Article

Multiwavelength Frequency Modulated CW Ladar: The Effect of Refractive Index

by
Mariano Barbieri
1,
Deborah Katia Pallotti
1,2,
Mario Siciliani de Cumis
1,2 and
Luigi Santamaria Amato
1,3,*
1
ASI, Italian Space Agency, Space Geodesy Center “G. Colombo”, 75100 Matera, Italy
2
CNR-INO, Istituto Nazionale di Ottica, Largo E. Fermi 6, 50125 Firenze, Italy
3
CNR-INO, Istituto Nazionale di Ottica, Via Campi Flegrei 34, 80078 Pozzuoli, Italy
*
Author to whom correspondence should be addressed.
Photonics 2020, 7(4), 90; https://doi.org/10.3390/photonics7040090
Submission received: 5 August 2020 / Revised: 15 September 2020 / Accepted: 4 October 2020 / Published: 8 October 2020

Abstract

:
Frequency modulated continuous wave (FMCW) laser detection and ranging is a technique for absolute distance measurements with high performances in terms of resolution, non-ambiguity range, accuracy and fast detection. It is based on a simple experimental setup, thus resulting in cost restraint with potential wide spread, not only limited to research institutions. The technique has been widely studied and improved both in terms of experimental setup by absolute reference or active stabilization and in terms of data analysis. Very recently a multi-wavelength approach has been exploited, demonstrating high precision and non ambiguity range. The variability of refractive index along the path was not taken into account with consequent degradation of range accuracy. In this work we developed a simple model able to take into account refractive index effect in multi-wavelength FMCW measurement. We performed a numerical simulation in different atmospheric conditions of temperature, pressure, humidity and CO2 concentration showing a net improvement of range accuracy when refractive index modeling is used.

1. Introduction

Precision measurement of non-cooperative target by laser ranging (LR) techniques had and continues to have applications in several research fields: from satellites flying in defined formation for large synthetic aperture telescope [1,2,3,4] used for fundamental physics test or extraterrestrial planets search to earth based laser ranging. The earth based satellite laser ranging was successfully demonstrated in 1964 and, since then, produced a huge amount of results in fundamental physics experiments like: general relativity test (gravitomagnetic effect [5], Lense-Thirring [6,7] non-Newtonian gravity [8]) or dark matter search [9,10]. However, it mostly finds application in fields like geodesy and geodynamics, allowing, for example, the length-of-day determination or earth gravitational field measurement [11] and contributing to the definition and the update of the terrestrial reference frame (TRF). In addition, there are perspectives in space debris applications and multipurpose communication experiments. The precise distance determination of near object has also several applications mostly in industrial fields [12], particularly in automotive [13,14]. The main characteristics of laser ranging measurements are: resolution of the detected length, non-ambiguity range and accuracy. A LR system is based on a cw or pulsed laser source, collimation optics, a detector and a timing system. A system based on pulsed source has large non-ambiguity range but low resolution whereas a cw system shows a very small non-ambiguity range (of the order of the wavelength) but extraordinary resolution. The accuracy, instead, does not depend on the particular laser source used. First of all, it is related to experimental setup calibration (e.g., linearity of frequency sweep, electrical cable length calibration or accuracy of timing system). Moreover, different environmental factors may influence the accuracy: one of the main contributor is the refractive index. A great advance was the introduction of multi-wavelength interferometry [15,16] that combines measurements at different wavelengths, generating a longer ‘synthetic wavelength’ and, therefore, a larger ambiguity range while maintaining same resolution. Unfortunately, extending the ambiguity range beyond a millimetre is very hard. In 2009 Coddington et al. [17] used dual coherent frequency combs approach for ranging experiments, obtaining at same time large non-ambiguity range, exceptional resolution and fast refresh rate. On the downside, the setup used is very complex and expensive and so available only in state-of-the-art laboratories. For this reason a very popular method that keeps costs low but good performance is the frequency modulated continuous wave ranging (FMCW) [18,19,20]. In this technique a cw laser is chirped at a constant rate K (Hz/s) and split in two parts: reference and probe beam. The probe beam is collimated and sent to object whose distance must be measured, and the reflection is mixed with reference beam on a photodetector. The electric signal is a beat note at frequency f = 2 L k / c where L is the optical path, and c the speed of light. The performance of FMCW depends on the coherence length of the used laser (related to spectral full width half maximum) and on total laser frequency excursion [21] that, in turn, depends on chirp rate K. The accuracy is related to linearity of chirp rate and to the equation used to model refractive index. The non-linearity of the chirp rate can be compensated in an active way using a real time detection system and actuators [22,23]. Another way is a post-processing correction by using a fine calibration up to 15 ppb accuracy level by using a frequency comb [24,25,26] or ppm level with a simple and cheaper reference against a molecular spectrum [27].
Recently multi-wavelength approach has been applied to FMCW [28,29] demonstrating high precision and non ambiguity range. In these experimental implementations the variability of refractive index is not considered. Consequently, even if the range uncertainty is low, the overall accuracy of range measurement is large since it does not consider the refractive index that masks the absolute range value introducing a systematic uncertainty. In this paper we develop a model to take into account the air refractive index in a FMCW multi-wavelength approach. The multiple observables introduced in multi-wavelength approach allow to measure not only the range but also the average atmospheric parameters (like pressure, temperature, humidity and CO2 concentration) along the path. Such parameters allow to calculate a sort of effective refractive index along the path enabling a more accurate estimation of real absolute distance. In other words, in single wavelength FMCW the atmospheric parameters are measured at laser station and used to calculate the refractive index. In this way the refractive index along the path is approximated with the one at laser station. The model developed in this paper allow to measure the effective refractive index along the path in multiwavelength FMCW. Consequently, the refractive index along the path is approximated with the effective refractive index (and not with index at laser station) with consequent improved accuracy. Finally a comparison between two possible atmospheric conditions and a comparison between possible multiple wavelengths has been performed.

2. Model

A general equation for object range (R) detection via beat frequency measurement of a chirped signal (linearly-chirped with constant K) is
{ (1) R 1 + n ( x , λ j ) = c 2 K · f j , j = 1 , , m (2) x = ( t [ C ] , p [ Pa ] , h [ % ] , x c [ ppm ] ) (3) λ j [ μ m ] = j - th laser wavelength , f j = j - th beat frequency observed
j means that we are applying the same equation for R but using different values of laser wavelength λ j hence measuring a different possibly specific beat frequency f j .
Following the formula proposed by Ciddor [30] for the excess component in air n ( x , λ ) compared to the refractive index of the vacuum, the former depends on λ j (from 0.3 to 1.69 μm in [30], formula validity extended in mid infrared [31]), t the temperature in Celsius ( 40 to + 100 C ), p the pressure in Pascal (80 to 120 kPa), h the fractional humidity in the [ 0 , 1 ] interval, x c the CO2 concentration (from 0 to 2000 ppm). Please see https://emtoolbox.nist.gov/Wavelength/Documentation.asp for detailed explanation of the Ciddor formula used.

3. First order Taylor Approximation

Using first order Taylor approximation around a known refractive index value n ( x ( 0 ) , λ j ) , and using the notation of sum over repeated indexes:
{ (4) R c j 0 + c j i · y i = c 2 K · f j (5) c j i = n ( x , λ j ) x i x ( 0 ) , i = 1 , , l (6) y i = x i x i ( 0 ) , x i i - th element of x (7) c j 0 = 1 + n ( x ( 0 ) , λ j )
In case of the Ciddor formula l = 4 , as we can see in Equation (2). Please notice that if we know c i j , i.e., x ( 0 ) and the derivatives of n, and if we have l + 1 = m , we can invert the equations isolating known and unknown quantities:
{ (8) P = R , R · y 1 , , R · y l (9) C = c 10 c 11 c 1 l c m 0 c m 1 c m l (10) F = c 2 · K f 1 , , f m (11) C · P T = F T P T = C 1 · F T , C 1 ( a ) i j
In the last equation we have inverted the linear system generated by the Taylor expansion of n ( C 1 is in fact the inverse of matrix C ). In Equation (11) we have redefined the index convention for ease of notation in the following treatment. The elements of C 1 are redefined as a i j , where now j sums up with the beat frequencies in F while i identifies the components of the parameters vector P where the range R and the atmospheric parameters x are stored. Such convention will be kept from now on unless otherwise stated.
From the above we can find R and x values in terms of only known parameters:
P i = a i j · F j P 1 = R = a 1 j · F j = a 1 j · c 2 · K f j , i = 1 x i 1 = a i j · F j / R + x i 1 ( 0 ) = a i j · c 2 · K f j / R + x i 1 ( 0 ) , i = 2 , , m
So the variance for R and x i will be (covariances excluded):
{ (13) σ R 2 = c 2 2 · ( a 1 j / K ) 2 σ f j 2 + ( a 1 j f j / K 2 ) 2 σ K 2 (14) σ x i 2 = c 2 2 · ( a i j / ( K · R ) ) 2 σ f j 2 + ( a i j f j / ( K 2 · R ) ) 2 σ K 2 + a i j f j K · R 2 2 σ R 2

4. Effect of Refractive Index Variability

In Equation (1) we assumed the refractive index has a constant value along the laser trajectory, depending only on the constant values of the atmospheric parameters x . In this section we start relaxing this assumption and provide an estimation of the error due to the refractive index variability.
We can think of the refractive index variation along the light path as an accumulation of d f j in Equation (1) along the way. Namely
{ (15) d f j = 2 K c · 1 + n ( x , λ j ) d r (16) f j ( R ) = R d f j
the longer the path, the bigger will be the beat frequency observed. To test the performance of the proposed approach, we evaluate the integral Equation (16) using simplified models for the variation of atmospheric parameters x between the target and the laboratory where the measure is performed. In the next section we consider the case of x fluctuating around an average value; this simplifies the analytical treatment and the exact solution, allowing for direct estimation over the entire validity range of Ciddor formula. Then we generalize to the case of a linear gradient of x between lab and target, solving numerically integral Equation (16) also for very extreme differences between lab and target in order to test the performance of the proposed approach in limit cases of input variability. We also sketch up the analytical treatment in the general case when x has an unknown three-dimensional spatial distribution to be chosen.

4.1. Refractive Index Fluctuation around an Average Value

In general the atmospheric variables x have a three-dimensional distribution in space. For instance, increasing altitude of the target means the light trajectory passes through a gradient of temperature; same usually happens with pressure and the other parameters x . In this section we will assume that the variability of n is dependent on fluctuations of the x spatial distribution around an average value with a certain distribution g. This allows an easier analytical treatment of Equation (16) producing a simple formula to test the performance of the proposed approach that can be evaluated over the entire x validity range of Ciddor formula. Such an assumption can be mathematically stated as follows:
R n ( r ) d r R d r g ( n ) n d n R · E [ n ]
Hence from Equation (16)
(18) f j ( R ) = 2 K c R 1 + n ( x ( r ) , λ j ) d r (19) = 2 K c R + R n ( x ( r ) , λ j ) d r (20) = 2 K c R + R d r g j ( n ) n d n (21) = 2 K c R 1 + E [ n j ]
This last equation gives an estimation of the beat frequency when R is known. We can substitute it in Equation (12)-dropping the first index for simplicity of notation so that now a 1 j is a j
(22) R * = a j f j · c 2 K = a j · R 1 + E [ n j ] (23) σ R * r e l = R * R R = j a j + a j · E [ n j ] 1
σ R * r e l can be seen as an estimate of the relative error due to the usage of Equation (12) for range estimation R * . Notice that even when E [ n j ] goes to zero, such error does not necessary go to zero as well. E [ n j ] could be estimated as n ( E [ x ] , λ j ) ; the a j are estimated from the derivatives x i n applied at x ( 0 ) = E [ x ] .
Similar results can be found for the x quantities substituting Equation (21) into Equation (12):
(24) x i 1 * = a i j · c 2 · K f j / R + x i 1 ( 0 ) = a i j · 1 + E [ n j ] + E [ x i 1 ] (25) x i 1 ( 0 ) E [ x i 1 ] (26) σ x * r e l = x i 1 * E [ x i 1 ] E [ x i 1 ] = 1 E [ x i 1 ] j a i j + a i j · E [ n j ]
Now we evaluate σ R * r e l varying the x i values across the range reported in Equation (2) and by calculating E [ n j ] . We see that σ R * r e l is never higher than 10 11 :
max x ( σ R * r e l ) < 1.5 · 10 12
In addition to the above uncertainties, the National Institute of Standards and Technology (NIST) estimates the error using the empirical Ciddor formula, that is unrelated to the uncertainties of the input atmospheric parameters. This takes the form of an adjustment to the formula accounting for various effects (please see https://emtoolbox.nist.gov/Wavelength/Documentation.asp#AppendixAV for detailed explanation and formula). By direct evaluation we estimate the error to be ∼10 8 , while assuming higher values only at the very extreme values of the formula’s validity.

4.2. General Case

In the general case, Equation (17) is not valid anymore and we need to consider the spatial distribution of n:
R n ( r ) d r Trajectory L n ( x ( l ) ) d l
and we can rewrite the equations above this way
{ (29) f j ( R ) = 2 K c L 1 + n ( x ( l ) , λ j ) d l (30) = 2 K c R 1 + L n ( x ( l ) , λ j ) R d l (31) = 2 K c R 1 + N j ( L , R ) (32) σ R * r e l = j a j + a j · N j ( L , R ) 1 (33) where N j ( L , R ) L n ( x ( l ) , λ j ) R d l

4.3. Linear Variation of Atmospheric Parameters

In order to have a numeric estimation of the error σ R * r e l in case of atmospheric parameters x that change values between the observer (the laboratory where the laser is positioned) and the target, we have chosen a specific value of the range R, a specific linear variation of x with light trajectory between the lab and the target, and performed numerically the integral in Equation (16). This provides an estimation of the real observed beat frequency f j .
Then we used Equation (12) to measure the range as per the multi-frequency method above exposed, and compared this prediction R * with the known value R. To this aim, we assigned to x ( 0 ) the x values as measured in the lab.
Similarly we used f j also to estimate the range in the single-frequency classical case R * * , using Equation (1), and compared it with the known R. For this case the refractive index used is calculated from the x in the lab conditions as well.
The results of the relative error are reported in Table 1. In the table we also report the example wavelengths of a Nd:YAG laser and its harmonics generated in nonlinear crystals. They produce beat frequencies of f b e a t = 13.5483 , 13.5484 , 13.5485 , 13.5487 , 13.5489 GHz for R = 10,101 m, with a chirp rate K = 201 MHz/μs, the same order of magnitude as in [27]. Similar relative errors are found for R = 100,301 m. Table 1 shows two sets of range measurements obtained from two linear variations of atmospheric parameters between lab and target. The two sets are chosen in order to have a large excursion in the lab-vs-target atmospheric value (case 1) and a more mild condition (case 2). The atmospheric parameters are shown in Table 2.
For λ = 1.0645 μ m and more extreme variation (case 1), the refractive index between the lab and the target varies from 1.00020745 to 1.000342853. Ps = 101,325 Pa is the standard atmospheric pressure. For completeness, we calculated the expected error using multiple wavelength generated by electro-optic modulator (see Appendix B). In this case the vicinity of the generated wavelengths limits the performance of the multi-wavelength approach. Furthermore also the error due to the numerical inversion of matrix C needs to be addressed. In fact matrix C is close to a singular matrix, due to the terms with the derivatives of n being very small compared with the terms with n. Hence the system is solved numerically, with a certain error on the solution. As described in Appendix A, this error turns out to be small compared with the errors above, and can be neglected.

5. Conclusions

In the last two decades, laser ranging techniques have been improved, due to fast development of frequency metrology tools [32,33,34] and performing laser sources (continuous wave [35,36,37,38,39] and pulsed [40,41]). Such enhancements affect particularly performances and reliability for end users. Multi-wavelength approach is widely used in interferometry and ranging fields with pulsed laser and recently has been applied to FMCW. In this context is fundamental a careful study of the refractive index variation due to atmospheric parameters (i.e., pressure, temperature, humidity and CO2 concentration). In our approach we focused on the accuracy of length estimation and we performed a numerical simulation in order to compare with the single wavelength approach. The result is an improvement on the accuracy limited only by accuracy of Ciddor formula. In addition, a careful modeling of the atmospheric influence on a continuous wave laser beam (amplitude or frequency modulated or even not modulated) enables new applications in the communication, meteorological and metrological fields.

Author Contributions

Conceptualization, L.S.A.; simulation and model development, M.B.; validation and discussion of the results, M.S.d.C. and D.K.P. All authors have read and agreed to the published version of the manuscript.

Funding

This research has been supported by: the Italian Space Agency through projects MOST and WhiTech and fruitful interaction with projects “OT4CLIMA” (D.D. 2261 del 6.9.2018, PON R&I 2014-2020 e FSC) from Ministero della Pubblica Istruzione and project QOMBS from EU commission.

Conflicts of Interest

The authors declare no conflict of interest.

Appendix A. Errors Due to Numerical Evaluation

The system (11) is more precisely solved to find P by means of numerical methods than by calculating the inverse C 1 , due to the high degree of singularity of matrix C . There are different techniques to do so. The technique, among those we tried, that finds the solution P * of the system (11) producing the smallest error on the vector F is the LU decomposition method. If ϵ is the error on F produced by the decomposition we can write:
C · P * T = F T + ϵ
To perform LU decomposition we used the numpy linalg.solve function that relies on the fortran LAPACK library-one of the standards for high precision and performance scientific computing [42,43], that returns the error ϵ 10 12 m for every component. We can use ϵ to propagate the relative error ϵ R on the range R. In fact we can manipulate the first equation of system (11) as follows
(A2) R ( c 10 + i = 1 l c 1 i y i ) = F 1 d R d F 1 = ( c 10 + i = 1 l c 1 i y i ) 1 (A3) ϵ R = 1 R d R d F 1 · ϵ 1 = ( c 10 + i = 1 l c 1 i y i ) 1 R · ϵ 1 10 16
since R 10 4 m and ( c 10 + i = 1 l c 1 i y i ) 1 1 .
The method proposed in this paper has been implemented as a python script, with the usage of standard routines of the SciPy ecosystem [43]. No proprietary and specifically dedicated software was needed, neither any particular setting. The numerical evaluation of integral in Equation (16) has been interpolated over 1000 points implementing a classical trapezoidal algorithm. No significant change is observed for higher number of points. Scripts and materials can be made available upon request to the authors.

Appendix B. Multiple Frequencies Generation by Means of Electro-Optic Modulator

The use of an electro-optic modulator is an alternative approach with respect to a non linear crystal for the generation of multiple frequencies from one laser frequency. These frequencies are symmetrically spaced by a fixed gap around a central frequency. A gap value of 40 GHz is now available in commercial devices. This produces very close values of the wavelength and increases the relative error σ R * r e l . In Figure A1 we report how the relative error decreases with the increase of the frequency gap for a central wavelength of 1.55 μ m. We can see that only for high not experimentally feasible gaps the error is sufficiently reduced. A slightly smaller, but quite similar, error is seen using a wavelength of 1.0645 μm.
Figure A1. σ R * r e l relative error (absolute value) vs. frequency gap via electro-optic modulator (central wavelength 1.55 μm). Different colors correspond to different number of frequencies used by the proposed approach to range estimation. It can be seen that for 3 and 4 (red and green colors) frequencies the method reaches good levels of sensitivity only for unfeasible high gap values, while for 2 frequencies only (blue color) the relative error never significantly declines in the interval considered, looking like a constant value in comparison to the other cases.
Figure A1. σ R * r e l relative error (absolute value) vs. frequency gap via electro-optic modulator (central wavelength 1.55 μm). Different colors correspond to different number of frequencies used by the proposed approach to range estimation. It can be seen that for 3 and 4 (red and green colors) frequencies the method reaches good levels of sensitivity only for unfeasible high gap values, while for 2 frequencies only (blue color) the relative error never significantly declines in the interval considered, looking like a constant value in comparison to the other cases.
Photonics 07 00090 g0a1

References

  1. Cash, W.; Shipley, A.; Osterman, S.; Joy, M. Laboratory detection of X-ray fringes with a grazing-incidence interferometer. Nature 2000, 407, 160–162. [Google Scholar] [CrossRef] [PubMed]
  2. Turyshev, S.G.; Shao, M. Laser astrometric test of relativity: Science, technology and mission design. Int. J. Mod. Phys. D 2007, 16, 2191–2203. [Google Scholar] [CrossRef] [Green Version]
  3. Fridlund, M. Future Space Missions to Search for Terrestrial Planets. Space Sci. Rev. 2008, 135, 355–369. [Google Scholar] [CrossRef]
  4. White, N. Imaging black holes. Nature 2000, 407, 146–147. [Google Scholar] [CrossRef]
  5. Nordtvedt, K. Gravitomagnetic interaction and laser ranging to Earth satellites. Phys. Rev. Lett. 1988, 61, 2647–2649. [Google Scholar] [CrossRef]
  6. Ciufolini, I. Test of General Relativity and Measurement of the Lense-Thirring Effect with Two Earth Satellites. Science 1998, 279, 2100–2103. [Google Scholar] [CrossRef] [Green Version]
  7. Ciufolini, I. Dragging of inertial frames. Nature 2007, 449, 41–47. [Google Scholar] [CrossRef]
  8. Lucchesi, D.M.; Peron, R. LAGEOS II pericenter general relativistic precession (1993–2005): Error budget and constraints in gravitational physics. Phys. Rev. D 2014, 89. [Google Scholar] [CrossRef]
  9. Yang, W.; Leng, J.; Zhang, S.; Zhao, J. Detecting Topological Defect Dark Matter Using Coherent Laser Ranging System. Sci. Rep. 2016, 6. [Google Scholar] [CrossRef] [Green Version]
  10. Tsujikawa, S. Lunar laser ranging constraints on nonminimally coupled dark energy and standard sirens. Phys. Rev. D 2019, 100. [Google Scholar] [CrossRef] [Green Version]
  11. Kaula, W.M. Earth science: The changing shape of the Earth. Nature 1983, 303, 756. [Google Scholar] [CrossRef]
  12. Bosch, T. Laser ranging: A critical review of usual techniques for distance measurement. Opt. Eng. 2001, 40, 10. [Google Scholar] [CrossRef]
  13. Urmson, C.; Anhalt, J.; Bagnell, D.; Baker, C.; Bittner, R.; Clark, M.N.; Dolan, J.; Duggins, D.; Galatali, T.; Geyer, C.; et al. Autonomous driving in urban environments: Boss and the Urban Challenge. J. Field Robot. 2008, 25, 425–466. [Google Scholar] [CrossRef] [Green Version]
  14. Behroozpour, B.; Sandborn, P.A.M.; Wu, M.C.; Boser, B.E. Lidar System Architectures and Circuits. IEEE Commun. Mag. 2017, 55, 135–142. [Google Scholar] [CrossRef]
  15. Dändliker, R.; Thalmann, R.; Prongué, D. Two-wavelength laser interferometry using superheterodyne detection. Opt. Lett. 1988, 13, 339. [Google Scholar] [CrossRef]
  16. Schuhler, N.; Salvadé, Y.; Lévêque, S.; Dändliker, R.; Holzwarth, R. Frequency-comb-referenced two-wavelength source for absolute distance measurement. Opt. Lett. 2006, 31, 3101. [Google Scholar] [CrossRef] [Green Version]
  17. Coddington, I.; Swann, W.C.; Nenadovic, L.; Newbury, N.R. Rapid and precise absolute distance measurements at long range. Nat. Photonics 2009, 3, 351–356. [Google Scholar] [CrossRef]
  18. Uttam, D.; Culshaw, B. Precision time domain reflectometry in optical fiber systems using a frequency modulated continuous wave ranging technique. J. Light Technol. 1985, 3, 971–977. [Google Scholar] [CrossRef]
  19. Cabral, A. Accuracy of frequency-sweeping interferometry for absolute distance metrology. Opt. Eng. 2007, 46, 073602. [Google Scholar] [CrossRef]
  20. Poulton, C.V.; Yaacobi, A.; Cole, D.B.; Byrd, M.J.; Raval, M.; Vermeulen, D.; Watts, M.R. Coherent solid-state LIDAR with silicon photonic optical phased arrays. Opt. Lett. 2017, 42, 4091. [Google Scholar] [CrossRef]
  21. Burdic, W. Radar Signal Analysis, Chapter 5; Prentice-Hall: Upper Saddle River, NJ, USA, 1968. [Google Scholar]
  22. Satyan, N.; Vasilyev, A.; Rakuljic, G.; Leyva, V.; Yariv, A. Precise control of broadband frequency chirps using optoelectronic feedback. Opt. Express 2009, 17, 15991. [Google Scholar] [CrossRef] [PubMed]
  23. Jiang, H.; Kéfélian, F.; Lemonde, P.; Clairon, A.; Santarelli, G. An agile laser with ultra-low frequency noise and high sweep linearity. Opt. Express 2010, 18, 3284. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  24. Baumann, E.; Giorgetta, F.R.; Deschênes, J.D.; Swann, W.C.; Coddington, I.; Newbury, N.R. Comb-calibrated laser ranging for three-dimensional surface profiling with micrometer-level precision at a distance. Opt. Express 2014, 22, 24914. [Google Scholar] [CrossRef] [PubMed]
  25. Barber, Z.W.; Giorgetta, F.R.; Roos, P.A.; Coddington, I.; Dahl, J.R.; Reibel, R.R.; Greenfield, N.; Newbury, N.R. Characterization of an actively linearized ultrabroadband chirped laser with a fiber-laser optical frequency comb. Opt. Lett. 2011, 36, 1152. [Google Scholar] [CrossRef] [PubMed]
  26. Baumann, E.; Giorgetta, F.R.; Coddington, I.; Sinclair, L.C.; Knabe, K.; Swann, W.C.; Newbury, N.R. Comb-calibrated frequency-modulated continuous-wave ladar for absolute distance measurements. Opt. Lett. 2013, 38, 2026. [Google Scholar] [CrossRef] [PubMed]
  27. Mateo, A.B.; Barber, Z.W. Precision and accuracy testing of FMCW ladar-based length metrology. Appl. Opt. 2015, 54, 6019. [Google Scholar] [CrossRef] [Green Version]
  28. Kuse, N.; Fermann, M.E. Frequency-modulated comb LIDAR. APL Photonics 2019, 4, 106105. [Google Scholar] [CrossRef]
  29. Vasilyev, A.; Satyan, N.; Xu, S.; Rakuljic, G.; Yariv, A. Multiple source frequency-modulated continuous-wave optical reflectometry: Theory and experiment. Appl. Opt. 2010, 49, 1932. [Google Scholar] [CrossRef]
  30. Ciddor, P.E. Refractive index of air: New equations for the visible and near infrared. Appl. Opt. 1996, 35, 1566–1573. [Google Scholar] [CrossRef]
  31. Mathar, R.J. Calculated refractivity of water vapor and moist air in the atmospheric window at 10 μm. Appl. Opt. 2004, 43, 928–932. [Google Scholar] [CrossRef] [Green Version]
  32. Cundiff, S.T.; Ye, J.; Hall, J.L. Optical frequency synthesis based on mode-locked lasers. Rev. Sci. Instrum. 2001, 72, 3749–3771. [Google Scholar] [CrossRef] [Green Version]
  33. Bellini, M.; Hänsch, T.W. Phase-locked white-light continuum pulses: Toward a universal optical frequency-comb synthesizer. Opt. Lett. 2000, 25, 1049. [Google Scholar] [CrossRef] [PubMed]
  34. Diddams, S.A.; Jones, D.J.; Ye, J.; Cundiff, S.T.; Hall, J.L.; Ranka, J.K.; Windeler, R.S.; Holzwarth, R.; Udem, T.; Hänsch, T.W. Direct Link between Microwave and Optical Frequencies with a 300 THz Femtosecond Laser Comb. Phys. Rev. Lett. 2000, 84, 5102–5105. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  35. Ricciardi, I.; Mosca, S.; Parisi, M.; Maddaloni, P.; Santamaria, L.; Natale, P.D.; Rosa, M.D. Sub-kilohertz linewidth narrowing of a mid-infrared optical parametric oscillator idler frequency by direct cavity stabilization. Opt. Lett. 2015, 40, 4743. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  36. Ricciardi, I.; Mosca, S.; Maddaloni, P.; Santamaria, L.; Rosa, M.D.; Natale, P.D. Phase noise analysis of a 10 Watt Yb-doped fibre amplifier seeded by a 1-Hz-linewidth laser. Opt. Express 2013, 21, 14618. [Google Scholar] [CrossRef]
  37. Gundavarapu, S.; Brodnik, G.M.; Puckett, M.; Huffman, T.; Bose, D.; Behunin, R.; Wu, J.; Qiu, T.; Pinho, C.; Chauhan, N.; et al. Sub-hertz fundamental linewidth photonic integrated Brillouin laser. Nat. Photonics 2018, 13, 60–67. [Google Scholar] [CrossRef] [Green Version]
  38. Borri, S.; de Cumis, M.S.; Viciani, S.; D’Amato, F.; Natale, P.D. Unveiling quantum-limited operation of interband cascade lasers. APL Photonics 2020, 5, 036101. [Google Scholar] [CrossRef] [Green Version]
  39. Borri, S.; de Cumis, M.S.; Insero, G.; Bartalini, S.; Pastor, P.C.; Mazzotti, D.; Galli, I.; Giusfredi, G.; Santambrogio, G.; Savchenkov, A.; et al. Tunable Microcavity-Stabilized Quantum Cascade Laser for Mid-IR High-Resolution Spectroscopy and Sensing. Sensors 2016, 16, 238. [Google Scholar] [CrossRef] [Green Version]
  40. Kwon, D.; Jeon, I.; Lee, W.K.; Heo, M.S.; Kim, J. Generation of multiple ultrastable optical frequency combs from an all-fiber photonic platform. Sci. Adv. 2020, 6, eaax4457. [Google Scholar] [CrossRef] [Green Version]
  41. Zhang, Z.; Han, H.; Wang, H.; Shao, X.; Fang, S.; Wei, Z. Ultra-low-noise carrier-envelope phase stabilization of a Kerr-lens mode-locked Yb:CYA laser frequency comb with a feed-forward method. Opt. Lett. 2019, 44, 5489. [Google Scholar] [CrossRef]
  42. Anderson, E.; Bai, Z.; Bischof, C.; Blackford, S.; Demmel, J.; Dongarra, J.; Du Croz, J.; Greenbaum, A.; Hammarling, S.; McKenney, A.; et al. LAPACK Users’ Guide, 3rd ed.; Society for Industrial and Applied Mathematics: Philadelphia, PA, USA, 1999. [Google Scholar]
  43. Virtanen, P.; Gommers, R.; Oliphant, T.E.; Haberland, M.; Reddy, T.; Cournapeau, D.; Burovski, E.; Peterson, P.; Weckesser, W.; Bright, J.; et al. SciPy 1.0: Fundamental Algorithms for Scientific Computing in Python. Nat. Methods 2020, 17, 261–272. [Google Scholar] [CrossRef] [PubMed] [Green Version]
Table 1. Relative errors obtained from one ( ( R * * R ) / R ) and multi-wavelength ( ( R * R ) / R ) approach in two different atmospheric conditions (Case 1 and Case 2 as reported in Table 2). The first column of the table reports the number of wavelengths used by the new method, while the second column reports their values.
Table 1. Relative errors obtained from one ( ( R * * R ) / R ) and multi-wavelength ( ( R * R ) / R ) approach in two different atmospheric conditions (Case 1 and Case 2 as reported in Table 2). The first column of the table reports the number of wavelengths used by the new method, while the second column reports their values.
Case 1Case 2
# f beat λ ( μ m ) R ( m ) ( R * R ) / R ( R * * R ) / R ( R * R ) / R ( R * * R ) / R
11.064510,101 6.4 × 10 5 1.8 × 10 5
21.0645, 0.5322510,101 2.5 × 10 5 1.9 × 10 6
31.0645, 0.53225, 0.3548310,101 5.0 × 10 14 3.1 × 10 13
41.0645, 0.53225, 0.35483, 0.26612510,101 5.6 × 10 14 6.6 × 10 14
51.0645, 0.53225, 0.35483, 0.266125, 0.212910,101 1.0 × 10 14 3.2 × 10 14
Table 2. Linear variation of atmospheric parameters x .
Table 2. Linear variation of atmospheric parameters x .
Case 1Case 2
LaboratoryTargetLaboratoryTarget
Temperature (°C)30−202010
Pressure (kPa) P s 20 P s + 10 P s 5 P s + 5
Humidity (fraction)100.80.5
CO2 (ppm)350550350550

Share and Cite

MDPI and ACS Style

Barbieri, M.; Pallotti, D.K.; Siciliani de Cumis, M.; Santamaria Amato, L. Multiwavelength Frequency Modulated CW Ladar: The Effect of Refractive Index. Photonics 2020, 7, 90. https://doi.org/10.3390/photonics7040090

AMA Style