Review

Flow-Injection Methods in Water Analysis—Recent Developments

by
Marek Trojanowicz
1,2,* and
Marta Pyszynska
1
1
Laboratory of Nuclear Analytical Methods, Institute of Nuclear Chemistry and Technology, Dorodna 16, 02-195 Warsaw, Poland
2
Department of Chemistry, University of Warsaw, Pasteura 1, 02-093 Warsaw, Poland
*
Author to whom correspondence should be addressed.
Molecules 2022, 27(4), 1410; https://doi.org/10.3390/molecules27041410
Submission received: 9 November 2021 / Revised: 11 February 2022 / Accepted: 15 February 2022 / Published: 19 February 2022
(This article belongs to the Special Issue Frontiers in Analytical Methods for Water Analysis)

Abstract

:
Widespread demand for the analysis and control of water quality and supply for human activity and ecosystem sustainability has necessitated the continuous improvement of water analysis methods in terms of their reliability, efficiency, and costs. To satisfy these requirements, flow-injection analysis using different detection methods has successfully been developed in recent decades. This review, based on about 100 original research papers, presents the achievements in this field over the past ten years. Various methodologies for establishing flow-injection measurements are reviewed, together with microfluidics and portable systems. The developed applications mostly concern not only the determination of inorganic analytes but also the speciation analysis of different elements, and the determination of several total indices of water quality. Examples of the determination of organic residues (e.g., pesticides, phenolic compounds, and surfactants) in natural surface waters, seawater, groundwater, and drinking water have also been identified. Usually, changes in the format of manual procedures for flow-injection determination results in the improvement of various operational parameters, such as the limits of detection, the sampling rate, or selectivity in different matrices.

1. Introduction

The protection of natural water resources on a global scale is one of the most significant challenges of civilization at the beginning of the third decade of the 21st century. Escalating climatic change is the largest contributor to the greenhouse effect, connected with the alteration of atmospheric composition and the increasing emission of anthropogenic pollutants into the environment, which contribute to a reduction in the available volume of fresh natural waters and the lowering of the below-ground water table. They are also responsible for biological degradation and the reduction in chemical quality of the whole aquatic environment, including surface water, groundwater, and the seas and oceans. This key problem for modern civilization is currently the object of very intensive biological, chemical and physical scientific investigation, as well as numerous initiatives and research programs in the fields of environmental engineering and the design of measuring instruments, and systems for monitoring environmental changes [1]. This also includes the search for legislative and technological solutions, in order to stop or reduce those catastrophic environmental changes.
One of the most significant factors affecting the current status of natural waters, which are very difficult to fully control and for which limitations need to be introduced, is the emission of anthropogenic pollutants into the environment. Although in different proportions, this problem concerns the behavior of individual members of the population as well as on a much larger scale, in terms of the processes associated not only with different branches of industry but also with agriculture, transport, and the mining of fossil-fuel materials and minerals. In the middle of the previous century, the main identified anthropogenic pollutants of natural waters were heavy metals ions, the residues of pesticides, and some groups of organic compounds of industrial origin, including primarily the so-called persistent organic pollutants (POPs), such as organic solvents or detergents. Now, the list of these anthropogenic chemical environmental pollutants is much longer, which is confirmed by the very extensive literature in the fields of environmental chemistry and engineering, toxicology, and health protection. Since the 1970s, particular attention has been focused on endocrine-disrupting compounds (EDCs). Apart from pesticides, these include not only plasticizers and phenolic compounds but also such emerging contaminants as human and veterinary drugs, and certain ingredients in personal care products and cosmetics [2,3]. The main reason for this interest is the fact that certain EDCs, for instance, steroids, nonsteroidal antibiotics, analgesics, and anti-inflammatories or anti-depressants, may induce physiological effects in humans, even at very low doses [4]. About 160 human and veterinary pharmaceutical products and about 30 by-products are considered to carry this risk, spread over 24 therapeutic classes [5]. The main source of their presence in the environment, and hence in drinking water, for example, is the discharge of municipal wastewater effluents [6]. Another group of organic anthropogenic environmental pollutants on which particular attention has been focused since the early 2000s is the poly- and perfluoroalkyl substances (PFASs). Due to their exceptional chemical stability and numerous commercial applications, they frequently occur in an aqueous environment [7]. Wastewater treatment plants are considered likely to be the main contributor to their presence in the surface waters and, hence, also in drinking water [8], as well as industrial sites, military firing ranges and training areas, and civilian airports [9].
In the past two decades, completely different anthropogenic pollutants, such as microplastics and nanomaterials, are pointed out as especially problematic in aquatic systems. The frequent discarding of plastics into natural water bodies, followed by degradation via mechanical and photochemical fragmentation, results in the formation of polymeric microparticles that may remain in the environment for hundreds of years, additionally accumulating other chemical pollutants [10]. The increasing importance of nanotechnology on an industrial scale means that engineered nanoparticles are commonly detected in natural water bodies [11]. Their high surface-to-volume ratio and outstanding reactivity are the sources of their highly dynamic transformations in the environment [12]. Special attention is also being paid to the presence in the aquatic environment of silver nanoparticles [13], and the transformation and bioavailability of metal oxide nanoparticles [14]. The latter, due to their electronic and magnetic properties, are widely employed as components in different commercial products (e.g., polymers and catalysts), finding numerous environmental applications (sensing, remediation); they are also employed in health care as antimicrobials and in cancer treatment.
Monitoring the aquatic environment and applying efficient methods for its protection is impossible without employing adequate chemical analytical methods. Besides the adaptation of analytical methods for laboratory, field, or process-control use, they should fulfill general criteria regarding the quality of chemical analysis. These include suitability for a given range of analyte concentrations, selectivity, the required precision of determination, and accuracy. In the present state of development in analytical chemistry, an exceptionally rich arsenal of measuring instruments and developed analytical methods is available for the analysis of water bodies and wastewaters of different origins. They also allow the virtual monitoring of each type of identified pollutant. Certain limitations to their practical use may be due to economic reasons in particular circumstances or to the availability of sufficiently trained personnel.
The literature available on water analysis is very wide-ranging and includes numerous dedicated books, original research articles, and reviews in scientific and technical journals, as well as the standard procedures reported in the regulations of various organizations, such as ISO, EPA, etc. Periodic reviews on the newest achievements in this field are published, e.g., in the ACS journal, Analytical Chemistry [15]. Determination of the trace content of organic micropollutants in samples with complex matrices is particularly challenging, for which high-resolution mass spectrometry and multidimensional chromatography are especially effective [16], as well as systems hyphenating high-performance liquid chromatography or gas chromatography instruments with mass spectrometry or nuclear magnetic resonance [17]. A particular difficulty in the trace determination of, e.g., emerging persistent organic pollutants is the development of efficient methods for the isolation and preconcentration of analytes [18,19]. In recent decades, one very favorable element of the scaling-down of measuring instrumentation, including their applications in water analysis, is the development of sensor technology. Generally speaking, this includes reagent-free measuring devices, which must be low-maintenance and autonomous; these are specific for a given analyte or group of analytes with the use of the appropriate detection technique [20]. Good examples of such devices that are reported in the literature and commercially available sensors and biosensors include semiconductor-based heavy metal ion sensors [21], membrane ion-selective electrodes, as well as biosensors based on the use of enzymes, antibodies, DNA, aptamers, and other synthetic biology tools [22]. New challenges in this field include the development of efficient methods for the determination of nanoparticles in the environment [23,24] and identifying microplastics [25]. A separate function, also contributing to water quality control, is microbial water analysis [26].

2. Flow Chemistry in Contemporary Chemical Analysis

Because of a widespread need for the analysis of waters and wastewaters and for monitoring of the aqueous environment, the miniaturization of measuring devices plays quite an essential role. This is manifested by the increasing application of integrated sensors and biosensors, as well as the design of flow analyzers [27]. Measuring devices for the continuous monitoring of aqueous ecosystems also need to take into account the rules of green analytical chemistry in their design as much as is technologically possible, particularly in terms of the limitation of reagent consumption and the generation of waste [28]. In this context, flow injection analysis (FIA) methods are considered to be especially suitable [29].
Flow chemistry is an increasingly recognized and exploited field of chemical science, in which the hydrodynamic properties of flowing liquids are utilized for conducting chemical reactions. This concerns the conducting of reactions for synthetic purposes (unfortunately, in this case, given the commonly but rather improperly used term “flow chemistry”), for conducting analytical determinations of flow conditions, for carrying out basic kinetic investigations, and also for performing industrial technological processes [30]. In all these fields, the flow conditions are utilized for regulating the transport of reagents, improving interphase contacts, for the enhancement of heat transfer, and for the safe manipulating of hazardous substances.
Pioneering steps in flow synthesis have been associated with the use of flow-through reactors (see, e.g., [31]). Regarding the first analytical applications of flow chemistry, one can refer to much earlier works on column chromatography [32,33,34]. Although they were, undoubtedly, analytical works, the pioneering studies by Skeggs on the development of “flow analysis” in the middle of the 1950s are well respected [35]. One bewildering situation must be acknowledged: even in very competent reviews, published in the most prestigious journals (see, e.g., [36]), one can find the opinion that “continuous flow has affected many fields over the last 20 years” without there being any acknowledgment that analytical applications have been developed for almost 70 years, in thousands of publications, and in widely and routinely used procedures. In those pioneering works by Skeggs, the essential instrumental novelty was the development of multistep analytical procedures for photometric determination. These were carried out under continuous-flow conditions, via the configuration of different flow-through modules, for example, for mixing solutions, incubation, dialysis, and detection. This method of mechanizing analytical determinations resulted in an evident improvement of the efficiency of determination, and in significant enhancement of the precision of analyses. This concept has been widely accepted, first in analytical clinical laboratories, and very soon after in environmental, agricultural, and industrial ones [37].
As a further essential improvement of that analytical methodology, which resulted in the reduction of the required sample volume, the decrease in consumption of reagents, and the shortening of measurement time, the invention of flow injection analysis is commonly recognized as significant. At the beginning of the 1970s, several research groups [38,39,40,41,42] invented the concept of flow analysis, with the injection of a small sample volume into a flowing stream of carrier solution. The term “flow-injection analysis” was introduced by Ruzicka and Hansen in a series of ten papers published in the journal Analytica Chimica Acta in 1975–1978; since then, it has been widely accepted by the analytical community. A further instrumental improvement of the flow-injection methodology was brought by the concept of sequential-injection analysis (SIA), mostly due to an even better cost-efficiency of determinations than that usually achieved by FIA systems [43]. In SIA systems, small segments of the sample and reagent(s) are introduced into the holding coil, which is then followed by pumping them in the opposite direction toward the detector. During this process, the partial overlapping of segments usually takes place, which allows the reaction to occur between the analyte and the reagent. The significant interest of analysts in flow analytical methods and the development of their numerous practical applications over six decades has resulted in the design of many types of flow analytical systems [44]; examples of those that are most commonly used are given in Figure 1.
The methodology of flow-injection measurements and their different variants have a very solid place in modern analytical chemistry [45]. It is the subject of thousands of original research papers and numerous books, and the apparatus for performing it is commercially available; however, it can be assembled relatively simply from parts and accessories usually found in chemical research laboratories. The main attributes of those methods are the small sample volume needed (usually 50 to 100 μL), detection during the flow of the sample solution through a detector, and the conducting of possibly all required sample processing operations in a properly designed flow-through module.
Figure 1. Schemes of the manifolds of basic flow injection systems [45]: (A)—normal flow injection analysis (FIA) system, (B)—reverse flow injection analysis (rFIA) system, (C)—sequential injection analysis (SIA) system, (D)—batch injection analysis (BIA) system, (E)—multisyringe flow injection analysis (MSFIA) system, (F)—multicommutated flow analysis with a multichannel pump system, (G)—multi-pumping flow analysis system. Reproduced with permission from Analyst; published by the Royal Society of Chemistry, 2016.
Figure 1. Schemes of the manifolds of basic flow injection systems [45]: (A)—normal flow injection analysis (FIA) system, (B)—reverse flow injection analysis (rFIA) system, (C)—sequential injection analysis (SIA) system, (D)—batch injection analysis (BIA) system, (E)—multisyringe flow injection analysis (MSFIA) system, (F)—multicommutated flow analysis with a multichannel pump system, (G)—multi-pumping flow analysis system. Reproduced with permission from Analyst; published by the Royal Society of Chemistry, 2016.
Molecules 27 01410 g001
In such cases, virtually all analytical techniques of detection that are available for taking measurements in solutions can be used, although the largest number of applications were developed for UV-Vis absorption spectrophotometry (see. e.g., review [46]). Such systems are usually constructed for the determination of single analytes, but the setups for flow multicomponent determination can also be designed in different ways. This can be achieved by a smart extension of the flow manifold with various detectors, or via the application of multicomponent detectors, as well as by suitable data processing using advanced chemometric methods (see, e.g., [47]). One example of such an approach is the so-called multicommuted system, where, with the application of a combination of single-channel solenoid micro-pumps and valves, or by the use of a multichannel pump with several solenoid valves (see Figure 1F,G), one can achieve time-based control of the sample injection and merging of segments of solutions [48]. In order to reduce additionally the consumption of reagents, a configuration of reverse flow-injection analysis can be employed, whereby a very small volume of reagent (usually also 50 to 100 μL) is injected into the setup along with a continuous flow of aspirated samples (Figure 1B) (see, e.g., review [49]). A very wide application of the FIA and SIA concepts is also reported in numerous papers presenting multisyringe pumping systems (Figure 1E) [45,50]. In such a configuration, the flow of solutions in a single or multichannel setup is propelled by the assembly of mechanically controlled syringes as, used commonly, for instance, in mechanized titrators. Each syringe head is equipped with a solenoid-driven valve to direct the solution flow. The scheme shown in Figure 1D depicts the principle of a so-called batch injection analysis (BIA) system, where the sample to be analyzed is injected directly onto the sensing surface of the detector. This provides a very fast response, but the possibility of on-line sample processing is very limited in such a case.
One very frequent misunderstanding is in calling flow analytical systems automated systems. In reality, this term should only be used in the situation when the employed computerized control system is able, with the use of a feedback loop in the software, to optimize the operating conditions of sample determination without the intervention of a human operator. It must be admitted that, in fact, in recent years there have been some examples of such smart systems (see, e.g., the recent review by [51]).
One important development trend in flow-injection instrumentation that has been observed in the past few decades is the miniaturization and integration of different components and modules of the measuring setup into a single instrument. This was initiated firstly by the development of integrated micro conduits [52], then, the design of multifunctional injection valves, the so-called lab-on-valve (LOV) with incorporated miniaturized detectors or microcolumns packed with solid sorbents [53]. A valuable innovation in this trend, from the point of view of different applications, including water analysis, is the adaptation of a design for such systems of a microfluidic device, with flow-through channels that are usually of a few tens of micrometers. This trend was begun at the end of the previous century (e.g., [54]), and then additionally enhanced by progress in nanotechnology [55], as well as by technology for their preparation in paper matrices [56].

3. Flow-Injection Methods in Water Analysis

Among the different applications of flow-injection analysis, their applications in water analysis are particularly broad. The method is favored first of all because of the wide demand for carrying out water quality control analyses for municipal needs, for industrial monitoring, and also for food production. In the first two decades of the development of laboratory flow analytical methods, with the segmentation of flowing streams and the use of the steady-state detector response as an analytical signal, their main field of application has been in clinical analysis. However, soon after, environmental applications have appeared, primarily in water analysis. The method’s medical applications were replaced in subsequent decades by the more efficient discrete analyzers, while from the middle of the 1970s, numerous applications of FIA methods have already focused on the analysis of waters of various origins.
Natural waters present a relatively simple type of matrix for the conducting of fast-flow analysis determinations since, except for seawater, they are not heavily loaded matrices requiring complex sample processing prior to performing the detection operation. The main advantages of using the various FIA methods include the mechanization of analytical procedures, leading to more efficient analyses and improvements in the precision of determination. The possibility of conducting the necessary sample processing in on-line mode, which results in the simplification of the whole multistep procedure, is especially attractive. The application of appropriate operations when isolating the analyte(s) from the sample matrix or the inclusion of a preconcentration step results in an enhancement of the selectivity of determination and an improvement in the limit of detection (LOD). Properly designed FIA systems can be fully integrated and mechanized complete analyzers or can serve as accessories for preliminary sample processing, prior to chromatographic, capillary electrophoretic, or mass spectrometry analyses.
A comprehensive description of the development of FIA methods for water analysis can be found in numerous books, dozens of review articles in scientific journals, and in the very rich literature of the original research works. The subject of this review is the presentation of development trends in this area of FIA applications in the past decade. Besides presenting the details of numerous original research works, it seems to be appropriate to mention several review papers that have been published on this subject since 2010 (Table 1). To some extent, these articles indicate the selection of the determined analytes, describe the detection methods employed, and show the configurations of the flow-injection systems developed. Among the analytes that have been determined, one can find trace amounts of numerous organic compounds [57,58], the residues of pharmaceuticals [59], and also many groups of inorganic analytes [60,61,62], including nutrients in oceanographic measurements [63], as well as the environmentally significant radionuclides [64,65,66]. Flow-injection analysis methods are also employed in the determination of the total indices of water quality [57,58,59,60,61,62,63,64,65,66,67]. Due to their usually rapid determinations, they are effective in the on-line monitoring of drinking water quality and can be employed as early warning analytical systems [68]. The recently published reviews also demonstrate one important role of the microfluidic systems developed for water analysis [69,70], and also, in recent years, microfluidic paper-based analytical devices [71], which offer particular advantages that, besides the miniaturization of measuring setups, should also include the low cost of their mass production.
Table 1. Review articles published in the past decade in scientific journals on the application of flow methods in water analysis.
Table 1. Review articles published in the past decade in scientific journals on the application of flow methods in water analysis.
Subject 1st Author,
Year of Publication
Reference
Automation of radiochemical analysis by applying flow techniques to environmental samplesFajardo, 2010[64]
Flow analysis techniques as effective tools for the improved environmental analysis of organic compounds, expressed as total indicesMaya, 2010[67]
Advances in on-line drinking water-quality monitoring and early warning systemsStorey, 2010[68]
Flow-based methods with chemiluminescence detection for food
and environmental analysis
Christodouleas, 2011[60]
Determination of total organic fluorine (TOF) in environmental samples, using flow-injection and chromatographic methodsMusijowski, 2011[57]
Advances in microfluidics for environmental analysisJokerst, 2012[69]
Flow-injection analysis as a tool for the determination of pharmaceutical residues in an aqueous environmentTrojanowicz, 2012[59]
Flow injection analysis as a tool for enhancing oceanographic nutrient measurementsWorsfold, 2013[63]
Application of microfluidics in waterborne pathogen monitoringBridle, 2014[70]
Application of flow analysis in the determination of selected radionuclidesKołacińska, 2014[65]
Analytical challenges and advantages of using flow-based methodologies for ammonia determination in estuarine and marine watersSraj, 2014[61]
Developments of microfluidic paper-based analytical devices (μPADs) for water analysisAlmeida, 2018[71]
Preconcentration of organic substances on low-polar adsorbents in the flow systems of analysisTsizin, 2018[58]
Dynamic flow, considered for automated radiochemical analysis in environmental, nuclear and medical applicationsQiao, 2020[66]
Development of testing methods for water quality using flow analysisTeshima, 2020[62]

3.1. Flow-Injection Systems with Spectroscopic Detection

In all the above-mentioned detection techniques employed in flow-injection analysis, the spectroscopic methods, both molecular and atomic spectroscopy, are the most widely used. This is particularly the case in the application of FIA methods to the analysis of natural waters and waters used after treatment for various purposes. As those applications have been developed over about half a century, they have gone through very different instrumental and methodological evolutions.
Without any doubt, the most commonly used detection methods are absorptive spectrophotometry and luminescence detections in the visible range of radiation; the UV or infrared range is much less frequently employed. This is also reflected in works published in the past decade [72,73,74,75,76,77,78,79,80,81,82,83,84,85,86,87,88,89,90,91,92,93,94,95,96,97,98,99,100,101,102,103,104,105,106,107,108,109,110,111,112,113,114,115,116], as can be seen in the data collected from the original research works in Table 2. The great popularity of these methods is generally the result of their wide use in analysis, not only in flow methods, because of their simplicity, the broad availability of the necessary apparatus and reagents, and the large number of already developed non-flow methods. Obviously, these recent developments are directed toward the improvement of detectability and selectivity, achieving a shorter time for analysis, the miniaturization of instrumentation, and the mechanization or automation of the whole procedure. Some basic information on recently developed procedures is given in Table 2.
In the determination of many analytes in natural waters, one crucial parameter, in many cases, is the possibility of achieving a low limit of detection (LOD). A simple instrumental tool to achieve this is to use long-path flow cells in the absorptive flow measurements [72,73]. In the determination of ammonia in seawater, the obtained LOD was at a nanomolar level [72], while in the case of nitrite and nitrate, it was even at a sub-nM level [73]. One especially attractive property of FIA systems is the possibility of carrying out different operations of sample processing in the flow mode, which practice is very widely used for the improvement of the detectability and selectivity of determinations. In the SIA-LOV system, a solid-phase extraction (SPE) with renewable beads was employed for the determination of the residues of chlorotriazine herbicides in ground- and tap water at a sub-μg L−1 level [74]. With the use of a 3D-printed disc module, SPE was employed for the trace analysis of Cr(VI) [75]. Then, in the SIA-LOV setup, the application of a complex formation on resin beads allowed researchers to obtain a sub-μg L−1 LOD in the determination of Pb(II) in both natural and tap water [76].
Particularly low levels of LOD (ng L−1) were reported in the determination of phosphate with preconcentration on a polymer inclusion membrane [77], as well as by coprecipitation in the determination of U(VI) in seawater [78]. Among other novel methods of on-line preconcentration of analytes in FIA systems with spectrophotometric detection, it is worth mentioning a smart in-syringe dispersive liquid-liquid extraction method for analyte isolation and enrichment [79], as well as magnetic-stirring-assisted dispersive liquid-liquid extraction [80]. Generally speaking, dispersive liquid-liquid extraction is a separation process that is based on the dispersion of fine droplets of the extracting solvent in an aqueous sample. Another relatively novel separation process employed for analytical purposes is pervaporation. This is a membrane-based process for the separation of liquid mixtures by partial vaporization, using non-porous membranes made of polymeric or ceramic materials. On-line pervaporation has been employed, for instance, for the elimination of turbidity and organic interference in the speciation analysis of arsenic in river water [81]. Several other spectrophotometric methods were also developed in the FIA and SIA systems [82,83,84,85,86]. Among general parameters characterizing the water quality, a novel attempt was reported regarding the FIA determination of total dissolved nitrogen, with the use of on-line UV digestion and reduction by VCl5 with an LOD of 40 μg L−1 [87]. The FIA method was also reported for the simultaneous determination of total dissolved nitrogen (TDN) and total dissolved phosphorus (TDP) [88]. As is shown in Figure 2, the developed system was equipped with one spectrophotometric detector, with two optical paths and on-line UV and thermal digestion. This is an especially interesting example of a smart system for a flow-injection procedure involving two complex on-line operations for sample processing. The reported sampling rate, of about 5 h−1, makes this setup particularly attractive for routine application.
The works listed in Table 2 indicate that FIA methods with spectrophotometric detection are much more frequently used for the determination of inorganic analytes than organic ones, due to the better selectivity of the reactions of inorganic analytes with color-forming reagents.
Systems developed for the multianalyte determinations seem to be especially attractive for routine applications, as also shown in Table 2. Besides the abovementioned system with one detector [88], another interesting setup is a reverse-FIA system for the determination of transient metal ions with LED-based multi-optical detection, which is depicted schematically in Figure 3 [89]. The use of a purpose-built multi-channel photometer allowed the researchers to obtain a very satisfactory efficiency of determination (4 analytes × 15 samples per h).
This methodology, which can also be employed for simultaneous determination under flow conditions, is called gradient titration. Such a system was reported, e.g., for the simultaneous determination of calcium and magnesium, with an LOD at sub-mg L−1 level, based on the use of complexometric titration in the SIA system with photometric detection [90]. It should be mentioned that the most common manual titrimetric methods with different detections are employed for millimolar rather than sub-mg L−1 concentrations of the analytes (except for some special techniques of microtitration). Very few reported determinations of organic analytes are carried out, mostly in FIA or SIA systems hyphenated to high-performance chromatographic separations [74,79,80,91].
Table 2. Application of molecular spectroscopy detection methods in flow-injection systems for water analysis.
Table 2. Application of molecular spectroscopy detection methods in flow-injection systems for water analysis.
Analyte(s)Type of WaterDetection MethodType of Flow SystemEmployed Sample Processing/RemarksLOD, mg L−11st Author, Year of Publ.Reference
Ag(I)Drinking waterTLSFIADetermination based on the formation of colloidal silver nanoparticles, via reduction with NaBH40.0015Korte, 2011[99,103]
AmmoniaSurface, sea, and tap watersFluoresc.SIASIA system with pressure-assisted dual-headspace gas-liquid microextraction module for membraneless gas separation50 ng L−1Giakisikli, 2018[93]
AmmoniumSea waterUV/Vis abs.FIAFIA system with on-line reaction in a knitted,
heated reaction coil and long-path cell
3.5 nMZhu, 2014[72]
AmmoniumRiver watersUV/Vis abs.Reverse FIA-70 nMLin, 2018[100]
Al(III)Surface and tap watersUV/Vis abs.Flow-batch systemSequential injection-mono-segmented flow system incorporating a mixing chamber0.020Khanhua-
thon, 2015
[101]
Al(III), Cr(V)Surface watersUV/Vis abs.Flow-batch systemMulti-commutation system, with a stop cell equipped
with a web camera for
digital imaging
Al: 0.00397
Cr: 0.00265
Andrade, 2013[102]
As speciation
(inorganic)
River waterUV/Vis abs.SIAOn-line pervaporation cell employed for the elimination of turbidity and organicsAs(III): 0.022 As(V): 0.051Boonjob, 2013[81]
Ca and MgSurface watersUV/Vis abs.SIASimultaneous determination using flow-injection gradient titrationCa 0.3
Mg 0.1
Kozak, 2018[90]
Cd(II) and Pb(II)Surface and well watersFluoresc.LOV-SIAOn-line SPE for separation and preconcentrationCd 0.0002
Pb 0.00017
Mattio, 2018[94]
Chlorotri-azine herbicidesGround and tapUV/Vis abs.LOV-SIA-
HPLC
SPE separation and preconcentration of analytes in LOV-SIA system and renewable beadsLOQ: 0.07–0.12
μg L−1
Boonjob, 2010[74]
Co(II)Well watersUV/Vis abs.Multisyringe FIAKinetic-catalytic determination with microconduit chip for efficient mixing20 ng L−1Abouhiat, 2017[104]
CODRiver and waste-watersChemlum.FIA-0.083Hue, 2017[97]
Cr(III), Cr(VI)River waterUV/Vis abs.FIAMicropumping multicommutated system with LED detectorCr(III) 2.05 Cr(VI) 1.0
g L−1
Pires, 2015[82]
Cr(VI)Surface and ground-
water
UV/Vis abs.Multisyringe FIAOn-line SPE preconcentration using a 3D-printed disc-based module 0.0005Calderilla, 2018[75]
Cu(II)River waterUV/Vis abs.LOV-SIASPE with renewable micro-beads0.003Yu, 2012[105]
CuRiver and waste waterUV/VIS abs.LOV-SIACatalytic method with SIA system, employing a micronduit for solution mixing 0.00012Phansi, 2014[117]
Cu and FeRiver watersUV/Vis abs.LOV-SIA-Cu 0.018
Fe 0.015
Gonzalez, 2017[106]
FeSea waterUV/Vis abs.LOV-FIAProcedure with stopped flow in a holding coil or flow-through cuvette0.00057Hatta, 2018[107]
Fe(II), Fe(III)Artesian waterUV/Vis abs.SIASingle peak procedure for the simultaneous determination of analytesFe(II) 0.04 Fe(III) 0.09 Kozak, 2016[83]
Fe(II), Fe(III)N.a.UV/Vis abs.FIA-ICLow-pressure chromatography with a post-column reaction and derivatization for detectionFe(II) 1.55 Fe(III) 3.09
μg L−1
Chen, 2015[84]
FluorideTap waterUV/Vis abs.FIAOn-line SPE preconcentration using a mini-column packed with layered double hydroxide sorbent 0.015Rocha, 2018[118]
Fluoroqui-
nolones
River waterFluoresc.FIA-HPLCOn-line SPE in FIA system hyphenated to HPLC for the determination of norfloxacin, ciprofloxacin and enrofloxacin6–19 ng L−1Peixoto, 2018[91]
Hg(II)Lake, river, tapChemlumFIAImmunoassay based on the use of resin beads, enzymatic amplification, and a novel monoclonal antibody15 ng L−1Xu, 2015[98]
NitrateTap, surface, waste-watersUV/Vis abs.FIA Detection with the use of paired emitter-detector diodes0.00073Cogan, 2013[108]
Nitrite, nitrateSeawaterUV/Vis abs.Multisyringe LOV-SIAExtended typical LOV by a chip, integrating reaction and mixing channelsNO2 30 nM NO3 100 nMHorstkotte, 2013[85]
Nitrite, nitrateSeawaterUV/Vis abs.Reverse FIADetection with a long path-length waveguide capillary cellBoth analytes 0.6 nMFeng, 2013[73]
Nitrite, nitrate, phosphate, silicateSeawaterUV/Vis abs.LOV-SIADeterminations based on a single, salinity-independent calibration, with standards prepared in distilled waterNO2 12 nM NO3 94 nM PO43− 47 nM SiO32− 0.24 μM Hatta, 2021[109]
Nitrite, phosphate, silicateArtificial seawaterUV/Vis abs.CFAFlow system with a bubble-free flow cellNO2 60 nM PO43− 80 nM SiO32− 110 nMLin, 2017[110]
Nitrite, nitrate, phosphate, Mn, Fe(II), Fe(III)Surface and tap watersUV/Vis abs.Reverse FIASequential determinations without sample processing0.03–0.7 μMLin, 2017[86]
Nitrogen, total dissolvedMineral, tap and surfaceUV/Vis abs.FIAOn-line UV digestion and reduction by VCl30.04 mg N L−1Lin, 2021[87]
ParabensTap- and seawaterUV SIA-HPLCSIA system with on-line in-syringe dispersive liquid-liquid extraction and robotic phase separator0.0003 to 0.0013Medina, 2018[79]
ParaquatRiver, tap, rice field waterUV/Vis abs.FIAOff-line preconcentration on cation exchanger 0.15Chuntib, 2015[111]
Pb(II)River watersUV/Vis abs.CFA3D-printed resin column for a sorbent, mixing coil and detection cell0.0027Mattio, 2017[112]
Pb(II)Ground, tap, well watersUV/Vis abs.Multisyringe LOV-SIAOn-line SPE of analyte, followed by complex formation on resin beads0.00079Rodriguez-
Maese, 2020
[76]
Phenolic compoundsMineral, well, tapUV/Vis abs.LOV-SIA-
HPLC
On-line magnetic-stirring-assisted dispersive liquid-liquid micro-extraction before HPLC0.04 to 0.46Gonzalez, 2015[80]
PhosphateSeawaterUV/Vis abs.Reverse FIA-50 ng L−1Li, 2012[119]
PhosphateMineral, melted snowUV/Vis abs.FIA Separation and preconcentration of an analyte with the use of a polymer inclusion membrane40 ng L−1 PO4-PNagul, 2013[77]
PhosphateSynthetic seawaterFluoresc.Reverse FIAFluorescence excited via an axially coupled fiber providing LED light and emission detected by photodiode0.00045 PO4-PKröckel, 2014[95]
Phosphate and silicateGround and river watersUV/Vis abs.FIASimultaneous determination based on single peak recordingPO4: 0.054 SiO3: 0.092Kozak, 2015[113]
SulfideHot spring watersFluorescenceFIASystem with on-line membrane-based micro-channel device for the extraction of free sulfides0.05 μMToda, 2012[96]
Surfactants, cationicMineral, tap, wellUV/Vis abs.Multisyringe FIAOn-line in-syringe dispersive liquid-liquid microextraction with solvent washing<30 nMHorstkotte, 2014[114]
TDN, TDPRiver waterUV/Vis abs.FIAOn-line UV and thermal digestion and reactions, leading to colored productsTDN 0.8 μM TDP 0.2 μMLin, 2018[88]
Th and USurface and tap watersUV/Vis abs.Multisyringe LOV-SIAOn-line SPE separation and preconcentrationTh 60 ng L−1 U 5.9 ng L−1Avivar, 2011[115]
Transition metal ions Cu, Fe(II), ZnMineral, river, tapUV/Vis abs.SIC
(FIA-HPLC)
Post-column derivatization for UV/Vis abs. detection 0.6–2.9 μM L−1Horstkotte, 2015[92]
Transition metal ions Cu, Mn, FeRiver waterUV/Vis abs.Reverse FIAMulticomponent reversed FIA system with LED-based multi-optical detection0.011–0.050Youngvises, 2017[89]
U(VI)SeawaterUV/Vis abs.FIAOn-line preconcentration by co-precipitation10 ng L−1Kuznetsov, 2014[78]
U(VI)Ground watersUV/Vis abs.Multisyringe FIA-0.3Danchana, 2019[116]
Abbreviations used: CFA—continuous flow analysis system, COD—chemical oxygen demand, ET–AAS—electrothermal atomic absorption spectrometry, FIA—flow injection analysis system, IC—ion-chromatography, LOV—lab-on-valve, SIA—sequential injection analysis system, SIC—sequential injection chromatography, SPE—solid-phase extraction, TDN—total dissolved nitrogen, TDP—total dissolved phosphorus, TLS—thermal lens spectrometry, UV/Vis abs.—UV/Vis absorption spectrophotometry.
Figure 2. Schematic diagram of an FIA system developed for the determination of the total dissolved nitrogen and phosphorus, with photometric detection and on-line UV and thermal digestion [88]. S—samples or standards; C—ultrapure water carrier stream; B—ammonium chloride buffer; OR—oxidation reagent; MRN—mixing reagent for nitrogen determination; MR—mixing reagent for phosphorus determination; AA—ascorbic acid; Cd—cadmium column; P1–P3—peristaltic pumps; IV—10-port injection valve; SV—8-position selector valve; DB—de-bubbler; L—light source; D—detector; FC—“U” shape flow cell; UV -UV-digester; HB—heated bath; RC—reaction coil; SL—sample loop; W—waste. The dashed line in IV represents the valve in position A, and the solid line represents the valve in position B. Reproduced with permission from Talanta, published by Elsevier, 2018.
Figure 2. Schematic diagram of an FIA system developed for the determination of the total dissolved nitrogen and phosphorus, with photometric detection and on-line UV and thermal digestion [88]. S—samples or standards; C—ultrapure water carrier stream; B—ammonium chloride buffer; OR—oxidation reagent; MRN—mixing reagent for nitrogen determination; MR—mixing reagent for phosphorus determination; AA—ascorbic acid; Cd—cadmium column; P1–P3—peristaltic pumps; IV—10-port injection valve; SV—8-position selector valve; DB—de-bubbler; L—light source; D—detector; FC—“U” shape flow cell; UV -UV-digester; HB—heated bath; RC—reaction coil; SL—sample loop; W—waste. The dashed line in IV represents the valve in position A, and the solid line represents the valve in position B. Reproduced with permission from Talanta, published by Elsevier, 2018.
Molecules 27 01410 g002
Determinations employing luminescence-based detection are much less frequently reported flow-injection methods than those with absorptive spectrophotometric detection, in spite of the wide availability of commercial instrumentation and what are usually much better limits of detection. Several examples of methods using fluorometric detection have been developed in the past decade; these are also listed in Table 2. A particularly low LOD was achieved, e.g., in the SIA determination of ammonia [93], Cd(II), and Pb(II) [94], and in the FIA system for the determination of phosphate [95] and sulfide [86]. In the method first mentioned, with a complex branched manifold (Figure 4), the pressure-assisted dual-headspace gas-liquid micro-extraction step, based on the lab-in-syringe concept, was employed on-line. The developed analytical procedure involved the in situ generation of ammonia vapor for its further dissolution and the formation of a fluorescence product. A sampling rate of 8 h−1 was reported for very sensitive determinations with an LOD of 50 ng L−1.
Figure 3. Schematic diagram of a multicomponent reversed-FIA system, with LED-based photometric multi-optical detection for the determination of transition metal ions in river water [89]. A—peristaltic pump; B—multisyringe pump (consisting of four syringes containing color-forming reagents; C—solenoid valves; D—mixing coil; E—multi-optical sensor and F—computer; S-sample; W-waste; (1),(2),(3),(4)—reagents for each analyte. Reproduced with permission from Talanta, published by Elsevier, 2017.
Figure 3. Schematic diagram of a multicomponent reversed-FIA system, with LED-based photometric multi-optical detection for the determination of transition metal ions in river water [89]. A—peristaltic pump; B—multisyringe pump (consisting of four syringes containing color-forming reagents; C—solenoid valves; D—mixing coil; E—multi-optical sensor and F—computer; S-sample; W-waste; (1),(2),(3),(4)—reagents for each analyte. Reproduced with permission from Talanta, published by Elsevier, 2017.
Molecules 27 01410 g003
Chemiluminescence detection, based on the principle of the permanganate method, was used for the determination of chemical oxygen demand in an FIA system [97], as well as for the very sensitive determination of Hg(II) in surface- and tap waters [98]. In the latter example, an immunoassay based on the use of resin beads, enzymatic amplification, and a novel monoclonal antibody with strong affinity recognition and high specificity for Hg(II), were used.
Figure 4. Schematic diagram of an SIA system developed for the determination of ammonia in waters with fluorimetric detection and a pressure-assisted dual-headspace gas-liquid micro-extraction module for membraneless gas separation [93]. SV—selection valve; V—valve; MS—magnetic stirrer; SP—syringe pump; MP—pump; TB—thermostat barrel; HC—holding coil; Ex—excitation; Em—emission; D—detector; W—waste. Reproduced with permission from Analytica Chimica Acta; published by Elsevier, 2018.
Figure 4. Schematic diagram of an SIA system developed for the determination of ammonia in waters with fluorimetric detection and a pressure-assisted dual-headspace gas-liquid micro-extraction module for membraneless gas separation [93]. SV—selection valve; V—valve; MS—magnetic stirrer; SP—syringe pump; MP—pump; TB—thermostat barrel; HC—holding coil; Ex—excitation; Em—emission; D—detector; W—waste. Reproduced with permission from Analytica Chimica Acta; published by Elsevier, 2018.
Molecules 27 01410 g004
Yet another application of molecular spectroscopy in the UV-Vis range is the use of thermal lens spectroscopy in an FIA system for the sensitive determination of Ag(I), based on the formation of colloidal silver nanoparticles [99]. This detection is based on the measurement of the temperature rise of an illuminated sample and the index of refraction modification as a result of nonradiative relaxation of the energy absorbed by the analyte from a laser.
Flow-injection systems with on-line sample-processing operations, employed mostly for analyte preconcentration or the elimination of interfering matrix components, are also combined with atomic spectroscopy instruments [120,121,122,123,124,125,126,127,128,129,130,131,132,133,134,135]. They are commonly used for determining numerous trace elements in water analysis (Table 3). In this case, the absorptive measurements are also most frequently exploited using different atomization techniques. Spectrometry with flame atomization was used in the FIA determination of Cd(II) [120], and Cr(III) [121], in both cases using on-line analyte enrichment with ion-imprinted polymers, with appropriate cavities produced during the polymerization for the selective uptake of the given ions. Determinations of Cd(II) were carried out, with preconcentration on magnetic ligand-functionalized nanoparticles [122], while for the determination of Cd, Co, and Cu, an SPE with ligand-functionalized silica gel was reported [123]. The original attempt at the use of flame-AAS detection is the indirect determination of sulfite, based on the on-line reduction of MnO2 and the measurement of quantitatively produced Mn(II) [124].
Table 3. The application of atomic spectroscopy detection methods in flow-injection systems for water analysis.
Table 3. The application of atomic spectroscopy detection methods in flow-injection systems for water analysis.
Analyte(s)Type of WaterDetection MethodType of Flow SystemEmployed Sample Processing/RemarksLOD, mg L−11st Author, Year of Publ.Reference
As, inorganicMineral, tap, river watersHG-AASSIASIA system equipped with on-line gas-liquid separator/reaction chamber and electrothermal quartz flow through atomizer50 ng L−1Anthemidis, 2014[127]
As, inorganicTap, ell and sea watersHG-AFSFIAFIA system equipped with microcolumn with immobilized tetrahydroborate and gas-liquid separator13 ng L−1 for As(III)Wang, 2014[133]
Bi, Hg, Sb, SnLake watersCVG-ET-
AAS
FIASequential determination of analytes using a high-resolution continuum source AASD and on-line SPE for separation and preconcentration Bi 1, Hg 170, Sb 9, Sn 180 ng L−1Guerrero, 2015[130]
Cd(II)Ground and wastewaterFlame AASFIACd(II)-imprinted polymer employed for on-line preconcentration0.00011Gawin, 2010[120]
Cd(II)Ground and sea watersCV-AASMultisyringe FIASystem with on-line gas-liquid separator5.8 ng L−1Silva, 2014[129]
Cd(II)Surface watersFlame AASFIAOn-line preconcentration using microcolumn with renewable magnetic ligand-functionalized nanoparticles0.002Rocha, 2021[122]
Cd(II), Co(II), Cu(II)Water standardFlame AASFIAOn-line preconcentration using microcolumn with ligand-functionalized silica gelN.a.Sivrikaya, 2018[123]
Co(II)Ground and tap waterET-AASLOV-SIAOn-line SPE extraction and preconcentration using functionalized magnetic nanoparticles6 ng L−1Wang, 2012[125]
Cr(III)Municipal wastewaterFlame AASFIACr(III) imprinted polymer for on-line separation and preconcentration0.0021Leśniewska, 2015[121]
Hg speciationLake waterAFSSIASpeciation of inorganic and organic mercury forms3 ng L−1Zhang, 2018[134]
Pb(II)Lake, sea, river watersHG-AASFIAOn-line SPE preconcentration ligand-functionalized sorbent82 ng L−1Trujillo, 2013[128]
Noble metalsLake, tap and sea watersICP-AESFIASimultaneous determination of Pt, Pd, Os, Ir, Rh, Ag and Au employing magnetic ligand-functionalized nanoparticles for on line SPEAg 0.03,
Pd 1.5,
Rh 100
mg L−1
Guerrero, 2017[131]
Sb speciationSurface, tap, ground waterHG-AFSMultisyringe FIAFIA system with cationic minicolumn for retaining trimethylantimony and gas-liquid separatorSb(III), (V) 30 TMSb 130
ng L−1
Portugal, 2015[135]
Se(IV), Se(VI)Tap waterHG-AASFIAAg nanoparticles employed for catalytic activation in Se(IV) determination0.0005Poonyaka 2017[126]
SulfiteMineral, tap, river watersFlame AASFIAOn-line reduction in a microcolumn packed with MnO20.08Zare-
Dorabei, 2018
[124]
TOCRiver watersAESCFAFlow system including miniaturized dielectric barrier carbon AES with on-line microwave-assisted oxidation0.01 as CHan, 2014[132]
Abbreviations used: AAS—atomic absorption spectrometry, AES—atomic emission spectrometry, CFA—continuous flow analysis, CV-AAS—cold vapor atomic absorption spectrometry, FIA—flow injection analysis, HG-AAS—hydride generation atomic absorption spectrometry, HG-AFS—hydride generation atomic fluorescence spectrometry, ICP—inductively coupled plasma, SIA –sequential injection analysis, SPE—solid-phase extraction, TOC—total organic carbon. N.a.—no information available.
Rather sporadically, one can also find examples of the hyphenation of an SIA system with an AAS instrument operating with electrothermal atomization (see, e.g., the schematic diagram of a manifold in Figure 5 [125]). With the use of on-line preconcentration on alumina-coated iron oxide nanoparticles that were functionalized with sodium dodecyl sulfate, cobalt (II) was determined in water samples in an SIA-LOV system connected on-line to an AAS spectrometer, with an impressive LOD of 6 ng L−1.
On the other hand, numerous flow-injection systems were reported that employed hydride generation. They were developed, for instance, for the speciation analysis of Se [126], and for very sensitive determinations of inorganic arsenic [127] and Pb(II) [128]. The cold vapor-AAS technique is used mostly in the trace determination of mercury, but in one work on a multisyringe FIA system, CV-AAS detection was employed in the determination of Cd(II) in ground- and seawater. A very low LOD of 5.8 ng L−1 was obtained without an additional preconcentration step, but with the addition of thiourea to increase the analytical signal [129]. A schematic diagram of the developed manifold is shown in Figure 6.
Figure 5. Schematic diagram of an SIA system incorporating a lab-on valve for the trace determination of cobalt in tap- and groundwater, with electrothermal-AAS detection [125]. SP, syringe pump; carrier, distilled water; HC, holding coil; eluent, ethanol; PP, peristaltic pump; ETAAS, electrothermal atomic absorption spectrometer. Reproduced with permission from Analytica Chimica Acta; published by Elsevier, 2012.
Figure 5. Schematic diagram of an SIA system incorporating a lab-on valve for the trace determination of cobalt in tap- and groundwater, with electrothermal-AAS detection [125]. SP, syringe pump; carrier, distilled water; HC, holding coil; eluent, ethanol; PP, peristaltic pump; ETAAS, electrothermal atomic absorption spectrometer. Reproduced with permission from Analytica Chimica Acta; published by Elsevier, 2012.
Molecules 27 01410 g005
Figure 6. Schematic diagram of the multisyrimge FIA system developed for trace determination of cadmium in ground and sea waters using cold vapor-AAS detection [129]. GL—gas–liquid separator; MP—micro-pump; QT—quartz tube; SC: sample coil; SV—solenoid valve, and TWC—three-way connector. Reproduced with permission from Journal of Atomic Absorption Spectrometry; published by the Royal Society of Chemistry, 2014.
Figure 6. Schematic diagram of the multisyrimge FIA system developed for trace determination of cadmium in ground and sea waters using cold vapor-AAS detection [129]. GL—gas–liquid separator; MP—micro-pump; QT—quartz tube; SC: sample coil; SV—solenoid valve, and TWC—three-way connector. Reproduced with permission from Journal of Atomic Absorption Spectrometry; published by the Royal Society of Chemistry, 2014.
Molecules 27 01410 g006
Then, in the determination of Bi and Pb with similar detectability, chemical vapor generation and an electrothermal-AAS with a continuum source of radiation were used, together with an on-line SPE on ligand-functionalized sorbent [130]. In the same setup configuration, the limits of detection for Hg and Sn were much poorer.
The chemical vapor generation of analytes prior to ICP-OAS detection allows the introduction of a larger sample to increase sensitivity, and, with the additional preconcentration of analytes on the magnetic nanoparticles functionalized with ligands, this was used in an FIA system for the simultaneous determination of several noble metals in natural waters [131]. Another mode of atomic spectroscopy detection has been employed very recently for the determination of total organic carbon (Figure 7). In this case, detection was based on the use of miniaturized dielectric barrier discharge carbon atomic emission, with the microwave-assisted oxidation of organic analytes [132]. The LOD for these determinations was evaluated at 10 μg L−1 C, and the developed system can also be used for in situ monitoring in routine applications.
Very low limits of detection were also reported for an FIA system with hydride generation-atomic fluorescence spectrometry (AFS) for the determination of inorganic arsenic [133]. Flow-injection systems with AFS detection were also reported for the determination of mercury [134] and antimony [135].
Figure 7. Schematic diagram of a flow system with on-line microwave-assisted oxidation and atomic spectrometry detection, based on the use of miniaturized dielectric barrier carbon developed for the determination of total organic carbon in river waters [132]. GLS—gas-liquid separator; CCD—charge-coupled device spectrometer. Reproduced with permission from Analytical Chemistry; published by the American Chemical Society, 2014.
Figure 7. Schematic diagram of a flow system with on-line microwave-assisted oxidation and atomic spectrometry detection, based on the use of miniaturized dielectric barrier carbon developed for the determination of total organic carbon in river waters [132]. GLS—gas-liquid separator; CCD—charge-coupled device spectrometer. Reproduced with permission from Analytical Chemistry; published by the American Chemical Society, 2014.
Molecules 27 01410 g007

3.2. Application of Electrochemical Detection

Although certain electroanalytical techniques find many common applications in routine laboratories for water analysis, such as potentiometry with membrane ion-selective electrodes (ISE) or voltammetry for trace metal analysis, the past decade did not bring too many examples of the application of those techniques in FIA systems for water analysis [136,137,138,139,140,141,142,143,144] (Table 4). The developed flow-injection systems dealt only with the determination of inorganic analytes, and the reported limits of detections very seldom competed with those obtained via the use of spectroscopic detection. An LOD level of ng L−1 for the determination of Hg(II) in river water was reached in an FIA system, with the use of anodic stripping voltammetry (ASV) with gold or platinum working electrodes, wherein metals were sputtered onto microporous membranes [136]. Then, detection with square wave voltammetry (SWV) was employed in the determination of phosphate in an SIA system, in which a glassy carbon electrode, modified with silanized carbon nanotubes and Au nanoparticles, was used as the working electrode [137]. The employed modification of the electrode surface enabled the researchers to obtain an increase in the catalytic response current of a molybdophosphate complex. A similar modification using the nanoparticles of a carbon screen-printed electrode was also reported in an FIA system with amperometric detection for the selective determination of As(III) in tap water and industrial wastewaters [138].
Table 4. Examples of the application of electroanalytical detection methods in flow-injection systems for water analysis.
Table 4. Examples of the application of electroanalytical detection methods in flow-injection systems for water analysis.
Analyte(s)Type of WaterDetection MethodType of Flow SystemEmployed Sample Processing/RemarksLOD, mg L−11st Author, Year of Publ.Referen- ce
Ammonium, sulfideCity canal waterConduc-tivityFIA-SIAHyphenation of FIA systems with membraneless vaporization units with SIA systems for contactless conductivity detectionNH4+ 2.0 S2−
1.9 μM
Alahmad, 2018[142]
Ammonium, TICTap, well seawatersConduc-tivityFIAA solenoid micropumping FIA system with a gas-diffusion module for CO2 and NH3 separation NH4+ 0.27
TIC 50 μM
Henriquez, 2014[141]
As(III)Tap and industrial wastewaterAmpero-
metry
FIAA carbon screen printed electrode modified with a gold nanoparticle/carbon nanofiber/chitosan0.0114Nellaiappan, 2018[138]
Ca, K, Na, chlorideCRM of different
waters
Potentio-
metry
FIA3D-printed multi-electrode flow through cell incorporating plasticized PVC-based ion-selective electrodesN.A.Dębosz, 2020[139]
Carbonate, NH3-N, salinityRiver, tap, seawaterConduct-
ivity
FIAAn FIA system incorporating membrane units for CO2 and NH3 separation and a dual-channel contactless conductivity detectorCO3 0.31 mM
N 1.85 μM Salinity 0.24%
Chaneam, 2018[140]
Cr(VI)Mineral waterCapacitanceFIADetection based on the electrostriction phenomenon, observed with a thiol modified gold electrode4.7 nMWieczorek, 2017[143]
Hg(II)River waterASVCFA and FIAAu or Pt working electrodes prepared by sputtering of metals onto microporous membranesCFA 40
FIA 50
ng L−1
Mizuguchi, 2013[136]
NitrateRiver, tap well waterPotentio- metrySIADetection with PVC-based nitrate ISE and sodium ISE used as reference electrode0.36 μMTossanaitada 2012[144]
PhosphateLake and pool waterSWVSIADetection with a glassy carbon electrode, modified with silanized carbon nanotubes and Au nanoparticles0.3Wu, 2021[137]
Abbreviations used: ASV—anodic stripping voltammetry, CRM—certified reference materials, FIA—flow injection analysis, ISE—ion-selective electrode, SIA—sequential injection analysis, SWV—square wave voltammetry, TIC—total inorganic carbon. N.A.—not available.
The application of potentiometric detection in an FIA system was reported for K+, Na+, Ca2+, and Cl ions commonly occurring in natural waters, for which membrane ISEs are available [139]. In the developed system, a 3D-printed flow cell was used, incorporating indicator electrodes with plasticized membranes; the measuring setup was employed successfully for the analysis of several certified reference materials of different water samples.
Detection with a non-selective conductivity technique in flow-injection systems requires the use of a suitable separation step or the isolation of the analyte. In three systems that can be found in the recent literature on water analysis, this detection method was employed for multicomponent determinations [140,141,142]. For instance, in the FIA system of configuration shown schematically in Figure 8, the simultaneous detection of carbonate, salinity, and ammonium nitrogen was achieved [140]. The measuring setup was equipped with a dual-channel contactless C4D conductivity detector and two gas diffusion modules, not only for the separation of the gaseous CO2 and NH3 but also for on-line dilution for salinity determination. Its successful application was described for the analysis of river, sea, and tap waters, with a sampling rate of 20 samples h−1, which can be considered very satisfactory. The developed system is able to tolerate most of the matrix constituents present in natural water samples. Another FIA system with conductivity detection has been developed for the simultaneous determination of ammonia and total inorganic carbon [141], as well as for ammonia and sulfide [142], with micromolar limits of detection.
Figure 8. Schematic diagram of an FIA system with conductivity detection, developed for the simultaneous determination of salinity (section A), carbonate (section B), and ammonium nitrogen (section C), using a dual-channel C4D detector [140]. P1 and P2—peristaltic pump; L1 and L2—sample loop at 200  µL (L1) and 1000  µL (L2); V1—10-port valve in “load” position, with an inset showing the “inject” position; V2—two-way right-angle switching valve; V—four-way diagonal switching valve; C1 and C2—water carrier; AC1, AC2, and AC3—water acceptor; BC1 and BC2—back-pressure coil (i.d.: 0.75  mm, length: 120  cm); DC—delay coil (i.d.: 1  mm, length: 320  cm); OD—on-line dilution unit; GD1 and GD2—gas diffusion unit; W1–W4—waste; CH1—channel 1 of C4D. CH2—channel 2 of C4D. Reproduced with permission from Talanta; published by Elsevier, 2018.
Figure 8. Schematic diagram of an FIA system with conductivity detection, developed for the simultaneous determination of salinity (section A), carbonate (section B), and ammonium nitrogen (section C), using a dual-channel C4D detector [140]. P1 and P2—peristaltic pump; L1 and L2—sample loop at 200  µL (L1) and 1000  µL (L2); V1—10-port valve in “load” position, with an inset showing the “inject” position; V2—two-way right-angle switching valve; V—four-way diagonal switching valve; C1 and C2—water carrier; AC1, AC2, and AC3—water acceptor; BC1 and BC2—back-pressure coil (i.d.: 0.75  mm, length: 120  cm); DC—delay coil (i.d.: 1  mm, length: 320  cm); OD—on-line dilution unit; GD1 and GD2—gas diffusion unit; W1–W4—waste; CH1—channel 1 of C4D. CH2—channel 2 of C4D. Reproduced with permission from Talanta; published by Elsevier, 2018.
Molecules 27 01410 g008
The very sensitive determination of Cr(VI) can be carried out in the FIA system with the use of detection exploiting the electrostriction phenomenon, based on the measurement of membrane capacitance [143]. This phenomenon refers to the effect of a slight change of shape, or mechanical deformation, under the application of an electric field. In the cited work, a gold electrode with a thiol self-assembled monolayer and additional functional groups was used, and the analyte interacted with that dielectric layer.

3.3. FIA Systems Using Other Detection Methods

In this group of flow-injection systems that were developed for water analysis and are listed in Table 5, the application of mass spectrometry predominates. These are not methods based on the interaction of electromagnetic radiation with matter; thus, according to the IUPAC recommendations for terms used in analytical spectroscopy [145], they are not included in Section 3.1. In only one case from the collected works [146,147,148,149,150,151,152,153,154,155,156,157], a flow system was designed for the determination of organic pollutants, namely, the SIA-LOV system, which was hyphenated to gas chromatography (GC) with MS detection [146]. In this complex instrumental configuration, shown schematically in Figure 9, the on-line sample processing was involved with the LOV-based dispersive liquid-liquid extraction carried out in the multisyringe SIA setup. For the determination of 16 priority polycyclic aromatic hydrocarbons in water, using 4 mL of sample volume, and good enrichment factors (27 to 38), the obtained LOD values at ng L−1 level were reported.
Table 5. Examples of the application of detection methods other than molecular and atomic spectroscopy or electrochemical detection in flow-injection systems for water analysis.
Table 5. Examples of the application of detection methods other than molecular and atomic spectroscopy or electrochemical detection in flow-injection systems for water analysis.
Analyte(s)Type of WaterDetection MethodType of Flow SystemEmployed Sample
Processing/Remarks
LOD, mg L−11st Author, Year of PublicationReference
Chloride, sulfateMineral watersPiezoelectricFIADetection using a flow-through acoustic sensor with a quartz crystal with a deposited ionophore layerCl 50 μM
SO4 42 μM
Venancio, 2018[155]
CODLake and river watersThermal sensorFIADetermination is based on measuring the heat generated when the sample passes through a column containing the periodic acid solution1.84Yao, 2014[156]
PAHsRiver, tap, rain watersMSMultisyringe SIA-LOV-GCSIA system with on-line dispersive liquid-liquid micro-extractionFrom 10 to 70 ng L−1Clavijo, 2014[146]
Pb, SrRiver and rainwaterICP-MSMultisyringe SIA-LOVOn-line preconcentration with
Sr-resin
Pb 4 Sr 12
ng L−1
Beltran, 2015[151]
ReSeawaterICP-MSFIAOn-line preconcentration on Dowex 1 × 8 anion-exchange resin0.1 ng L−1Zhu, 2017[150]
SrRiver waterICP-MSSIA-LOVApplication of commercial Sr-resin for off-line separation and preconcentration of radionuclide 90Sr in SIA system2.9 ng L−1
14.5 Bq L−1
Kołacińska, 2017[147]
TcMineral, tap surfaceICP-MSMultisyringe SIA-LOVApplication of commercial resin TEVA for on-line separation and preconcentration of radionuclide 99Tc in SIA system49 pg L−1Rodriguez, 2015[149]
TcRiver and wastewaterICP-MSSIA-LOVApplication of commercial resins Dowex 1 and TEVA for off-line separation and preconcentration of radionuclide 99Tc in SIA system9.55 pg L−1 6.0mBq L−1Kołacińska, 2018[148]
TcRiver, sea, groundwaterICP-MSFIAApplication of commercial resin TK201 for on-line separation and preconcentration of radionuclide 99Tc 9.3 pg L−1 5.9mBq L−1Matsueda, 2021[157]
Trace metalsSeawaterICP-MSFIA-ICFIA system coupled with ion chromatography with ICP-MS detection and on-line preconcen-tration on chelating resin for determination of Al, Cd, Co, Cr, Fe, Mn, Ni, Pb, Ti, V, ZnFrom ~0.03 (Cd) to ~100 (Al) ng L−1Ho, 2010[154]
Trace metalsOpen-ocean waterICP-MSFIAOn-line preconcentration on minicolumn packed with a chelating sorbent for determination of Cd, Co, Cu, Ni, Pb, Zn1.5 (Pb) to
71 (Zn) pM
O’Sullivan, 2013[152]
Trace metalsSeawaterICP-MSFIAApplication of commercial resins for on-line separation and preconcentration of Co, Fe, Pb and V in the flow systemFrom 0.0058 (Co) to 0.34 (Fe) nM Clough, 2015[153]
Abbreviations used: COD—chemical oxygen demand, FIA—flow injection analysis, GC—gas chromatography, IC—ion chromatography, ICP-MS—inductively coupled plasma—mass spectrometry, SIA—sequential injection analysis, LOV—lab-on-valve, MS—mass spectrometry, PAHs—polycyclic aromatic hydrocarbons.
Figure 9. Schematic diagram of the LOV-SIA system developed for on-line sample processing prior to the gas chromatography setup, for the determination of polycyclic hydrocarbons in river, tap, and rain waters [146]. MPV—multiposition valve; MSM—multisyringe module; MIV—micro-injection valve; RD—reagent dispersant (acetonitrile); RE—reagent extractive (trichloroethylene); HC 1–2—holding coils; SV—solenoid valve; S 1–2—syringe pumps; V 1–2—valves; EC—extraction chamber. Reproduced with permission from Analytical Methods; published by the Royal Society of Chemistry, 2014.
Figure 9. Schematic diagram of the LOV-SIA system developed for on-line sample processing prior to the gas chromatography setup, for the determination of polycyclic hydrocarbons in river, tap, and rain waters [146]. MPV—multiposition valve; MSM—multisyringe module; MIV—micro-injection valve; RD—reagent dispersant (acetonitrile); RE—reagent extractive (trichloroethylene); HC 1–2—holding coils; SV—solenoid valve; S 1–2—syringe pumps; V 1–2—valves; EC—extraction chamber. Reproduced with permission from Analytical Methods; published by the Royal Society of Chemistry, 2014.
Molecules 27 01410 g009
Much more frequently, FIA and SIA systems are being developed for elemental analysis with the use of very sensitive, inductively coupled plasma-mass spectrometry (ICP-MS). The main purpose of combining a very selective ICP-MS detection with flow-injection systems is usually the possibility of the additional elimination of isobaric interference in MS detection, elimination of the interfering effects of sample matrices, and also an additional preconcentration of analytes. These goals can be achieved in an off-line configuration without instrumental hyphenation, which was reported, for instance, for the determination of 90Sr [147], and 99Tc [148] radionuclides, especially in the second case with SPE preconcentration, where an LOD was achieved that was as low as 9.55 pg L−1. For the determination of 99Tc, on-line setups were also developed, with an on-line SPE using different commercial sorbents (e.g., [149]). Similar flow systems were described for the trace determination of rhenium in seawater [150], of Pb and Sr in river and rain waters [151], and also for the simultaneous determination of many trace elements in seawater [152,153]. In all those systems, the analytes were preconcentrated on-line on solid sorbents. For a similar purpose, an FIA setup with on-line enrichment can be hooked up to a high-performance ion-chromatograph and ICP-MS spectrometer, which allows sample processing for the simultaneous determination of 12 analytes in seawater, taking about 10 min [154].
Among other detection techniques employed in FIA systems for water analysis, one can find the application of piezoelectric detection [155], as well as thermal sensing [156]. In the first study, the detection method was based on the use of a quartz crystal oscillator attached to an electrode covered with a layer containing the appropriate ionophore for the detection of chloride and sulfate, with micromolar LODs. In the second study, a determination of the chemical oxygen demand in water was based on measuring the heat generated by the reaction of organic analytes with periodic acid, packed into a mini-column.

3.4. Microfluidic Flow-Injection Systems Dedicated to Water Analysis

One particular instrumental novelty in the field of water analysis in recent years seems to be an upsurge of interest in the construction of microfluidics, mostly used so far for capillary electrophoretic determinations and for clinical and biochemical determinations. They can be used for flow-injection measurements on a much smaller scale, in terms of the sample volume needed and the consumption of reagents. It must be realized, however, that in the present stage of their development, the microfluidic chip usually replaces the hydraulic part of the flow system, eventually including some integrated miniaturized flow-through detection cells or sample processing module. However, it still needs the typical periphery devices in a common macroscale (sampling units, pumps, detection transducers, control units, etc.). This means that, in fact, it is very far from a realistic micro total analytical system. Obviously, the examples collected in Table 6 [158,159,160,161,162,163,164,165] do not include microfluidic devices involving, e.g., capillary electrophoretic separations, but they do include the reported miniaturized flow-injection setups.
Similar to typical flow-injection systems, in the case of microfluidics, spectroscopic detections are also commonly used. Among the listed works, in only two cases, electrochemical detection was conducted. The potentiometric setup was used with a chip incorporating an ammonium membrane ion-selective electrode, a gas diffusion unit, and a screen-printed reference electrode [158]. A very sensitive determination of the herbicide residues in groundwater was reported, using an amperometric immunoassay carried out with a microfluidic chip and a screen-printed carbon working electrode [159].
Table 6. Examples of application of microfluidics in the flow-injection systems for water analysis.
Table 6. Examples of application of microfluidics in the flow-injection systems for water analysis.
Analyte(s)Type of WaterDetection MethodEmployed Sample
Processing/Remarks
LOD, mg L−11st Author, Year of Publ.Reference
AmmoniaLake waterUV/VIS absorptionFIA system with microfabricated electroosmotic pump coupled to a gas-diffusion microchip0.10Zhu, 2015[160]
AmmoniumTreatment plantPotentio-
metric
On-line integration of gas-diffusion step with a flow cell with ammonium ISE and screen-printed reference electrode0.07Calvo-Lopez, 2015[158]
BacteriaLake, wastewaterFluorescenceEnrichment and determination of Escherichia coli using an antibody-modified microfluidic chip, and benchtop real-time PCR6 clsDharmasiri, 2010[166]
Cr(VI)River waterUV/VIS absorptionMonolithically integrated system based on green tape ceramic technology, involving an LED-based detector0.05Alves Segundo 2011[161]
Cr(III), Cr(VI)River and tap watersFluorescenceApplication of the magnetic-carbon nanotube nanocomposite for off-line solid-phase extraction preconcentration94 pMPeng, 2017[165]
HerbicideGround-
water
Ampero-
metric
Immunoassay of 2,6-dichlorobenzamide residues in a microfluidic setup with a screen-printed electrode for amperometric detection20 ng L−1Uthuppu, 2015[159]
Iodine, total inorganicGround-
water
FluorescenceDetermination based on the use of the Sandell–Kolthoff reaction and fluorimetric detection of Cr(III)0.0077Inpota, 2018[164]
NitriteRiver waterUV/VIS absorptionMicrofluidic system with fluid manipulation using a biomimetic photoresponsive ionogel microvalve0.034Czugala, 2013[162]
NitrateTap, sea and surface watersUV/VIS absorptionIntegrated nitrate analyzer with microfluidic and LED light source with a photodiode detector0.70Cogan, 2015[163]
Cls—colony-forming units.
Numerous interesting innovations have been reported in the design of microfluidics with spectrophotometric detection. For the determination of ammonia, a gas-diffusion microchip was coupled with a microfabricated electroosmotic pump, to perform FIA measurements [160]. The microfluidic device was produced with the use of green tape ceramic technology for the determination of Cr(VI) in river water, with an LOD of 50 μg L−1 [161]. Microfluidic chips were also designed for the determination of nitrite [162] and nitrate [163] in water. In the first study, the manipulation of solutions was carried out by the use of a biomimetic photoresponsive ionogel microvalve, controlled by a white light-emitting diode [162].
Several examples in the literature over the last decade can also be found for the application of fluorimetric detection in flow-injection microfluidics. A very sensitive (the LOD 7.7 μg L−1) determination of total inorganic iodine was reported, using a microfluidic chip, based on the reaction between Ce(IV) and As(III), catalyzed by iodide [164]. The determination of Cr(III) and Cr(VI) with on-line fluorescent derivatization was conducted in a dedicated microfluidic chip, following off-line solid-phase extraction using magnetic carbon nanotubes [165]. A microfluidic chip with immobilized antibodies was employed prior to performing the determination of bacteria via an off-line benchtop real-time quantitative polymerase chain reaction.

4. Portable Flow-Injection Instrumentation for Water Analysis

Water monitoring requires not only the development of efficient analytical methods for use in specialized laboratories but also portable instrumentation with remote control for field applications. Such an apparatus is often needed for carrying out process analysis at technological installations (e.g., for the continuous monitoring of processing in wastewater treatment plants), as well as for in-field applications related to environmental monitoring, such as the control of pollution in river waters, or for monitoring natural processes occurring in seas and oceans. In the design of such portable instruments, since the early 1980s, a flow-injection methodology has been exploited (see, e.g., review [167]). The construction of such devices is based on the suitable choice and integration of a propulsion system, sampling and injection tools, and selection of the most satisfactory detection technique, depending on the target analyte(s) and the required selectivity. Such a system should be computer-controlled (see, for instance, [168]), although in many of the works cited in this review that include the use of commercial equipment, the flow-injection systems that are employed are fully computerized.
The analytical literature on the development of flow-injection methods and instrumentation from the 1980s provides numerous examples of such portable systems constructed for water analysis. Several such systems have also been developed in the past decade. The FIA system for monitoring total organic carbon has already been mentioned above [132]. It is based on the use of carbon atomic emission spectrometry for detection, as well as sample processing using the on-line microwave-assisted oxidation of analytes by persulfate. The obtained LOD of 10 μg L−1 is satisfactory for various environmental applications.
In the past several decades, numerous measuring systems have already been designed for submersible measurement, as well as setups for shipboard measurement on research vessels, with underway-sampling probes. In recent years, several such units have been developed by a research group from the State Key Laboratory of Marine Environmental Science in Xiamen University, Xiamen, China [169,170,171,172,173]. These were developed using absorptive spectrophotometric detection to identify trace dissolved manganese in estuarine and coastal waters [169], for the determination of dissolved sulfide [170], and for the redox speciation of iron, with sub-nanomolar LODs [171]. The configuration of the manifold of the latter study is shown in Figure 10. This reverse-FIA setup employs a liquid waveguide capillary flow cell, and the detection of Fe(II) is based on a reaction with ferrozine, while Fe(III) was reduced by ascorbate. In two other systems that have been mentioned [169,170], reverse-FIA systems were also used, with reported nanomolar limits of detection. The ultra-trace on-board determination of ammonium was based on the use of the fluorimetric detection of the reaction product of ammonia with o-phthalaldehyde [172,173]. In a flow-batch system via the SPE preconcentration of the fluorescent product, an LOD of 1.2 nM was achieved [172]. The developed systems were successfully used for the determination of ammonia in seawater [172], in estuarine, and in coastal waters [173].
Figure 10. Manifold of the reversed FIA setup for a real-time iron speciation shipboard-use system [171]. The dashed box indicates the towed fish-based underway-sampling system, through which unit surface seawater was pumped continuously onboard, with pump A, at a flow rate of about 1 L min−1. MV, IV, and SV are the multiposition valve, six-port injection valve, and three-way solenoid switching valve; KR is the knitted reaction coil; FZ and AA are ferrozine and ascorbic acid. Reproduced with permission from Environmental Science and Technology; published by the American Chemical Society, 2015.
Figure 10. Manifold of the reversed FIA setup for a real-time iron speciation shipboard-use system [171]. The dashed box indicates the towed fish-based underway-sampling system, through which unit surface seawater was pumped continuously onboard, with pump A, at a flow rate of about 1 L min−1. MV, IV, and SV are the multiposition valve, six-port injection valve, and three-way solenoid switching valve; KR is the knitted reaction coil; FZ and AA are ferrozine and ascorbic acid. Reproduced with permission from Environmental Science and Technology; published by the American Chemical Society, 2015.
Molecules 27 01410 g010
One very challenging attempt was also reported on the design of a potentiometric microsystem for the determination of nitrate and potassium when monitoring a water recycling process, oriented toward its use in manned space missions [174]. It integrates a ceramic microfluidic chip, with two solid-contact polymeric ion-selective electrodes and a screen-printed reference electrode. LODs of below 1 mg L−1 were reported for both analytes, which is satisfactory for the given application. A further perspective in the design of that type of setup involves exchanging the conventional pumps and valves for miniaturized, fully integrated devices. A similar system was also reported for the determination of ammonium [158].

5. Conclusions

Flow-injection analytical methods, which have been developed for almost half a century, find a wide range of applications in water analysis, usually due to the large sampling rate, the possibility of the application of different detection techniques, and better precision compared to similar manual procedures. The on-line conducting of various sample processing operations is especially attractive and, in flow-injection systems, does not need to be carried out until a steady state of equilibrium is reached, without any loss of precision. These features mean that with an appropriate construction of the measuring system, they can also be used in remote and portable configurations, and they may also serve as early warning systems in processing or environmental applications [68].
In a further improvement of those methods. Novel sorption materials are being introduced, along with new methodologies for undertaking various sample-processing operations. The primary directions of further development seem to be the miniaturization of measuring setups, with the integration of various modules, the introduction of elements of real automation, especially those involving the optimization of measuring conditions without human effort, and the further popularization of those methods as a convenient tool for sampling and sample processing, prior to high-performance separations using chromatographic and electrophoretic methods.

Author Contributions

Conceptualization, M.T.; validation, M.T. and M.P.; writing—original draft preparation, M.T. and M.P.; writing—review and editing, M.T.; supervision, M.T. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not Applicable.

Informed Consent Statement

Not Applicable.

Data Availability Statement

Not Applicable.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Grant, S.B.; Saphores, J.-D.; Feldman, D.L.; Hamilton, A.J.; Fletcher, T.D.; Cook, P.L.; Stewardson, M.; Sanders, B.F.; Levin, L.A.; Ambrose, R.F.; et al. Taking the “waste” out of “wastewater” for human water security and ecosystem sustainability. Science 2012, 337, 681–686. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  2. Wee, S.Y.; Aris, A.Z. Occurrence and public-perceived risk of endocrine disrupting compounds in drinking water. Clean Water 2019, 2, 4. [Google Scholar] [CrossRef] [Green Version]
  3. Petrie, B.; Barden, R.; Kasprzyk-Houdern, B. A review on emerging contaminants in wastewaters and the environment: Current knowledge, understudied areas and recommendations for future monitoring. Water Res. 2015, 72, 3–27. [Google Scholar] [CrossRef]
  4. Ebele, A.J.; Abdallah, M.A.; Harrad, S. Pharmaceuticals and personal care products (PPCPs) in the freshwater aquatic environment. Emerg. Contam. 2017, 3, 1–16. [Google Scholar] [CrossRef]
  5. Mompelat, S.; Le Bot, B.; Thomas, O. Occurrence and fate of pharmaceutical products and by products, from resource to drinking water. Environ. Intern. 2009, 35, 803–814. [Google Scholar] [CrossRef] [PubMed]
  6. Pedrouzo, M.; Borrull, F.; Pocurull, E.; Marce, R.M. Presence of pharmaceuticals and hormones in waters from sewage of treatment plants. Water Air Soil Pollut. 2011, 217, 267–281. [Google Scholar] [CrossRef]
  7. Xiao, F. Emerging poly- and perfluoroalkyl substances in the aquatic environment: A review of current literature. Water Res. 2017, 124, 482–495. [Google Scholar] [CrossRef]
  8. Eschauzier, C.; Raat, K.J.; Struyfzand, P.J.; DeVoogt, P. Perfluorinated alkylated acids in groundwater and drinking water. Sci. Total Environ. 2013, 458–460, 477–485. [Google Scholar] [CrossRef]
  9. Hu, X.C.; Andrews, D.Q.; Lindstrom, A.B.; Bruton, T.A.; Schaider, L.A.; Grandjean, P.; Lohmann, R.; Carignan, C.C.; Blum, A.; Balan, S.A.; et al. Detection of poly- and perfluoroalkyl substances (PFASs) in U.S. drinking water linked to industrial sites, military fire training areas, and wastewater treatment plants. Environ. Sci. Technol. Lett. 2016, 3, 344–350. [Google Scholar] [CrossRef]
  10. Issac, M.N.; Kandassubramanian, R. Effect of microplastics in water and aquatic systems. Environ. Sci. Pollut. Res. 2021, 28, 19544–19562. [Google Scholar] [CrossRef]
  11. Weinberg, H.; Galyean, A.; Leopold, M. Evaluating engineered nanoparticles in natural waters. Trends Anal. Chem. 2011, 30, 72–83. [Google Scholar] [CrossRef]
  12. Lowry, G.V.; Gregory, K.B.; Apte, S.C.; Lead, J.R. Transformations of nanomaterials in the environment. Environ. Sci. Technol. 2012, 46, 6893–6899. [Google Scholar] [CrossRef] [PubMed]
  13. Fabrega, J.; Luoma, S.N.; Tyler, C.R.; Galloway, T.S.; Lead, J.R. Silver nanoparticles: Behaviour and effects in the aquatic environment. Environ. Int. 2011, 37, 517–531. [Google Scholar] [CrossRef] [PubMed]
  14. Amde, M.; Liu, J.; Tan, Z.; Bekana, D. Transformation and bioavailability of metal oxide nanoparticles in aquatic and terrestrial environments. A review. Environ. Int. 2017, 230, 250–267. [Google Scholar] [CrossRef]
  15. Richardson, S.D.; Kimura, S.Y. Water analysis: Emerging contaminants and current issues. Anal. Chem. 2020, 92, 473–505. [Google Scholar] [CrossRef]
  16. Schmidt, T. Recent trends in water analysis triggering future monitoring of organic micropollutants. Anal. Bioanal. Chem. 2018, 410, 3933–3941. [Google Scholar] [CrossRef] [Green Version]
  17. Van Leerdam, J.A.; Vervoort, J.; Stroomberg, G.; de Voogt, P. Identification of unkwon microcontaminants in Dutch river water by liquid chromatography-high resolution mass spectrometry and nuclear magnetic resonance spectroscopy. Environ. Sci. Technol. 2014, 48, 12791–12799. [Google Scholar] [CrossRef]
  18. Lorezno, M.; Campo, J.; Pico, Y. Analytical challenges to determine persistent organic pollutants in aquatic ecosystems. Trends Anal. Chem. 2018, 103, 137–155. [Google Scholar]
  19. Loh, S.H.; Yahaya, N.; Ishak, S.M.; Khalik, W.M.A.W.M.; Abdullah, N.S.C.; Aboul-Enein, H.Y.; Ong, M.C. Recent trends in adsorbent-based microextraction of micropollutants in environmental waters. Curr. Pollut. Rep. 2021, 7, 89–103. [Google Scholar] [CrossRef]
  20. Kruse, P. Review on water quality sensors. J. Phys. D Appl. Phys. 2018, 51, 203002. [Google Scholar] [CrossRef] [Green Version]
  21. Nigam, A.N.; Sharma, N.; Tripathy, S.; Kumar, M. Development of semiconductor based heavy metal ion sensors for water analysis: A review. Sens. Actuat. A Phys. 2021, 330, 112879. [Google Scholar] [CrossRef]
  22. Thavarajah, W.; Verosloff, M.S.; Jung, J.K.; Alan, K.K.; Miller, J.D.; Jewett, M.C.; Young, S.L.; Lucks, J.B. A primer on emerging field-deployable synthetic biology tools for global water quality monitoring. Clean Water 2020, 3, 18. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  23. Simonet, B.M.; Valcarcel, M. Monitoring nanoparticles in the environment. Anal. Bioanal. Chem. 2009, 393, 17–21. [Google Scholar] [CrossRef] [PubMed]
  24. Da Silva, B.F.; Perez, S.; Gardinalli, P.; Singhal, R.K.; Mozeto, A.A.; Barcelo, D. Analytical chemistry of metallic nanoparticles in natural waters. Trends Anal. Chem. 2011, 30, 528–540. [Google Scholar] [CrossRef]
  25. Wu, M.; Yang, C.; Du, C.; Liu, H. Microplastics in waters and soils: Occurrence, analytical methods and ecotoxicological effects. Ecotoxicol. Environ. Saf. 2020, 202, 110910. [Google Scholar] [CrossRef]
  26. Kelly, E.R.; Cronk, R.; Kumpel, E.; Howard, G.; Bartram, J. How we assess water safety: A critical review of sanitary inspection and water quality analysis. Sci. Total Environ. 2020, 718, 137237. [Google Scholar] [CrossRef]
  27. Pena-Pereira, F.; Bendicho, C.; Pavlovic, D.M.; Martin-Esteban, A.; Diaz-Alvarez, M.; Pan, Y.; Cooper, J.; Yang, Z.; Safarik, I.; Pospiskova, K.; et al. Miniaturized analytical methods for determination of environmental contaminants of emerging concern—A review. Anal. Chem. Acta 2021, 1158, 238108. [Google Scholar] [CrossRef]
  28. Armenta, S.; Garriguers, S.; de la Guardia, M. Green analytical chemistry. Trends Anal. Chem. 2008, 27, 497–511. [Google Scholar] [CrossRef]
  29. Bezerra, M.A.; Santelli, R.E.; Lems, V.A.; dos Santos Alves, J.P.; Breaz, B.F.; Santos, L.B. Strategies to make methods based on flow injection analysis greener. Clean–Soil Air Water 2020, 43, 2000007. [Google Scholar] [CrossRef]
  30. Trojanowicz, M. Flow chemistry in contemporary chemical sciences: A real variety of its applications. Molecules 2020, 25, 1434. [Google Scholar] [CrossRef] [Green Version]
  31. Karnatz, F.A.; Whitmore, F.C. Dehydration of diethylcarbinol. J. Am. Chem. Soc. 1932, 54, 3461. [Google Scholar] [CrossRef]
  32. Reed, L. Notes on capillary separation of substances in solution. Proc. Chem. Soc. 1893, 9, 123–126. [Google Scholar]
  33. Cwiet, M.S. О нoвoй категoрии адсoрбциoнных явлений и o применении их к биo-химическoму анализу. Proc. Warsaw Soc. Nat. Sci. Biol. Sec. 1903, 14, 20–39. [Google Scholar]
  34. Tswett, M. Adsorptionsanalyse und chromatographishe Methode. Anwendung auf die Chemie des Chlorophylls. Ber. Dtsch. Botan. Ges. 1906, 24, 385–392. [Google Scholar]
  35. Skeggs, L.T., Jr. An automatic method for colorimetric analysis. Clin. Chem. 1956, 2, 241, Erratum in Am. J. Clin. Path. 1957, 28, 311–322. [Google Scholar] [CrossRef] [PubMed]
  36. Plutschack, M.B.; Pieber, B.; Gilmotre, K.; Seeberger, P.H. The Hitchhiker’s guide to flow chemistry. Chem. Rev. 2017, 117, 11796–11893. [Google Scholar] [CrossRef]
  37. Furman, W.B. Continuous Flow Analysis. Theory and Practice; Marcel Dekker: New York, NY, USA, 1976. [Google Scholar]
  38. Pungor, E.; Feher, Z.; Nagy, G. Application of silicone rubber-based graphite electrodes for continuous flow measurements. Part I. general consideration. Anal. Chim. Acta 1970, 51, 431–436. [Google Scholar] [CrossRef]
  39. Nagy, G.; Feher, Z.; Pungor, E. Application of silicone rubber-based graphite electrodes for continuous flow measurements: Part II. Voltammetric study of active substances injected into electrolyte streams. Anal. Chim. Acta 1970, 52, 47–54. [Google Scholar] [CrossRef]
  40. Stewart, K.K.; Beeher, G.R.; Hare, P.E. Automated high-speed analyses of discrete samples—Use of nonsegmented, continuous-flow systems. Fed. Proc. 1974, 33, 1439. [Google Scholar]
  41. Stewart, K.K.; Beecher, G.R.; Hare, P.E. Rapid analysis of discrete samples: The use of nonsegmented, continuous flow. Anal. Biochem. 1976, 70, 167–173. [Google Scholar] [CrossRef]
  42. Ruzicka, J.; Hansen, E.H. Flow injection analysis. 1. New concept of fast continuous-flow analysis. Anal. Chim. Acta 1975, 78, 145–157. [Google Scholar]
  43. Ruzicka, J.; Marshall, G.D. Sequential injection: A new concept for chemical sensors, process analysis and laboratory assays. Anal. Chim. Acta 1990, 237, 329–343. [Google Scholar] [CrossRef]
  44. Vakh, C.; Falkova, M.; Timofeeva, I.; Moskvin, A.; Moskvin, L.; Bulatov, A. Flow analysis: A novel approach for classification. Crit. Rev. Anal. Chem. 2016, 46, 374–388. [Google Scholar] [CrossRef] [PubMed]
  45. Trojanowicz, M.; Kołacińska, K. Recent advances in flow injection analysis. Analyst 2016, 141, 2085–2139. [Google Scholar] [CrossRef]
  46. Cerda, V.; Ferrer, L.; Kamogawa, L.A.; de Souza, C.T.; Ferreira, S.L.C. Multisyringe flow injection analysis in spectroanalytical techniques—A review. Trends Anal. Chem. 2018, 98, 1–18. [Google Scholar] [CrossRef]
  47. Saurina, J. Flow-injection analysis for multi-component determinations of drugs based on chemometric approaches. Trends Anal. Chem. 2010, 29, 1027–1037. [Google Scholar] [CrossRef]
  48. Reis, B.F.; Gine, M.F.; Zagatto, E.A.G.; Lima, J.L.F.C.; Lapa, R.A. Multicommutation in flow analysis. Part 1. Binary sampling concepts, instrumentation and spectrophotometric determination of iron in plant digests. Anal. Chim. Acta 1994, 293, 129–138. [Google Scholar] [CrossRef]
  49. Mansour, F.R.; Danielson, N.D. Reverse flow-injection analysis. Trends Anal. Chen. 2012, 40, 1–14. [Google Scholar] [CrossRef]
  50. Albertus, F.; Horstkotte, B.; Cladera, A.; Cerda, V. A robust multisyringe system for process flow analysis. Part I. Online dilution and single point titration of protolytes. Analyst 1999, 124, 1373–1381. [Google Scholar] [CrossRef]
  51. Zagatto, E.A.G.; Rocha, F.R.P. Large-scale flow analysis: From repetitive assays to expert analyzers. Talanta 2021, 233, 122479. [Google Scholar] [CrossRef]
  52. Ruzicka, J.; Hansen, E.H. Integrated microconduits for flow-injection analysis. Anal. Chim. Acta 1984, 161, 1–25. [Google Scholar] [CrossRef]
  53. Ruzicka, J. Lab-on-valve: Universal microflow analyzer based on sequential and bead injection. Analyst 2000, 125, 1053–1060. [Google Scholar] [CrossRef]
  54. Hadd, A.G.; Jacobson, S.C.; Ramsey, J.M. Microfluidic assays of acetylcholinesterase inhibitors. Anal. Chem. 1999, 71, 5206–5212. [Google Scholar] [CrossRef]
  55. Badilescu, S.; Packirisany, M. Microfluidic-nano-integration for synthesis and sensing. Polymers 2012, 4, 1278–1310. [Google Scholar] [CrossRef] [Green Version]
  56. He, Y.; Wu, Y.; Fu, J.; Wu, W. Fabrication of paper-based microfluidic analysis devices: A review. RSC Adv. 2015, 5, 78109–78127. [Google Scholar] [CrossRef]
  57. Musijowski, J.; Koc, M.; Donten, M.; Szostek, B.; Trojanowicz, M. Determination of total organic fluorine (TOF) in environmental samples using flow-injection and chromatographic methods. Anal. Methods 2011, 3, 1039–1045. [Google Scholar]
  58. Tsizin, G.I.; Statkus, M.A.; Zolotov, Y.A. Preconcentration of organic substances on low-polar adsorbents in the flow systems of analysis. J. Anal. Chem. 2018, 73, 1030–1042. [Google Scholar] [CrossRef]
  59. Trojanowicz, M. Flow-injection analysis as a tool for determination of pharmaceutical residues in aqueous environment. Talanta 2012, 96, 3–10. [Google Scholar] [CrossRef]
  60. Christodouleas, D.; Fotakis, C.; Economou, A.; Papadopoulos, K.; Timotheou-Potamia, M.; Calokerinos, A. Flow-based methods with chemiluminescence detection for food and environmental analysis: A review. Anal. Lett. 2011, 44, 176–215. [Google Scholar] [CrossRef]
  61. Sraj, L.O.; Almeida, M.I.G.C.; Swearer, S.E.; Kolev, S.D.; McKelvie, I.D. Analytical challenges and advantages of using flow-based methodologies for ammonia determination in estuarine and marine waters. Trends Anal. Chem. 2014, 59, 83–92. [Google Scholar] [CrossRef]
  62. Teshima, N.; Murakami, H.; Sakai, T. Development of testing methods for water quality by flow analysis. Bunseki Kagaku 2020, 69, 257–269. [Google Scholar] [CrossRef]
  63. Worsfold, P.J.; Clough, R.; Lohan, M.C.; Monbet, P.; Ellis, P.S.; Quetel, C.R.; Floor, G.H.; McKelvie, I.D. Flow injection analysis as a tool for enhancing oceanographic nutrient measurements—A review. Anal. Chim. Acta 2013, 803, 15–40. [Google Scholar] [CrossRef] [PubMed]
  64. Fajardo, Y.; Avivar, J.; Ferrer, L.; Gomez, E.; Casas, M.; Cerda, V. Automation of radiochemical analysis by applying flow techniques to environmental samples. Trends Anal. Chem. 2010, 29, 1399–1408. [Google Scholar] [CrossRef]
  65. Kołacińska, K.; Trojanowicz, M. Application of flow analysis in determination of selected radionuclides. Talanta 2014, 125, 131–145. [Google Scholar] [CrossRef]
  66. Qiao, J. Dynamic flow approached for automated radiochemical analysis in environ-mental, nuclear and medical applications. Molecules 2020, 25, 1462. [Google Scholar] [CrossRef] [Green Version]
  67. Maya, F.; Estela, J.M.; Cerda, V. Flow analysis techniques as effective tools for the improved environmental analysis of organic compounds expressed as total indices. Talanta 2010, 81, 1–8. [Google Scholar] [CrossRef]
  68. Storey, M.V.; van der Gaag, B.; Burns, B.P. Advances in on-line drinking water quality monitoring and early warning systems. Water Res. 2010, 45, 741–757. [Google Scholar] [CrossRef]
  69. Jokerst, J.C.; Emory, J.M.; Henry, C.S. Advances in microfluidics for environmental analysis. Analyst 2012, 137, 24–34. [Google Scholar] [CrossRef]
  70. Bridle, H.; Miller, B.; Desmulliez, M.P.Y. Application of microfluidics in waterborne pathogen monitoring: A review. Water Res. 2014, 55, 256–271. [Google Scholar] [CrossRef]
  71. Almeida, M.I.G.S.; Jayawardane, B.M.; Kolev, S.D.; McKelvie, I.D. Developments of microfluidic paper-based analytical devices (μPADs) for water analysis: A review. Talanta 2018, 177, 176–190. [Google Scholar] [CrossRef]
  72. Zhu, Y.; Yuan, D.; Huang, Y.; Ma, J.; Feng, S.; Lin, K. A modified method for on-line determination of trace ammonium in seawater with a long-path liquid waveguide capillary cell and spectrophotometric detection. Marine Chem. 2014, 162, 114–121. [Google Scholar] [CrossRef]
  73. Feng, S.; Zhang, M.; Huang, Y.; Yuan, D.; Zhu, Y. Simultaneous determination of nanomolar nitrite and nitrate in seawater using reverse flow injection analysis coupled with a long path length liquid waveguide capillary cell. Talanta 2013, 117, 456–462. [Google Scholar] [CrossRef] [PubMed]
  74. Boonjob, W.; Yu, Y.; Miro, M.; Segundo, M.A.; Wang, J.; Cerda, V. Online hyphenation of multimodal microsolid phase extraction involving renewable molecularly imprinted and reversed-phase sorbents to liquid chromatography for automatic multiresidue assays. Anal. Chem. 2010, 82, 3052–3060. [Google Scholar] [CrossRef]
  75. Calderilla, C.; Maya, F.; Cerda, V.; Leal, L.O. 3D printed device for the automated preconcentration and determination of chromium (VI). Talanta 2018, 184, 15–22. [Google Scholar] [CrossRef]
  76. Rodriguez-Maese, R.; Ferrer, R.; Cerda, V.; Lral, L.O. Fully automatic system for lead monitoring in water. Microchem. J. 2020, 154, 104550. [Google Scholar] [CrossRef]
  77. Nagul, E.A.; Fontas, C.; McKelvie, I.D.; Cattrall, R.W.; Kolev, S.D. The use of a polymer inclusion membrane for separation and preconcentration of orthophosphate in flow analysis. Anal. Chim. Acta 2013, 803, 82–90. [Google Scholar] [CrossRef] [PubMed]
  78. Kuznetsov, V.V.; Zemyatova, S.V.; Kornev, K.A. Automated determination of uranium(VI) in seawater using on-line preconcentration by coprecipitation. J. Anal. Chem. 2014, 69, 105–110. [Google Scholar] [CrossRef]
  79. Medina, D.A.V.; Santos-Neto, A.J.; Cerda, V.; Maya, F. Automated dispersive liquid-liquid microextraction based on the solidification of the organic phase. Talanta 2018, 189, 241–248. [Google Scholar] [CrossRef]
  80. Gonzalez, A.; Avivar, J.; Cerda, V. Determination of priority phenolic pollutants exploiting an in-syringe dispersive liquid-liquid microextraction-multisyringe chromatography system. Anal. Bioanal. Chem. 2015, 407, 2013–2022. [Google Scholar] [CrossRef]
  81. Boonjob, W.; Miro, M.; Kolev, S.D. On-line speciation analysis of inorganic arsenic in complex environmental aqueous samples by pervaporation sequential injection analysis. Talanta 2013, 117, 8–13. [Google Scholar] [CrossRef]
  82. Pires, C.K.; Reis, B.F.; Morales-Rubio, A.; de la Guardia, M. Speciation of chromium in natural waters by micropumping multicommutated light emitting diode photometry. Talanta 2015, 144, 432–438. [Google Scholar] [CrossRef] [PubMed]
  83. Kozak, J.; Paluch, J.; Węgrzecka, A.; Kozak, M.; Wieczorek, M.; Kochana, J.; Kościelniak, P. Single peak parameters technique for simultaneous measurements: Spectrophotometric sequential injection determination of Fe(II) and Fe(III). Talanta 2016, 148, 626–632. [Google Scholar] [CrossRef] [PubMed]
  84. Chen, S.; Li, N.; Zhang, X.; Yang, D.; Jiang, H. Online spectrophotometric determination of Fe(II) and Fe(III) by flow injection combined with low pressure ion chromatography. Spectrochim. Acta Part A Mol. Biomol. Spectr. 2015, 138, 375–380. [Google Scholar] [CrossRef] [PubMed]
  85. Horstkotte, B.; Duarte, C.M.; Cerda, V. Chip-on-valve concept: An integrated platform for multisyringe flow injection analysis: Application to nitrite and nitrate determination in seawater. Anal. Lett. 2013, 46, 2345–2358. [Google Scholar] [CrossRef] [Green Version]
  86. Lin, K.N.; Ma, J.; Yuan, D.X.; Feng, S.C.; Su, H.T.; Huang, Y.M.; Shangguan, Q.P. Sequential determination of multi-nutrient elements in natural water samples with a reverse flow injection system. Talanta 2017, 167, 166–171. [Google Scholar] [CrossRef]
  87. Lin, K.N.; Xu, J.; Guo, H.G.; Huo, Y.L.; Zhang, Y.B. Flow injection analysis method for determination of total dissolved nitrogen in natural waters using on-line ultraviolet digestion and vanadium chloride reduction. Microchem. J. 2021, 164, 105993. [Google Scholar] [CrossRef]
  88. Lin, K.N.; Pei, J.X.; Li, P.S.; Ma, J.; Li, Q.L.; Yuan, D.X. Simultaneous determination of total dissolved nitrogen and total dissolved phosphorus in natural waters with an on-line UV and thermal digestion. Talanta 2018, 185, 419–426. [Google Scholar] [CrossRef]
  89. Youngvises, N.; Suwannasaroj, K.; Jakmunee, J.; Al Suhaimi, A. Multi-reverse flow injection analysis integrated with multi-optical sensor for simultaneous determination of Mn(II), Fe(II), Cu(II) and Fe(III) in natural waters. Talanta 2017, 166, 369–374. [Google Scholar] [CrossRef]
  90. Kozak, J.; Paluch, J.; Kozak, M.; Kochana, J.; Wieczorek, M.; Kościelniak, P. Novel simultaneous determination of calcium and magnesium based on flow injection gradient titration. Anal. Lett. 2018, 51, 2157–2172. [Google Scholar] [CrossRef]
  91. Peixoto, P.S.; Silva, E.M.P.; Osorio, M.V.; Barreiros, L.; Lima, J.L.F.C.; Segundo, M.A. Automatic solid-phase extraction by programmable flow injection coupled to chromatographic fluorimetric determination of fluoroquinolones. Anal. Methods 2018, 10, 2180–2186. [Google Scholar] [CrossRef]
  92. Horstkotte, B.; Jarosova, P.; Chocholous, P.; Sklenarova, H.; Solich, P. Sequential injection chromatography with post-column reaction/derivatization for the determination of transition metal cations in natural water samples. Talanta 2015, 136, 75–83. [Google Scholar] [CrossRef] [PubMed]
  93. Giakisikli, G.; Anthemidis, A.N. Automatic pressure-assisted dual-headspace gas-liquid microextraction. Lab-in-syringe platform for membraneless gas separation of ammonia coupled with fluorimetric sequential injection analysis. Anal. Chim. Acta 2018, 1033, 73–80. [Google Scholar] [CrossRef] [PubMed]
  94. Mattio, E.; Robert-Peillard, F.; Vassalo, L.; Branger, C.; Margaillan, A.; Brach-Papa, C.; Knoery, J.; Boudenne, J.-L.; Coulomb, B. 3D-printed lab-on-valve for fluorescent determination of cadmium and lead in water. Talanta 2018, 183, 201–208. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  95. Kröckel, L.; Lehmann, H.; Wieduwilt, T.; Schmidt, M.A. Fluorescence detection for phosphate monitoring using reverse injection analysis. Talanta 2014, 125, 107–113. [Google Scholar] [CrossRef] [PubMed]
  96. Toda, K.; Ebisu, Y.; Hirota, K.; Ohira, S.-I. Membrane-based microchannel device for continuous quantitative extraction of dissolved free sulfide from water and from oil. Anal. Chim. Acta 2012, 741, 38–46. [Google Scholar] [CrossRef]
  97. Hue, D.T.K.; Shiba, T.; Maeda, Y.; Takenaka, N. A 3-step chemiluminescence method for Chemical Oxygen Demand measurement with dichromate oxidizing reagent. Anal. Methods 2017, 9, 5797–5805. [Google Scholar] [CrossRef]
  98. Xu, M.; Chen, M.; Dong, T.; Zhao, K.; Deng, A.; Li, J. Flow injection chemiluminescence immunoassay based on resin beads, enzymatic amplification and a novel monoclonal antibody for determination of Hg2+. Analyst 2015, 140, 6373–6378. [Google Scholar] [CrossRef]
  99. Korte, D.; Bruzzoniti, M.C.; Sarzanini, C.; Franko, M. Thermal lens spectrometric determination of colloidal and ionic silver in water. Int. J. Thermophys. 2011, 32, 818–827. [Google Scholar] [CrossRef]
  100. Lin, K.N.; Li, P.C.; Wu, Q.L.; Feng, S.C.; Ma, J.; Yuan, D.X. Automated determination of ammonium in natural waters with reverse flow injection analysis based on the indophenol blue method with o-phenylphenol. Microchem. J. 2018, 138, 519–525. [Google Scholar] [CrossRef]
  101. Khanhuathon, Y.; Siriangkhawut, W.; Chantiratikul, P.; Grudpan, K. Spectrophotometric method for determination of aluminum content in water and beverage samples employing flow-batch sequential injection system. J. Food Comp. Anal. 2015, 41, 45–53. [Google Scholar] [CrossRef]
  102. Andrade, S.I.E.; Lima, M.B.; Barreto, I.S.; Lyra, W.S.; Almeida, L.F.; Araujo, M.C.U.; Silva, E.C. A digital image-based flow-batch analyzer for determining Al(III) and Cr (VI) in water. Microchem. J. 2013, 109, 106–111. [Google Scholar] [CrossRef]
  103. Korte, D.; Bruzzoniti, M.C.; Sarzanini, C.; Franko, M. Influence of foreign ions on determination of ionic Ag in water by formation of nanoparticles in FIA-TLS system. Anal. Lett. 2011, 44, 2901–2910. [Google Scholar] [CrossRef]
  104. Abouhiat, F.Z.; Henriques, C.; El Yousfi, F.; Cerda, V. Sensitive kinetic-catalytic spectrophotometric method for cobalt determination using a chip coupled to a multisyringe flow injection analysis system. Talanta 2017, 166, 405–411. [Google Scholar] [CrossRef] [PubMed]
  105. Yu, Y.-L.; Jiang, Y.; He, R.-H. Development of a miniature analytical system in a lab-on-valve determination of trace copper by bead injection spectroscopy. Talanta 2012, 88, 352–357. [Google Scholar] [CrossRef]
  106. Gonzalez, A.; Mesquita, R.B.R.; Avivar, J.; Moniz, T.; Rangel, M.; Cerda, V.; Rangel, A.O.S.S. Microsequential injection lab-on-valve system for the spectrophotometric bi-parametric determination of iron and copper in natural waters. Talanta 2017, 167, 703–708. [Google Scholar] [CrossRef]
  107. Hatta, M.; Measures, C.I.; Ruzicka, J. Programmable flow injection. Principle, methodology and application for trace analysis of iron in a sea water matrix. Talanta 2018, 178, 698–703. [Google Scholar] [CrossRef]
  108. Cogan, D.; Cleary, J.; Phelan, T.; McNamara, E.; Bowkett, M.; Diamond, D. Integrated flow analysis platform for the direct detection of nitrate in water using a simplified chromotropic acid method. Anal. Methods 2013, 5, 4798–4804. [Google Scholar] [CrossRef] [Green Version]
  109. Hatta, M.; Ruzicka, J.; Measurers, C.I.; Davies, M. Programmable flow injection in batch mode: Determination of nutrients in sea water by using a single, salinity independent calibration line, obtained with standards prepared in distilled water. Talanta 2021, 232, 122354. [Google Scholar] [CrossRef]
  110. Lin, K.N.; Ma, J.; Pai, S.C.; Huang, Y.M.; Feng, S.C.; Yuan, D.X. Determination of nitrite, phosphate, and silicate by valveless continuous analysis with a bubble-free flow cell and spectrophotometric detection. Anal. Lett. 2017, 50, 510–529. [Google Scholar] [CrossRef]
  111. Chuntib, P.; Jakmunee, J. Simple flow injection colorimetric system for determination of paraquat in natural waters. Talanta 2015, 144, 432–438. [Google Scholar] [CrossRef]
  112. Mattio, E.; Robert-Peillard, F.; Branger, C.; Puzio, K.; Margaillan, A.; Brach-Papa, C.; Knoery, J.; Boudenne, J.-L.; Coulomb, B. 3D-printed flow system for determination of lead in natural waters. Talanta 2017, 168, 298–302. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  113. Kozak, J.; Latocha, K.; Kochana, J.; Wieczorek, M.; Kościelniak, P. Simultaneous spectrophotometric flow injection determination of phosphate and silicate. Talanta 2015, 133, 150–154. [Google Scholar] [CrossRef] [PubMed]
  114. Horstkotte, B.; Suarez, R.; Solich, P.; Cerda, V. In-syringe magnetic stirring assisted dispersive liquid-liquid micro-extraction with solvent washing for fully automated determination of cationic surfactants. Anal. Methods 2014, 6, 9601–9609. [Google Scholar] [CrossRef]
  115. Avivar, J.; Ferrer, L.; Casas, M.; Cerda, V. Smart thorium and uranium determination exploiting renewable solid-phase extraction applied to environmental samples in a wide concentration range. Anal. Bioanal. Chem. 2011, 400, 3585–3594. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  116. Danchana, K.; de Souza, C.T.; Palacio, E.; Cerda, V. Multisyringe flow injection analysis for the spectrophotometric determination of uranium (VI) with 2-(5-bromo-2-pyridylazo)-5-diethylaminophenol. Microchem. J. 2019, 150, 104148. [Google Scholar] [CrossRef]
  117. Phansi, P.; Henriquez, C.; Palacio, E.; Nacapricha, D.; Cera, V. An automated in-chip-catalytic-spectrophotometric method for determination of copper(II) using a multisyringe flow injection analysis-multipumping flow system. Anal. Methods 2014, 6, 8494–8504. [Google Scholar] [CrossRef]
  118. Rocha, D.P.; Terra, G.C.; Nery, T.S.; Tronto, J.; Pinto, F.G.; Silva, S.G.; Coelho, N.M.M. A flow injection procedure using Layered Double Hydroxide for on line pre-concentration of fluoride. Talanta 2018, 178, 102–108. [Google Scholar] [CrossRef]
  119. Li, H.; Zhang, X.; Yu, L.; Wang, L.; Chen, S. The determination of phosphate in seawater by reverse flow injection spectrophotometry. Rev. Anal. Chem. 2012, 31, 99–105. [Google Scholar] [CrossRef]
  120. Gawin, M.; Konefał, J.; Trzewik, B.; Walas, S.; Tobiasz, A.; Mrowiec, H.; Witek, E. Preparation of a new Cd(II)-imprinted polymer and its application to determination of cadmium(II) via flow-injection-flame atomic absorption spectrometry. Talanta 2010, 80, 1305–1310. [Google Scholar] [CrossRef]
  121. Leśniewska, B.; Trzonkowska, L.; Zambrzycka, E.; Godlewska-Żyłkiewicz, B. Multi-commutation flow system with on-line solid phase extraction exploiting the ion-imprinted polymer and FAAS detection for chromium speciation analysis in sewage samples. Anal. Methods 2015, 7, 1517–1526. [Google Scholar] [CrossRef]
  122. Rocha, P.S.M.; Araujo, A.S.; Cassella, R.J. Renewable column on-line magnetic preconcentration of Cd(II) using Fe3O4 nanoparticles functionalized with L-glutamine for determination by flame atomic absorption spectrometry. Talanta 2021, 222, 121519. [Google Scholar] [CrossRef] [PubMed]
  123. Sivrikaya, S.; Imamoglu, M. Online solid-phase extraction of Cd(II), Cu(II), and Co(II) using covalently attached bis(salicylaldimine) to silica gel for determination in food and water by flame atomic absorption spectrometry. Anal. Lett. 2018, 51, 773–791. [Google Scholar] [CrossRef]
  124. Zare-Dorabei, R.; Boroun, S.; Noroozifar, M. Flow injection analysis-flame atomic absorption spectrometry system for indirect determination of sulfite after on-line reduction of solid-phase manganese (IV) dioxide reactor. Talanta 2018, 178, 722–727. [Google Scholar] [CrossRef] [PubMed]
  125. Wang, Y.; Luo, X.; Tang, J.; Hu, X.; Xu, Q.; Yang, C. Extraction and preconcentration of trace levels of cobalt using functionalized magnetic nanoparticles in a sequential injection lab-on-valve system with detection by electrothermal atomic absorption spectrometry. Anal. Chim. Acta 2012, 713, 92–96. [Google Scholar] [CrossRef]
  126. Poonyaka, R.; Chanthai, S. The effect of silver nanoparticles on selenium speciation analysis in water sample by flow injection-hydride generation atomic absorption spectrometry. Orient. J. Chem. 2017, 33, 1171–1181. [Google Scholar] [CrossRef] [Green Version]
  127. Anthemidis, A.N.; Daftsis, E.I.; Kalogiuri, N.P. A sequential injection lab-at-valve (SI-LAV) platform for hydride generation atomic absorption spectrometry (HG-AAS): On-line determination of inorganic arsenic. Anal. Methods 2014, 6, 2745–2750. [Google Scholar] [CrossRef]
  128. Trujillo, I.S.; Castillo, M.L.A.; Pavon, J.M.C.; Alonso, E.I.V.; de Torres, A.G. Flow injection on-line solid phase extraction for ultra-trace lead determination with hydride generation electrothermal atomic absorption spectrometry. Anal. Methods 2013, 5, 2551–2557. [Google Scholar] [CrossRef]
  129. Silva, L.O.B.; Portugal, L.A.; Palacio, E.; Ferrer, L.; Cerda, V.; Ferreira, S.L.C. Multi-commutated flow system for cadmium determination in natural water by cold vapour atomic absorption spectrometry. J. Anal. At. Spectrom. 2014, 29, 2398–2404. [Google Scholar] [CrossRef]
  130. Guerrero, M.M.L.; Cordero, M.T.S.; Alonso, E.V.; Pavon, J.M.C.; de Torres, A.A.G. High resolution continuum source atomic absorption spectrometry and solid phase extraction for the simultaneous separation/preconcentration and sequential monitoring of Sb, Bi, Sn and Hg in low concentrations. J. Anal. At. Spectrom. 2015, 30, 1169–1178. [Google Scholar] [CrossRef]
  131. Guerrero, M.M.L.; Alonao, E.V.; de Torres, A.G.; Pavon, J.M.C. Simultaneous determination of traces of Pt, Pd, Os, Ir, Rh, Ag and Au metals by magnetic SPE ICP OES and in situ chemical vapour generation. J. Anal. At. Spectr. 2017, 32, 2281–2291. [Google Scholar] [CrossRef]
  132. Han, B.; Jiang, X.; Hou, X.; Zheng, C. Miniaturized dielectric barrier discharge carbon atomic emission spectrometry with online microwave-assisted oxidation for determination of total organic carbon. Anal. Chem. 2014, 86, 6214–6219. [Google Scholar] [CrossRef] [PubMed]
  133. Wang, N.; Tyson, J. Non-chromatographic speciation of inorganic arsenic by atomic fluorescence spectrometry with flow injection hydride generation with a tetrahydroborate-form anion-exchanger. J. Anal. At. Spectrom. 2014, 29, 665–673. [Google Scholar] [CrossRef] [Green Version]
  134. Zhang, Y.L.; Miro, M.; Kolev, S.D. A novel on-line organic mercury digestion method combined with atomic fluorescence spectrometry for automatic mercury speciation. Talanta 2018, 189, 220–224. [Google Scholar] [CrossRef] [PubMed]
  135. Portugal, L.A.; Ferrer, L.; Serra, A.M.; Da Silva, D.G.; Ferreira, S.L.C.; Cerda, V. A non-chromatographic automated system for antimony speciation in natural water exploiting multisyringe flow injection analysis coupled with online hydride generation—Atomic fluorescence spectrometry. J. Anal. At. Spectrom. 2015, 30, 133–1141. [Google Scholar] [CrossRef] [Green Version]
  136. Mizuguchi, H.; Numata, K.; Monma, C.; Iiyama, M.; Tachibana, K.; Nishina, T.; Shida, J. Determination of ultra-trace mercury(II) by flow-injection/anodic stripping voltammetry using a track-etched microporous membrane electrode. Anal. Sci. 2013, 29, 949–954. [Google Scholar] [CrossRef] [Green Version]
  137. Wu, T.; Xia, T.; Xu, J.; Ye, C.; Zhang, D.; Deng, D.; Zhang, J.; Huang, G. Sequential injection-square wave voltammetric sensor for phosphate detection in freshwater using silanized multi-walled carbon nanotubes and gold nanoparticles. Microchem. J. 2021, 167, 106311. [Google Scholar] [CrossRef]
  138. Nellaiappan, S.; Pillai, K.C.; Kumar, A.S. Flow-injection analysis coupled with electrochemical detection of poisonous inorganic arsenic(III) species using a gold nanoparticle/carbon nanofiber/chitosan chemically modified carbon screen printed electrode in neutral pH solution. Anal. Methods 2018, 10, 799–808. [Google Scholar] [CrossRef]
  139. Dębosz, M.; Wieczorek, M.; Paluch, J.; Migdalski, J.; Baś, B.; Kościelniak, P. 3D-printed flow manifold based on potentiometric measurements with solid-state ion-selective electrodes and dedicated to multicomponent water analysis. Talanta 2020, 217, 121092. [Google Scholar] [CrossRef]
  140. Chaneam, S.D.; Inpota, P.; Saisarai, S.; Wilairat, P.; Ratanawimarnwong, N.; Uraisin, K.; Meesiri, W.; Nacapricha, D. Green analytical method for simultaneous determination of salinity, carbonate and ammoniacal nitrogen in waters using flow injection coupled dual-channel C4D. Talanta 2018, 189, 196–204. [Google Scholar] [CrossRef]
  141. Henriquez, C.; Horstkotte, B.; Cerda, V. A highly reproducible solenoid micropump system for the analysis of total inorganic carbon and ammonium using gas-diffusion with conductimetric detection. Talanta 2014, 118, 186–194. [Google Scholar] [CrossRef]
  142. Alahmad, W.; Pluangklang, T.; Mantim, T.; Cefrda, V.; Wilairat, P.; Ratanawimarnwong, N.; Nacapricha, D. Development of flow systems incorporating membraneless vaporization units and flow-through contactless conductivity detector for determination of dissolved ammonium and sulfide in canal water. Talanta 2018, 177, 34–40. [Google Scholar] [CrossRef] [PubMed]
  143. Wieczorek, M.; Kochana, J.; Knihnicki, P.; Wapiennik, K.; Kościelniak, P. Novel electroanalytical method based on the electrostriction phenomenon and its application to determination of Cr(VI) by the flow injection technique. Talanta 2017, 166, 383–390. [Google Scholar] [CrossRef] [PubMed]
  144. Tossanaitada, R.; Masadome, T.; Imato, T. Sequential injection analysis of nitrate ions using a microfluidic polymer chip with an embedded ion-selective electrode. Anal. Methods 2012, 4, 4384–4388. [Google Scholar] [CrossRef]
  145. Infante, H.G.; Warren, J.; Chalmers, J.; Dent, G.; Todoli, J.L.; Collingwood, J.; Telling, N.; Resano, M.; Limbeck, A.; Schoenberger, T.; et al. Glossary methods and terms used in analytical spectroscopy (IUPAC Recommendations 2019). Pure Appl. Chem. 2022, 93, 647–776. [Google Scholar] [CrossRef]
  146. Clavijo, S.; Fernandez, M.; Forteza, R.; Del Rosario Brunetto, M.; Cerda, V. Online coupling lab on valve-dispersive liquid-liquid microextraction-multisyringe flow injection with gas chromatography-mass spectrometry for the determination of sixteen priority PAHs in water. Anal. Methods 2014, 6, 3335–3344. [Google Scholar] [CrossRef]
  147. Kołacińska, K.; Chajduk, E.; Dudek, J.; Samczyński, Z.; Łokas, E.; Bojanowska-Czajka, A.; Trojanowicz, M. Automation of sample processing for ICP-MS determination of 90Sr radionuclide at ppq level for nuclear technology and environmental purposes. Talanta 2017, 169, 216–226. [Google Scholar] [CrossRef]
  148. Kołacińska, K.; Samczyński, A.; Dudek, J.; Bojanowska-Czajka, A.; Trojanowicz, M. A comparison study on the use of Dowex 1 and TEVA-resin in determination of 99Tc in environmental and nuclear coolant samples in a SIA system with ICP-MS detection. Talanta 2018, 184, 527–536. [Google Scholar] [CrossRef]
  149. Rodriguez, R.; Leal, L.; Miranda, S.; Ferrer, L.; Avivar, J.; Garcia, A.; Cerda, V. Automation of 99Tc extraction by LOV prior to ICP-MS detection: Application to environmental samples. Talanta 2015, 133, 88–93. [Google Scholar] [CrossRef]
  150. Zhu, Z.H.; Zheng, A.R. Determination of rhenium in seawater from the Jiulong river estuary and Taiwan Strait, China by automated flow injection inductively coupled plasma-mass spectrometry. Anal. Lett. 2017, 50, 1422–1434. [Google Scholar] [CrossRef]
  151. Beltran, B.; Avivar, J.; Mola, M.; Ferrer, L.; Cerda, V.; Leal, L.O. Automated method for simultaneous lead and strontium isotopic analysis applied to rain water samples and airborne particulate filters (PM10). Environ. Sci. Technol. 2015, 47, 9850–9857. [Google Scholar] [CrossRef]
  152. O’Sullivan, J.E.; Watson, R.J.; Butler, E.C.V. An ICP-MS procedure to determine Cd, Co, Cu, Ni, Pb and Zn in oceanic waters using in-line flow-injection with solid-phase extraction for preconcentration. Talanta 2013, 115, 999–1010. [Google Scholar] [CrossRef] [PubMed]
  153. Clough, R.; Sela, H.; Milne, A.; Lohan, M.C.; Tokalioglu, S.; Worsfold, P.J. Uncertainty contributions to the measurement of dissolved Co, Fe, Pb and V in seawater using flow injection with solid phase preconcentration and detection by collision/reaction cell-quadrupole ICP-MS. Talanta 2015, 133, 162–169. [Google Scholar] [CrossRef] [PubMed]
  154. Ho, T.-Y.; Chien, C.-T.; Wang, B.-N.; Siriraks, A. Determination of trace metals in seawater by an automated flow injection ion chromatography pretreatment system with ICPMS. Talanta 2010, 82, 1478–1484. [Google Scholar] [CrossRef] [PubMed]
  155. Venancio, L.V.L.; Farinha, A.S.F.; Gomes, M.T.S.R. Analysing sulphate and chloride in mineral drinking water by flow injection analysis with a single acoustic wave sensor. Talanta 2018, 189, 65–70. [Google Scholar] [CrossRef]
  156. Yao, N.; Wang, J.; Zhou, Y. Rapid determination of the Chemical Oxygen Demand of water using a thermal biosensor. Sensors 2014, 14, 9949–9960. [Google Scholar] [CrossRef] [Green Version]
  157. Calvo-Lopez, A.; Ymbern, O.; Puyol, M.; Casalta, J.M.; Alonso-Chamarro, J. Potentiometric analytical microsystem based on the integration of a gas-diffusion step for on-line ammonium determination in water recycling in manned space missions. Anal. Chim. Acta 2015, 874, 26–32. [Google Scholar] [CrossRef]
  158. Uthuppu, B.; Heiskanen, A.; Kofoed, D.; Aarnand, J.; Jorgensen, C.; Dufva, M.; Jakobsen, M.H. Micro-flow-injection analysis (µFIA) immunoassay of herbicide residue 2,6-dichlorobenzamide—Towards automated at-line monitoring using modular microfluidics. Analyst 2015, 140, 1616–1623. [Google Scholar] [CrossRef] [Green Version]
  159. Zhu, Z.; Lu, J.J.; Almeida, M.I.G.S.; Pu, Q.; Kolev, S.D.; Liu, S. A microfabricated electroosmotic pump coupled to a gas-diffusion microchip for flow injection analysis of ammonia. Microchim. Acta 2015, 182, 1063–1070. [Google Scholar] [CrossRef]
  160. Alves-Segundo, R.; Ibanez-Garcia, N.; Baeza, M.; Puyol, M.; Alonso-Chamaro, J. Towards a monolithically integrated microsystem based on the green tape ceramics technology for spectrophotometric measurements. Determination of chromium (VI) in water. Microchim. Acta 2011, 172, 225–232. [Google Scholar] [CrossRef]
  161. Czugala, M.; Fay, C.; O’Conner, N.E.; Corcoran, B.; Benito-Lopez, F.; Diamond, D. Portable integrated microfluidic analytical; platform for the monitoring and detection of nitrite. Talanta 2013, 116, 997–1004. [Google Scholar] [CrossRef]
  162. Cogan, D.; Fay, C.; Boyle, D.; Osborne, C.; Kent, N.; Cleary, J.; Diamond, D. Development of a low cost microfluidic sensor for the direct determination of nitrate using chromotropic acid in natural waters. Anal. Methods 2015, 7, 5396–5405. [Google Scholar] [CrossRef]
  163. Inpota, P.; Strzelak, K.; Koncki, R.; Sripunkhai, W.; Jaemsaksiri, W.; Ratanawimarwong, N.; Chantiwas, R.; Nacapricha, D. Microfluidic analysis with front-face fluorimetric detection for the determination of total inorganic iodine in drinking water. Anal. Sci. 2018, 34, 161–167. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  164. Peng, G.L.; He, Q.; Lu, Y.; Huang, J.; Lin, J.M. Flow injection microfluidic device with on-line fluorescent derivatization for the determination of Cr(III) and Cr(VI) in water samples after solid phase extraction. Anal. Chim. Acta 2017, 955, 58–66. [Google Scholar] [CrossRef] [PubMed]
  165. Dharmasiri, U.; Witek, M.A.; Adams, A.A.; Osiri, J.K.; Hupert, M.L.; Bianchi, T.S.; Roelke, D.L.; Soper, S.A. Enrichment and detection of Escherichia coli O157:H7 from water samples using an antibody modified microfluidic chip. Anal. Chem. 2010, 82, 2844–2849. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  166. Matsueda, M.; Yanagisawa, K.; Koarai, K.; Terashima, M.; Fujiwara, K.; Abe, H.; Kitamura, A.; Takagai, Y. Online solid-phase extraction-inductively coupled plasma-quadrupole mass spectrometry with oxygen dynamic reaction for quantification of technetium-99. ACS Omega 2021, 6, 19281–19290. [Google Scholar] [CrossRef]
  167. Andrew, K.N.; Blundell, N.J.; Price, D.; Worsfold, P.J. Flow injection techniques for water monitoring. Anal. Chem. 1994, 66, 916A–922A. [Google Scholar] [CrossRef]
  168. Motomizu, S. Computer-controlled fluid-flow chemical analysis (CC-FCA) and its application to environmental analytical chemistry. Pure Appl. Chem. 2012, 84, 1999–2013. [Google Scholar] [CrossRef]
  169. Feng, S.; Huang, Y.; Yuan, D.; Zhu, Y.; Zhou, T. Development and application of a shipboard method for spectrophotometric determination of trace dissolved manganese in estuarine and coastal waters. Cont. Shelf Res. 2015, 92, 37–43. [Google Scholar] [CrossRef]
  170. Wang, T.; Huang, Y.; Xu, J.; Guo, W.; Yuan, D. Development and application of a shipboard method for spectrophotometric determination of nanomolar dissolved sulfite in estuarine surface waters using reverse flow injection analysis coupled with a long path length liquid waveguide capillary cell. Microchem. J. 2021, 168, 106522. [Google Scholar] [CrossRef]
  171. Huang, Y.; Yuan, D.; Zhu, Y.; Feng, S. Real-time redox speciation of iron in estuarine and coastal waters. Environ. Sci. Technol. 2015, 49, 3619–3627. [Google Scholar] [CrossRef]
  172. Zhu, Y.; Yuan, D.; Huang, Y.; Ma, J.; Feng, S. A sensitive flow-batch system for on board determination of ultra-trace ammonium in seawater: Method development and shipboard application. Anal. Chim. Acta 2013, 794, 47–54. [Google Scholar] [CrossRef] [PubMed]
  173. Zhu, Y.; Chen, J.; Shi, X.; Yuan, D.; Feng, S.; Zhou, T.; Huang, Y. Development and application of a portable fluorescence detector for shipboard analysis of ammonium in estuarine and coastal waters. Anal. Methods 2018, 10, 1781–1787. [Google Scholar] [CrossRef]
  174. Calvo-Lopez, A.; Arasa-Puig, E.; Puyol, M.; Casalta, J.M.; Alonso-Chamarro, J. Biparametric potentiometric analytical microsystem for nitrate and potassium monitoring in water recycling processes for manned space missions. Anal. Chim. Acta 2013, 804, 190–196. [Google Scholar] [CrossRef] [PubMed]
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Trojanowicz, M.; Pyszynska, M. Flow-Injection Methods in Water Analysis—Recent Developments. Molecules 2022, 27, 1410. https://doi.org/10.3390/molecules27041410