mm when calculated with 2DSTT (p = 0.999) (Table 1).
The Bland–Altman analysis for FP = 75 min−1 and FP = 90 min−1 indicated that the records gathered with the set-up and 2DSTT for the iliac arteries were comparable. The difference between the set-up and 2DSTT was equal to 0.02 mm for the range equal to 0.04 mm (Figure 3a), while for FP = 90 min−1, it was 0.04 mm for the range equal to 0.13 mm (Figure 3b). Moreover, the difference between the set-up and 2DSTT for ePTFE vascular prostheses was equal to 0.01 mm for the range equal to 0.03 mm (Figure 3c), while for FP = 90 min−1, it was 0.01 mm for the range equal to 0.04 mm (Figure 3d).

3.2. Wall Displacement

Firstly, wall displacement was measured with the Set up (Table 2). For iliac arteries with different frequencies of pulsation, a similar wall displacement was observed (for 60 s−1, 75 s−1, 90 s−1, 105 s−1, and 120 s−1, 0.142 ± 0.002 mm, 0.144 ± 0.003 mm, 0.144 ± 0.004 mm, 0.142 ± 0.002 mm, and 0.141 ± 0.004 mm, respectively) (Figure 4a,b). For ePTFE vascular prostheses, an increase in the frequency of pulsation led to a decrease in wall displacement (for 60 s−1, 75 s−1, 90 s−1, 105 s−1 and 120 s−1, 0.741 ± 0.190 mm, 0.683 ± 0.107 mm, 0.670 ± 0.095 mm, 0.593 ± 0.103 mm and 0.600 ± 0.071 mm, respectively) (Figure 4c,d).
An example of the analyzed objects (iliac artery and ePTFE vascular prosthesis) is presented in Figure 5 and Figure 6. The degree of deformation was illustrated with triangles. Each time, greater flexibility of the iliac arteries was observed compared to the ePTFE vascular prostheses.

3.3. Tortuosity

Initially, the analysis focused on the tortuosity and tortuosity index. Notably, the average tortuosity and tortuosity index for the iliac arteries were comparable, reaching the level of 0.86 ± 0.02. Conversely, for ePTFE vascular prostheses, both the average tortuosity and tortuosity index were almost 3.5 times lower (0.24 ± 0.06) compared to the iliac arteries (p = 0.0001). Subsequently, the modified tortuosity index was examined, revealing that, for iliac arteries, the average modified tortuosity index was 0.87 ± 0.02. In contrast, for ePTFE vascular prostheses, the average modified tortuosity index and tortuosity indexes were 0.28 ± 0.05 (p = 0.0001).
The Bland–Altman analysis demonstrated minimal significant differences between Ti and MTi for iliac arteries. The discrepancy between Ti and MTi was 0.01, within the range of 0.02 (Figure 7a). In contrast, for ePTFE vascular prostheses, the difference between Ti and Mti was 0.04, within the range of 0.09 (Figure 7b).
Moreover, Wd was compared to T, Ti as well as MTi. Since the results of T and Ti were the same, Ti was used for further calculations. For the iliac artery, a positive correlation between Wd and Ti was calculated (R2 = 0.981). Similarly, for ePTFE vascular prostheses, a positive correlation between Wd and MTi (R2 = 0.994) (Figure 8a), Wd and Ti (R2 = 0.942), and Wd and MTi was observed (R2 = 0.922) (Figure 8b).

4. Discussion

In our earlier investigations, we demonstrated the ability of an ex vivo bioreactor to recreate hemodynamics in elastic 3D-printed models of abdominal aortic aneurysms [24,25] and human grafts [22]. In those papers, we showed that our self-made system allows us to successfully track hemodynamic changes in the abdominal aortic aneurysm (AAA) model with thrombus or stent-graft under different hemodynamic conditions. Here, we aimed to verify the accuracy of our platform to simulate the movement of artificial vessels in relation to a real vessel under changing cardiovascular parameters such as heartbeat, ejection fraction, and total peripheral resistance.
The integration of engineering and medical tools could offer a valuable method for the ex vivo analysis of biomaterials. [26]. Nevertheless, artificial systems have limitations, such as the specificity of their construction, the absence of a pulsating flow [27], or their reliance on animal tissues rather than human tissues [28]. Additionally, while MRI techniques offer data regarding the spatial configuration of vessels, they do not provide information about their stability or rupture [29]. The present research confirms that the Artificial Circulatory Phantom effectively replicates blood hemodynamics, allowing for the recreation of physiological conditions, such as the ejection volume, pulsation frequency, and ejection pressure for iliac arteries and ePTFE vascular prostheses. This corresponds with findings from comparable studies conducted by other researchers [11,30,31].
Significantly, we noted that the achievement of physiological conditions required the utilization of a water environment, an innovative approach with no prior literature precedent for simulating organ tension. For instance, Piola et al. devised a culture system to replicate hemodynamic conditions in saphenous veins following coronary artery bypass grafting [16,32]. Mundagi et al. investigated cellular interactions in pig carotid arteries using an ex vivo system, with flow reconstructed using a peristaltic pump at a constant speed [33]. However, these studies lacked an authentic representation of the heartbeat, which is a distinctive feature of our approach. Similarly, Bihari et al. examined abdominal aortic aneurysm displacement using 3D speckle-tracking ultrasound [31], but their horizontal vessel configuration differed from our vertical positioning.
Moreover, Genovese et al. applied a dedicated system for mouse artery stress analysis [34]. Gulan et al. studied the transparent part of the cardiac system using 3D particle-tracking velocimetry [35], but their methodology lacked a water environment and the measurement of vessel spatial configuration.
Taking into account the significance of the vessel’s physical properties for the reconstruction of clinical conditions [36,37], our study is consistent with the observations made by Pahlevan et al. [37], who identified changes in the aortic pressure wave associated with varying aortic stiffness, and Li et al. [38], who documented the age-related stiffening of the ascending aorta. Kashyap et al. [39] analyzed the accuracy of the tortuosity index in the left main coronary artery using computational modeling, contrasting with our results, which indicated a good agreement between the experimental and clinical data. Additionally, our study differs from the findings of Malve et al. and Pinho et al. [40,41], who noted higher averages and wider spreads of the WSS values.
In light of our presented results, we suggest that it is essential to examine other geometric parameters, such as angulations, in our system to accurately predict the occurrence of thrombus or prosthesis movement.

Limitation of the Study

The examined arteries and ePTFE vascular prostheses were limited to a length of 100 mm due to the dimensions of the chamber. This constraint was dictated by the design of the Artificial Circulatory Phantom (ACP). To increase the flexibility of the ACP, modifications should be considered for vessels of both shorter and longer lengths in future iterations. Furthermore, the inlet and outlet pipes were designed for vessels within the size range of the iliac arteries, limiting their suitability for smaller vessels. To expand the applicability of the ACP, additional adjustments to accommodate smaller vessels are necessary.

5. Conclusions

The Artificial Circulatory Phantom (ACP) presented in this article demonstrates the potential to be a valuable tool for evaluating different types of vascular prostheses under defined boundary and initial conditions. The ability to independently vary the frequency of pulsation and the ejection volume within the ACP unveiled distinct mechanical responses in iliac artery grafts and vascular prostheses. These insights offer valuable information that can contribute to the exploration of novel approaches in the design of vascular grafts, ultimately aiming to improve their patency rates in future applications. In the future, we plan to use our self-made equipment to test selected types of vascular prostheses.

Author Contributions

Conceptualization, A.P., A.W.P.-C. and A.P.-P.; methodology, I.H. and C.N.; software, M.S.; validation, A.P., A.P.-P. and M.S.; formal analysis, A.P.; investigation, A.P.; resources, A.P.; data curation, I.H. and C.N.; writing—original draft preparation, A.P., A.P.-P., I.H., C.N. and M.S.; writing—review and editing, A.P., A.P.-P., I.H., C.N., A.W.P.-C. and M.S.; visualization, A.P., A.P.-P. and J.B.; supervision, M.S.; project administration, A.P. and M.S.; funding acquisition, M.S. All authors have read and agreed to the published version of the manuscript.

Funding

The study was supported by the Polish National Centre for Research and Development (501/10-34-19-605 to AP) and by grant number 181110 from the Medical University of Vienna, Department of Surgery, Division of Vascular Surgery (to IH).

Institutional Review Board Statement

The study was approved by the local Institutional Review Board (2069/2012) of the Medical University of Vienna.

Informed Consent Statement

Informed consent was obtained from all subjects involved in the study.

Data Availability Statement

The data used to support the findings of this study are available from the corresponding author upon request.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Astashkina, A.; Mann, B.; Grainger, D.W. A Critical Evaluation of In Vitro Cell Culture Models for High-Throughput Drug Screening and Toxicity. Pharmacol. Ther. 2012, 134, 82–106. [Google Scholar] [CrossRef]
  2. Jung, S.-M.; Chung, G.Y.; Shin, H.S. Artificial Controlled Model of Blood Circulation System for Adhesive Evaluation. Sci. Rep. 2017, 7, 16720. [Google Scholar] [CrossRef] [PubMed]
  3. Sakolish, C.M.; Esch, M.B.; Hickman, J.J.; Shuler, M.L.; Mahler, G.J. Modeling Barrier Tissues In Vitro: Methods, Achievements, and Challenges. EBioMedicine 2016, 5, 30–39. [Google Scholar] [CrossRef]
  4. Ratcliffe, A.; Niklason, L.E. Bioreactors and Bioprocessing for Tissue Engineering. Ann. N. Y. Acad. Sci. 2002, 961, 210–215. [Google Scholar] [CrossRef]
  5. Fish, J.; Halberstadt, C.; McCoy, D.W.; Robbins, N. Bioreactor Design Considerations for Hollow Organs. Methods Mol. Biol. 2013, 1001, 207–214. [Google Scholar] [CrossRef]
  6. Diamantouros, S.E.; Hurtado-Aguilar, L.G.; Schmitz-Rode, T.; Mela, P.; Jockenhoevel, S. Pulsatile Perfusion Bioreactor System for Durability Testing and Compliance Estimation of Tissue Engineered Vascular Grafts. Ann. Biomed. Eng. 2013, 41, 1979–1989. [Google Scholar] [CrossRef]
  7. Maschhoff, P.; Heene, S.; Lavrentieva, A.; Hentrop, T.; Leibold, C.; Wahalla, M.; Stanislawski, N.; Blume, H.; Scheper, T.; Blume, C. An Intelligent Bioreactor System for the Cultivation of a Bioartificial Vascular Graft. Eng. Life Sci. 2016, 17, 567–578. [Google Scholar] [CrossRef]
  8. Karatolios, K.; Wittek, A.; Nwe, T.H.; Bihari, P.; Shelke, A.; Josef, D.; Schmitz-Rixen, T.; Geks, J.; Maisch, B.; Blase, C.; et al. Method for Aortic Wall Strain Measurement with Three-Dimensional Ultrasound Speckle Tracking and Fitted Finite Element Analysis. Ann. Thorac. Surg. 2013, 96, 1664–1671. [Google Scholar] [CrossRef]
  9. Rouet, L.; Mory, B.; Attia, E.; Bredahl, K.; Long, A.; Ardon, R. A Minimally Interactive and Reproducible Method for Abdominal Aortic Aneurysm Quantification in 3D Ultrasound and Computed Tomography with Implicit Template Deformations. Comput. Med. Imaging Graph. 2017, 58, 75–85. [Google Scholar] [CrossRef]
  10. Derwich, W.; Wittek, A.; Pfister, K.; Nelson, K.; Bereiter-Hahn, J.; Fritzen, C.P.; Blase, C.; Schmitz-Rixen, T. High Resolution Strain Analysis Comparing Aorta and Abdominal Aortic Aneurysm with Real Time Three Dimensional Speckle Tracking Ultrasound. Eur. J. Vasc. Endovasc. Surg. 2016, 51, 187–193. [Google Scholar] [CrossRef]
  11. Fadel, B.M.; Bakarman, H.; Al-Admawi, M.; Bech-Hanssen, O.; Di Salvo, G. Pulse-Wave Doppler Interrogation of the Abdominal Aorta: A Window to the Left Heart and Vasculature. Echocardiography 2014, 31, 543–547. [Google Scholar] [CrossRef]
  12. Klepaczko, A.; Szczypiński, P.; Strzelecki, M.; Stefańczyk, L. Simulation of Phase Contrast Angiography for Renal Arterial Models. BioMed. Eng. Online 2018, 17, 41. [Google Scholar] [CrossRef] [PubMed]
  13. Polanczyk, A.; Piechota-Polanczyk, A.; Huk, I.; Neumayer, C.; Strzelecki, M. Computational Fluid Dynamics as an Engineering Tool for the Reconstruction of Endovascular Prosthesis Endoleaks. IEEE Access 2022, 10, 18873–18885. [Google Scholar] [CrossRef]
  14. Polanczyk, A.; Podgorski, M.; Wozniak, T.; Stefanczyk, L.; Strzelecki, M. Computational Fluid Dynamics as an Engineering Tool for the Reconstruction of Hemodynamics after Carotid Artery Stenosis Operation: A Case Study. Medicina 2018, 54, 42. [Google Scholar] [CrossRef]
  15. Akcimen, M.; Gungor, F.; Beydilli, I.; Soydam Guven, D.; Kesapli, M.; Goksu, E. The Effects of Classic Two-Minute Pressing and Commercial Adhesive Bandages on Ecchymosis Development Following Removal of Intravenous Catheters. Emerg. Med. Investig. 2016, 110, 4. [Google Scholar] [CrossRef]
  16. Piola, M.; Ruiter, M.; Vismara, R.; Mastrullo, V.; Agrifoglio, M.; Zanobini, M.; Pesce, M.; Soncini, M.; Fiore, G.B. Full Mimicking of Coronary Hemodynamics for Ex-Vivo Stimulation of Human Saphenous Veins. Ann. Biomed. Eng. 2017, 45, 884–897. [Google Scholar] [CrossRef]
  17. Thomas, A.; Daniel Ou-Yang, H.; Lowe-Krentz, L.; Muzykantov, V.R.; Liu, Y. Biomimetic Channel Modeling Local Vascular Dynamics of Pro-Inflammatory Endothelial Changes. Biomicrofluidics 2016, 10, 014101. [Google Scholar] [CrossRef]
  18. Winkler, B.; Reineke, D.; Heinisch, P.P.; Schönhoff, F.; Huber, C.; Kadner, A.; Englberger, L.; Carrel, T. Graft Preservation Solutions in Cardiovascular Surgery. Interact. CardioVasc. Thorac. Surg. 2016, 23, 300–309. [Google Scholar] [CrossRef]
  19. Hicks, C.W.; Obeid, T.; Arhuidese, I.; Qazi, U.; Malas, M.B. Abdominal Aortic Aneurysm Repair in Octogenarians Is Associated with Higher Mortality Compared with Nonoctogenarians. J. Vasc. Surg. 2016, 64, 956–965.e1. [Google Scholar] [CrossRef]
  20. Janke, D.; Jankowski, J.; Rüth, M.; Buschmann, I.; Lemke, H.-D.; Jacobi, D.; Knaus, P.; Spindler, E.; Zidek, W.; Lehmann, K.; et al. The “Artificial Artery” as In Vitro Perfusion Model. PLoS ONE 2013, 8, e57227. [Google Scholar] [CrossRef]
  21. Ebrahimi, A.P. Mechanical Properties of Normal and Diseased Cerebrovascular System. J. Vasc. Interv. Neurol. 2009, 2, 155–162. [Google Scholar] [PubMed]
  22. Polanczyk, A.; Klinger, M.; Nanobachvili, J.; Huk, I.; Neumayer, C. Artificial Circulatory Model for Analysis of Human and Artificial Vessels. Appl. Sci. 2018, 8, 1017. [Google Scholar] [CrossRef]
  23. Asgharzadeh, H.; Borazjani, I. A Non-Dimensional Parameter for Classification of the Flow in Intracranial Aneurysms. I. Simplified Geometries. Phys. Fluids 2019, 31, 031904. [Google Scholar] [CrossRef]
  24. Polanczyk, A.; Podgorski, M.; Polanczyk, M.; Piechota-Polanczyk, A.; Stefanczyk, L.; Strzelecki, M. A Novel Vision-Based System for Quantitative Analysis of Abdominal Aortic Aneurysm Deformation. BioMed. Eng. OnLine 2019, 18, 56. [Google Scholar] [CrossRef] [PubMed]
  25. Polanczyk, A.; Podgorski, M.; Polanczyk, M.; Piechota-Polanczyk, A.; Neumayer, C.; Stefanczyk, L. A Novel Patient-Specific Human Cardiovascular System Phantom (HCSP) for Reconstructions of Pulsatile Blood Hemodynamic Inside Abdominal Aortic Aneurysm. IEEE Access 2018, 6, 61896–61903. [Google Scholar] [CrossRef]
  26. Iwasaki, K.; Kojima, K.; Kodama, S.; Paz, A.C.; Chambers, M.; Umezu, M.; Vacanti, C.A. Bioengineered Three-Layered Robust and Elastic Artery Using Hemodynamically-Equivalent Pulsatile Bioreactor. Circulation 2008, 118, S52–S57. [Google Scholar] [CrossRef]
  27. Tong, J.; Holzapfel, G.A. Structure, Mechanics, and Histology of Intraluminal Thrombi in Abdominal Aortic Aneurysms. Ann. Biomed. Eng. 2015, 43, 1488–1501. [Google Scholar] [CrossRef]
  28. Beenakker, J.-W.M.; Ashcroft, B.A.; Lindeman, J.H.N.; Oosterkamp, T.H. Mechanical Properties of the Extracellular Matrix of the Aorta Studied by Enzymatic Treatments. Biophys. J. 2012, 102, 1731–1737. [Google Scholar] [CrossRef]
  29. Asgharzadeh, H.; Asadi, H.; Meng, H.; Borazjani, I. A Non-Dimensional Parameter for Classification of the Flow in Intracranial Aneurysms. II. Patient-Specific Geometries. Phys. Fluids 2019, 31, 031905. [Google Scholar] [CrossRef] [PubMed]
  30. Urbina, J.; Sotelo, J.A.; Springmüller, D.; Montalba, C.; Letelier, K.; Tejos, C.; Irarrázaval, P.; Andia, M.E.; Razavi, R.; Valverde, I.; et al. Realistic Aortic Phantom to Study Hemodynamics Using MRI and Cardiac Catheterization in Normal and Aortic Coarctation Conditions. J. Magn. Reson. Imaging 2016, 44, 683–697. [Google Scholar] [CrossRef] [PubMed]
  31. Bihari, P.; Shelke, A.; Nwe, T.H.; Mularczyk, M.; Nelson, K.; Schmandra, T.; Knez, P.; Schmitz-Rixen, T. Strain Measurement of Abdominal Aortic Aneurysm with Real-Time 3D Ultrasound Speckle Tracking. Eur. J. Vasc. Endovasc. Surg. 2013, 45, 315–323. [Google Scholar] [CrossRef]
  32. Piola, M.; Soncini, M.; Prandi, F.; Polvani, G.; Beniamino Fiore, G.; Pesce, M. Tools and Procedures for Ex Vivo Vein Arterialization, Preconditioning and Tissue Engineering: A Step Forward to Translation to Combat the Consequences of Vascular Graft Remodeling. Recent Pat. Cardiovasc. Drug Discov. 2012, 7, 186–195. [Google Scholar] [CrossRef]
  33. Mundargi, R.; Venkataraman, D.; Kumar, S.; Mogal, V.; Ortiz, R.; Loo, J.; Venkatraman, S.; Steele, T. Novel Sensor-Enabled Ex Vivo Bioreactor: A New Approach towards Physiological Parameters and Porcine Artery Viability. BioMed Res. Int. 2015, 2015, 958170. [Google Scholar] [CrossRef]
  34. Genovese, K.; Lee, Y.-U.; Lee, A.Y.; Humphrey, J.D. An Improved Panoramic Digital Image Correlation Method for Vascular Strain Analysis and Material Characterization. J. Mech. Behav. Biomed. Mater. 2013, 27, 132–142. [Google Scholar] [CrossRef]
  35. Gülan, U.; Lüthi, B.; Holzner, M.; Liberzon, A.; Tsinober, A.; Kinzelbach, W. Experimental Investigation of the Influence of the Aortic Stiffness on Hemodynamics in the Ascending Aorta. IEEE J. Biomed. Health Inform. 2014, 18, 1775–1780. [Google Scholar] [CrossRef]
  36. Kang, J.; Aghilinejad, A.; Pahlevan, N.M. On the Accuracy of Displacement-Based Wave Intensity Analysis: Effect of Vessel Wall Viscoelasticity and Nonlinearity. PLoS ONE 2019, 14, e0224390. [Google Scholar] [CrossRef] [PubMed]
  37. Pahlevan, N.M.; Tavallali, P.; Rinderknecht, D.G.; Petrasek, D.; Matthews, R.V.; Hou, T.Y.; Gharib, M. Intrinsic Frequency for a Systems Approach to Haemodynamic Waveform Analysis with Clinical Applications. J. R. Soc. Interface 2014, 11, 20140617. [Google Scholar] [CrossRef]
  38. Li, Y.; Hickson, S.S.; McEniery, C.M.; Wilkinson, I.B.; Khir, A.W. Stiffening and Ventricular–Arterial Interaction in the Ascending Aorta Using MRI: Ageing Effects in Healthy Humans. J. Hypertens. 2019, 37, 347–355. [Google Scholar] [CrossRef]
  39. Kashyap, V.; Gharleghi, R.; Li, D.D.; McGrath-Cadell, L.; Graham, R.M.; Ellis, C.; Webster, M.; Beier, S. Accuracy of Vascular Tortuosity Measures Using Computational Modelling. Sci. Rep. 2022, 12, 865. [Google Scholar] [CrossRef] [PubMed]
  40. Pinho, N.; Castro, C.F.; António, C.C.; Bettencourt, N.; Sousa, L.C.; Pinto, S.I.S. Correlation between Geometric Parameters of the Left Coronary Artery and Hemodynamic Descriptors of Atherosclerosis: FSI and Statistical Study. Med. Biol. Eng. Comput. 2019, 57, 715–729. [Google Scholar] [CrossRef]
  41. Malvè, M.; Gharib, A.M.; Yazdani, S.K.; Finet, G.; Martínez, M.A.; Pettigrew, R.; Ohayon, J. Tortuosity of Coronary Bifurcation as a Potential Local Risk Factor for Atherosclerosis: CFD Steady State Study Based on In Vivo Dynamic CT Measurements. Ann. Biomed. Eng. 2015, 43, 82–93. [Google Scholar] [CrossRef]
Figure 1. An artificial circulatory phantom (ACP). The system comprised the following: 1—an artificial heart; 2—a transparent, rectangular chamber for the vessel; 3—a transparent container with distilled water and glycerol; 4—temperature sensors; 5—a camera; 6—a pump for distilled water and glycerol; 7—computer.
Figure 1. An artificial circulatory phantom (ACP). The system comprised the following: 1—an artificial heart; 2—a transparent, rectangular chamber for the vessel; 3—a transparent container with distilled water and glycerol; 4—temperature sensors; 5—a camera; 6—a pump for distilled water and glycerol; 7—computer.
Applsci 14 02600 g001
Figure 2. An artificial circulatory phantom (ACP): (a) iliac artery; (b) expanded polytetrafluoroethylene arterial (ePTFE) vascular prosthesis.
Figure 2. An artificial circulatory phantom (ACP): (a) iliac artery; (b) expanded polytetrafluoroethylene arterial (ePTFE) vascular prosthesis.
Applsci 14 02600 g002
Figure 3. Comparison of set-up and 2DSTT for: (a) iliac arteries for FP = 75 min−1; (b) iliac arteries for FP = 90 min−1; (c) ePTFE vascular prosthesis for FP = 75 min−1; (d) ePTFE vascular prosthesis for FP = 90 min−1 with the use of Bland–Altman analysis. The analyzed group consisted of n = 6. Statistical significance was considered for all analyses with a threshold of p < 0.05.
Figure 3. Comparison of set-up and 2DSTT for: (a) iliac arteries for FP = 75 min−1; (b) iliac arteries for FP = 90 min−1; (c) ePTFE vascular prosthesis for FP = 75 min−1; (d) ePTFE vascular prosthesis for FP = 90 min−1 with the use of Bland–Altman analysis. The analyzed group consisted of n = 6. Statistical significance was considered for all analyses with a threshold of p < 0.05.
Applsci 14 02600 g003
Figure 4. The wall displacement (WD) of: (a) iliac arteries as function of frequency of pulsation (FP) for FP = 75 min−1; (b) iliac arteries as a function of frequency of pulsation (FP) for FP = 90 min−1; (c) ePTFE vascular prosthesis as a function of frequency of pulsation (FP) for FP = 75 min−1; (d) ePTFE vascular prosthesis as a function of frequency of pulsation (FP) for FP = 90 min−1; analyzed cases consisted of n = 120.
Figure 4. The wall displacement (WD) of: (a) iliac arteries as function of frequency of pulsation (FP) for FP = 75 min−1; (b) iliac arteries as a function of frequency of pulsation (FP) for FP = 90 min−1; (c) ePTFE vascular prosthesis as a function of frequency of pulsation (FP) for FP = 75 min−1; (d) ePTFE vascular prosthesis as a function of frequency of pulsation (FP) for FP = 90 min−1; analyzed cases consisted of n = 120.
Applsci 14 02600 g004
Figure 5. An example of the iliac artery movement for: (a) EV = 70 mL, FP = 60 min−1; (b) EV = 70 mL, FP = 75 min−1; (c) EV = 85 mL, FP = 60 min−1; (d) EV = 85 mL, FP = 75 min−1. For each analyzed case, two centerlines were added. Centerlines illustrate how the vessel works for different hemodynamic parameters. Moreover, triangles were added to illustrate the degree of the vessel’s deformation/bent as area under the centerlines. Colors for each case were as follows: orange—EV = 70 mL, FP = 60 min−1; blue—EV = 70 mL, FP = 75 min−1; green—EV = 85 mL, FP = 60 min−1; red—EV = 85 mL, FP = 75 min−1.
Figure 5. An example of the iliac artery movement for: (a) EV = 70 mL, FP = 60 min−1; (b) EV = 70 mL, FP = 75 min−1; (c) EV = 85 mL, FP = 60 min−1; (d) EV = 85 mL, FP = 75 min−1. For each analyzed case, two centerlines were added. Centerlines illustrate how the vessel works for different hemodynamic parameters. Moreover, triangles were added to illustrate the degree of the vessel’s deformation/bent as area under the centerlines. Colors for each case were as follows: orange—EV = 70 mL, FP = 60 min−1; blue—EV = 70 mL, FP = 75 min−1; green—EV = 85 mL, FP = 60 min−1; red—EV = 85 mL, FP = 75 min−1.
Applsci 14 02600 g005
Figure 6. An example of ePTFE vascular prosthesis movement for: (a) EV = 70 mL, FP = 60 min−1; (b) EV = 70 mL, FP = 75 min−1; (c) EV = 85 mL, FP = 60 min−1; (d) EV = 85 mL, FP = 75 min−1. For each analyzed case, two centerlines were added. The centerlines illustrate how ePTFE vascular prosthesis works for different hemodynamic parameters. Moreover, triangles were added to illustrate the degree of ePTFE vascular prosthesis’s deformation/bent as the area under the centerlines. Colors for each case are as follows: orange—EV = 70 mL, FP = 60 min−1; blue—EV = 70 mL, FP = 75 min−1; green—EV = 85 mL, FP = 60 min−1; red—EV = 85 mL, FP = 75 min−1.
Figure 6. An example of ePTFE vascular prosthesis movement for: (a) EV = 70 mL, FP = 60 min−1; (b) EV = 70 mL, FP = 75 min−1; (c) EV = 85 mL, FP = 60 min−1; (d) EV = 85 mL, FP = 75 min−1. For each analyzed case, two centerlines were added. The centerlines illustrate how ePTFE vascular prosthesis works for different hemodynamic parameters. Moreover, triangles were added to illustrate the degree of ePTFE vascular prosthesis’s deformation/bent as the area under the centerlines. Colors for each case are as follows: orange—EV = 70 mL, FP = 60 min−1; blue—EV = 70 mL, FP = 75 min−1; green—EV = 85 mL, FP = 60 min−1; red—EV = 85 mL, FP = 75 min−1.
Applsci 14 02600 g006
Figure 7. Comparison of tortuosity index (Ti) and modified tortuosity index (MTi) for: (a) iliac arteries with the use of Bland–Altman analysis; (b) ePTFE vascular prosthesis with the use of Bland–Altman analysis. The analyzed cases consisted of n = 120. Statistical significance was considered for all analyses with a threshold of p < 0.05.
Figure 7. Comparison of tortuosity index (Ti) and modified tortuosity index (MTi) for: (a) iliac arteries with the use of Bland–Altman analysis; (b) ePTFE vascular prosthesis with the use of Bland–Altman analysis. The analyzed cases consisted of n = 120. Statistical significance was considered for all analyses with a threshold of p < 0.05.
Applsci 14 02600 g007
Figure 8. Correlation for: (a) the iliac arteries for the tortuosity index as function of wall deformation; (b) for the ePTFE vascular prosthesis for the tortuosity index as function of wall deformation. Scatterplot graphic; analyzed cases consisted of n = 120.
Figure 8. Correlation for: (a) the iliac arteries for the tortuosity index as function of wall deformation; (b) for the ePTFE vascular prosthesis for the tortuosity index as function of wall deformation. Scatterplot graphic; analyzed cases consisted of n = 120.
Applsci 14 02600 g008
Table 1. Diameter dilatation for the iliac artery and ePTFE vascular prosthesis. Values are measured in millimeters.
Table 1. Diameter dilatation for the iliac artery and ePTFE vascular prosthesis. Values are measured in millimeters.
Frequency of Pulsation [1/s]Ejection Volume [mL]Diameter of Iliac Artery (Measured with Set-Up) [mm]Diameter of Iliac Artery (Measured with 2DSTT) [mm]Diameter of ePTFE Vascular Prosthesis (Measured with Set-Up) [mm]Diameter of ePTFE Vascular Prosthesis (Measured with 2DSTT) [mm]
75700.890.910.420.56
75700.920.930.410.53
75700.930.950.430.51
75700.930.950.420.52
75700.920.950.440.54
75700.920.940.430.51
90700.991.040.480.48
90701.021.030.480.49
90701.011.030.490.48
90701.011.030.490.48
90701.041.040.480.49
90701.031.060.480.49
Table 2. Wall displacement for the iliac artery and ePTFE vascular prosthesis. Values are measured in millimeters.
Table 2. Wall displacement for the iliac artery and ePTFE vascular prosthesis. Values are measured in millimeters.
Frequency of Pulsation [1/s]Ejection Volume [mL]Wall Displacement for Iliac Artery (Measured with Set-Up) [mm]Wall Displacement for ePTFE Vascular Prosthesis (Measured with Set-Up) [mm]
75700.140.56
75700.150.53
75700.150.51
75700.150.52
75700.150.54
75700.150.51
90700.140.39
90700.140.38
90700.140.40
90700.140.40
90700.140.39
90700.140.40
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Polanczyk, A.; Piechota-Polanczyk, A.; Piastowska-Ciesielska, A.W.; Huk, I.; Neumayer, C.; Balcer, J.; Strzelecki, M. Reconstruction of the Physiological Behavior of Real and Synthetic Vessels in Controlled Conditions. Appl. Sci. 2024, 14, 2600. https://doi.org/10.3390/app14062600

AMA Style

Polanczyk A, Piechota-Polanczyk A, Piastowska-Ciesielska AW, Huk I, Neumayer C, Balcer J, Strzelecki M. Reconstruction of the Physiological Behavior of Real and Synthetic Vessels in Controlled Conditions. Applied Sciences. 2024; 14(6):2600. https://doi.org/10.3390/app14062600

Chicago/Turabian Style

Polanczyk, Andrzej, Aleksandra Piechota-Polanczyk, Agnieszka W. Piastowska-Ciesielska, Ihor Huk, Christoph Neumayer, Julia Balcer, and Michal Strzelecki. 2024. "Reconstruction of the Physiological Behavior of Real and Synthetic Vessels in Controlled Conditions" Applied Sciences 14, no. 6: 2600. https://doi.org/10.3390/app14062600

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop