Article

Determination of d-Cycloserine Impurities in Pharmaceutical Dosage Forms: Comparison of the International Pharmacopoeia HPLC–UV Method and the DOSY NMR Method

1
Slovenian NMR Centre, National Institute of Chemistry, Hajdrihova 19, SI-1000 Ljubljana, Slovenia
2
Faculty of Chemistry and Chemical Technology, University of Ljubljana, Večna pot 113, SI-1001 Ljubljana, Slovenia
3
Lek Pharmaceuticals d.d., Sandoz Development Center Slovenia, Verovškova ulica 57, SI-1526 Ljubljana, Slovenia
4
Department of Food Chemistry, National Institute of Chemistry, Hajdrihova 19, SI-1000 Ljubljana, Slovenia
5
EN-FIST Centre of Excellence, Trg Osvobodilne fronte 13, SI-1000 Ljubljana, Slovenia
6
University of Ljubljana, Faculty of Pharmacy, Aškerčeva cesta 7, SI-1000 Ljubljana, Slovenia
*
Author to whom correspondence should be addressed.
These authors contributed equally.
Current position at: Biotechnology Development Unit, International Institute for Genetic Engineering and Biotechnology, Trieste, Italy
Molecules 2020, 25(7), 1684; https://doi.org/10.3390/molecules25071684
Submission received: 21 March 2020 / Revised: 1 April 2020 / Accepted: 3 April 2020 / Published: 7 April 2020
(This article belongs to the Special Issue Drug Analysis in Pharmaceutical Development and Drug Manufacturing)

Abstract

:
d-cycloserine is a broad-spectrum antibiotic that is currently being used as a secondary choice in the treatment of tuberculosis. In recent years, it has become more popular, due to its effect on the nervous system. In this current study, we provide evidence that The International Pharmacopoeia HPLC–UV method for d-cycloserine impurity profiling is not repeatable due to the variable response of cycloserine dimer, one of d-cycloserine impurities. Therefore, we introduced the DOSY (diffusion ordered spectroscopy) NMR (nuclear magnetic resonance) technique to determine the levels of d-cycloserine impurities in pharmaceutical dosage forms. The DOSY NMR technique allowed separation of d-cycloserine, its degradation products, and key process impurities in concentrations below pharmacopoeial specification limits. The proposed DOSY NMR method allowed accurate identification and quantification of the cycloserine dimer, which was not possible through the use of the pharmacopoeial HPLC method. The current method has the potential for practical use in analytical laboratories of the pharmaceutical industry.

Graphical Abstract

1. Introduction

d-Cycloserine is a broad-spectrum antibiotic, which can be biosynthesized by Streptomyces garyphalus, Streptomyces orchidaceous, and Streptomyces lavendulae, but can also be obtained by synthesis. It was first isolated in 1955 and in vitro tested against M. tuberculosis in 1966 [1]. It is classified as a second-line antitubercular agent for the treatment of multidrug-resistant tuberculosis and is used in combined therapies, along with other antitubercular drugs to bypass its toxic effects (i.e., neurotoxicity) [1]. In addition, it could be potentially used to treat neuropsychological disorders (e.g., schizophrenia, anxiety, major depression, autism, and Alzheimer’s disease), due to its agonistic effect on N-methyl-d-aspartate (NMDA) receptors [1,2].
To date, several analytical methods have been reported for the determination of d-cycloserine in different matrices, mostly using high-performance liquid chromatography (HPLC). Different approaches like ion-pair chromatography [3,4,5], aqueous normal phase (ANP) chromatography [6], or hydrophilic interaction liquid chromatography (HILIC) [7] have been applied so far to retain the highly hydrophilic analyte and its analogues, on a column.
Since the d-cycloserine molecule does not contain chromophores, it has been usually derivatized prior to (by p-benzoquinone [8] or 9-chloro-10-methyl acrydinium triflate [9]) or after (by o-phtalaldehyde [5]) its HPLC column separation and, thus, detected by a fluorescence detector (FLD). However, none of these methods included determination of its main degradation product, cycloserine dimer (3,6-bis(aminooxymethyl)piperazine-2,5-dione, for structure see Figure 1), which does not form a fluorescent product, and consequently, cannot be detected by FLD.
Several issues regarding the HPLC–UV method for the determination of impurities in pharmaceutical dosage forms containing d-cycloserine have been met so far. Particularly, the quantification of the cycloserine dimer has been challenging. An ion-pair HPLC–UV method was proposed to identify and quantify d-cycloserine and its degradation products [4]. The aforementioned work reported unrepeatable cycloserine dimer peak areas and consequently hampered quantification thereof. It was speculated that the reason might lie in the nucleophilic nature of this compound, causing its binding to free silanol groups of the column [4]. To solve the problem, column conditioning through successive injections of the cycloserine dimer in a high concentration of 0.5 mg mL−1 to 2.0 mg mL−1 was suggested [4]. This method was further optimized [3] and later published in The International Pharmacopoeia (Ph. Int.) [10,11]. The International Pharmacopoeia mentions two known impurities, the d-serine (for structure see Figure 1) and cycloserine dimer [10,11].
Kaushal et al. developed another HPLC–UV method for d-cycloserine impurity profiling [12]. This method showed insufficient resolution between the peaks of d-cycloserine, d-serine, and hydroxylamine, while the dimer was not included in the study [12]. The United States Pharmacopoeia (USP) contains only a method for d-cycloserine assay determination [13]. In addition, it has been reported that the USP method does not provide a stable baseline [3]. Only ANP chromatography coupled with mass spectrometry (MS) showed a successful determination of d-cycloserine and its dimer [6].
Several methods, including HPLC–UV [14], HPLC coupled with mass spectrometry (MS) [15] or tandem mass spectrometry (MS/MS) [7,16,17,18], for the determination of d-cycloserine in human plasma have also been reported. However, these approaches do not consider nor measure cycloserine dimer as an impurity. Similarly, in another study, only cycloserine was determined in microdialysis samples through LC–MS/MS, after derivatization with benzoyl chloride [19]. A (reversed-phase) RP–HPLC method for the determination of enantiomeric purity of cycloserine, using chiral derivatizing reagents, o-phthalaldehyde, and N-acetyl-cysteine was also proposed [20].
Two studies reported on the dimerization of d-cycloserine initiated by acetonitrile when used as a sample solvent and strongly recommended methanol use instead [6,21]. This dimerization could be further enhanced in the electrospray ionization (ESI) source [21]. In addition, it was found that d-cycloserine is unstable under acidic conditions (pH 1–2) [12,22], converting into its dimer in the solid-state and in solution [22]. Interestingly, a study from 1962 proposed that the active form of d-cycloserine in vivo is actually its dimer [23]. However, it has become imperative to purify cycloserine from the dimer, which is nowadays considered to be a contaminant [23]. Recently, a new study reported on severe peak tailing of a d-cycloserine peak, due to its on-column dimerization, promoted by stationary phase metal oxides, which was resolved by addition of citrate (chelating agent) to the HILIC mobile phase [24].
The present study showed that d-cycloserine impurities could not be reliably quantified in a final dosage form using The International Pharmacopoeia method [10,11]. Therefore, a new approach was proposed using the DOSY NMR technique to determine the levels of its impurities. Few reports exist in the literature, where the quality of different products (e.g., Fluoxetine, Fluvoxamine [25], heroin [26] and libido improving food supplements [27]) was determined by DOSY NMR. To this end, the concentration of the analytes was determined by 1H spectra, while using 3-(trimethylsilyl)propionic-2,2,3,3-d4 acid sodium salt (TMSP) and maleic acid as standards.

2. Results and Discussion

2.1. Verification of The International Pharmacopoeia (Ph. Int.) HPLC Method for Determination of Impurities in Samples Containing d-Cycloserine

2.1.1. Repeatability of d-Cycloserine Chromatographic Peak Areas

Standard d-cycloserine in concentration of 0.0015 mg mL−1 was injected ten times in sequence. Relative standard deviation (RSD) for the area of d-cycloserine peaks obtained from all ten injections was calculated to be 0.58%, which proved the repeatability of the analytical system for determination of d-cycloserine. Moreover, retention times of d-cycloserine were repeatable as well.

2.1.2. Determination of the Limit of Quantification (LOQ) for d-Cycloserine

For determination of LOQ, five different concentrations of d-cycloserine in the range of 0.001% to 0.05%, relative to its concentration in the sample were prepared and the peak signal to noise ratio (S/N) was checked. The concentration at 0.05% (S/N = 52.5) was determined as LOQ. At this level, the peak was easily integrable, while the concentration was still below the specification limit (no more than (NMT) 0.15% in active pharmaceutical ingredient and NMT 0.4% in final dosage form [10,11]).

2.1.3. Linearity of the d-cycloserine Signal

The solutions for the linearity study were prepared, as shown in Table 1 (Lin1–Lin9), and the linearity of the d-cycloserine signal in the range of 0.05% to 0.53% d-cycloserine, relative to its concentration in the sample was tested. The equation of the curve was y = 91799x + 404.25, where y stands for area and x for concentration of d-cycloserine in µg mL−1. R2 was 0.9999 (R2 ≥ 0.99), which confirmed the linear response of the d-cycloserine signal in the determined range.

2.1.4. Determination of Retention Times of d-Cycloserine and Its Impurities

Retention times for d-cycloserine and its impurities were checked at the wavelength of 219 nm, separately for each compound. Among the evaluated impurities, we considered the cycloserine dimer and d-serine (Figure 1), which are both known process impurities (related substances), as well as the degradation products of the final dosage product from the Ph. Int. monographs [10,11]. In addition, we also examined the 3-chloro-d-alanine methyl ester hydrochloride (for structure see Figure 1), which is a well-known process impurity of d-cycloserine [28,29,30,31]. The solutions were injected with concentrations of about 0.77 mg mL−1. The d-serine peak area was significantly lower in comparison to that of d-cycloserine. The peak of the 3-chloro-d-alanine methyl ester hydrochloride was not observed, while the cycloserine dimer gave two chromatographic peaks that have been reported to correspond to its isomeric forms [3]. Depending on the injection number, the symmetry of cycloserine dimer peaks was varied, but in general the peaks were not bell-shaped (Gaussian). Interestingly, when another chromatographic (ANP) method was coupled with mass spectrometric detection was applied, only one peak corresponding to the cycloserine dimer was observed [6].
Since the 3-chloro-d-alanine methyl ester hydrochloride peak was not observed at 219 nm, UV spectra of all compounds that were separately dissolved in mobile phase A, were acquired in the range of 190 nm to 400 nm. Figure 2 clearly shows that the d-cycloserine impurities (cycloserine dimer, d-serine, and 3-chloro-d-alanine methyl ester hydrochloride) absorb at lower wavelengths in comparison to the d-cycloserine molecule, which was most likely the reason why the 3-chloro-d-alanine methyl ester hydrochloride peak was not observed previously and the peak for d-serine was of really low intensity. Although the issue with the d-serine peak could be resolved by including its response factor in the calculations, the tested method [3] was not suitable for the determination of 3-chloro-d-alanine methyl ester hydrochloride. Since d-serine and the 3-chloro-d-alanine methyl ester hydrochloride absorb at lower wavelengths, chromatograms of all compounds were recorded again at 190 nm, despite the noisy baseline and incompatibility of the mobile phase at this wavelength.
Using the Ph. Int. HPLC method, the elution order of the tested compounds starting from the lowest retention time was as follows—3-chloro-d-alanine methyl ester hydrochloride < d-serine < d-cycloserine < cycloserine dimer 1 < cycloserine dimer 2. The cycloserine dimer in the acidic mobile phase (pH 2.8) harbored two positively charged amino groups, which might result in high affinity to the negatively charged groups of sodium 1-octanesulfonate (SOS) in the stationary phase (ion-exchange chromatography) and, thus, the highest retention time. Both, d-cycloserine and d-serine had only one positively charged amino group. The reason for the difference in their retention times lies in their polarity. Namely, d-serine was more polar with a polar surface area of 83.5 Å, in comparison to d-cycloserine with a polar surface area of 64.4 Å. Although 3-Chloro-d-alanine methyl ester hydrochloride has a polar surface area of 52.3 Å, it is present in the form of an ionic salt, which could be the reason for its early elution. Indeed, 3-Chloro-d-alanine methyl ester hydrochloride elutes with the dead volume, which makes its quantification with this method impossible. Data for the polar surface areas of the molecules were obtained with the software ACD/Labs Perceptas Platform [32].

2.1.5. Saturation of Column Free Silanol Groups with Cycloserine Dimer and Repeatability of Cycloserine Dimer Chromatographic Peak Areas

As for d-cycloserine, linearity, repeatability, and other validation parameters should also be tested for its impurities, which are known to be present in the final dosage form. To date, the linearity of the reported HPLC method for the determination of d-cycloserine-related substances [3], which was later adopted as the official method and was included in the Ph. Int., was performed only by evaluating the response of d-cycloserine.
Burge et al. (2005) reported an issue with the repeatability of the cycloserine dimer chromatographic peak areas, speculating that this could be due to the fact that the nucleofilic cycloserine dimer could bind to the free silanol groups of the column [4]. Therefore, it was suggested to saturate the column with cycloserine dimer, before the analyses [4]. Pendela et al. (2008) reported that they have not observed this issue with an Hypersil BDS column [3] and their method was later included in The International Pharmacopoeia [10,11].
The Ph. Int. HPLC method for determination of d-cycloserine-related substances was tested using the new Hypersil BDS column with included column saturation, using two times 50 µL of the cycloserine dimer solution (0.5 mg mL−1) [4]. Then, cycloserine dimer in the same concentration and volume was consecutively injected twenty times from the same HPLC vial. The obtained chromatograms of the cycloserine dimer are shown in Figure 3. The first four injections resulted in no or very little response. In the chromatogram corresponding to the fifth injection, two chromatographic peaks (isomer 1 and 2 [6] at retention times 24.3 and 25.3 min, respectively) appeared and stayed present in the next chromatograms, up to chromatogram 16. The problem reoccurred again from 17th to 20th injection. In Table 2, results from the repeatability study for cycloserine dimer are summarized. Based on the RSD results shown in Table 2, the Ph. Int. HPLC method was not found to be repeatable for the determination of the cycloserine dimer. Therefore, further verification of the Ph. Int. HPLC method was skipped.
When designing a new HPLC quantification study of d-cycloserine and its related compounds, one should be careful, as d-cycloserine might not be stable in the solvent used (mobile phase A; see Section 3.3.2), since it contains acetonitrile and a buffer with a pH of 2.8. According to the literature, dimerization of d-cycloserine was initiated by acetonitrile as a sample solvent, therefore, promoting methanol as a solvent of choice [6,21]. Conversion of d-cycloserine to the cycloserine dimer was found to be further enhanced in the ESI source of the MS [21]. Moreover, d-cycloserine was unstable under acidic conditions [12,22], and converted to its dimer in the solid-state and in solution [22]. The dimer was also shown to exist in a pH-dependent equilibrium (pH 1-2) with cycloserine [22]. The addition of citrate or ethylenediaminetetraacetic acid into the mobile phase might prevent the potential on-column dimerization of d-cycloserine, as reported by Heaton et al. (2019) [24]. Therefore, a future study, where the sample solvent and the pH are carefully controlled might solve some of the issues with the quantification of d-cycloserine and its related substances in pharmaceutical dosage forms. Furthermore, a systematic method development and thorough troubleshooting approach are needed to reveal the right reasons for the unrepeatability of d-cycloserine dimer areas. This would aid the development of a reliable HPLC–UV method for the determination of d-cycloserine-related compounds, thus, increasing the safety of the produced d-cycloserine drugs. However, development of new HPLC method was out of the scope of this manuscript and alternative NMR method for the determination of d-cycloserine impurities in final dosage forms, was examined.

2.2. Development and Verification of the 1H DOSY NMR Method for Determination of Impurities in Samples Containing d-Cycloserine

The HPLC method performed according to The International Pharmacopoeia failed to show repeatability of the cycloserine dimer chromatographic peak areas and, therefore, was not considered to be suitable for quantification of the cycloserine dimer. Therefore, a 1H DOSY NMR method was developed as an alternative method for the determination of d-cycloserine impurities in its formulations.
However, NMR spectroscopy was less sensitive with respect to the HPLC method. Therefore, the limit of detection (LOD) and LOQ were determined for the cycloserine dimer, with initial NMR experiments. Table 3 presents the S/N ratio of selected NMR signals at different concentrations of the cycloserine dimer in the 1H and 1H DOSY NMR spectra. Impurity profiling methods should achieve S/N ≥ 10, according to The International Pharmacopoeia, which was satisfied at 4 μg mL−1 for 1H NMR and 10 or 20 μg mL−1 for the DOSY NMR spectrum (Table 3, NMR spectra of the cycloserine dimer are shown in the Supplementary Materials, Figures S1–S3).
The established HPLC method in The International Pharmacopoeia [10,11] asks for the preparation of capsule solution containing 0.5 mg mL−1 of d-cycloserine. For the 1H DOSY NMR method, concentrations of d-cycloserine and impurity solutions were increased due to the lower sensitivity of NMR. The high solubility of d-cycloserine in water allowed preparation of samples, where concentration of the d-cycloserine solution was 15 mg mL−1, and the specification limit for impurities was achieved at 60 μg mL−1 for the final dosage form (0.4%), and at 22.5 μg mL−1 (0.15%) for API (Active Pharmaceutical Ingredient). Signals of substances at these concentrations gave the appropriate S/N ratio for the DOSY NMR technique. Figure 4 shows the 1H DOSY NMR spectrum of the 0.15% cycloserine dimer, relative to the concentration of d-cycloserine in the sample solution. Well-resolved NMR signals of the d-cycloserine and cycloserine dimer allowed the determination of the translational diffusion coefficient for API and dimer impurity, with the values of 5.5 × 10−10 m2 s−1 and 3.2 × 10−10 m2 s−1 for the d-cycloserine and the cycloserine dimer, respectively.
In addition to the d-cycloserine and cycloserine dimer signals, NMR signals of the unidentified impurities were observed in the 1H spectrum, which most likely corresponded to the degradation products or process-related impurities of the d-cycloserine, as they were also observed in the 1H NMR spectra of the pure active substance d-cycloserine.
Additionally, 0.05% cycloserine dimer, relative to the concentration of d-cycloserine in the sample solution was prepared and analyzed to reach the quantification limit. The 1H NMR and 1H DOSY NMR spectra are shown in the Supplementary Materials (Figure S4), and once again the NMR signals of the d-cycloserine and cycloserine dimer were well-resolved.
Two other impurities were potentially found in the d-cycloserine products—3-chloro-d-alanine methyl ester hydrochloride and d-serine. Both impurities are process impurities of d-cycloserine, whereas the second is also the decomposition product of the d-cycloserine final dosage form. The method for determining the impurity profile must allow sufficient resolution between the peaks or otherwise enable proper identification and quantification of impurities. The 1H NMR spectra of all three impurities and the d-cycloserine are shown in Supplementary Materials (Figure S5).
Quantitative NMR analysis was performed for the sample solution containing all three impurities and d-cycloserine. Well-resolved 1H NMR signals of each compound were used (Figure 5), with the addition of internal standard maleic acid, to determine the content of each compound in the mixture according to well-known equation:
P x = I x × N R S × M x × P R S × m R S I R S × N x × M R S × m x
where mx is the (calculated) mass of the analyte; mRS is the mass of the reference substance; Ix is the integral of the analyte signal; IRS is the integral of the reference substance (= 2.000); Nx is the number of analyte proton nuclei; NRS is the number of reference substance proton nuclei (= 2); Mx is the molecular mass of the analyte; MRS is the molecular mass of the reference substance (= 116.07 gmol−1); Px is the purity of the analyte as mass fraction; and PRS is the purity of the reference substance as mass fraction.
The accuracy and reliability of the 1H NMR quantification was established for the d-cycloserine and its impurities, in comparable concentrations. We demonstrated that measurements of the intended analytes’ signals were free of interference from the reference substance, as well as from impurities (Figure 5). The values of the integrals were reported relative to the signal of maleic acid, which was arbitrarily set at a value of 2. The calculated and actual contents of d-cycloserine and its impurities in the sample solution are given in Table 4. It was shown by Malz and Jancke that single pulse 1H NMR fulfils all requirements to be used as a validated method for quantitative determinations of the amount fractions of dissolved sample mixtures, where all issues regarding linearity, robustness, specificity, selectivity, and accuracy were considered [33]. On the other hand, the NMR stability test (Tables S1–S4) and T1 relaxation times (Figure S6) for each compound were determined and are reported in the Supplementary Materials.
The targeted NMR method should be able to characterize and quantify d-cycloserine and its impurities at the specification limit (NMT 0.15% with respect to concentration of d-cycloserine in API and NMT 0.4% with respect to concentration of d-cycloserine in final dosage form [10,11]). However, to employ the DOSY technique for characterization based on different molecular weight, one has to meet the following criterion [25]:
M W a M W b M ¯ W   0.10
Certain pairs of compounds satisfy this condition easily (d-cycloserine − cycloserine dimer = 0.67), whereas others fail to do so (e.g., d-serine − d-cycloserine = 0.03). Presence of all three impurities in a final dosage form is the worst-case scenario and most likely will not occur. In addition, it can be controlled, as the 3-chloro-d-alanine methyl ester hydrochloride is a process impurity that can be limited below the specification limit, during the synthesis of pure active substance d-cycloserine. As a result, a 1H NMR spectrum of high resolution and acceptable selectivity can be obtained (Figure S5).
The suitability of the method was eventually evaluated on a real sample. d-cycloserine capsule solution was prepared to obtain d-cycloserine concentration of about 5 mg mL−1. Specification limit for each impurity was 20 μg mL−1, which satisfied the S/N requirement. Analysis of sample solutions through NMR can be performed for the samples, where at least 80% of API is still intact. The stress stability procedure usually results in 90% of intact API. Therefore, the d-cycloserine capsules packed in a primary container were exposed to accelerated stress conditions (at 40 °C, RH 30%) for 10 days, which resulted in 87% of intact API, as established by the 1H NMR spectrum. The 1H DOSY NMR spectrum of the d-cycloserine capsule solution did not show the presence of process impurity (3-chloro-d-alanine methyl ester hydrochloride, Figure 6). Therefore, DOSY allowed us to distinguish NMR signals attributed to API, its key degradation products, and related substances. NMR assays were calculated based on the 1H NMR spectra. The 1H NMR spectra were used to establish a concentration of cycloserine dimer (0.1 mg mL−1) in a sample solution, which corresponded to 2% of d-cycloserine and was above the specification limit (Figure S7).

3. Materials and Methods

3.1. Chemicals

Acetonitrile (ACN) HPLC gradient grade was purchased from Fisher Scientific (Leicester, UK). Sodium 1-octanesulfonate (SOS) was acquired from Sigma-Aldrich (St. Louis, MO, USA), while potassium dihydrogen phosphate (KH2PO4), and phosphoric acid (H3PO4; 85%, analytical grade) were purchased from Merck (Darmstadt, Germany). Demineralized water was supplied by the Milli-Q water purification system (Millipore, Milford, MA, USA).
d-cycloserine (99.6%; USP standard) was used as a standard for the active pharmaceutical ingredient. Other standards used in the study were d-serine (100%), which was purchased from Sigma-Aldrich, cycloserine dimer (3,6-bis(aminooxymethyl)piperazine-2,5-dione) (95%) from Toronto Research Chemicals Inc. (North York, ON, Canada), and 3-chloro-d-alanine methyl ester hydrochloride (97.3%) from Sigma-Aldrich (St. Louis, MO, USA).
d-cycloserine capsules were manufactured by The Chao Center (West Lafayette, IN, USA).
For the NMR experiments, 99.9% deuterated water from Merck was used as a solvent. Maleic acid (99.98%; TraceCERT® CRM) was purchased from Sigma-Aldrich (St. Louis, MO, USA). Deuterated sodium 3-(trimethylsilyl)propionate (TMSP) was purchased from ARMAR Chemicals (Döttingen, Switzerland).

3.2. Instrumentation

HPLC chromatograms were acquired using the Waters 2695 Alliance liquid chromatographic system (Milford, MA, USA), equipped with the UV Waters 2996 photodiode array detector (Milford, MA, USA). Empower3 software (Waters, Milford, MA, USA) was used to record and process the signals.
1D 1H and 1H-DOSY NMR spectra were collected on an Agilent Technologies DD2 600 MHz NMR spectrometer (Santa Clara, CA, USA), equipped with the HCN Cold probe.

3.3. Verification of The International Pharmacopoeia (Ph. Int.) HPLC Method for Determination of Impurities in Samples Containing d-Cycloserine

3.3.1. Chromatographic Conditions

Potassium phosphate monobasic buffer (0.2 M) was prepared by dissolving 27.2 g of KH2PO4 in 800 mL water, adjusting the pH at 2.8 ± 0.05 with 85% H3PO4 and adding water to the total volume of 1000 mL. SOS solution (20 mM) was prepared by dissolving 4.7 g SOS in 1000 mL water. Mobile phases A and B, both consisting of ACN, 20 mM SOS, 0.2 M KH2PO4 buffer at pH 2.8 and water were prepared as specified in the Ph. Int. in the following ratios—mobile phase A: ACN/SOS/KH2PO4 buffer/water = 4/70/10/16 v/v/v/v; and mobile phase B: ACN/SOS/KH2PO4 buffer/water = 17/70/10/3 v/v/v/v.
A Hypersil BDS C18 HPLC column (250 × 4.6 mm; 5 µm) from Thermo Fischer Scientific (San Jose, CA, USA) was used for the analysis. Gradient elution was applied at 45 °C and a flow rate of 1 mL min−1: 0% B (0–16 min), 0%–100%B (16–18 min), 100% B (18–22 min), 100%–0%B (22–24 min), and 0% B (24–30 min). The detection wavelength was set at 219 nm and the injection volume was 50 µL.

3.3.2. Standard and Sample Preparation

d-cycloserine capsule content, corresponding to 50 mg of the active compound was weighted and dissolved in 100 mL mobile phase A, to obtain a solution with a concentration of 0.5 mg mL−1. This solution was mixed and filtered through 0.45 µm Millipore Millex HV, PVDF filter, Merck.
A standard stock solution of d-cycloserine was prepared in a concentration of 0.5 mg mL−1 in mobile phase A (50 mg d-cycloserine in 100 mL mobile phase A). This solution was further diluted to obtain a standard solution of 0.1% d-cycloserine, relative to its concentration in the sample (0.1 mL from the standard stock solution was transferred to a 100 mL flask and mobile phase A was added to the mark).
To check the repeatability of d-cycloserine chromatographic peak areas, 10 mg of the d-cycloserine standard was dissolved in 50 mL mobile phase A and was ultrasonicated for 5 min until complete dissolution. It was further diluted (0.75 mL from the stock solution was transferred to 100 mL flask and mobile phase A was added to the mark) to obtain a solution with a concentration of 0.0015 mg mL−1.
To check the linearity and limit of quantification (LOQ), 2 mg of the d-cycloserine standard were dissolved in 10 mL mobile phase A, and ultrasonicated for 5 min until complete dissolution. The solution was further diluted (2 mL from the stock solution was transferred to a 20 mL flask and mobile phase A was added to the mark), to obtain a working solution (WS). The working solution was again diluted for LOQ and linearity studies, as shown in Table 1.
To check the retention times of d-cycloserine, 3-chloro-d-alanine methyl ester hydrochloride, and d-serine, 7.7 mg of each standard was dissolved in 10 mL of mobile phase A and ultrasonicated for 5 min until complete solvation, to obtain concentrations of about 0.77 mg mL−1.
For column conditioning with cycloserine dimer and repeatability check of the cycloserine dimer chromatographic peak areas, the cycloserine dimer standard was prepared as the previous three standards in mobile phase A, with a concentration of about 0.77 mg mL−1.
To acquire UV spectra, separate solutions containing d-cycloserine, 3-chloro-d-alanine methyl ester hydrochloride, and d-serine with concentrations of about 0.02 mg mL−1 were prepared. Each standard (2 mg) was dissolved in 70 mL mobile phase A, and the solution was ultrasonicated for 5 min, until complete solvation and finally filled up to the volume (100 mL) with mobile phase A. UV spectra were acquired in the range from 190 nm to 400 nm, using a quartz cuvette. A spectrum of the blank containing only mobile phase A, was also acquired.

3.4. Development and Verification of the 1H DOSY NMR Method for Determination of Impurities in Samples Containing d-Cycloserine

Since the active pharmaceutical ingredient, d-cycloserine, as well as all impurities were well-soluble in water, the samples were dissolved in D2O, using ultrasonic bath to obtain concentrations of 2, 4, 10, and 20 μg mL−1. To each sample, the TMSP salt was added as the internal standard for a relative chemical shift and maleic acid, for quantification.
Diffusion coefficients were collected using pulsed-field gradient echo pulse sequence. The acquisition time for the 1H and 1H DOSY spectra were 4.0 s, and the recorded spectra were the result of 64 scans. Recycle delay of 10 s was used for the 1H spectra. The chemical shifts were given in parts per million (ppm), relative to the TMSP salt at 0 ppm.

4. Conclusions

The Ph. Int. HPLC method for the d-cycloserine capsule impurity profiling showed good repeatability, linearity, and S/N ratio for d-cycloserine. However, the method failed to show the repeatability of the chromatographic peak areas corresponding to the cycloserine dimer—one of the known d-cycloserine impurities—and was, therefore, not considered to be suitable for its quantification. Thus, the DOSY NMR method for determination of d-cycloserine impurities in its formulations was developed. Specification limit of d-cycloserine was achieved at 22.5 μg mL−1, which gave an appropriate S/N ratio for the DOSY NMR technique. Well-resolved NMR signals of the d-cycloserine and cycloserine dimer allowed for the determination of translational diffusion coefficient for API and dimer impurity, with the values 5.5 × 10−10 m2 s−1 and 3.2 × 10−10 m2 s−1 for d-cycloserine and cycloserine dimer, respectively. Furthermore, the accuracy and reliability of the 1H NMR quantification was established for d-cycloserine and its impurities, in comparable concentrations. We demonstrated that measurements of the intended analytes’ signals were free of interference from the reference substance, as well as from impurities. Suitability of the NMR method was eventually evaluated on a real sample. d-cycloserine capsules packed in the primary container were exposed to accelerated stress conditions (10 days at 40 °C, RH 30%). DOSY allowed us to distinguish NMR signals attributed to API and its degradation products. The NMR assay was calculated, based on the 1H NMR spectra. The 1H NMR spectra were used to establish the concentration of the cycloserine dimer, which corresponded to 2% of d-cycloserine and was above the specification limit.
The method described in this study, which combined the 1H and DOSY NMR techniques, offers a new approach for practical use in analytical laboratories of the pharmaceutical industry. DOSY NMR has been found to be particularly useful due to its ability to perform virtual separation of the d-cycloserine and cycloserine dimer impurity, without engaging in a physical separation scheme for the purification of each components. A similar approach could be used for other compounds, as long as the specificity and selectivity were achieved for both 1H and DOSY NMR.

Supplementary Materials

The following are available online. Figure S1: 1H NMR spectra of the cycloserine dimer in D2O at 25 °C, concentrations between 2 and 20 μg mL−1. Figure S2: 1H NMR and 1H DOSY NMR spectra of the cycloserine dimer at concentrations of 20 μg mL−1. Figure S3: 1H NMR and 1H DOSY NMR spectra of the cycloserine dimer at concentration 10 μg mL−1. Figure S4: 1H NMR and 1H DOSY NMR spectra of the d-cycloserine and cycloserine dimer, within the quantification limit (0.05% relative to the concentration of d-cycloserine in the sample solution). Figure S5: 1H NMR spectra of d-cycloserine and impurities in D2O at 25 °C. Figure S6: T1 relaxation times for the d-cycloserine and all three impurities, as well as for maleic acid, using the inversion recovery experiment. Figure S7: 1H NMR spectrum to establish the concentration of the cycloserine dimer (0.1 mg mL−1) in a sample solution, which corresponded to 2% of d-cycloserine. Tables S1–S4: NMR stability tests for all analytes.

Author Contributions

Conceptualization, J.P. and Z.Č.; methodology, D.M., Ž.Š., K.N., J.P., and Z.Č.; data curation, D.M., Ž.Š., and K.N.; formal analysis, D.M., Ž.Š., K.N., J.P., and Z.Č.; investigation, D.M., Ž.Š., and K.N.; project administration, Z.Č.; resources, K.N., J.P., and Z.Č.; supervision, J.P. and Z.Č.; validation, D.M., Ž.Š., and K.N.; visualization, D.M., Ž.Š., K.N., J.P., and Z.Č.; writing—original draft, D.M., Ž.Š., K.N., J.P., and Z.Č.; writing—review & editing, D.M., Ž.Š., K.N., J.P., and Z.Č. All authors have read and agreed to the published version of the manuscript.

Funding

The authors would like to acknowledge the financial support from the Slovenian Research Agency (research core funding No. P1-0005 and No. P1-0242) and the Slovenian Ministry of Education Science and Sport & EU—European Regional Development Fund for the funding of the research project C3330-17-529013.

Acknowledgments

The authors gratefully acknowledge M. Tržan, J. Lalić, and S. Pergar, for support in the procurement of some of the materials.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Hong, W.; Cheng, L.; Xie, J. Molecular basis underlying Mycobacterium tuberculosis d-cycloserine resistance. Is there a role for ubiquinone and menaquinone metabolic pathways? Exp. Opin. Ther. Targets 2014, 18, 691–701. [Google Scholar] [CrossRef] [PubMed]
  2. Schade, S.; Paulus, W. d-cycloserine in neuropsychiatric diseases: A systematic review. Int. J. Neuropsychopharmacol. 2016, 19, 1–7. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  3. Pendela, M.; Dragovic, S.; Bockx, L.; Hoogmartens, J.; Van Schepdael, A.; Adams, E. Development of a liquid chromatographic method for the determination of related substances and assay of d-cycloserine. J. Pharm. Biomed. Anal. 2008, 47, 807–811. [Google Scholar] [CrossRef] [PubMed]
  4. Burge, L.; Sorgen, C.M. Evaluation of polar-embedded and polar-endcapped C12-C18 stationary phase vulnerabilities toward highly nucleophilic compounds: A case study. LCGC N. Am. 2005, 23, 762–765. [Google Scholar]
  5. Musson, D.G.; Maglietto, S.M.; Hwang, S.S.; Gravellese, D.; Bayne, W.F. Simultaneous quantification of cycloserine and its prodrug acetylacetonylcycloserine in plasma and urine by high-performance liquid chromatography using ultraviolet absorbance and fluorescence after post-column derivatization. J. Chromatogr. 1987, 414, 121–129. [Google Scholar] [CrossRef]
  6. Pesek, J.J.; Matyska, M.T.; Dang, A. Analysis of cycloserine and related compounds using aqueous normal phase chromatography/mass spectrometry. J. Pharm. Biomed. Anal. 2012, 64–65, 72–76. [Google Scholar] [CrossRef]
  7. Yaroshenko, D.V.; Grigoriev, A.V.; Sidorova, A.A. Development and validation of a LC-MS/MS method for d-cycloserine determination in human plasma for bioequivalence study. Anal. Bioanal. Chem. 2014, 406, 923–927. [Google Scholar] [CrossRef]
  8. David, V.; Ionescu, M.; Dumitrescu, V. Determination of cycloserine in human plasma by high-performance liquid chromatography with fluorescence detection, using derivatization with p-benzoquinone. J. Chromatogr. B 2001, 761, 27–33. [Google Scholar] [CrossRef]
  9. Yoo, G.S.; Choi, K.; Stewart, J.T. Second derivative ultraviolet spectrophotometry and high performance liquid chromatography with fluorometric detection of cycloserine using 9-chloro-10-methyl acridinium triflate as a new UV and fluorescent labeling agent. Anal. Lett. 1990, 23, 1245–1263. [Google Scholar] [CrossRef]
  10. World Health Organization. Monographs: Pharmaceutical substances: Cycloserine (Cycloserinum). In The International Pharmacopoeia, 9th ed.; 2019; Available online: http://apps.who.int/phint/en/p/docf/ (accessed on 20 March 2020).
  11. World Health Organization. Monographs: Dosage forms: Specific monographs: Cycloserine capsules (Cycloserini capsulae). In The International Pharmacopoeia, 9th ed.; 2019; Available online: http://apps.who.int/phint/en/p/docf/ (accessed on 20 March 2020).
  12. Kaushal, G.; Ramirez, R.; Alambo, D.; Taupradist, W.; Choksi, K.; Sirbu, C. Initial characterization of d-cycloserine for future formulation development for anxiety disorders. Drug Discov. Ther. 2011, 5, 253–260. [Google Scholar] [CrossRef] [Green Version]
  13. The United States Pharmacopeia: Cycloserine; United States Pharmacopoeial convention, Inc.: Rockville, MD, USA, 2019; Volume 42, pp. 1187–1188.
  14. Hemanth Kumar, A.K.; Kumar Polisetty, A.; Sudha, V.; Vijayakumar, A.; Ramachandran, G. A selective and sensitive high performance liquid chromatography assay for the determination of cycloserine in human plasma. Indian J. Tuberc. 2018, 65, 118–123. [Google Scholar] [CrossRef] [PubMed]
  15. Stepanova, E.S.; Ovcharov, M.V.; Barsegyan, S.S.; Chistyakov, V.V. Structure of chemical compounds, methods of analysis and process control. Determination of cycloserine in blood plasma by HPLC/MS: Application to bioequivalence studies. Pharm. Chem. J. 2016, 50, 195–199. [Google Scholar] [CrossRef]
  16. Patel, D.S.; Sharma, N.; Patel, M.C.; Patel, B.N.; Shrivastav, P.S.; Sanyal, M. Development and validation of a selective and sensitive LC–MS/MS method for determination of cycloserine in human plasma: Application to bioequivalence study. J. Chromatogr. B 2011, 879, 2265–2273. [Google Scholar] [CrossRef] [PubMed]
  17. Polagani, S.R.; Pilli, N.R.; Maddela, R.; Gajula, R.; Gandu, V. A rapid and sensitive liquid chromatography–tandem mass spectrometric assay for cycloserine in 50μL of human plasma: Its pharmacokinetic application. J. Pharm. Biomed. Anal. 2013, 76, 21–27. [Google Scholar] [CrossRef]
  18. Mao, Z.; Wang, X.; Li, B.; Jin, J.; Xu, M.; Liu, Y.; Di, X. A simplified LC-MS/MS method for rapid determination of cycloserine in small-volume human plasma using protein precipitation coupled with dilution techniques to overcome matrix effects and its application to a pharmacokinetic study. Anal. Bioanal. Chem. 2017, 409, 3025–3032. [Google Scholar] [CrossRef]
  19. Zhang, W.; Wan, S.; Chen, L.; Wang, X.; Wang, Z.; Huang, Y. Determination of cycloserine in microdialysis samples using liquid chromatography-tandem mass spectrometry with benzoyl chloride derivatization. Biomed. Chromatogr. 2018, 32, e4187. [Google Scholar] [CrossRef]
  20. Karthikeyan, K.; Arularasu, G.T.; Ramadhas, R.; Chandrasekara Pillai, K. Development and validation of indirect RP-HPLC method for enantiomeric purity determination of d-cycloserine drug substance. J. Pharm. Biomed. Anal. 2011, 54, 850–854. [Google Scholar] [CrossRef]
  21. Fraschetti, C.; Filippi, A.; Mannina, L.; Sobolev, A.P.; Speranza, M. Role of the solvent on the stability of cycloserine under ESI-MS conditions. J. Mass. Spectrom. 2014, 49, 608–612. [Google Scholar] [CrossRef]
  22. Lassen, F.O.; Stammer, C.H. Cycloserine dimer hydrolysis and its equilibration with cycloserine. J. Org. Chem. 1971, 36, 2631–2634. [Google Scholar] [CrossRef]
  23. Lynch Miller, J.; Neuhaus, F.C.; Lassen, F.O.; Stammer, C.H. The reactions of 3,6-bis(aminooxymethyl)-2,5-piperazinedione with acid and alkali. A kinetic study. J. Org. Chem. 1968, 33, 3908–3913. [Google Scholar] [CrossRef]
  24. Heaton, J.C.; Smith, N.W.; McCalley, D.V. Retention characteristics of some antibiotic and anti-retroviral compounds in hydrophilic interaction chromatography using isocratic elution, and gradient elution with repeatable partial equilibration. Anal. Chim. Acta 2019, 1045, 141–151. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  25. Trefi, S.; Gilard, V.; Balayssac, S.; Malet-Martino, M.; Martino, R. Quality assessment of fluoxetine and fluvoxamine pharmaceutical formulations purchased in different countries or via the internet by 19F and 2D DOSY 1H NMR. J. Pharm. Biomed. Anal. 2008, 46, 707–722. [Google Scholar] [CrossRef] [PubMed]
  26. Balayssac, S.; Retailleau, E.; Bertrand, G.; Escot, M.P.; Martino, R.; Malet-Martino, M.; Gilard, V. Characterization of heroin samples by 1H NMR and 2D DOSY 1H NMR. Forensic Sci. Int. 2014, 234, 29–38. [Google Scholar] [CrossRef] [PubMed]
  27. Balayssac, S.; Trefi, S.; Gilard, V.; Malet-Martino, M.; Martino, R.; Delsuc, M.A. 2D and 3D DOSY 1H NMR, a useful tool for analysis of complex mixtures: Application to herbal drugs or dietary supplements for erectile dysfunction. J. Pharm. Biomed. Anal. 2009, 50, 602–612. [Google Scholar] [CrossRef] [PubMed]
  28. Kim, H.-K.; Jenny Park, K.-J. Simple and efficient synthetic routes to d-cycloserine. Tetrahedron Lett. 2012, 53, 1668–1670. [Google Scholar] [CrossRef]
  29. Kim, H.-K.; Jenny Park, K.-J. Novel practical synthesis of d-cycloserine. Tetrahedron Lett. 2012, 53, 4090–4092. [Google Scholar] [CrossRef]
  30. Fu, D.; Wu, Z.; Yang, Y.; Wen, W. Process of Preparation of 3-halo-d-alanine Methyl Ester or Acid Salt Thereof. Patent Application No. WO 2016155596, 6 October 2016. [Google Scholar]
  31. Liu, H. Synthesis Method of 3-chloro-d-alanine Methyl Ester Hydrochloride. Chinese Patent Application CN 106518695, 22 March 2017. [Google Scholar]
  32. ACD/Labs Percepta Predictors. Software Modules to Profile Compounds for Physicochemical, ADME, and Toxicity Properties. Available online: http://www.acdlabs.com/products/percepta/predictors.php (accessed on 20 March 2020).
  33. Malz, F.; Jancke, H. Validation of quantitative NMR. J. Pharm. Biomed. Anal. 2005, 38, 813–823. [Google Scholar] [CrossRef]
Sample Availability: Samples of the compounds are not available from the authors.
Figure 1. Chemical structure of the d-cycloserine and its impurities.
Figure 1. Chemical structure of the d-cycloserine and its impurities.
Molecules 25 01684 g001
Figure 2. UV spectra of d-cycloserine, d-serine, cycloserine dimer, and 3-chloro-d-alanine methyl ester hydrochloride.
Figure 2. UV spectra of d-cycloserine, d-serine, cycloserine dimer, and 3-chloro-d-alanine methyl ester hydrochloride.
Molecules 25 01684 g002
Figure 3. The chromatograms of twenty injections of the cycloserine dimer in sequential injection order.
Figure 3. The chromatograms of twenty injections of the cycloserine dimer in sequential injection order.
Molecules 25 01684 g003
Figure 4. 1H NMR (top) and 1H DOSY NMR spectra (bottom) of the d-cycloserine and cycloserine dimer within the specification limit (0.15% relative to the concentration of d-cycloserine in the sample solution). The diffusion coefficient (D) is reported in m2 s−1 × 10−10.
Figure 4. 1H NMR (top) and 1H DOSY NMR spectra (bottom) of the d-cycloserine and cycloserine dimer within the specification limit (0.15% relative to the concentration of d-cycloserine in the sample solution). The diffusion coefficient (D) is reported in m2 s−1 × 10−10.
Molecules 25 01684 g004
Figure 5. 1H NMR spectrum of d-cycloserine and three impurities 3-chloro-d-alanine methyl ester hydrochloride, d-serine, and cycloserine dimer, together with the integral values for each of the signals that was used to calculate the content of each component (replicate 2).
Figure 5. 1H NMR spectrum of d-cycloserine and three impurities 3-chloro-d-alanine methyl ester hydrochloride, d-serine, and cycloserine dimer, together with the integral values for each of the signals that was used to calculate the content of each component (replicate 2).
Molecules 25 01684 g005
Figure 6. The 1H DOSY NMR spectrum of the d-cycloserine capsule solution exposed to accelerated stress conditions. Diffusion coefficient (D) was reported in m2 s−1 × 10−10.
Figure 6. The 1H DOSY NMR spectrum of the d-cycloserine capsule solution exposed to accelerated stress conditions. Diffusion coefficient (D) was reported in m2 s−1 × 10−10.
Molecules 25 01684 g006
Table 1. Dilution protocol to obtain the solutions for the limit of quantification (LOQ) and linearity studies.
Table 1. Dilution protocol to obtain the solutions for the limit of quantification (LOQ) and linearity studies.
Solution NameVolume WS [mL]% d-Cycloserine 1Concentration
[µg mL−1]
LOQ10.0050.0010.005
LOQ20.050.010.05
LOQ30.100.020.10
LOQ40.150.030.15
LOQ5 = Lin10.250.050.25
Lin20.300.060.30
Lin30.350.070.35
Lin41.00.21.0
Lin51.20.241.2
Lin61.50.31.5
Lin72.00.42.0
Lin82.40.482.4
Lin92.650.532.65
1 Relative to its concentration in the sample.
Table 2. Repeatability of the cycloserine dimer chromatographic peak areas.
Table 2. Repeatability of the cycloserine dimer chromatographic peak areas.
InjectionArea [µV*sec]
Cycloserine Dimer 1
Area [µV*sec]
Cycloserine Dimer 2
Total Area
[µV*sec]
Cycloserine Dimers 1 & 2
103279432794
20136069136069
394450435062529512
4000
5157113112619732833104
628135141581982971712
7353295617869995319955
8401522719606355975862
9440210921179706520079
10472343022604726983902
11499538323924547387837
12522289125168707739761
13541918426348018053985
14559395627480048341960
15574512328555498600672
16550969327790198288712
17136259151090287349
18100282132602232884
1979771124348204119
2065357119038184395
RSD (%) (1–20):90.087.788.5
RSD (%) (8–20):68.965.867.8
RSD (%) (8–16):11.612.711.9
Table 3. S/N ratios at different concentrations of cycloserine in the 1H and 1H DOSY NMR spectra.
Table 3. S/N ratios at different concentrations of cycloserine in the 1H and 1H DOSY NMR spectra.
Concentration [μg mL−1]S/N at
δ 4.4 ppm
S/N at
δ 4.1 ppm
S/N at
δ 4.0 ppm
1H2039.142.463.2
1017.526.130.9
411.313.112.3
29.97.77.7
1H DOSY 12012.520.723.9
106.611.016.4
1 First increment of the magnetic field gradient in the DOSY NMR spectrum.
Table 4. Calculated vs. actual content of d-cycloserine and its impurities in the sample solution 1.
Table 4. Calculated vs. actual content of d-cycloserine and its impurities in the sample solution 1.
CompoundReplicatemxIxNxmRSCalculated Px (%)Actual Px (%)
d-cycloserine 1replicate 12.9081.36512.25693.12299.6
replicate 21.9541.15811.78893.18199.6
replicate 32.0840.90912.44593.78299.6
average:93.499.6
Cycloserine dimer 2replicate 11.4470.71222.25697.62795.0
replicate 21.5810.96121.78895.58295.0
replicate 31.5840.71422.44596.92695.0
average:96.795.0
3-chloro-d-alanine methyl ester HCl 3replicate 11.4911.25532.25694.88697.3
replicate 21.5601.66731.78895.47297.3
replicate 31.9881.61032.44598.94397.3
average:96.497.3
d-serine 4replicate 11.5570.81012.256106.240100.0
replicate 22.7281.73311.788102.820100.0
replicate 33.5311.63112.445102.233100.0
average:103.8100.0
1 M(d-cycloserine) = 102.09 gmol−1. 2 M(dimer) = 204.20 gmol−1. 3 M(ester) = 174.03 gmol−1. 4 M(d-serine) = 105.09 gmol−1; For all replicates: MRS = 116.07 gmol−1, PRS = 99.98, IRS = 2.000, and NRS = 2.

Share and Cite

MDPI and ACS Style

Makuc, D.; Švab, Ž.; Naumoska, K.; Plavec, J.; Časar, Z. Determination of d-Cycloserine Impurities in Pharmaceutical Dosage Forms: Comparison of the International Pharmacopoeia HPLC–UV Method and the DOSY NMR Method. Molecules 2020, 25, 1684. https://doi.org/10.3390/molecules25071684

AMA Style

Makuc D, Švab Ž, Naumoska K, Plavec J, Časar Z. Determination of d-Cycloserine Impurities in Pharmaceutical Dosage Forms: Comparison of the International Pharmacopoeia HPLC–UV Method and the DOSY NMR Method. Molecules. 2020; 25(7):1684. https://doi.org/10.3390/molecules25071684

Chicago/Turabian Style

Makuc, Damjan, Živa Švab, Katerina Naumoska, Janez Plavec, and Zdenko Časar. 2020. "Determination of d-Cycloserine Impurities in Pharmaceutical Dosage Forms: Comparison of the International Pharmacopoeia HPLC–UV Method and the DOSY NMR Method" Molecules 25, no. 7: 1684. https://doi.org/10.3390/molecules25071684

Article Metrics

Back to TopTop