Article

Evaluation of the Biocompatibility and Osteogenic Properties of Metal Oxide Coatings Applied by Magnetron Sputtering as Potential Biofunctional Surface Modifications for Orthopedic Implants

by
Mariana Fernández-Lizárraga
1,2,3,
Julieta García-López
2,
Sandra E. Rodil
4,
Rosa María Ribas-Aparicio
1,3,* and
Phaedra Silva-Bermudez
2,*
1
Posgrado de Doctorado en Ciencias en Biomedicina y Biotecnología Molecular, Escuela Nacional de Ciencias Biológicas, Instituto Politécnico Nacional, Mexico City 11340, Mexico
2
Unidad de Ingeniería de Tejidos, Terapia Celular y Medicina Regenerativa, Instituto Nacional de Rehabilitación Luis Guillermo Ibarra Ibarra, Mexico City 14389, Mexico
3
Laboratorio de Producción y Control de Biológicos, Departamento de Microbiología, Escuela Nacional de Ciencias Biológicas, Instituto Politécnico Nacional, Mexico City 11340, Mexico
4
Instituto de Investigaciones en Materiales, Universidad Nacional Autónoma de México, Mexico City 04510, Mexico
*
Authors to whom correspondence should be addressed.
Materials 2022, 15(15), 5240; https://doi.org/10.3390/ma15155240
Submission received: 10 June 2022 / Revised: 2 July 2022 / Accepted: 5 July 2022 / Published: 29 July 2022

Abstract

:
Biomaterials with adequate properties to direct a biological response are essential for orthopedic and dental implants. The surface properties are responsible for the biological response; thus, coatings with biologically relevant properties such as osteoinduction are exciting options to tailor the surface of different bulk materials. Metal oxide coatings such as TiO2, ZrO2, Nb2O5 and Ta2O5 have been suggested as promising for orthopedic and dental implants. However, a comparative study among them is still missing to select the most promising for bone-growth-related applications. In this work, using magnetron sputtering, TiO2, ZrO2, Ta2O5, and Nb2O5 thin films were deposited on Si (100) substrates. The coatings were characterized by Optical Profilometry, Scanning Electron Microscopy, Energy-Dispersive X-ray Spectroscopy, X-ray Photoelectron Spectroscopy, X-ray Diffraction, Water Contact Angle measurements, and Surface Free Energy calculations. The cell adhesion, viability, proliferation, and differentiation toward the osteoblastic phenotype of mesenchymal stem cells plated on the coatings were measured to define the biological response. Results confirmed that all coatings were biocompatible. However, a more significant number of cells and proliferative cells were observed on Nb2O5 and Ta2O5 compared to TiO2 and ZrO2. Nevertheless, Nb2O5 and Ta2O5 seemed to induce cell differentiation toward the osteoblastic phenotype in a longer cell culture time than TiO2 and ZrO2.

1. Introduction

In the orthopedic and dental fields, different efforts have been done to develop improved biomaterials capable of fulfilling the different requirements for their use as temporal osteosynthesis implants to treat bone fractures or critical bone defects [1,2], or as permanent implants such as knee total replacement prosthesis or dental implants [3,4,5,6].
For bone-substituting implants, the standard requirements are mechanical resistance and load-bearing capacity, which are associated with the material’s bulk properties [7,8,9]. However, the bone-implant junction (osseointegration), the osteosynthesis performance (osteogenesis capability), the ability to prevent infections, and the corrosion resistance are surface-related properties [4,10,11,12,13].
An interesting strategy to tailor the surface properties is the development of biofunctional coatings. In the last few years, increasing interest has been focused on developing coatings to harness the surface properties of different bulk materials and their interaction with the biological media [4,13,14,15,16,17,18]. Titanium oxide coatings with different morphologies, crystalline structures, Ti:O ratios, etc., have been the most evaluated after demonstrating that the native oxide of Ti-based implants is responsible for their biocompatibility (chemical inertness) and excellent osseointegration observed in dental implants [19,20,21,22,23,24,25,26]. However, the extensive use of Ti-based and TiO2 materials has raised concerns regarding their low bioactivity, decreasing corrosion resistance in F or Cl containing media after long-term use, and the consequent adverse effects of titanium accumulation and its effects on the human body [3,10,27,28,29]. Reports about Ti-associated allergic reactions and hypersensitivity demand research on alternative materials [30,31,32].
Other biocompatible transition metal oxides have also shown promising biological properties such as osseointegration, improved cell adhesion and proliferation, decreased inflammatory response and antibacterial properties, along with exhibiting very good corrosion and wear resistance [10,33,34,35,36,37,38,39,40,41,42,43]. However, a significantly smaller number of works have studied the biological response of promising oxides, such as Nb2O5 and Ta2O5 [44,45,46,47,48,49]. Another possibility is zirconium oxide, ZrO2, which presents mechanical strength, appropriate corrosion resistance and adequate biological response for intraosseous use [10,34,40,50,51,52,53,54].
A recent review shows that pure NbxOy and NbxOy-containing coatings deposited using dry and/or wet techniques affect the overall characteristics of the underlying bulk materials, where the deposited layers not only address the deficiencies of the biomaterials, such as corrosion resistance but also induce some new properties, e.g., antibacterial activity [55]. For sputtered deposited NbxOy coatings, the improvement in the biological properties is a function of surface topography, wettability, and atomic order [55]. The amorphous atomic order showed superior human fibroblast cell adhesion and antibacterial activity [56], similarly to our previous results for TiOx and ZrOx [57,58]. Moreover, an important finding was the anti-inflammatory properties observed for NbxOy-coated metallic substrates, which decreased the toxicity [55].
Yin-Yu et al. [59] compared amorphous and crystalline tantalum oxide coatings deposited by magnetron sputtering, finding that the hydrophilic crystalline β-Ta2O5 coating presents good biocompatibility for human skin fibroblast cells. In contrast, the amorphous tantalum oxide coating possessed antibacterial properties, in agreement with N. Donkov et al. [60]. Horandghadim et al. [48] showed that the higher the content of Ta2O5 on hydroxyapatite-Ta2O5 coatings, the higher the osteoblast-like cell attachment and the bone-like apatite growth. In agreement, F. Wang et al. [47] demonstrated higher cell adhesion and viability, and enhanced calcium deposition and expression of osteogenic genes on Ta2O5-coated titanium nanotubes compared to the bare nanotubes. In the same trend, Almeida-Alves et al. [44] and H.-L. Huang et al. [61] reported higher MC3T3, human skin fibroblasts, and human osteosarcoma MG-63 cells viability on Ta2O5 coatings compared to bare Ti.
More research has been performed on ZrO2 as a bulk and coating material [10,34,50]. The use of zirconia in the medical field has expanded over the past twenty years, driven by its advantageous physical, biological, esthetic, and corrosion properties [10,34]. For orthopedic hip replacements, it has superior wear-resistance compared to metals, and the esthetic factor has benefited its use in dentistry for all-ceramic crowns [50]. However, as a bulk material, the risk of catastrophic fracture remains a concern [34]. Concerning ZrO2 coatings, M. Peron et al. showed enhanced cell viability for L929 fibroblasts in contact with lixiviates from ZrO2-coated Mg alloys compared to TiO2-coated and bare Mg alloys’ lixiviates [53,54]. In the same trend, J. Maminskas et al. demonstrated more significant cell adhesion and growth of HGF-1 cells on ZrO2 coatings than on bare Ti and Ti alloys [52]. On the other hand, S. Saleem et al. [51] demonstrated a lower friction coefficient and wear rate for ZrO2-coated titanium substrates than for the bare substrates.
These ceramic oxides could be applied as coatings with the advantage that they modify the surface properties preserving the bulk properties because they can be deposited over any substrate material that fulfills: (1) the mechanical requirements of the application, either a rigid polymer or a metallic substrate, and (2) the micro roughness requirements for osseointegration [14,18,62]. Coatings can be deposited using techniques such as Atomic Layer Deposition, Electron Beam, sol-gel, etc. In particular, magnetron sputtering is a versatile, clean deposition technique that allows the deposition of conformal and homogeneous coatings, with controlled chemical composition and thickness, at an appropriate deposition rate, and it can be industrially scalable to coat three-dimensional substrates [63,64,65,66]. In the industry, magnetron sputtering has been successfully used to coat a variety of substrates for different applications, e.g., architectural glass for energy efficiency buildings, electronics for the communication industry, or cutting tools for the construction industry [64,67,68,69]. By controlling the deposition parameters such as the power supply, the deposition time, the cathode-anode distance, the magnetron sputtering discharge configuration (planar, cylindrical, rotating, single or multiple blanks, etc.), the sample holder (rotating or static), etc. three dimensional structures of different materials can be conformally and homogeneously coated [64,67,68,70,71].
Surface roughness in the micro range (1–100 µm; substrate roughness) modifies cell proliferation and viability, and osseointegration [58,72,73,74]. Nevertheless, surface nano roughness also modifies the cell response by controlling the protein layer adsorbed on the material surface and, consequently, the extracellular matrix composition [58,75,76,77]. Thus, it is important to deposit the metal oxide coatings on nano-flat substrates such as atomically flat Si (100) wafers for evaluating the biological response to the properties of the oxide coating without interference by the substrate micro roughness.
Despite the different reports evaluating the application of metal oxide coatings for potential application in the biomedical area, to the best of our knowledge, there is no report comparing the biological and osteogenic response under the same experimental conditions.
In the present work, titanium oxide, zirconium oxide, niobium oxide and tantalum oxide thin films were deposited by magnetron sputtering on Si (100) substrates to study their potential as surface modifications (metal oxide coatings) for orthopedic and dental implants. The physical–chemical properties of the coatings were characterized by optical profilometry, Scanning Electron Microscopy (SEM), Energy-Dispersive X-ray Spectroscopy (EDS), X-ray Photoelectron Spectroscopy (XPS), X-ray Diffraction (XRD), water contact angle measurements (wettability) and Surface Free Energy (SFE) calculations. The biocompatibility and osteogenic properties of the coatings were characterized by using bone marrow-derived mesenchymal stem cells, and studying their adhesion, viability, proliferation, and differentiation toward the osteoblastic phenotype upon culture on the coatings.

2. Materials and Methods

2.1. Metal Oxide Coatings Deposition

Metal oxide thin films were deposited on Si N/PH (100) wafers (UniSil Corporation, Sta. Clara, CA, USA) and cut into 1 cm2 substrates. Atomically flat Si wafers were used as the substrates to examine the influence of the properties of the metal oxide thin films deposited, mainly their roughness and wettability, on the biological response, with no influence of the substrate roughness. Deposition of the metal oxide thin films, titanium oxide (TiOx), tantalum oxide (TaOx), niobium oxide (NbOx), and zirconium oxide (ZrOx), was carried out by magnetron sputtering using 4” in diameter, high purity, metallic targets of either Ti, Ta, Nb or Zr (SCI Engineered Materials, Ohio, MA, USA). Deposition conditions were the same for all coatings deposited; base pressure was below 2 × 10−6 Torr, a reactive Ar:O2 (80:20) atmosphere was used to a working pressure of 22 mTorr, 200 W incident RF power was applied, and deposition was carried out for 45 min. Figure 1 shows the macroscopic appearance of the coated samples.
Si (100) samples with the middle section covered by permanent marker ink were coated in the same batches as the non-covered, bare Si (100) substrates. After the coating process, ink was removed, exposing the bare Si surface, and creating an appropriate step (bare Si surface- metal oxide-coated Si) to measure the thickness of the metal oxide coatings by profilometry.

2.2. Metal Oxide Coatings Characterization

The morphology of the coated samples and their elemental composition were characterized using a JEOL 7600F Field Emission-Scanning Electron Microscope (FE-SEM) (JEOL USA Inc., Peabody, MA, USA) coupled to EDS, using a voltage of 5.0 kV. For the characterization of the elemental composition by EDS, three different areas (≈10 µm2) on two independent samples for each coating group were analyzed, and the average elemental composition (in at.% and wt.%) is presented along with the standard deviation. Elemental and chemical surface composition analyses were performed by X-ray Photoelectron Spectroscopy in a Physical Electronics Versa ProbeTM II system with a scanning XPS microprobe (Physical Electronics Inc., Chanhassen, MN, USA), using an Al Kα X-ray source (1486.6 eV and 100 μm beam). To assure the homogeneity and reproducibility of the coatings, different zones on two independent samples per coating group were analyzed. Survey spectra were recorded at 117.4 eV pass energy and high-resolution spectra at 23 eV. Data were analyzed in the Multipack© version 9.6.0.15 software.
Wettability and Surface Free Energy were characterized in an OCA 15EC equipment (DataPhysics Instruments, Filderstadt, Germany) using the Sessil drop contact angle method, and 4 µL drops of distilled water for wettability determinations and 4 µL drops of distilled water, glycerol, isopropanol and dimethyl sulfoxide (DMSO) for surface energy calculations. Contact angles were measured using the SCA_20® software, and the Owens, Wendt, Rabel & Kälble (OWRK) equation was applied to calculate the SFE. For wettability measurements, the stability of the water droplet size over time was measured using the same droplet volume and technique described above. Droplet size and shape were continuously monitored and recorded during the first 5 min immediately after droplet deposition on the coating surface. Images of the droplet at 0, 1, 2, 3, 4 and 5 min after droplet deposition were analyzed. All measurements were done in triplicate.
Crystallinity was characterized by X-ray Diffraction in grazing incidence mode using an Ultima IV diffractometer (Rigaku, Corporation, Tokyo, Japan) with Cu Kα radiation (40 kV, 44 mA). To perform the measurements, the following experimental parameters were set: 0.5° incidence angle, 0.5°/min scanning speed, acquisition step of 0.02°, and 2θ range from 20–70°. TiO2 and ZrO2 crystalline phases in the TiOx and ZrOx coatings were identified by comparing the X-ray diffraction patterns obtained with the 00-021-1276 and 00-021-1272, and 00-007-0343 PDF cards, correspondingly for TiOx and ZrOx. Crystalline phase identification, average crystal grain size, and crystallinity analysis were performed using the PDXL2® software. Crystal grain size for crystalline samples was calculated by two different methods: the Scherer equation indexing the (1,1,1) and (1,1,−1) diffraction peaks for ZrO2 (Baddeleyite), and the (1,1,0) and (2,1,1) diffraction peaks for Rutile and (2,1,1) diffraction peak for Anatase for TiO2. The crystal grain size was also calculated by the Williamson–Hall method. Average crystal grain sizes are presented.
Coatings thickness, surface topography imaging, and average roughness (Sa) were measured in a Zygo NexviewTM optical profilometer (Zygo Corporation, Middlefield, CT, USA) using the MxTM software (version 6.4.0.21, Zygo Corporation, Middlefield, CT, USA) for data analysis.

2.3. Cells

Human Bone Marrow-derived Mesenchymal Stem Cells (BM-MSC; PCS-500-012, ATCC®, Manassas, VA, USA) were used for all biological experiments. Cells were expanded in a Mesenchymal Stem Cell Basal Medium (PCS-500-030; ATCC®, Manassas, VA USA) supplemented with Mesenchymal Stem Cell Growth Kit (PCS-500-041; ATCC®, Manassas, VA, USA), Penicillin-Streptomycin-Amphotericin B solution (PCS-999-002; ATCC®, Manassas, VA, USA) and phenol red (PCS-999-001; ATCC®, Manassas, VA, USA), in the concentrations recommended by the cells’ supplier. Cells were used within a maximum of 9 passages, as indicated by the quality control certificate. Cells were incubated at 37 °C and 5% CO2, and the medium was changed every three days.
Even though BM-MSC in passage 1 were obtained with a quality control certificate from ATCC®, their expression of positive and negative mesenchymal stem cells markers and their differentiation capacity towards chondrogenic, adipogenic and osteogenic lineages were evaluated at passage 6 by flow cytometry and histological staining, respectively. Expression of positive CD90, CD105, CD73, and negative CD45, CD34 and HLA MSC surface markers from BM-MSC were measured in a flow cytometer FACSCaliburTM (Beckton Dickinson, Franklin Lakes, NJ, USA) using the Cell Quest Pro software (V.5.2.1., Beckton Dickinson, Franklin Lakes, NJ, USA) for data acquisition and the FlowJoTM software for data analysis (V.10.8.1., Beckton Dickinson, Franklin Lakes, NJ, USA). BM-MSC cells were independently incubated in specific (osteogenic, adipogenic, and chondrogenic) differentiation-inducing culture media, and cells differentiation toward the specific phenotype (depending on the specific differentiation-inducing media used) was corroborated by histological staining; that is, alizarin red, alcian blue and oil red staining were used to identify osteoblastic, chondrogenic and adipogenic differentiation, respectively. Details of methods and results for BM-MSC characterization are presented in Supplementary Materials: “Bone-marrow derived mesenchymal stem cells characterization”.

2.4. Cell Attachment and Viability of BM-MSC on the Metal Oxide Coatings

Cell adhesion and viability upon culture on the metal oxide coatings were qualitatively and quantitatively evaluated. The coatings samples were sterilized with UV light on both sides for 20 min each side. Then, samples were placed in 24-well tissue culture plates and BM-MSC were drop-seeded on the samples at a density of 4 × 10 3 cells/cm2. To allow the cells’ attachment to the sample’s surface, cells-drop-seeded samples were incubated at 37 °C and 5% CO2 for 90 min. Then, DMEM/F-12 supplemented with 10% v/v Fetal Bovine Serum (FBS; Gibco, Thermo Fisher Scientific, Waltham, MA, USA) and 1% v/v antibiotic-antimycotic was added to the wells up to a 400 µL volume, and culture plates were placed back into the incubator. Independent samples for each coating group were cultured for 3, 7 and 14 days, changing the culture medium every three days. Three different culture time points were chosen to evaluate cell adhesion, viability, and morphology over time up to the cell culture time span when cell differentiation is expected to occur: 14 days.
To evaluate the adhesion and morphology of the BM-MSC at the specific culture time points, independent cell-cultured samples were rinsed twice with PBS 1X and fixed overnight with 2.5% glutaraldehyde (Sigma Aldrich, Waltham, MA, USA). Then, samples were rinsed again with PBS 1X and progressively dehydrated in 20%, 40%, 60%, 80% and 100% ethanol solutions (JT BakerTM, Radnor, PE, USA). Samples were allowed to dry at room temperature (RT), and images were acquired in a Zygo NexviewTM optical profilometer (Zygo Corporation, Middlefield, CT, USA) using the MxTM software (version 6.4.0.21, Zygo Corporation, Middlefield, CT, USA).
To evaluate the cells’ viability at the specific culture time points, the LIVE/DEADTM Viability/Cytotoxicity Kit for mammalian cells (Invitrogen®, Massachusetts, USA) was used on independent cell-cultured samples following the kit manufacturer’s instructions. After incubation with the LIVE/DEADTM kit (Invitrogen®, Waltham, MA, USA), samples were rinsed twice with PBS and immediately visualized by Fluorescence Microscopy in an Axio Observer microscope (Carl Zeiss, Jena, Germany) using the AxioVision® software (Version Rel 4.8.2, Carl Zeiss, Jena, Germany) to acquire the images. Cell viability was also quantitatively evaluated on independent cells-cultured samples using the Alamar BlueTM cell viability kit (Thermo Fisher Scientific, Waltham MA, USA, USA) according to the kit’s manufacturer protocol. Briefly, at the specific culture time points established, the culture medium was removed and replaced in the darkness by 10:1 DMEM/F-12:Alamar BlueTM solution. Then, samples were placed back in the incubator for 2 h, and finally, 100 µL aliquots of the supernatants were collected, and the absorbance was read at 570 nm in a Synergy-HTX multi-mode reader spectrophotometer (Bio Tek Instruments, Winooski, VT, USA). A calibration curve was obtained by seeding different known numbers of cells on independent wells in standard tissue culture plates (TCP). After 24 h of cell culture, cell viability was evaluated following the same procedure as described above for the Alamar Blue assay. The calibration curve calculated the number of viable cells from absorbance readings.
For all techniques described, but Alamar Blue assay, BM-MSC were also cultured on uncoated Si samples for comparison purposes. Cell viability positive controls (Ctrl+) corresponded to cells cultured on TCP under standard culture conditions. All experiments were carried out in triplicate.

2.5. Proliferation of BM-MSC on the Metal Oxide Coatings

Procedures used for cell seeding and culture on the coatings were the same as described in Section 2.4, but cell seeding density was 3.5 × 10 3 cells/cm2 and two different culture times were handled (3 and 5 days) in order to avoid reaching a high cell density on the surface, and the consequent inhibition of cell proliferation due to cell–cell contact phenomenon. After the culture time-points indicated, cell proliferation was evaluated using the Cell Proliferation ELISA, BrdU (colorimetric) kit (Roche, Basilea, Switzerland), following the kit’s manufacturer instructions and setting the incubation time for BrdU labeling solution at 4 h. After removing the labeling medium, samples were dried using a hair dryer for 15 min and stored at 4 °C to continue with the Cell Proliferation ELISA BrdU kit protocol the next day. All experiments were carried out by triplicate.

2.6. Evaluation of the Osteogenic Properties of the Metal Oxide Coatings

An evaluation of cell differentiation toward the osteogenic lineage upon culture on the metal oxide coatings was performed by Immunofluorescence (IF) and ELISA assays.
For IF assays, cells were seeded on the coated samples under the same procedure as described in Section 2.5. Three different osteogenic markers were chosen to be evaluated: RUNX2, osteopontin (OP), and osteocalcin (OC). For RUNX2, OP, and OC expression characterization, independent cell-seeded samples were evaluated after 5, 7 and 14 days of cell culture, respectively. Different days of culture were chosen for the evaluation of the different markers due to their expression occurring at different stages of the cell differentiation process. The culture medium was DMEM/F-12 supplemented with 2% v/v FBS and 1% v/v antibiotic-antimycotic, and incubation conditions were 37 °C and 5% CO2. Human osteoblasts and fibroblasts cells cultured in TCP were used as controls of positive and negative markers expression, respectively; Supplementary information: Immunofluorescence assays controls. BM-MSC cells culture in TCP under standard culture conditions were used as comparative controls of the natural level of markers expression from BM-MSC.
For RUNX2 and OP, the following IF protocol was followed: after corresponding culture time, the culture medium was removed, samples were rinsed twice with a Phosphate Buffer Saline solution (PBS; Gibco Thermo Fisher Scientific, Waltham, MA, USA), incubated in methanol (Sigma–Aldrich, Michigan, USA) for 5 min at RT, rinsed three times with PBS, incubated in Triton 100 × 0.2% (Sigma-Aldrich, St. Louis, MO, USA) in PBS for 30 min, rinsed three times with PBS, and incubated in a PBS-Tween-BSA-FBS-glycine blocking solution (PBS, Tween 0.1% (USB Corporation, Cleveland, OH, USA), BSA 1% (Sigma-Aldrich, St. Louis, MO, USA), FBS 10%, and glycine 0.3 M (MP Biomedicals, Irvine, CA, USA)) for 1 h at RT in a humid chamber. Then, a primary antibody was added, either anti-RUNX2 ab76956 (Abcam, Cambridge, UK) at 1:50 dilution in PBS-1% BSA solution or Anti-Osteopontin ab8448 (Abcam, Cambridge, UK) at 1:100 dilution in PBS-1% BSA solution and samples were incubated for 1 h at RT. After incubation, samples were rinsed five times with the 0.2% PBST (PBS-Tween) solution and incubated for 1 h with the corresponding secondary antibody, either Anti-Mouse ab150105, Alexa Fluor® 488 (Abcam, Cambridge, UK) at 1:1000 dilution in PBS-1% BSA solution for RUNX2 or Donkey Anti-Rabbit Alexa Fluor® 488 (Abcam, Cambridge, UK) at 1:2000 dilution in the PBS-1% BSA solution (Abcam, Cambridge, UK) for OP. Finally, samples were rinsed five times with PBST 0.2%, cells nuclei were counterstained with Hoechst 33342 (InvitrogenTM, Carlsbad, CA, USA) at a 1:1000 dilution in PBS, and samples were placed on a slide with a mounting medium for microscope observation. IF protocol for OC was as follows: after 14 days of cells culture on the coating samples, the culture medium was removed, samples were rinsed twice with PBS, fixed with PFA 4% for 10 min at RT, and rinsed three times with PBS 1X. Then, samples were incubated in a blocking solution (PBST 0.2% and BSA 8%) for 30 min, rinsed three times with PBS 1X, and incubated for 1 h with the primary antibody anti-Osteocalcin ab93876 (Abcam, Cambridge, UK) at a 1:200 dilution in blocking solution at RT in a humid chamber. Then, samples were rinsed five times with PBST 0.2%, and incubated for 1 h at RT with the secondary antibody Donkey Anti-Rabbit Alexa Fluor® 488 (Abcam, Cambridge, UK) at a 1:2000 dilution in PBS-1% BSA solution. Finally, samples were rinsed five times with PBST 0.2%, cell nuclei were counterstained with Hoescht solution, and samples were placed on microscope slides with mounting medium. Mounted samples were visualized by Fluorescence Microscopy using an Axio.Observer microscope (Carl Zeiss, Jena, Germany) and the AxioVision® software, using 100× magnification. Experiments were carried out in triplicate.
The expression of osteogenic markers was quantified using ELISA assays. BM-MSC were seeded on coated samples and cultured for 14 days in DMEM-F/12 medium supplemented with 10% FBS and 1% antibiotic-antimycotic at 37 °C and 5% CO2. The culture medium was replaced every three days. BM-MSC cultured on TCP with Mesenchymal Stem Cells basal medium were used as controls of negative expression of osteogenic markers. After culture time, the culture medium was replaced by fresh DMEM-F/12 with no FBS, and samples were further incubated for 24 h. Then, supernatants were collected and used to quantify marker expression following the ELISA kits manufacturer’s instructions. ELISA kits used were: Human Osteocalcin Simple Step ELISA® kit ab270202 (Abcam, Cambridge, UK); Osteoprotegerin Human ELISA kit ab100617 (Abcam, Cambridge, UK); and Osteopontin Human ELISA kit ab100618 (Abcam, Cambridge, UK). All assays were normalized to the number of viable cells using the Alamar Blue assay for quantification of the number of viable cells on the samples. Experiments were carried out in triplicate. Alkaline Phosphatase Diethanolamine Detection kit AP0100 (Sigma–Aldrich, St. Louis, MO, USA) was also used according to the manufacturer’s instructions to quantify ALP expression at 7 and 14 days for cell culture.

2.7. Statistical Analysis

All data are presented as the mean ± standard error using three independent experiments per variable. Statistical significance was determined by a one-way ANOVA analysis with a Dunnett’s post-hoc test, using the Graph Pad Prism 9.1.0 software (Graph Pad by Dotmatics, Boston, MA, USA). Values of p < 0.05 were considered statistically significant.

3. Results

3.1. Metal Oxides Coatings Characterization

Representative SEM micrographs exhibiting the microscopic morphology of the samples are shown in Figure 2, along with their elemental composition spectra as obtained by EDS. From micrographs in Figure 2, it is possible to observe a smooth morphology for all metal oxide coatings studied, which was later corroborated by a profilometry roughness analysis. Chemical elements identified for each coating are shown in the EDS spectra (Figure 2), exhibiting the expected elements depending on the oxide coating and no signs of trace contamination within the EDS detection limit. Table 1 shows the atomic and weight percentage of the identified chemical elements as calculated from EDS.
Table 2 includes the results from the optical profilometry images (topography), arithmetical average roughness (Sa), and thickness of the metal oxide coatings. Average roughness is expressed as the average of the Sa measured on the whole surface of three different coated samples per metal oxide group of study. As Table 2 shows, all Sa values were smaller than 1 nm, indicating conformal deposition of metal oxide films on the atomically flat substrates (Si (100) wafers) used. The coatings thickness values represent the average of at least eight measurements performed on three independent samples per metal oxide group of study. In all cases, the average thickness was larger than 50 nm. The thinnest coating was TiOx (54.53 nm), while thickness increased, respectively, for NbOx, ZrOx, and TaOx, which corresponded to the thickest coating (346.49 nm). The thickness is not expected to influence the early biological response toward the metal oxide coatings, but latter phenomena such as corrosion or biodegradation might be sensitive to the thickness and homogeneity of the coatings.
Representative survey XPS spectra of the metal oxide coatings after Ar+ ion cleaning are shown in Figure 3. In all cases, the spectra evidenced photoelectron peaks corresponding only to O and to the metal in the oxide: Ti, Ta, Nb, or Zr, corroborating no trace elements contamination in the coatings. A minor C component was observed for ZrOx due to a shorter Ar+ ion cleaning performed on this coating to avoid sub-stoichiometric oxides formation, which occurred upon longer Ar+ ion cleaning; deformation of the Zr photoelectron peaks was clearly observed upon long-time Ar+ ion cleaning. There were no signals of the Si substrate for any of the metal oxide coatings (different regions in at least two different samples per oxide group were analyzed), indicating that the coatings homogeneously and uniformly cover the substrate. Even the thinnest coating deposited, TiOx, did not show any evidence of photoelectron peaks from the Si substrate. Average metal:oxygen ratio for the coatings, as calculated from XPS spectra, were 0.48 ± 0.02 for TiOx; 0.40 ± 0.01 for TaOx; 0.40 ± 0.03 for NbOx and 0.48 ± 0.08 for ZrOx. Evidencing the formation of nearly stoichiometric oxides for all coatings deposited; that is, TiO2 for TiOx, Ta2O5 for TaOx, Nb2O5 for NbOx, and ZrO2 for ZrOx. Metal:oxygen ratios calculated by XPS differed from those as calculated from the at.% elemental composition obtained by EDS. Smaller metal:oxygen ratios (around 0.25) are obtained from EDS data. The differences in the measurement techniques can address this. Firstly, the volume of analysis in EDS is deeper, and some detected oxygen might be associated with the native SiO2 layer. Secondly, during EDS measurements, samples are in vacuum but are not Ar+ ion cleaned before the measurements; thus, traces of organic compounds (containing oxygen) and oxygen from the atmosphere might remain adsorbed on the surface, increasing the amount of oxygen in the coatings’ surface, and consequently decreasing the calculated metal:oxygen ratio from EDS data, in comparison to that obtained from XPS measurements.
The GI-XRD diffraction patterns shown in Figure 4 indicate that both TaOx and NbOx coatings were amorphous. Two broad peaks were observed for these two coatings; however, the average crystallite size calculated from these peaks would be smaller than 1 nm. Consequently, a significantly large number of crystal grains frontiers would exist within the material, causing significant disorder in it so that it can be confidently considered as X-ray amorphous. On the other hand, TiOx and ZrOx showed a nanocrystalline structure with an average crystallite size of 5 ± 1 and 7 ± 2 nm for TiOx and ZrOx, respectively. ZrOx nanocrystalline structure corresponded to the monoclinic phase, baddeleyite, while the TiOx XRD pattern indicated a mixture of the tetragonal crystalline phases, anatase (50.5%) and rutile (49.5%). ZrOx can be considered crystalline with no amorphous part since there was no observable amorphous contribution evidenced by the quite linear background observed. On the other hand, a slight non-linearity can be observed in the background of the XRD spectra for TiOx. Calculating the amorphous-crystalline contributions for this coating, an 87% crystallinity was obtained.
The wettability (WCA) and SFE of the coatings were evaluated by contact angle measurements. The stability of the water droplets over time on the coating’s surface is shown in Table 3, exhibiting the image footage at initial water droplet deposition, and at 1, 2, 3, 4, and 5 min after droplet deposition. From Table 3, a decrease in droplet size with time can be observed for all coatings. Nevertheless, variations in droplet shape, and consequently contact angle, were not relevant; thus, a decrease in droplet size can be mainly attributed to evaporation effects due to water angle measurements performed in an open chamber. From results obtained from water droplet stability measurements over time, it was established to measure WCA at 30 s after droplet deposition on the coating’s surface. In the case of the liquids used for calculating SFE, the same time span was considered except for isopropanol, where contact angles were calculated at 15 s after droplet deposition. Isopropanol has a higher vapor pressure and, consequently, a higher evaporation rate [78], drastically decreasing the droplet’s time to remain stable.
Table 4 shows the average WCA values, as well as the total SFE (γd/p) and the analysis of the polar (γp) and dispersive (γd) components of the total SFE. According to the results obtained for WCA, all coatings can be conventionally considered as having a hydrophilic nature, exhibiting WCA < 90°, being ZrOx the one closest to the hydrophilic–hydrophobic limit. Total SFE was similar among all coatings, ≈24 mN/m with a larger dispersive component compared to the polar component. Total SFE increased from TaOx and ZrOx, to TiOx and NbOx, presenting similar total SFE. γp/(γpd) is known as the polarity factor and indicates the polar fraction of the surface, where TaOx exhibited the highest polar character and decreased to NbOx and ZrOx that exhibited similar polar characters, and to TiOx, which presented the smallest polar character.

3.2. BM-MSC Characterization

BM-MSC cells in passage 6 were characterized by evaluating their expression of the main positive and negative mesenchymal stem cell surface markers, as well as evaluating their capacity to differentiate towards the chondrogenic, adipogenic, and osteogenic lineages, upon incubation with specific differentiation-inducing media; according to the International Society of Cell Therapy. BM-MSC cells exhibited more than 95% positive expression of surface mesenchymal stem markers CD90, CD105, and CD73, and less than 2% of expression of the hematopoietic stem cells markers CD45, CD34, and HLA; Supplementary Figure S1. BM-MSC cells exhibited a fibroblast-like cell morphology (passage 8) incubated in supplemented mesenchymal stem cell basal medium (Supplementary Figure S2a). After 14 days of culture in a chondrogenic-inducing medium, BM-MSC cells were positively stained with alcian blue, indicating synthesis of proteoglycans and corroborating chondrogenic differentiation (Supplementary Figure S2b). After 17 days of incubation in an osteogenic-inducing differentiation medium, BM-MSC were positively stained with alizarin red, exhibiting the formation of calcium deposits, indicative of osteogenic differentiation (Supplementary Figure S2c). Finally, after 12 days of incubation in an adipogenic-inducing differentiation medium, BM-MSC showed lipid vesicles positively stained with oil red dye, demonstrating adipogenic differentiation (Supplementary Figure S2d). It was confirmed that BM-MSC cells used in the present study preserved the main characteristics of mesenchymal stem cells up to passage 8.

3.3. Viability of BM-MSC Cultured on the Surface of the Coatings

Cell adhesion, morphology, and viability of BM-MSC cultured on uncoated Si surfaces and the coatings were evaluated at 3, 7 and 14 days of culture. Figure 5 shows representative micrographs of the cell cultures (after cells fixation) acquired in an optical profilometer. At 5× magnification, main micrographs, the entire sample surface can be observed, which allows for visually evaluating the density and distribution of cells over the entire surface of the samples, and with culture time. These images allow qualitative evaluation of cell proliferation (increment of cell density on the surface) and adhesion over the whole sample surface. Higher magnification images, 50×, allow for precise observation of cells morphology. In the case of BM-MSC cultured on the uncoated Si substrates (Figure 5a–c), the cells initially adhered to the surface, displaying a similar morphology to those of the cells cultured on the coatings. However, at 7 days of culture, fewer cells are observed on Si compared to the coatings, and by day 14, cells detachment from Si was clearly observed. On the coatings, at 3 days of culture (Figure 5d,g,j,m), cells were well-adhered to the surfaces displaying the expected fibroblast-like morphology of well-adhered MSC. For the coatings, as culture time increased, cell density raised to cover the entire surface at 14 days of culture, indicating cell proliferation. At 7 days of culture (Figure 5e,h,k,n), cells exhibited similar extended morphologies on all the coatings, signaling appropriate adhesion to the surface. At 14 days of culture, cells cultured on ZrOx (Figure 5o) exhibited a slight cell detachment phenomenon occurring in the center of the cell monolayer, probably due to a significantly higher cell density in that area. Cells appeared to be well adhered to the ZrOx surface, other than the center; however, cell morphology seemed to be more elongated and in a more compact array compared to cells cultured on TiOx, TaOx, or NbOx.
Figure 6 presents a qualitative evaluation of cell viability by the LIVE/DEAD® fluorescence assay. Viable cells are marked in green, and dead cells are marked in red. Upon cell culture on uncoated Si samples, cells were viable; however, they presented a more rounded morphology and became easily detached from the surface compared to the cells cultured on the coatings. In addition, a smaller number of viable cells and a higher number of dead cells were observed at all culture times on Si compared to the coatings. Upon culture on the coatings, cells were viable at all culture times studied, showing just a few dead cells. In concordance with what was observed by optical profilometry (Figure 5), LIVE/DEAD assay showed that cell density increased on the coatings surfaces with culture time, and corroborated that cells were not only well-adhered to the coatings and increasing their population with culture time, but also were viable at all culture times on the coatings.
Cell viability was also quantitatively assessed using the Alamar Blue reagent. For this assay, BM-MSC cultured on uncoated Si were not considered due to the negative results observed for this surface in the qualitative evaluations of cell adhesion and viability (Figure 5 and Figure 6). According to Figure 7a, the number of viable cells at 3 days of culture was similar between all coatings and the Ctrl; no significant differences were found in the statistical analysis. The Ctrl corresponded to BM-MSC cultured on the surface of standard TCP. After 7 days of culture (Figure 7b), the number of viable cells increased for all coated surfaces and the Ctrl; the number of viable cells was comparable for all coatings studied and the Ctrl with no significant differences. An increase in the number of cells with culture days indirectly indicates cell proliferation. At 14 days of culture (Figure 7c), the number of viable cells increased on all coatings compared to 7 days of culture; however, the number of viable cells on the coatings was significantly smaller than that on the Ctrl. It is essential to mention that in no case the number of viable cells on the coatings was smaller than 50% of the number of viable cells on the Ctrl. TaOx and NbOx showed the highest number of viable cells (there were no significant differences between them) among the coatings studied, followed by ZrOx, and finally TiOx, which showed the smallest number of viable cells; the number of viable cells was not significantly different between TiOx and ZrOx.

3.4. Proliferation of BM-MSC Cultured on the Coatings

Cell proliferation upon culture on the coatings was evaluated using the Cell Proliferation ELISA, BrdU (colorimetric) kit at 3 and 5 days of BM-MSC culture. Ctrl corresponds to BM-MSC cultured on the surface of standard TCP. Figure 8a shows that at 3 days of culture, a higher percentage of the viable cells on TaOx and NbOx were in a proliferative state (BrdU assay absorbance read) since, according to Figure 7, the total number of cells was similar to the Ctrl up to 7 days of culture. However, there was no significant difference in the percentage of viable cells in a proliferative state among the different coatings studied. At 5 days of culture (Figure 8b), cells cultured on all the coatings exhibited a significantly smaller percentage of viable cells in a proliferative state compared to the Ctrl. The observed decrease in cell proliferation rate at 5 days compared to the control might be that cells attached to the coating’s surface have already initiated the differentiation process, as observed in the next section. Among the cells cultured on the coatings, NbOx showed the largest percentage of viable cells in a proliferative state at 5 days of culture, and this difference was significant against TiOx, TaOx, and ZrOx. After NbOx, the second-largest percentage of viable cells in a proliferative state among the coatings was observed on ZrOx, which exhibited a significantly larger percentage of viable cells in a proliferative state in comparison to TiOx and TaOx. There was no significant difference in the percentage of viable cells in a proliferative state between TiOx and TaOx.

3.5. Osteogenic Properties of the Coatings

Cell differentiation toward the osteogenic lineage was evaluated by qualitative IF and quantitative ELISA assays. Figure 9 displays representative micrographs of the expression of RUNX2, OP, and OC osteogenic markers from BM-MSC cultured on the coatings. RUNX2 is a transcription factor; thus, secondary antibody labeling is observed in and around the cell nucleus (Figure 9b–e). OP (Figure 9g–j) and OC proteins are expressed mainly in the cell cytoplasm (Figure 9l–o). Figure 9a,f,k correspond to BM-MSC cultured on TCP (Ctrl), which allowed for the evaluation of the basal expression of the osteogenic markers studied from BM-MSC. It has been described that FBS used in the culture medium might have induced a certain level of osteogenic differentiation [79]. Then, for the present evaluation, cells were incubated with DMEM/F-12 supplemented with 2% FBS; thus, the potential of osteogenic differentiation observed can be mainly attributed to the external stimulus studied, in this case, the contact with the coatings. According to the results observed in Figure 9, there was a positive expression of RUNX2, OP and OC in all the coatings studied. This positive expression of osteogenic differentiation markers can be confidently correlated to the effect of the coatings since the Ctrl presented a minimal expression of OP and OC, and no expression of RUNX2. Cellular controls for the IF assays are shown in the Supplementary Information (Supplementary Figures S3 and S4).
Osteogenic differentiation was also quantitatively evaluated. Figure 10 shows the quantification of some proteins involved in the differentiation process of MSC toward the osteogenic lineage. For protein quantification, supernatants were collected, and ELISA commercial kits were used; a colorimetric assay kit was used for quantification of ALP activity. Two independent comparison cell controls were set, consisting of BM-MSC cultured on TCP with a supplemented Mesenchymal Stem Basal Medium (Ctrl-) and BM-MSC cultured on TCP with DMEM/F-12 supplemented with 10% v/v FBS (Ctrl). So, it was possible to compare the proteins intrinsically expressed by mesenchymal stem cells (Ctrl-) with the basal expression of osteogenic markers expressed by mesenchymal stem cells incubated in FBS-supplemented medium (Ctrl), which might per se induced a certain level of osteogenic differentiation.
Regarding the results for ALP activity at 7 days of cell culture (Figure 10a), there was no ALP activity detected in the Ctrl-, and the largest ALP activity was observed on TiOx and ZrOx, which presented similar ALP activity between them, but significantly larger ALP activity compared to the Ctrl. ALP activity on NbOx was smaller than that TiOx and ZrOx but still significantly larger than the Ctrl. ALP activity on TaOx was similar to that of the Ctrl. At 14 days of cell culture (Figure 10b), there was no ALP activity detected in the Ctrl-, while ALP activity was detected on all coatings, with TaOx, presenting the largest value. The ALP activity decreased from TaOx to NbOx, TiOx and ZrOx; NbOx and TiOx presented a higher ALP than the Ctrl, while ZrOx exhibited a similar ALP activity to that of the Ctrl.
In the case of OP (Figure 10c), protein expression in the Ctrl- was significantly smaller than that for the Ctrl and the coatings. TaOx and NbOx expressed the highest amount of OP, exhibiting significantly larger OP expression than cells cultured on TiOx, ZrOx or the Ctrl. Cells cultured on ZrOx and TiOx expressed a significantly smaller amount of OP than the Ctrl. For OPG (Figure 10d), the most extensive protein expression was observed for cells cultured on NbOx, where OPG expression was significantly larger compared to all the other coatings studied and to the Ctrl. OPG expression from cells cultured on TaOx was similar to that observed for the Ctrl. OPG expression on ZrOx and TiOX was significantly smaller than that observed for the Ctrl and similar to the OPG expression observed for the Ctrl-. Finally, for OC (Figure 10e), protein expression was similar between the Ctrl- and the Ctrl. OC expression was significantly larger on all the coatings studied in comparison to the Ctrl. The most significant OC expression among the coatings corresponded to cells cultured on NbOx and TiOx, which expressed similar levels of OC between them, followed by OC expression from cells cultured on ZrOx and TaOx. Cells culture on TaOx expressed the smallest amount of OC among the different coatings studied; however, OC expression on TaOx was still significantly larger in comparison to the Ctrl.

4. Discussion

Several studies have searched for appropriate surfaces to promote the osseointegration process in orthopedic implants. It is known that surface modifications allow harnessing of the biological response that may significantly reduce implant failure or patient rejection. Although appropriate biological responses being triggered by certain surface modifications are known, there is still a long way to go in developing biomaterials with the optimal surface features that ensure a successful osseointegration process [80]. Many techniques have been described for obtaining positive surface modifications. Some of them add structures on the substrate surface (coatings), while others modify the existing ones (acid etching, sandblasting, etc.) [80]. In this work, surfaces coated with metal oxide thin films (Ti, Ta, Nb and Zr oxides) were generated by magnetron sputtering. This technique allows the deposition of coatings on various materials that meet the desirable mechanical properties for orthopedic implants and allows tailored surface chemistry by varying the deposition conditions [67,68,69,70,71]. Therefore, this technique represents a promising proposal for in vitro and in vivo investigation, and for its application at an industrial level in the generation of orthopedic implants.
To elucidate the correlations of the biological response observed toward the properties of the coating (TiO2, Ta2O5, Nb2O5, ZrO2), an appropriate physical and chemical characterization of the oxide coatings is very important. The present work characterized the coatings by SEM, EDS, XPS, XRD, wettability, and SFE calculations. The EDS and XPS analysis (Figure 2 and Figure 3) corroborated that the deposited oxide coatings only contained the expected chemical elements, and no traces of other elements were observed. Thus, the cellular response presented in this work can be precisely correlated to the corresponding chemical and elemental composition of the metal oxide coatings and the physical properties emerging from this.
The difficulty in modifying a particular property without altering others has caused the heterogeneity observed in published results, making it challenging to reach a consensus about the most significant surface properties a biomaterial must meet for developing orthopedic implants [58]. In this sense, to evaluate the effect of the chemical composition on the biocompatibility and differentiation of MSC cells, four metal oxide coatings were deposited on atomically flat Si (100) substrates. Coatings deposition on Si (100) substrates ruled out the influence of roughness and wettability in the biological response, as both could directly affect cell adhesion and proliferation [58,81,82,83]. Wettability due to variations in the surface roughness can directly affect cell adhesion and proliferation [58,81,82,83]. Meanwhile, it has been reported that roughness values between 0.1 and 100 µm favor cell adhesion (specifically in the range of 10–30 µm) [84], while cell adhesion decreased on rougher surfaces (≥100 µm) [58]. At the nanoscale (1–100 nm), roughness indirectly affects cell adhesion by controlling the adsorbed protein layer on the surface [75,76,77,85]. Results in Figure 5 and SFE data suggested that the excellent adhesion and proliferation of the BM-MSC on the coatings are related to the surface chemistry, the rougher surface, or the more hydrophobic character (protein adsorption related) of the coatings since, on the silicon surface (Figure 5b,c), cells did not remain adhered after 7 days of culture.
Regarding the coatings thickness values obtained (Table 2), all the coatings were thicker than 50 nm, being TiO2 the thinnest coating (54.5 nm). Coatings thickness can be correlated to corrosion rates and coating adhesion strength, which are desirable characteristics in an in vivo system. Generally, the thinner the coating, the greater its adhesion strength to the substrate, and the poorer its corrosion protection [86]. In addition, detachment of the coatings might cause the accumulation of debris in the surrounding tissue and consequently cause inflammation and exacerbate immunological responses [32]. Nevertheless, some reports relate the thickness of the coatings with the enhancement of osteogenic properties [87], and the increasing bone attachment to the implant in an in vivo system [88] but always related to another surface feature influenced by coatings thickness, such as surface zeta-potential and porous surface structure. For the evaluations carried out in this work, the thicknesses of the different coatings are not expected to directly affect the biological response as long as the coatings homogeneously cover the substrate surface. Thus, it was important to corroborate that covering of the substrate surface by the metal oxide coatings was complete and homogeneous, a characteristic corroborated by the replicated XPS and EDS measurements on different areas of different coated samples. According to our results, the thickness of the coatings does not seem to have a direct relationship with osteogenic differentiation, since both TiO2 and ZrO2 (Figure 10) presented a faster osteogenic induction than the other coatings, with TiO2 being the thinnest coating and ZrO2 one of the thickest.
Although the differences were not too significant, cell numbers after 14 days of incubation were larger on the amorphous Ta2O5 and Nb2O5 surfaces compared to the nanocrystalline TiO2 and ZrO2 (Figure 7) coatings. In some cases, it has been reported that crystalline structures favor cell adhesion and proliferation; in the case of TiO2 it has been described that the anatase phase increases cell adhesion [89,90]. However, the native TiO2 surface has been reported as amorphous [91], and in previous works, we showed that cell adhesion and proliferation were promoted on the amorphous TiO2 [58] and ZrO2 [50] coatings in comparison with their crystalline counterpart.
Other properties also crucial in cellular adhesion are wettability and surface-free energy. It is known that cells preferentially adhere to hydrophilic surfaces, while proteins tend to adsorb on hydrophobic surfaces efficiently [92]. Nevertheless, on hydrophilic surfaces, protein adsorption depends on the electrostatic interactions, and thus, pH variations in the physiological environment might improve protein adsorption by causing proteins to easily undergo conformational changes that might change their positive/negative charge facilitating their interaction with hydrophilic surfaces; depending on the surface charge [93]. According to the present wettability results (Table 4), hydrophilic oxide coatings seemed to favor cell adhesion and proliferation. All the coatings studied presented a hydrophilic nature, being ZrO2 the closest to the hydrophilic–hydrophobic limit. Of the coatings studied, Nb2O5 presented the highest hydrophilic character, and on this metal oxide, cells covered the largest surface area upon seeding (Figure 5j), and cells proliferation in the first days of incubation (3 and 5 days) was favored (Figure 8). These results are consistent with those reported in previous studies, where a higher cell proliferation was observed at 3 days of incubation on Nb2O5 compared to TiO2 [94].
Surface-free energy is involved in protein adsorption and cell adhesion. Proteins are known to more readily adhere to hydrophobic surfaces, but it has been described that on hydrophilic surfaces, the polar component of the surface free energy is directly correlated with the amount of protein adsorbed. In this way, even if the surface is hydrophilic, if its polar component is considerable within the value of the total surface free energy, adsorption of proteins will be favored. Cell adhesion and proliferation are enhanced with increasing SFE, where the polar component has an important role. As the SFE polar component increases, cell adhesion is favored due to easily established bonds between the carboxyl and hydroxyl groups present on the hydrated surface and the lipids and ions in the cell membrane [95]. Table 4, shows the SFE values of the coatings and the polarity factor (γp/(γp +γd)). Ta2O5 and Nb2O5 presented the highest contribution of the polar component to the SFE. In the viability assay at 3 days of incubation (Figure 7a), more viable cells were observed on Ta2O5 and Nb2O5, in comparison to the other oxides, which can be addressed to the fact that these two oxides have the largest value of the polarity factor. At 7 days of incubation (Figure 7b), a higher number of viable cells was observed on Ta2O5 compared to the other oxides. All the coatings generated in the present study can be expected to favor cell adhesion and proliferation, being hydrophilic. However, according to the coatings’ surface energy values (dispersive and polar component), it can be hypothesized that, although Ta2O5 presented a hydrophilic nature, its higher polar component of the surface free energy might have favored appropriate adsorption of proteins, and the consequent appropriate biological response observed [93].
When evaluating the viability of BM-MSC cultured on the coatings (Figure 6 and Figure 7), in all cases, cells remained viable up to 14 days of culture, with Ta2O5 being the coating with the most significant number of viable cells at day 14, followed by Nb2O5. On the other hand, Nb2O5, presented a larger number of proliferative cells at 14 days of cell culture, which had also been previously reported [94]. It is essential to notice that cell proliferation decreases as cell differentiation starts; thus, decreased cell number observed on TiO2, ZrO2 and on Nb2O5 at 14 days of cell culture might be also correlated with a faster differentiation of BM-MSC cells occurring on TiO2, ZrO2, Ta2O5 and finally Nb2O5
It is not only desirable that a biomaterial intended to be used in orthopedic implants allows cell adhesion and proliferation, but it is also desirable that it favors the whole osseointegration process, for example, by improving osteoinduction. Figure 9 shows that all the coatings induced a positive expression of the evaluated osteogenic markers (RUNX2, OP, OC, and ALP), which qualitatively demonstrated that on all the four coatings studied, BM-MSC were differentiating towards the osteogenic lineage. Figure 10 shows the quantification of some proteins that participate at different stages of the osteogenesis (osteoinduction) process, which helps us to infer how the differentiation process took place on each of the coatings studied. RUNX2 and ALP are early osteogenic differentiation markers, while OP, OPG, and OC are considered late differentiation markers. RUNX2 is a transcription factor that regulates the differentiation of MSC toward preosteoblasts and regulates the expression of latter osteogenic markers [96]. ALP is a glycoprotein secreted by preosteoblasts and osteoblasts, and it is directly involved in mineralization, in addition to the fact that during osteogenesis, the presence of this enzyme is indicative of a differentiation process of MSC toward osteoblasts [79,97]. OP is an extracellular matrix protein secreted by mature osteoblasts and is involved in various processes, including bone mineralization [98,99]. It has been described that OP is directly involved in improving the osteogenesis process during osseointegration in dental implants [100]. OPG is a protein found in various organs, mainly bone and it is known to be released by osteoblasts, and function as an antagonist of osteclastogenesis [101]. Finally, OC is the most abundant non-collagenous protein in mature bone, and osteoblasts specifically synthesize it; although it has been mainly related to bone formation, several processes in which it could be involved have been also described [102]. According to the protein quantification results in the present study (Figure 10), after 7 days of BM-MSC culture, ZrO2 and TiO2 exerted the highest ALP activity, indicating that on these two coatings, the induction of osteogenesis might have started earlier than on Ta2O5 and Nb2O5. However, by 14 days of culture Ta2O5 was the one presenting the highest ALP activity, as on this coating the differentiation process started later than in the other coatings. When comparing ZrO2 and TiO2, only TiO2 was the one that presented the highest amount of OC at 14 days of cultures. Thus, probably ZrO2 stopped, or was in a later state of, differentiation by 14 days of culture; ALP activity on ZrO2 at 7 days of culture was the largest. In Figure 10a, it is observed that Nb2O5 presented a higher ALP activity with respect to Ta2O5, at 14 days of incubation, it presented one of the highest expressions of OP, OPG and OC. Finally, Ta2O5 at 14 days of incubation presented the highest activity of ALP and OP. For OPG it had a lower expression than Nb2O5, while for OC, Ta2O5 was the coating with the lowest expression. In this way, it is possible to propose that the induction of osteogenesis was faster in ZrO2, followed by TiO2, then Nb2O5 and finally Ta2O5. This osteogenic differentiation process is also supported by the viability assay, in which at 14 days of incubation (Figure 7c), a lower number of cells on TiO2 and ZrO2, followed by Nb2O5 and Ta2O5 was observed. It has been reported that after 9 days, cell line changes phenotype from proliferative to differentiative [103], process in which a decrease in proliferation occurs to allow differentiation.
Although TiO2 continues to be one of the best options for manufacturing orthopedic implants, reports warn about its toxicity [30]. In most cases, these problems are related to the damage in the oxide layer due to corrosion, causing the release of particles that lead to peri-implantitis [32]. According to the present results, ZrO2 represents a potential alternative to TiO2, displaying ZrO2 a similar osteoinduction process to that of TiO2; nevertheless, further studies on the mechanisms and timewise proceeding of the differentiation process on these oxides are needed. Different studies have reported characteristics of ZrO2 that can be comparable with those of TiO2. Among the most relevant features, it was reported a higher adhesion of MG63 cells and higher ALP activity at 4 days cultured on ZrO2 films compared to TiO2 films [104], in addition ZrO2has a higher corrosion resistance due to its higher electrochemical stability and surface integrity, which significantly reduces toxicity problems [54]. It has also been described that zirconia layers inhibit bacterial adhesion, which is a very desirable characteristic [105].

5. Conclusions

This study shows that cell adhesion, proliferation, and osteogenic differentiation can be promoted by the four metal oxide coatings evaluated: TiO2, which can be considered the goal standard to compare, ZrO2, Ta2O5, and Nb2O5. Small but statistically significant differences were observed in their biological responses, suggesting ZrO2 as an alternative substitute for TiO2, with the advantage of a faster differentiation process. ZrO2 presented the larger water contact (close to 90°) and nano crystalline structure, but intermediate SFE, polarity factor, and thickness among the oxides. Thus, the first two parameters seem to be more critical for osteogenic differentiation.
From our analysis, it can also be observed that the most polar surface (polarity factor of 34%), Ta2O5, promoted a large number of attached cells, which might be relevant for cell culture applications.
To date, few studies have studied various metal oxides under comparable experimental conditions as possible candidates for potential orthopedic implant applications. Therefore, the results presented in this work represent a starting point for knowing and improving the surface properties of metal oxides for orthopedic applications.

Supplementary Materials

The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/ma15155240/s1, Figures S1 and S2: Bone-marrow derived mesenchymal stem cells characterization and Figures S3 and S4: Immunofluorescence assays controls.

Author Contributions

Conceptualization, S.E.R. and P.S.-B.; Data curation, M.F.-L.; Formal analysis, M.F.-L., J.G.-L., R.M.R.-A. and P.S.-B.; Funding acquisition, S.E.R., R.M.R.-A. and P.S.-B.; Investigation, M.F.-L., S.E.R., R.M.R.-A. and P.S.-B.; Methodology, M.F.-L. and J.G.-L.; Resources, S.E.R., R.M.R.-A. and P.S.-B.; Supervision, J.G.-L. and P.S.-B.; Writing—original draft, M.F.-L. and P.S.-B.; Writing—review & editing, S.E.R., R.M.R.-A. and P.S.-B. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by CONACyT-CB-288101 and partially funded by CONACYT-Frontera-1740.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The data presented in this study are available on request from the corresponding author. The data are not publicly available due to ongoing research.

Acknowledgments

This work was supported by CONACyT-CB-288101 and partially supported by CONACYT-Frontera-1740, and DGAPA-PAPIIT-IN101419. R.M.R.-A. is recipient of Comisión de Operación y Fomento de Actividades Académicas (COFAA-IPN) and Estímulo al Desempeño Docente-IPN grants, and part of this work was supported by her SIP-IPN grants 20211631 and 20221624. We thank technical support from O. Novelo, L. Huerta and A. Tejeda from the IIM-UNAM, and technical support from X. Guerrero-Alva from the INR-LGII. The author also acknowledges the Laboratorio de Biotecnología in the INR-LGII for the facilities (Axi.Observer Microscope and Synergy-HTX multi-mode reader spectrophotometer) to carry out some of the experimental parts presented in this research. Advice on XRD data analysis from O. Deplablos-Rivera from the Chemistry Faculty at UNAM is acknowledged. Technical support and advice for cytometry experiments from V. Martínez-López from the INR-LGII is also acknowledged. Fernández-Lizárraga M. acknowledges the support from the program Posgrado de Doctorado en Ciencias en Biomedicina y Biotecnología Molecular, Escuela Nacional de Ciencias Biológicas at the Instituto Politécnico Nacional and the Ph.D. scholarship from CONACyT (CVU 739515).

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Grün, N.G.; Holweg, P.L.; Donohue, N.; Klestil, T.; Weinberg, A.-M. Resorbable Implants in Pediatric Fracture Treatment. Innov. Surg. Sci. 2018, 3, 119–125. [Google Scholar] [CrossRef] [PubMed]
  2. Chandra, G.; Pandey, A. Design Approaches and Challenges for Biodegradable Bone Implants: A Review. Expert Rev. Med. Devices 2021, 18, 629–647. [Google Scholar] [CrossRef] [PubMed]
  3. Siti Nur Hazwani, M.R.; Lim, L.X.; Lockman, Z.; Zuhailawati, H. Fabrication of Titanium-Based Alloys with Bioactive Surface Oxide Layer as Biomedical Implants: Opportunity and Challenges. Trans. Nonferrous Met. Soc. China Engl. Ed. 2022, 32, 1–44. [Google Scholar] [CrossRef]
  4. Liu, Y.; Rath, B.; Tingart, M.; Eschweiler, J. Role of Implants Surface Modification in Osseointegration: A Systematic Review. J. Biomed. Mater. Res. Part A 2020, 108, 470–484. [Google Scholar] [CrossRef] [Green Version]
  5. Rizzo, P. A Review on the Latest Advancements in the Non-Invasive Evaluation/Monitoring of Dental and Trans-Femoral Implants. Biomed. Eng. Lett. 2020, 10, 83–102. [Google Scholar] [CrossRef]
  6. Silva, R.C.S.; Agrelli, A.; Andrade, A.N.; Mendes-Marques, C.L.; Arruda, I.R.S.; Santos, L.R.L.; Vasconcelos, N.F.; Machado, G. Titanium Dental Implants: An Overview of Applied Nanobiotechnology to Improve Biocompatibility and Prevent Infections. Materials 2022, 15, 3150. [Google Scholar] [CrossRef]
  7. Yadav, R.; Meena, A.; Patnaik, A. Biomaterials for Dental Composite Applications: A Comprehensive Review of Physical, Chemical, Mechanical, Thermal, Tribological, and Biological Properties. Polym. Adv. Technol. 2022, 33, 1762–1781. [Google Scholar] [CrossRef]
  8. Wang, L.; Ding, X.; Feng, W.; Gao, Y.; Zhao, S.; Fan, Y. Biomechanical Study on Implantable and Interventional Medical Devices. Acta Mech. Sin. Xuebao 2021, 37, 875–894. [Google Scholar] [CrossRef]
  9. Tipan, N.; Pandey, A.; Mishra, P. Selection and Preparation Strategies of Mg-Alloys and Other Biodegradable Materials for Orthopaedic Applications: A Review. Mater. Today Commun. 2022, 31, 103658. [Google Scholar] [CrossRef]
  10. Yoshinari, M. Future Prospects of Zirconia for Oral Implants—A Review. Dent. Mater. J. 2020, 39, 37–45. [Google Scholar] [CrossRef] [Green Version]
  11. Albrektsson, T.; Johansson, C. Osteoinduction, Osteoconduction and Osseointegration. Eur. Spine J. 2001, 10, S96–S101. [Google Scholar] [CrossRef] [Green Version]
  12. Melo-Fonseca, F.; Gasik, M.; Madeira, S.; Silva, F.S.; Miranda, G. Surface Characterization of Titanium-Based Substrates for Orthopaedic Applications. Mater. Charact. 2021, 177, 111161. [Google Scholar] [CrossRef]
  13. Romanò, C.L.; Scarponi, S.; Gallazzi, E.; Romanò, D.; Drago, L. Antibacterial Coating of Implants in Orthopaedics and Trauma: A Classification Proposal in an Evolving Panorama. J. Orthop. Surg. Res. 2015, 10, 1–11. [Google Scholar] [CrossRef] [Green Version]
  14. Sultana, A.; Zare, M.; Luo, H.; Ramakrishna, S. Surface Engineering Strategies to Enhance the in Situ Performance of Medical Devices Including Atomic Scale Engineering. Int. J. Mol. Sci. 2021, 22, 11788. [Google Scholar] [CrossRef]
  15. Stepanovska, J.; Matejka, R.; Rosina, J.; Bacakova, L.; Kolarova, H. Treatments for Enhancing the Biocompatibility of Titanium Implants. Biomed. Pap. 2020, 164, 23–33. [Google Scholar] [CrossRef] [Green Version]
  16. Dehghanghadikolaei, A.; Fotovvati, B. Coating Techniques for Functional Enhancement of Metal Implants for Bone Replacement: A Review. Materials 2019, 12, 1795. [Google Scholar] [CrossRef] [Green Version]
  17. Kumar, M.; Kumar, R.; Kumar, S. Coatings on Orthopedic Implants to Overcome Present Problems and Challenges: A Focused Review. Mater. Today Proc. 2021, 45, 5269–5276. [Google Scholar] [CrossRef]
  18. Liu, Z.; Liu, X.; Ramakrishna, S. Surface Engineering of Biomaterials in Orthopedic and Dental Implants: Strategies to Improve Osteointegration, Bacteriostatic and Bactericidal Activities. Biotechnol. J. 2021, 16, 2000116. [Google Scholar] [CrossRef]
  19. Blendinger, F.; Seitz, D.; Ottenschläger, A.; Fleischer, M.; Bucher, V. Atomic Layer Deposition of Bioactive TiO2Thin Films on Polyetheretherketone for Orthopedic Implants. ACS Appl. Mater. Interfaces 2021, 13, 3536–3546. [Google Scholar] [CrossRef]
  20. Chen, C.S.; Chang, J.H.; Srimaneepong, V.; Wen, J.Y.; Tung, O.H.; Yang, C.H.; Lin, H.C.; Lee, T.H.; Han, Y.; Huang, H.H. Improving the in Vitro Cell Differentiation and in Vivo Osseointegration of Titanium Dental Implant through Oxygen Plasma Immersion Ion Implantation Treatment. Surf. Coatings Technol. 2020, 399, 126125. [Google Scholar] [CrossRef]
  21. Chernozem, R.V.; Surmeneva, M.A.; Ignatov, V.P.; Peltek, O.O.; Goncharenko, A.A.; Muslimov, A.R.; Timin, A.S.; Tyurin, A.I.; Ivanov, Y.F.; Grandini, C.R.; et al. Comprehensive Characterization of Titania Nanotubes Fabricated on Ti-Nb Alloys: Surface Topography, Structure, Physicomechanical Behavior, and a Cell Culture Assay. ACS Biomater. Sci. Eng. 2020, 6, 1487–1499. [Google Scholar] [CrossRef]
  22. Gomez Sanchez, A.; Katunar, M.R.; Pastore, J.I.; Tano de la Hoz, M.F.; Ceré, S. Evaluation of Annealed Titanium Oxide Nanotubes on Titanium: From Surface Characterization to in Vivo Assays. J. Biomed. Mater. Res. Part A 2021, 109, 1088–1100. [Google Scholar] [CrossRef]
  23. İzmir, M.; Ercan, B. Anodization of Titanium Alloys for Orthopedic Applications. Front. Chem. Sci. Eng. 2019, 13, 28–45. [Google Scholar] [CrossRef]
  24. Mathai, S.; Shaji, P.S. Different Coating Methods of Titanium Dioxide on Metal Substrates for Orthopedic and Dental Applications: A Review. Asian J. Chem. 2022, 34, 9–17. [Google Scholar] [CrossRef]
  25. Pesode, P.; Barve, S. Surface Modification of Titanium and Titanium Alloy by Plasma Electrolytic Oxidation Process for Biomedical Applications: A Review. Mater. Today Proc. 2021, 46, 594–602. [Google Scholar] [CrossRef]
  26. Kurup, A.; Dhatrak, P.; Khasnis, N. Surface modification techniques of titanium and titanium alloys for biomedical dental applications: A review. Mater. Today Proc. 2020, 39, 84–90. [Google Scholar] [CrossRef]
  27. Prestat, M.; Thierry, D. Corrosion of Titanium under Simulated Inflammation Conditions: Clinical Context and in Vitro Investigations. Acta Biomater. 2021, 136, 72–87. [Google Scholar] [CrossRef]
  28. Burstein, G.T.; Souto, R.M. Observations of Localised Instability of Passive Titanium in Chloride Solution. Electrochim. Acta 1995, 40, 1881–1888. [Google Scholar] [CrossRef]
  29. Vrchovecká, K.; Weiser, A.; Přibyl, J.; Kuta, J.; Holzer, J.; Pávková-Goldbergová, M.; Sobola, D.; Dlouhý, A. A Release of Ti-Ions from Nanostructured Titanium Oxide Surfaces. Surf. Interfaces 2022, 29, 101699. [Google Scholar] [CrossRef]
  30. Kim, K.T.; Eo, M.Y.; Nguyen, T.T.H.; Kim, S.M. General Review of Titanium Toxicity. Int. J. Implant Dent. 2019, 5, 10. [Google Scholar] [CrossRef] [Green Version]
  31. Comino-Garayoa, R.; Brinkmann, J.C.B.; Peláez, J.; López-Suárez, C.; Martínez-González, J.M.; Suárez, M.J. Allergies to Titanium Dental Implants: What Do We Really Know about Them? A Scoping Review. Biology 2020, 9, 904. [Google Scholar] [CrossRef] [PubMed]
  32. Xu, A.; Alhamad, M.; Ampadi Ramachandran, R.; Shukla, A.; Barão, V.A.; Sukotjo, C.; Mathew, M.T. Peri-Implantitis in Relation to Titanium Corrosion: Current Status and Future Perspectives. J. Bio-Tribo-Corros. 2022, 8, 46. [Google Scholar] [CrossRef]
  33. Ding, Z.; Tang, Y.; Liu, L.; Ding, Z.; Tan, Y.; He, Q. Improving the Adhesive, Mechanical, Tribological Properties and Corrosion Resistance of Reactive Sputtered Tantalum Oxide Coating on Ti6Al4V Alloy via Introducing Multiple Interlayers. Ceram. Int. 2022, 48, 5983–5994. [Google Scholar] [CrossRef]
  34. Chen, Y.W.; Moussi, J.; Drury, J.L.; Wataha, J.C. Zirconia in Biomedical Applications. Expert Rev. Med. Devices 2016, 13, 945–963. [Google Scholar] [CrossRef]
  35. Mani, G.; Porter, D.; Grove, K.; Collins, S.; Ornberg, A.; Shulfer, R. A Comprehensive Review of Biological and Materials Properties of Tantalum and Its Alloys. J. Biomed. Mater. Res. Part A 2022, 110, 1291–1306. [Google Scholar] [CrossRef]
  36. Grobelny, M.; Kalisz, M.; Mazur, M.; Wojcieszak, D.; Kaczmarek, D.; Domaradzki, J.; Świniarski, M.; Mazur, P. Functional Nb2O5 Film and Nb2O5 + CuO, Nb2O5 + Graphene, Nb2O5 + CuO + Graphene Composite Films to Modify the Properties of Ti6Al4V Titanium Alloy. Thin Solid Films 2016, 616, 64–72. [Google Scholar] [CrossRef]
  37. Fomina, M.; Koshuro, V.; Shumilin, A.; Voyko, A.; Zakharevich, A.; Skaptsov, A.; Steinhauer, A.; Fomin, A. Functionally Graded “Ti-Base + (Ta, Ta2O5)-Coatings” Structure and Its Production Using Induction Heat Treatment. Compos. Struct. 2020, 234, 111688. [Google Scholar] [CrossRef]
  38. Sarraf, M.; Razak, B.A.; Nasiri-Tabrizi, B.; Dabbagh, A.; Kasim, N.H.A.; Basirun, W.J.; Bin Sulaiman, E. Nanomechanical Properties, Wear Resistance and in-Vitro Characterization of Ta2O5 Nanotubes Coating on Biomedical Grade Ti–6Al–4V. J. Mech. Behav. Biomed. Mater. 2017, 66, 159–171. [Google Scholar] [CrossRef]
  39. Pereira, B.L.; da Luz, A.R.; Lepienski, C.M.; Mazzaro, I.; Kuromoto, N.K. Niobium Treated by Plasma Electrolytic Oxidation with Calcium and Phosphorus Electrolytes. J. Mech. Behav. Biomed. Mater. 2018, 77, 347–352. [Google Scholar] [CrossRef]
  40. Pieralli, S.; Kohal, R.J.; Lopez Hernandez, E.; Doerken, S.; Spies, B.C. Osseointegration of Zirconia Dental Implants in Animal Investigations: A Systematic Review and Meta-Analysis. Dent. Mater. 2018, 34, 171–182. [Google Scholar] [CrossRef]
  41. de Almeida Bino, M.C.; Eurídice, W.A.; Gelamo, R.V.; Leite, N.B.; da Silva, M.V.; de Siervo, A.; Pinto, M.R.; de Almeida Buranello, P.A.; Moreto, J.A. Structural and Morphological Characterization of Ti6Al4V Alloy Surface Functionalization Based on Nb2O5 Thin Film for Biomedical Applications. Appl. Surf. Sci. 2021, 55, 149739. [Google Scholar] [CrossRef]
  42. Beline, T.; da Silva, J.H.D.; Matos, A.O.; Azevedo Neto, N.F.; de Almeida, A.B.; Nociti Júnior, F.H.; Leite, D.M.G.; Rangel, E.C.; Barão, V.A.R. Tailoring the Synthesis of Tantalum-Based Thin Films for Biomedical Application: Characterization and Biological Response. Mater. Sci. Eng. C 2019, 101, 111–119. [Google Scholar] [CrossRef]
  43. Wang, X.; Ning, B.; Pei, X. Tantalum and Its Derivatives in Orthopedic and Dental Implants: Osteogenesis and Antibacterial Properties. Colloids Surf. B Biointerfaces 2021, 208, 112055. [Google Scholar] [CrossRef]
  44. Almeida Alves, C.F.; Fialho, L.; Marques, S.M.; Pires, S.; Rico, P.; Palacio, C.; Carvalho, S. MC3T3-E1 Cell Response to Microporous Tantalum Oxide Surfaces Enriched with Ca, P and Mg. Mater. Sci. Eng. C 2021, 124, 112008. [Google Scholar] [CrossRef]
  45. Antonini, L.M.; Menezes, T.L.; dos Santos, A.G.; Takimi, A.S.; Villarinho, D.J.; dos Santos, B.P.; Camassola, M.; Marcuzzo, J.S.; de Fraga Malfatti, C. Osteogenic Differentiation of Bone Marrow-Derived Mesenchymal Stem Cells on Anodized Niobium Surface. J. Mater. Sci. Mater. Med. 2019, 30, 2–13. [Google Scholar] [CrossRef]
  46. Fialho, L.; Grenho, L.; Fernandes, M.H.; Carvalho, S. Porous Tantalum Oxide with Osteoconductive Elements and Antibacterial Core-Shell Nanoparticles: A New Generation of Materials for Dental Implants. Mater. Sci. Eng. C 2021, 120, 111761. [Google Scholar] [CrossRef]
  47. Wang, F.; Li, C.; Zhang, S.; Liu, H. Tantalum Coated on Titanium Dioxide Nanotubes by Plasma Spraying Enhances Cytocompatibility for Dental Implants. Surf. Coatings Technol. 2020, 382, 125161. [Google Scholar] [CrossRef]
  48. Horandghadim, N.; Khalil-Allafi, J.; Urgen, M. Effect of Ta2O5 Content on the Osseointegration and Cytotoxicity Behaviors in Hydroxyapatite-Ta2O5 Coatings Applied by EPD on Superelastic NiTi Alloys. Mater. Sci. Eng. C 2019, 102, 683–695. [Google Scholar] [CrossRef]
  49. Huang, H.L.; Tsai, M.T.; Chang, Y.Y.; Lin, Y.J.; Hsu, J.T. Fabrication of a Novel Ta(Zn)O Thin Film on Titanium by Magnetron Sputtering and Plasma Electrolytic Oxidation for Cell Biocompatibilities and Antibacterial Applications. Metals 2020, 10, 649. [Google Scholar] [CrossRef]
  50. Vitti, R.P.; Catelan, A.; Amaral, M.; Pacheco, R.R. Zirconium in dentistry. Adv. Dent. Biomater. 2019, 317–345. [Google Scholar] [CrossRef]
  51. Saleem, S.; Ahmad, R.; Ayub, R.; Ikhlaq, U.; Jin, W.; Chu, P.K. Investigation of Nano-Structured Zirconium Oxide Film on Ti6Al4V Substrate to Improve Tribological Properties Prepared by PIII&D. Appl. Surf. Sci. 2017, 394, 586–597. [Google Scholar] [CrossRef]
  52. Maminskas, J.; Pilipavicius, J.; Staisiunas, E.; Baranovas, G.; Alksne, M.; Daugela, P.; Juodzbalys, G. Novel Yttria-Stabilized Zirconium Oxide and Lithium Disilicate Coatings on Titanium Alloy Substrate for Implant Abutments and Biomedical Application. Materials 2020, 13, 2070. [Google Scholar] [CrossRef]
  53. Peron, M.; Bertolini, R.; Cogo, S. On the Corrosion, Stress Corrosion and Cytocompatibility Performances of ALD TiO2 and ZrO2 Coated Magnesium Alloys. J. Mech. Behav. Biomed. Mater. 2022, 125, 104945. [Google Scholar] [CrossRef]
  54. Peron, M.; Cogo, S.; Bjelland, M.; Bin Afif, A.; Dadlani, A.; Greggio, E.; Berto, F.; Torgersen, J. On the Evaluation of ALD TiO2, ZrO2 and HfO2 Coatings on Corrosion and Cytotoxicity Performances. J. Magnes. Alloy. 2021, 9, 1806–1819. [Google Scholar] [CrossRef]
  55. Safavi, M.S.; Walsh, F.C.; Visai, L.; Khalil-Allafi, J. Progress in Niobium Oxide-Containing Coatings for Biomedical Applications: A Critical Review. ACS Omega 2022, 7, 9088–9107. [Google Scholar] [CrossRef]
  56. Senocak, T.C.; Ezirmik, K.V.; Aysin, F.; Simsek Ozek, N.; Cengiz, S. Niobium-Oxynitride Coatings for Biomedical Applications: Its Antibacterial Effects and in-Vitro Cytotoxicity. Mater. Sci. Eng. C 2021, 120, 111662. [Google Scholar] [CrossRef]
  57. Almaguer-Flores, A.; Silva-Bermudez, P.; Galicia, R.; Rodil, S.E. Bacterial Adhesion on Amorphous and Crystalline Metal Oxide Coatings. Mater. Sci. Eng. C 2015, 57. [Google Scholar] [CrossRef]
  58. Silva-Bermudez, P.; Almaguer-Flores, A.; Garcia, V.I.; Olivares-Navarrete, R.; Rodil, S.E. Enhancing the Osteoblastic Differentiation through Nanoscale Surface Modifications. J. Biomed. Mater. Res. Part A 2017, 105, 498–509. [Google Scholar] [CrossRef]
  59. Chang, Y.-Y.; Huang, H.-L.; Chen, H.-J.; Lai, C.-H.; Wen, C.-Y. Antibacterial Properties and Cytocompatibility of Tantalum Oxide Coatings. Surf. Coatings Technol. 2014, 259, 193–198. [Google Scholar] [CrossRef]
  60. Donkov, N.; Zykova, A.; Safonov, V.; Yakovin, S.; Dudin, S.; Melnikova, G.; Petrovskaya, A.; Kuznetsova, T.; Chizhik, S.A. Antibacterial Properties of Ta-Based Ceramic Coatings Deposited by Magnetron Sputtering. J. Phys. Conf. Ser. 2021, 1859, 012062. [Google Scholar] [CrossRef]
  61. Huang, H.L.; Tsai, M.T.; Lin, Y.J.; Chang, Y.Y. Antibacterial and Biological Characteristics of Tantalum Oxide Coated Titanium Pretreated by Plasma Electrolytic Oxidation. Thin Solid Films 2019, 688, 137268. [Google Scholar] [CrossRef]
  62. Chen, X.; Fan, H.; Deng, X.; Wu, L.; Yi, T.; Gu, L.; Zhou, C.; Fan, Y.; Zhang, X. Scaffold Structural Microenvironmental Cues to Guide Tissue Regeneration in Bone Tissue Applications. Nanomaterials 2018, 8, 960. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  63. Sarakinos, K.; Alami, J.; Konstantinidis, S. High Power Pulsed Magnetron Sputtering: A Review on Scientific and Engineering State of the Art. Surf. Coat. Technol. 2010, 204, 1661–1684. [Google Scholar] [CrossRef]
  64. Greene, J.E. Tracing the Recorded History of Thin-Film Sputter Deposition: From the 1800s to 2017. J. Vac. Sci. Technol. A Vac. Surf. Film. 2017, 35, 05C204. [Google Scholar] [CrossRef] [Green Version]
  65. Gudmundsson, J.T. Physics and Technology of Magnetron Sputtering Discharges. Plasma Sources Sci. Technol. 2020, 29, 113001. [Google Scholar] [CrossRef]
  66. Lin, J.; Sproul, W.D.; Moore, J.J.; Wu, Z.; Lee, S.; Chistyakov, R.; Abraham, B. Recent Advances in Modulated Pulsed Power Magnetron Sputtering for Surface Engineering. JOM 2011, 63, 48–58. [Google Scholar] [CrossRef]
  67. Vetter, J.; Eriksson, A.O.; Reiter, A.; Derflinger, V.; Kalss, W. Quo Vadis: Alcr-based Coatings in Industrial Applications. Coatings 2021, 11, 344. [Google Scholar] [CrossRef]
  68. Larhlimi, H.; Ghailane, A.; Makha, M.; Alami, J. Magnetron Sputtered Titanium Carbide-Based Coatings: A Review of Science and Technology. Vacuum 2022, 197, 110853. [Google Scholar] [CrossRef]
  69. Behera, A.; Rajak, D.K.; Kolahchi, R.; Scutaru, M.L.; Pruncu, C.I. Current Global Scenario of Sputter Deposited NiTi Smart Systems. J. Mater. Res. Technol. 2020, 9, 14582–14598. [Google Scholar] [CrossRef]
  70. Deng, Y.; Chen, W.; Li, B.; Wang, C.; Kuang, T.; Li, Y. Physical Vapor Deposition Technology for Coated Cutting Tools: A Review. Ceram. Int. 2020, 46, 18373–18390. [Google Scholar] [CrossRef]
  71. Baptista, A.; Silva, F.; Porteiro, J.; Míguez, J.; Pinto, G. Sputtering Physical Vapour Deposition (PVD) Coatings: A Critical Review on Process Improvement Andmarket Trend Demands. Coatings 2018, 8, 402. [Google Scholar] [CrossRef] [Green Version]
  72. Olivares-Navarrete, R.; Hyzy, S.L.; Berg, M.E.; Schneider, J.M.; Hotchkiss, K.; Schwartz, Z.; Boyan, B.D. Osteoblast Lineage Cells Can Discriminate Microscale Topographic Features on Titanium–Aluminum–Vanadium Surfaces. Ann. Biomed. Eng. 2014, 42, 2551–2561. [Google Scholar] [CrossRef] [Green Version]
  73. Berger, M.B.; Slosar, P.; Schwartz, Z.; Cohen, D.J.; Goodman, S.B.; Anderson, P.A.; Boyan, B.D. A Review of Biomimetic Topographies and Their Role in Promoting Bone Formation and Osseointegration: Implications for Clinical Use. Biomimetics 2022, 7, 7020046. [Google Scholar] [CrossRef]
  74. Anselme, K.; Ploux, L.; Ponche, A. Cell/Material Interfaces: Influence of Surface Chemistry and Surface Topography on Cell Adhesion. J. Adhes. Sci. Technol. 2010, 24, 831–852. [Google Scholar] [CrossRef]
  75. Luo, J.; Walker, M.; Xiao, Y.; Donnelly, H.; Dalby, M.J.; Salmeron-Sanchez, M. The Influence of Nanotopography on Cell Behaviour through Interactions with the Extracellular Matrix—A Review. Bioact. Mater. 2022, 15, 145–159. [Google Scholar] [CrossRef]
  76. Silva-Bermudez, P.; Muhl, S.; Rodil, S.E. A Comparative Study of Fibrinogen Adsorption onto Metal Oxide Thin Films. Appl. Surf. Sci. 2013, 282, 351–362. [Google Scholar] [CrossRef]
  77. Silva-Bermudez, P.; Rodil, S.E. An Overview of Protein Adsorption on Metal Oxide Coatings for Biomedical Implants. Surf. Coat. Technol. 2013, 233, 147–158. [Google Scholar] [CrossRef]
  78. Akterian, S. Evaluating the Vapour Evaporation from the Surface of Pure Organic Solvents and Their Mixtures. Food Sci. Appl. Biotechnol. 2020, 3, 77. [Google Scholar] [CrossRef]
  79. Ansari, S.; Ito, K.; Hofmann, S. Alkaline Phosphatase Activity of Serum Affects Osteogenic Differentiation Cultures. ACS Omega 2022, 7, 12724–12733. [Google Scholar] [CrossRef]
  80. Nobles, K.P.; Janorkar, A.V.; Williamson, R.S. Surface Modifications to Enhance Osseointegration–Resulting Material Properties and Biological Responses. J. Biomed. Mater. Res. Part B Appl. Biomater. 2021, 109, 1909–1923. [Google Scholar] [CrossRef]
  81. Lincks, J.; Boyan, B.D.; Blanchard, C.R.; Lohmann, C.H.; Liu, Y.; Cochran, D.L.; Dean, D.D.; Schwartz, Z. Response of MG63 Osteoblast-like Cells to Titanium and Titanium Alloy Is Dependent on Surface Roughness and Composition. Biomater. Silver Jubil. Compend. 1998, 19, 147–160. [Google Scholar] [CrossRef]
  82. Gittens, R.A.; Olivares-Navarrete, R.; McLachlan, T.; Cai, Y.; Hyzy, S.L.; Schneider, J.M.; Schwartz, Z.; Sandhage, K.H.; Boyan, B.D. Differential Responses of Osteoblast Lineage Cells to Nanotopographically-Modified, Microroughened Titanium-Aluminum-Vanadium Alloy Surfaces. Biomaterials 2012, 33, 8986–8994. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  83. Gittens, R.A.; Olivares-Navarrete, R.; Cheng, A.; Anderson, D.M.; McLachlan, T.; Stephan, I.; Geis-Gerstorfer, J.; Sandhage, K.H.; Fedorov, A.G.; Rupp, F.; et al. The Roles of Titanium Surface Micro/Nanotopography and Wettability on the Differential Response of Human Osteoblast Lineage Cells. Acta Biomater. 2013, 9, 6268–6277. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  84. Zareidoost, A.; Yousefpour, M.; Ghasemi, B.; Amanzadeh, A. The Relationship of Surface Roughness and Cell Response of Chemical Surface Modification of Titanium. J. Mater. Sci. Mater. Med. 2012, 23, 1479–1488. [Google Scholar] [CrossRef]
  85. Silva-Bermudez, P.; Rodil, S.E.; Muhl, S. Albumin Adsorption on Oxide Thin Films Studied by Spectroscopic Ellipsometry. Appl. Surf. Sci. 2011, 258, 1711–1718. [Google Scholar] [CrossRef]
  86. Al-Amin, M.; Abdul-Rani, A.M.; Danish, M.; Rubaiee, S.; Mahfouz, A.B.; Thompson, H.M.; Ali, S.; Unune, D.R.; Sulaiman, M.H. Investigation of Coatings, Corrosion and Wear Characteristics of Machined Biomaterials through Hydroxyapatite Mixed-Edm Process: A Review. Materials 2021, 14, 3597. [Google Scholar] [CrossRef]
  87. Torres-Sanchez, C.; Alabort, E.; Wang, J.; Norrito, M.; Conway, P.P. In-Silico Design and Experimental Validation of TiNbTaZrMoSn to Assess Accuracy of Mechanical and Biocompatibility Predictive Models. J. Mech. Behav. Biomed. Mater. 2021, 124, 104858. [Google Scholar] [CrossRef]
  88. Sul, Y.T.; Johansson, C.B.; Kang, Y.; Jeon, D.G.; Albrektsson, T. Bone Reactions to Oxidized Titanium Implants with Electrochemical Anion Sulphuric Acid and Phosphoric Acid Incorporation. Clin. Implant Dent. Relat. Res. 2002, 4, 78–87. [Google Scholar] [CrossRef]
  89. Shokuhfar, T.; Hamlekhan, A.; Chang, J.Y.; Choi, C.K.; Sukotjo, C.; Friedrich, C. Biophysical Evaluation of Cells on Nanotubular Surfaces: The Effects of Atomic Ordering and Chemistry. Int. J. Nanomed. 2014, 9, 3737–3748. [Google Scholar] [CrossRef] [Green Version]
  90. He, J.; Zhou, W.; Zhou, X.; Zhong, X.; Zhang, X.; Wan, P.; Zhu, B.; Chen, W. The Anatase Phase of Nanotopography Titania Plays an Important Role on Osteoblast Cell Morphology and Proliferation. J. Mater. Sci. Mater. Med. 2008, 19, 3465–3472. [Google Scholar] [CrossRef]
  91. Jarmar, T.; Palmquist, A.; Brånemark, R.; Hermansson, L.; Engqvist, H.; Thomsen, P. Characterization of the Surface Properties of Commercially Available Dental Implants Using Scanning Electron Microscopy, Focused Ion Beam, and High-Resolution Transmission Electron Microscopy. Clin. Implant Dent. Relat. Res. 2008, 10, 11–22. [Google Scholar] [CrossRef]
  92. Arima, Y.; Iwata, H. Effect of Wettability and Surface Functional Groups on Protein Adsorption and Cell Adhesion Using Well-Defined Mixed Self-Assembled Monolayers. Biomaterials 2007, 28, 3074–3082. [Google Scholar] [CrossRef]
  93. Silva-Bermudez, P.; Muhl, S.; Rivera, M.; Rodil, S.E. A look into the interaction of metal oxide thin films with biological media: Albumin and Fibrinogen adsorption. MRS Proc. 2011, 1376, 45–52. [Google Scholar] [CrossRef]
  94. Ochsenbein, A.; Chai, F.; Winter, S.; Traisnel, M.; Breme, J.; Hildebrand, H.F. Osteoblast Responses to Different Oxide Coatings Produced by the Sol-Gel Process on Titanium Substrates. Acta Biomater. 2008, 4, 1506–1517. [Google Scholar] [CrossRef]
  95. Majhy, B.; Priyadarshini, P.; Sen, A.K. Effect of Surface Energy and Roughness on Cell Adhesion and Growth-Facile Surface Modification for Enhanced Cell Culture. RSC Adv. 2021, 11, 15467–15476. [Google Scholar] [CrossRef]
  96. Thiagarajan, L.; Abu-Awwad, H.A.D.M.; Dixon, J.E. Osteogenic Programming of Human Mesenchymal Stem Cells with Highly Efficient Intracellular Delivery of RUNX2. Stem Cells Transl. Med. 2017, 6, 2146–2159. [Google Scholar] [CrossRef]
  97. SIFFERT, R.S. The Role of Alkaline Phosphatase in Osteogenesis. J. Exp. Med. 1951, 93, 415–426. [Google Scholar] [CrossRef] [Green Version]
  98. Giachelli, C.M.; Steitz, S. Osteopontin: A Versatile Regulator of Inflammation and Biomineralization. Matrix Biol. 2000, 19, 615–622. [Google Scholar] [CrossRef]
  99. Si, J.; Wang, C.; Zhang, D.; Wang, B.; Hou, W.; Zhou, Y. Osteopontin in Bone Metabolism and Bone Diseases. Med. Sci. Monit. 2020, 26, 1–9. [Google Scholar] [CrossRef]
  100. Makishi, S.; Yamazaki, T.; Ohshima, H. Osteopontin on the Dental Implant Surface Promotes Direct Osteogenesis in Osseointegration. Int. J. Mol. Sci. 2022, 23, 1039. [Google Scholar] [CrossRef]
  101. Udagawa, N.; Koide, M.; Nakamura, M.; Nakamichi, Y.; Yamashita, T.; Uehara, S.; Kobayashi, Y.; Furuya, Y.; Yasuda, H.; Fukuda, C.; et al. Osteoclast Differentiation by RANKL and OPG Signaling Pathways. J. Bone Miner. Metab. 2021, 39, 19–26. [Google Scholar] [CrossRef]
  102. Komori, T. Functions of Osteocalcin in Bone, Pancreas, Testis, and Muscle. Int. J. Mol. Sci. 2020, 21, 7513. [Google Scholar] [CrossRef]
  103. Quarles, L.D.; Yohay, D.A.; Lever, L.W.; Caton, R.; Wenstrup, R.J. Distinct Proliferative and Differentiated Stages of Murine MC3T3-E1 Cells in Culture: An in Vitro Model of Osteoblast Development. J. Bone Miner. Res. 1992, 7, 683–692. [Google Scholar] [CrossRef]
  104. Zhang, S.; Sun, J.; Xu, Y.; Qian, S.; Wang, B.; Liu, F.; Liu, X. Biological Behavior of Osteoblast-like Cells on Titania and Zirconia Films Deposited by Cathodic Arc Deposition. Biointerphases 2012, 7, 1–10. [Google Scholar] [CrossRef] [Green Version]
  105. Jo, Y.; Kim, Y.T.; Cho, H.; Ji, M.K.; Heo, J.; Lim, H.P. Atomic Layer Deposition of ZrO2 on Titanium Inhibits Bacterial Adhesion and Enhances Osteoblast Viability. Int. J. Nanomed. 2021, 16, 1509–1523. [Google Scholar] [CrossRef]
Figure 1. Representative pictures of the macroscopic appearance of Si (100) wafers coated with metal oxide thin films by magnetron sputtering. Coated samples were named as TiOx, TaOx, NbOx, ZrOx, for titanium, tantalum, niobium and zirconium oxide coatings, while Si samples correspond to uncoated Si (100) wafers.
Figure 1. Representative pictures of the macroscopic appearance of Si (100) wafers coated with metal oxide thin films by magnetron sputtering. Coated samples were named as TiOx, TaOx, NbOx, ZrOx, for titanium, tantalum, niobium and zirconium oxide coatings, while Si samples correspond to uncoated Si (100) wafers.
Materials 15 05240 g001
Figure 2. Representative SEM images of the areas analyzed (pink rectangles marked on the micrographs) for EDS elemental composition studies and of the corresponding EDS spectra obtained for the (a) TiOx; (b) TaOx; (c) NbOx; and (d) ZrOx coated samples.
Figure 2. Representative SEM images of the areas analyzed (pink rectangles marked on the micrographs) for EDS elemental composition studies and of the corresponding EDS spectra obtained for the (a) TiOx; (b) TaOx; (c) NbOx; and (d) ZrOx coated samples.
Materials 15 05240 g002
Figure 3. Example low resolution XPS spectra of TiOX (TiO2), ZrOX (ZrO2), NbOX (Nb2O5) and TaOX (Ta2O5).
Figure 3. Example low resolution XPS spectra of TiOX (TiO2), ZrOX (ZrO2), NbOX (Nb2O5) and TaOX (Ta2O5).
Materials 15 05240 g003
Figure 4. Representative grazing incidence XRD patterns of (a) TiOx, (b) ZrOx, (c) NbOx, and (d) TaOx coatings. In Figure 3a, the A and R stand for the expected XRD pattern for Anatase and Rutile crystalline phases of TiO2, respectively. In Figure 3b, the B stands for the expected XRD pattern for the Baddeleyite crystalline phase of ZrO2.
Figure 4. Representative grazing incidence XRD patterns of (a) TiOx, (b) ZrOx, (c) NbOx, and (d) TaOx coatings. In Figure 3a, the A and R stand for the expected XRD pattern for Anatase and Rutile crystalline phases of TiO2, respectively. In Figure 3b, the B stands for the expected XRD pattern for the Baddeleyite crystalline phase of ZrO2.
Materials 15 05240 g004
Figure 5. Representative micrographs, obtained by an optical profilometer, of BM-MSC cell cultured on uncoated Si (ac) and coated, TiOx (df), TaOx (gi), NbOx (jl), and ZrOx (mo), samples. Three incubation times were handled, 3, 7, and 14 days. Micrographs were acquired at two different magnifications, 5× for main micrographs and 50× for insert magnifications.
Figure 5. Representative micrographs, obtained by an optical profilometer, of BM-MSC cell cultured on uncoated Si (ac) and coated, TiOx (df), TaOx (gi), NbOx (jl), and ZrOx (mo), samples. Three incubation times were handled, 3, 7, and 14 days. Micrographs were acquired at two different magnifications, 5× for main micrographs and 50× for insert magnifications.
Materials 15 05240 g005
Figure 6. Qualitative evaluation of BM-MSC viability at 3, 7, and 14 days of culture on uncoated Si (df) and TiOx (gi), TaOx (jl), NbOx (mo), and ZrOx (pr) coated samples. Ctrl (ac) corresponds to BM-MSC culture on the surface of standard tissue culture plates. Micrographs were acquired at two different magnifications, 5× and 20×. The fluorescent LIVE/DEADTM Viability/Cytotoxicity Kit (Invitrogen®, Waltham, MA, USA) for mammalian cells was used to mark viable (green) and dead (red nuclei) cells.
Figure 6. Qualitative evaluation of BM-MSC viability at 3, 7, and 14 days of culture on uncoated Si (df) and TiOx (gi), TaOx (jl), NbOx (mo), and ZrOx (pr) coated samples. Ctrl (ac) corresponds to BM-MSC culture on the surface of standard tissue culture plates. Micrographs were acquired at two different magnifications, 5× and 20×. The fluorescent LIVE/DEADTM Viability/Cytotoxicity Kit (Invitrogen®, Waltham, MA, USA) for mammalian cells was used to mark viable (green) and dead (red nuclei) cells.
Materials 15 05240 g006
Figure 7. Quantitative evaluation of BM-MSC viability cultured on the coatings at (a) 3 days, (b) 7days, (c) 14 days of culture. Thermo Fisher Alamar BlueTM cell viability reagent kit was used. Ctrl corresponds to BM-MSC cultured on the surface of standard tissue culture plates. Data are presented as the mean ± standard error using three independent cultures per variable and analyzed by one-way ANOVA using a Dunnett’s post hoc test. * p ≤ 0.05 vs. Ctrl; ** p ≤ 0.001 vs. Ctrl; **** p ≤ 0.0001; & p ≤ 0.05 vs. TiOx; && p ≤ 0.01 vs. TiOx.
Figure 7. Quantitative evaluation of BM-MSC viability cultured on the coatings at (a) 3 days, (b) 7days, (c) 14 days of culture. Thermo Fisher Alamar BlueTM cell viability reagent kit was used. Ctrl corresponds to BM-MSC cultured on the surface of standard tissue culture plates. Data are presented as the mean ± standard error using three independent cultures per variable and analyzed by one-way ANOVA using a Dunnett’s post hoc test. * p ≤ 0.05 vs. Ctrl; ** p ≤ 0.001 vs. Ctrl; **** p ≤ 0.0001; & p ≤ 0.05 vs. TiOx; && p ≤ 0.01 vs. TiOx.
Materials 15 05240 g007
Figure 8. Evaluation of cell proliferation for BM-MSC cultured on the coatings studied. (a) 3 days and (b) 5 days of culture. The Roche Cell Proliferation ELISA, BrdU (colorimetric) kit was used. Ctrl corresponds to BM-MSC cultured on standard TCP. Data are presented as the mean ± standard error using three independent cultures per variable and analyzed by one-way ANOVA using a Dunnett’s post hoc test. * p ≤ 0.05 vs. Ctrl; *** p ≤ 0.005 and **** p ≤ 0.0005; &&& p ≤ 0.001 vs. TiOx; &&&& p ≤ 0.0001 vs. TiOx; $$ p ≤ 0.01 vs. TaOx; $$$$ p ≤ 0.001 vs. TaOx; % p ≤ 0.05 vs. NbOx.
Figure 8. Evaluation of cell proliferation for BM-MSC cultured on the coatings studied. (a) 3 days and (b) 5 days of culture. The Roche Cell Proliferation ELISA, BrdU (colorimetric) kit was used. Ctrl corresponds to BM-MSC cultured on standard TCP. Data are presented as the mean ± standard error using three independent cultures per variable and analyzed by one-way ANOVA using a Dunnett’s post hoc test. * p ≤ 0.05 vs. Ctrl; *** p ≤ 0.005 and **** p ≤ 0.0005; &&& p ≤ 0.001 vs. TiOx; &&&& p ≤ 0.0001 vs. TiOx; $$ p ≤ 0.01 vs. TaOx; $$$$ p ≤ 0.001 vs. TaOx; % p ≤ 0.05 vs. NbOx.
Materials 15 05240 g008
Figure 9. RUNX2 (ae), OP (fj) and OC (ko) cellular expression characterized by IF assays in BM-MSC cultured on the coatings. In green, positive expression of the corresponding osteogenic marker is observed, while in blue, cell nuclei stained with Hoechst are observed. Ctrl corresponds to cells cultured on standard tissue culture plates.
Figure 9. RUNX2 (ae), OP (fj) and OC (ko) cellular expression characterized by IF assays in BM-MSC cultured on the coatings. In green, positive expression of the corresponding osteogenic marker is observed, while in blue, cell nuclei stained with Hoechst are observed. Ctrl corresponds to cells cultured on standard tissue culture plates.
Materials 15 05240 g009
Figure 10. Quantification of proteins involved in osteogenic differentiation. Quantification of ALP activity, at (a) 7 days, and (b) 14 days of cells culture. Quantification of (c) osteopontin (OP) expression, (d) osteoprotegerin (OPG), and (e) osteocalcin (OC) expression after 14 days of cells culture. Ctrl- and Ctrl correspond to BM-MSC cultured on TCP surfaces with supplemented Mesenchymal Stem Basal Medium (Ctrl−), or DMEM/F-12 supplemented with 10% v/v FBS (Ctrl). Data are presented as the mean ± standard error using three independent cultures per variable and analyzed by one-way ANOVA using a Dunnett’s post hoc test. * p ≤ 0.05 vs. Ctrl; ** p ≤ 0.01 vs. Ctrl; *** p ≤ 0.001; and **** p ≤ 0.0001 vs. Ctrl vs Ctrl.
Figure 10. Quantification of proteins involved in osteogenic differentiation. Quantification of ALP activity, at (a) 7 days, and (b) 14 days of cells culture. Quantification of (c) osteopontin (OP) expression, (d) osteoprotegerin (OPG), and (e) osteocalcin (OC) expression after 14 days of cells culture. Ctrl- and Ctrl correspond to BM-MSC cultured on TCP surfaces with supplemented Mesenchymal Stem Basal Medium (Ctrl−), or DMEM/F-12 supplemented with 10% v/v FBS (Ctrl). Data are presented as the mean ± standard error using three independent cultures per variable and analyzed by one-way ANOVA using a Dunnett’s post hoc test. * p ≤ 0.05 vs. Ctrl; ** p ≤ 0.01 vs. Ctrl; *** p ≤ 0.001; and **** p ≤ 0.0001 vs. Ctrl vs Ctrl.
Materials 15 05240 g010
Table 1. Coatings elemental composition as obtained from EDS analysis.
Table 1. Coatings elemental composition as obtained from EDS analysis.
CoatingElementWeight %Atomic %
TiOxO57.8 ± 0.580.4 ± 0.3
Ti42.2 ± 0.519.6 ± 0.3
TaOxO17.5 ± 0.370.5 ± 0.5
Ta82.6 ± 0.329.5 ± 0.5
NbOxO31.7 ± 0.272.9 ± 0.2
Nb68.3 ± 0.227.1 ± 0.2
ZrOxO26.2 ± 0.266.9 ± 0.2
Zr73.8 ± 0.233.1 ± 0.2
Table 2. Roughness, representative images of the surface topography, and thickness of the metal oxide coatings.
Table 2. Roughness, representative images of the surface topography, and thickness of the metal oxide coatings.
Coating
Property
TiOxTaOxNbOxZrOx
Sa * (nm)0.5 ± 0.1
Materials 15 05240 i001
0.4 ± 0.2
Materials 15 05240 i002
0.8 ± 0.1
Materials 15 05240 i003
0.3 ± 0.1
Materials 15 05240 i004
Thickness (nm)54.5 ± 7.3346.5 ± 31.5175.9 ± 15227.7 ± 18.5
* Sa, arithmetical mean of the roughness.
Table 3. Droplet size stability over time for water contact angle measurements.
Table 3. Droplet size stability over time for water contact angle measurements.
SampleInitial Measurement
(0 min)
1 min2 min3 min4 min5 min
Si Materials 15 05240 i005 Materials 15 05240 i006 Materials 15 05240 i007 Materials 15 05240 i008 Materials 15 05240 i009 Materials 15 05240 i010
TiOx Materials 15 05240 i011 Materials 15 05240 i012 Materials 15 05240 i013 Materials 15 05240 i014 Materials 15 05240 i015 Materials 15 05240 i016
TaOx Materials 15 05240 i017 Materials 15 05240 i018 Materials 15 05240 i019 Materials 15 05240 i020 Materials 15 05240 i021 Materials 15 05240 i022
NbOx Materials 15 05240 i023 Materials 15 05240 i024 Materials 15 05240 i025 Materials 15 05240 i026 Materials 15 05240 i027 Materials 15 05240 i028
ZrOx Materials 15 05240 i029 Materials 15 05240 i030 Materials 15 05240 i031 Materials 15 05240 i032 Materials 15 05240 i033 Materials 15 05240 i034
Table 4. Water Contact Angle (WCA) and Surface Free Energy (SFE) of uncoated and coated silicon surfaces with metal oxides thin films.
Table 4. Water Contact Angle (WCA) and Surface Free Energy (SFE) of uncoated and coated silicon surfaces with metal oxides thin films.
SamplesWCA (°)Surface Free Energy (mN/m)γp/(γpd)
(%)
γd/pγdγp
Si79.8 ± 0.9
Materials 15 05240 i035
30.0 ± 0.121.2 ± 0.48.8 ± 0.529.3
TiOx87.6 ± 0.5
Materials 15 05240 i036
26.2 ± 0.121.2 ± 0.35.1 ± 0.219.4
TaOx88.3 ± 0.9
Materials 15 05240 i037
22.6 ± 0.214.9 ± 0.37.7 ± 0.534.0
NbOx86.3 ± 1.2
Materials 15 05240 i038
26.1 ± 0.119.5 ± 0.46.7 ± 0.525.7
ZrOx89.2 ± 0.9
Materials 15 05240 i039
23.7 ± 0.118.0 ± 0.45.7 ± 0.424.0
γd/p, total SFE; γd, SFE dispersive component; γp, SFE polar component; γp/(γp + γd), polarity factor.
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Fernández-Lizárraga, M.; García-López, J.; Rodil, S.E.; Ribas-Aparicio, R.M.; Silva-Bermudez, P. Evaluation of the Biocompatibility and Osteogenic Properties of Metal Oxide Coatings Applied by Magnetron Sputtering as Potential Biofunctional Surface Modifications for Orthopedic Implants. Materials 2022, 15, 5240. https://doi.org/10.3390/ma15155240

AMA Style

Fernández-Lizárraga M, García-López J, Rodil SE, Ribas-Aparicio RM, Silva-Bermudez P. Evaluation of the Biocompatibility and Osteogenic Properties of Metal Oxide Coatings Applied by Magnetron Sputtering as Potential Biofunctional Surface Modifications for Orthopedic Implants. Materials. 2022; 15(15):5240. https://doi.org/10.3390/ma15155240

Chicago/Turabian Style

Fernández-Lizárraga, Mariana, Julieta García-López, Sandra E. Rodil, Rosa María Ribas-Aparicio, and Phaedra Silva-Bermudez. 2022. "Evaluation of the Biocompatibility and Osteogenic Properties of Metal Oxide Coatings Applied by Magnetron Sputtering as Potential Biofunctional Surface Modifications for Orthopedic Implants" Materials 15, no. 15: 5240. https://doi.org/10.3390/ma15155240

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop