Article

Response Surface Methodology as a Useful Tool for Evaluation of the Recovery of the Fluoroquinolones from Plasma—The Study on Applicability of Box-Behnken Design, Central Composite Design and Doehlert Design

Chair and Department of Physical Pharmacy and Pharmacokinetics, Poznań University of Medical Sciences, Święcickiego 6 Street, 60-781 Poznań, Poland
*
Author to whom correspondence should be addressed.
Processes 2020, 8(4), 473; https://doi.org/10.3390/pr8040473
Submission received: 3 March 2020 / Revised: 10 April 2020 / Accepted: 11 April 2020 / Published: 17 April 2020

Abstract

:
The aim of this study was to find the best design that is suitable for optimizing the recovery of the representatives of the 2nd, 3rd and 4th generation of fluoroquinolones. The following designs were applied: Central Composite Design, Box–Behnken Design and Doehlert Design. The recovery, which was a dependent variable, was estimated for liquid–liquid extraction. The time of shaking, pH, and the volume of the extracting agent (dichloromethane) were the independent variables. All results underwent the statistical analysis (ANOVA), which indicated Central Composite Design as the best model for evaluation of the recovery. For each analyte, an equation was generated that enabled to estimate the theoretical value for the applied conditions. The graphs for these equations were provided by the Response Surface Methodology. The statistical analysis also estimated the most significant factors that have an impact on the liquid–liquid extraction, which occurred to be pH for ciprofloxacin and moxifloxacin and the volume of an extracting solvent for levofloxacin.

1. Introduction

Fluoroquinolones are the antibacterial agents that possess a wide bactericidal activity against Gram-positive (for example S. Pneumoniae) and Gram-negative (for example P. aeruginosa or B. Fragilis) bacteria [1,2]. They can be found in different matrices such as pharmaceutical formulations, biological samples (blood, serum, saliva) or food samples (meat). Different techniques of sample preconditioning such as liquid–liquid extraction, solid-phase extraction, microextraction, dilution, or protein precipitation can be applied. The applied technique depends on the matrix used [3]. Liquid–liquid extraction (LLE) is a very common technique of isolation. The analyte is extracted from a hydrophilic matrix with an organic solvent. The extraction procedure purifies the sample prior to injecting onto the chromatographic system [4]. LLE is more favorable than solid-phase extraction due to its simplicity—the sample is only shaken with an organic solvent. The factors that may influence the analyte’s recovery are pH, the volume of the extracting solvent, and time of extraction. According to FDA guidelines, the recovery should be consistent and reproducible but it does not have to reach the value of 100% [5]. The evaluation of the best conditions for recovery might be time-consuming. In the optimization, the response surface methodology (RSM) in which the surface can take the different shapes, is useful [6]. It is based on mathematical and statistical techniques. The response (dependent variable) is described quantitatively with the polynomial equation. The dependent variable is optimized when several factors are taken into consideration simultaneously. It can also be analyzed as one variable at a time. In this situation, only one parameter is optimized and the others should remain on the same level. It generates too much raw data and it might be impossible to interpret the optimal conditions properly. The application of chemometric design can overcome this problem. The number of experiments is limited and it is possible to estimate the mutual interactions between the independent variables [7,8].
In the optimization process, it is necessary to apply a proper design. The data can be correlated linearly or by the quadratic models. The models that are desirable for the factorial analysis are Box–Behnken Design (BBD), Doehlert Design (DD) and Central Composite Design (CCD), which are classified as response surface designs. They are the symmetrical designs [9]. They make it possible to approximate the response function to experimental data with the quadratic functions. They are useful when the linear functions cannot be applied [8]. These three designs analyze the responses of the independent variables on different levels. In BBB, the experimental points are located on a hypersphere equidistant from the central point. The CCD has the following parts: the points of factorial design, the axial points, and the central point. In the DD, the intervals between the levels present a uniform distribution [8]. The abovementioned models can be found in the literature in optimizing the different extracting procedures such as LLE for 1.4-dihydropyridines [4], ultrasound-assisted extraction for colchicine [10], microextraction for psychoactive substances [11], microvave extraction of the essential oil [12], LLE of fluoroquinolones in food [13], LLE of gallic acid [14] or cloud point extraction in analysis of the metal’s traces [15], stir bar sorptive extraction in the analysis of the volatile compounds in vinegar [16] or liquid-phase extraction of metamizole from biological samples and animal feed [17].
The aim of this work was to find the best design that might be useful in the analysis of the recovery of fluoroquinolones. This paper presents the statistical evaluation of the three factorial designs in the optimization of recovery. It indicates which design is suitable and evaluates the significant interactions between the independent variables.

2. Materials and Methods

2.1. Reagents

Levofloxacin (LEVO), Ciprofloxacin (CIPRO), sodium monophosphate and orthophosphoric acid were purchased by SigmaAldrich (Steiheim, Germany). Moxifloxacin (MOXI) was purchased by SantaCruz Biotechnology (Dallas, TX, USA). Acetonitrile (ACN), methanol (MeOH), triethylamine (TEA), and dichloromethane (DCM) were purchased by Merck (Darmstadt, Germany). All reagents were of HPLC grade. The plasma was obtained from the Local Blood Centre.

2.2. The Design Matrices

The following designs were analyzed: Box–Behnken Design, Doehlert Design, and Central Composite Design. In our study, we applied circumscribed CCD for which │α│ = 1.68 for three factors [9]. They are listed in Table 1, Table 2 and Table 3. The values of each coded factor are presented in Table 4. The method for codification of the values was described by Bezerra et al. [8].

2.3. The Optimization of the Extraction Procedure

Two-hundred microliters of the human blank plasma was spiked with 25 μL of the standard (CIPRO, LEVO or MOXI) stock solution, in which the concentration of the analyte was 100 mg/L, and the 25 μL of internal standard solution (LEVO for CIPRO and MOXI, CIPRO for LEVO determination) which concentration of the stock solution was 60 mg/L. The goal concentration of the internal standard in the plasma sample was 6 mg/L, and the analyte 10 mg/L. Two-hundred-and-fifty microliters of the phosphate buffer (50 mM) was added to the proper volume of DCM (VDCM) defined in Table 4. The proper pH was shifted with a concentrated orthophosphoric acid. The mixture was shaken on a shaker. The organic layer was transferred to the tube and evaporated to dryness in the sample concentrator in 45 °C. The dry residue was dissolved in 200 μL of the mobile phase and injected onto the chromatographic column. The VDCM, the pH of the buffer, and the time of the shaking were the independent variables. The recovery was the response in this study. According to Dejaegher et al. [9], the number of repeats in the central point was three. The samples, for which the recovery was tested for the optimized conditions, were done in five replicates for three concentrations of the analyte in plasma i.e., 2 mg/L, 6 mg/L and 8 mg/L.

2.4. Statistical Analysis

The statistical analysis was done by Design Expert ver. 11, Statistica ver. 13.1 and MiniTab ver. 18. The ANOVA test was applied. The results are presented in Table 5. The p < 0.05 was considered as statistically significant.
The R2 close to 1 implies a better fit design to experimental data, which is also supported with the R2 adjusted. The R2-adjusted value also explains the suitability of the design. It increases with the addition of independent variables, which are significant to a dependent variable. If the non-significant variables are added into the design, the adjusted R2 value decreases contrary to the R2 value that increases with the introduction of the new variables regardless of their importance. The gap between R2 and adjusted R2 should be as small as possible and not exceed 0.2. The importance of F-value implies the significance of the variable in single or in interaction. The higher the F-value observed, the more significant the factor is.

2.5. The Chromatographic Analysis

Twenty microliters of the sample was injected onto the chromatographic system. The analytes were detected with a UV detector (detection wavelength λ = 280 nm), the temperature was ambient. The elution was isocratic. The LiChroCART column (250 × 4 mm, 5 μm, Merck, Darmstadt, Germany) with Purospher RP—18 guard column (4 × 4 mm, 5 μm, Merck, Darmstadt, Germany) were applied for chromatographic separation. The analysis for LEVO and CIPRO was done on the following mobile phase: ACN—MeOH—50mM NaH2PO4—0.7% TEA (12:17:35.5:35.5; v/v/v/v), the pH of the mobile phase was 2.5 and flow 1 mL/min. Retention time for LEVO was ca. 5.4 min., and for CIPRO 6.2 min. The total run time did not exceed 9 min. MOXI was analyzed on the following mobile phase which consisted of ACN—1.5% TEA—33.8 mM NaH2PO4 (27:36.5:36.5; v/v/v), the pH of the mobile phase was 2.5 and flow was 1 mL/min. The retention time for MOXI was ca. 4.8 min., and for LEVO ca. 2.8 min. The total time of analysis did not exceed 7 min. The components and proportions of the constituents of the mobile phase were optimized using Box–Behnken Design. The HPLC separation was isocratic in both cases [18].

3. Results

3.1. The Statistical Evaluation of the Model

The aim of this study was to compare the suitability of the different chemometric models for the optimization of the recovery of the representatives of the fluoroquinolones. The following models were tested: BBD, CCD and DD. The design matrices are presented in Table 1, Table 2 and Table 3. The following parameters were taken into consideration: the p-value, the F-value of the ANOVA test, the coefficients of determination (R2-value, the R2-adjusted), and lack of fit. The p-value should be below 0.05. The F-value should take the highest values, the lack of fit should not be below 0.05. The coefficient of variations R2 is a measure of the amount of variation around the model, and the R2-adjusted is adjusted for the number of terms in the model. It decreases as the number of terms in the model increases on the condition that these terms do not add value to the model. The R2 value should be higher than 0.9 and the differences between them should not exceed 0.2 [18].
The model which was the least suitable for estimation of the recovery was the BBD. The p-values for CIPRO and LEVO were not significant (Table 5). For MOXI, it was on the threshold of statistical significance. The F-values were the lowest when compared with the other models. Moreover, none of the R2 values for each analyte exceeded the value 0.9. The R2-adjusted values were the lowest when compared with other models and the difference between them was higher than 0.2 for CIPRO and LEVO. For MOXI, it was close to this value. The lack-of-fit was significant for all analytes. That excluded the BBD model from further analysis.
The p-values in CCD and DD implied the significance of the models for all analytes, however, for CCD they took the lowest values (below 0.0001). The lack of fit in all cases was not significant. The coefficients of determinations R2 values exceeded 0.95, however, it took the higher values for CCD. The lower values for R2 adjusted were noted for CIPRO and LEVO for DD. For CCD, they were higher—they exceeded the value of 0.94, which revealed a good correlation between the independent variables. The differences between R2 and R2-adjusted did not exceed the value 0.2 for all analytes in CCD and DD models.
The analysis of F-values proved that the CCD was suitable for the optimization of the recovery of analyzed fluoroquinolones—they were the highest for CCD and they exceeded 25. For DD, they did not exceed 21, and the lowest one was noted for MOXI. For BBD, the F-value took the lowest values, which did not exceed 5 (Table 5).
The standardized Pareto Charts for recovery of the analyzed compounds for CCD are in Figure 1. The length of the bar is proportional to the standardized effect [19]. The line indicates the threshold p = 0.05. For CIPRO and MOXI, the most significant factors are the quadratic terms of pH, VDCM and time. For LEVO, the most significant factors are VDCM, the quadratic term of time and the VDCM × time interaction. The mutual interaction between two different independent variables for MOXI and CIPRO compounds is not as significant as for LEVO. The Pareto chart for LEVO indicates as significant all possible mutual interactions between the independent variables.

3.2. Response Surface Methodology (RSM) in the Optimization of Recovery

The statistical analysis for CIPRO and MOXI proved that F-values are the highest for pH2, VDCM2, and time2. For LEVO, the highest values are observed for VDCM, time2, and interaction VDCM × time (Table 6). The statistical analysis also proved that for LEVO all the mutual interactions between independent variables are statistically significant.
The shapes of the RSM graphs for CIPRO, LEVO and MOXI are presented in Figure 2, Figure 3 and Figure 4.
The theoretical equations for each of the analytes are listed below:
RecoveryCIPRO = 91.79 + 9.62 pH + 3.40 VDCM − 12.19 time − 61.33 pH2 − 53.47 VDCM2 − 46.64 time2 − 2.32 pH × VDCM − 14.11 pH × time + 18.34 VDCM × time
RecoveryLEVO = 98.59 − 0.83 pH + 17.75 VDCM + 2.83 time − 6.80 pH2 − 7.28 VDCM2 − 9.17 time2 + 5.11 pH × VDCM + 4.26 pH × time − 8.22 VDCM × time
RecoveryMOXI = 88.29 + 3.29 pH − 8.72 VDCM + 2.08 time − 19.49 pH2 − 17.08 VDCM2 − 15.80 time2 + 1.16 pH × VDCM − 0.94 pH × time − 3.64 VDCM × time

4. Discussion

The analysis of the shapes of the RSM graphs for CIPRO proved that the maximum value of recovery was located inside the experimental region for all configurations (pH and time vs. recovery, VDCM and time vs. recovery, and pH and VDCM vs. recovery). In Figure 2a, we can see that the maximum recovery observed is for pH 6.0–7.0 and the VDCM 2.0–3.5 mL. When pH is considered with the time of shaking, we can see that the maximum recovery is observed for time 8–12 min (Figure 2b). Figure 2c gives an information that the maximum recovery is noted for the volume range of an extracting agent 2.0–3.5 mL. Too high or too low pH reduces the recovery because of the ionization of the carboxyl group or nitrogen atom in the CIPRO molecule. The analysis of the data for CIPRO showed that the maximum recovery is observed for the intermediate values of the variables. The pH 6.0–7.0 combined with the VDCM 2–3 mL and the time of shaking 7.5–12.5 min. resulted in high recovery. The critical parameter that influences most the recovery of CIPRO is pH. The further statistical analysis with the software indicated the following conditions for which the recovery should be the highest: 6.6 for pH, 2.5 mL for VDCM, and 9 min. for the time of shaking. The predicted recovery for CIPRO estimated with equation (1) for optimal conditions should take the value 92.9%. The experimental values were 92.2% ± 0.8%, 92.4% ± 1.0% and 92.9% ± 1.1% for the following concentrations 2 mg/L, 6 mg/L and 8mg/L respectively. The mean value for these three concentrations was 92.5%. It also confirmed the suitability of the applied design.
The analysis of the RSM graph (pH and VDCM vs. Recovery) showed that the maximum recovery of LEVO is observed for VDCM 3.0–4.0 mL and pH 6.0–7.0 (Figure 3a). The major restricting agent of LEVO recovery is the VDCM. It decreases with the decrease in the volume of the extracting solvent. However, the higher volumes of the extracting agent also reduces the recovery. It was also noted by Asadollahzadeh et al. [19], who investigated the impact of an extracting agent on the microextraction of arsenic compounds, that the enrichment factor decreases with the increase of the extracting agent. The impact of the time is illustrated in Figure 3b,c. It indicated that the region with the maximum recovery is observed for the time of shaking 7.5–14.0 min, pH from 6.0 to 7.0, and VDCM 3.0–4.0 mL. When the time of shaking shortens, and the applied VDCM is 2.0 mL and lower, the recovery decreases especially for the higher pH levels. The impact of pH on the recovery is significant for low VDCM. In all graphs, the maximum is located inside the investigated region. The VDCM and pH vs. Recovery, and pH and time vs. Recovery graphs indicate that variation of the pH level impacts the recovery significantly for higher values of pH when considered with low VDCM and short time of shaking. The software analysis indicated that the most optimal conditions for recovery are pH 6.4, VDCM 3.5 mL and time of shaking 13 min. The expected value for LEVO recovery calculated with equation (2) was 101.6%. The observed value for the concentration 2 mg/L was 99.9% ± 0.2%, and for concentrations 6 mg/L and 8 mg/L was 100% ± 0.1%. The mean value for these concentrations was 100%.
The highest recovery for MOXI is observed for pH 6.0–7.0 and the VDCM 2–3 mL (Figure 4a). According to the graph in Figure 4b, the highest recovery is noted for pH 6.0–7.0 and time of shaking 10–12 min. In Figure 4a,c, the maximum of the recovery is stretched to low values of DCM, which suggests the theoretical possibility of the use of relatively small amounts (less than 3 mL) of the organic solvent at the intermediate values of pH (Figure 4a). The most optimal conditions for MOXI recovery are the following: pH 6.6, 2.3 mL for VDCM and 10 min for the time of shaking. The predicted value of MOXI recovery with Equation (3) is 89.5%. The observed values were 90.5% ± 1.1%, 90.9% ± 0.9% and 92.2% ± 0.7% for the following concentrations: 2 mg/L, 6 mg/L and 8 mg/L respectively. The observed mean value for these concentrations is 91.2%. The shapes of the response plots indicated for all analytes that the maximum recovery is located inside the experimental region.

5. Conclusions

According to the literature data, BBD, DD and CCD can be applied in the analysis of the extraction step for different compounds [20]. This study gives an answer to which of the investigated model is suitable for the analysis of the fluorquinolones’ recovery from plasma. The statistical analysis excluded BBD. For DD, the F-values indicated that this design is better than BBD. However, the detailed statistical analysis proved that in the optimization fluoroquinolones, the most suitable is CCD; the highest values for the ANOVA test were noted for CCD. Szerkus et al. also used the CCD in the extraction of levofloxacin—it was an extraction with methanol from prostatic tissue [21]. Ferreira et al. indicated that DD for the k = 3 is the same as BBD [20]. The suitability of the chosen model was confirmed by the measurements—the predicted values of recovery were similar to the calculated ones.
The statistical analysis proved that the pH is the factor that limits the recovery of CIPRO and MOXI. In the case of LEVO, the factor that most influences the recovery is the VDCM.
The conducted statistical analysis not only indicated the most relevant factors that influence the recovery of the analytes but also makes it possible to reduce the use of the reagents. It is a very important issue from both the economic and ecological point of view.

Author Contributions

A.C.—conceived and designed the experiments, analysed the data, contributed the reagents, materials, analysis tools and wrote the paper; A.C. and H.J. prepared the samples for HPLC analysis and did the measurements. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding

Conflicts of Interest

The authors declare no conflict of interest

References

  1. Czyrski, A.; Kondys, K.; Szałek, E.; Karbownik, A.; Grześkowiak, E. The pharmacokinetic interaction between levofloxacin and sunitinib. Pharmacol. Rep. 2015, 67, 542–544. [Google Scholar] [CrossRef] [PubMed]
  2. Czyrski, A.; Anusiak, K.; Teżyk, A. The degradation of levofloxacin in infusions exposed to daylight with an identification of a degradation product with HPLC-MS. Sci. Rep. 2019, 9, 3621. [Google Scholar] [CrossRef] [PubMed]
  3. Czyrski, A. Analytical Methods for Determining Third and Fourth Generation Fluoroquinolones: A Review. Chromatographia 2017, 80, 181–200. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  4. Baranda, A.B.; Etxebarria, N.; Jiménez, R.M.; Alonso, R.M. Development of a liquid-liquid extraction procedure for five 1,4-dihydropyridines calcium channel antagonists from human plasma using experimental design. Talanta 2005, 67, 933–941. [Google Scholar] [CrossRef] [PubMed]
  5. Bioanalytical Method Validation Guidance for Industry; 2018; pp. 1–44. Available online: https://www.fda.gov/files/drugs/published/Bioanalytical-Method-Validation-Guidance-for-Industry.pdf (accessed on 3 March 2020).
  6. Hibbert, D.B. Experimental design in chromatography: A tutorial review. J. Chromatogr. B Analyt. Technol. Biomed. Life Sci. 2012, 910, 2–13. [Google Scholar] [CrossRef] [PubMed]
  7. Sahu, P.K.; Ramisetti, N.R.; Cecchi, T.; Swain, S.; Patro, C.S.; Panda, J. An overview of experimental designs in HPLC method development and validation. J. Pharm. Biomed. Anal. 2018, 147, 590–611. [Google Scholar] [CrossRef] [PubMed]
  8. Bezerra, M.A.; Santelli, R.E.; Oliveira, E.P.; Villar, L.S.; Escaleira, L.A. Response surface methodology (RSM) as a tool for optimization in analytical chemistry. Talanta 2008, 76, 965–977. [Google Scholar] [CrossRef] [PubMed]
  9. Dejaegher, B.; Heyden, Y.V. Experimental designs and their recent advances in set-up, data interpretation, and analytical applications. J. Pharm. Biomed. Anal. 2011, 56, 141–158. [Google Scholar] [CrossRef] [PubMed]
  10. Khodadoust, S.; Sadeghi, H.; Pebdani, A.A.; Mohammadi, J.; Salehi, A. Optimization of ultrasound-assisted extraction of colchicine compound from Colchicum haussknechtii by using response surface methodology. J. Saudi Soc. Agric. Sci. 2017, 16, 163–170. [Google Scholar] [CrossRef] [Green Version]
  11. Ares, A.M.; Fernández, P.; Regenjo, M.; Fernández, A.M.; Carro, A.M.; Lorenzo, R.A. A fast bioanalytical method based on microextraction by packed sorbent and UPLC-MS/MS for determining new psychoactive substances in oral fluid. Talanta 2017, 174, 454–461. [Google Scholar] [CrossRef] [PubMed]
  12. Tran, Q.T.; Vu Thi, T.L.; Do, T.L.; Pham Thi, H.M.; Hoang Thi, B.; Chu, Q.T.; Lai Phuong, P.T.; Do, H.N.; Hoang Than, H.T.; Ta Thi, T.T.; et al. Optimization of Microwave-Assisted Extraction Process of Callicarpa candicans (Burm. f.) Hochr Essential Oil and Its Inhibitory Properties against Some Bacteria and Cancer Cell Lines. Processes 2020, 8, 173. [Google Scholar] [CrossRef] [Green Version]
  13. Gao, M.; Wang, H.; Ma, M.; Zhang, Y.; Yin, X.; Dahlgren, R.A.; Du, D.; Wang, X. Optimization of a phase separation based magnetic-stirring salt-induced liquid-liquid microextraction method for determination of fluoroquinolones in food. Food Chem. 2015, 175, 181–188. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  14. Athankar, K.K.; Wasewar, K.L.; Varma, M.N.; Shende, D.Z. Reactive extraction of gallic acid with tri-n-caprylylamine. New J. Chem. 2016, 40, 2413–2417. [Google Scholar] [CrossRef]
  15. Bezerra, M.A.; Bruns, R.E.; Ferreira, S.L.C. Statistical design-principal component analysis optimization of a multiple response procedure using cloud point extraction and simultaneous determination of metals by ICP OES. Anal. Chim. Acta 2006, 580, 251–257. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  16. Durán Guerrero, E.; Natera Marín, R.; Castro Mejías, R.; García Barroso, C. Optimisation of stir bar sorptive extraction applied to the determination of volatile compounds in vinegars. J. Chromatogr. A 2006, 1104, 47–53. [Google Scholar] [CrossRef] [PubMed]
  17. Ebrahimzadeh, H.; Asgharinezhad, A.A.; Adlnasab, L.; Shekari, N. Optimization of ion-pair based hollow fiber liquid phase microextraction combined with HPLC-UV for the determination of methimazole in biological samples and animal feed. J. Sep. Sci. 2012, 35, 2040–2047. [Google Scholar] [CrossRef] [PubMed]
  18. Czyrski, A.; Sznura, J. The application of Box-Behnken-Design in the optimization of HPLC separation of fluoroquinolones. Sci. Rep. 2019, 9, 19458. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  19. Asadollahzadeh, M.; Tavakoli, H.; Torab-Mostaedi, M.; Hosseini, G.; Hemmati, A. Response surface methodology based on central composite design as a chemometric tool for optimization of dispersive-solidification liquid-liquid microextraction for speciation of inorganic arsenic in environmental water samples. Talanta 2014, 123, 25–31. [Google Scholar] [CrossRef] [PubMed]
  20. Ferreira, S.L.C.; Bruns, R.E.; da Silva, E.G.P.; Dos Santos, W.N.L.; Quintella, C.M.; David, J.M.; de Andrade, J.B.; Breitkreitz, M.C.; Jardim, I.C.S.F.; Neto, B.B. Statistical designs and response surface techniques for the optimization of chromatographic systems. J. Chromatogr. A 2007, 1158, 2–14. [Google Scholar] [CrossRef] [PubMed]
  21. Szerkus, O.; Jacyna, J.; Wiczling, P.; Gibas, A.; Sieczkowski, M.; Siluk, D.; Matuszewski, M.; Kaliszan, R.; Markuszewski, M.J. Ultra-high performance liquid chromatographic determination of levofloxacin in human plasma and prostate tissue with use of experimental design optimization procedures. J. Chromatogr. B Anal. Technol. Biomed. Life Sci. 2016, 1029–1030, 48–59. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Standardized Pareto charts for recovery of CIPRO, LEVO and MOXI.
Figure 1. Standardized Pareto charts for recovery of CIPRO, LEVO and MOXI.
Processes 08 00473 g001
Figure 2. The RSM diagrams illustrating the change of recovery of CIPRO when analyzed: (a) VDCM and pH; (b) time and pH; (c) time and VDCM for CCD.
Figure 2. The RSM diagrams illustrating the change of recovery of CIPRO when analyzed: (a) VDCM and pH; (b) time and pH; (c) time and VDCM for CCD.
Processes 08 00473 g002
Figure 3. The RSM diagrams illustrating the change of recovery of LEVO when analyzed: (a) VDCM and pH; (b) time and pH; (c) time and VDCM for CCD.
Figure 3. The RSM diagrams illustrating the change of recovery of LEVO when analyzed: (a) VDCM and pH; (b) time and pH; (c) time and VDCM for CCD.
Processes 08 00473 g003
Figure 4. The RSM diagrams illustrating the change of recovery of MOXI when analyzed: (a) VDCM and pH; (b) time and pH; (c) time and VDCM for CCD.
Figure 4. The RSM diagrams illustrating the change of recovery of MOXI when analyzed: (a) VDCM and pH; (b) time and pH; (c) time and VDCM for CCD.
Processes 08 00473 g004
Table 1. The Central Composite Design.
Table 1. The Central Composite Design.
Coded Factors
Buffer pHVolume of DCMTime
−1−1−1
1−1−1
−11−1
11−1
−1−11
1−11
−111
111
−α00
α00
0−α0
0α0
00−α
00α
000
Table 2. The Doehlert Design.
Table 2. The Doehlert Design.
Coded Factors
Buffer pHVolume of DCMTime
100
0.50.8660
0.50.2890.817
−100
−0.5−0.8660
−0.5−0.289−0.817
0.5−0.8660
0.5−0.289−0.817
−0.50.8660
00.577−0.817
−0.50.2890.817
0−0.5770.817
000
Table 3. The Box–Behnken Design.
Table 3. The Box–Behnken Design.
Coded Factors
Buffer pHVolume of DCMTime
−1−10
1−10
−110
110
−10−1
10−1
−101
101
0−1−1
01−1
0−11
011
000
Table 4. The levels of the factors.
Table 4. The levels of the factors.
The Coded LevelBuffer pHVolume of DCM (mL)Time (min)
−α4.820.8201.60
−15.501.5005.00
−0.8661.634
−0.8175.92
−0.5771.923
−0.56.00
−0.2892.211
06.502.50010.00
0.2892.789
0.57.00
0.5773.077
0.81714.08
0.8663.360
17.503.50015.00
α8.184.18018.40
Table 5. The statistical analysis of the applied models.
Table 5. The statistical analysis of the applied models.
Central Composite Design
CIPROLEVOMOXI
F29.2727.1448.500
p<0.0001<0.0001<0.0001
R20.97770.97600.9842
R2-adjusted0.94430.94010.9639
Lack of fitNSNSNS
Doehlert Design
CIPROLEVOMOXI
F14.39020.52013.210
p0.0110.0020.019
R20.96640.96360.9824
R2-adjusted0.89930.92620.9689
Lack of fitNSNSNS
Box–Behnken Design
CIPROLEVOMOXI
F2.3203.4304.920
p0.1840.0540.047
R20.80660.86050.8986
R2-adjusted0.48540.60940.7160
Lack of fit0.006<0.0010.018
Table 6. The statistical analysis of CCD for the analytes.
Table 6. The statistical analysis of CCD for the analytes.
CIPROLEVOMOXI
Fp Fp Fp
Model29.27<0.001Model27.14<0.001Model48.50<0.001
linear9.230.012linear67.11<0.001linear22.780.001
pH10.100.019pH0.400.552pH8.090.025
VDCM1.260.305VDCM185.80<0.001VDCM57.02<0.001
time16.230.007time15.150.008time3.230.115
quadratic69.58<0.001quadratic18.510.002quadratic120.46<0.001
pH × pH142.69<0.001pH × pH20.920.004pH × pH234.47<0.001
VDCM × VDCM108.46<0.001VDCM × VDCM16.240.007VDCM × VDCM180.08<0.001
time × time82.52<0.001time × time38.080.001time × time154.15<0.001
2-FI3.010.1162FI12.540.0052-FI2.260.168
pH × VDCM0.110.753pH × VDCM8.790.025pH × VDCM0.590.466
pH × time4.000.092pH × time6.090.049pH × time0.390.555
VDCM × time6.760.041VDCM × time22.740.003VDCM × time5.810.047

Share and Cite

MDPI and ACS Style

Czyrski, A.; Jarzębski, H. Response Surface Methodology as a Useful Tool for Evaluation of the Recovery of the Fluoroquinolones from Plasma—The Study on Applicability of Box-Behnken Design, Central Composite Design and Doehlert Design. Processes 2020, 8, 473. https://doi.org/10.3390/pr8040473

AMA Style

Czyrski A, Jarzębski H. Response Surface Methodology as a Useful Tool for Evaluation of the Recovery of the Fluoroquinolones from Plasma—The Study on Applicability of Box-Behnken Design, Central Composite Design and Doehlert Design. Processes. 2020; 8(4):473. https://doi.org/10.3390/pr8040473

Chicago/Turabian Style

Czyrski, Andrzej, and Hubert Jarzębski. 2020. "Response Surface Methodology as a Useful Tool for Evaluation of the Recovery of the Fluoroquinolones from Plasma—The Study on Applicability of Box-Behnken Design, Central Composite Design and Doehlert Design" Processes 8, no. 4: 473. https://doi.org/10.3390/pr8040473

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.