Article

Properties of Titanium Oxide Coating on MgZn Alloy by Magnetron Sputtering for Stent Application

1
School of Chemistry and Chemical Engineering, Henan Normal University, Xinxiang 453007, China
2
School of Mechanical and Electrical Engineering, Xinxiang University, Xinxiang 453003, China
*
Author to whom correspondence should be addressed.
Coatings 2020, 10(10), 999; https://doi.org/10.3390/coatings10100999
Submission received: 14 September 2020 / Revised: 13 October 2020 / Accepted: 16 October 2020 / Published: 19 October 2020
(This article belongs to the Special Issue Bioactive Coatings for Implantable Devices)

Abstract

:
Constructing surface coatings is an effective way to improve the corrosion resistance and biocompatibility of magnesium alloy bioabsorbable implants. In this present work, a titanium oxide coating with a thickness of about 400 nm was successfully prepared on a MgZn alloy surface via a facile magnetron sputtering route. The surface features were characterized using scanning electron microscopy (SEM), atomic force microscopy (AFM), X-ray diffraction (XRD), X-ray photoelectron spectroscopy (XPS), and the contact angle method. The corrosion behavior and biocompatibility were evaluated. The results indicated that the amorphous TiO2 coating with a flat and dense morphology was obtained by magnetron-sputtering a titanium oxide target. The corrosion current density decreased from 1050 (bare MgZn alloy) to 49 μA/cm2 (sample with TiO2 coating), suggesting a significant increase in corrosion resistance. In addition, the TiO2 coating showed good biocompatibilities, including significant reduced hemolysis and platelet adhesion, and increased endothelial cell viability and adhesion.

1. Introduction

Magnesium alloy (Mg alloy) is an excellent biological material for bioabsorbable implant applications, due to its low corrosion resistance, good biocompatibility, and mechanical properties [1,2]. In recent years, cardiovascular stents made from absorbable Mg alloys have been developed to overcome the drawbacks of permanent metallic stents, including late-term restenosis, delayed re-endothelialization, and persistent inflammation [3,4]. However, the main disadvantage of Mg alloys for medical applications is the rapid and inhomogeneous corrosion in the physiological environment; hence, Mg alloy implants may lose their mechanical integrity before the complete healing of the tissue. In order to solve this problem, surface modifications have been widely carried out to prepare suitable coatings onto Mg alloys as corrosion protective layers, and also to improve their biocompatibility and mechanical stability [5,6,7,8,9]. Therefore, modifying the surface with ideal coatings is of high significance for Mg alloy stents.
For Mg alloy cardiovascular stents, some organic-based coatings, such as polylactic acid (PLA) [7], polydopamine (PDA) [10,11], poly (lactic-co-glycolic acid) (PLGA) [12], and polytrimethylene carbonate (PTMC) [13], have been commonly used as corrosion protection layers or drug delivery carriers to prevent the in-stent restenosis (ISR). However, the adhesion of organic coatings still needs to be improved. On the other hand, inorganic coatings can be prepared by various methods, providing a superior adhesion with substrates. More importantly, inorganic coatings are remarkably corrosion-resistant, and thus, they should be suitable for Mg alloy stents to improve the rapid corrosion behavior. Some inorganic coatings including magnesium fluoride (MgF2) [14] and titanium dioxide (TiO2) [15,16] have been used to endow the Mg alloy with greater corrosion resistance and improved biocompatibility.
It is well known that titanium oxide is chemically stable, and the biocompatibility of a titanium oxide coating has been demonstrated [17,18,19]. In a representative literature, Huang et al. [20] synthesized titanium oxide film on a cobalt alloy by ion beam-enhanced deposition at below 180 °C, and proved that titanium oxide was an excellent blood contact material because of the semiconductor nature. In our previous studies, the anatase titanium dioxide coating with sheet-like nanoscale features was fabricated via the solvothermal route at 160 °C, which improved the corrosion resistance of degradable MgZn alloy [21].
In view of the current research status and progress, there is still a lack of comprehensive evaluation on the application prospects of the titanium oxide-coated absorbable Mg alloy stent. In this paper, titanium oxide was prepared on Mg alloy substrates at room temperature by a facile and clean magnetron sputtering. The comprehensive properties including corrosion resistance, blood compatibility, and endothelial cells (ECs) adhesion of the as-prepared coating were discussed.

2. Materials and Methods

2.1. Preparation of the Coating

In the current study, titanium oxide coatings were deposited onto the surface of biomedical MgZn alloy (composition: 2.0 wt.% Zn, 0.46 wt.% Y, 0.5 wt.% Nd, balance Mg), developed by Zhengzhou University, China [12,22]. Firstly, rectangular samples with dimensions of 10 × 10 × 1 mm3 were cut from the as-cast MgZn ingots by wire-electrode machining, then mechanically polished and sonicated in ethanol for 5 min. Finally, samples were dried before magnetron sputtering. Titanium oxide was deposited on the MgZn samples for 2 h, with a radio-frequency magnetron sputtering system by sputtering a high-purity TiO2 target under a constant power of 150 W. The working pressure was 0.6 Pa with an argon flow of 30 sccm.

2.2. Characterizations

The characteristics of the as-prepared coatings were analyzed by a scanning electron microscope (SEM, Hitachi SU8000, Tokyo, Japan), energy-dispersive spectrometer (EDS), and atomic force microscope (AFM, Bruker MultiMode8, Billerica, MA, USA). The crystal structure was determined by employing an X-ray diffractometer (XRD, PANalytical X’Pert3 Powder, Malvern Instruments, Malvern, UK) with Cu Kα radiation at 45 kV and 40 mA. The valence states of the elements were detected using an X-ray photoelectron spectrometer (XPS). The water contact angle (CA) of samples (n = 3) was measured with 10 μL droplets of deionized water at ambient temperature, using a contact angle meter.

2.3. Corrosion Tests

For corrosion tests, bare and Ti-O-coated MgZn samples (n = 3) were immersed in simulated body fluids (SBFs, containing 8.035 g of NaCl, 0.355 g of NaHCO3, 0.225 g of KCl, 0.231 g of K2HPO4·3H2O, 0.311 g of MgCl2·6H2O, 40 mL of 1.0M-HCl, 0.292 g of CaCl2, 0.072 g of Na2SO4, and 6.118 g of Na2HPO4, in 1000 mL of deionized water, pH = 7.4, 37 °C) [23]. The solution volume-to-specimen area ratio was 40 mL/cm2. After a decent interval, samples were rinsed with deionized water and dried. Then, the corrosion products were analyzed by SEM and XRD.
The electrochemical corrosion behavior of samples (n = 3) was measured through a three-electrode system (CorrTest CS2350, Corrtest Instruments, Wuhan, China). A saturated calomel electrode, platinum electrode, and the sample were respectively utilized as the reference electrode, counter-electrode, and working electrode. The polarization curve of tested samples (with a 1 cm2 area exposed in 150 mL of SBF) was measured with a scanning rate of 0.5 mV/s, after a steady open-circuit potential (OCP) value was achieved.

2.4. Biocompatibility Evaluation

Blood compatibility was evaluated including hemolysis and platelet adhesion. The samples (n = 3) were put into the test tubes containing 10 mL normal saline at 37 °C for 30 min. Subsequently, 0.2 mL of diluted blood was added into every tube and incubated for 60 min at 37 °C. The solution was centrifuged at 3000 rpm for 5 min and the absorbance (ODt) of the supernatant at 545 nm was determined by a microplate reader. Normal saline was used as a negative control (ODnc), and deionized water as a positive control (ODpc). The hemolysis ratio was calculated by the following equation:
Hemolysis   ratio   % = OD t OD nc OD pc OD nc × 100 %
For the platelet adhesion test, fresh human anticoagulant blood was centrifuged at 1500 rpm for 15 min and the upper yellow liquid was separated to obtain platelet-rich plasma (PRP). Then, 0.5 mL of PRP was taken and dropped onto every sample (10 × 10 × 1 mm3, n = 3) in 24-well culture plates, incubated at 37 °C for 120 min. After that, samples were taken out, rinsed with normal saline, and fixed with 0.2% glutaraldehyde solution for 6 h. Finally, samples were washed with normal saline and dehydrated with 50%, 75%, 90%, and 100% ethanol for 15 min, respectively, then dried and examined by SEM.
Human umbilical vein endothelial cells (HUVECs, Ea.hy926, Cell Bank of the Chinese Academy of Sciences, Shanghai, China) were selected to evaluate the cytocompatibility. The sterilized samples (10 × 10 × 1 mm3) were put into 24-well culture plates (n = 3). The cells were cultured in Dulbecco’s modified Eagle medium (DMEM), then digested by 0.25% trypsin and seeded on the surface of the samples at a concentration of 2 × 104 cells/mL. After incubating for 1 and 3 d, 20 μL of MTT solution (5 mg/mL in PBS) was added to the wells for 4 h incubation at 37 °C, then 150 μL of DMSO was added and shaken for 10 min. The absorbance (ODt) of 100 μL of supernatant in 96-well plates was tested by a microplate reader at 490 nm. DMEM solution with 10% serum was used as a negative control. Cell viability was calculated by the following equation:
Viability   % = OD t OD nc   ×   100 %
To observe the adhered cells, samples were taken out and washed with PBS buffer several times, then put into PBS buffer containing 2.5% glutaraldehyde for 12 h at 4 °C, and dehydrated in increasing concentrations of ethanol. After drying and spraying with gold, the cell morphology on the surface of samples was observed by SEM.

3. Results and Discussion

3.1. Coating Characteristics

Figure 1a,b present the surface and cross-sectional microstructure of the 400 nm-thick titanium oxide coating deposited on the MgZn substrate after 2 h of magnetron sputtering. The obtained coating was flat macroscopically but slightly rough on the microscopic level. There is a clear boundary visible in the SEM image between the substrate and the coating. This is determined by the sputtering process. Under the action of the electric field, the electrons collide with argon atoms in the process of flying to the substrate, which makes them ionize to produce Ar+ ions and new electrons. The new electrons fly to the substrate, and Ar+ ions accelerate to fly to the cathode TiO2 target and bombard the target surface with high energy to make the target sputtering. The sputtered neutral target molecules, i.e., TiO2, deposited and gathered more and more on the substrate to form a thin coating in the form of nanoparticles. According to the results of AFM measurement, as shown in Figure 2c, the particle size was ~100 nm and the coating showed a smooth surface with a roughness (Ra) of 51 nm. The micro-characteristic of the coating caused a change in the surface water contact angle (CA), from 8.4° of the polished MgZn substrate to 79.6° after magnetron sputtering, indicating a decreased hydrophilicity, as shown in Figure 2a,b.
In Figure 1c, Ti and O were detected in the coating by EDS, which confirmed the existence of the Ti-O compound. Meanwhile, the phase composition of the coating and the chemical state of Ti were investigated by XRD and XPS (see Figure 3). The result of XPS shows that all Ti in the coating exists in the form of TiO2, because the peak positions of Ti 2p3/2 and Ti 2p1/2 are 458.4 and 464.1 eV, respectively, which are consistent with the electronic binding peak position of Ti4+ in TiO2 [17]. Figure 3a shows the XRD pattern of the TiO2 coating deposited on the Mg alloy substrate. Clear peaks are found in the XRD pattern, which are all characteristics of the Mg alloy substrate. In addition, there are no other diffraction peaks corresponding to the titanium dioxide. A noisy pattern proves the amorphous structure of the coating. The result shows that the structure of the TiO2 coating deposited by magnetron sputtering is amorphous at a low temperature [18].

3.2. Corrosion Behavior

The corrosion behavior was studied by the immersion test and electrochemical technique. The surface micrographs of the bare and TiO2-coated Mg alloys after 14 d immersion in SBF are shown in Figure 4a,b respectively. The surface of the bare MgZn sample is heavily corroded, with abundant corrosion products deposited. It has been evidenced that during the corrosion process of Mg alloys, insoluble Mg(OH)2 is formed and falls off the surface of the matrix gradually with time, leaving an uneven surface morphology [14]. By contrast, the corrosion extent of the TiO2-coated sample is obviously lesser than that of the substrate. It can be seen that the surface was flat and the corrosion products were smaller and discontinuous. In fact, the better corrosion resistance is mainly conditional on the chemical stability of the TiO2 coating. In Figure 4c, it was observed that the uncorroded area is where the TiO2 coating remains, even if the coating has been broken and incomplete. The results prove that the 400 nm-thick TiO2 coating can provide effective corrosion protection for the Mg alloy. The corrosion products are primarily Mg(OH)2, as shown in Figure 4d.
A more quantitative assessment of the corrosion resistance of the MgZn substrate with and without the TiO2 coating was performed using potentiodynamic polarization curve measurements in SBF, as shown in Figure 4e. From the polarization curves, the corrosion potentials (Ecorr) and corrosion current densities (Icorr) were obtained. As is known, good corrosion resistance and low corrosion rate can be reflected in a higher corrosion potential and a lowered corrosion current density. The Ecorr of the MgZn sample increased from −1.70 to −1.65 V after the TiO2 coating deposition. The coated sample exhibited a more positive corrosion potential than the bare substrate, so the substrate will be corroded more easily than the TiO2-coated sample under the same conditions. In addition, it is found that the Icorr of the TiO2-coated sample markedly decreased compared with that of the bare sample (i.e., 49 vs. 1050 μA/cm2). Hence, the constructed TiO2 coating is a superior protective layer to retard the corrosion rate of the MgZn substrate.

3.3. Biocompatibility

Materials with a poor hemolysis ratio could cause rupture of red blood cells, which may lead to thrombosis and the implantation failure. The hemolysis ratio of biomaterials in contact with blood should be less than 5%. The results in Table 1 show that serious hemolysis (47.23%, hemolysis ratio) occurred in MgZn alloys without surface treatment, while the hemolysis ratio of samples modified by the TiO2 coating was only 0.1%. The anti-hemolytic property of the Mg alloy was significantly improved by TiO2 surface modification. Gao et al. [24] believed that the poor corrosion resistance of the unmodified magnesium alloy increased the pH value of the blood rapidly and promoted the combination of red blood cells and Ca2+ in the solution, finally resulting in the rupture of red blood cells and serious hemolytic reaction. The above results demonstrate that the TiO2 coating can effectively delay the corrosion of substrates and thus improve the hemolysis of Mg alloy.
Platelets play an important role in promoting blood coagulation. The platelet adhesion is an essential reaction corresponding to the formation of thrombus and occlusion of arterial. Figure 5 provides the morphology of the platelets immobilized on samples. Figure 5a shows the adhesion of platelets to MgZn alloy after 1 h of culture. The magnesium alloy surface was rapidly corroded in plasma; hence, obvious cracks can be seen after the corrosion product layer drying. In respect of the adhered platelets characteristic, a large number and serious aggregation of platelets was observed. The adhesion and aggregation of platelets will be further activated to induce thrombosis. In addition, the agglomerated platelets are likely to become fibrinogen adsorption sites, and the adsorbed fibrinogen may also lead to thrombosis. The results indicate that the unmodified Mg alloy is easy to adsorb platelets and has poor blood compatibility.
Figure 5b shows the adhesion of platelets on the surface of the TiO2 coating. The corrosion resistance of the MgZn alloy was enhanced by surface modification, so after soaking in platelet-rich plasma for 1 h, the sample was not corroded. Compared with the case of bare substrate, a remarkable reduced number of adhered platelets could be seen on the surface of TiO2 samples, and no agglomeration occurred. The nearly round shape and well-distributed morphology implied that the fixed platelets were not activated during the culture. The results suggest that the TiO2 coating has a better ability of anti-platelet adhesion.
Accordingly, in consideration of the hemolysis ratio and the adsorbed platelets morphology of different samples, the TiO2-coated MgZn alloy not only possesses excellent properties in cutting down the hemolysis ratio but also has satisfied characteristics of not immobilizing or activating the platelets.
The TiO2-coated magnesium alloys exhibit good cytocompatibility. The viability of endothelial cells reduced in the first 3 days when exposed to the bare magnesium extract. By contrast, the cell viability of the coated group was above 90% at any time point, showing the desirable cytocompatibility, as presented in Table 1. The results could be attributed to the lower values of Mg2+ ion concentration, pH value, and osmolality in the extracts of TiO2-coated samples than those of bare magnesium [25]. Figure 6 shows the morphologies of Ea.hy926 cells cultured on the coated and uncoated groups for 1 day. For the uncoated samples, no cell was observed on the surface. As for the TiO2-coated sample, it can be seen that the Ea.hy926 cells attached and proliferated well on the sample surface, and spread in the shape of a spindle.
The endothelial cell plays a key role in constructing vascular tissue, and cell growth and endothelialization can be controlled by the surface on which it attaches [26]. Generally, a smaller contact angle can correspond to a stronger hydrophilicity of materials, which is more conducive to cell adhesion [27]. The contact angle of the TiO2-coated sample is 79.6°, much higher than that of the uncoated MgZn alloy (8.4°, see Figure 3a,b). However, although the uncoated MgZn alloy has the better hydrophilicity, the cell adhesion on the surface is poor. In this paper, the results indicated that the TiO2-coated sample was stable in the physiological environment and conducive for cell growth and proliferation. Therefore, we believe that the effect of the alkaline environment and hydrogen evolution by the rapid corrosion might worsen the cell adherence and spreading on the uncoated Mg alloy surface.

4. Conclusions

In this work, the titanium oxide coating was prepared on the MgZn alloy surface by magnetron sputtering. The corrosion behavior and biocompatibility have been evaluated. The major conclusions of the present work are as follows.
(1)
A 400 nm-thick titanium oxide coating with a smooth surface was deposited on the MgZn substrate after 2 h magnetron sputtering at room temperature. The coating was composed of dense amorphous TiO2 nanoparticles.
(2)
The corrosion resistance of MgZn alloy was improved apparently by the TiO2 coating. After 14 d of immersion in SBF, the surface of the TiO2-coated sample was less corroded than that of the substrate.
(3)
The uncoated Mg alloys caused serious hemolysis and aggregation of platelets, whereas the TiO2-coated sample had a hemolysis ratio of less than 1% and showed a better ability of anti-platelet adhesion. The TiO2-coated MgZn alloy exhibited lower cytotoxicity and the endothelial cells attached well on the surface, indicating good cytocompatibility.

Author Contributions

Conceptualization, S.H. and L.Y.; methodology, S.H. and W.Y.; formal analysis, Z.Y. and S.L.; writing—original draft preparation, S.H. and Y.L.; writing—review and editing, S.H. All authors have read and agreed to the published version of the manuscript.

Funding

The research was funded by the Henan Center for Outstanding Overseas Scientists (grant no. GZS2018003), Key Scientific Research Projects of Colleges and Universities in Henan Province (grant no. 19B430008), and National Natural Science Foundation of China (grant no. 51801171).

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Chagnon, M.; Guy, L.G.; Jackson, N. Evaluation of magnesium-based medical devices in preclinical studies: Challenges and points to consider. Toxicol. Pathol. 2019, 47, 390–400. [Google Scholar] [CrossRef] [PubMed]
  2. Rahim, M.; Ullah, S.; Mueller, P. Advances and challenges of biodegradable implant materials with a focus on magnesium-alloys and bacterial infections. Metals 2018, 8, 532. [Google Scholar] [CrossRef] [Green Version]
  3. Haude, M.; Ince, H.; Abizaid, A.; Toelg, R.; Lemos, P.A.; von Birgelen, C.; Christiansen, E.H.; Wijns, W.; Neumann, F.; Kaiser, C.; et al. Safety and performance of the second-generation drug-eluting absorbable metal scaffold in patients with de-novo coronary artery lesions (BIOSOLVE-II): 6 month results of a prospective, multicentre, non-randomised, first-in-man trial. Lancet 2016, 387, 31–39. [Google Scholar] [CrossRef]
  4. Ho, M.; Chen, C.; Wang, C.; Chang, S.; Hsieh, M.; Lee, C.; Wu, V.C.; Hsieh, I. The development of coronary artery stents: From bare-metal to bio-resorbable types. Metals 2016, 6, 168. [Google Scholar] [CrossRef] [Green Version]
  5. Soleymani, F.; Emadi, R.; Sadeghzade, S.; Tavangarian, F. Bioactivity behavior evaluation of PCL-chitosan-nanobaghdadite coating on AZ91 magnesium alloy in simulated body fluid. Coatings 2020, 10, 231. [Google Scholar] [CrossRef] [Green Version]
  6. Saadati, A.; Hesarikia, H.; Nourani, M.R.; Taheri, R.A. Electrophoretic deposition of hydroxyapatite coating on biodegradable Mg-4Zn-4Sn-0.6Ca-0.5Mn alloy. Surf. Eng. 2020, 36, 908–918. [Google Scholar] [CrossRef]
  7. Dong, H.; Li, D.; Mao, D.; Bai, N.; Chen, Y.; Li, Q. Enhanced performance of magnesium alloy for drug-eluting vascular scaffold application. Appl. Surf. Sci. 2018, 435, 320–328. [Google Scholar] [CrossRef]
  8. Zhang, D.; Liu, Y.; Liu, Z.; Wang, Q. Advances in antibacterial functionalized coatings on Mg and its alloys for medical use—A review. Coatings 2020, 10, 828. [Google Scholar] [CrossRef]
  9. Elkaiam, L.; Hakimi, O.; Aghion, E. Stress corrosion and corrosion fatigue of biodegradable Mg-Zn-Nd-Y-Zr alloy in in-vitro conditions. Metals 2020, 10, 791. [Google Scholar] [CrossRef]
  10. Song, C.; Yang, Y.; Zhou, Y.; Wang, L.; Zhu, S.; Wang, J.; Zeng, R.; Zheng, Y.; Guan, S. Electrochemical polymerization of dopamine with/without subsequent PLLA coating on Mg-Zn-Y-Nd alloy. Mater. Lett. 2019, 252, 202–206. [Google Scholar] [CrossRef]
  11. Liu, X.; Zhen, Z.; Liu, J.; Xi, T.; Zheng, Y.; Guan, S.; Zheng, Y.; Cheng, Y. Multifunctional MgF2/polydopamine coating on mg alloy for vascular stent application. J. Mater. Sci. Technol. 2015, 31, 733–743. [Google Scholar] [CrossRef]
  12. Liu, J.; Zheng, B.; Wang, P.; Wang, X.; Zhang, B.; Shi, Q.; Xi, T.; Guan, S. Enhanced in vitro and in vivo performance of Mg-Zn-Y-Nd alloy achieved with APTES pretreatment for drug-eluting vascular stent application. ACS Appl. Mater. Interfaces 2016, 8, 17842–17858. [Google Scholar] [CrossRef] [PubMed]
  13. Yuan, T.; Yu, J.; Cao, J.; Gao, F.; Zhu, Y.; Cheng, Y.; Cui, W. Fabrication of a delaying biodegradable magnesium alloy-based esophageal stent via coating elastic polymer. Materials 2016, 9, 384. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  14. Wang, P.; Liu, J.; Shen, S.; Li, Q.; Luo, X.; Xiong, P.; Gao, S.; Yan, J.; Cheng, Y.; Xi, T. In vitro and in vivo studies on two-step alkali-fluoride-treated Mg-Zn-Y-Nd alloy for vascular stent application: Enhancement in corrosion resistance and biocompatibility. ACS Biomater. Sci. Eng. 2019, 5, 3279–3292. [Google Scholar] [CrossRef]
  15. Hou, S.; Mi, L.; Wang, L.; Zhu, S.; Hu, J.; Ding, Q.; Guan, S. Corrosion protection of Mg-Zn-Y-Nd alloy by flower-like nanostructured TiO2 film for vascular stent application. J. Chem. Technol. Biotechnol. 2013, 88, 2062–2066. [Google Scholar]
  16. Chen, S.; Guan, S.; Hou, S.; Wang, L.; Zhu, S.; Wang, J.; Li, W. Characterization and corrosion properties of Ti-O/HA composite coatings on Mg-Zn alloy. Surf. Interface Anal. 2011, 43, 1575–1580. [Google Scholar] [CrossRef]
  17. Zhao, A.; Wang, Z.; Zhou, S.; Xue, G.; Wang, Y.; Ye, C.; Huang, N. Titanium oxide films with vacuum thermal treatment for enhanced hemocompatibility. Surf. Eng. 2014, 31, 898–903. [Google Scholar] [CrossRef]
  18. Ramos-Corella, K.J.; Sotelo-Lerma, M.; Gil-Salido, A.A.; Rubio-Pino, J.L.; Auciello, O.; Quevedo-López, M.A. Controlling crystalline phase of TiO2 thin films to evaluate its biocompatibility. Mater. Technol. 2019, 34, 455–462. [Google Scholar] [CrossRef]
  19. Lin, Z.; Zhao, Y.; Chu, P.K.; Wang, L.; Pan, H.; Zheng, Y.; Wu, S.; Liu, X.; Cheung, K.M.; Wong, T.; et al. A functionalized TiO2/Mg2TiO4 nano-layer on biodegradable magnesium implant enables superior bone-implant integration and bacterial disinfection. Biomaterials 2019, 219, 119372. [Google Scholar] [CrossRef]
  20. Huang, N.; Yang, P.; Cheng, X.; Leng, Y.; Zheng, X.; Cai, G.; Zhen, Z.; Zhang, F.; Chen, Y.; Liu, X.; et al. Blood compatibility of amorphors titanium oxide films synthesized by ion beam enhanced deposition. Biomaterials 1998, 19, 771–776. [Google Scholar]
  21. Hou, S. Solvothermal fabrication of TiO2 nanosheet films on degradable Mg-Zn alloys. Surf. Eng. 2016, 32, 745–749. [Google Scholar] [CrossRef]
  22. Wang, J.; Wang, L.; Guan, S.; Zhu, S.; Ren, C.; Hou, S. Microstructure and corrosion properties of as sub-rapid solidification Mg-Zn-Y-Nd alloy in dynamic simulated body fluid for vascular stent application. J. Mater. Sci. Mater. Med. 2010, 21, 2001–2008. [Google Scholar] [CrossRef]
  23. Kokubo, T.; Takadama, H. How useful is SBF in predicting in vivo bone bioactivity? Biomaterials 2006, 27, 2907–2915. [Google Scholar] [CrossRef] [PubMed]
  24. Gao, F.; Hu, Y.; Li, G.; Liu, S.; Quan, L.; Yang, Z.; Wei, Y.; Pan, C. Layer-by-layer deposition of bioactive layers on magnesium alloy stent materials to improve corrosion resistance and biocompatibility. Bioact. Mater. 2020, 5, 611–623. [Google Scholar] [CrossRef] [PubMed]
  25. Gu, X.N.; Guo, H.M.; Wang, F.; Lu, Y.; Lin, W.T.; Li, J.; Zheng, Y.F.; Fan, Y.B. Degradation, hemolysis, and cytotoxicity of silane coatings on biodegradable magnesium alloy. Mater. Lett. 2017, 193, 266–269. [Google Scholar] [CrossRef]
  26. Guo, X.; Wang, X.; Li, X.; Jiang, Y.C.; Han, S.; Ma, L.; Guo, H.; Wang, Z.; Li, Q. Endothelial cell migration on poly(epsilon-caprolactone) nanofibers coated with a nanohybrid shish-kebab structure mimicking collagen fibrils. Biomacromolecules 2020, 21, 1202–1213. [Google Scholar] [CrossRef]
  27. Yeh, H.I.; Lu, S.K.; Tian, T.Y.; Hong, R.C.; Lee, W.H.; Tsai, C.H. Comparison of endothelial cells grown on different stent materials. J. Biomed. Mater. Res. A 2006, 76, 835–841. [Google Scholar] [CrossRef]
Figure 1. SEM morphologies of (a) surface and (b) cross-sectional microstructures, and (c) the EDS spectra of the titanium oxide-coated MgZn alloy.
Figure 1. SEM morphologies of (a) surface and (b) cross-sectional microstructures, and (c) the EDS spectra of the titanium oxide-coated MgZn alloy.
Coatings 10 00999 g001
Figure 2. The contact angles of the (a) MgZn substrate, (b) TiO2 coating, and (c) the AFM micrograph of the TiO2 coating.
Figure 2. The contact angles of the (a) MgZn substrate, (b) TiO2 coating, and (c) the AFM micrograph of the TiO2 coating.
Coatings 10 00999 g002
Figure 3. (a) XRD pattern of the titanium oxide coating along with the substrate and (b) XPS pattern of high-resolving Ti 2p on the titanium oxide surface.
Figure 3. (a) XRD pattern of the titanium oxide coating along with the substrate and (b) XPS pattern of high-resolving Ti 2p on the titanium oxide surface.
Coatings 10 00999 g003
Figure 4. SEM images of (a) bare and (b,c) TiO2-coated MgZn samples after 14 d immersion in simulated body fluids (SBFs); (d) XRD pattern of corrosion products and (e) potentiodynamic polarization curves in SBF of samples (tested after the coating preparation).
Figure 4. SEM images of (a) bare and (b,c) TiO2-coated MgZn samples after 14 d immersion in simulated body fluids (SBFs); (d) XRD pattern of corrosion products and (e) potentiodynamic polarization curves in SBF of samples (tested after the coating preparation).
Coatings 10 00999 g004
Figure 5. Typical morphology of adherent platelets on surface of (a) MgZn substrate and (b) TiO2 coating.
Figure 5. Typical morphology of adherent platelets on surface of (a) MgZn substrate and (b) TiO2 coating.
Coatings 10 00999 g005
Figure 6. SEM images of Ea.hy926 cells after culturing for 1 d on (a) MgZn substrate and (b) TiO2 coating.
Figure 6. SEM images of Ea.hy926 cells after culturing for 1 d on (a) MgZn substrate and (b) TiO2 coating.
Coatings 10 00999 g006
Table 1. List of the results of hemolysis and cytocompatibility tests.
Table 1. List of the results of hemolysis and cytocompatibility tests.
SamplesHemolysis Ratio (%)Cell Viability (%)
1 d3 d
Control100100
Bare MgZn substrate47.2394.189.5
MgZn substrate with TiO2 coating0.1093.494.8
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Hou, S.; Yu, W.; Yang, Z.; Li, Y.; Yang, L.; Lang, S. Properties of Titanium Oxide Coating on MgZn Alloy by Magnetron Sputtering for Stent Application. Coatings 2020, 10, 999. https://doi.org/10.3390/coatings10100999

AMA Style