Review

Integrated Resonant Micro/Nano Gravimetric Sensors for Bio/Chemical Detection in Air and Liquid

State Key Lab of Transducer Technology, Shanghai Institute of Microsystem & Information Technology, Chinese Academy of Sciences, 865 Changning Road, Shanghai 200050, China
*
Author to whom correspondence should be addressed.
Micromachines 2021, 12(6), 645; https://doi.org/10.3390/mi12060645
Submission received: 7 May 2021 / Accepted: 27 May 2021 / Published: 31 May 2021
(This article belongs to the Special Issue Integrated MEMS Resonators)

Abstract

:
Resonant micro/nanoelectromechanical systems (MEMS/NEMS) with on-chip integrated excitation and readout components, exhibit exquisite gravimetric sensitivities which have greatly advanced the bio/chemical sensor technologies in the past two decades. This paper reviews the development of integrated MEMS/NEMS resonators for bio/chemical sensing applications mainly in air and liquid. Different vibrational modes (bending, torsional, in-plane, and extensional modes) have been exploited to enhance the quality (Q) factors and mass sensing performance in viscous media. Such resonant mass sensors have shown great potential in detecting many kinds of trace analytes in gas and liquid phases, such as chemical vapors, volatile organic compounds, pollutant gases, bacteria, biomarkers, and DNA. The integrated MEMS/NEMS mass sensors will continuously push the detection limit of trace bio/chemical molecules and bring a better understanding of gas/nanomaterial interaction and molecular binding mechanisms.

1. Introduction

The past two decades have witnessed significant development in sensor technologies for the recognition and detection of chemical (e.g., volatile organic compounds, VOCs) and biological species (e.g., cells, proteins) in ambient and liquid environments. In contrast to the optical methods, micro-/nanoelectromechanical systems (e.g., microcantilevers), featured by miniature device size, exquisite detection limit (from part per million to part per trillion, or micromolar to femtomolar), and easy to be on-chip integrated, have offered better opportunities to detect the trace analytes and capture the molecular interaction processes in gas and liquid phases [1,2,3].
Taking the prevailing microcantilever sensors as an example, two different operational modes (i.e., static and dynamic modes) have been engineered for bio/chemical sensing. In static mode, the cantilever surface is functionalized to have a good affinity to the target molecules. The stress change at the cantilever surface (owing to electrostatic repulsion or attraction, steric interactions, hydration, and entropic effects) causes the cantilever bending, which is usually measured by an optical lever system [4]. Since the responses of surface-stress sensors are difficult to interpret, a dynamic scheme has been proposed by operating a cantilever at resonance. A tiny mass loading (Δm) at a cantilever free end can lead to a resonant frequency downshift (Δf), which defines the mass sensitivity: = Δ f / Δ m = f / 2 M , where f = 1 / 2 π k / M is the resonant frequency, k and M are the effective spring constant and effective mass of the resonant device [5]. Therefore, one can perceive that micro-and nanoelectromechanical (MEMS/NEMS) resonators promise gravimetric detection of trace analytes down to a fraction of the sensor mass. In comparison with the conventional quartz crystal microbalances (QCM), the silicon micro-machined resonant cantilevers possess finer mass resolution, smaller size, low-cost batch fabrication, and easy IC-compatible integration [6]. Although pioneering efforts have been made towards attogram and atomic-level resonant mass sensing, with outstanding device performance often achieved by off-sensor optical detection methods in high vacuum or at low temperature [7,8,9,10,11,12,13,14,15,16], there is an increasing need for potable micro- and nano-gravimetric sensors with integrated resonance excitation and readout elements on-chip, toward real-world bio/chemical sensing applications.
To date, several on-chip resonance excitation and readout mechanisms have been developed, considering their compatibility with ambient and liquid conditions (especially conductive biosolutions). As to resonance readout, most microcantilevers use piezoresistive effect, as the doped semiconducting materials (mostly silicon) provide very large gauge factors, and the fabrication process has been well-developed and optimized [17,18]. The piezoresistors are often patterned at the device’s clamping points (e.g., fixed end of the cantilever), where the largest stress is expected at resonance. Wheatstone bridge with 4 terminal piezoresistors is also desirable to improve the displacement sensitivity to nanometer scale in air and liquid.
As to resonance excitation, electrothermal excitation has been widely adopted on many resonant MEMS/NEMS prototypes, such as microcantilevers [19], ‘dog-bone’ resonators [20], and disk resonators [21]. A DC + AC voltage signal is typically applied to a heating resistor patterned at the device’s clamping points, and the time-varying heating power whose cycle matches the device’s resonant frequency induces the mechanical vibration. On the other hand, electromagnetic excitation has been achieved with metal loops patterned around the cantilevers. With the presence of an external magnetic field (tens of mT) and an AC electrical current flowing through the metal loops, Lorentzian force drives the bending or torsional vibration, depending on the positions of the metal loops on the resonators [17,22].
Alternatively, the piezoelectric approach emerges with the successful synthesis of ceramics with high piezoelectric constants (such as aluminum nitride, AlN, Pb-based lanthanum-doped zirconate titanates, PZT). The piezoelectric effect enables simultaneous self-excitation/readout. A pure AC input signal creates time-varying stress at the clamping points, while another AC piezoelectric current signal is generated to read out the resonant frequency [23,24,25].
Although on-chip, all-electrical integration brings immense simplification to the bulky, expensive measurement system that is inevitable for optical readout, researchers are still confronted with technical challenges of operating resonant sensors in viscous media that dissipates resonating energy, and great endeavors have been made to achieve better device performance (e.g., higher Q, better mass sensitivity and mass resolution) in air and liquid by studying different vibrational modes, such as bending, torsional, in-plane, and extensional modes (as shown in Figure 1 and Table 1).

2. Integrated Resonant Gravimetric Sensors Using Different Vibrational Modes

2.1. Integrated Resonant Gravimetric Sensors Using Fundamental Bending Modes

Microcantilever vibrating at its fundamental bending mode is the most common type of resonant mass sensor. Figure 2b shows a typical cantilever mass sensor with integrated thermoelectric excitation and Wheatstone bridge readout components [18]. As illustrated in Figure 2a, 4 piezoresistors and 1 heating resistor near the fixed end of the cantilever are created by boron-doping through thermally grown silicon oxide (SiO2) windows. With patterned aluminum interconnects covered by PECVD SiO2 passivation layer and Au/Cr sensing pad at the free end, the cantilever is finally released by the backside deep reactive ion etching (DRIE), followed by the removal of SiO2 by hydrofluoric acid (HF). The heating power with a DC + AC voltage is given by P = ( V D C + V A C cos ω t ) 2 / R , where R is the resistance of the piezoresistive heater, ω = 2 π f , where f is the driving frequency. One can perceive that the mechanical resonance is excited by the V D C V A C cos ω t / R component (given that V D C > > V A C ). The sensing region at the cantilever free end is covered by an Au film for subsequent chemical functionalization. Such resonant cantilevers with dimensions of 200 × 100 × 3 μm have exhibited fundamental-mode resonant frequencies f ~100 kHz, and quality factors Q > 100. With a mass sensitivity ~1.5 Hz/pg and mass resolution ~0.1 pg, various hazardous chemical vapors, e.g., Trinitrotoluene (TNT), dimethyl methyl phosphonate (DMMP), VOCs and biological species (e.g., bacteria, proteins, DNA) with very low concentrations have been successfully detected.
It is worth noting that with all-electrical integration, a resonator is readily connected to a phase-lock loop (PLL) for real-time frequency tracking. Hence, the minimum detectable mass is defined by the minimum detectable phase change of the system, given by δ m = ( δ θ / Q ) M [30]. Therefore, the mass resolution is significantly dependent on the Q factor, which is inevitably deteriorated by the strong viscous damping in air and liquid. Therefore, researchers have been exploring other high-order resonance modes to improve the Q factor and mass sensing performance.

2.2. Integrated Resonant Gravimetric Sensors Using Higher-Order Bending Modes

The enhancement in Q factor and mass sensing performance when using high-order flexural modes has been demonstrated in many cases [9,27,31,32,33,34,35]. Taking cantilever resonators in air as an example, a ~5-fold increase in Q factors, ~6-fold increase in mass sensitivity, and >8-folder improvement in mass resolution have been demonstrated using the 2nd-order bending mode than using fundamental mode (as shown in Table 1). The increase in mass sensitivity (from 0.43 to 2.7 Hz/pg) can be attributed to the higher resonant frequency of 2nd-order bending mode (~298 kHz), which is ~6-folder higher than that of the fundamental mode (~47.8 kHz). The improvement in Q factor (from 867 to 168) has been analyzed from the perspective of flow pattern over the width of the cantilever. Using finite element simulation, less energy dissipations into the viscous media have been visualized for high order bending modes, leading to higher Qs [28,36,37].

2.3. Integrated Resonant Gravimetric Sensors Using Torsional Modes

Torsional modes have also been exploited to improve the sensing performance of resonant mass sensors [17,22,28,38]. Figure 3b shows a typical T-shaped torsional-mode cantilever with on-chip integrated electromagnetic excitation and piezoresistive readout components [17]. The fabrication process of torsional-mode resonators is similar to that of bending-mode resonators. 4 piezoresistors are formed by boron-doping using SiO2 mask. Aluminum is patterned not only as the interconnects but also as the metal loop for electromagnetic excitation. After the Cr/Au sensing pads are defined by liftoff, the device is released by both frontside RIE, backside DRIE, and SiO2 removal by HF. Together with a decent increase in Q factor, Xia et al. have reported a significant enhancement in mass sensing performance using torsional modes. The >10-fold increase in mass sensitivity is attributed to the much higher resonant frequencies of the torsional modes than that of the fundamental mode. Benefited from high-order torsional vibrations, the mass detection limit has been improved by almost 30 times, down to 9 fg in air [22].

2.4. Integrated Resonant Gravimetric Sensors Using In-Plane Modes

Resonant gravimetric sensors using in-plane modes have been studied [26,39,40], as shown in Figure 4b. To achieve effective excitation and detection of in-plane vibration, thin beam legs are doped to form resistors for electrothermal driving and piezoresistive readout. The fabrication process is similar to that for bending-mode cantilevers, as shown in Figure 4a. An additional oxygen annealing step is often performed to protect the vertical sidewalls of the cantilever and tiny beams against electric leakage in conductive solutions since these devices are often designed for bio/chemical sensing in liquid. An important merit of using in-plane mode is the high Q factors >2000 in air, which is >10-fold better than that of the fundamental mode (as shown in Table 1). As this type of resonator is often used for biosensing applications in liquid, the sensing performance will be detailed in Section 3.3.

2.5. Integrated Resonant Gravimetric Sensors Using Extensional Modes

Extensional-mode resonators typically have ‘dog-bone’ structure [20,29,41], which is quite different from a cantilever, as shown in Figure 5. Two large sensing pads are connected by 2 or 3 thin beams and vibrate oppositely along the device length. The piezoresistive arms are designed for electrothermal excitation and piezoresistive readout of the extensional mode. The fabrication process of the ‘dog-bone’ resonator is quite similar to that of in-plane mode cantilevers. Such mass sensors using bulk mode surpass the aforementioned bending, torsional, and in-plane mode resonators in terms of Q factor and mass sensitivity. Q factors > 11,000 and mass sensitivity up to 10.6 Hz/pg have been reported, which are ~65-time and ~25-time higher than those from fundamental-mode cantilevers (refer to Table 1).
Overall, resonant mass sensors, with integrated excitation and readout schemes and enhanced mass sensing performance by different vibrational modes, have opened up new possibilities for ultrasensitive bio/chemical detection in gas- and liquid phases, which will be detained in the following section.

3. Integrated Resonant Gravimetric Sensors for Bio/Chemical Detection

3.1. Integrated Resonant Gravimetric Sensors for Gas Detection

The ability to detect trace chemical vapors (e.g., TNT, DMMP), VOCs (e.g., aniline, xylene), and other pollutant gases (e.g., carbon monoxide, CO, sulfur oxide, SO2) is highly demanded in environmental protection, industrial pollution control, biomedical systems, and public safety. However, the very low concentration of target gases, interfering gases, and potable sensor design have imposed great challenges in on-the-spot, rapid detection. Therefore, great attention has been made to engineer integrated resonant gravimetric sensors for gas sensing, thanks to their small size, ultra-high sensitivity, and scalability for mass production.
Figure 6a shows an example of a cantilever gas sensor, with nanomaterials (hexafluoro-2-propanol-functionalized mesoporous silica, HFMS, shown in the inset) loaded to the sensing region near the free end [42]. The cantilever is electrothermally excited at its fundamental mode, and the resonant frequency is monitored in real-time using PLL control. When the target gas molecules flow over the cantilever (e.g., 45 ppt, 90 ppt, 135 ppt TNT in Figure 6b), the molecules are adsorbed onto the nanomaterials, hence decrease the resonant frequency. The presence of functionalized nanomaterials can greatly improve the sensor selectivity of the target gas. As shown in Figure 6c, the HFMS-based cantilever resonator is highly responsive to TNT than other interference gases.
So far, resonant cantilever sensors have shown great potential in detecting many kinds of trace analytes, such as chemical vapors, VOCs, and pollutant gases, as summarized in Table 2. In general, the cantilever mass sensors exhibit detection limits from ~100 pb to ~10 ppt level for detecting chemical vapors, such as TNT, DMMA, TMA. The detection limit for VOCs falls in ~1 ppm–100 ppb level, such as aniline, xylene. As to pollutant gases, such as CO, SO2, the minimum detectable concentrations as low as ~10 ppb have been demonstrated. The cantilever resonators also promise fast response time from tens of seconds to ~10 min, thanks to the optimized design of resonators and nanomaterials.

3.2. Integrated Resonant Gravimetric Sensors for Biosensing in Air

The detection of biological species, such as bacteria, viruses, and proteins is of great importance for disease diagnosis, food safety, and fundamental research. Luckily, some of these species can survive in humid air, hence they can be detected by resonant mass sensors. In these cases, the resonators are operated in ‘dip-and-dry’ mode, without suffering from strong viscous damping. After sufficient time for immobilization of bacteria or antigen-antibody interactions in solution, the resonators are dried, and the resonant frequencies before and after dipping are recorded.
Using such an approach, Xu et al. have reported the detection of Escherichia coli (E. coli) O157:H7 down to 103 CFU/mL, and Bacillus Anthracis as low as 103 spores/mL using cantilever mass sensors [57]. With an ultra-high mass resolution down to 9 fg, Xia et al. have demonstrated the detection of 60 ng/mL alpha-fetoprotein (AFP) using torsional-mode resonators [22].

3.3. Resonant Gravimetric Sensors for Bio/Chemical Detection in Liquid

Although the ‘dip-and-dry’ method has been proved to be effective for certain cases that avert device immersion in liquid, direct sensing in liquid is highly desirable, especially for biological applications since most biological processes take place in liquid. The much stronger viscous damping in liquid than in air has imposed great challenges in detecting trace bio/chemical analytes in liquid using resonant gravimetric sensors. For example, fundamental-mode cantilever resonators typically exhibit Qs ~10–20 when operated in liquid (e.g., water, phosphate buffered saline, PBS) (refer to Table 3). Still, prostate-specific antigen (PSA), C-reactive protein (CRP), DNA with concentrations between 10 μg/mL to 10 ng/mL are detectable using such gravimetric sensors.
Several approaches have been taken to improve the sensing performance of resonant mass sensors in liquid. First, as mentioned in Section 2, Q factor and mass sensitivity can be improved by exploiting vibrational modes beyond fundamental mode. Figure 7 shows an example of monitoring heavy metal-ion (Hg2+) in water (mimicking ion pollution to water resource) using extensional-mode ‘dog-bone’ resonators [20]. Q factor ~256 and mass sensitivity ~9.76 Hz/pg have been observed, which are >10-time and ~100-time higher than those of fundamental-mode cantilevers. With –SH modified mesoporous silica loaded to the sensing regions, 500 ppb Hg2+ can be easily discerned from frequency response (as shown in Figure 7b). Second, new device structures can be engineered to isolate the resonators from strong liquid damping. For example, suspended microchannel resonators, SMRs [58,59,60,61,62,63,64], have drawn considerable attention because of their unique way of minimizing viscous damping by fabricating fluidic channels inside the cantilevers. Therefore, these devices are operated in vacuum, allowing ultra-high Qs on the order of 1000–10,000, and unprecedented mass resolutions < 1 ag. Although this type of device has yet been on-chip fully integrated (resonances are often excited by off-chip piezoshaker, and detected by optical lever), and shown limitations in cases, for example, monitoring adherent mass rather than floating mass, they still hold the records for mass sensing in liquid using cantilevers.
Alternatively, resonant mass sensors can be directly immersed in liquid with decent Qs with the help of hydrophobic shells. Such a sensor platform is more suitable for measuring adherent masses, such as adherent cells and molecular binding. Yu et al. have shown the detection of E. coli. down to 102 CFU/mL using a parylene-shell encased bending-mode cantilevers [65]. More recently, Wang et al. have extended such a technique to in-plane cantilevers [66]. Q factor has been significantly improved to >200, and mass sensitivity has increased to 1.2 Hz/pg. Both are one order of magnitude better than conventional bending-mode cantilevers (refer to Table 3).

4. NEMS Resonators for Ultrasensitive Gravimetric Sensing

To continuously pushing the detection limit of the integrated mass sensors, the device sizes have been miniaturized to the nanoscale. For example, nanobeams, nanowires, and nanotubes have shown effective masses reduced by orders of magnitude to attogram. Subsequently, the resonance frequencies have increased by orders of magnitude up to 1 GHz. These devices have been used for atomic and molecular level mass sensing in vacuum, e.g., Au [8], Xe [12,69], Cr [14] atoms, C10H8 molecule [16], bovine serum albumin (BSA), and β-amylase [15].
However, it worth noting that the limit of detection at single-molecule/atom level has been achieved through operating those NEMS resonators at very-high/ultra-high frequency (VHF/UHF) bands, but in stringent experimental conditions (e.g., high vacuum and low temperature). Meanwhile, the quality (Q) factor decreases as the device sizes get smaller, which is due to the increased energy dissipation at the nanoscale than at the microscale.
Some mass sensing attempts have been made with nanocantilevers in air (as shown in Figure 8), which indeed show exquisite gravimetric sensitivities down to ~0.7 Hz/zg and mass resolution ~0.1 ag [70,71]. More attention still needs to be made to tackle the challenges of detecting VHF/UHF resonances with picometer displacement sensitivity while maintaining sufficient sensing area and decent Qs factors in air and liquid, such that the NEMS resonators can be widely used for bio/chemical sensing applications in gas and liquid-phases.

5. Conclusions

This paper reviews the development of integrated resonant gravimetric resonators for bio/chemical sensing applications in the past two decades. Bending-, torsional-, in-plane-, and extensional-mode resonators have been studied to enhance the mass sensing performance in viscous media. Thanks to the ultra-high mass sensitivities (typically ~Hz/pg) and mass resolution (pg to fg), trace gas molecules and biological species using integrated resonant mass sensors have been demonstrated. More attention still needs to be made to tackle the challenges of detecting VHF/UHF resonances with picometer displacement sensitivity in air and liquid, such that the nanoresonators (including nanocantilevers and those made of low dimensional materials [14,69,72,73] can be widely used for bio/chemical sensing applications in gas and liquid-phases. In summary, the resonant MEMS/NEMS gravimetric sensors hold promise to continuously push the bio/chemical detection limits and bring a better understanding of gas/nanomaterial interaction and molecular binding mechanisms.

Author Contributions

H.J. prepared the manuscript. P.X. and X.L. reviewed and edited the manuscript. All authors have read and agreed to the published version of the manuscript.

Funding

The authors acknowledge funding support from the National Key R&D Program of China (2020YFB2008603), Shanghai Pujiang Program (20PJ1415600), Key Research Program of Frontier Sciences of the Chinese Academy of Sciences (QYZDJ-SSW-JSC001), Shanghai “Road and Belt” International Young Scientist Exchange Program (19510744600).

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Lavrik, N.V.; Sepaniak, M.J.; Datskos, P. Cantilever transducers as a platform for chemical and biological sensors. Rev. Sci. Instrum. 2004, 75, 2229–2253. [Google Scholar] [CrossRef]
  2. Arlett, J.; Myers, E.; Roukes, M. Comparative advantages of mechanical biosensors. Nat. Nanotechnol. 2011, 6, 203–215. [Google Scholar] [CrossRef] [Green Version]
  3. Johnson, B.N.; Mutharasan, R. Biosensing using dynamic-mode cantilever sensors: A review. Biosens. Bioelectron. 2012, 32, 1–18. [Google Scholar] [CrossRef]
  4. Fritz, J.; Baller, M.K.; Lang, H.P.; Rothuizen, H.; Vettiger, P.; Meyer, E.; Güntherodt, H.-J.; Gerber, C.; Gimzewski, J.K. Translating biomolecular recognition into nanomechanics. Science 2000, 288, 316–318. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  5. Thundat, T.; Wachter, E.A.; Sharp, S.L.; Warmack, R.J. Detection of mercury vapor using resonating microcantilevers. Appl. Phys. Lett. 1995, 66, 1695–1697. [Google Scholar] [CrossRef]
  6. Li, X.; Lee, D.-W. Integrated microcantilevers for high-resolution sensing and probing. Meas. Sci. Technol. 2011, 23, 022001. [Google Scholar] [CrossRef]
  7. Ono, T.; Li, X.; Miyashita, H.; Esashi, M. Mass sensing of adsorbed molecules in sub-picogram sample with ultrathin silicon resonator. Rev. Sci. Instrum. 2003, 74, 1240–1243. [Google Scholar] [CrossRef]
  8. Ekinci, K.L.; Huang, X.M.H.; Roukes, M.L. Ultrasensitive nanoelectromechanical mass detection. Appl. Phys. Lett. 2004, 84, 4469–4471. [Google Scholar] [CrossRef] [Green Version]
  9. Dohn, S.; Sandberg, R.K.; Svendsen, W.E.; Boisen, A. Enhanced functionality of cantilever based mass sensors using higher modes. Appl. Phys. Lett. 2005, 86, 233501. [Google Scholar] [CrossRef] [Green Version]
  10. Ilic, B.; Yang, Y.; Aubin, K.; Reichenbach, R.; Krylov, S.; Craighead, H.G. Enumeration of DNA molecules bound to a nanomechanical oscillator. Nano Lett. 2005, 5, 925–929. [Google Scholar] [CrossRef]
  11. Nishio, M.; Sawaya, S.; Akita, S.; Nakayama, Y. Carbon nanotube oscillators toward zeptogram detection. Appl. Phys. Lett. 2005, 86, 133111. [Google Scholar] [CrossRef]
  12. Yang, Y.T.; Callegari, C.; Feng, X.L.; Ekinci, K.L.; Roukes, M.L. Zeptogram-scale nanomechanical mass sensing. Nano Lett. 2006, 6, 583–586. [Google Scholar] [CrossRef] [PubMed]
  13. Jensen, K.H.; Kim, K.; Zettl, A. An atomic-resolution nanomechanical mass sensor. Nat. Nanotechnol. 2008, 3, 533–537. [Google Scholar] [CrossRef]
  14. Lassagne, B.; Garcia-Sanchez, D.; Aguasca, A.; Bachtold, A. ultrasensitive mass sensing with a nanotube electromechanical resonator. Nano Lett. 2008, 8, 3735–3738. [Google Scholar] [CrossRef]
  15. Naik, A.; Hanay, M.S.; Hiebert, W.K.; Feng, X.L.; Roukes, M.L. Towards single-molecule nanomechanical mass spectrometry. Nat. Nanotechnol. 2009, 4, 445–450. [Google Scholar] [CrossRef]
  16. Chaste, J.; Eichler, A.; E. Moser, J.; Ceballos, G.; Rurali, R.; Bachtold, A. A nanomechanical mass sensor with yoctogram resolution. Nat. Nanotechnol. 2012, 7, 301–304. [Google Scholar] [CrossRef]
  17. Jin, D.; Li, X.; Bao, H.; Zhang, Z.; Wang, Y.; Yu, H.; Zuo, G. Integrated cantilever sensors with a torsional resonance mode for ultraresoluble on-the-spot bio/chemical detection. Appl. Phys. Lett. 2007, 90, 041901. [Google Scholar] [CrossRef]
  18. Yu, H.; Li, X.; Gan, X.; Liu, Y.; Liu, X.; Xu, P.; Li, J.; Liu, M. Resonant-cantilever bio/chemical sensors with an integrated heater for both resonance exciting optimization and sensing repeatability enhancement. J. Micromech. Microeng. 2009, 19, 045023. [Google Scholar] [CrossRef]
  19. Cai, S.; Li, W.; Xu, P.; Xia, X.; Yu, H.; Zhang, S.; Li, X. In situ construction of metal-organic framework (MOF) UiO-66 film on parylene-patterned resonant microcantilever for trace organophosphorus molecules detection. Analyst 2019, 144, 3729–3735. [Google Scholar] [CrossRef]
  20. Yu, F.; Wang, J.; Xu, P.; Li, X. A tri-beam dog-bone resonant sensor with high-Q in liquid for disposable test-strip detection of analyte droplet. J. Microelectromech. Syst. 2016, 25, 244–251. [Google Scholar] [CrossRef]
  21. Mehdizadeh, E.; Chapin, J.; Gonzales, J.; Rahafrooz, A.; Abdolvand, R.; Purse, B.; Pourkamali, S.; Chapin, J. Direct detection of biomolecules in liquid media using piezoelectric rotational mode disk resonators. In Proceedings of the 2012 IEEE Sensors, Taipei, Taiwan, 28–31 October 2012; pp. 1–4. [Google Scholar] [CrossRef]
  22. Xia, X.; Zhang, Z.; Li, X. A Latin-cross-shaped integrated resonant cantilever with second torsion-mode resonance for ultra-resoluble bio-mass sensing. J. Micromech. Microeng. 2008, 18, 035028. [Google Scholar] [CrossRef]
  23. Hwang, K.S.; Lee, J.H.; Park, J.; Yoon, D.S.; Park, J.H.; Kim, T.S. In-situ quantitative analysis of a prostate-specific antigen (PSA) using a nanomechanical PZT cantilever. Lab Chip 2004, 4, 547–552. [Google Scholar] [CrossRef]
  24. Kwon, T.Y.; Eom, K.; Park, J.H.; Yoon, D.S.; Kim, T.S.; Lee, H.L. In situ real-time monitoring of biomolecular interactions based on resonating microcantilevers immersed in a viscous fluid. Appl. Phys. Lett. 2007, 90, 223903. [Google Scholar] [CrossRef]
  25. Kwon, T.; Eom, K.; Park, J.; Yoon, D.S.; Lee, H.L.; Kim, T.S. Micromechanical observation of the kinetics of biomolecular interactions. Appl. Phys. Lett. 2008, 93, 173901. [Google Scholar] [CrossRef]
  26. Tao, Y.; Li, X.; Xu, T.; Yu, H.; Xu, P.; Xiong, B.; Wei, C. Resonant cantilever sensors operated in a high-Q in-plane mode for real-time bio/chemical detection in liquids. Sens. Actuators B Chem. 2011, 157, 606–614. [Google Scholar] [CrossRef]
  27. Jin, D.; Li, X.; Liu, J.; Zuo, G.; Wang, Y.; Liu, M.; Yu, H. High-mode resonant piezoresistive cantilever sensors for tens-femtogram resoluble mass sensing in air. J. Micromech. Microeng. 2006, 16, 1017–1023. [Google Scholar] [CrossRef]
  28. Xia, X.; Li, X. Resonance-mode effect on microcantilever mass-sensing performance in air. Rev. Sci. Instrum. 2008, 79, 074301. [Google Scholar] [CrossRef] [PubMed]
  29. Yu, F.; Yu, H.; Xu, P.; Li, X. Precise extension-mode resonant sensor with uniform and repeatable sensitivity for detection of ppm-level ammonia. J. Micromech. Microeng. 2014, 24, 045005. [Google Scholar] [CrossRef]
  30. Ekinci, K.L.; Yang, Y.T.; Roukes, M.L. Ultimate limits to inertial mass sensing based upon nanoelectromechanical systems. J. Appl. Phys. 2004, 95, 2682–2689. [Google Scholar] [CrossRef] [Green Version]
  31. Ghatkesar, M.K.; Barwich, V.; Braun, T.; Ramseyer, J.-P.; Gerber, C.; Hegner, M.; Lang, H.P.; Drechsler, U.; Despont, M. Higher modes of vibration increase mass sensitivity in nanomechanical microcantilevers. Nanotechnology 2007, 18, 445502. [Google Scholar] [CrossRef] [Green Version]
  32. Jia, H.; Lee, J.; Wang, Z.; Feng, P.X.-L. High-frequency SiC microdisk resonators operating in water with responses to H2O2 and NH4OH. In Proceedings of the 2014 IEEE International Frequency Control Symposium (FCS), IEEE, Taipei, Taiwan, 19–22 May 2014; pp. 1–4. [Google Scholar] [CrossRef]
  33. Jia, H.; Wu, X.; Tang, H.; Lu, Z.-R.; Feng, P.X.-L. Culturing and probing physical behavior of individual breast cancer cells on SiC microdisk resonators. In Proceedings of the 2015 IEEE International Conference on Micro Electro Mechanical Systems (MEMS), Estoril, Portugal, 18–22 January 2015; pp. 698–701. [Google Scholar] [CrossRef]
  34. Jia, H.; Lu, X.; Main, C.; Lin, Q.; Feng, P.X.-L. Mode-dependent anchor loss in silicon carbide micromechanical disk resonators. In Proceedings of the 2019 Joint Conference of the IEEE International Frequency Control Symposium and European Frequency and Time Forum (EFTF/IFC), Orlando, FL, USA, 14–18 April 2019; pp. 1–2. [Google Scholar] [CrossRef]
  35. Chen, H.; Jia, H.; Zorman, C.A.; Feng, P.X.L. Determination of elastic modulus of silicon carbide (SiC) thin diaphragms via mode-dependent Duffing non-linear resonances. J. Microelectromech. Syst. 2020, 29, 783–789. [Google Scholar] [CrossRef]
  36. Jia, H.; Feng, P.X.-L. Very high-frequency silicon carbide microdisk resonators with multimode responses in water for particle sensing. J. Microelectromech. Syst. 2019, 28, 941–953. [Google Scholar] [CrossRef]
  37. Jia, H.; Feng, P.X.-L. Tracing and resolving microparticle aquatic mass motion and distribution on multimode silicon carbide microdisk resonators. In Proceedings of the 2019 IEEE 32nd International Conference on Micro Electro Mechanical Systems (MEMS), Seoul, Korea, 27–31 January 2019; pp. 529–532. [Google Scholar] [CrossRef]
  38. Liu, Y.; Li, X.; Zhang, Z.; Zuo, G.; Cheng, Z.; Yu, H. Nanogram per milliliter-level immunologic detection of alpha-fetoprotein with integrated rotating-resonance microcantilevers for early-stage diagnosis of hepatocellular carcinoma. Biomed. Microdevices 2008, 11, 183–191. [Google Scholar] [CrossRef] [PubMed]
  39. Xu, T.; Yu, H.; Xu, P.; Xu, W.; Chen, W.; Chen, C.; Li, X. Real-time enzyme-digesting identification of double-strand DNA in a resonance-cantilever embedded microchamber. Lab Chip 2014, 14, 1206–1214. [Google Scholar] [CrossRef]
  40. Beardslee, L.A.; Addous, A.M.; Heinrich, S.; Josse, F.; Dufour, I.; Brand, O. Thermal excitation and piezoresistive detection of cantilever in-plane resonance modes for sensing applications. J. Microelectromech. Syst. 2010, 19, 1015–1017. [Google Scholar] [CrossRef] [Green Version]
  41. Yu, F.; Xu, P.; Wang, J.; Li, X. Length-extensional resonating gas sensors with IC-foundry compatible low-cost fabrication in non-SOI single-wafer. Microelectron. Eng. 2015, 13, 1–7. [Google Scholar] [CrossRef]
  42. Xu, P.; Yu, H.; Li, X. Functionalized mesoporous silica for microgravimetric sensing of trace chemical vapors. Anal. Chem. 2011, 83, 3448–3454. [Google Scholar] [CrossRef]
  43. Xu, P.; Li, X.; Yu, H.; Liu, M.; Li, J. Self-assembly and sensing-group graft of pre-modified CNTs on resonant micro-cantilevers for specific detection of volatile organic compound vapors. J. Micromech. Microeng. 2010, 20, 115003. [Google Scholar] [CrossRef]
  44. Xu, P.; Yu, H.; Li, X. In situ growth of noble metal nanoparticles on graphene oxide sheets and direct construction of functionalized porous-layered structure on gravimetric microsensors for chemical detection. Chem. Commun. 2012, 48, 10784–10786. [Google Scholar] [CrossRef]
  45. Xu, P.; Guo, S.; Yu, H.; Li, X. Mesoporous silica nanoparticles (MSNs) for detoxification of hazardous organophorous chemicals. Small 2014, 10, 2404–2412. [Google Scholar] [CrossRef]
  46. Guo, S.; Xu, P.; Yu, H.; Li, X.; Cheng, Z. Hyper-branch sensing polymer batch self-assembled on resonant micro-cantilevers with a coupling-reaction route. Sens. Actuators B Chem. 2015, 209, 943–950. [Google Scholar] [CrossRef]
  47. Guo, S.; Xu, P.; Yu, H.; Cheng, Z.; Li, X. Synergistic improvement of gas sensing performance by micro-gravimetrically extracted kinetic/thermodynamic parameters. Anal. Chim. Acta 2015, 863, 49–58. [Google Scholar] [CrossRef] [PubMed]
  48. Yu, H.; Yang, T.; Chen, Y.; Xu, P.; Lee, D.-W.; Li, X. Chemo-mechanical joint detection with both dynamic and static microcantilevers for interhomologue molecular identification. Anal. Chem. 2012, 84, 6679–6685. [Google Scholar] [CrossRef] [PubMed]
  49. Yu, H.; Xu, P.; Lee, D.-W.; Li, X. Porous-layered stack of functionalized AuNP–rGO (gold nanoparticles–reduced graphene oxide) nanosheets as a sensing material for the micro-gravimetric detection of chemical vapor. J. Mater. Chem. A 2013, 1, 4444–4450. [Google Scholar] [CrossRef]
  50. Xia, X.; Guo, S.; Zhao, W.; Xu, P.; Yu, H.; Xu, T.; Li, X. Carboxyl functionalized gold nanoparticles in situ grown on reduced graphene oxide for micro-gravimetric ammonia sensing. Sens. Actuators B Chem. 2014, 202, 846–853. [Google Scholar] [CrossRef]
  51. Liu, M.; Guo, S.; Xu, P.; Yu, H.; Xu, T.; Zhang, S.; Li, X. Revealing humidity-enhanced NH3 sensing effect by using resonant microcantilever. Sens. Actuators B Chem. 2018, 257, 488–495. [Google Scholar] [CrossRef]
  52. Lv, Y.; Yu, H.; Xu, P.; Xu, J.; Li, X. Metal organic framework of MOF-5 with hierarchical nanopores as micro-gravimetric sensing material for aniline detection. Sens. Actuators B Chem. 2018, 256, 639–647. [Google Scholar] [CrossRef]
  53. Xu, T.; Xu, P.; Zheng, D.; Yu, H.; Li, X. Metal–organic frameworks for resonant-gravimetric detection of trace-level xylene molecules. Anal. Chem. 2016, 88, 12234–12240. [Google Scholar] [CrossRef] [PubMed]
  54. Lv, Y.; Xu, P.; Yu, H.; Xu, J.; Li, X. Ni-MOF-74 as sensing material for resonant-gravimetric detection of ppb-level CO. Sens. Actuators B Chem. 2018, 262, 562–569. [Google Scholar] [CrossRef]
  55. Wang, X.; Yao, F.; Xu, P.; Li, M.; Yu, H.; Li, X. Quantitative structure-activity relationship of nanowire adsorption to SO2 revealed by in situ TEM technique. Nano Lett. 2021, 21, 1679–1687. [Google Scholar] [CrossRef]
  56. Yu, H.; Xu, P.; Xia, X.; Lee, D.-W.; Li, X. Micro-/nanocombined gas sensors with functionalized mesoporous thin film self-assembled in batches onto resonant cantilevers. IEEE Trans. Ind. Electron. 2011, 59, 4881–4887. [Google Scholar] [CrossRef]
  57. Xu, T.; Yu, H.; Xu, P.; Li, X. A chelating-bond breaking and re-linking technique for rapid re-immobilization of immune micro-sensors. Biomed. Microdevices 2011, 14, 303–311. [Google Scholar] [CrossRef] [PubMed]
  58. Olcum, S.; Cermak, N.; Wasserman, S.C.; Christine, K.S.; Atsumi, H.; Payer, K.R.; Shen, W.; Lee, J.; Belcher, A.M.; Bhatia, S.N.; et al. Weighing nanoparticles in solution at the attogram scale. Proc. Natl. Acad. Sci. USA 2014, 111, 1310–1315. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  59. Lee, J.; Shen, W.; Payer, K.; Burg, T.P.; Manalis, S.R. Toward attogram mass measurements in solution with suspended nanochannel resonators. Nano Lett. 2010, 10, 2537–2542. [Google Scholar] [CrossRef] [Green Version]
  60. Godin, M.; Delgado, F.F.; Son, S.; Grover, W.H.; Bryan, A.K.; Tzur, A.; Jorgensen, P.; Payer, K.; Grossman, A.D.; Kirschner, M.W.; et al. Using buoyant mass to measure the growth of single cells. Nat. Methods 2010, 7, 387–390. [Google Scholar] [CrossRef] [Green Version]
  61. Olcum, S.; Cermak, N.; Wasserman, S.C.; Payer, K.; Shen, W.; Lee, J.; Manalis, S.R. Suspended nanochannel resonators at attogram precision. In Proceedings of the 2014 IEEE International Conference on Micro Electro Mechanical Systems (MEMS), San Francisco, CA, USA, 26–30 January 2014; pp. 116–119. [Google Scholar] [CrossRef] [Green Version]
  62. Barton, R.A.; Ilic, B.; Verbridge, S.S.; Cipriany, B.R.; Parpia, J.M.; Craighead, H.G. Fabrication of a nanomechanical mass sensor containing a nanofluidic channel. Nano Lett. 2010, 10, 2058–2063. [Google Scholar] [CrossRef] [PubMed]
  63. Burg, T.P.; Mirza, A.R.; Milovic, N.; Tsau, C.H.; Popescu, G.A.; Manalis, S.R.; Foster, J.S. Vacuum-packaged suspended microchannel resonant mass sensor for biomolecular detection. J. Microelectromech. Syst. 2006, 15, 1466–1476. [Google Scholar] [CrossRef]
  64. Cermak, N.; Olcum, S.; Delgado, F.F.; Wasserman, S.C.; Payer, K.R.; A. Murakami, M.; Knudsen, S.M.; Kimmerling, R.J.; Stevens, M.M.; Kikuchi, Y.; et al. High-throughput measurement of single-cell growth rates using serial microfluidic mass sensor arrays. Nat. Biotechnol. 2016, 34, 1052–1059. [Google Scholar] [CrossRef] [Green Version]
  65. Yu, H.; Chen, Y.; Xu, P.; Xu, T.; Bao, Y.; Li, X. μ-‘Diving suit’ for liquid-phase high-Q resonant detection. Lab Chip 2016, 16, 902–910. [Google Scholar] [CrossRef]
  66. Wang, X.; Cheng, Y.; Cai, S.; Chen, J.; Xu, P.; Chen, Y.; Yu, H.; Xu, T.; Zhang, S.; Li, X. Resonant-cantilever-detected kinetic/thermodynamic parameters for aptamer–ligand binding on a liquid-solid interface. Anal. Chem. 2020, 92, 11127–11134. [Google Scholar] [CrossRef]
  67. Vancura, C.; Li, Y.; Lichtenberg, J.; Kirstein, K.U.; Hierlemann, A.; Josse, F. Liquid-phase chemical and biochemical detection using fully integrated magnetically actuated complementary metal oxide semiconductor resonant cantilever sensor systems. Anal. Chem. 2007, 79, 1646–1654. [Google Scholar] [CrossRef]
  68. Li, Y.; Vancura, C.; Kirstein, K.-U.; Lichtenberg, J.; Hierlemann, A. monolithic resonant-cantilever-based CMOS microsystem for biochemical sensing. IEEE Trans. Circuits Syst. I Regul. Pap. 2008, 55, 2551–2560. [Google Scholar] [CrossRef]
  69. Chiu, H.-Y.; Hung, P.; Postma, H.W.C.; Bockrath, M. Atomic-scale mass sensing using carbon nanotube resonators. Nano Lett. 2008, 8, 4342–4346. [Google Scholar] [CrossRef] [PubMed]
  70. Li, M.; Tang, H.X.; Roukes, M.L. Ultra-sensitive NEMS-based cantilevers for sensing, scanned probe and very high-frequency applications. Nat. Nanotechnol. 2007, 2, 114–120. [Google Scholar] [CrossRef] [PubMed]
  71. Yang, Y.; Xia, X.; Gan, X.; Xu, P.; Yu, H.; Li, X. Nano-thick resonant cantilevers with a novel specific reaction-induced frequency-increase effect for ultra-sensitive chemical detection. J. Micromech. Microeng. 2010, 20, 055022. [Google Scholar] [CrossRef]
  72. Jia, H.; Yang, R.; Nguyen, A.E.; Alvillar, S.N.; Empante, T.; Bartels, L.; Feng, P.X.-L. Large-scale arrays of single- and few-layer MoS2nanomechanical resonators. Nanoscale 2016, 8, 10677–10685. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  73. Wang, Z.; Jia, H.; Zheng, X.-Q.; Yang, R.; Ye, G.; Chen, X.; Feng, P.X.-L. Resolving and tuning mechanical anisotropy in black phosphorus via nanomechanical multimode resonance spectromicroscopy. Nano Lett. 2016, 16, 5394–5400. [Google Scholar] [CrossRef] [PubMed] [Green Version]
Figure 1. Integrated resonant gravimetric sensors using different vibrational modes, such as bending, torsional, in-plane, and extensional modes. Reprinted with permission from [17,19,20,26]. Copyright 2007 American Institute of Physics, 2019 Royal Society of Chemistry, 2016 IEEE, 2011 Elsevier B.V.
Figure 1. Integrated resonant gravimetric sensors using different vibrational modes, such as bending, torsional, in-plane, and extensional modes. Reprinted with permission from [17,19,20,26]. Copyright 2007 American Institute of Physics, 2019 Royal Society of Chemistry, 2016 IEEE, 2011 Elsevier B.V.
Micromachines 12 00645 g001
Figure 2. Bending-mode cantilever mass sensors. (af) Fabrication process of microcantilevers showing on-chip integration of electrothermal excitation and piezoresistive readout. (g) Typical resonant sensor with sensing region at the free end for nanomaterial loading. Reprinted with permission from [18]. Copyright 2009 IOP Publishing Ltd.
Figure 2. Bending-mode cantilever mass sensors. (af) Fabrication process of microcantilevers showing on-chip integration of electrothermal excitation and piezoresistive readout. (g) Typical resonant sensor with sensing region at the free end for nanomaterial loading. Reprinted with permission from [18]. Copyright 2009 IOP Publishing Ltd.
Micromachines 12 00645 g002
Figure 3. Torsional-mode cantilever mass sensors. (ae) Fabrication process of the microcantilevers showing on-chip integration of electromagnetic excitation and piezoresistive readout. (f) Typical resonant sensor with 2 sensing regions for nanomaterial loading. Reprinted with permission from [17]. Copyright 2007 American Institute of Physics.
Figure 3. Torsional-mode cantilever mass sensors. (ae) Fabrication process of the microcantilevers showing on-chip integration of electromagnetic excitation and piezoresistive readout. (f) Typical resonant sensor with 2 sensing regions for nanomaterial loading. Reprinted with permission from [17]. Copyright 2007 American Institute of Physics.
Micromachines 12 00645 g003
Figure 4. In-plane-mode resonant microcantilever mass sensors. (ad) Fabrication process of microcantilevers showing on-chip integration of electrothermal excitation and piezoresistive readout. (e) Typical resonant sensor with sensing region for loading functional nanomaterials. Reprinted with permission from [26]. Copyright 2011 Elsevier B.V.
Figure 4. In-plane-mode resonant microcantilever mass sensors. (ad) Fabrication process of microcantilevers showing on-chip integration of electrothermal excitation and piezoresistive readout. (e) Typical resonant sensor with sensing region for loading functional nanomaterials. Reprinted with permission from [26]. Copyright 2011 Elsevier B.V.
Micromachines 12 00645 g004
Figure 5. Extensional-mode ‘dog-bone’ mass sensors. (a,b) Two types of devices with sensing pads connected by dual- and tri-beams. Reprinted with permission from [20]. Copyright 2016 IEEE.
Figure 5. Extensional-mode ‘dog-bone’ mass sensors. (a,b) Two types of devices with sensing pads connected by dual- and tri-beams. Reprinted with permission from [20]. Copyright 2016 IEEE.
Micromachines 12 00645 g005
Figure 6. TNT vapor detection using a microcantilever sensor. (a) SEM images showing a microcantilever sensor with loaded HEMS nanomaterials. (b) Frequency responses of the microcantilever to TNT vapors at different contractions. (c) Sensor responses to various kinds of interfering gases compared with 380 ppt TNT vapor. Reprinted with permission from [42]. Copyright 2011 American Chemical Society.
Figure 6. TNT vapor detection using a microcantilever sensor. (a) SEM images showing a microcantilever sensor with loaded HEMS nanomaterials. (b) Frequency responses of the microcantilever to TNT vapors at different contractions. (c) Sensor responses to various kinds of interfering gases compared with 380 ppt TNT vapor. Reprinted with permission from [42]. Copyright 2011 American Chemical Society.
Micromachines 12 00645 g006
Figure 7. Detection of Hg2+ in liquid droplets using a tri-beam extensional-mode resonator. (a) SEM images showing a microcantilever sensor with loaded mesoporous silica. (b) Frequency responses of the resonator to 500 ppb and 1 ppm Hg2+ ion in solution. Reprinted with permission from [20]. Copyright 2016 IEEE.
Figure 7. Detection of Hg2+ in liquid droplets using a tri-beam extensional-mode resonator. (a) SEM images showing a microcantilever sensor with loaded mesoporous silica. (b) Frequency responses of the resonator to 500 ppb and 1 ppm Hg2+ ion in solution. Reprinted with permission from [20]. Copyright 2016 IEEE.
Micromachines 12 00645 g007
Figure 8. NEMS cantilevers for ultrasensitive gravimetric detection in air. (a) A T-shaped nano-thick cantilever with torsional resonant mode ~37.65 kHz exhibits mass sensitivity ~860 Hz/pg, and TMA detection limit <0.1 ppm. (b) A SiC nano cantilever with fundamental bending mode in the VHF band (up to 127 MHz), exhibits mass sensitivity ~0.7 Hz/zg, and mass resolution ~0.1 ag. Reprinted with permission from [70,71]. Copyright 2007 Nature Publishing Group, 2010 IOP Publishing Ltd.
Figure 8. NEMS cantilevers for ultrasensitive gravimetric detection in air. (a) A T-shaped nano-thick cantilever with torsional resonant mode ~37.65 kHz exhibits mass sensitivity ~860 Hz/pg, and TMA detection limit <0.1 ppm. (b) A SiC nano cantilever with fundamental bending mode in the VHF band (up to 127 MHz), exhibits mass sensitivity ~0.7 Hz/zg, and mass resolution ~0.1 ag. Reprinted with permission from [70,71]. Copyright 2007 Nature Publishing Group, 2010 IOP Publishing Ltd.
Micromachines 12 00645 g008
Table 1. Performance of typical integrated resonant gravimetric sensors in air using different vibrational modes.
Table 1. Performance of typical integrated resonant gravimetric sensors in air using different vibrational modes.
DevicesResonant FrequencyQ FactorSensing PerformanceRefs
Fundamental Bending-Mode Cantilevers47.838 kHz1680.43 Hz/pg0.26 pg[27]
High-Order Bending-Mode Cantilevers298.132 kHz8672.7 Hz/pg30 fg[27]
Torsional-Mode Cantilevers114.805 kHz2520.9 Hz/pg23 fg[28]
High-Order Torsional-Mode Cantilevers508.082 kHz2865.1 Hz/pg9 fg[28]
In-Plane Mode Cantilevers536 kHz2096//[26]
Extensional-Mode Resonators4.1 MHz1115710.617 Hz/pg0.94 pg[29]
Table 2. Integrated Resonant Gravimetric Sensors for Gas Sensing.
Table 2. Integrated Resonant Gravimetric Sensors for Gas Sensing.
TargetsSensing MaterialsGas Sensing PerformanceRefs
Micromachines 12 00645 i001
TNT Vapor
HFIP-MWCNTs~1.2 ppb~5 min[43]
HFIP-Mesoporous Silica~20.8 ppt~2 min[42]
HFIP-GO/Au-
NPs hybrid
~60 ppt~1 min[44]
Micromachines 12 00645 i002
DMMP Vapor
Siloxane-Hyperbranched
Polymers
~5 ppb~3 min[18]
Mesoporous Silica Nanoparticles<1 ppb<6 min[45]
HB-PCSOX-BHPHFB<300 ppb5 min[46]
BHPF-KIT-5 Mesoporous Silica~30 ppb~10 min[47]
UiO-66 film<5 ppb~10 min[19]
Micromachines 12 00645 i003
TMA Vapor
−COOH-Mesoporous Silica~0.8 ppm~6 min[48]
AuNP–rGO~0.5 ppm<30 s[49]
Micromachines 12 00645 i004
NH3 Gas
−COOH-Au-NPs/rGO<10 ppm~5 min[50]
Carboxyl- Mesoporous Silica Nanoparticles<5 ppb~1.8 min[51]
Micromachines 12 00645 i005
Aniline Vapor
MOF-5<1.4 ppm~1.8 min[52]
Micromachines 12 00645 i006
p-xylene
HKUST-1~120 ppb~15 min[53]
Micromachines 12 00645 i007
CO
Ni-MOF-74<10 ppb~5 min[54]
Micromachines 12 00645 i008
SO2
ZnO Nanowires~70 ppb~10 min[55]
Micromachines 12 00645 i009
CO2 Gas
−NH2-MTF30 ppm<60 s[56]
Table 3. Integrated Resonant Microgravimetric Sensors for Biological Detection in Liquid.
Table 3. Integrated Resonant Microgravimetric Sensors for Biological Detection in Liquid.
DeviceDimensionsfQSensing PerformanceRefs
PZT cantilever300 × 100 × 0.65 μm30.95 kHz--1–100 ng/mL PSA[23]
Si Cantilever150 × 140 × 8.2 µm200 kHz100.02 Hz/pg10–100 ng/mL PSA[67]
PZT Cantilever500 × 500 × 35 μm36.11 kHz15–250.118 Hz/pg10 µg/mL CRP[24]
Si Cantilever150 × 140 × 8.2 µm250 kHz200.1 Hz/pg10–100 ng/mL PSA[68]
PZT Cantilever500 × 500 × 32 μm59 kHz20-100 ng/mL CRP
5 µM ssDNA
[25]
In-Plane Mode
Cantilever
190 × 310 × 5 μm406 kHz148.8 Hz/pg2 × 103 CFU/mL E. Coli
EcoRV-enzyme digestion of dsDNA
[26,39]
Rotational Diskd = 500 μm3.44 MHz20–80-Hybridization between ssDNA (1.0 μM) and HS-ssDNA (2.0 μM)[21]
Encased
Bending-Mode Cantilever
200 × 100 × 3 μm50.615 kHz23-102 CFU/mL E. Coli[65]
Encased In-Plane Mode Cantilever190 × 310 × 3.7 μm576 kHz2081.23 Hz/pg1–10 μM ATP-Aptamer Interaction[66]
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Jia, H.; Xu, P.; Li, X. Integrated Resonant Micro/Nano Gravimetric Sensors for Bio/Chemical Detection in Air and Liquid. Micromachines 2021, 12, 645. https://doi.org/10.3390/mi12060645

AMA Style

Jia H, Xu P, Li X. Integrated Resonant Micro/Nano Gravimetric Sensors for Bio/Chemical Detection in Air and Liquid. Micromachines. 2021; 12(6):645. https://doi.org/10.3390/mi12060645

Chicago/Turabian Style

Jia, Hao, Pengcheng Xu, and Xinxin Li. 2021. "Integrated Resonant Micro/Nano Gravimetric Sensors for Bio/Chemical Detection in Air and Liquid" Micromachines 12, no. 6: 645. https://doi.org/10.3390/mi12060645