Next Article in Journal
Influence of Agricultural Activity on the Surface Climate of New Delta of Egypt Using the RegCM4
Previous Article in Journal
Wearable Preventive Pressure Ulcer System Using Embroidered Textile Electrodes
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Proceeding Paper

Design and Test of E-Textiles for Stroke Rehabilitation †

1
Electronics and Computer Science, University of Southampton, Southampton SO17 IBJ, UK
2
Winchester School of Art, University of Southampton, Southampton SO23 8DL, UK
3
Odstock Medical Ltd., Salisbury SP2 8BJ, UK
*
Author to whom correspondence should be addressed.
Presented at the 4th International Conference on the Challenges, Opportunities, Innovations and Applications in Electronic Textiles, Nottingham, UK, 8–10 November 2022.
Eng. Proc. 2023, 30(1), 16; https://doi.org/10.3390/engproc2023030016
Published: 6 February 2023

Abstract

:
This work presents the design and test of an e-textile based functional electrical stimulation system for post-stroke upper limb rehabilitation. The prototype was tested on five stroke survivors to assess stimulation comfort, the stimulation intensity required to achieve hand opening, and ease of use. Wrist extension was measured using two inertial measurement units. The wearable e-textile prototype achieved similar stimulation comfort compared to high-quality hydrogel electrodes with a score difference of between 0 and 1. The stimulation intensity to achieve full hand opening was the same for the hydrogel electrodes and the e-textiles for all five participants. A second design based on a knitted sleeve has been assessed in terms of usability. Additional new designs have been proposed to improve the usability.

1. Introduction

Stroke occurs when there is a blockage or bleeding of the blood vessels affecting the supply of blood to the brain. There are 1.3 million stroke survivors in the UK [1] and stroke costs the UK National Health Service and wider society £26 billion per annum [2]. Over half of stroke survivors have weak arm/hand movement affecting their independence and quality of life. Functional Electrical Stimulation (FES) is a technology used for stroke rehabilitation. It applies a safe electrical impulse through electrodes placed on the skin to strengthen weak muscles and improve movement functions. FES has been used to exercise muscles and assist walking for people with mobility issues since the 1960s [3]. Systematic reviews with meta-analysis have concluded that FES improves the ability to perform activities [4,5,6]. Existing FES products are difficult to set-up by stroke survivors without help from their carers or healthcare professionals which significantly constrains usage. Our previous work has received positive feedback regarding wearable e-textile FES for home-based stroke rehabilitation [7]. This work presents the test results of a fabric electrode based wearable FES in terms of user comfort, stimulation intensity, and functional movement (wrist extension for hand opening) on five stroke survivors (ethics approval ID: University of Southampton ERGO 70296). Ease of use has been assessed. A second knitted design has been assessed and additional new designs to improve the usability have been proposed for future study.

2. Materials and Methods

2.1. Electrodes

Wearable electrodes (5 cm × 5 cm) were fabricated by stacking in turn: a non-woven fabric, conductive wires leading to a connector, a conductive carbon film and a carbon rubber electrode layer. Encapsulating the edges holds the entire assembly together. The electrode was attached to a fabric arm band as shown in Figure 1. The wearable electrodes were compared with commercial high quality hydrogel electrodes (PALS, Axelgaard, Fallbrook, CA, USA).

2.2. FES

The OML Microstim 2V2 neuromuscular stimulator (Figure 2) was used in this study. Stimulation was set-up by a clinician to optimise the electrode positions and stimulation intensity. Water was sprayed on the electrodes and the skin before applying the fabric electrode to improve user comfort and stimulation effectiveness. Stimulation comfort was rated in a scale from 0 to 10 with 0 is the most comfortable and 10 being very painful. The stimulation intensity was recorded.

2.3. Inertial Measurement Unit (IMU) Sensors for Wrist Bending Measurement

The wrist angle was measured using a pair of BNO055 IMU sensor breakouts from Adafruit (Figure 3) which reports the absolute angular position of the sensor. The IMUs were attached to the user by using Velcro straps with one attached to the hand and the other to the arm. An Arduino Micro was used to gather the data from the sensors and sent it to the computer recording the data. The computer then calculated the wrist angle by taking the difference between the two absolute positions from the sensors.

2.4. Usability Test

Participants were asked to put on and take off both the hydrogel electrodes and fabric electrode arm bands to assess their capability of using the products independently.

3. Test Results and Discussion

Five participants were recruited for the testing. Age: 56–76, Years of stroke: 3–17 years. Genders: 1 female and 4 males.

3.1. Stimulation Comfort

All participants reported the same or similar stimulation comfort with only 0 to 1 score difference between the two electrode types as shown in Table 1. One participant (P1) reported the fabric electrode was more comfortable than the hydrogel electrodes. One participant (P4) reported the same comfort scale. The other three reported the hydrogel electrodes were more comfortable than fabric electrodes. No pain sensation was reported.

3.2. Stimulation Intensity

There was no difference in the stimulation level required to achieve full hand opening between the two types of electrodes (Table 2). The required stimulation levels vary from 40 mA to 50 mA. This indicates the fabric electrodes were as effective as the hydrogel electrodes in generating a functional movement.

3.3. Wrist Bending Measurement

Wrist movement was measured during the stimulation. All participants achieved a similar movement for the two types of electrodes. Figure 4 is a representative example of the wrist bending angle for the two electrode types.

3.4. Usability

Researchers observed that it was a challenging task for participants to peel off the hydrogel electrodes from the protective plastic film because the hydrogel electrodes are very sticky. It was even more challenging for them to take out the electrodes from the sealed bag and put them back after use. With the electrode arm band, although all participants were able to put it on and take it off independently, they found it challenging to keep the electrode in place because it moved around before it was fastened and secured in place. In addition, it was challenging to put the Velcro hook fastener through a loop. All participants were able to spray water on the electrodes and the arm using a single hand.

3.5. Second Design: Knitted Sleeve

Two pairs of electrodes were printed on a knitted fabric made of wool and Lycra yarns to stimulate muscles for both hand extension and flexion (Figure 5). The electrode fabric was assembled to form a pull-on sleeve. It was noticed the electrode sleeve was difficult to put on or take off because of the strong friction between the electrodes inside the sleeve and the skin. Discussions with the stroke survivors have indicated that an open, or partially open structure, would allow the user to put on the electrode sleeve easily, then tighten the sleeve to ensure the electrodes and skin contact sufficiently.

4. Conclusions and Future Work

This work has demonstrated the stimulation comfort and functional movement delivered by a wearable e-textile activated with the OML FES stimulator Microstim 2V2. The fabric electrode achieved similar stimulation performance while providing the advantage of being suitable for wearable application and offering a long service life. All participants were able to put on and take off the fabric electrode arm bands independently, but it is time consuming and a better design is required. A new design with electrodes printed on a knitted stretchable sleeve was investigated but it was difficult to put on and take off due to the friction between the electrodes inside the sleeve and the skin. New designs, required to improve the ease of use, will be addressed in future work.

Author Contributions

Conceptualization, M.L., J.T. and K.Y.; methodology, M.L., T.W., O.K., P.T. and K.Y.; software, T.W.; validation, M.L., Y.Y., P.T. and K.Y.; formal analysis, T.W. and K.Y.; investigation, M.L. and K.Y.; resources, P.T. and K.Y; writing—original draft preparation, M.L. and K.Y.; review and editing, All; funding acquisition, J.T. and K.Y. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the SIAH HEIF Research Stimulus Fund at the University of Southampton and the Medical Research Council (MRC) under grant number MR/N027841/1.

Institutional Review Board Statement

The study was approved by the Faculty of Arts and Humanities Ethics Committee at the University of Southampton on 21/02/2022 with a submission ID of ERGO 70296.

Informed Consent Statement

Informed consent was obtained from all subjects involved in the study.

Data Availability Statement

No new date was created.

Acknowledgments

The authors want to thank the stroke survivors at Different Strokes Southampton for attending the focus group studies and prototype testing.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Stroke Statistics, Stroke Association. Available online: https://www.stroke.org.uk/what-is-stroke/stroke-statistics (accessed on 4 December 2022).
  2. Patel, A.; Berdunov, V.; Quayyum, Z.; King, D.; Knapp, M.; Wittenberg, R. Estimated societal costs of stroke in the UK based on a discrete event simulation. Age Ageing 2020, 49, 270–276. [Google Scholar] [CrossRef] [PubMed]
  3. Vincent, G. Electrical Control of Partially Denervated Muscles. U.S. Patent 2737183A, 6 March 1956. [Google Scholar]
  4. Veerbeek, J.M.; van Wegen, E.; van Peppen, R.; van der Wees, P.J.; Hendriks, E.; Rietberg, M.; Kwakkel, G. What Is the Evidence for Physical Therapy Poststroke? A Systematic Review and Meta-Analysis. PLoS ONE 2014, 9, e87987. [Google Scholar]
  5. Howlett, O.A.; Lannin, N.A.; Ada, L.; McKinstry, C. Functional Electrical Stimulation Improves Activity after Stroke: A Systematic Review With Meta-Analysis. Arch. Phys. Med. Rehabil. 2015, 96, 934–943. [Google Scholar] [CrossRef] [PubMed]
  6. Sun, J.; Yan, F.; Liu, A.; Liu, T.; Wang, H. Electrical Stimulation of the Motor Cortex or Paretic Muscles Improves Strength Production in Stroke Patients: A Systematic Review and Meta-Analysis. J. Inj. Funct. Rehabil. 2020, 13, 171–179. [Google Scholar] [CrossRef] [PubMed]
  7. Liu, M.; Ward, T.; Keim, O.; Yin, Y.; Tudor, M.; Yang, K. Smart textile with integrated functional electrical stimulation and movement sensor for hand exercise. In Proceedings of the Annual Conference of the International Functional Electrical Stimulation Society (IFESS), RehabWeek, Rotterdam, The Netherlands, 25–29 July 2022. [Google Scholar]
Figure 1. (a) fabric electrode arm band. (b) fabric electrode arm bands being worn by a stroke survivor. (c) hydrogel electrodes being worn by a stroke survivor.
Figure 1. (a) fabric electrode arm band. (b) fabric electrode arm bands being worn by a stroke survivor. (c) hydrogel electrodes being worn by a stroke survivor.
Engproc 30 00016 g001
Figure 2. OML Microstim 2V2.
Figure 2. OML Microstim 2V2.
Engproc 30 00016 g002
Figure 3. BNO055 IMU sensor.
Figure 3. BNO055 IMU sensor.
Engproc 30 00016 g003
Figure 4. Wrist bending angle.
Figure 4. Wrist bending angle.
Engproc 30 00016 g004
Figure 5. (a) electrodes printed on knitted fabric. (b) electrode fabric with snap buttons. (c) electrode sleeve worn on an arm and connected to a stimulator via snap buttons.
Figure 5. (a) electrodes printed on knitted fabric. (b) electrode fabric with snap buttons. (c) electrode sleeve worn on an arm and connected to a stimulator via snap buttons.
Engproc 30 00016 g005
Table 1. Stimulation comfort scale results by participant.
Table 1. Stimulation comfort scale results by participant.
ParticipantHydrogel ElectrodeFabric Electrode
P132
P22–33
P356
P455
P52–33–4
Table 2. Stimulation intensity required to achieve hand opening.
Table 2. Stimulation intensity required to achieve hand opening.
ParticipantHydrogel ElectrodeFabric Electrode
P150 mA50 mA
P240 mA40 mA
P340 mA40 mA
P450 mA50 mA
P545 mA45 mA
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Liu, M.; Ward, T.; Keim, O.; Yin, Y.; Taylor, P.; Tudor, J.; Yang, K. Design and Test of E-Textiles for Stroke Rehabilitation. Eng. Proc. 2023, 30, 16. https://doi.org/10.3390/engproc2023030016

AMA Style

Liu M, Ward T, Keim O, Yin Y, Taylor P, Tudor J, Yang K. Design and Test of E-Textiles for Stroke Rehabilitation. Engineering Proceedings. 2023; 30(1):16. https://doi.org/10.3390/engproc2023030016

Chicago/Turabian Style

Liu, Meijing, Tyler Ward, Odina Keim, Yuanyuan Yin, Paul Taylor, John Tudor, and Kai Yang. 2023. "Design and Test of E-Textiles for Stroke Rehabilitation" Engineering Proceedings 30, no. 1: 16. https://doi.org/10.3390/engproc2023030016

Article Metrics

Back to TopTop