Glass-Ceramic Fillers Based on Zinc Oxide–Silica Systems for Dental Composite Resins: Effect on Mechanical Properties

by
Peyman Torkian
1,
SayedMohsen Mortazavi Najafabadi
2,*,
Majid Ghashang
3 and
Dariusz Grzelczyk
2
1
Department of Manufacturing, Faculty of Mechanical Engineering, Babol Noshirvani University of Technology, Babol 47148-71167, Iran
2
Department of Automation, Biomechanics and Mechatronics, Lodz University of Technology, 1/15 Stefanowski Street, 90-537 Lodz, Poland
3
Department of Chemistry, Najafabad Branch, Islamic Azad University, Najafabad 15847-43311, Iran
*
Author to whom correspondence should be addressed.
Materials 2023, 16(18), 6268; https://doi.org/10.3390/ma16186268
Submission received: 9 August 2023 / Revised: 11 September 2023 / Accepted: 16 September 2023 / Published: 19 September 2023

Abstract

:
The potential of glass ceramics as applicable materials in various fields including fillers for dental restorations is our guide to present a new procedure for improvements of the mechanical properties of dental composites. This work aims to use Zn2SiO4 and SiO2–ZnO nano-materials as fillers to improve the mechanical properties of Bis-GMA/TEGDMA mixed dental resins. Zn2SiO4 and SiO2–ZnO samples were prepared and characterized by using XRD, FE-SEM, EDX, and FT-IR techniques. The XRD pattern of the SiO2–ZnO sample shows that ZnO crystallized in a hexagonal phase, while the SiO2 phase was amorphous. Similarly, the Zn2SiO4 sample crystallized in a rhombohedral crystal system. The prepared samples were used as fillers for the improvement of the mechanical properties of Bis-GMA/TEGDMA mixed dental resins. Five samples of dental composites composed of Bis-GMA/TEGDMA mixed resins were filled with 2, 5, 8, 10, and 15 wt% of SiO2–ZnO, and similarly, five samples were filled with Zn2SiO4 samples (2, 5, 8, 10, and 15 wt%). All of the 10 samples (A1A10) were characterized by using different techniques including FT-IR, FE-SEM, EDX, and TGA analyses. According to the TGA analysis, all samples were thermally stable up to 200 °C, and the thermal stability increased with the filler percent. Next, the mechanical properties of the samples including the flexural strength (FS), flexural modulus (FM), diameter tensile strength (DTS), and compressive strength (CS) were investigated. The obtained results revealed that the samples filled with 8 wt% of SiO2–ZnO and 10 wt% of Zn2SiO4 had higher FS values of 123.4 and 136.6 MPa, respectively. Moreover, 8 wt% of both fillers displayed higher values of the FM, DTS, and CS parameters. These values were 8.6 GPa, 34.2 MPa, and 183.8 MPa for SiO2–ZnO and 11.3 GPa, 41.2 MPa, and 190.5 MPa for the Zn2SiO4 filler. Inexpensive silica-based materials enhance polymeric mechanics. Silica–metal oxide nanocomposites improve dental composite properties effectively.

1. Introduction

For years, scientists’ efforts have been focused on providing polymer composites with improved chemical and mechanical properties for applications such as the construction and restoration of dentals. These activities have resulted in products with improved resistance and strength. Polymer composites used in dentistry include three different components: The first component includes organic monomers that form the polymer mixture and usually form the main component of the composite. The second component is the radical initiators, which include a small percentage of the mixture. Finally, the third includes fillers that are usually used to increase the mechanical strength of dental composites [1]. Among the used monomers, bisphenylglycidyl dimethacrylate (Bis-GMA) and triethyleneglycol dimethacrylate (TEGDMA) have been widely used due to their easy polymerization, the high chemical stability of the produced polymers, and finally, their suitable waterproof properties. On the other hand, the main role of fillers is to improve the mechanical properties of the composite, including its tensile and bending strength, and silica-based fillers have been used the most [1].
The development of fillers can be considered the main research field of researchers to provide a method to improve the mechanical properties of dental composites. Revisiting the research conducted in the field of optimizing the properties of dental polymers, there are many interesting reports. In other words, many fillers have been used to improve the properties of dental polymers, some of which are polyacrylonitrile (PAN) nanofibers [2], carbon nanotubes modified by silane coupling agents [3], Ag nanoparticles [4], quaternary ammonium polyethylenimine nanoparticles (QA-PEI-NPs) [5], bioactive glass [6], SiO2 nanofibers [7], alumina particles [8], spherical silica [9], hydroxyapatite whisker [10], nano fibrillar silicate [11], nano-zirconia fillers [12], electrospun nano-scaled glass fiber [13], NaF-loaded core–shell PAN–PMMA nanofibers [14], and APTES- or MPTS-conditioned nanozirconia [15].
In addition, glass ceramics have found many applications in various industries due to their low prices, easy production methods especially on a commercial scale, and high crystallization ability. Notably, glass ceramics composed of ZnO and SiO2 phases have suitable potentials as luminescence materials to be used in various optoelectronic devices; thermal, chemical, and mechanical stability; and excellent transparency for glass phosphorous materials. Glass ceramics are suitable candidates for use as fillers in dental composites due to their high transparency and sufficient strength and biocompatibility with the tooth. So far, many cases of glass ceramics have been used as fillers in dental composites. However, efforts to develop dental fillings based on glass ceramics have never ended [16,17,18].
The aim of this work is to develop two new systems to be used as a filler for dental composite reinforcements. In continuation of our studies on the preparation and application of nanostructured materials [19,20,21,22,23], here Zn2SiO4 and SiO2–ZnO nano-materials were used as resistant reinforcing particles in a dental composite to measure mechanical properties such as flexural strength, etc., after construction.

2. Materials and Methods

2.1. Reagents and Instrumentation

The salts, chemicals, and solvents used for this project were purchased from a commercial seller of Merck Chemicals (Merck Company, Darmstadt, Germany). The Bis-GMA, TEGDMA, CQ, and 3-NDADM were purchased from a commercial seller of Aldrich chemicals (Sigma-Aldrich, St. Louis, MO, USA). The thermal behavior was investigated to confirm the thermal stability of the samples using TGA analysis on a Shimadzu TGA-DTG-60H instrument (Kagoshima, Japan). The FT-IR analysis was performed on a Bruker spectrophotometer (BRUKER TEMSOR27, Borken, Germany) using the KBr disk in the range of 400 to 4000 cm−1. The morphology and chemical composition of the samples were studied by FE-SEM and EDX analysis using a JEOL JSM-IT 100 (Tokyo, Japan) instrument. The mechanical properties were measured on a universal testing machine (Hounsfield, Redhill, UK). The XRD patterns of the Zn2SiO4 and SiO2–ZnO samples were analyzed on a Bruker XRD D8/Advance/axs instrument (Bruker AXS GmbH, Karlsruhe, Germany).

2.2. Preparation of SiO2–ZnO Nano-Composite

To a diluted aqueous solution of liquid glass (25 mL) in ethanol (50 mL), ZnCl2.4H2O (previously dissolved in 50 mL of deionized water) was added drop-wise under magnetic stirring for 30 min and then neutralized using a hydrochloric acid solution (10%). The mixture was slowly combined with aminoethanol (100 mmol) to adjust the pH at 9. The precipitate was filtered, washed with distilled water, and dried and subsequently was calcined at 600 °C in an electric furnace for 2 h. The resultant matter was powdered and analyzed.

2.3. Preparation of Zn2SiO4

To a diluted aqueous solution of liquid glass in ethanol (50 mL), ZnCl2.4H2O (previously dissolved in 50 mL of deionized water) was added drop-wise under magnetic stirring for 30 min and then neutralized using a hydrochloric acid solution (10%). The ratio of Si/Zn was adjusted to 1:1. The mixture was slowly combined with aminoethanol (100 mmol) to adjust the pH at 9. The precipitate was filtered, washed with distilled water, dried, and subsequently calcined at 1200 °C in an electric furnace for 2 h. The resultant matter was powdered and analyzed.

2.4. Sample Preparation

The starting materials including the fillers were mixed with initial ratio percentages depicted in Table 1 and homogenized by an ultrasonic homogenizer. The homogenized mixture was transferred into aluminum alloy molds and vacuumed. Then the light emitting technique using a Coltolux LED (300 mW/cm2, Coltene/Whaledent Inc., Cuyahoga Falls, OH, USA) was applied, and samples were cured for 60 s on each side. Each sample was prepared in 3 × 6 mm diameters. The FS results were determined using a universal testing machine (SANTAM Co. STM-400 serial No:818408 (SANTAM ENG. DESIGN, Tehran, Iran)) at a speed of 1 mm/min. The DTS test was conducted according to ADA Standard No. 27 [24]. The FS, FM, and CS measurements were performed with dimensions according to ISO 10477:92 standards [25]. The tests were repeated three times, and with the assistance of Minitab statistical software, the analysis of variance (ANOVA) (p < 0.05) was performed. Tukey’s test was used for homogeneous results, and the Dunnett post hoc 3 test was used to analyze the non-homogenous data.
The samples with Zn2SiO4 were prepared following a similar procedure.

3. Result

3.1. Filler Characterization Results

The phase composition of Zn2SiO4 and SiO2–ZnO nano-materials was primarily recognized through the XRD pattern measurement. The XRD patterns are depicted in Figure 1. The FE-SEM image of the SiO2–ZnO nano-composite shows that the particles with a spherical morphology exhibited a particle size distribution of less than 100 nm, while Zn2SiO4 had larger and more uniform particles (Figure 2). The chemical composition of Zn2SiO4 and SiO2–ZnO was determined by EDX analysis as depicted in Figure 2. Zn2SiO4 has an atomic ratio of 4/1/2 and a chemical composition of 28.75, 12.49, and 57.35 wt% for O, Si, and Zn elements, respectively. Similarly, SiO2–ZnO has an atomic ratio of 3.4/1.2/1 and a chemical composition of 33.02, 20.58, and 45.11 wt% for O, Si, and Zn elements, respectively.

3.2. Dental Composite Samples Characterization Results

Ten samples of dental composites filled with 2, 5, 8, 10, and 15 wt% of SiO2–ZnO and Zn2SiO4 were prepared. The samples were coded as A1A10 as shown in Table 1. All samples were characterized by different techniques including FT-IR, FE-SEM, EDX, and TGA.
Figure 3 shows the FT-IR analyses of Zn2SiO4 and SiO2–ZnO and dental composite samples. The TGA measurements for all samples were carried out in a nitrogen atmosphere to characterize the thermal stability behaviors of the dental composite samples (Table 1, A1A10). The obtained results are depicted in Figure 4. The morphology and chemical compositions of all samples were investigated by FESEM images and EDX analyses, and the results are shown in Figure 5, Figure 6 and Figure 7. All samples were irregular in shape and composed of C, O, S, and Zn atoms, which confirmed the corrected structure of the dental composites and the presence of inorganic fillers in organic polymer mixtures.

3.3. Mechanical Properties Results

All samples were evaluated for their mechanical characteristics including the flexural strength (FS), flexural modulus (FM), diameter tensile strength (DTS), and compressive strength (CS). Figure 8, Figure 9, Figure 10 and Figure 11 show the results of other mechanical parameters including the flexural strength (FS), flexural modulus (FM), diameter tensile strength (DTS), and compressive strength (CS).

4. Discussion

The XRD pattern of SiO2–ZnO showed a broad peak below 30 [2θ°] corresponding to the silica (SiO2) phase, which seemed to be amorphous (Figure 1). The sharp peaks that appeared after that were around 31.8, 34.4, 36.2, 47.5, 56.6, 62.8, 67.9, and 69.1 [2θ°], which corresponded to the hexagonal structure (JCPDS: 01-079-2205) of zinc oxide (ZnO) with the P63mc space group. The XRD pattern of Zn2SiO4 showed characteristic peaks at 22.2, 25.4, 31.5, 34.1, 38.8, 47.1, 49.3, 66.0, 70.2, and 77.7 [2θ°] corresponding to a rhombohedral crystal system [Reference code: 00-002-1412] (Figure 1).
The crystal size of the samples was calculated using the Scherrer Equation (1).
D = K λ β   Cos θ
where K is the Scherrer constant, λ is the wavelength of the X-ray beam used (1.54 Å). β is the full-width at half maximum (FWHM) of the peak, and θ is the Bragg angle [19,20,21,22,23]. The main crystal sizes of Zn2SiO4 and SiO2–ZnO samples were calculated as 65 and 59 nm (for the ZnO phase), respectively.
As shown in Figure 3, The FT-IR adsorption spectra of the SiO2–ZnO sample show distinctive adsorption bonds at 1618, 1070, 942, and 560 cm−1 for stretching and bending vibration of Si–O, Zn–O, and Si–OH bonds. The broad adsorption bond at 3100–3600 cm−1 could be referred to OH groups inserted to the surface of SiO2 and adsorbed water by KBr discs. Similarly, the Zn2SiO4 sample has adsorption bands centered at 1626, 1485, 1012, and 510 cm−1 for their Si-O, SiO4, and Zn–O groups.
Next, the FT-IR analyses of all samples were performed, and their infrared spectra are depicted in Figure 3. All samples showed a broad adsorption bond from 3200 to 3600 cm−1 referred to OH groups and adsorbed water by KBr discs. The adsorption bonds at 3031 (stretching vibration C–H, sp2), 2945, 2870 (stretching vibration C–H, sp3), 1769, 1725 (stretching vibration COO groups), 1630, 1601, 1565 (stretching vibration C=C bonds), 1354 (bending vibration C–H, sp3), 1170, 1135, and 1067 (stretching and bending vibration C–C and C–O bonds) could confirm that organic parts of the composite were present. The filled samples with a lower 10% concentration did not show adsorption bonds related to metal oxide bonds, while in higher dosages, the bonds were broadened at the range 1000–1200 cm−1, and a distinctive bond below 600 cm−1 confirmed that the inorganic parts of the sample were present.
According to the TGA measurements as shown in Figure 4, the weight losses of all samples had a similar pattern from about 200 to 600 °C due to the thermal decomposition of the organic polymeric networks. Also, a maximum of 2% of weight loss from 50 to 200 °C was due to the adsorbed water and impurities evaporation. The weight loss percents could be matched with the composition of the samples at the first-time construction of the samples (Table 1, A1A10). In addition, the filled samples showed higher thermal stability than those of the unfilled ones.
As can be seen from Figure 8, the FS of the samples was increased when the concentration of the filler was increased, and the highest increase was observed in 8 wt% (A3, 123.4 MPa) of SiO2–ZnO (p > 0.05). This value for Zn2SiO4 as a filler was 10 wt% (A9, 136.6 MPa). With the further increase in filler dosages, the FS values decreased.
Generally, compared with the dental resin base sample with a compressive strength value of 97.6 MPa, the samples containing SiO2–ZnO as a filler with 2, 5, 8, 10, and 15 wt% showed higher strength. Zn2SiO4 as a filler showed higher values of FS than SiO2–ZnO.
A similar trend was found when other mechanical parameters including FM, DTS, and CS were investigated. Notably, with the increase in filler content, an increase in the FM, DTS, and CS characteristics was observed. Accordingly, 8 wt% of fillers (samples SiO2–ZnO (A3) and Zn2SiO4 (A8)) showed higher values of FM, DTS, and CS characteristics as 8.6 GPa, 34.2 MPa, and 183.8 MPa values for SiO2–ZnO and 11.3 GPa, 41.2 MPa, and 190.5 MPa values for Zn2SiO4 samples (Figure 9, Figure 10 and Figure 11). The improvement of the mechanical properties of dental resins was due to the adhesion properties between the resins and fillers. In the meantime, due to the characteristic of high polarity, Zn2SiO4 created the possibility of a more suitable connection and more delay compared with SiO2–ZnO. This characteristic showed itself in the higher values of the measured FS, FM, DTS, and CS characteristics. The acceptable values for the DTS test of a dental composite were 30–55 MPa [26], which was consistent with the results of SiO2–ZnO and Zn2SiO4 in the current study (samples A3 (34.2 MPa) and A8 (41.2 MPa)). In all cases, the p value was significant (p ˂ 0.05).
In comparison, SiO2 and ZnO as fillers alone or together did not improve the mechanical properties as well as SiO2–ZnO or Zn2SiO4 (Table 2). The results showed that the filler dosages and types significantly influenced the mechanical characteristics of the dental resins. On the other hand, Zn2SiO4 as a filler performed better results than the SiO2–ZnO sample. A historical view of point revealed that a glass (Specialty Glass, USA, 2–4m, 73 wt%) was used by Zandinejad et al. for reinforcing and improving the mechanical characteristics of bis-GMA/TEGDMA (70/30) mixed resins and showed suitable improvement in the FS, FM, and DTS mechanical properties [27]. Samuel and co-workers reported that mesoporous and nonporous spherical silica (70 wt%) could improve the mechanical properties of dental resins [28]. In another study, silica nanoparticle loading showed suitable increasing in the FS in bis-GMA/TEGDMA mixed resins [29]. However, when compared, the results of the SiO2–ZnO and Zn2SiO4 samples were comparable and, in some cases, better than those of the fillers reported in the literature (Table 2).
This work suggested that Zn2SiO4 and SiO2–ZnO could effectively improve the thermal stability and mechanical properties of Bis-GMA/ TEGDMA dental resins when used as reinforcing filler. In comparison with the fillers reported in the literature, Zn2SiO4 and SiO2–ZnO had reasonably good potential to improve the FS, FM, DTS, and CS characteristics of dental resins. Worth noting about ZnO and its compounds is their bioactivity, which may increase the applications of this work. For future research, Zn2SiO4 and SiO2–ZnO could be used for the improvement of different dental resins and polymers. In addition, experiments related to biological tests could be the next study.
The main limitation of this work is the use of only Bis-GMA/TEGDMA dental resins, and also, this work did not investigate the dental tests. Also, the cytotoxicity and biological effects of these fillers were not studied. Thus, this work could be expanded by other researchers for dental applications. In addition, the studies with CAD/CAM blocks and color stability of the samples are of interest for future investigations [33].

5. Conclusions

Zn2SiO4 and SiO2–ZnO samples were prepared and characterized by XRD, SEM, EDX, and FT-IR techniques. The samples were used as a filler in a dental composite constructed from Bis-GMA and TEGDMA copolymers. Compared with non-composited dental resins, a dental composite containing Zn2SiO4 and SiO2–ZnO as a filler had superior values for FS, FM, DTS, and CS characteristics. As found, 8 wt% of SiO2–ZnO and 10 wt% of Zn2SiO4 showed higher values of FS of 123.4 and 136.6 MPa, respectively. The optimized values of the FM, DTS, and CS parameters for Zn2SiO4 (8 wt%) as a filler were 131.3 MPa, 11.33 GPa, 41.25 MPa, and 190.56 MPa, and those for SiO2–ZnO (8 wt%) were 123.4 MPa, 8.64 GPa, 34.24 MPa, and 183.83 MPa, respectively. To date, various fillers have been used to improve the mechanical properties of dental resins. Among the different fillers, nanocomposites seem to provide promising results for enhanced mechanical properties. Silica nanocomposites are inexpensive and show reasonable improvements regarding the mechanical properties of various polymeric systems. Thus, the use of silica–metal oxide nanocomposites could definitely improve the mechanical properties of dental composites.

Author Contributions

P.T.: methodology and performing experiments. S.M.N.: formal analysis, investigation, and writing—original draft. D.G.: formal analysis and writing—original draft. M.G.: supervising, methodology, formal analysis, and writing the paper. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Data will be made available on request.

Acknowledgments

We are thankful to the Najafabad Branch, Islamic Azad University research council, for partial support of this research. This article was completed while the second author, its a doctoral candidate in the Interdisciplinary Doctoral School at the Lodz University of Technology, Poland.

Conflicts of Interest

The authors declared no potential conflict of interest concerning this article’s research, authorship, and/ or publication.

References

  1. Habib, E.; Wang, R.; Wang, Y.; Zhu, M.; Zhu, X.X. Inorganic fillers for dental resin composites: Present and future. ACS Biomater. Sci. Eng. 2016, 2, 1–11. [Google Scholar] [CrossRef] [PubMed]
  2. Amiri, P.; Talebi, Z.; Semnani, D.; Bagheri, R.; Fashandi, H. Improved performance of Bis-GMA dental composites reinforced with surface-modified PAN nanofibers. J. Mater. Sci. Mater. Med. 2021, 32, 82. [Google Scholar] [CrossRef] [PubMed]
  3. Zeng, W.; Liu, F.; He, J. Physicochemical properties of Bis-GMA/TEGDMA dental resin reinforced with silanized multi-walled carbon nanotubes. Silicon 2019, 11, 1345–1353. [Google Scholar] [CrossRef]
  4. Barszczewska-Rybarek, I.; Chladek, G. Studies on the curing efficiency and mechanical properties of Bis-GMA and TEGDMA nanocomposites containing silver nanoparticles. Int. J. Mol. Sci. 2018, 19, 3937. [Google Scholar] [CrossRef]
  5. Barszczewska-Rybarek, I.M.; Chrószcz, M.W.; Chladek, G. Physicochemical and mechanical properties of bis-gma/tegdma dental composite resins enriched with quaternary ammonium polyethylenimine nanoparticles. Materials 2021, 14, 2037. [Google Scholar] [CrossRef]
  6. Par, M.; Plančak, L.; Ratkovski, L.; Tauböck, T.T.; Marovic, D.; Attin, T.; Tarle, Z. Improved flexural properties of experimental resin composites functionalized with a customized low-sodium bioactive glass. Polymers 2022, 14, 4289. [Google Scholar] [CrossRef]
  7. Wang, X.; Cai, Q.; Zhang, X.; Wei, Y.; Xu, M.; Yang, X.; Ma, Q.; Cheng, Y.; Deng, X. Improved performance of Bis-GMA/TEGDMA dental composites by net-like structures formed from SiO2 nanofiber fillers. Mater. Sci. Eng. C 2016, 59, 464–470. [Google Scholar] [CrossRef]
  8. Alsharif, S.O.; Akil, H.B.; Abd El-Aziz, N.A.; Ahmad, Z.A. Effect of alumina particles loading on the mechanical properties of light-cured dental resin composites. Mater. Des. 2014, 54, 430–435. [Google Scholar] [CrossRef]
  9. Habib, E.; Wang, R.; Zhu, X.X. Monodisperse silica-filled composite restoratives mechanical and light transmission properties. Dent. Mater. 2017, 33, 280–287. [Google Scholar] [CrossRef]
  10. Zhang, H.; Darvell, B.W. Mechanical properties of hydroxyapatite whisker-reinforced bis-GMA-based resin composites. Dent. Mater. 2012, 28, 824–830. [Google Scholar] [CrossRef]
  11. Tian, M.; Gao, Y.; Liu, Y.; Liao, Y.; Hedin, N.E.; Fong, H. Fabrication and evaluation of Bis-GMA/TEGDMA dental resins/composites containing nano fibrillar silicate. Dent. Mater. 2008, 24, 235–243. [Google Scholar] [CrossRef] [PubMed]
  12. Yang, J.; Shen, J.; Wu, X.; He, F.; Xie, H.; Chen, C. Effects of nano-zirconia fillers conditioned with phosphate ester monomers on the conversion and mechanical properties of Bis-GMA-and UDMA-based resin composites. J. Dent. 2020, 94, 103306. [Google Scholar] [CrossRef] [PubMed]
  13. Gao, Y.; Sagi, S.; Zhang, L.; Liao, Y.; Cowles, D.M.; Sun, Y.; Fong, H. Electrospun nano-scaled glass fiber reinforcement of bis-GMA/TEGDMA dental composites. J. Appl. Polym. Sci. 2008, 110, 2063–2070. [Google Scholar] [CrossRef]
  14. Cheng, L.; Zhou, X.; Zhong, H.; Deng, X.; Cai, Q.; Yang, X. NaF-loaded core–shell PAN–PMMA nanofibers as reinforcements for Bis-GMA/TEGDMA restorative resins. Mater. Sci. Eng. C 2014, 34, 262–269. [Google Scholar] [CrossRef] [PubMed]
  15. Yang, J.; Liao, M.; Hong, G.; Dai, S.; Shen, J.; Xie, H.; Chen, C. Effect of APTES-or MPTS-conditioned nanozirconia fillers on mechanical properties of Bis-GMA-based resin composites. ACS Omega 2020, 5, 32540–32550. [Google Scholar] [CrossRef] [PubMed]
  16. Fu, L.; Engqvist, H.; Xia, W. Glass–ceramics in dentistry: A review. Materials 2020, 13, 1049. [Google Scholar] [CrossRef]
  17. Elsaka, S.E.; Elnaghy, A.M. Mechanical properties of zirconia reinforced lithium silicate glass-ceramic. Dent. Mater. 2016, 32, 908–914. [Google Scholar] [CrossRef]
  18. GKaur, G.; Kumar, V.; Baino, F.; Mauro, J.C.; Pickrell, G.; Evans, I.; Bretcanu, O. Mechanical properties of bioactive glasses, ceramics, glass-ceramics and composites: State-of-the-art review and future challenges. Mater. Sci. Eng. C 2019, 104, 109895. [Google Scholar] [CrossRef]
  19. Bahramian, F.; Fazlinia, A.; Mansoor, S.S.; Ghashang, M.; Azimi, F.; Najafi Biregan, M. Preparation of 3, 4, 5-substituted furan-2 (5 H)-ones using HY Zeolite nano-powder as an efficient catalyst. Res. Chem. Intermed. 2016, 42, 6501–6510. [Google Scholar] [CrossRef]
  20. Roudbaraki, S.J.; Mansoor, S.S.; Ghashang, M. Aqueous media synthesis of pyrano [3, 2-c] chromen derivatives using magnesium oxide nanoparticles as a recyclable catalyst. Polycycl. Aromat. Compd. 2021, 41, 211–222. [Google Scholar] [CrossRef]
  21. Jafari, M.; Hassanzadeh-Tabrizi, S.A.; Ghashang, M.; Pournajaf, R. Characterization of Ba2+-added alumina/cobalt nanoceramic pigment prepared by polyacrylamide gel method. Ceram. Int. 2014, 40, 11877–11881. [Google Scholar] [CrossRef]
  22. Reza Mohammad Shafiee, M.; Kargar, M.; Sadat Hashemi, M.; Ghashang, M. Green synthesis of NiFe2O4/Fe2O3/CeO2 nanocomposite in a Walnut Green Hulls extract medium: Magnetic properties and characterization. Curr. Nanosci. 2016, 12, 645–649. [Google Scholar] [CrossRef]
  23. Ghashang, M. An Aurivillius perovskite nano-structure of SrBi4Ti4O15: Efficient catalyst for the preparation of novel dihydronaphtho [2′, 1′: 4, 5] thieno [2, 3-d] pyrimidin-7 (6 H)-one derivatives using HSBM technique. J. Iran. Chem. Soc. 2018, 15, 55–60. [Google Scholar] [CrossRef]
  24. New American Dental Association specification no. 27 for direct filling resins. Council on Dental Materials and Devices. J. Am. Dent. Assoc. 1977, 94, 1191–1194.
  25. ISO 10477:1992(E); Dentistry—Polymer-based crown and bridge materials. International Organization for Standardization: Geneva, Switzerland, 1992.
  26. Alkemper, J.; Rentsch, H.; Dermann, K.; Ritzhaupt-Kleissl, H.J.; Hausselt, J.; Albert, P.; Gall, C.; Aktiengesellschaft, D.-H. Dental Material Comprising Porous Glass Ceramics, Porous Glass Ceramics, Processes and Use. United States Patent US 6362251, 26 March 2002. [Google Scholar]
  27. Zandinejad, A.A.; Atai, M.; Pahlevan, A. The effect of ceramic and porous fillers on the mechanical properties of experimental dental composites. Dent. Mater. 2006, 22, 382–387. [Google Scholar] [CrossRef]
  28. Samuel, S.P.; Li, S.; Mukherjee, I.; Guo, Y.; Patel, A.C.; Baran, G.; Wei, Y. Mechanical properties of experimental dental composites containing a combination of mesoporous and nonporous spherical silica as fillers. Dent. Mater. 2009, 25, 296–301. [Google Scholar] [CrossRef]
  29. Hosseinalipour, M.; Javadpour, J.; Rezaie, H.; Dadras, T.; Nemati Hayati, A. Investigation of Mechanical Properties of Experimental Bis-GMA/TEGDMA Dental Composite Resins Containing Various Mass Fractions of Silica Nanoparticles. J. Prosthodont. 2010, 19, 112–117. [Google Scholar] [CrossRef]
  30. Krause, W.R.; Park, S.H.; Straup, R.A. Mechanical properties of BIS-GMA resin short glass fiber composites. J. Biomed. Mater. Res. 1989, 23, 1195–1211. [Google Scholar] [CrossRef]
  31. Liu, F.; Sun, B.; Jiang, X.; Aldeyab, S.S.; Zhang, Q.; Zhu, M. Mechanical properties of dental resin/composite containing urchin-like hydroxyapatite. Dent. Mater. 2014, 30, 1358–1368. [Google Scholar] [CrossRef]
  32. Fong, H. Electrospun nylon 6 nanofiber reinforced BIS-GMA/TEGDMA dental restorative composite resins. Polymer 2004, 45, 2427–2432. [Google Scholar] [CrossRef]
  33. Paolone, G.; Mandurino, M.; De Palma, F.; Mazzitelli, C.; Scotti, N.; Breschi, L.; Gherlone, E.; Cantatore, G.; Vichi, A. Color Stability of Polymer-Based Composite CAD/CAM Blocks: A Systematic Review. Polymers 2023, 15, 464. [Google Scholar] [CrossRef] [PubMed]
Figure 1. XRD patterns of Zn2SiO4 and SiO2–ZnO.
Figure 1. XRD patterns of Zn2SiO4 and SiO2–ZnO.
Materials 16 06268 g001
Figure 2. FE-SEM images and EDX analyses of Zn2SiO4 and SiO2–ZnO.
Figure 2. FE-SEM images and EDX analyses of Zn2SiO4 and SiO2–ZnO.
Materials 16 06268 g002
Figure 3. FT-IR analyses of Zn2SiO4 and SiO2–ZnO and dental composite samples. The left image illustrates the FT-IR analyses of dental samples (A3A5A7A10, and X), The right image shows FT-IR analyses of Zn2SiO4 and ZnO-SiO2.
Figure 3. FT-IR analyses of Zn2SiO4 and SiO2–ZnO and dental composite samples. The left image illustrates the FT-IR analyses of dental samples (A3A5A7A10, and X), The right image shows FT-IR analyses of Zn2SiO4 and ZnO-SiO2.
Materials 16 06268 g003
Figure 4. TGA analyses of dental composite samples (A1A10) filled with Zn2SiO4 and SiO2–ZnO. In accordance with Table 1, the image on the left displays TGA analyses (A1A5) conducted on dental composite samples, while the image on the right features samples (A6A10 and X).
Figure 4. TGA analyses of dental composite samples (A1A10) filled with Zn2SiO4 and SiO2–ZnO. In accordance with Table 1, the image on the left displays TGA analyses (A1A5) conducted on dental composite samples, while the image on the right features samples (A6A10 and X).
Materials 16 06268 g004
Figure 5. Selected FE-SEM images of dental composite samples (A1A10).
Figure 5. Selected FE-SEM images of dental composite samples (A1A10).
Materials 16 06268 g005
Figure 6. FE-SEM and EDX analyses of dental composite samples (A3, A4, A5, A8, A9, and A10).
Figure 6. FE-SEM and EDX analyses of dental composite samples (A3, A4, A5, A8, A9, and A10).
Materials 16 06268 g006
Figure 7. FE-SEM and EDX analyses of dental composite samples (A1, A2, A6, and A7).
Figure 7. FE-SEM and EDX analyses of dental composite samples (A1, A2, A6, and A7).
Materials 16 06268 g007
Figure 8. Flexural strength (FS) of dental composites (A1A10).
Figure 8. Flexural strength (FS) of dental composites (A1A10).
Materials 16 06268 g008
Figure 9. Flexural modulus (FM) of dental composites (A1A10).
Figure 9. Flexural modulus (FM) of dental composites (A1A10).
Materials 16 06268 g009
Figure 10. Flexural diameter tensile strength (DTS) of dental composites (A1A10).
Figure 10. Flexural diameter tensile strength (DTS) of dental composites (A1A10).
Materials 16 06268 g010
Figure 11. Compressive strength (CS) of dental composites (A1A10).
Figure 11. Compressive strength (CS) of dental composites (A1A10).
Materials 16 06268 g011
Table 1. Coded samples of dental composites: the starting materials were mixed according to the wt% of each material.
Table 1. Coded samples of dental composites: the starting materials were mixed according to the wt% of each material.
Sample CodeBis-GMA (wt%)TEGDMA (wt%)CQ/3-NDADM (wt%)Filler (wt%)
X70290.5/0.5-
A169280.5/0.5SiO2–ZnO (2)
A267.526.50.5/0.5SiO2–ZnO (5)
A366250.5/0.5SiO2–ZnO (8)
A465240.5/0.5SiO2–ZnO (10)
A562.521.50.5/0.5SiO2–ZnO (15)
A669280.5/0.5Zn2SiO4 (2)
A767.526.50.5/0.5Zn2SiO4 (5)
A866250.5/0.5Zn2SiO4 (8)
A965240.5/0.5Zn2SiO4 (10)
A1062.521.50.5/0.5Zn2SiO4 (15)
Table 2. Comparison results.
Table 2. Comparison results.
FillerBis-GMA/TEGDMA wt%FS [MPa]FM [GPa]DTS [MPa]CS [MPa]Ref.
-70/29.4597.93.419.8144.9This work
A365/26.45123.48.6434.24183.83This work
A865/26.45131.311.3341.25190.56This work
SiO2–ZnO bulk (8 wt%)65/24.45105.746.4524.14155.31This work
SiO2 (6wt%) + ZnO (2 wt%)65/24.45115.027.8231.21177.24This work
ZnO (6 wt%)68/29.4591.253.2617.09142.02This work
SiO2 (6 wt%)68/29.45103.494.926.13155.67This work
Glass (Specialty Glass, Willow Grove, PA, USA) 2–4 m. (73 wt%)70/3043.68.642.5-Zandinejad et al. [27]
Mesoporous and nonporous spherical silica (70 wt%)50/50726.7-191Samuel et al. [28]
Silica nanoparticles modified with γ-methacryloxy propyl trimethoxy silane (40 wt%)70/30149.74---Hosseinalipour et al. [29]
Quaternary ammonium polyethylenimine nanoparticles (2 wt%)60/4063.093.712--Barszczewska-Rybarek et al. [5]
Short glass fibers (70 wt%) 70/30-4.59-66.6Krause et al. [30]
Urchin-like hydroxyapatite (10 wt%) 49.5/49.5123.5-42.7363.5Liu et al. [31]
Electrospun nylon 6 nanofiber (7.5 wt%)50/50112.12.726--Fong [32]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Torkian, P.; Mortazavi Najafabadi, S.; Ghashang, M.; Grzelczyk, D. Glass-Ceramic Fillers Based on Zinc Oxide–Silica Systems for Dental Composite Resins: Effect on Mechanical Properties. Materials 2023, 16, 6268. https://doi.org/10.3390/ma16186268

AMA Style