Efficacy of Filter Trocar for Clear Visualization during Laparoscopic Cholecystectomy: A Prospective Randomized Controlled Trial

1
Samsung Medical Center, Division of Hepatobiliary-Pancreatic Surgery, Department of Surgery, Sungkyunkwan University School of Medicine, Seoul 06351, Republic of Korea
2
Department of Surgery, Seoul Song-do Hospital, Seoul 04597, Republic of Korea
3
Department of Surgery, Chung-Ang University College of Medicine, Seoul 06974, Republic of Korea
*
Author to whom correspondence should be addressed.
These authors contributed equally to this work.
J. Pers. Med. 2024, 14(2), 204; https://doi.org/10.3390/jpm14020204
Submission received: 18 January 2024 / Revised: 2 February 2024 / Accepted: 12 February 2024 / Published: 13 February 2024
(This article belongs to the Section Methodology, Drug and Device Discovery)

Abstract

:
Filter trocar designed to eliminate harmful smoke is also regarded as effective for improving surgical visualization. The aim of this study is to evaluate the efficacy of filter trocar in maintaining clear operative view. From 2019 to 2020, 100 patients underwent laparoscopic cholecystectomy and they were randomized to either the control or filter group. The primary end point was a laparoscopic operative view score (1, clear; 2, slightly blurry; 3, completely blurry) during gallbladder dissection from the liver bed when dissection was started (LV1), when dissection was half completed (LV2) and when dissection was completed (LV3). Between the control and filter groups, there were no significant differences in mean LV1 (1.44 vs. 1.40, p = 0.234) and LV3 (1.86 vs. 2.01, p = 0.880). There was no significant difference in the mean duration of suction after dissection (3.82 s vs. 3.67 s, p = 0.097) and the mean number of laparoscope removals from inside to outside the body to clean during gallbladder dissection from the liver bed (0.55 vs. 0.22, p = 0.963) or the mean amount of time required to dissect the gallbladder from the liver bed (221.58 s vs. 177.09 s, p = 0.253). The study demonstrated that filter trocar is not as effective as expected in the maintenance of clear operative view. Further study is needed to develop devices to improve clear surgical visualization.

1. Introduction

Maintaining a clear field of view is paramount in laparoscopic procedures not only for safety by preventing inadvertent injury, but also to improve precision and reduce operative time. Recently, with technological advancement, various energy-based surgical devices, such as monopolar or bipolar electrocautery devices, ultrasonic devices and laser systems, have shortened operation time. However, these devices also produce large amounts of surgical smoke during surgery which contains a mixture of water vapor, ultrafine particles and vaporized materials [1]. Compared to open surgery, confined spaces associated with laparoscopic procedure worsen the surgical view. The smoke generated from use of energy-based surgical devices often interferes with clear visualization of the surgical field in laparoscopic surgery and potentially increases risk of intra-operative complications [2]. The aerosolized particles in smoke can impair clear visualization by sticking to the lens of the laparoscope or by floating in the closed body space. Removing particles from the lens of a laparoscope requires pulling the laparoscope out from the body cavity for cleaning and removing floating particles or smoke involves venting or aspiration. These interventions take time to re-establish the adequate visualization of the surgical field and can impede surgical progress.
The smoke generated during laparoscopic surgery contains various harmful gases such as benzene, toluene, ethyl benzene styrene and formaldehyde. These are known as volatile organic compounds (VOCs), and they have been reported to cause acute or delayed toxicity and have potential carcinogenic effects on humans [3,4]. When smoke escapes from the abdominal cavity, the risk of inhalation by the surgical team increases and especially during the COVID-19 pandemic, the surgical team has become more aware of the dangers associate with surgical smoke, recognizing it as a potential carrier of harmful particles [5,6,7,8]. Various laparoscopic surgical smoke removal solutions have been developed, including filters and suction devices [9,10,11,12]. Filters made of various materials and thicknesses of fibers that can block these harmful gases are developed and have been used [9,10,11]. Several reports have demonstrated that the use of a filter in laparoscopic surgery significantly removes and reduces these harmful gases [9,10,11]. Generally, these laparoscopic filters are regarded as effective instruments for improving surgical visualization via smoke elimination and actually reduce the overall operative time. However, to our knowledge, no published studies have evaluated the efficacy of laparoscopic filters for maintaining a clear surgical view.
Laparoscopic cholecystectomy is among the most common laparoscopic procedures in the world. In laparoscopic cholecystectomy, during gallbladder dissection from the liver bed, the tissue generates a lot of smoke and this is the stage of the procedure most likely to be associated with surgical smoke disrupting a clear laparoscopic view. In general, we can remove the smoke using suction or opening the vent valve of the trocar, but these interventions rapidly decrease intra-abdominal pressure requiring the surgeon to wait until the CO2 is fully insufflated and the proper level of pressure is achieved. Moreover, moisture is usually collected that smudges the laparoscope lens, interfering with visibility, requiring repeated camera withdrawal to clean the lens and often adding to the operative time. However, when a filter is connected to the trocar, the intra-abdominal pressure does not decrease so fast and the surgeon does not need to wait for CO2 insufflation. Therefore, this study aimed to evaluate the efficacy of the filter trocar for maintaining a clear operative view in laparoscopic surgery.

2. Materials and Methods

A single-center prospective randomized controlled trial was conducted from September 2019 to August 2020 in Chung-Ang University Hospital. The study protocol was approved by the institutional review board of Chung-Ang University Hospital (IRB No. 1961-003-375) and was registered with Clinical Research Information Service (CRiS) of the Republic of Korea (identifier:KCT0004435, 14 November 2019). The study was reported in accordance with the rigor of the CONSORT guideline, and all experimental conditions conformed to the Declaration of Helsinki. All patients signed a written informed consent form before enrolment in the study. The filter used in this study was the SMOKE FILTERTM (Endovision, Seoul, Republic of Korea) as a complex filter, which has multiple layers of special filters, including activated carbon fiber, ultra-low particulate air (ULPA) and anti-viral filters. The filter was connected to the channel for the gas vent of the trocar.

2.1. Inclusion and Exclusion Criteria

Patients scheduled for elective laparoscopic cholecystectomy for gallstone disease or gallbladder polyp removal between 18 and 85 years of age with the capacity to provide informed consent were included. Exclusion criteria were those who were less than 18 years old; lactating or pregnant; had evidence of previous extensive abdominal surgery; or current enrollment in another drug or device study.

2.2. Sample Size Calculation

The sample size was calculated using the difference of time spent waiting or suctioning smoke for clear visualization. There has been no similar studies to guide the calculation, but a difference of 2 min was considered to be clinically significant. To detect a difference of 2 min as statistically significant, more than 45 patients would be required in each study arm based on the following assumption: α = 0.05, power = 0.80, within-group standard deviation = 2 min and ratio of control to experimental patients = 1. Assuming a dropout rate of 10%, a total of 100 participants was required, with 50 participants per group.

2.3. Randomization

A block randomization method was used. Enrolled patients were randomized at the time of their operation to either the trocar with filter group or trocar without filter group (control group) in a 1:1 ratio using a randomization table created before the surgery. This was undertaken by an independent surgical research nurse.

2.4. Surgical Procedure

All surgical procedures were performed by two surgeons in the same manner as the three-trocar technique. These surgeons had each performed more than 1000 laparoscopic cholecystectomies; this acted to standardize the variability of surgical skills. Pneumoperitoneum was maintained between 12 and14 mmHg. Under the xyphoid process, a 5 mm trocar was inserted for the operator’s right-hand access. Along the mid-clavicular line, below the costal margin, another 5 mm trocar with or without a filter was inserted for the operator’s left-hand access. Gallbladder dissection from the liver bed was carried out with a monopolar cauterizer. In the control group, the channel for a gas vent in a trocar was opened during gallbladder dissection from the liver bed. In the filter group, the connected channel to filter was opened during the whole procedure. Operative procedures were video recorded onto a standard high-definition recording device and subsequently transferred to individual high-definition memory cards, labeled with the participant’s identification number. The videos began with the start of monopolar hook dissection of the gallbladder from the liver bed, and they ended when the gallbladder was completely free. The suction of smoke using a suction device was performed only once after gallbladder removal.

2.5. Endpoints

The primary endpoint of this study was the laparoscopic view score during dissection of the gallbladder from the liver bed. The scoring system consisted of a 3-point scale according to blurriness, at three different moments. A score of 1 was applied when the laparoscopic view was clear, a score of 2 was applied when the laparoscopic view was slightly blurry and a score of 3 was applied when laparoscopic view was completely blurry (Figure 1). The laparoscopic view (LV) was also scored at different moments. LV1 was at the beginning of dissection, LV2 was when dissection was half completed and LV3 was when dissection was completed. Two surgeons who did not participate in the surgery reviewed the surgery videos and determined the scores.
The secondary endpoints were the duration of suction after gallbladder removal to maintain a clear surgical view, the number of laparoscope removals from inside to outside of body for cleaning during gallbladder dissection from the liver bed and the total time required for gallbladder dissection from the liver bed. Two surgeons who did not participate in the surgery reviewed the surgery videos and assessed these endpoints.

2.6. Data Collection and Statistical Analysis

Patient demographic and clinical data (age, gender, body mass index, history of prior abdominal surgery) were collected at the time of enrollment. Procedural data such as American Society of Anesthesiology Classification (ASA), estimated blood loss (EBL) and intra-operative complications were collected at the time of the procedure. Statistical analysis was performed with two-sided χ2 test using SPSS Statistics for Windows, version 24.0 (IBM Corp., Armonk, NY, USA). All analyses were performed according to the intention-to-treat principle. Data are presented as mean ± standard deviation (SD) or median. Student’s t test or the Mann–Whitney test was used to compare means or medians, and the χ2 or Fisher’s exact test was used to compare the frequency distributions between the categorical variables, as appropriate. A p-value < 0.05 was considered statistically significant.

3. Results

One hundred patients were assessed for eligibility. No patients were excluded. Therefore, 100 patients were randomized, with 50 patients allocated to the filter group and the other 50 allocated to the control group (Figure 2). All participants completed the intervention and there was no exclusion from analysis.

3.1. Clinicopathologic Characteristics

There were no significant differences between the filter group and the control group in age, sex, comorbidities, previous operative history, body mass index, total operation time, histopathologic results and postoperative complications (Table 1). There was one postoperative complication (incisional hernia through the umbilical port site in a patient in the filter group).

3.2. Primary Endpoint

In the filter group, the laparoscopic operative view was significantly blurrier when dissection was half completed (LV2; 1.94 vs. 1.88, p = 0.027). However, there were no significant intergroup differences in LV1 score (1.4 vs. 1.44, p = 0.234), LV3 score (2.01 vs. 1.86, p = 0.880) and overall LV score (1.65 vs. 1.56, p = 0.397) (Table 2).

3.3. Secondary Endpoint

There were no significant intergroup differences in the duration of suction after gallbladder removal to maintain a clear surgical view (3.82 s vs. 3.67 s, p = 0.097), the number of laparoscope removals from inside to outside of the body for cleaning during gallbladder dissection from the liver bed (0.55 vs. 0.22, p = 0.963) and the total time required for gallbladder dissecting from the liver bed (221.58 s vs. 177.09 s, p = 0.253) (Table 3).

4. Discussion

Ensuring surgical field visualization is important for safe surgery. Compared with open surgery, laparoscopic surgery is associated with more difficulties with maintaining surgical field visualization due to uncomfortable laparoscopic tools and limited peritoneal space [2]. Furthermore, with technological advancements, newly developed energy-based surgical instruments are widely used and these produce large quantities of smoke [1]. This smoke disturbs the operator’s view of the surgical field, interferes with proper dissection, increases operating time and potentially increases the risk of inadvertent intraoperative organ injury. Recently, the World Society of Emergency Surgery (WSES) expert panel has suggested laparoscopy as the first approach for stable patients undergoing emergency abdominal surgery for general surgery emergencies and abdominal trauma [13]. During emergency laparoscopy, tissue edema or inflammation can make more smoke or evaporated particles often causing an obscure surgical view. Therefore, intraoperative smoke removal is important to facilitate safe laparoscopic procedures during both scheduled and emergency operations and now is the right time to consider and develop effective methods. Although filters for trocar were originally developed for the removal of harmful gases generated during laparoscopic surgery, they are generally regarded to improve surgical visualization via smoke removal. However, the effect of filter trocar to remove and enhance laparoscopic visibility has not been evaluated and to our knowledge, this was the first randomized controlled study to evaluate the efficacy of the filter trocar for maintaining a clear operative view.
For smoke removal from the peritoneal space during laparoscopic surgery, opening of the vent valve in the trocar or suctioning the smoke using a suction device are the most commonly used methods. However, opening of the vent valve in the trocar or suctioning the smoke rapidly deflates the required pneumoperitoneum, making the abdominal cavity smaller and narrowing the surgical field of view. To recover sufficient intra-abdominal pressure and clear visualization, surgeons must wait for several seconds. If these steps are applied several times, it becomes time consuming and impedes operative efficiency. However, when a filter is connected to the trocar, since the intra-abdominal pressure is not decreasing as fast and the surgeon need not wait for CO2 insufflation, we hypothesized that a filter is effective for removing smoke and maintaining a clear surgical view. This is the first study to evaluate the potential advantage of filtered trocar for a clear surgical view by removing smoke and surgical plumes. We found that the filter did not improve surgical visualization during laparoscopic surgery and did not decrease the frequency or time of suctioning of smoke and the total time taken to dissect the gallbladder from the liver bed. Although recently developed smoke evacuation systems are theoretically considered effective for maintaining a surgical view during laparoscopic surgery by providing stable pneumoperitoneum and constant smoke evacuation, there has been no comparative study in terms of clear vision [14,15,16,17]. The theoretical advantage of the smoke evacuation system is that this system responds immediately to the slightest changes in intra-abdominal pressure, maintaining a stable-pressure pneumoperitoneum and continuous smoke evacuation even under difficult surgical conditions and constant suction, ensuring adequate visibility [14,15,16,17]. When comparing smoke evacuation systems to the filter trocar in the present study, the latter presents a list of benefits including acceptable price, no need for external system and complicated set-up and no requirement for additional space.
Although filter use was not so effective in maintaining surgical visualization, keeping the trocar open without a filter during surgery is associated with safety problems for medical staff because the smoke generated by using energy-based devices during surgery contains various harmful substances including carcinogens such as toluene, ethyl benzene and butadiene [3,4,18]. Baggish et al. confirmed a relationship between inhalation of unfiltered surgical smoke and changes in lung conditions including alveolar congestion and emphysema in a rat model [19,20] and several reports demonstrated that surgical smoke has potentially dangerous effects on medical staff [5,6,7,8,21]. The US Centers for Disease Control recommended the installation of smoke removal devices in the operating room and the use of N95 masks by medical staff during surgery [22]. In this regard, the use of filter trocar during laparoscopic surgery should be considered because the effect of filter trocar in term of removal of surgical smoke has been proven [9,10,11] and especially in Korea, the price of filter trocar is the same as regular trocar.
There are several limitations in this study. Firstly, since laparoscopic cholecystectomy was performed by an experienced surgeon, the total duration of gallbladder dissection from the liver bed was only within 4 min (221 s in the filter group and 177 s in the control group). Therefore, the difference in overall time and the time required for smoke suction between the two groups was likely to be minimal. For less-experienced staff, to achieve clear surgical visualization, removal of smoke may be more time consuming and a statistical significance could be achieved. It is necessary to evaluate the outcome of filter trocar on complex procedures with longer operating time and higher smoke production. Secondly, our scoring system for clear visualization was a subjective categorical measurement and dependent on the experimenters. However, to overcome this limitation, two surgeons evaluated the clarity of visualization.
Recently, a substantial number of computer-based visualization techniques which help in modelling different smoke components and lead to smoke-free images have been proposed to restore visibility in the context of hazy surgical fields [23,24,25]. However, they are undergoing validation studies, and it is expected to take some time before such technology is commercialized.

5. Conclusions

The present study demonstrated that the laparoscopic operative view during dissection of the gallbladder from the liver bed was not clearer in the filter group than the in-control group. Therefore, a filter for trocar use does not represent an improvement over a lack of use to maintain a clear operative view in laparoscopic surgery. However, we believe that further research and development should enhance the filtering efficiency and further study is needed to develop or identify the new devices or methods for smoke clearance and improvement of clear surgical visualization.

Author Contributions

H.-C.C., B.-J.K. and S.E.L. participated in the conception and design of the study, and the drafting of the manuscript B.-J.K. and S.-W.S. lead the data processing, statistical analysis and prepared the figures. Y.S.C. and S.E.L. supervised the study. All authors reviewed, edited and approved the manuscript. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

The study was conducted in accordance with the Declaration of Helsinki, and approved by the Institutional Review Board of Chung-Ang University Hospital (IRB No. 1961-003-375) and was registered with Clinical Research Information Service (CRiS) of Republic of Korea (identifier: KCT0004435, 14 November 2019).

Informed Consent Statement

Informed consent was obtained from all subjects involved in the study. Written informed consent has been obtained from the patients to publish this paper.

Data Availability Statement

Data related to the manuscript are available upon request to corresponding author.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Spearman, J.; Tsavellas, G.; Nichols, P. Current attitudes and practices towards diathermy smoke. Ann. R. Coll. Surg. Engl. 2007, 89, 162–165. [Google Scholar] [CrossRef]
  2. Weld, K.J.; Dryer, S.; Ames, C.D.; Cho, K.; Hogan, C.; Lee, M.; Biswas, P.; Landman, J. Analysis of surgical smoke produced by various energy-based instruments and effect on laparoscopic visibility. J. Endourol. 2007, 21, 347–351. [Google Scholar] [CrossRef] [PubMed]
  3. Choi, S.H.; Kwon, T.G.; Chung, S.K.; Kim, T.H. Surgical smoke may be a biohazard to surgeons performing laparoscopic surgery. Surg. Endosc. 2014, 28, 2374–2380. [Google Scholar] [CrossRef] [PubMed]
  4. Baan, R.; Grosse, Y.; Straif, K.; Secretan, B.; El Ghissassi, F.; Bouvard, V.; Benbrahim-Tallaa, L.; Guha, N.; Freeman, C.; Galichet, L.; et al. A review of human carcinogens—Part F: Chemical agents and related occupations. Lancet Oncol. 2009, 10, 1143–1144. [Google Scholar] [CrossRef]
  5. Mintz, Y.; Arezzo, A.; Boni, L.; Chand, M.; Brodie, R.; Fingerhut, A.; the Technology Committee of the European Association for Endoscopic Surgery. A Low-cost, Safe, and Effective Method for Smoke Evacuation in Laparoscopic Surgery for Suspected Coronavirus Patients. Ann. Surg. 2020, 272, e7–e8. [Google Scholar] [CrossRef]
  6. The American College of Surgeons (ACS). COVID-19 and Surgery. 2020. Available online: https://www.facs.org/covid-19/clinical-guidance (accessed on 11 February 2024).
  7. The Society of American Gastrointestinal and Endoscopic Surgeons (SAGES). Recommendations Regarding Surgical Response to COVID-19 Crisis 2020. Available online: https://www.sages.org/recommendations-surgical-response-covid-19/ (accessed on 11 February 2024).
  8. The Royal College of Surgeons. Updated Intercollegiate General Surgery Guidance on COVID-19. 2020. Available online: https://www.rcseng.ac.uk/coronavirus/joint-guidance-for-surgeons-v2/ (accessed on 11 February 2024).
  9. Hahn, K.Y.; Kang, D.W.; Azman, Z.A.M.; Kim, S.Y.; Kim, S.H. Removal of Hazardous Surgical Smoke Using a Built-in-Filter Trocar: A Study in Laparoscopic Rectal Resection. Surg. Laparosc. Endosc. Percutan Tech. 2017, 27, 341–345. [Google Scholar] [CrossRef]
  10. Choi, S.H.; Choi, D.H.; Kang, D.H.; Ha, Y.-S.; Lee, J.N.; Kim, B.S.; Kim, H.T.; Yoo, E.S.; Kwon, T.G.; Chung, S.K.; et al. Activated carbon fiber filters could reduce the risk of surgical smoke exposure during laparoscopic surgery: Application of volatile organic compounds. Surg. Endosc. 2018, 32, 4290–4298. [Google Scholar] [CrossRef]
  11. Ha, H.I.; Choi, M.C.; Jung, S.G.; Joo, W.D.; Lee, C.; Song, S.H.; Park, H. Chemicals in Surgical Smoke and the Efficiency of Built-in-Filter Ports. JSLS 2019, 23, e2019.00037. [Google Scholar] [CrossRef]
  12. Robertson, D.; Sterke, F.; van Weteringen, W.; Arezzo, A.; Mintz, Y.; Technology Committee of the European Association for Endoscopic Surgery (EAES); Horeman, T. Escape of surgical smoke particles, comparing conventional and valveless trocar systems. Surg. Endosc. 2023, 37, 8552–8561. [Google Scholar] [CrossRef]
  13. Sermonesi, G.; Tian, B.W.C.A.; Vallicelli, C.; Abu-Zidan, F.M.; Damaskos, D.; Kelly, M.D.; Leppäniemi, A.; Galante, J.M.; Tan, E.; Kirkpatrick, A.W.; et al. Cesena guidelines: WSES consensus statement on laparoscopic-first approach to general surgery emergencies and abdominal trauma. World J. Emerg. Surg. 2023, 18, 57. [Google Scholar] [CrossRef]
  14. Annino, F.; Topazio, L.; Autieri, D.; Verdacchi, T.; De Angelis, M.; Asimakopoulos, A.D. Robotic partial nephrectomy performed with Airseal versus a standard CO2 pressure pneumoperitoneum insufflator: A prospective comparative study. Surg. Endosc. 2017, 31, 1583–1590. [Google Scholar] [CrossRef]
  15. Hamed, H. Underwater-seal evacuation of surgical smoke in laparoscopy during the COVID-19 pandemic: A feasibility report of a simple technique. Br. J. Surg. 2020, 107, e640–e641. [Google Scholar] [PubMed]
  16. da Costa, K.M.; Saxena, A.K. Coronavirus disease 2019 pandemic and identifying insufflators with desufflation mode and surgical smoke evacuators for safe CO2 removal. Asian J. Endosc. Surg. 2021, 14, 165–169. [Google Scholar] [CrossRef]
  17. Hirota, M.; Takahashi, H.; Takahashi, T.; Kurokawa, Y.; Yamasaki, M.; Eguchi, H.; Doki, Y.; Nakajima, K. A Smoke Evacuator Equipped With a Filter Contributes Enough to Safe Gas Exhaustion From the Abdominal Cavity. Asian J. Endosc. Surg. 2022, 15, 427–431. [Google Scholar] [CrossRef]
  18. Ansell, J.; Warren, N.; Wall, P.; Cocks, K.; Goddard, S.; Whiston, R.; Stechman, M.; Scott-Coombes, D.; Torkington, J. Electrostatic precipitation is a novel way of maintaining visual field clarity during laparoscopic surgery: A prospective double-blind randomized controlled pilot study. Surg. Endosc. 2014, 28, 2057–2065. [Google Scholar] [CrossRef] [PubMed]
  19. Baggish, M.S.; Elbakry, M. The effects of laser smoke on the lungs of rats. Am. J. Obstet. Gynecol. 1987, 156, 1260–1265. [Google Scholar] [CrossRef] [PubMed]
  20. Baggish, M.S.; Baltoyannis, P.; Sze, E. Protection of the rat lung from the harmful effects of laser smoke. Lasers Surg. Med. 1988, 8, 248–253. [Google Scholar] [CrossRef]
  21. Alp, E.; Bijl, D.; Bleichrodt, R.P.; Hansson, B.; Voss, A. Surgical smoke and infection control. J. Hosp. Infect. 2006, 62, 1–5. [Google Scholar] [CrossRef]
  22. Sehulster, L.; Chinn, R.Y.; CDC; HICPAC. Guidelines for environmental infection control in health-care facilities: Recommendations of CDC and the Healthcare Infection Control Practices Advisory Committee (HICPAC). MMWR Recomm. Rep. 2003, 52, 1–42. [Google Scholar]
  23. He, K.; Sun, J.; Tang, X. Single Image Haze Removal Using Dark Channel Prior. IEEE Trans. Pattern Anal. Mach. Intell. 2011, 33, 2341–2353. [Google Scholar]
  24. Zhu, H.; Cheng, Y.; Peng, X.; Zhou, J.T.; Kang, Z.; Lu, S.; Fang, Z.; Li, L.; Lim, J.-H. Single-Image Dehazing via Compositional Adversarial Network. IEEE Trans. Cybern. 2021, 51, 829–838. [Google Scholar] [CrossRef] [PubMed]
  25. Pan, Y.; Bano, S.; Vasconcelos, F.; Park, H.; Jeong, T.T.; Stoyanov, D. DeSmoke-LAP: Improved unpaired image-to-image translation for desmoking in laparoscopic surgery. Int. J. Comput. Assist. Radiol. Surg. 2022, 17, 885–893. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Laparoscopic view score. (a) Score 1: the laparoscopic view is clear. (b) Score 2: the laparoscopic view is slightly blurry. (c) Score 3: the laparoscopic view is completely blurry.
Figure 1. Laparoscopic view score. (a) Score 1: the laparoscopic view is clear. (b) Score 2: the laparoscopic view is slightly blurry. (c) Score 3: the laparoscopic view is completely blurry.
Jpm 14 00204 g001aJpm 14 00204 g001b
Figure 2. CONSORT patient inclusion flow chart. This study was registered with Clinical Research Information Service (#KCT0004435).
Figure 2. CONSORT patient inclusion flow chart. This study was registered with Clinical Research Information Service (#KCT0004435).
Jpm 14 00204 g002
Table 1. Clinicopathologic characteristics of the patients.
Table 1. Clinicopathologic characteristics of the patients.
Filter Group (n = 50)Control Group (n = 50)p-Value
Age (years)47.750.50.300
Sex, M:F21:2925:250.433
Comorbidity15 (29.4%)12 (24.5%)0.579
DM7 (13.7%)4 (8.2%)0.374
Hepatitis2 (3.9%)4 (8.2%)0.372
Hypertension7 (13.7%)6 (12.2%)0.826
Previous op hx.12 (23.5%)15 (30.6%)
BMI25.224.70.491
Operation time (min)43.241.60.560
Pathologic result
Chronic cholecystitis42 (82.4%)38 (79.2%)0.338
GB polyp9 (17.6%)8 (16.7%)0.230
GB cancer0 (0.0%)2 (4.2%)0.535
Postoperative complication1 (2%) *0 (0%)0.325
*; incisional hernia through umbilical port site.
Table 2. Primary endpoints: the laparoscopic view scores.
Table 2. Primary endpoints: the laparoscopic view scores.
Filter Group (n = 50)Control Group (n = 50)p-Value
LV* 1 score1.41.440.234
LV2 score1.941.880.027
LV3 score2.011.860.880
LV average score1.651.560.397
LV*; laparoscopic view.
Table 3. Secondary endpoints: duration of suction after gallbladder removal to maintain a clear surgical view, numbers of laparoscope removals from inside to outside of body for lens cleaning during gallbladder dissection from the liver bed and total time required for gallbladder dissection from the liver bed.
Table 3. Secondary endpoints: duration of suction after gallbladder removal to maintain a clear surgical view, numbers of laparoscope removals from inside to outside of body for lens cleaning during gallbladder dissection from the liver bed and total time required for gallbladder dissection from the liver bed.
Filter Group (n = 50)Control Group (n = 50)p-Value
Duration of suction after gallbladder removal (s)3.823.670.097
Number of laparoscope removals (times)0.550.220.963
Total time required for gallbladder dissection (s)221.58177.090.253
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Chae, H.-C.; Kim, B.-J.; Choi, Y.S.; Suh, S.-W.; Lee, S.E. Efficacy of Filter Trocar for Clear Visualization during Laparoscopic Cholecystectomy: A Prospective Randomized Controlled Trial. J. Pers. Med. 2024, 14, 204. https://doi.org/10.3390/jpm14020204

AMA Style

Chae H-C, Kim B-J, Choi YS, Suh S-W, Lee SE. Efficacy of Filter Trocar for Clear Visualization during Laparoscopic Cholecystectomy: A Prospective Randomized Controlled Trial. Journal of Personalized Medicine. 2024; 14(2):204. https://doi.org/10.3390/jpm14020204

Chicago/Turabian Style

Chae, Ho-Chang, Beom-Jin Kim, Yoo Shin Choi, Suk-Won Suh, and Seung Eun Lee. 2024. "Efficacy of Filter Trocar for Clear Visualization during Laparoscopic Cholecystectomy: A Prospective Randomized Controlled Trial" Journal of Personalized Medicine 14, no. 2: 204. https://doi.org/10.3390/jpm14020204