Estimation of Tibiofemoral Joint Contact Forces Using Foot Loads during Continuous Passive Motions

1
State Key Laboratory of Robotics and System, Harbin Institute of Technology, Harbin 150001, China
2
Orthopaedical Rehabilitation Centre, Weihai Municipal Hospital, Weihai 264209, China
3
Department of Mechanical Engineering, School of Naval Architecture and Ocean Engineering, Harbin Institute of Technology (Weihai), Weihai 264209, China
4
Department of Neurology, The First Affiliated Hospital of Dalian Medical University, Dalian 116011, China
5
Tianzhi Institute of Innovation and Technology, Weihai 264209, China
*
Author to whom correspondence should be addressed.
Sensors 2022, 22(13), 4947; https://doi.org/10.3390/s22134947
Submission received: 15 May 2022 / Revised: 21 June 2022 / Accepted: 27 June 2022 / Published: 30 June 2022

Abstract

:
Continuous passive motion (CPM) machines are commonly used after various knee surgeries, but information on tibiofemoral forces (TFFs) during CPM cycles is limited. This study aimed to explore the changing trend of TFFs during CPM cycles under various ranges of motion (ROM) and body weights (BW) by establishing a two-dimensional mathematical model. TFFs were estimated by using joint angles, foot load, and leg–foot weight. Eleven healthy male participants were tested with ROM ranging from 0° to 120°. The values of the peak TFFs during knee flexion were higher than those during knee extension, varying nonlinearly with ROM. BW had a significant main effect on the peak TFFs and tibiofemoral shear forces, while ROM had a limited effect on the peak TFFs. No significant interaction effects were observed between BW and ROM for each peak TFF, whereas a strong linear correlation existed between the peak tibiofemoral compressive forces (TFCFs) and the peak resultant TFFs (R2 = 0.971, p < 0.01). The proposed method showed promise in serving as an input for optimizing rehabilitation devices.

Graphical Abstract

1. Introduction

The continuous passive motion (CPM) device for the lower limb is a motorized device that passively moves the knee joint through a preset range of motion. It is widely used in the rehabilitation process of knee arthroplasty (KA), early rehabilitation of stroke, and other knee surgeries [1,2,3]. The CPM devices are usually used to eliminate swelling, heal joint soft-tissue, restore joint flexibility and function [4], and reduce the workload of physiotherapists [5]. However, its effectiveness and discomfort have been controversial [6,7]; for example, patients experience pain during rehabilitation due to stretching surgical wounds [8].
A conventional CPM device with a linear electrical drive affects the foot [9] and causes impact loading [10] because of the lack of a bionic design. In recent years, pneumatic actuators [11,12] and optimization of the motors’ location [13,14,15] have been applied to improve the comfort and safety of CPM devices to some extent. However, no conclusive consensus has been reached on reducing impact force because the mechanism for these improvements is still not well understood, and systematic studies of tibiofemoral forces (TFFs) during CPM cycles are lacking. The postoperative rehabilitation of fractures and KA are sensitive to knee joint stress [16]; thus, it is necessary to quantify and decrease this stress during early rehabilitation [17,18]. In addition, the estimated TFFs can be used to plan rehabilitation treatments and monitor abnormal forces during the rehabilitation process. Moreover, TFFs can also be used to optimize the design of CPM devices. However, neither the changing trend of the foot pressure nor the effects of ranges of motion (ROM) on TFFs during CPM cycles have been investigated.
Although in vivo and in vitro studies have estimated TFFs, most of these systems are non-portable, non-real-time assessment, and laboratory-based. An implantable telemetry electronic prosthesis was designed to directly measure TFFs during the activities of daily living [19,20]. Moreover, musculoskeletal models [21], finite element models [22], and wearable inertial devices [23,24] have been applied to estimate TFFs. However, without complex experimental settings, it is difficult to use an electromyogram-driven musculoskeletal model, smart implant, or finite element model to analyze the influence of external forces on TFFs.
Pressure-sensitive insoles have been widely used in both gait assessment [25,26] and the measurement of plantar pressure during rehabilitation [27,28,29]. A general two-dimensional knee model [30,31] and pressure-sensitive insoles were employed in this study to estimate TFFs. The aims of this study were as follows: (1) establishing a general two-dimensional mathematical knee model, (2) estimating TFFs using the established model and the directly measured foot pressure, and (3) exploring the influence of ROM and body weight (BW) on the peak resultant TFFs and peak tibiofemoral shear forces (TFSFs).

2. Materials and Methods

2.1. Participants

The data were collected from eleven healthy male participants (Table 1) with no disorder of the lower extremities or any symptoms of muscle soreness, where L indicated the distance between the plantar and the knee axis. All participants were informed about the test procedures, and they provided informed written consent prior to the study. The study was conducted in accordance with the Declaration of Helsinki, and the protocol was approved by the local Ethics Committee (Ethics Committee of Weihai Municipal Hospital; Ref. No. 2020055). The participants were divided into three groups according to their BWs.

2.2. Experimental Data Collection

Because of the limitation of the CPM device, the ROM of the lower extremities ranged from 60° to 120° at intervals of 20°. Each trial was repeated four times and included more than three cycles. The participants were required to relax their bodies and refrain from voluntary movement during CPM cycles, and only the left leg was tested. All the experiments were controlled by a physical therapist (not a member of the authorship), and the foot loads and joint angles were measured simultaneously.
The CPM device (The KH-11, Model JK-A lower-limb CPM; CANWELL Technologies, Jinhua, China) contained a handheld digital display unit. The display angle of the CPM device was verified with an inertial sensor. To avoid fatigue, each experiment was limited to 5 min, and the interval was more than 5 min. A motion-capture system with eight cameras (Mars2H, Nokov, Beijing, China, sampling rate 50 Hz) was used to estimate the knee joint angle (ψk) and the hip joint angle (Figure 1). The CPM device moved at a relatively low speed of 6°/s to decrease the influence of the inertia of the CPM on the foot load.
The pressure between the foot and the support plate was measured using a plantar pressure system (T&T Medilogic Medizintechnik GmbH, Schönefeld, Germany, 50 Hz) [32]. The plantar pressure insole started to record the pressure data before starting the CPM machine. The insole was selected according to the foot length (sizes of 39–44 were available) and calibrated using participants’ weights [25,29,33] before the experiment. Participants’ weights were used to calibrate the digital output of sensors while the participant was standing still. Each participant stood still for more than 60 s, and the digital output was subtracted from the time series. Each trial was repeated three times. The calibration factor (qx) can be estimated using Equation (1):
q x = [ i = 1 n i ( l i , j 64   255 ) 1 . 125 ]   / W x g
where Wx is the mass of participant x, g is the acceleration of gravity, and li,j denotes the 0–255 digital value output of sensor i at instant j. The variability of the active sensors summed for participant A4 is shown in Figure 2a, which shows that the variability is small.

2.3. Biomechanical Model

Four assumptions were made to consider the anatomical structure and physiological characteristics of the knee joint to establish a general lower limb model.
(1)
The lower limb was regarded as a rigid body model in the sagittal plane, the lower leg and foot (foot–leg system) were regarded as one body, and the average angle (θ) between the tibial plane and its long axis was 76.8° [30].
(2)
The center of pressure (CoP) distribution of insole measurements in the insole coordinate system was around the tibial axis, with the ankle angle kept at 0° (Figure 3). The moments induced by the position error of CoP could be presented by Me in Nm.
(3)
The center point, O, of the intercondylar contact coincided with the origin of moments and force, guaranteeing that the moment generated by the pressure and shear stresses was equivalent to zero [31,34].
(4)
The knee joint had no active force and was affected only by the passive torque, Mk, generated by the tissue [35]. It moved in the sagittal plane with only one degree of freedom, and the knee axis was unicentric (Figure 4).
Considering the low speed of the rehabilitation movement, the lower limb was regarded as quasi-static, and point O was selected as the origin of the torque calculation. The following equations provide the resultant force and moment around O:
{ F n d n + M k + M e = G d G     ( knee   flexion ) F n d n M k + M e = G d G     ( knee   extention )
where dn denotes the distance from the center of mass of the leg–foot system to the knee axis; for more details, please refer to [36]. G is the gravity of the foot–leg system, which was estimated using kinematic and anthropometric data [37]. dG could be calculated using Equation (3):
d G = d n cos β
The passive torque Mk (Nm) of the knee could be calculated using:
M k = e 1 . 8 0 . 0352 ψ k + 0 . 0217 ψ h e 3 . 971 + 0 . 0495 ψ k 0 . 0128 ψ h + e 2 . 22 0 . 15 ψ k 4 . 82  
where ψk and ψh are knee and hip joint angles, respectively, measured by the motion-capture system [35]. Angles are presented in degrees. F1 in N was the resultant measured plantar force at an instant and can be calculated using Equation (5):
F 1 = ( i = 1 n i ( l i , j 64   N / cm 2 255 ) 1 . 125 cm 2 ) / q x
where li,j denotes the 0–255 digital value output of sensor i at instant j [32], and qx is the calibration factor.
Me can be calculated using Equation (6):
M e = F 1 y
where y denotes the CoP location shift in the sagittal plane. The tibiofemoral compressive force Fc and the tibiofemoral shear force Fs could be calculated using Equation (7).
{ F s = F 1 cos θ + F n sin θ G cos α F c = F 1 sin θ + F n cos θ + G sin α    ( α = { π 2 θ + β , β > 0 θ π 2 + β , β < 0 )
All equations were combined and numerically solved using the computational software Maple (Maple 2020).

2.4. Data Analysis

The duration of each CPM cycle was normalized to 100% for comparison among conditions and across participants. All statistical analyses were performed with SPSS 24.0 (version IBM SPSS 24; SPSS Inc., Armonk, NY, USA). A one-way analysis of variance (ANOVA) was used to test the significance of the differences in forces among the various ROM, and Bonferroni post hoc tests were performed if the ANOVA result was significant (p < 0.05). The effects of BW and ROM on the peak resultant TFFs and peak TFSFs were analyzed by two-way ANOVA. Significant interactions and main effects were post hoc analyzed using Bonferroni correction.

3. Results

3.1. Typical Peak Resultant TFFs

Representative CPM cycles from three participants (that is, A1, B1, and C1) demonstrated the existence of two major peak forces during each complete cycle (Figure 5), with forces represented as the multiples of BW. Although the peak resultant TFFs of these participants were different, the changing trend of the estimated TFFs was similar. Figure 5 depicts the major peak knee contact forces during a CPM cycle ranging from 0.24 × BW during cycles of 60° to 0.43 × BW at a ROM of 120°. The TFFs during a cycle were asymmetrical, and the values of the peak resultant TFFs during knee flexion were higher than those during knee extension. In all cases, the first peak force appeared when the knee flexion approximated 20 degrees.
The intragroup statistical analysis of the peak resultant TFFs of the three groups was conducted to evaluate the effect of ROM alone (Figure 6). Significant differences were observed between the ROM of 60° and that of 80° (p < 0.05) for all groups. For each group, one-way ANOVA intragroup comparison indicated that the lowest mean value of the peak resultant TFFs was recorded during the ROM of 60°, varying from 0.16 (BW) to 0.31 (BW). Moreover, the results showed that the highest mean peak resultant TFFs was observed during the ROM of 80° rather than during the ROM of 120°. When the ROM was larger than 80°, the mean peak resultant TFFs experienced a slight decrease.

3.2. TFCFs and TFSFs

Figure 7 presents the mean estimated values of the TFCFs and TFSFs in three participants (that is, A1, B1, and C1); significant differences were observed between the peaks of the TFCFs and the TFSFs (p < 0.05). The average value of the peak compressive forces was approximately two or three times the magnitude of the peak shear forces. Compared with the TFSFs, the TFCFs changed significantly. Similar to the trend of the resultant TFFs, the peaks of the TFSFs during knee flexion were typically higher than those during knee extension.

3.3. Interactions of BW and ROM on the Peak Resultant TFFs and TFSFs

The two-way ANOVA indicated no significant interaction effects between BW and ROM for the peak resultant TFFs. Generally, the mean value of the peak resultant TFFs during the CPM motion arc of 60° was the lowest (Table 2) and was significantly different from that during the CPM motion arc of 80° (p < 0.05). Moreover, the BW had significant main effects on the major peak of the resultant TFFs (p < 0.01). The peak resultant TFFs increased with the BW, which caused a significant difference between groups A and B/C (p < 0.01).
The peak resultant TFFs were in a significant linear relationship with the peak TFCFs (R2 = 0.971, p < 0.01), which was obtained using linear regression analysis, indicating the approximate statistical properties of the TFFs and TFCFs. Additionally, no significant interactions were detected between BW and ROM for the peak TFSFs (two-way measure ANOVA), whereas the BW had a significant main effect on the peak TFSFs when comparing group A with B/C (p < 0.05). Compared with the mean value of the peak resultant TFFs, the mean peak of TFSFs was almost one-third to one-half of the former (Table 2). Details of the data for plantar pressure and video of data acquisition experiments are reported in the Supplementary Material.

4. Discussion

This exploratory study examined the effects of BW and ROM on the changing trend of peak resultant TFFs and peak TFSFs during CPM cycles. The ROM had a limited effect on the peak resultant TFFs, while the BW had a significant main effect on the peak resultant TFFs and peak TFSFs. No significant interactions between BW and ROM were witnessed, but a strong linear correlation between the peak resultant TFFs and peak TFCFs was observed.
The results in Figure 6 indicate that TFFs did not continually increase with ROM; the biggest mean peak resultant TFFs occurred during the ROM of 80° rather than the ROM of 120°. The reason was not fully understood, but the mechanisms of the therapeutic action of CPM and physiologic explanations might account for it. Few studies have systematically investigated the effect of ROM on TFFs during CPM cycles. We cannot directly compare our results with the findings of other studies carried out without rehabilitation devices, such as walking and running. However, previous studies seem consistent with our findings to some extent. Bergmann et al. [38] showed that the peak resultant force occurred with a flexion angle of 94° rather than a bigger knee-flexion angle. In addition, the peak resultant TFFs during the flexion phase were smaller than those during the extension phase, indicating that the resultant TFFs in a CPM cycle were asymmetric, which was in agreement with a prior study of the intra-articular pressure of the human knee during CPM [39]. It is worth noting that the observed trend may be affected by the motion trajectory of rehabilitation devices; thus, our method can serve to plan the trajectory of rehabilitation robots.
Unlike the peak TFCFs, the peak TFSFs were not affected by ROM. The peak TFSFs were substantially lower than the peak TFCFs; a similar pattern of results was observed during walking [40] and isokinetic knee extension [31]. However, the peak TFSFs in the present study had relatively low levels compared with previously reported values (5% vs. 25%). This could be explained since the knee extensor muscles and patellar tendon of participants in our study were not activated during the knee joint passive motion. Thus, although the location of the tibiofemoral contact point (C) and the knee axis were not exactly coincident as hypothesized, the error induced by this bias would be lower than 5%.
The BW was the main factor affecting the peak resultant TFFs, and the peak resultant TFFs increased with the BW when comparing group A with groups B and C. The TFFs increased with the BW of leg–foot systems. This changing trend was consistent with the results obtained in previous studies [19,41] in which the knee joint contact forces depended on BW. However, the predicted peak TFFs (Table 2) in this study were considerably lower than the reported results by Damm et al. [19] (TFFs ≥ 2.5 BW). Compared with the study [19], the load only exerted by the leg may account for the difference in TFFs. Obesity increased the prevalence of diagnosed knee abnormalities [42] and might impair the early outcome of KA [43,44]. Thus, weight management before KA might decrease the pain during CPM after KA, which coincided with the results reported in previous studies [45,46]. In addition, under various BWs, the changing trends of the estimated forces versus time in CPM cycles were similar, and this was similar to that observed during walking [47,48].
The directly measured plantar pressure changed with time, as shown in Figure 8, indicating that the soft tissues might deform along the tibial axis; otherwise, the curve should change symmetrically and smoothly during each CPM cycle. The CPM equipment could not align with complex knee movements [49,50] and stretch surgical wounds [8]. Hence, patients suffered from pain during CPM cycles [51]. Therefore, reducing the deformations of soft tissues by improving kinematic compatibilities might decrease pain during CPM cycles. The estimated TFFs can be utilized as optimization metrics to minimize the risk of a knee injury. In conclusion, reducing the effect of CPM on surgical wounds during knee rehabilitation is one key aspect of our future investigations. Additionally, exploring the skin deformation of patients and detecting the relationship between skin deformation and pain during CPM cycles will be the focus of future studies.
Overall, the proposed mathematical method may decrease the performance of some movements but provides the advantage that not every movement must be modeled. This generic model is easier to apply in clinical and home-based rehabilitation owing to the ease of measurement of BW and height compared with the musculoskeletal model [47] and the visual motion capture system [29,52]. Measured foot pressure has been used to evaluate abnormal walking behavior [53,54] and the CoP of active motions [28]. Furthermore, the estimated TFFs using measured foot pressure can act as a biofeedback method to monitor abnormal force during rehabilitation exercises to prevent injuries, implying that the proposed method could be applied in both active and passive motions.
This study has a few limitations. First, most CPM machines are similar in mechanical design. However, only one brand of CPM machine was tested. Second, considering that the CPM has a wide range of users, we recruited only healthy male individuals to better explore the changing trend of TFFs. To compensate for this difference between healthy male individuals and KA population [55], this method can also be applied to patients after KA by modifying the values of θ and Mk. In the future, both healthy individuals and patients undergoing knee surgeries in different stages will be recruited to verify the proposed method. Third, the numbers cited for the mass, location of the center of mass, and passive elastic joint moments were estimated statistically from sampled individuals. Thus, they were regarded as estimates.

5. Conclusions

This study described a novel integrated approach with a general two-dimensional knee model and pressure-sensitive insoles to explore the changing trend of TFFs versus ROM and BW during CPM cycles. The ROM had a limited effect on the peak resultant TFFs, whereas TFFs did not increase linearly with ROM. The BW had a significant positive effect on both TFSFs and TFFs. No significant interactions between BW and ROM were observed. The proposed method may serve as an input for optimizing rehabilitation devices. Moreover, this method showed promise for clinical and long-term home-based rehabilitation owing to its portable and configurable properties.

Supplementary Materials

The following supporting information can be downloaded at: https://figshare.com/articles/dataset/Supplementary_Materials/19766644/2 (accessed on 15 May 2022). Table S1. Data for Figure 4 and Figure 6. (XLSX); Table S2. Data for plantar pressure of A1, B1, and C1. (CSV); Video S1. Video of data acquisition experiments (MP4).

Author Contributions

Conceptualization, Y.Y. (Yunlong Yang); data curation, Y.Y. (Yunlong Yang), H.H., J.G., F.Y. and Y.Y. (Yufeng Yao); formal analysis, J.G.; funding acquisition, Y.Y. (Yufeng Yao); methodology, Y.Y. (Yunlong Yang), H.H. and F.Y.; project administration, Y.Y. (Yufeng Yao); software, Y.Y. (Yunlong Yang); supervision, Y.Y. (Yufeng Yao); writing–original draft, Y.Y. (Yunlong Yang); writing–review and editing, Y.Y. (Yunlong Yang), H.H., J.G., F.Y. and Y.Y. (Yufeng Yao). All authors have read and agreed to the published version of the manuscript.

Funding

This work is supported by the National Key R&D Program of China (Grant No. 2018YFB1309400).

Institutional Review Board Statement

The study was conducted according to the guidelines of the Declaration of Helsinki, and the protocol was approved by the local ethics committee (Ref. No. 2020055; Weihai Municipal Hospital).

Informed Consent Statement

Informed consent was obtained from all subjects involved in the study.

Acknowledgments

We would like to thank the staff of the rehabilitation center for their cooperation during the implementation of this study.

Conflicts of Interest

The authors declare no conflict of interest.

Abbreviations

CPMcontinuous passive motion
KAknee arthroplasty
TFFstibiofemoral forces
TFCFstibiofemoral compressive forces
TFSFstibiofemoral shear forces
BWbody weight
CoPcenter of pressure
Mkpassive torque generated by the tissue
ANOVAanalysis of variance

References

  1. Karnes, J.M.; Harris, J.D.; Griesser, M.J.; Flanigan, D.C. Continuous passive motion following cartilage surgery: Does a common protocol exist? Physician Sportsmed. 2013, 41, 53–63. [Google Scholar] [CrossRef] [PubMed]
  2. Plessis, M.; Eksteen, E.; Jenneker, A.; Kriel, E.; Mentoor, C.; Stucky, T.; Staden, D.V.; Morris, L. The effectiveness of continuous passive motion on range of motion, pain and muscle strength following rotator cuff repair: A systematic review. Clin. Rehabil. 2011, 25, 291–302. [Google Scholar] [CrossRef] [PubMed]
  3. Howard, J.S.; Mattacola, C.G.; Romine, S.E.; Lattermann, C. Continuous passive motion, early weight bearing, and active motion following knee articular cartilage repair: Evidence for clinical practice. Cartilage 2010, 1, 276–286. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  4. McCarthy, M.R.; O’Donoghue, P.C.; Yates, C.K.; Yates-McCarthy, J.L. The clinical use of continuous passive motion in physical therapy. J. Orthop. Sports Phys. Ther. 1992, 15, 132–140. [Google Scholar] [CrossRef] [PubMed]
  5. Brattig, B.; Schablon, A.; Nienhaus, A.; Peters, C. Occupational accident and disease claims, work-related stress and job satisfaction of physiotherapists. J. Occup. Med. Toxicol. 2014, 9, 36. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  6. Yang, X.; Li, G.; Wang, H.; Wang, C. Continuous passive motion after total knee arthroplasty: A systematic review and meta-analysis of associated effects on clinical outcomes. Arch. Phys. Med. Rehabil. 2019, 100, 1763–1778. [Google Scholar] [CrossRef]
  7. Bible, J.E.; Simpson, A.K.; Biswas, D.; Pelker, R.R.; Grauer, J.N. Actual knee motion during continuous passive motion protocols is less than expected. Clin. Orthop. Relat. Res. 2009, 467, 2656–2661. [Google Scholar] [CrossRef] [Green Version]
  8. Yu, R.; Zhuo, Y.; Feng, E.; Wang, W.; Lin, W.; Lin, F.; Li, Z.; Lin, L.; Xiao, L.; Wang, H.; et al. The effect of musical interventions in improving short-term pain outcomes following total knee replacement: A meta-analysis and systematic review. J. Orthop. Surg. Res. 2020, 15, 1–16. [Google Scholar] [CrossRef]
  9. Ivlev, O.; Baiden, D.; Wilkening, A.; Koch, C.; Haas, H.-D. Compact assistive rehabilitation devices–concept and preliminary function test. In Proceedings of the World Congress on Medical Physics and Biomedical Engineering, Munich, Singapore, 12–17 June 2022; pp. 88–91. [Google Scholar]
  10. Näf, M.B.; Junius, K.; Rossini, M.; Rodriguez-Guerrero, C.; Vanderborght, B.; Lefeber, D. Misalignment compensation for full human-exoskeleton kinematic compatibility: State of the art and evaluation. Appl. Mech. Rev. 2018, 70, 050802. [Google Scholar] [CrossRef] [Green Version]
  11. Ioana, P.; Andrea, D.; Liliana, R.; Deaconescu, T.I. Orthopaedic rehabilitation device actuated with pneumatic muscles. Int. J. Adv. Robot. Syst. 2014, 11, 105. [Google Scholar]
  12. Rupp, R.; Plewa, H.; Hofer, E.P.; Knestel, M. MotionTherapy@Home—A robotic device for automated locomotion therapy at home. In Proceedings of the 2009 IEEE International Conference on Rehabilitation Robotics, Kyoto, Japan, 23–26 June 2009. [Google Scholar]
  13. Trochimczuk, R.; Kuźmierowski, T.; Anchimiuk, P. Study of CPM Device used for Rehabilitation and Effective Pain Management Following Knee Alloplasty. Int. J. Appl. Mech. Eng. 2017, 22, 241–251. [Google Scholar] [CrossRef] [Green Version]
  14. Trochimczuk, R.; Kuźmierowski, T. Kinematic analysis of CPM machine supporting to rehabilitation process after surgical knee arthroscopy and arthroplasty. Int. J. Appl. Mech. Eng. 2014, 19, 841–848. [Google Scholar] [CrossRef]
  15. Ivlev, O. Soft fluidic actuators of rotary type for safe physical human-machine interaction. In Proceedings of the 2009 IEEE International Conference on Rehabilitation Robotics, Kyoto, Japan, 23–26 June 2009; pp. 1–5. [Google Scholar]
  16. Wolff, J. The Law of Bone Remodeling; Springer: Berlin/Heidelberg, Germany, 1986. [Google Scholar]
  17. Calabrò, R.S.; Billeri, L.; Andronaco, V.A.; Accorinti, M.; Milardi, D.; Cannavò, A.; Aliberti, E.; Militi, A.; Bramanti, P.; Naro, A. Walking on the Moon: A randomized clinical trial on the role of lower body positive pressure treadmill training in post-stroke gait impairment. J. Adv. Res. 2020, 21, 15–24. [Google Scholar] [CrossRef] [PubMed]
  18. Gojkovic, Z.; Ivancevic, T.; Jovanovic, B. Biomechanical model of swimming rehabilitation after hip and knee surgery. J. Biomech. 2019, 94, 165–169. [Google Scholar] [CrossRef]
  19. Damm, P.; Kutzner, I.; Bergmann, G.; Rohlmann, A.; Schmidt, H. Comparison of in vivo measured loads in knee, hip and spinal implants during level walking. J. Biomech. 2017, 51, 128–132. [Google Scholar] [CrossRef]
  20. Jung, Y.; Phan, C.-B.; Koo, S. Intra-articular knee contact force estimation during walking using force-reaction elements and subject-specific joint model. J. Biomech. Eng. 2016, 138. [Google Scholar] [CrossRef]
  21. Eskinazi, I.; Fregly, B.J. A computational framework for simultaneous estimation of muscle and joint contact forces and body motion using optimization and surrogate modeling. Med. Eng. Phys. 2018, 54, 56–64. [Google Scholar] [CrossRef]
  22. Ali, A.A.; Mannen, E.M.; Rullkoetter, P.J.; Shelburne, K.B. Validated Computational Framework for Evaluation of In Vivo Knee Mechanics. J. Biomech. Eng. 2020, 142, 81003. [Google Scholar] [CrossRef]
  23. Stetter, B.J.; Ringhof, S.; Krafft, F.C.; Sell, S.; Stein, T. Estimation of knee joint forces in sport movements using wearable sensors and machine learning. Sensors 2019, 19, 3690. [Google Scholar] [CrossRef] [Green Version]
  24. Liu, K.; Yan, J.; Liu, Y.; Ye, M. Noninvasive Estimation of Joint Moments with Inertial Sensor System for Analysis of STS Rehabilitation Training. J. Healthc. Eng. 2018, 18, 971–986. [Google Scholar] [CrossRef] [Green Version]
  25. DeBerardinis, J.; Trabia, M.B.; Dufek, J.S.; Le Gall, Y.; Da Silva Sacoto, N. Enhancing the Accuracy of Vertical Ground Reaction Force Measurement During Walking Using Pressure-Measuring Insoles. J. Biomech. Eng. 2021, 143, 011010. [Google Scholar] [CrossRef] [PubMed]
  26. Martini, E.; Fiumalbi, T.; Dell’Agnello, F.; Ivanić, Z.; Munih, M.; Vitiello, N.; Crea, S. Pressure-Sensitive Insoles for Real-Time Gait-Related Applications. Sensors 2020, 20, 1448. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  27. Stolz, B.; Grim, C.; Lutter, C.; Gelse, K.; Schell, M.; Swoboda, B.; Carl, H.-D.; Hotfiel, T. Assessing Foot Loads in Continuous Passive Motion (CPM) and Active Knee Joint Motion Devices. Sportverletz. Sportschaden 2019, 35, 18–23. [Google Scholar] [CrossRef] [PubMed]
  28. Jnsson, M.; Munkhammar, T.; Norrbrand, L.; Berg, H.E. Foot centre of pressure and ground reaction force during quadriceps resistance exercises; a comparison between force plates and a pressure insole system. J. Biomech. 2019, 87, 206–210. [Google Scholar] [CrossRef]
  29. Koch, M.; Lunde, L.-K.; Ernst, M.; Knardahl, S.; Veiersted, K.B. Validity and reliability of pressure-measurement insoles for vertical ground reaction force assessment in field situations. Appl. Ergon. 2016, 53 (Pt A), 44–51. [Google Scholar] [CrossRef]
  30. Kellis, E. Tibiofemoral joint forces during maximal isokinetic eccentric and concentric efforts of the knee flexors. Clin. Biomech. 2001, 16, 229–236. [Google Scholar] [CrossRef]
  31. Nisell, R.; Ericson, M.O.; Nemeth, G.; Ekholm, J. Tibiofemoral joint forces during isokinetic knee extension. Am. J. Sports Med. 1989, 17, 49–54. [Google Scholar] [CrossRef]
  32. Lidstone, D.E.; DeBerardinis, J.; Dufek, J.S.; Trabia, M.B. Electronic measurement of plantar contact area during walking using an adaptive thresholding method for Medilogic® pressure-measuring insoles. Foot 2019, 39, 1–10. [Google Scholar] [CrossRef]
  33. Deberardinis, J.; Dufek, J.S.; Trabia, M.B.; Lidstone, D.E. Assessing the validity of pressure-measuring insoles in quantifying gait variables. J. Rehabil. Assist. Technol. Eng. 2018, 5, 2055668317752088. [Google Scholar] [CrossRef] [Green Version]
  34. Nisell, R.; Németh, G.; Ohlsén, H. Joint forces in extension of the knee: Analysis of a mechanical model. Acta Orthop. Scand. 1986, 57, 41–46. [Google Scholar] [CrossRef]
  35. Riener, R.; Edrich, T. Identification of passive elastic joint moments in the lower extremities. J. Biomech. 1999, 32, 539–544. [Google Scholar] [CrossRef]
  36. Adolphe, M.; Clerval, J.; Kirchof, Z.; Lacombe-Delpech, R.; Zagrodny, B. Center of mass of human’s body segments. Mech. Mech. Eng. 2017, 21, 485–497. [Google Scholar]
  37. GB/T 17245-2004; Inertial Parameters of Adult Human Body. China National Standardization Administration Committee: Beijing, China, 2004; pp. 1–16.
  38. Bergmann, G.; Bender, A.; Graichen, F.; Dymke, J.; Rohlmann, A.; Trepczynski, A.; Heller, M.O.; Kutzner, I. Standardized loads acting in knee implants. PLoS ONE 2014, 9, e86035. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  39. Pedowitz, R.; Gershuni, D.; Crenshaw, A.; Petras, S.; Danzig, L.; Hargens, A. Intraarticular pressure during continuous passive motion of the human knee. J. Orthop. Res. 1989, 7, 530–537. [Google Scholar] [CrossRef]
  40. D’Lima, D.D.; Patil, S.; Steklov, N.; Chien, S., Jr.; Colwell, C.W. In vivo knee moments and shear after total knee arthroplasty. J. Biomech. 2007, 40, S11–S17. [Google Scholar] [CrossRef] [PubMed]
  41. Kutzner, I.; Heinlein, B.; Graichen, F.; Bender, A.; Rohlmann, A.; Halder, A.; Beier, A.; Bergmann, G. Loading of the knee joint during activities of daily living measured in vivo in five subjects. J. Biomech. 2010, 43, 2164–2173. [Google Scholar] [CrossRef]
  42. Laberge, M.A.; Baum, T.; Virayavanich, W.; Nardo, L.; Nevitt, M.C.; Lynch, J.; Mcculloch, C.E.; Link, T.M. Obesity increases the prevalence and severity of focal knee abnormalities diagnosed using 3T MRI in middle-aged subjects—Data from the Osteoarthritis Initiative. Skelet. Radiol. 2012, 41, 633–641. [Google Scholar] [CrossRef] [Green Version]
  43. Werner, B.C.; Rawles, R.B.; Jobe, J.T.; Chhabra, A.B.; Freilich, A.M. Obesity is associated with increased postoperative complications after operative management of distal humerus fractures. J. Shoulder Elb. Surg. 2015, 24, 1602–1606. [Google Scholar] [CrossRef]
  44. Järvenpää, J.; Kettunen, J.; Kröger, H.; Miettinen, H. Obesity may impair the early outcome of total knee arthroplasty a prospective study of 100 patients. Scand. J. Surg. 2010, 99, 45–48. [Google Scholar] [CrossRef] [Green Version]
  45. Benotti, P.N.; Still, C.D.; Wood, G.C.; Seiler, J.L.; Seiler, C.J.; Thomas, S.P.; Petrick, A.T.; Suk, M.; Irving, B.A.; Samuels, J. Surgical weight-loss to improve functional status trajectories following total knee arthroplasty: SWIFT trial: Rationale, design, and methods. Contemp. Clin. Trials 2018, 69, 1–9. [Google Scholar] [CrossRef]
  46. McLawhorn, A.S.; Southren, D.; Wang, Y.C.; Marx, R.G.; Dodwell, E.R. Cost-effectiveness of bariatric surgery prior to total knee arthroplasty in the morbidly obese: A computer model-based evaluation. J. Bone Jt. Surg. 2016, 98, e6. [Google Scholar] [CrossRef] [PubMed]
  47. Peng, Y.; Zhang, Z.; Gao, Y.; Chen, Z.; Xin, H.; Zhang, Q.; Fan, X.; Jin, Z. Concurrent prediction of ground reaction forces and moments and tibiofemoral contact forces during walking using musculoskeletal modelling. Med. Eng. Phys. 2018, 52, 31–40. [Google Scholar] [CrossRef] [PubMed]
  48. Lenton, G.K.; Bishop, P.J.; Saxby, D.J.; Doyle, T.L.; Pizzolato, C.; Billing, D.; Lloyd, D.G. Tibiofemoral joint contact forces increase with load magnitude and walking speed but remain almost unchanged with different types of carried load. PLoS ONE 2018, 13, e0206859. [Google Scholar] [CrossRef] [PubMed]
  49. Yi, C.; Zhang, S.; Jiang, F.; Liu, J.; Ding, Z.; Yang, C.; Zhou, H. Enable Fully Customized Assistance: A Novel IMU-based Motor Intent Decoding Scheme. IEEE Trans. Cogn. Dev. Syst. 2021. early access. [Google Scholar] [CrossRef]
  50. Feng, Y.; Tsai, T.-Y.; Li, J.-S.; Rubash, H.E.; Li, G.; Freiberg, A. In-vivo analysis of flexion axes of the knee: Femoral condylar motion during dynamic knee flexion. Clin. Biomech. 2016, 32, 102–107. [Google Scholar] [CrossRef] [Green Version]
  51. Mallat, R.; Khalil, M.; Venture, G.; Bonnet, V.; Mohammed, S. Human-Exoskeleton Joint Misalignment: A Systematic Review. In Proceedings of the 2019 Fifth International Conference on Advances in Biomedical Engineering (ICABME), Tripoli, Lebanon, 17–19 October 2019; pp. 1–4. [Google Scholar]
  52. Johnson, W.R.; Mian, A.; Donnelly, C.J.; Lloyd, D.; Alderson, J. Predicting athlete ground reaction forces and moments from motion capture. Med. Biol. Eng. Comput. 2018, 56, 1781–1792. [Google Scholar] [CrossRef] [Green Version]
  53. Yi, C.; Park, S.; Yang, C.; Jiang, F.; Ding, Z.; Zhu, J.; Liu, J. Muscular Human Cybertwin for Internet of Everything: A Pilot Study. IEEE Trans. Ind. Inform. 2022. early access. [Google Scholar] [CrossRef]
  54. Zhang, W.; Tomizuka, M.; Byl, N. A Wireless Human Motion Monitoring System for Smart Rehabilitation. J. Dyn. Syst. Meas. Control 2016, 138, 111004. [Google Scholar] [CrossRef] [Green Version]
  55. Alice, B.-M.; Stéphane, A.; Yoshisama, S.J.; Pierre, H.; Domizio, S.; Hermes, M.; Katia, T. Evolution of knee kinematics three months after total knee replacement. Gait Posture 2015, 41, 624–629. [Google Scholar] [CrossRef]
Figure 1. Acquisition system used in the experiments. (a) Sketch of the experimental setup; (b) graphical user interface and receiving device.
Figure 1. Acquisition system used in the experiments. (a) Sketch of the experimental setup; (b) graphical user interface and receiving device.
Sensors 22 04947 g001
Figure 2. Curve of pressure soles during calibration. (a) Curve of the sensors summed for participant A4; (b) Example curve of foot pressure.
Figure 2. Curve of pressure soles during calibration. (a) Curve of the sensors summed for participant A4; (b) Example curve of foot pressure.
Sensors 22 04947 g002
Figure 3. CoP distribution of the foot pressure.
Figure 3. CoP distribution of the foot pressure.
Sensors 22 04947 g003
Figure 4. Free-body diagram of the femur and tibia during CPM cycles. (a) F0, the force generated by plantar pressure; F1, the resultant force of F0; Fn, the normal force imparted by the tibia brace; G, the weight of the leg–foot system; Mk, the passive torque of the knee generated by the tissue; β, the angle of tibia long-axis incline; α, the angle between Fs and G; dn, the distance from the center of mass of the leg–foot system to the knee axis [36]; dG, the moment arm of G around knee axis; ψk, the knee flexion angle; Fc, the tibiofemoral compressive force; Fs, the tibiofemoral shear force; ψh, the hip flexion angle. (b) Forces and distances for the free-body diagram of the femur and tibia during knee extension.
Figure 4. Free-body diagram of the femur and tibia during CPM cycles. (a) F0, the force generated by plantar pressure; F1, the resultant force of F0; Fn, the normal force imparted by the tibia brace; G, the weight of the leg–foot system; Mk, the passive torque of the knee generated by the tissue; β, the angle of tibia long-axis incline; α, the angle between Fs and G; dn, the distance from the center of mass of the leg–foot system to the knee axis [36]; dG, the moment arm of G around knee axis; ψk, the knee flexion angle; Fc, the tibiofemoral compressive force; Fs, the tibiofemoral shear force; ψh, the hip flexion angle. (b) Forces and distances for the free-body diagram of the femur and tibia during knee extension.
Sensors 22 04947 g004
Figure 5. Mean (and standard error) of the estimated resultant TFFs from three subjects during typical CPM cycles. (ad) represent 60°−120°, respectively; 1 = first major peak during knee flexion; 2 = second major peak during knee extension. The shaded area represents ±1 standard deviation.
Figure 5. Mean (and standard error) of the estimated resultant TFFs from three subjects during typical CPM cycles. (ad) represent 60°−120°, respectively; 1 = first major peak during knee flexion; 2 = second major peak during knee extension. The shaded area represents ±1 standard deviation.
Sensors 22 04947 g005aSensors 22 04947 g005b
Figure 6. Mean of the major peak resultant TFFs for four ROM across CPM cycles. All significant results were compared with the ROM of 80° (* p < 0.05).
Figure 6. Mean of the major peak resultant TFFs for four ROM across CPM cycles. All significant results were compared with the ROM of 80° (* p < 0.05).
Sensors 22 04947 g006
Figure 7. Mean of the estimated TFCFs (solid lines) and TFSFs (dashed lines) from participants during three repeated typical CPM cycles at different ROMs; 1 = First major peak during knee flexion; 2 = second major peak during knee extension. A1 CF represents the compressive forces of A1, and A1 SF represents the shear forces of A1. (ad) represent 60°–120°, respectively.
Figure 7. Mean of the estimated TFCFs (solid lines) and TFSFs (dashed lines) from participants during three repeated typical CPM cycles at different ROMs; 1 = First major peak during knee flexion; 2 = second major peak during knee extension. A1 CF represents the compressive forces of A1, and A1 SF represents the shear forces of A1. (ad) represent 60°–120°, respectively.
Sensors 22 04947 g007
Figure 8. Measured plantar pressure curves.
Figure 8. Measured plantar pressure curves.
Sensors 22 04947 g008
Table 1. Overview of participant data.
Table 1. Overview of participant data.
GroupL (mm)Weight (kg)Age (Years)Height (cm)
A (60−68 kg)A148561.922178
A246463.826170
A346563.926171
A447864.825175
B (70−78 kg)B148070.726176
B251073.426185
B349075.223180
C (80−88 kg)C146582.723172
C248583.327178
C348186.323176
C446587.826170
Table 2. Mean values (+SD) of maximum resultant TFFs, TFCFs, and TFSFs (×BW) during various testing conditions.
Table 2. Mean values (+SD) of maximum resultant TFFs, TFCFs, and TFSFs (×BW) during various testing conditions.
Max Force (×BW)Group60° (SD)80° (SD)100° (SD)120° (SD)
Resultant TF forceA0.16 (0.05)0.26 (0.02)0.23 (0.04)0.24 (0.03)
B0.24 (0.05)0.35 (0.11)0.33 (0.09)0.33 (0.13)
C0.31 (0.10)0.40 (0.09)0.31 (0.03)0.31 (0.04)
Compressive forceA0.16 (0.04)0.26 (0.04)0.21 (0.04)0.21 (0.04)
B0.22 (0.05)0.34 (0.08)0.30 (0.09)0.31 (0.12)
C0.30 (0.11)0.36 (0.10)0.29 (0.03)0.29 (0.05)
Shear forceA0.08 (0.03)0.09 (0.03)0.10 (0.04)0.10 (0.03)
B0.10 (0.01)0.12 (0.02)0.12 (0.03)0.11 (0.05)
C0.12 (0.01)0.13 (0.02)0.11 (0.01)0.11 (0.01)
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Yang, Y.; Huang, H.; Guo, J.; Yu, F.; Yao, Y. Estimation of Tibiofemoral Joint Contact Forces Using Foot Loads during Continuous Passive Motions. Sensors 2022, 22, 4947. https://doi.org/10.3390/s22134947

AMA Style

Yang Y, Huang H, Guo J, Yu F, Yao Y. Estimation of Tibiofemoral Joint Contact Forces Using Foot Loads during Continuous Passive Motions. Sensors. 2022; 22(13):4947. https://doi.org/10.3390/s22134947

Chicago/Turabian Style

Yang, Yunlong, Huixuan Huang, Junlong Guo, Fei Yu, and Yufeng Yao. 2022. "Estimation of Tibiofemoral Joint Contact Forces Using Foot Loads during Continuous Passive Motions" Sensors 22, no. 13: 4947. https://doi.org/10.3390/s22134947